K Number
K133012
Device Name
SENTIERO
Manufacturer
Date Cleared
2014-06-09

(257 days)

Product Code
Regulation Number
874.1050
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Sentiero is a portable instrument to diagnose all ages or hearing loss. The instrument offers different test methods which can be configured to fit the professional's needs for screening or diagnostic purposes. It offers physiological test methods such as:

· Distortion Product Otoacoustic Emissions (DPOAE)

· Transient Evoked Otoacoustic Emissions (TEOAE)

· Auditory Brainstem Response (ABR)

· Auditory Steady State Response (ASSR)

· Auditory Impedance and acoustic reflex (TYMP)
Additionally if offers standard audiometry (psycho-acoustical). All physiological test methods are especially indicated for use in defining the type and configuration of hearing loss particularly for individuals whose behavioral audiometric results are deemed unreliable or to assist in the diagnosis of otologic disorders. Estimation of cochlear hearing thresholds (DPTHRESH) is possible at various frequencies without the need of cooperative interaction with the patient. Acoustic reflex and tympanometry (TYMP) are featured to evaluate the function of the middle and outer ear. For each method, several protocols can be configured. The results are to be used to make further recommendations regarding appropriate intervention strategies. Therefore, Sentiero is intended for use by trained personnel such as audiologists. pediatricians. ENT doctors and other health care professionals in a medical or home environment. In the United States of America. Federal law restricts this device to sale by or on the order of a licensed physician.

Available psycho-acoustical methods on Sentiero are especially indicated for use with cooperative patients starting at the age of 2 years or adequate development age, which enables them to do play/interactive audiometry. All other modules are suitable to be used for all ages elder than.infants from 34 weeks (gestational age) that are ready for discharge from the hospital.

Sentiero is designed for:

  1. Diagnostics, monitoring and follow-up after newborn hearing screening
  2. Pre-school, school, and adult hearing screening
  3. ENT diagnostics based on measurement of
    a) Otoacoustic emissions
    b) Tympanometry and acoustic reflex
    c) Auditory Brainstem Responses
    d) Auditory Steady State Responses

Sentiero must not be used in cases of external otitis (outer ear canal infection) or in any case which yields to pain when inserting the ear probe or applying any other transducer.

Device Description

Sentiero is an audiometric examination platform which consists of the Sentiero device with a touch screen display together with different accessories such as mains adapter, OAE probes, headphones, bone conductor, electrode cable, ear coupler cable, patient response switch. All connectors and transducers have a special plug in order to ensure the correct connection to the device. All plugs of the transducers have a memory chip inside which stores the information about the respective transducer (including type of connector, calibration table). As a result, the Sentiero instrument can be connected flexibly to different kind of transducers while enabling the instrument to 'know' the features of the connected transducer. This information is used within the different modules (test methods which are configured) to guide the user (feedback via display) and help to ensure correct performance of the test methods.

Sentiero is available in two different models: handheld or desktop version. The desktop version is labeled Sentiero (Type Desktop). Both versions are standalone examination platforms and can be connected to a personal computer (PC) via USB for data review and management. The handheld version is portable and is meant to be mainly used as mobile device. The desktop version is portable as well but is meant to be mainly used as stationary device. Both models base on a common hardware platform (printed circuit board, PCB) but with different configurations. Materials in contact with humans are selected to be biocompatible.

Furthermore, each model can be configured to allow different test methods and features (modules) by a license key in the device. Sentiero is based on configurable modules. Sentiero can have one single module or a combination of multiple of modules described in the following intended use.

The measurement application is controlled from a self-contained firmware (software installed on the instrument). The measurement flow is menu guided on a touch screen. Evaluation of test results is based on signal statistics (if available for the test method). Besides that wave forms and result information is displayed for the user's evaluation.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study information for the Sentiero device, based on the provided 510(k) summary:

Acceptance Criteria and Device Performance for Sentiero

The provided document describes the Sentiero device and its comparison to predicate devices, primarily focusing on its substantial equivalence through adherence to standards and qualitative clinical evaluations. There are no explicit, quantifiable acceptance criteria tables with specific thresholds and reported performance metrics as one might typically find for an AI/ML device.

Instead, the acceptance criteria are implicitly defined by compliance with a comprehensive set of national and international standards related to audiological devices, electrical safety, EMC, software lifecycle, and biocompatibility, as well as qualitative clinical equivalence to predicate devices demonstrated through user feedback and expert evaluation.

1. Table of Acceptance Criteria and Reported Device Performance

Given the lack of explicit performance metrics for comparison against acceptance criteria in the provided text for the Sentiero device itself, the table below will summarize the types of criteria and the manner in which performance was assessed and deemed acceptable, primarily through equivalence to predicate devices and compliance with standards.

