(239 days)
Closed System drug Transfer Device (CSTD) for safe preparation, reconstitution, compounding and administration of drugs, including antineoplastic and hazardous drugs. This closed system mechanically prohibits the transfer of environmental contaminants into the system and the escape of drug or vapor concentrations outside the system, thereby minimizing individual and environmental exposure to drug vapor, aerosols, and spills and also prevents microbial ingress.
The Equashield system consists of a piston syringe (Syringe Unit), an adaptor to the medication vial (Vial Adaptor), an adaptor for the infusion bag for injection (Spike Adaptor 1), an adaptor for the infusion bag for withdrawal (Spike Adaptor W), an adaptor for female luer lock ports (Luer Lock Adaptor), nursing pair connectors for male and female luer lock ports (Female LL Connector and Male LL Connector), and a protective plug (Protective Plug). The Equashield closed system consists of 8 components that are dedicated to each other to create a system. The components are not intended to be used individually. Vial Adaptor is 48 x 31mm, 23mm at distal end. Plug is 51 x 17mm large. Female luer lock is 50 x 23 mm with a 5.2mm ID. Male luer lock is 42 x 21mm with a 15mm depth. Spike Adaptor is 75 x 42 mm and the height is 22mm.
The provided text is a 510(k) Summary for a medical device (Equashield Closed System Drug Transfer Device), which focuses on establishing substantial equivalence to predicate devices rather than directly presenting acceptance criteria and detailed study results. Therefore, much of the requested information cannot be extracted from this document.
Here's what can be inferred or explicitly stated based on the provided text, and where information is missing:
Acceptance Criteria and Reported Device Performance
The document does not explicitly list "acceptance criteria" in a numerical or categorical format with corresponding "reported device performance." Instead, it relies on demonstrating adherence to established consensus standards and substantial equivalence to predicate devices. The core performance claims are related to preventing microbial ingress and minimizing exposure to hazardous drugs.
| Acceptance Criterion (Implied) | Reported Device Performance (Implied from the "Principle of Operation" and "Characteristics" sections) |
|---|---|
| Prevention of microbial ingress | Substantiated to prevent microbial ingress up to a time period of 7 days. |
| Multiple reconnections capability | Substantiated for multiple reconnections of components up to 10 times. |
| Prohibition of environmental contaminants into the system | Mechanically prohibits the transfer of environmental contaminants into the system. Uses double membrane method to create a leak-proof connection. |
| Prevention of drug/vapor escape outside the system | Mechanically prohibits the escape of drug or vapor concentrations outside the system, thereby minimizing individual and environmental exposure to drug vapor, aerosols, and spills. System has integrated closed pressure equalization. Syringe barrel is sealed airtight also at its rear end. The closed expansion chamber makes the system airtight consequently containing all aerosols, particles and vapors. |
| Integration with standard medical equipment (Luer fittings) | Follows ISO 594-1:1998 and ISO 594-2:1998 (Conical fittings with a 6% (Luer) taper). |
| Biocompatibility | Follows ISO 10993-1:2009, ISO 10993-4:2002, ISO 10993-5:2009 for biological evaluation. |
| Sharps injury protection | Follows ISO 23908:2011 (Sharps injury protection). System contains needle stick safe design. |
| Sterility | Provided sterile as long as the package is intact, undamaged, and protective caps are secure. Adheres to EO/SAL 10-6 sterilization. |
Missing Information (Not available in the provided text):
- Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective): The document mentions "performance data" in the context of substantial equivalence but does not provide details on specific studies, sample sizes, or data provenance.
- Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience): Not applicable for this type of device and submission. There isn't "ground truth" established by experts in the sense of image interpretation or diagnosis. The "ground truth" here would be objective physical and chemical tests verifying the device's functionality.
- Adjudication method (e.g., 2+1, 3+1, none) for the test set: Not applicable as there's no human adjudication of results in the traditional sense for this device. Performance is measured through empirical tests.
