K Number
K132944
Device Name
ADVANTAGESIM MD MR PELVIC ORGAN AT RISK SEGMENTATION
Manufacturer
Date Cleared
2014-03-14

(176 days)

Product Code
Regulation Number
892.5840
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
AdvantageSim™ MD is used to prepare geometric and anatomical data relating to a proposed external beam radiotherapy treatment prior to dosimetry planning. Anatomical volumes can be defined automatically or manually in three dimensions using a set of CT images acquired with the patient in the proposed treatment position. Definition of the anatomical volumes may be assisted by additional CT, MR or PET studies that have been co-registered with the planning CT scan. Additionally, CT & PET data from a respiratory tracked examination may be used to allow the user to define the target or treatment volume over a defined range of the respiratory cycle. The geometric parameters of a proposed treatment field are selected to allow non-dosimetric, interactive optimization of field coverage. Defined anatomical structures and geometric treatments fields are displayed on transverse images, on reformatted sagittal, coronal or oblique images, on 3 D views created from the images, or on a beam eye's view display with or without the display of defined structures with or without the display of digitally reconstructed radiograph.
Device Description
AdvantageSim™ MD is a CT/MR/PET oncology application used by clinicians (radiologist, radiation oncologist, medical oncologist nuclear medicine physicians and trained healthcare professional) to assist treatment planning. AdvantageSim MD with MR pelvic organ at risk segmentation Option is used to provide MR based prostate and pelvic organs-at-risk segmentation. A suite of semi-automated MR based organ segmentation contouring allows generating complex structures around organs at risk. These contours overlay on the co-registered CT planning image. The segmentation methods in the modified device are semi-automatic. The user has to place seed points to identify an inner point of the organ to contour. The software offers a suite of manual contour editing tools enabling the user to edit, modify, or change contours generated from the MR segmentation tools to their desired configuration based on their medical and clinical knowledge and experience. The results provided by the software needs to be approved by the experienced clinician and can always be modified or corrected by him/her. It is up to the expert user to accept the result without any change, reject it completely and delineate manually, or modify the result and then save it. The software does not provide any auto-detection or auto-saving functionalities. Same as the predicate devices, the clinician retains the ultimate responsibility for making the pertinent diagnosis and patient management decisions based on their standard practices and visual comparison of the individual images, regardless of the accuracy of the output generated by the software.
More Information

No
The description explicitly states the segmentation methods are "semi-automatic" and require user input (seed points). It also emphasizes manual editing tools and the clinician's ultimate responsibility, indicating a lack of fully automated, AI/ML-driven processes. The absence of mentions of AI, DNN, or ML, and the lack of details on training/test sets further support this conclusion.

No.
This device is a software application used for treatment planning in radiotherapy by assisting with anatomical volume definition and segmentation; it does not directly apply therapy to a patient.

No

The device is described as an oncology application used to assist treatment planning by preparing geometric and anatomical data for external beam radiotherapy and segmenting anatomical volumes. It does not provide a diagnosis or aid in making clinical diagnostic decisions. The text states that "the clinician retains the ultimate responsibility for making the pertinent diagnosis and patient management decisions."

Yes

The device description explicitly states it is a "CT/MR/PET oncology application" and "software," and the entire summary focuses on the software's functionality for image processing and segmentation for treatment planning. There is no mention of any accompanying hardware component that is part of the medical device itself.

Based on the provided information, this device is NOT an IVD (In Vitro Diagnostic).

Here's why:

  • IVD Definition: In vitro diagnostics are tests performed on samples taken from the human body, such as blood, urine, or tissue, to detect diseases, conditions, or infections. They are used to provide information for diagnosis, monitoring, or screening.
  • AdvantageSim™ MD's Function: AdvantageSim™ MD is a software application used for radiotherapy treatment planning. It processes medical images (CT, MR, PET) to define anatomical volumes and geometric parameters for external beam radiation therapy. It assists clinicians in preparing for treatment, not in diagnosing a condition based on a biological sample.
  • Intended Use: The intended use clearly states it's for "preparing geometric and anatomical data relating to a proposed external beam radiotherapy treatment prior to dosimetry planning." This is a treatment planning function, not a diagnostic one.
  • Device Description: The description reinforces its role in "assist treatment planning" and "provide MR based prostate and pelvic organs-at-risk segmentation" for this purpose.
  • Lack of Biological Sample Analysis: There is no mention of analyzing biological samples or performing tests on bodily fluids or tissues.