Acceptance Criterion TypeSpecific Criteria/Standard (as implied)Reported Device Performance/Evidence
SafetyIEC 60601-1 (various editions: 1988, 2005)Sentiero successfully passed tests for these standards, demonstrating safety.
Electromagnetic Compatibility (EMC)IEC 60601-1-2:2007Sentiero successfully passed tests for this standard, demonstrating EMC.
Programmable Electrical Systems (Software)IEC 60601-1-4:1996, IEC 62304:2006 (Software lifecycle)Sentiero successfully passed tests for these standards, confirming proper functioning of programmable systems and adherence to software lifecycle requirements. PATH medical GmbH is ISO 9001:2008 and ISO 13485:2003 certified, indicating a robust quality management system for medical devices, including software.
BiocompatibilityISO 10993-1:2009Sentiero's materials in contact with humans are selected to be biocompatible, and the device passed tests for this standard.
Audiometer StandardsIEC 60645-1:2001 (Pure-tone audiometer), ISO 389 series (reference levels for various transducers)Sentiero successfully passed tests for these standards.
Otoacoustic Emissions (OAE)IEC 60645-6:2009Sentiero successfully passed tests for this standard. Performance for TEOAE, DPOAE, DPTHRES against predicate devices (Cochlea-Scan, AccuScreen) was validated in-house using a simulator, and in external clinical evaluations. Considered "similar safe, effective and performant."
Evoked Potentials (ABR, ASSR)IEC 60601-2-40:1998 (Safety of electromyographs and devices for evoked potentials), IEC 60645-7:2009 (Acoustically evoked potentials)Sentiero successfully passed tests for these standards. A clinical evaluation in 2011 compared Sentiero and Chartr EP 200 (predicate ABR device) on the same test persons. Morphology and latencies of wave V were analyzed. "Both systems, Sentiero and Chartr EP 200, were described as reliable clinical ABR devices, which supports the substantial equivalence between both instruments with respect to clinical performance." External tests involved comparison with Chartr EP 200 and Eclipse.
Impedance Audiometry (TYMP)IEC 60645-5/ANSI S3.39, Type 1 (impedance audiometry), IEC 60645-1/ANSI S3.6, IEC 60645-3 (test signals)Sentiero's TYMP module is conform to IEC 60645-5 / ANSI S3.39 Type 1. Probe tone frequency and level are identical to the predicate (Titan IMP440). Pressure safety limits (-700...+500 daPa) are within the required range (-800...+600 daPa), considered "Equivalent" with increased safety due to a smaller range. Transducers and reflex stimuli are "Same" or "Equivalent", complying with standards. "Overall Conclusion" explicitly states similar safety, effectiveness, and performance to predicate devices.
Clinical Performance/UsabilityEquivalence to predicate devices (e.g., Cochlea-Scan, AccuScreen, Chartr EP 200, Titan IMP440) in clinical application/usage, usability for trained personnel.Clinical tests "showed that Sentiero can be used by professionals as efficiently as the respective predicate devices." External users validated measurement results and device usage, comparing against predicate devices with normal hearing patients and those with hearing loss.

2. Sample Size Used for the Test Set and Data Provenance

  • Test Set Sample Size: For the clinical evaluations, a "minimum number of participants per evaluation was 20." This applies to the external user validation studies for various modules (PTA, MAGIC, MATCH, DPTHRES, ABR, TEOAE, DPQUICK, ASSR).
    • For the ABR clinical evaluation comparing Sentiero and Chartr EP 200, the sample size is not explicitly stated but implies "the same test persons."
    • For the in-house validation using a simulator device, the "test protocol is attached in the Appendix" but not provided.
  • Data Provenance: The external clinical evaluations involved various institutions, predominantly in Europe (Germany, Italy, Switzerland).
    • Center of Communication Disorders, ENT Clinic, University Medical Center of the Johannes-Guttenberg-University Mainz, Mainz, Germany
    • Department of Audiology and Speech Therapy, University of Ferrara, Ferrara, Italy
    • ENT Clinic, Klinikum rechts der Isar, Technische Universität München, Munich, Germany
    • ENT Clinic, Kantonsspital Liestal, Liestal, Switzerland
    • Department of Phoniatry and Pedaudiology, ENT Clinic, St. Elisabeth-Hospital, Bochum, Germany
  • Retrospective or Prospective: The text implies prospective clinical evaluations where external users "were asked to evaluate the performance" and "Comparison measurements had to be done." The "first clinical evaluation in 2011" for ABR also suggests a prospective study explicitly comparing the two devices.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

  • Number of Experts: For the ABR clinical evaluation (Sentiero vs. Chartr EP 200), "All results have been evaluated by an expert in audiology." While only one explicit expert is mentioned for this specific evaluation, the list of external testers includes multiple individuals with titles like "Prof. Dr. med." and "Ph.D.", suggesting multiple expert evaluators across the various module assessments.
  • Qualifications of Experts:
    • "expert in audiology" (for ABR evaluation)
    • Prof. Dr. med. Annerose Keilmann
    • Ph.D. Stavrous Hatzopoulos (Department of Audiology and Speech Therapy)
    • Prof. Dr. med. Hans Peter Niedermeyer (ENT Clinic)
    • Dr. Thomas Rosner (ENT Clinic)
    • PD Dr. med. Nicolas Schmuziger (ENT Clinic)
    • Prof. Dr. Katrin Neumann (Department of Phoniatry and Pedaudiology. ENT Clinic)
    • These qualifications (Prof. Dr. med., Ph.D., associated with university medical centers and ENT clinics) indicate highly qualified professionals in audiology, ENT, and related fields.

4. Adjudication Method for the Test Set

The document does not explicitly describe an adjudication method like 2+1 or 3+1 for resolving discrepancies among multiple readers or interpretations of results. The clinical evaluations suggest expert review and comparison to predicate devices, where the "results have been evaluated by an expert" or "out of these studies reports were generated or published, which indicated that Sentiero is an equivalence to the given predicate devices." This implies a form of expert consensus or direct comparison, rather than a formal adjudication process for disagreements.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No formal MRMC comparative effectiveness study, as typically understood in the context of AI performance improvement, is described. The clinical evaluations involved "multiple users" to verify usability and comparison measurements with predicate devices. However, the goal was equivalence in clinical application and usage, not quantifying an "effect size of how much human readers improve with AI vs without AI assistance" because the Sentiero device is not an AI assistant for human readers; it's a diagnostic instrument. The "AI" (advanced algorithms/signal processing) is embedded within the device's diagnostic output, rather than providing assistance to a human making a read.