- If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is a physical medical device, not an AI-powered diagnostic tool engaging human readers.
- If a standalone (i.e., algorithm only without human-in-the-loop performance) was done: Not applicable. This is a physical medical device.
- The type of ground truth used (expert consensus, pathology, outcomes data, etc.): For this device, the "ground truth" would be established through objective laboratory testing against specified standards (e.g., measuring microbial ingress, assessing leak-proof connections, verifying pressure equalization, physical integrity tests for sharps protection). The document refers to adherence to ISO standards (e.g., ISO 22413:2010 for transfer sets, ISO 10993 for biocompatibility).
- The sample size for the training set: Not applicable. This is a physical medical device, not a machine learning model.
- How the ground truth for the training set was established: Not applicable. This is a physical medical device.
Study/Evidence that Proves the Device Meets Acceptance Criteria:
The document states that the Equashield System Generation 2, its accessories, and components follow FDA recognized consensus standards. This implies that studies (tests) were conducted to demonstrate compliance with these standards. The specific standards listed are:
- ISO 594-1:1998, ISO 594-2:1998 (Luer fittings)
- ISO 7886-1:1993 (Sterile hypodermic syringes)
- ISO 8536-4:2010 (Infusion equipment, gravity feed)
- ISO 10993-1:2009, ISO 10993-4:2002, ISO 10993-5:2009 (Biological evaluation)
- ISO 14971:2007 (Risk management)
- ISO 22413:2010 (Transfer sets for pharmaceutical preparations)
- ISO 23908:2011 (Sharps injury protection)
The document explains that "Equivalence was determined using a side by side tabular comparison between the predicate and proposed devices which included: Features, Intended Use, Labeling, Materials, Specifications, Performance Data, and Technological Aspects." This comparison, along with adherence to recognized standards, forms the basis of their claim for substantial equivalence and, by extension, meeting the necessary performance expectations for safety and effectiveness. However, detailed results of these performance studies or comparisons are not provided in this summary.
{0}------------------------------------------------
Medical Ltd.
POB 12 , Tefen Industrial Park Migdal Tefen 24959 | Israel www.equashieid.com
510(k) Summary K132899
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
Date Prepared: May 12, 2014
Applicant Equashield Medical Ltd. POB 12 , Tefen Industrial Park Migdal Tefen 24959 , Israel
Contact Person Raymond Kelly Licensale Inc. 57 Lazy Brook Rd Monroe, CT 06468 USA Phone: (203) 880-4091
Alternate Contact Dorit Valal Projects Manager Equashield Medical Ltd. POB 12 , Tefen Industrial Park Migdal Tefen 24959 . Israel Phone: +972-49873737 (ex. 114) Cell: +972-549952810
Device Information Trade Name: Equashield Closed System drug Transfer Device (CSTD) Model Numbers: Generations 1 & 2: (VA-x, SU-x, SA-x, LL-x, MC-x, FC-x, PP-x) Classification Name: Closed Antineoplastic and Hazardous Drug Reconstitution and Transfer System Review Panel: General Hospital Product Code: ONB Device Class: Class II Regulation: 21 C.F.R. §880.5440
Predicate Device Information
K101940 Equashield Closed System, cleared November 19, 2010 K 130197 PhaSeal Closed System, cleared February 27, 2013
{1}------------------------------------------------
Intended Use/Indications for Use
Closed System drug Transfer Device (CSTD) for safe preparation, reconstitution, compounding and administration of drugs, including antineoplastic and hazardous drugs. This closed system mechanically prohibits the transfer of environmental contaminants into the system and the escape of drug or vapor concentrations outside the system, thereby minimizing individual and environmental exposure to drug vapor, aerosols, and spills and also prevents microbial ingress.
