K Number
K132987
Manufacturer
Date Cleared
2014-07-03

(282 days)

Product Code
Regulation Number
886.1120
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The CSO Cobra Fundus Camera is intended for taking digital images of a human retina with or without the use of mydriatic agent. The retinal image can be stored to an image filing device.

Device Description

Cobra is intended for taking digital images of a human retina with or without the use of a mydriatic agent. A function for retinal plane capturing is provided. The instrument is furnished with an integrated 5MP CCD camera, and uses one (1) white LED for flashing, one (1) IR led for alignment and Infrared acquisitioning.

Cobra can be used with pupil diameters starting from 2.5 mm, and therefore can be used without the need for a mydriatic agent. Acquisition transfer is performed from the instrument to the accompanying PC via a Firewire cable.

AI/ML Overview

This document is a 510(k) premarket notification for the Cobra Fundus Camera, seeking clearance from the FDA. The information provided focuses on demonstrating substantial equivalence to a predicate device rather than presenting a detailed independent clinical study with specific acceptance criteria in the typical sense for an AI/CADe device.

Therefore, many of the requested sections (e.g., effect size of human readers with AI assistance, detailed sample sizes for test and training sets with ground truth establishment for AI, number of experts for ground truth, adjudication method) are not applicable or not explicitly stated in this type of submission, as the device itself is a fundus camera, not an AI diagnostic system. The "performance test" section refers to engineering and regulatory compliance tests, not a clinical study to evaluate diagnostic accuracy against a ground truth.

Here's the closest interpretation of the requested information based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria CategorySpecific Acceptance Criteria (Standards Met)Reported Device Performance
Electrical SafetyIEC 60601-1 (International Standard for Medical Electrical Equipment - Part 1: General Requirements for Basic Safety and Essential Performance)Cobra met all requirements.
Electromagnetic Compatibility (EMC)IEC 60601-1-2 (International Standard for Medical Electrical Equipment - Part 1-2: General Requirements for Basic Safety and Essential Performance - Collateral Standard: Electromagnetic Compatibility - Requirements and Tests)Cobra met all requirements.
Ophthalmic Instrument SafetyISO 15004-2 (Ophthalmic instruments — Fundus cameras — Part 2: Requirements for safety) - Specifically "Group 1 instruments"Cobra met all requirements.
Software ValidityFDA Guidance for the content of premarket submissions for software contained in medical devicesEvaluation performed; all functional tests and unit level tests passed, meeting test criteria. Level of concern: Moderate.
BiocompatibilityAssessment performedMaterials used are the same as other legally marketed devices in US.
Risk ManagementISO 14971:2012 (Medical devices - Application of risk management to medical devices)Cobra met all requirements; risk management deemed satisfactory.
Image Transfer HardwareNot explicitly stated as a separate standard, but comparison to predicate and impact on safety/effectiveness.Different from predicate, but not considered critical to intended use or affecting safety/effectiveness.
Substantial EquivalenceEquivalence to Kowa nonmyd WX Ophthalmic Camera (K101628) in functionality, design verification, and intended use, with no effect on safety/effectiveness due to differences.Concluded to be substantially equivalent.

2. Sample size used for the test set and the data provenance

  • Test Set Sample Size: Not applicable in the context of this 510(k) submission. The "performance tests" mentioned are engineering and regulatory compliance studies, not clinical trials with patient data.
  • Data Provenance: Not applicable.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

  • Not applicable. This submission doesn't describe a clinical study requiring ground truth established by experts.

4. Adjudication method for the test set

  • Not applicable.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • No. This device is a fundus camera, not an AI/CADe system. Therefore, an MRMC study comparing human readers with and without AI assistance is not relevant or described.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • No. This is a hardware device (fundus camera), not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

  • Not applicable in the context of a clinical performance evaluation. The "ground truth" for the performance tests would be the fulfillment of established engineering standards (e.g., electrical safety, EMC requirements).

8. The sample size for the training set

  • Not applicable. This device is a fundus camera, not an AI system that requires a "training set."

9. How the ground truth for the training set was established

  • Not applicable.

{0}------------------------------------------------

r

510(k) Summary 5

5.1 General information

General information
ManufacturerC.S.O. SrlVia degli Stagnacci 12/E50018 Badia a Settimo Scandicci (FI)ITALY
Contact personPiet de Jong,+39 055 722 1958p.dejong@csoitalia.it
Date Prepared16 July 2013

Device information 5,2

Device information
Trade nameCobra Fundus Camera
Common NameFundus Camera
Classification NameOphthalmic Camera, AC-Powered
Regulation Number(s)21 CFR 886.1120
21 CFR 892.2050
Product Code(s)HKI, NFJ
PanelOphthalmic

5.3 Predicate devices

The Cobra Fundus Camera is compared for substantial equivalence with Kowa nonmyd WX Ophthalmic Camera, manufactured by Kowa Company Ltd, Tokyo, Japan (K101628).