6. Standalone Performance Study (Algorithm Only)

The Sentiero device is presented as a "standalone examination platform" for measurements. The document indicates that "The measurement application is controlled from a self-contained firmware (software installed on the instrument)." The clinical evaluations and in-house validations assessed the device as a whole, including its algorithms.

  • "Performance tests were also conducted on a test bench. For this reason a simulator device, which simulates TEOAE, DPOAE or ABR output was used for proof of validity of the derived signal." This simulator was used to validate the performance of the test modules, which implies assessing the algorithms' ability to correctly process simulated signals and produce valid results, thus serving as a form of standalone testing for the core algorithmic functionality.
  • The "signal processing information is used in Sentiero only as an indicator (traffic light or indicator for valid response)." This suggests the internal algorithms generate definitive outputs (pass/refer type indicators, waveform analysis) that are then interpreted by the user.

7. Type of Ground Truth Used

The ground truth used for the evaluations appears to be a combination:

  • Expert Consensus/Clinical Judgment: The ABR evaluation states an "expert in audiology" evaluated results. The external testers, who are medical professionals and researchers, made judgments on equivalence based on clinical application/usage.
  • Comparison to Predicate Devices: Clinical evaluations explicitly involved comparison measurements with "predicate devices" on "normal hearing patients with hearing loss." The predicate devices represent the established standard of care and thus serve as a form of ground truth or reference for demonstrating equivalence.
  • Simulator Data: For in-house validation, a "simulator device" delivered "noise or dedicated normative answers derived from literature." Here, the "normative answers" serve as a controlled, known ground truth for validating the device's signal processing.
  • Established Algorithms/Literature: "all implemented modules on Sentiero are derived from well-known algorithms, which are described in detail in published literature... The design process of Sentiero included in-house verification tests also with respect to the mentioned literature and and establishment of normative data." This means the foundational correctness of the algorithms is rooted in established audiology science.

8. Sample Size for the Training Set

The document does not mention a "training set" in the context of modern machine learning that would require a specific sample size. The Sentiero device's modules are described as being based on "well-known algorithms" and "normative data," rather than being "trained" on a dataset in the AI/ML sense. The development process involved "in-house verification tests" and "establishment of normative data," which implies calibration and validation against known physiological responses, but not a distinct training phase with a dedicated dataset as might be found in deep learning.

9. How Ground Truth for the Training Set Was Established

As noted above, there is no explicit "training set" described for the Sentiero device in the AI/ML sense. The algorithms are based on established audiological science. The "normative data" and "well-known algorithms" from literature effectively serve as the "ground truth" for the device's inherent design and calibration. This would have been established through years of audiology research and clinical practice, rather than a specific set of labeled data for this device.

{0}------------------------------------------------

510(k) Summary

SUBMISSION INFORMATION

Date of preparation:May 4th, 2014
510(k) Submitter:PATH medical GmbHLandsberger Str. 6382110 GermeringGermanyPhone: ++49-89-80076502Fax: ++49-89-80076503
Contact Person:Dr. Johann Oswald, DirectorLandsberger Str. 6382110 GermeringGermanyPhone: ++49-89-80076502Fax: ++49-89-80076503Email: oswald@pathme.de

DEVICE INFORMATION

Device Name:Sentiero
Device Trade Names:Sentiero Basic, Sentiero Advanced, Sentiero Desktop
Device Identification Codes:SOH100098, SOH100360, SOD100497
Common Name:Evoked Response Auditory Stimulator
Classification Name:Evoked Response Auditory Stimulator, Audiometer, AuditoryImpedance TesterFDA 21 CFR section 882.1900 and 874.1050 and 874.1090

PREDICATE DEVICES

Cochlea-Scan510(k) number: K061744
AccuScreen510(k) number: K122067
Chartr EP 200510(k) number: K092373
TITAN IMP440510(k) number: K083861
Eclipse510(k) number: K070696

DEVICE DESCRIPTION

Sentiero is an audiometric examination platform which consists of the Sentiero device with a touch screen display together with different accessories such as mains adapter, OAE probes, headphones, bone conductor, electrode cable, ear coupler cable, patient response switch. All connectors and transducers have

a special plug in order to ensure the correct connection to the device. All plugs of the transducers have a memory chip inside which stores the information about the respective transducer (including type of

{1}------------------------------------------------

connector, calibration table). As a result, the Sentiero instrument can be connected flexibly to different kind of transducers while enabling the instrument to 'know' the features of the connected transducer. This information is used within the different modules (test methods which are configured) to guide the user (feedback via display) and help to ensure correct performance of the test methods.

Sentiero is available in two different models: handheld or desktop version. The desktop version is labeled Sentiero (Type Desktop). Both versions are standalone examination platforms and can be connected to a personal computer (PC) via USB for data review and management. The handheld version is portable and is meant to be mainly used as mobile device. The desktop version is portable as well but is meant to be mainly used as stationary device. Both models base on a common hardware platform (printed circuit board, PCB) but with different configurations. Materials in contact with humans are selected to be biocompatible.

Furthermore, each model can be configured to allow different test methods and features (modules) by a license key in the device. Sentiero is based on configurable modules. Sentiero can have one single module or a combination of multiple of modules described in the following intended use.

The measurement application is controlled from a self-contained firmware (software installed on the instrument). The measurement flow is menu guided on a touch screen. Evaluation of test results is based on signal statistics (if available for the test method). Besides that wave forms and result information is displayed for the user's evaluation.