Principle of Operation
Equashield (first and second generation) is a device which has been substantiated to prevent microbial ingress up to a time period of 7 days and has been substantiated for multiple reconnections of components up to 10 times. Equashield is a Closed System drug Transfer Device (CSTD) which does not require priming, does not include an Insertion Tube, and is not a "Gravity Feed" System. Equashield does contain accessories which contain tubing, one accessory uses "drip therapy" but no tubing contains plasticizer DEHP. Equashield is a single use, nontoxic, non-pyrogenic fluid path system which is provided sterile as long as the package is intact, undamaged, and protective caps are secure. This closed system mechanically prohibits the transfer of environmental contaminants into the system and the escape of drug or vapor concentrations outside the system, thereby minimizing individual and environmental exposure to drug vapor, aerosols, and spills and also prevents microbial ingress. System contains filters and needles using a needle stick safe design,
Device Description
The Equashield system consists of a piston syringe (Syringe Unit), an adaptor to the medication vial (Vial Adaptor), an adaptor for the infusion bag for injection (Spike Adaptor 1), an adaptor for the infusion bag for withdrawal (Spike Adaptor W), an adaptor for female luer lock ports (Luer Lock Adaptor), nursing pair connectors for male and female luer lock ports (Female LL Connector and Male LL Connector), and a protective plug (Protective Plug). The Equashield closed system consists of 8 components that are dedicated to each other to create a system. The components are not intended to be used individually. Vial Adaptor is 48 x 31mm, 23mm at distal end. Plug is 51 x 17mm large. Female luer lock is 50 x 23 mm with a 5.2mm ID. Male luer lock is 42 x 21mm with a 15mm depth. Spike Adaptor is 75 x 42 mm and the height is 22mm.
Accessories Description
Y-Site Tubing - this accessory consists of a Luer Lock Adaptor permanently attached to a Y-Site. Short pieces of tubing extend from each of the other two Y-Site ports, creating one tubing line. One tube is terminated with a female luer lock and the other is terminated with a male luer lock. The Y-Site Tubing Accessory is intended to be attached in series to an infusion line and the Equashield Luer Lock Adaptor enables safe connection and injection with the Equashield Syringe Unit or Female Connector. Priming volume: 1.9ml, length 25cm, Y-angle 30 degrees.
{2}------------------------------------------------
Spike Adaptor Secondary Tubing - consists of a Spike Adaptor-1 for injection with a drip chamber at its outlet. The drip chamber with the attached tubing, the roller clamp and the male luer lock at the tubing end, all constitute standard secondary infusion tubing sets. The integration of the standard infusion tubing set with the Equashield Spike Adaptor provides a closed system protection during injection of drugs into the infusion bag and during their administration. It reduces the assembly steps and the risk of exposure to drugs associated with such assembly procedures. Length is 100 cm, priming volume 7.5ml.
Performance
No performance standards or special controls have been developed under Section 514 of the Federal Food, Drug, and Cosmetic Act ("FD&C Act") for Closed System Transfer Devices (CSTD). However, the Equashield System Generation 2, its accessories, and its components follow FDA recognized consensus standards:
- . Recognition Number 6-11; ISO 594-1:1998, Conical fittings with a 6% (Luer) taper for syringes, needles and certain other medical equipment - Part 1: General requirements
- . Recognition Number 6-129: ISO 594-2:1998, Conical fittings with a 6% (Luer) taper for syringes, needles and certain other medical equipment - Part 2: Lock fittings
- . Recognition Number 6-170: ISO 7886-1:1993, Sterile hypodermic syringes for single use - Part 1: Syringes for manual use
- Recognition Number 6-245: ISO 8536-4:2010, Infusion equipment for medical use Part . 4: Systems for single use, gravity feed
- Recognition Number 2-156: ISO 10993-1:2009 Biological evaluation of medical devices . -- Part 1: Evaluation and testing within a risk management process
- ISO 10993-4:2002, Biological evaluation of medical devices -- Part 4: Selection of tests . for interactions with blood
- Recognition Number 2-153: ISO 10993-5:2009, Biological evaluation of medical devices . -- Part 5: Tests for In Vitro cytotoxicity