In summary, AdvantageSim™ MD is a medical device used in the field of radiation oncology for treatment planning, which is distinct from in vitro diagnostics.

N/A

Intended Use / Indications for Use

AdvantageSim™ MD is used to prepare geometric and anatomical data relating to a proposed external beam radiotherapy treatment prior to dosimetry planning. Anatomical volumes can be defined automatically or manually in three dimensions using a set of CT images acquired with the patient in the proposed treatment position.

Definition of the anatomical volumes may be assisted by additional CT, MR or PET studies that have been co-registered with the planning CT scan. Additionally, CT & PET data from a respiratory tracked examination may be used to allow the user to define the target or treatment volume over a defined range of the respiratory cycle.

The geometric parameters of a proposed treatment field are selected to allow non-dosimetric, interactive optimization of field coverage. Defined anatomical structures and geometric treatments fields are displayed on transverse images, on reformatted sagittal, coronal or oblique images, on 3 D views created from the images, or on a beam eye's view display with or without the display of defined structures with or without the display of digitally reconstructed radiograph.

AdvantageSim™ MD with MR pelvic organ at risk segmentation Option is used to provide MR based prostate and pelvic organs-at-risk segmentation. A suite of semi-automated MR based organ segmentation contouring allows generating complex structures around organs at risk. These contours overlay on the co-registered CT planning image.

Product codes

KPQ

Device Description

AdvantageSim™ MD is a CT/MR/PET oncology application used by clinicians (radiologist, radiation oncologist, medical oncologist nuclear medicine physicians and trained healthcare professional) to assist treatment planning.

AdvantageSim MD with MR pelvic organ at risk segmentation Option is used to provide MR based prostate and pelvic organs-at-risk segmentation. A suite of semi-automated MR based organ segmentation contouring allows generating complex structures around organs at risk. These contours overlay on the co-registered CT planning image.

The segmentation methods in the modified device are semi-automatic. The user has to place seed points to identify an inner point of the organ to contour.

The software offers a suite of manual contour editing tools enabling the user to edit, modify, or change contours generated from the MR segmentation tools to their desired configuration based on their medical and clinical knowledge and experience. The results provided by the software needs to be approved by the experienced clinician and can always be modified or corrected by him/her. It is up to the expert user to accept the result without any change, reject it completely and delineate manually, or modify the result and then save it. The software does not provide any auto-detection or auto-saving functionalities.

Same as the predicate devices, the clinician retains the ultimate responsibility for making the pertinent diagnosis and patient management decisions based on their standard practices and visual comparison of the individual images, regardless of the accuracy of the output generated by the software.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

CT, MR, PET

Anatomical Site

pelvic organs, prostate

Indicated Patient Age Range

Not Found

Intended User / Care Setting

clinicians (radiologist, radiation oncologist, medical oncologist nuclear medicine physicians and trained healthcare professional)

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

A usability study using consented clinical images was conducted by three board certified Radiation Oncologists who were considered experts. The study was meant to assess the following:

  • Accuracy of measurement
  • Precision of the measurement
  • Efficiency, as a comparison of amount of time
  • General user Qualitative feedback

The study results demonstrated that when used by qualified clinician the new software device has the potential to reduce inter-operator variability, provides statistically significant and practically meaningful clinical efficiency improvements and substantiates the characteristics of this feature, among others, as easy to learn, useful, efficient and providing increased throughput.

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

Not Found

Predicate Device(s)

K132045, K111311

Reference Device(s)

Not Found

Predetermined Change Control Plan (PCCP) - All Relevant Information

Not Found

§ 892.5840 Radiation therapy simulation system.

(a)
Identification. A radiation therapy simulation system is a fluoroscopic or radiographic x-ray system intended for use in localizing the volume to be exposed during radiation therapy and confirming the position and size of the therapeutic irradiation field produced. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.(b)
Classification. Class II.