See also "Indications for Use".

The following accessories are available to conduct the different measurement modules:

  • · TEOAE probe: PATH EP-TE
    • · Clearance by FDA following the submission of a 510(k) K100661 and update on K131141
  • · DPOAE probe: PATH EP-DP
    • · Clearance by FDA following the submission of a 510(k) K100661 and update on K131141
    • In combination with a silicon tube attached to Sentiero SOD and Earprobe EP-EP. the probe . can be used as stimulator for the TYMP module.
  • Electrode cable: PATH .
    • · shielded, passive cable to connect the instrument to electrodes for ABR or ASSR
  • Headphones: Sennheiser HDA 280, Sennheiser HDA 200, Interacoustics DD 45, Holmco PD 81, GN . Otometrics otoInsert. PATH Ear Coupler Cable
    • · Only certified standard audiometry headphones are used in conjunction with Sentiero.
  • · Bone conductor: RadioEar B71
    • · As stated by RadioEar (FDA establishment registration number 2516347): "The Bone Conductor does not come under the FDA Requiations for Medical Devices nor the FDA 510(k) Pre-market Approval provisions."
  • · Patient response switch
    • Passive switch for the patient to feedback his response to pure tone audiometry (heard or not heard equals pressed or not pressed as indicated in standard ISO 8253-1)

These accessories can be connected to Sentiero based on a special plug, which holds the information about the connected transducer / cable. Therefore the firmware can make use of this information and adapt the measurement procedures accordingly or provide information to the user via its display,

INTENDED USE

Sentiero is a portable instrument to diagnose all ages or hearing loss. The instrument offers different test methods which can be configured to fit the professional's needs for screening or diagnostic purposes. It offers physiological test methods such as:

  • . Distortion Product Otoacoustic Emissions (DPOAE)
  • . Transient Evoked Otoacoustic Emissions (TEOAE)
  • · Auditory Brainstem Response (ABR)

{2}------------------------------------------------

  • · Auditory Steady State Response (ASSR)
  • · Auditory Impedance and acoustic reflex (TYMP)

Additionally if offers standard audiometry (psycho-acoustical).

All physiological test methods are especially indicated for use in defining the type and configuration of hearing loss particularly for individuals whose behavioral audiometric results are deemed unreliable or to assist in the diagnosis of otologic disorders. Estimation of cochlear hearing thresholds (DPTHRESH) is possible at various frequencies without the need of cooperative interaction with the patient. Acoustic reflex and tympanometry (TYMP) are featured to evaluate the function of the middle and outer ear. For each method, several protocols can be configured. The results are to be used to make further recommendations regarding appropriate intervention strategies. Therefore, Sentiero is intended for use by trained personnel such as audiologists. pediatricians. ENT doctors and other health care professionals in a medical or home environment. In the United States of America. Federal law restricts this device to sale by or on the order of a licensed physician.

Available psycho-acoustical methods on Sentiero are especially indicated for use with cooperative patients starting at the age of 2 years or adequate development age, which enables them to do play/interactive audiometry. All other modules are suitable to be used for all ages elder than.infants from 34 weeks (gestational age) that are ready for discharge from the hospital.

Sentiero is designed for:

    1. Diagnostics, monitoring and follow-up after newborn hearing screening
    1. Pre-school, school, and adult hearing screening
    1. ENT diagnostics based on measurement of
  • a) Otoacoustic emissions
  • b) Tympanometry and acoustic reflex
  • c) Auditory Brainstem Responses
    • d) Auditory Steady State Responses

Sentiero must not be used in cases of external otitis (outer ear canal infection) or in any case which vields to pain when inserting the ear probe or applying any other transducer.

COMPARISON TO PREDICATE DEVICES

In terms of technology, workflow, and accessories, the predicate device AccuScreen shares most of the characteristics of Sentiero. In contrast to AccuScreen, which is a screening device, Sentiero provides diagnostic features based on sophisticated algorithms.

With respect to the intended use, combining psycho-acoustical and physiological test modules in a single handheld device, the Cochlea-Scan is similar.

With respect to diagnostic features on evoked potentials, Chartr EP 200 (ABR and ASSR) and Eclipse (OAE, ABR, and ASSR) can be seen as predicate devices.

With respect to the auditory impedance and acoustic reflex (TYMP), the Titan IMP440 can be seen as predicate device.

In order to facilitate the comparison, first the device setup and technical features and then the modules' features are compared.

  1. Device setup and technical features (including accessories)

Design and working principles of Sentiero are very similar to Cochlea-Scan as well as AccuScreen. The similarity can be attributed to the fact that the same engineers at Fischer-Zoth Diagnosesvsteme GmbH (Germany), who designed the Cochlea-Scan at that time, also designed the Sentiero as well as the AccuScreen at PATH medical GmbH. Accessories like probe and cables were designed and are used for both AccuScreen and Sentiero identically. These two systems were also designed in the same period of time using the same technology standards. In contrast thereto, Cochlea-Scan uses older technology standards. Therefore, the comparison to the predicate device with respect to technological characteristics is focused on the comparison between AccuScreen and Sentiero. Non-clinical performance data are also considered substantially equivalent: the development process was conducted by the same engineers and the resulting devices were tested to be compliant to the same standards by independent laboratories.

{3}------------------------------------------------

Both systems, Sentiero and AccuScreen, are battery driven handheld devices with the user interface on a 3.5" touch screen (Sentiero type MOD 100497: 5.0"). Probes and electrode cables are connected to the devices with similar plugs. Probes, electrode cables, ear coupler cable to the PC (USB) are identical.