- . Recognition Number 5-40: ISO 14971:2007, Medical devices - Application of risk management to medical devices
- . ISO 22413:2010, Transfer sets for pharmaceutical preparations -- Requirements and test methods
- Recognition Number 6-273: ISO 23908:2011, Sharps injury protection Requirements . and test methods - Sharps protection features for single-use hypodermic needles, introducers for catheters and needles used for blood sampling
Substantial Equivalence
Equivalence was determined using a side by side tabular comparison between the predicate and proposed devices which included: Features, Intended Use, Labeling, Materials, Specifications, Performance Data, and Technological Aspects
{3}------------------------------------------------
| Predicate Device (K13019 | Same | system and the escape of drum vapor contractions outside of drug system thereby minimizing ndividual and environmentsSystem is an airtight and leak-sposure to drug vapor, aeroso and spills. The system also prevents microbial ingress.vironmental contaminants | Class II | 888.544( | ONB |
|---|---|---|---|---|---|
| Predicate Device (K101940 | Same | nacists or other healthcarfessionals to prepare drugsned system to be used | Class II | 888.5440 | r H |
| Proposed Device | Same | 888.544( | ONB | ||
| tended Us | dications forUse | assificatio | Regulation Number | roduct Cod |
SE Table of Attributes
Page 32
{4}------------------------------------------------
| Proposed Device | Predicate Device (K101940) | Predicate Device (K130197) | |
|---|---|---|---|
| Vial Adaptor | Vial Adaptor | Vial Adaptor ("Protector") | |
| Syringe Unit | Syringe Unit | Syringe connector ("Injector"with female luer lock) | |
| Spike Adaptor I- for injection | Spike Adaptor I- for injection | Spike Adaptor ("InfusionAdaptor") | |
| Spike Adaptor W- for withdrawal | Spike Adaptor W- for withdrawal | ||
| Luer Lock Adaptor | Luer Lock Adaptor | Male "Connector Luer Lock" | |
| Systemcomponents | Male Luer Lock Connector | Male Luer Lock Connector | Male "Connector Luer Lock" |
| Female Luer Lock Connector | Female Luer Lock Connector | Female luer lock connector("Injector") | |
| Protective Plug | Protective Plug | NA | |
| Y-Site Tubing(Accessory) | NA | Luer Lock Adaptor with Y-sitetubing ("Y-Site Connector") | |
| Secondary Tubing(Accessory) | NA | Spike adaptor with secondarytubing ("Secondary Set with DripChamber") | |
| Characteristics | Closed System used forantineoplastic and hazardous drugreconstitution, transfer andadministration, in order to preventcontamination of the surroundingenvironment and of the drug | Closed System used forantineoplastic and hazardous drugreconstitution, transfer andadministration, in order to preventcontamination of the surroundingenvironment and of the drug | Closed System used forantineoplastic and hazardous drugreconstitution, transfer andadministration, in order to preventcontamination of the surroundingenvironment and of the drug |
| Principles ofOperation | Multi-component system.Components are intended to be usedas a system. | Multi-component system.Components are intended to be usedas a system. | Multi-component system.Components are intended to beused as a system. |
| TechnologicalCharacteristics | Proposed Device | Predicate Device (K101940) | Predicate Device (K130197) |
| A leak-proof connector with asingle use syringe permanentlyattached to it as part of the system. | A leak-proof connector with asingle use syringe permanentlyattached to it as part of the system. | A leak-proof connector("Injector") that needs to beconnected to an off the shelfsingle use syringe. | |
| All system components are sealedwith resealing membranes(Septum). When components arejoined together the two membranesare pressed together and thenpierced by needles. System hasintegrated closed pressureequalization. | All system components are sealedwith resealing membranes(Septum). When components arejoined together the two membranesare pressed together and thenpierced by needles. System hasintegrated closed pressureequalization. | All system components are sealedwith resealing membranes(Septum). When components arejoined together the twomembranes are pressed togetherand then pierced by needles.System has integrated closedpressure equalization. | |
| The system syringe is closed fromall sides. The syringe barrel issealed airtight also at its rear end,thereby isolating the plunger rodand the interior of the barrel. | The system syringe is closed fromall sides. The syringe barrel issealed airtight also at its rear end,thereby isolating the plunger rodand the interior of the barrel. | NA | |
