K Number
K132848
Manufacturer
Date Cleared
2013-11-29

(79 days)

Product Code
Regulation Number
888.4580
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Piezosurgery FLEX is an ultrasonic surgical system consisting of handpieces and associated tips for cutting bone, osteotomy, osteoplasty and drilling in variety of surgical procedures, including but not limited to otolaryngological, oral/maxillofacial, hand, foot, neurosurgical, spine, and paalstic/reconstructive surgery. It may also be used with endoscopic visual assistance to perform the above listed procedures.

Device Description

Piezosurgery FLEX is a piezoelectric ultrasonic device that uses ultrasonic energy to generate mechanical micro-vibration of tip inserts intended for bone cutting. The device consists of a base control unit with intearated peristaltic pump, a sterilizable handpiece with connecting cord, a range of sterile, single use tip inserts, a torque wrench, a foot-pedal, an irrigation set, and other minor attachments.

The console is equipped with a touch panel that allows the surgeon to control device operation. The control panel also provides monitoring information, by means icons, in case of malfunction of the device. Inside the console are located the ultrasonic generator, the electrical power supply module and the micro-processor electronic board that controls and supervises the functional parameters of the device. The console is connected to mains power by an electrical cord, and includes connectors for the handpiece and for the foot-pedal. The console incorporates a peristaltic pump which provides, through the irrigation tubing set, a sterile fluid supply to the surgical site. Ultrasonic power and irrigation flow to the handpiece are simultaneously activated by pressing the foot-pedal. The handpiece contains a piezoelectric ultrasonic transducer which attaches to the generator (inside the console) by a cable at one end of the handpiece. Tip inserts are attached to the other end of the handpiece.

The touch panel of the console allows the operator to set the treatment parameters, as follows:

  • Seven levels of power: .
  • Six levels of irrigation: .
  • Three ultrasonic output modes; .
  • · The pump function.

The Piezosurgery FLEX uses piezoelectric ultrasonic technology to generate mechanical microvibrations of the tip insert connected to the handpiece, the piezoelectric transducer converting the electrical voltage supplied by the ultrasonic generator into mechanical energy that induces vibration of the tip insert at the resonant frequency of the tip insert.

AI/ML Overview

This 510(k) summary describes a traditional submission for the Mectron Spa Piezosurgery FLEX, an ultrasonic surgical system.


1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (Non-Clinical Testing)Reported Device Performance
Compliance with safety requirements of standard IEC 60601-1Confirmed
Compliance with EMC requirements of standard EN 60601-1-2Confirmed
Similar intended use/indications of use to predicate devicesConfirmed (demonstrated through submission)
Similar technology to predicate devicesConfirmed (demonstrated through submission)
No significant effect on safety and effectiveness due to technological differences from predicate devicesConfirmed (stated in submission)

2. Sample Size Used for the Test Set and Data Provenance

The provided document describes non-clinical testing focused on compliance with electrical safety and electromagnetic compatibility (EMC) standards. It does not contain information about a "test set" in the context of clinical performance evaluation (e.g., patient data). The testing performed was related to device characteristics.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

Not applicable. The document describes compliance testing against established engineering standards (IEC 60601-1, EN 60601-1-2), not a study requiring expert-established ground truth for a clinical dataset.

4. Adjudication Method for the Test Set

Not applicable. Adjudication methods (e.g., 2+1, 3+1) are typically used in clinical studies involving human interpretation or performance, which is not described here.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve With AI vs Without AI Assistance

Not applicable. This device is an ultrasonic surgical system, not an AI-based diagnostic or assistive technology that would involve human "readers" or an MRMC study.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

Not applicable. This device is a surgical instrument. "Standalone" performance testing would relate to its engineering specifications and functionality, which are covered by the non-clinical testing mentioned (IEC and EMC standards compliance).

7. The Type of Ground Truth Used

For the non-clinical testing: The "ground truth" was defined by the requirements and specifications outlined in the international standards IEC 60601-1 (medical electrical equipment safety) and EN 60601-1-2 (electromagnetic compatibility).

8. The Sample Size for the Training Set

Not applicable. The Piezosurgery FLEX is a hardware surgical device, not a machine learning or AI model that requires a training set.

9. How the Ground Truth for the Training Set Was Established

Not applicable. As noted above, this device does not utilize a "training set" in the context of machine learning.

{0}------------------------------------------------

K132848

510(k) Summary in accordance with 21 CFR 807.92
Proprietary Name:Piezosurgery
FLEX Type of 510(k) submission:Traditional
Date of Submission:September 5, 2013
Manufacturer:Mectron SpaVia Loreto, 15, 16042 Carasco- (GE) Italy
FDA Registration Number:3003933619
510(k) Owner:Mectron SpaVia Loreto, 15, 16042 Carasco- (GE) Italy
510(k) Contact:Roger GrayVP, Quality and RegulatoryDonawa Lifescience ConsultingPiazza Albania, 10, 00153 Rome, ItalyTel: +39 06 5782665Fax: +39 06 5743786
Common Name:Bone cutting instrument and accessories
Classification Regulation:21 CFR 888.4850Sonic surgical instrument andaccessories/attachments.
Product Codes:JDX, DZI, ERL, HBE, HWE
Class:Class II

ﺰ ﺗﻌ

Predicate devices: The Substantial Equivalence of the subject device of this submission is
based on the predicate devices listed in Table 5.1. :

Table 5.1: Predicate devices
Trade NameManufacturerProduct Code/s510(k)NumberDecisionDate
Piezosurgery MedicalMECTRON SpaJDX, DZI, ERL, HBE, HWEK083284April 8, 2009
Piezoelectric SystemSATELEC-ACTEON GroupJDX, DZI, ERL, HBE, HWEK100410April 28, 2010

{1}------------------------------------------------

Intended use/Indications for Use:

Piezosurgery FLEX is an ultrasonic surgical system consisting of handpieces and associated tips for cutting bone, osteotomy, osteoplasty and drilling in variety of surgical procedures, including but not limited to otolaryngological, oral/maxillofacial, hand, foot, neurosurgical, spine, and plastic/reconstructive surgery. It may

also be used with endoscopic visual assistance to perform the above listed procedures.