Sentiero and AccuScreen provide the same measurement methods (DPOAE, TEOAE and ABR) in order to provide statistically sound results. In contrary to the AccuScreen, which provides a pass/refer result as an automated screening result, the signal processing information is used in Sentiero only as an indicator (traffic light or indicator for valid response). In addition, wave forms (ABR) or frequency domain information (TEOAE and DPOAE) are provided together with additional diagnosic information and environmental information of the measurement to inform the user about the quality of the measurement.

The primary mechanical difference between AccuScreen and Sentiero is the housing which contains the printed circuit board (PCB) and the charging of the built-in rechargeable batteries. AccuScreen can be connected to a docking station for charging and data transfer. In contrast thereto, Sentiero must be connected to a custom made plug for connecting to the charger is possible via a USB cable connected directly to the device. Both housings are made of biocompatible material to fulfill the standards accordingly.

Both devices, Sentiero and AccuScreen, show similar performance data and compliance to the following standards:

  • · IEC 60601-1:2005
  • · IEC 60601-1-2:2007
  • · IEC 60601-1-4:1996
  • · IEC 62304:2006
  • · ISO 10993-1:2009

Safety Electromagnetic compatibility Programmable electrical systems Software lifecycle Biocompatibility

In conclusion, Sentiero is substantially equivalent to the predicate device AccuScreen with respect to technological characteristics and non-clinical performance data.

2) Device measurement features

From a measurement perspective, the methods of predicting a behavioral threshold by means of DPOAE are similar to the predicate device Cochlear-Scan. Also the feature to combine psycho-acoustical test methods like pure tone audiometry and TEOAE or DPOAE on a diagnostic level is similar between both devices. Again, this is supported by the fact, that the same engineers developed both devices. In comparison to Cochlea-Scan, the most significant difference is the technological improvements added to the design of Sentiero. Nevertheless, substantial equivalence is given with respect to the diagnostic features of OAE measurements and hearing threshold predictions.

Additional diagnostic features and measurement techniques were added to Sentiero in comparison to AccuScreen or Cochlea-Scan, which are similar to Chartr EP 200 and Eclipse. Namely the ABR and ASSR features of Sentiero are similar to these predicate devices.

Sentiero uses up to two channels of data recording to run several diagnostic protocols and modalities. The firmware can be configured to run ABR with a traditional click stimulus but also with a tone burst stimulus to predict frequency specific behavioral hearing thresholds. This is similar to the predicate devices Chartr EP 200 and Eclipse. Besides that, Eclipse also provides similar run-time compensated stimuli - similar as Sentiero does (time-delay compensation of the traveling wave on the basilar membrane).

Sentiero can conduct ASSR with a 40 or 80 Hz repetition rate. Several stimuli can be combined together simultaneously and also on both ears to assess responses to various frequencies at the same lime. The evoked response spectrum can yield the information about the presence of the response to stimuli so that ASSR thresholds are supposed to hold a predictable relationship to behavioral thresholds that is dependent on stimulus frequency and the presence or degree of hearing loss. This is similar to the predicate devices Chartr EP 200 and Eclipse.

Besides the difference of dimensions, another difference is that Sentiero can do ABR or ASSR measurements as a standalone system and is therefore not dependent on a PC during the measurement. Chartr EP 200 and Eclipse are all PC-controlled devices.

In a first clinical evaluation in 2011, Sentiero and the predicate device Chartr EP 200 have been tested on the same test persons by different users (trained technicians in audiology). Morphology and latencies of

{4}------------------------------------------------

wave V have been analyzed. Furthermore, the usage of the device was evaluated by the feedback of the users. All results have been evaluated by an expert in audiology. Both systems, Sentiero and Chartr EP 200, were described as reliable clinical ABR devices, which supports the substantial equivalence between both instruments with respect to clinical performance. The run time compensated stimuli and algorithms are similar between Sentiero and Eclipse as they were taken out of the descriptions in published literature.

Sentiero also complies to the relevant standards 60645-1, 60645-7 which displays that TEOAE, DPOAE and evoked potentials can be recorded as safe and as effective as the predicate devices.

Sentiero (Type MOD 100497) can also be configured to allow measurements of auditory impedance and acoustic reflex, which are also present on the predicate device TITAN IMP 440. The other types of Sentiero cannot be configured to use the TYMP module. The stimulus is transmitted via the EarProbe EP-DP in conjunction with a silicon tube in parallel to the probe to deliver the air pressure, which is produced by a pump inside the housing of Sentiero. The TYMP is conform to a IEC 60645-5 / ANSI S3.39 Type 1 acoustic impedance instrument.

In order to facilitate the comparison, the device setup and technical features are compared in the following table to the respective predicate device.