| Proposed Device | Predicate Device (K101940) | Predicate Device (K130197) | |
| Secondary Tubing("Secondary Set with Drip Chamber") | Spike Adaptor with drip chamber and secondary tubing attached. | NA | Spike Adaptor ("infusion adaptor") with drip chamber and secondary tubing attached. |
| System | A fully encapsulated Syringe Unit is the active transfer device of this closed system.The syringe barrel is sealed airtight also at its rear end, thereby isolating the plunger rod and the interior of the barrel. A leak-proof connector is permanently welded to the syringe. The closed pressure equalization system with a chamber containing sterile air is built-in the Syringe Unit and makes the system airtight consequently containing all aerosols, particles and vapors.For transfer of fluids the Syringe Unit connects to the passive Vial Adaptors, infusion bag and infusion tubing adaptors of the system, using the double membrane method to create a leak-proof and drug residual-free connection. | A fully encapsulated Syringe Unit is the active transfer device of this closed system.The syringe barrel is sealed airtight also at its rear end, thereby isolating the plunger rod and the interior of the barrel. A leak-proof connector is permanently welded to the syringe. The closed pressure equalization system with a chamber containing sterile air is built-in the Syringe Unit and makes the system airtight consequently containing all aerosols, particles and vapors.For transfer of fluids the Syringe Unit connects to the passive Vial Adaptors, infusion bag and infusion tubing adaptors of the system, using the double membrane method to create a leak-proof and drug residual-free connection. | The system uses a double membrane to create a leak-proof connection. The closed expansion chamber makes the system airtight consequently containing all aerosols, particles and vapors. The expansion chamber also equalizes the pressure within the system. It can handle most types of IV sets and syringes to enable a totally safe and convenient administration of drugs. |
| Device Type | Rx/Single Use | Rx/Single Use | Rx/Single Use |
| Target Users | Licensed Pharmacists/Health Care Professionals | Licensed Pharmacists/Health Care Professionals | Licensed Pharmacists/Health Care Professionals |
| Environment | Hospitals and clinics | Hospitals and clinics | Hospitals and clinics |
| Sterilization | EO/SAL 10-6 | EO/SAL 10-6 | EO/SAL 10-6 |
{5}------------------------------------------------
...
.
Page 34
the country of the county of
and the comments of the country of
{6}------------------------------------------------
:
:
:
.
ﺮ
.
Page 35
{7}------------------------------------------------
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 13, 2014
Equashield Medical Ltd. c/o Raymond J Kelly IV Arazy Group 57 Lazy Brook Rd MONROE, CT 06468
Re: K132899
Trade/Device Name: Equashield Closed System drug Transfer Device (CSTD) Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular administration set Regulatory Class: II Product Code: ONB Dated: April 14, 2014 Received: April 17, 2014
Dear Mr. Kelly,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{8}------------------------------------------------
Page 2 - Mr. Kelly
forth in the quality systems (QS) regulation (2) CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
· Sincerely yours,
Mary S. Runner -S
Erin I. Keith, M.S. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{9}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known) K 132899
Device Name
Equashield Closed System drug Transfer Device (CSTD)
Indications for Use (Describe)
Closed System drug Transfer Device (CSTD) for preparation, compounding and administration of drugs, including antineoplastic and hazardous drugs. This closed system mechanically prohibits the transfer of environmental contaminants into the system and the escape of drug or vapor concentrations outside the system, thereby minimizing individual and environmental exposure to drug vapor, aerosols, and spills and also prevents microbial ingress.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
. . . .
Digitally signed by Richard C. Chapman Date: 2014.05.13 14:21:18 -04'00'
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
§ 880.5440 Intravascular administration set.
(a)
Identification. An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.(b)
Classification. Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.