Device Description:

Piezosurgery FLEX is a piezoelectric ultrasonic device that uses ultrasonic energy to generate mechanical micro-vibration of tip inserts intended for bone cutting. The device consists of a base control unit with intearated peristaltic pump, a sterilizable handpiece with connecting cord, a range of sterile, single use tip inserts, a torque wrench, a foot-pedal, an irrigation set, and other minor attachments.

The console is equipped with a touch panel that allows the surgeon to control device operation. The control panel also provides monitoring information, by means icons, in case of malfunction of the device. Inside the console are located the ultrasonic generator, the electrical power supply module and the micro-processor electronic board that controls and supervises the functional parameters of the device. The console is connected to mains power by an electrical cord, and includes connectors for the handpiece and for the foot-pedal. The console incorporates a peristaltic pump which provides, through the irrigation tubing set, a sterile fluid supply to the surgical site. Ultrasonic power and irrigation flow to the handpiece are simultaneously activated by pressing the foot-pedal. The handpiece contains a piezoelectric ultrasonic transducer which attaches to the generator (inside the console) by a cable at one end of the handpiece. Tip inserts are attached to the other end of the handpiece.

The touch panel of the console allows the operator to set the treatment parameters, as follows:

  • Seven levels of power: .
  • Six levels of irrigation: .
  • Three ultrasonic output modes; .
  • · The pump function.

The Piezosurgery FLEX uses piezoelectric ultrasonic technology to generate mechanical microvibrations of the tip insert connected to the handpiece, the piezoelectric transducer converting the electrical voltage supplied by the ultrasonic generator into mechanical energy that induces vibration of the tip insert at the resonant frequency of the tip insert.

Technological characteristics:

The technological characteristics of the Piezosurgery Flex are very similar or identical to the predicate devices. Piezosurgery FLEX offers a simpler, lower cost alternative to the predicate devices, however, having a single peristaltic pump and handpiece instead of the two pumps and handpieces that are included with the predicate

devices. Piezosurgery FLEX also offers users three output modes instead of the four modes available in the

predicate devices. For the user interface, Piezosurgery FLEX has a touch panel, whereas the predicate devices incorporate an LCD touch screen. The operational ultrasonic frequency range has been extended in Piezosurgery FLEX to 24 - 36 kHz, from 24 - 29.5 kHz in one predicate device and approximately 28 kHz in the other predicate device. The footswitch in one predicate device offers a wider range of operational options that in Piezosurgery FLEX.

None of these technological differences between Piezosurgery Flex and the predicate devices has any significant effect on the safety and effectiveness of the subject device.

Non-clinical testing:

{2}------------------------------------------------

Non clinical testing of the device components covered by this 510(k) submission includes:

  • · Testing to confirm compliance with the safety requirements of standard IEC 60601-1
  • · Testing to confirm compliance with EMC requirements of standard EN 60601-1-2

Substantial Equivalence

The data presented in this submission demonstrate the similarities in the intended use/indications of use and technology between the Piezosurgery Flex and the referenced predicate devices, and thus supports a finding of substantial equivalence between the subject device and the predicate devices which are already in commercial distribution in the United States.

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of an eagle or bird-like figure with three curved lines representing its body and wings. The logo is surrounded by a circular border containing the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" in capital letters.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WOnt-Golly Silver Spring, MI) 20993-0002

November 29, 2013

MECTRON SPA % Mr. Roger Gray Donawa Lifescience Consulting Srl Piazza Albania, 10 Rome 00153 ITALY

Re: K132848

Trade/Device Name: Piezosurgery FLEX Regulation Number: 21 CFR 888.4580 Regulation Name: Sonic surgical instrument and accessories/attachments Regulatory Class: Class II Product Code: JDX Dated: September 26, 2013 Received: September 30, 2013

Dear Mr. Gray:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act s requirements. including, but not limited to: registration and listing (21

{4}------------------------------------------------

Page 2 - Mr. Roger Gray

CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYow/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Binita S. Ashar - S 2013.11.29 10:58:54 -05'00'

Binita S. Ashar, M.D., M.B.A., F.A.C.S. Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications for Use

510(k) Number (if known): K132848

Device Name: Piezosurgery FLEX

Indications For Use: The Piezosurgery FLEX is an ultrasonic surgical system consisting of handpieces and associated tips for cutting bone, osteotomy, osteoplasty and drilling in variety of surgical procedures, including but not limited to otolaryngological, oral/maxillofacial, hand, foot, neurosurgical, spine, and paalstic/reconstructive surgery. It may also be used with endoscopic visual assistance to perform the above listed procedures.

Prescription Use X______________________________________________________________________________________________________________________________________________________

AND/OR

Over-The-Counter Use

(Part 21 CFR 801 Subpart D)

(21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of Center for Devices and Radiological Health (CDRH)

.

Binita S. Ashar -S. 2013.11.29 11:04:05 -05'00'

(Division Sign-Off) for BSA

Division of Surgical Devices

510(k) Number K132848

§ 888.4580 Sonic surgical instrument and accessories/attachments.

(a)
Identification. A sonic surgical instrument is a hand-held device with various accessories or attachments, such as a cutting tip that vibrates at high frequencies, and is intended for medical purposes to cut bone or other materials, such as acrylic.(b)
Classification. Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 888.9.