Parameter forComparisonCochlea-ScanAccuScreenChartr EP 200Eclipse
Intended UseSame:Hearingscreening anddiagnostics bymeans of OAEand pure toneaudiometry.Same:Hearing screeningwith statisticalevaluation ofresults by meansof OAE and ABR.Different:Limited purpose(screening) andage group(newborns).Same:Hearingdiagnostics withABR and ASSR.Different:No screening, noOAE tests forspecificexamination ofinner ear hearingstatus.Same:Hearingdiagnostics andneurologicalscreening withOAE, ABR, andASSR.
PatientPopulationSame:All agesDifferent:Newborns onlySame:All agesSame:All ages
DeviceHardware SetupSame:Standalone,handheld /portable device,battery-powered.Same:Same touchscreen displaywith sameresolution,connectors tocables andtransducers.Different:Housing, dockingstationDifferent:Connected to PC,not standalone.Different:Connected to PC,not standalone.
Ear ProbeSame:See K131141.Same:Same probe,special plugs
Parameter forComparisonCochlea-ScanAccuScreenChartr EP 200Eclipse
including memory.
Electrode CableSame:Same:Same cable,special plugsincluding memory.
Headphones,InsertEarphonesSame:Additionalheadphones orinserts available- automaticdetectionthrough memoryin plugDifferent:No headphone orinsert earphone tobe connected.
Safety,Characteristics,PerformanceSame:IEC 10993-1,IEC 60601-1,IEC 60601-1-2,IEC 60601-1-4,EN 1041, ISO980 (ISO15223-1), IEC60645-1,Same:IEC 10993-1, IEC60601-1, IEC60601-2, 60601-1-4, IEC 60645-6,IEC 606045-7,IEC 62304, IEC62366 (IEC60601-1-6)Same:IEC 60601-1-2, IEC60601-2-40Different:Noencephalographyapplication ismentioned for thepredicate device.Nevertheless, thepredicate device iscompliant with IEC60601-2-26.Sentiero does notcomply with EN60601-2-26 asintended use is notfor cerebralmonitoring orelectro-encephalography.Same:IEC 60601-1, IEC60601-2, IEC60645-1,Different:IEC 60645-3(only relevant toclick stimuli -predicate to timecompensatedstimuli is notaffected by thisstandard)IEC 60601-2-26(see Chartr EP200)
Workflow,MenuSame:GeneralworkflowDifferent:Lower screenresolution,additional keysto enter dataSame:General workflow,online informationand device controlvia header/footerDifferent:PC-controlled withseparatemeasurementplatform.Different:PC-controlled withseparatemeasurementplatform.
Parameter forComparisonCochlea-ScanAccuScreenChartr EP 200Eclipse
Availablemodules:Same:DPOAE,TEOAE,DPTHRESHOLD, StandardAudiometrySame:TEOAE, DPOAE,ABRDifferent:Sentiero offersdiagnosticfeatures insteadof screening onlySame:ABR, ASSRSame:ABR, ASSR
Interface toComputer,Software onComputerSame:USB/RS232connectionSame:USB.connection,data transfer toPCDifferent: dockingstation.Different:Runs on PCDifferent:Runs on PC

5

{5}------------------------------------------------

{6}------------------------------------------------

Table 6-1: Comparison of device setup and technical features (Sentiero against different predicate devices)

1

.

Feature / ComparisonTitan with IMP440 (Predicate) 510(k) No. K083861SentieroComments
Indications for useThe Titan Impedance System is an electroacoustic test instrument that produces controlled levels of test tones and signals intended for use in conduction diagnostic hearing evaluations and assisting in the diagnosis of possible otologic disorders. It features tympanometry and acoustic reflexes.Sentiero features tympanometry and acoustic reflexes.Same
Target populationThe devices are suitable for all populations including new-born infantsThe devices are suitable for all populations including new-born infantsSame
Intended userThe devices are to be used by trained personnel onlyThe devices are to be used by trained personnel onlySame
Safety standardIEC 60601-1IEC 60601-1Same
EMC standardIEC 60601-1-2IEC 60601-1-2Same
Standards for impedance audiometryIEC 60645-5/ANSI S3.39, Type 1IEC 60645-5/ANSI S3.39, Type 1Same
Test signalIEC 60645-1/ANSI S3.6, IEC 60645-3IEC 60645-1/ANSI S3.6, IEC 60645-3Same
Probe tone frequency (tympanometry)226 Hz. Optional 678 Hz, 800 Hz and 1000 Hz226 Hz. 678 Hz, 800 Hz and 1000 Hz optionally availableSame

{7}------------------------------------------------

Probe tone level(tympanometry)69 dB HL with AGC69 dB HL with AGCSame
Pressure safety limits-750...+550 daPa-700...+500 daPaEquivalentSafety limits need to bewithin the range of -800...+600 daPaaccording to IEC60645-5/ANSI S3.39, asmaller range meansincreased safety
TransducersEar probe fortympanometry, earprobe andheadphone/insert earphone for reflexmeasurementEar probe fortympanometry, earprobe andheadphone/insert earphone for reflexEquivalentDifferent models of earprobe and headphonesare used, all of themgenerating test signalscomplying to IEC60645-5/ANSI S3.39standards
Reflex stimuliPure tones ataudiometricfrequencies, narrowand wide band noisePure tones ataudiometricfrequencies, narrowand wide band noiseSame

TABLE 6-3: COMPARISON OF TEST MODULE TYMP

OVERALL CONCLUSION

The clinical tests showed that Sentiero can be used by professionals as efficiently as the respective predicate devices. The nonclinical tests shows that Sentiero is as similar safe, as effective and as performant as the predicate devices.