0

Image /page/0/Picture/1 description: The image shows a document page with handwritten text and a logo. The text at the top reads "K132 944 page 1 of 4". Below the text is the GE logo, which consists of the letters "GE" inside a circular design.

GE Healthcare AdvantageSim™ MD with MR pelvic organ at risk segmentation Option 510(k) Premarket Notification Submission

510(k) Summary

In accordance with 21 CFR 807.92 the following summary of information is provided:

Date:September 16, 2013
Submitter:GE Hungary Kft.
Akron utca 2
2040 Budörs, Hungary
Primary Contact
Person:Angelique Cruz
Regulatory Affairs Leader
Tel: 0033 (0)1 30 70 47 55
Fax: (262)364-2506
Secondary Contact
Person:Huy Doan
Regulatory Affairs Director
Tel: 262-312-7751
Mobile: 414-5818553
Fax:: (262)364-2506

Helen Peng
Regulatory Affairs Manager
Tel: (262)5485091
Fax: (262)364-2506 |
| Device Trade Name: | AdvantageSim™ MD MR pelvic organ at risk segmentation |
| Common/Usual Name: | AdvantageSim™ MD MR pelvic organ at risk segmentation |
| Classification Names: | 21CFR 892.5840, Radiology |
| Product Code: | KPQ |
| Predicate Device(s): | K132045 - AdvantageSim™ MD with CT Atlas-based Contouring and
Replanning Options from GE Healthcare
K111311 Segasist Prostate Auto-Contouring Software from Segasist
Technologies |
| Device Description /
Intended Use: | AdvantageSim™ MD is a CT/MR/PET oncology application used by
clinicians (radiologist, radiation oncologist, medical oncologist |
| | |
| nuclear medicine physicians and trained healthcare professional)
to assist treatment planning. | |
| AdvantageSim MD with MR pelvic organ at risk segmentation
Option is used to provide MR based prostate and pelvic organs-at-
risk segmentation. A suite of semi-automated MR based organ
segmentation contouring allows generating complex structures
around organs at risk. These contours overlay on the co-registered
CT planning image. | |
| The segmentation methods in the modified device are semi-
automatic. The user has to place seed points to identify an inner
point of the organ to contour. | |
| The software offers a suite of manual contour editing tools enabling
the user to edit, modify, or change contours generated from the MR
segmentation tools to their desired configuration based on their
medical and clinical knowledge and experience. The results
provided by the software needs to be approved by the experienced
clinician and can always be modified or corrected by him/her. It is
up to the expert user to accept the result without any change,
reject it completely and delineate manually, or modify the result
and then save it. The software does not provide any auto-detection
or auto-saving functionalities. | |
| Same as the predicate devices, the clinician retains the ultimate
responsibility for making the pertinent diagnosis and patient
management decisions based on their standard practices and
visual comparison of the individual images, regardless of the
accuracy of the output generated by the software. | |
| Indications for Use: | AdvantageSim™ MD is used to prepare geometric and anatomical
data relating to a proposed external beam radiotherapy treatment
prior to dosimetry planning. Anatomical volumes can be defined
automatically or manually in three dimensions using a set of CT
images acquired with the patient in the proposed treatment
position. |
| | Definition of the anatomical volumes may be assisted by additional
CT, MR or PET studies that have been co-registered with the
planning CT scan. Additionally, CT & PET data from a respiratory
tracked examination may be used to allow the user to define the
target or treatment volume over a defined range of the respiratory
cycle. |
| | The geometric parameters of a proposed treatment field are
selected to allow non-dosimetric, interactive optimization of field
coverage. Defined anatomical structures and geometric treatments
fields are displayed on transverse images, on reformatted sagittal,
coronal or oblique images, on 3 D views created from the images,
or on a beam eye's view display with or without the display of
defined structures with or without the display of digitally
reconstructed radiograph. |
| Technology: | The AdvantageSim MD with MR pelvic organ at risk segmentation
Option software employs the same fundamental scientific
technology as that of its predicate devices. |
| Determination of
Substantial
Equivalence: | Summary of Non-Clinical Tests:
The AdvantageSim MD MR pelvic organ at risk segmentation Option
software complies with voluntary standards as detailed in Section
9, 11 and 16 of this premarket submission. The following quality
assurance measures were applied to the development of the
system:
• Risk Analysis
• Requirements Reviews
• Design Reviews
• Integration testing (System verification)
• Performance testing (Bench testing, verification)
• Safety testing (Verification) |
| | Summary of Clinical Tests: |
| | A usability study using consented clinical images was conducted
by three board certified Radiation Oncologists who were
considered experts. The study was meant to assess the following: |