5.4 Device description

Cobra is intended for taking digital images of a human retina with or without the use of a mydriatic agent. A function for retinal plane capturing is provided. The instrument is furnished with an integrated 5MP CCD camera, and uses one (1) white LED for flashing, one (1) IR led for alignment and Infrared acquisitioning.

Cobra can be used with pupil diameters starting from 2.5 mm, and therefore can be used without the need for a mydriatic agent. Acquisition transfer is performed from the instrument to the accompanying PC via a Firewire cable.

5.5 Intended use

Cobra is an electromedical system that allows for digitally retrieving, acquiring and processing of a retinal image to the extent permitted by the laws and regulations for the exercise of their profession.

The instrument allows for a direct, 'live' monitoring on the computer screen, and capturing also in cases where the pupil diameter starts from 2.5 mm.

{1}------------------------------------------------

5.6 Substantial equivalence

In accordance with the predicate device standards tests have been executed, resulting in the same technological characteristics.

Cobra met all the requirements regarding to electrical safety (tests according to IEC 60601-1)

Cobra met all the requirements to ensure electromagnetic compatibility, according to IEC 60601-1-2.

Cobra met all the requirements of Group 1 instruments according to ISO 15004-2 in order to be in compliance to the standard.

Concerning software validity of the CSO software, an evaluation was made based on FDA guidance, Guidance for the content of premarket submissions for software contained in medical devices, was performed.

The validation of Cobra acquisition software is performed as a part of a system function test. All functions are tested and confirmed working with traceability of remaining anomalies in a tracking system. Two tests were performed, one on functional level and one on system unit level. Both tests have been verified as meeting test criteria. All functional tests have been passed, as well as unit level tests.

The level of concern of Cobra acquisition software is moderate.

To evaluate biocompatibility, biocompatibility assessment was performed. All materials used are the same as other legally marked devices in US.

Cobra was evaluated for risk management in accordance with ISO 14971:2012, and met all requirements of the standard. The risk management of the device was deemed satisfactory.

The hardware interface for image transfer is different from the predicate device, but it cannot be considered critical to the intended therapeutic, diagnostic, prosthetic or surgical use of the device, nor can it affect the safety and effectiveness of the device when used as labeled.

Remaining risks will be noted in the user manual, so users will be able to avoid them.

It is concluded that Cobra is substantially equivalent with regards to functionality, design verification and intended use, to the identified predicate device already in interstate commerce within the USA, and that any differences that do exist have no effect on the safety and effectiveness of the device

5.7 Performance test

Performance testing was conducted on the subject device and was found to perform as intended.

Cobra met all the requirements regarding to electrical safety (tests according to IEC 60601-1) and electromagnetic compatibility, according to IEC 60601-1-2.

Cobra met all the requirements of Group 1 instruments according to ISO 15004-2 in order to be in compliance to the standard.

{2}------------------------------------------------

Concerning software validity of the Cobra acquisition software, an evaluation was made based on FDA guidance, Guidance for the content of premarket submissions for software contained in medical devices, was performed.

To evaluate biocompatibility, biocompatibility assessment was performed. All materials used are the same as other legally marked devices in US.

Cobra was evaluated for risk management in accordance with ISO 14971:2012, and met all requirements of the standard. The risk management of the device was deemed satisfactory.

Remaining risks will be noted in the user manual, so users will be able to avoid them.

In conclusion we can say that Cobra is substantially equivalent to the predicate device since the same requirements to referenced standards regarding safety and effectiveness were met.

{3}------------------------------------------------

Image /page/3/Picture/11 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is a stylized symbol that resembles an abstract human figure with outstretched arms.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

July 3, 2014

C.S.O. Srl. c/o Mr. Claude Berthoin Regulatory Consultant Thema USA 110 East Granada Blvd., Suite 209 Ormond Beach, FL 32716

Re: K132987

Trade/Device Name: Cobra Fundus Camera Regulation Number: 21 CFR 886.1120 Regulation Name: Ophthalmic Camera Regulatory Class: Class II Product Code: HKI, NFJ Dated: May 28, 2014 Received: June 2, 2014

Dear Mr. Berthoin:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2 - Mr. Claude Berthoin

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://yww.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely vours.

Kesia Y. Alexander -S

for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

510(k) Number (if known) K132987

Device Name CSO Cobra Fundus Camera

Indications for Use (Describe)

The CSO Cobra Fundus Camera is intended for taking digital images of a human retina with or without the use of mydriatic agent. The retinal image can be stored to an image filing device.

Type of Use (Select one or both, as applicable)

2 Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

Leonid Livshitz -S 2014.06.27 13:20:03 -04'00'

This section applies only to requirements of the Paperwork Reduction Act of 1995.

"DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

§ 886.1120 Ophthalmic camera.

(a)
Identification. An ophthalmic camera is an AC-powered device intended to take photographs of the eye and the surrounding area.(b)
Classification. Class II (special controls). The device, when it is a photorefractor or a general-use ophthalmic camera, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.