Target populationThe devices aresuitable for allpopulations includingnew-born infantsThe devices aresuitable for allpopulations includingnew-born infantsSame
Intended userThe devices are to beused by trainedpersonnel onlyThe devices are to beused by trainedpersonnel onlySame
Safety standardIEC 60601-1IEC 60601-1Same
EMC standardIEC 60601-1-2IEC 60601-1-2Same
Standards forimpedanceaudiometryIEC 60645-5/ANSIS3.39, Type 1IEC 60645-5/ANSIS3.39, Type 1Same
Test signalIEC 60645-1/ANSIS3.6, IEC 60645-3IEC 60645-1/ANSIS3.6, IEC 60645-3Same
Probe tone frequency(tympanometry)226 Hz. Optional 678Hz, 800 Hz and 1000Hz226 Hz. 678 Hz, 800Hz and 1000 Hzoptionally availableSame
Probe tone level(tympanometry)69 dB HL with AGC69 dB HL with AGCSame
Pressure safety limits$-750...+550$ daPa$-700...+500$ daPaEquivalentSafety limits need to bewithin the range of -$-800...+600$ daPaaccording to IEC60645-5/ANSI S3.39, asmaller range meansincreased safety
TransducersEar probe fortympanometry, earprobe andheadphone/insert earEar probe fortympanometry, earprobe andheadphone/insert earEquivalentDifferent models of earprobe and headphonesare used, all of them

{8}------------------------------------------------

phone for reflexmeasurementphone for reflexgenerating test signalscomplying to IEC60645-5/ANSI S3.39standards
Reflex stimuliPure tones ataudiometricfrequencies, narrowand wide band noisePure tones ataudiometricfrequencies, narrowand wide band noiseSame

TABLE 10-3: COMPARISON OF TEST MODULE TYMP

OVERALL CONCLUSION

Sentiero shows similar safety, effectiveness and performance data as the respective predicate devices.

CONCISE SUMMARY OF PERFORMANCE TESTING

Sentiero (all types and accessories) was tested by independent external laboratories with respect to the following standards:

  • · IEC 60601-1:1988
  • · IEC 60601-1:2005
  • · · IEC 60601-1-2:2007
    • · IEC 60601-2-40:1998

Safety Safety Electromagnetic compatibility Safety of electromyographs and devices for evoked potentials

PATH medical GmbH is certified according to Medical Device Directive 93/42 ECC and holds certificates for compliance with respect to the following standards:

  • · ISO 9001:2008 Quality management system
  • · ISO 13485:2003 Quality management system (medical devices)

Basing on this, reqular audits and compliance checks are performed by the following notified bodies:

  • · Notified body for certification in Europe: DEKRA Certification GmbH (Germany), (see certificate in the Appendix).
  • · Notified body for certification in Canada: DEKRA Certification B.V. (Netherlands), (see certificate in the Appendix).

Performance tests were also conducted on a test bench. For this reason a simulator device, which simulates TEOAE, DPOAE or ABR output was used for proof of validity of the derived signal. The same simulator device was distributed by Natus Europe GmbH as a testing device for the predicate devices AccuScreen and Cochlea-Scan.

The device under test (DUT) had to be connected to deliver the stimulus (via insert earphone or ear probe) into a cavity of the simulator. The simulator was connected to the electrode cable of the DUT to record the test result. The stimulator was able to deliver noise or dedicated normative answers derived from literature. The same simulator was used for Sentiero as well as the predicate devices Cochlea-Scan and AccuScreen. The simulator was developed by the same engineers who developed the Cochlea-Scan as well as the AccuScreen. Passing the requirements given by the simulator was part of the acceptance criteria for the verification phase of the device development cycle. Consequently the performance of the test

modules TEOAE, DPOAE, DPTHRES. ABR were validated in comparison to the predicate devices during the in-house validation phase of Sentiero. The test protocol is attached in the Appendix.

Additionally Sentiero was tested against the following standards to show compliance and performance as follows:

  • · ISO 10993-1:2009 Biocompatibility
  • · IEC 60601-1-4:1996 Programmable electrical system
  • · IEC 60601-2-40:1998 Safety of electromyographs and devices for evoked potentials
  • · IEC 62304:2006 Software lifecycle

{9}------------------------------------------------

  • · IEC 60645-1:2001 Pure-tone audiometer
  • · IEC 60645-6:2009 Otoacoustic emissions
  • · IEC 60645-7:2009 Acoustically evoked potentials
  • · ISO 389-1:1998 Reference levels for pure tones and supra aural headphones
  • · ISO 389-2:1994 Reference levels for pure tones and insert earphones
  • · ISO 389-3:1994 Reference force levels for pure tones and bone conductors
  • · ISO 389-4:1994 Reference levels for narrow band masking noise
  • · ISO 389-5:2006 Reference levels for pure tones from 8 to 16 kHz
  • · ISO 389-8:2004 Reference levels for pure tones and circumaural headphones

Sentiero successfully passed the tests and checklists of the aforementioned standards. Thus it is shown that Sentiero is as safe, efficient and performing as the predicate devices which also fulfill the standards as mentioned in the comparison table above.

In preparation of the first application for receiving the CE mark for Sentiero, additional test sites were asked to evaluate the performance of relevant test modules and the handling of the instrument. To state clearly: all implemented modules on Sentiero are derived from well-known algorithms, which are described in detail in published literature. The design process of Sentiero included in-house verification tests also with respect to the mentioned literature and and establishment of normative data. In addition, external potential users were asked to validate the measurement results and the usage of the device and test modules (usability tests and performance tests). Comparison measurements had to be done with predicate devices and normal hearing patients with hearing loss. Minimum number of participants per evaluation was 20. Multiple users should be involved to verify usability of the device in comparison to experience level of the user.

No tests were conducted with respect to animal testing.

The test sites, which participated in the evaluation of test modules on Sentiero are shown in Table 10-4. Please note that not all these testers and studies did receive financial support from PATH medical. Most of the evaluation data and feedback was contributed by initiative from the tester. The tester a financial arrangement with PATH medical - except the two sites mentioned in section 9 of this document, where PATH was the sponsor and initiator of the study.