1

GE Healthcare AdvantageSim™ MD with MR pelvic organ at risk segmentation Option 510(k) Premarket Notification Submission

y

Image /page/1/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters "G" and "E" intertwined within a circular border. The letters and the border are black, contrasting with the white background.

2

GE Healthcare AdvantageSim™ MD with MR pelvic organ at risk segmentation Option 510(k) Premarket Notification Submission

Image /page/2/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'G' and 'E' intertwined within a circular border. The letters and the border are black, contrasting with the white background.

:

3

Image /page/3/Picture/0 description: The image shows the text "K132944 Page 4 of 4" in a handwritten style. The text is arranged in two lines, with "K132944" on the top line and "Page 4 of 4" on the bottom line. Below the text, there is a logo of GE (General Electric) in a circular shape.

·

GE Healthcare

AdvantageSim™ MD with MR pelvic organ at risk segmentation Option 510(k) Premarket Notification Submission

Accuracy of measurementPrecision of the measurementEfficiency, as a comparison of amount of timeGeneral user Qualitative feedback The study results demonstrated that when used by qualified clinician the new software device has the potential to reduce inter- operator variability, provides statistically significant and practically meaningful clinical efficiency improvements and substantiates the characteristics of this feature, among others, as easy to learn, useful, efficient and providing increased throughput.
Conclusion:GE Healthcare considers the AdvantageSim MD with MR pelvic organ at risk segmentation Option software application to be as safe, as effective, and performance is substantially equivalent to the predicate devices.

4

Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized symbol resembling a person with outstretched arms, with three wavy lines below it. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the symbol.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

March 14, 2014

GE Hungary Kft. % Mr. Stephen Slavens Regulatory Affairs Director GE Medical Systems, LLC dba GE Healthcare 3000 N Grandview WAUKESHA WI 53188

Re: K132944

Trade/Device Name: AdvantageSim" MD MR pelvic organ at risk segmentation Option Regulation Number: 21 CFR 892.5840 Regulation Name: Radiation therapy simulation system Regulatory Class: II Product Code: KPQ Dated: January 20, 2014 Received: January 22, 2014

Dear Mr. Slavens:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

5

Page 2-Mr. Slavens

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/dcfault.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Michael D. O'Hara

for

Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

6

Indications for Use

510(k) Number (if known): K132944

Device Name: AdvantageSim™ MD with MR pelvic organ at risk segmentation Option

Indications For Use:

AdvantageSim™ MD is used to prepare geometric and analomical data relating to a proposed external beam radiotherapy treatment prior to dosimetry planning. Anatomical volumes can be defined automatically or manually in three dimensions using a set of CT images acquired with the patient in the proposed treatment position of the anatomical volumes may be assisted by additional CT, MR or PET studies that have been co-registered with the planning CT scan. Additionally, CT & PET data from a respiratory tracked examination may be used to allow the user define the target or treatment volume over a defined range of the respiratory cycle.

The geometric parameters of a proposed treatment field are selected to allow non-dosimetric. interactive optimization of field coverage. Defined anatomical structures and geometric treatments fields are displayed on transverse images, on reformatted sagittal, coronal or oblique images, on 3D views created from the images, or on a beam eye's view display with or without the display of defined structures with or without the display of digitally reconstructed radiograph.

Prescription Use V (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of Center for Devices and Radiological Health (CDRH)

Michael D. O'Hara

Page 1 of __ 1 _______________________________________________________________________________________________________________________________________________________________

(Division Sign-Off) Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health K132944 510(k)