Sentiero ModulePredicate deviceTester
PTA class 3Cochlea-Scan(DPOAEthresholdestimation)Prof. Dr. med. Annerose Keilmann
MAGIC (image-based pure toneaudiometry)Center of Communication Disorders, ENT Clinic,University Medical Center of the Johannes-Guttenberg-University Mainz, Mainz, Germany
MATCH (image-based speech test)Establishedspeech tests(e.g. Mainzerspeech test forchildren)
DPTHRES (DPOAEthreshold estimation)
ABRChartr EP 200Ph.D. Stavrous Hatzopoulos
Department of Audiology and Speech Therapy,University of Ferrara, Ferrara, Italy
TEOAEAccuScreen(TEOAE, DPOAE)Prof. Dr. med. Hans Peter Niedermeyer
DPQUICK (DPOAEat fixed levels)ENT Clinic, Klinikum rechts der Isar, Technische
DPTHRES (DPOAEthreshold estimation)Universität München, Munich, Germany
ASSRCochlea-Scan(DPOAEthresholdestimation)Dr. Thomas Rosner
ECLIPSEENT Clinic, Klinikum rechts der Isar. Technische

{10}------------------------------------------------

Sentiero ModulePredicate deviceTester
Universität München, Munich, Germany
ABRABR device,unknown type.PD Dr. med. Nicolas Schmuziger
ENT Clinic, Kantonsspital Liestal, Liestal, Switzerland
PTAEchoScreen/AccuScreen andCochleaScanProf. Dr. Katrin Neumann
MAGIC (image-based pure toneaudiometry)Department of Phoniatry and Pedaudiology. ENTClinic, St .- Elisabeth-Hospital, Bochum, Germany
TEOAE
DPQUICK (DPOAEat fixed levels)
DPTHRES (DPOAEthreshold estimation)
ABR

Table 10-4: List of external tests including tested module and the corresponding tester

Out of these studies reports were generated or published, which indicated that Sentiero is an equivalence to the given predicate devices with respect to clinical application/usage.

{11}------------------------------------------------

Image /page/11/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services (HHS). The logo features a stylized depiction of an eagle or bird in flight, composed of three curved lines that suggest movement and dynamism. The bird is positioned above a circular arrangement of text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES USA".

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring. MD 20993-002

June 9, 2014

PATI Medical GmbH c/o Dr. Johann Oswald Director of PATH Medical GmbH Landsberger Str.63 Germering. Bavaria GM D-82110

Rc: K133012

Trade/Device Name: Sentiero Regulation Number: 21 CFR 874.1050 Regulation Name: Audiometer Regulatory Class: Class II Product Code: EWO. GWJ. ETY Dated: May 1. 2014 Received: May 6. 2014

Dear Dr. Oswald:

We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28. 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into cither class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.

{12}------------------------------------------------

Page 2 - Dr. Johann Oswald

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Eric A. Mann -S

for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{13}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

510(k) Number (if known)

Device Name

Sentiero

MICRO

Indications for Use (Describe)

Sentiero is a portable instrument to diages for hearing loss. The instrument offers different test nethods which can be configured to fit the professional's needs for screening or diagnostic purposes. It offers physiological test methods such as:

· Distortion Product Otoacoustic Emissions (DPOAF)

· Transient Evoked Otoacoustic Emissions (TEOAF)

· Auditory Brainstem Response (ABR)

· Auditory Steady State Response (ASSR)

· Auditory Impedance and acousiic relles (TYMP) Additionally if offers standard audiometry (psycho-acousiveal). All physiological test methods are especially indicated for use in defining the type and configuration of hearing loss particularly for individuals whose behavioral audiometric results are deemed unreliable or to assist in the diagnosis of otologic disorders. Estination of cochlear hearing thresholds (DP THRESI) is possible at various frequencies without the need of cooperative interaction with the pairent. Acoussic rellex and tympanomery (TYMP) are featured the functional condition of the middle and outer car. For each method, several protocols can be configured. The results are to be used to make further recommendations regarding appropriate interention strategies. Therefore, Sentiero is intended for use by trained personnel such as audiologiss, pediatricians. ENT doctors and other health care professionals. In the United States of America. Federal law restricts this device to sale by or on the order of a licensed physician.

A vailable psycho-acoustical methods on Sentiero are especially indicated for use with cooperative parting at the age of 2 years or adequate development age, which enables them to do playinters. All other modules are suitable to be used for all ages.

Sentiero is designed for:

  1. Diagnostics, monitoring and follow-up after newborn hearing screening

  2. Pre-school, school. and adult hearing screening

  3. ENT diagnostics based on measurement of

· Otoacoustic emissions

· Tympanometry and acoustic reflex

· Auditory Brainstem Responses

· Auditory Steady State Responses

Sentiero must not be used in cases of external otitis (outer car canal intection) or in any case which yields to pain when inserting the ear probe or applying any other transducer.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpan D)

Over-The-Counter Use (21 CFR 801 Subpart C)

{14}------------------------------------------------

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLYConcurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response. including the time to review instructions. search existing data sources. gather and mainlain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@ida.hhs.gov

*An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

§ 874.1050 Audiometer.

(a)
Identification. An audiometer or automated audiometer is an electroacoustic device that produces controlled levels of test tones and signals intended for use in conducting diagnostic hearing evaluations and assisting in the diagnosis of possible otologic disorders.(b)
Classification. Class II. Except for the otoacoustic emission device, the device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter, if it is in compliance with American National Standard Institute S3.6-1996, “Specification for Audiometers,” and subject to the limitations in § 874.9.