K Number
K132828
Manufacturer
Date Cleared
2013-12-04

(86 days)

Product Code
Regulation Number
862.1140
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Elecsys Calcitonin Immunoassay:
The Calcitonin Immunoasay is intended for the in vitro quantitative determination of human calcitonin (thyrocalcitonin) in serum and plasma. The calcitonin determination is intended to be used as an aid in the diagnosis and treatment of diseases involving the thyroid and parathyroid glands, including carcinoma and hyperparathyroidism in conjunction with other clinical and laboratory findings.

Elecsys Calcitonin CalSet:
Calcitonin CalSet is used for calibrating the quantitative Elecsys Calcitonin assay on the Elecsys and cobas e immunoassay analyzers.

Elecsys Calcitonin CalCheck 5:
The Elecsys CalCheck 5 is an assayed control for use in calibration verification and for use in the verification of the calcitonin assay range established by the Elecsys and cobas e immunoassay analvzers.

Elecsys PreciControl Varia:
The Elecsys PreciControl Varia is used for quality control of the Elecsys immunoassays on Elecsys and cobas e immunoassay analyzers.

Device Description

The Calcitonin Assay employs monoclonal antibodies specifically directed against hCT. The antibodies labeled with a ruthenium complex and biotin, respectively consist of mouse-specific components.
Results are determined using a calibration curve that is generated specifically on each instrument by a 2 point calibration and a master curve (5-point-calibration) provided with the reagent bar code.
The Elecsys Calcitonin CalSet is a lyophilized product based on buffered equine serum. It has been standardized to IRP WHO Reference Standard 89/620.
The Elecsys Calcitonin CalCheck 5 is a lyophilized product based on buffered equine serum. It has been standardized to IRP WHO Reference Standard 89/620.
The PreciControl Varia is a multi-composite lyophilized 3 level control set which has been previously cleared (K111506), with added calcitonin analyte.

AI/ML Overview

This document describes the specifications and comparative characteristics of the Elecsys Calcitonin Immunoassay and its associated calibrators and quality controls against predicate devices. Here's a breakdown of the requested information:

1. Table of Acceptance Criteria and Reported Device Performance (Elecsys Calcitonin Immunoassay vs. Predicate Device)

FeaturePredicate Device: Immulite 1000 Calcitonin, Model LKCL (K023304)Elecsys Calcitonin Immunoassay (New Device)
Intended UseTo measure calcitonin levels in serum for diagnosis and treatment of thyroid and parathyroid gland diseases.Quantitative determination of human calcitonin in serum and plasma for diagnosis and treatment of thyroid and parathyroid gland diseases (including carcinoma and hyperparathyroidism) in conjunction with other clinical and laboratory findings.
Assay ProtocolSolid-phase, enzyme-labeled, two-site chemiluminescent immunometric assay; 2 x 30 minutes incubation, 60 minutes total duration.Murine monoclonal antibodies (ruthenium complex and biotin labeled) directed against hCT. Streptavidin coated microparticles as solid phase. Sandwich principle. 18 minutes total duration.
Detection ProtocolChemiluminescent Immunometric assayElectrochemiluminescent Immunoassay
Instrument PlatformImmulite 1000cobas e 411
Sample Volume75 µL50 µL
Sample TypeHuman serum or heparinized plasmaHuman serum and plasma treated with K2-EDTA, K3-EDTA, lithium heparin, and lithium heparin plasma gel separation tubes.
CalibratorSiemens Calcitonin Adjustors (LCLL, LCLH)Elecsys Calcitonin CalSet
Calibration IntervalRecommended: 4 weeksOnce per reagent lot (fresh reagent, <24 hours since registered on analyzer). Renewed: After 2 months (56 days) for same reagent lot; After 7 days for same reagent kit on analyzer; As required (e.g., QC outside limits).
ControlsTwo levels (low and high) of calcitonin.Elecsys PreciControl Varia
Traceability/StandardizationStandardized against WHO 2nd IRP 89/620Same (Standardized against WHO 2nd IRP 89/620)
Reagent StabilityStable at 2-8°C until expiration dateUnopened: 2-8°C until expiration date. Opened: 2-8°C for 12 weeks. On Analyzers: 4 weeks.
Measuring Range0.6 - 2000 pg/mL1.0 - 2000 pg/mL
Precision (Within-run)Sample 1: 22 pg/mL, 3.9% CVSample 2: 44 pg/mL, 3.0% CVSample 3: 145 pg/mL, 2.3% CVSample 4: 621 pg/mL, 2.2% CVSample 5: 1029 pg/mL, 2.4% CVSample 6: 1207 pg/mL, 2.3% CVcobas e 411 (Repeatability):0.03 SD @ 1.21 pg/mL3.0% CV @ 11.5 pg/mL2.5% CV @ 48.5pg/mL2.9% CV @ 482 pg/mL2.2% CV @ 1910 pg/mL0.19 SDPCV1 @ 8.88 pg/mL1.5% CVPCV2 @ 97.7 pg/mL
Precision (Total/Intermediate)Sample 1: 22 pg/mL, 11.4% CVSample 2: 44 pg/mL, 7.5% CVSample 3: 145 pg/mL, 7.5% CVSample 4: 621 pg/mL, 10.6% CVSample 5: 1029 pg/mL, 8.5% CVSample 6: 1207 pg/mL, 9.5% CV(Intermediate):0.04 SD @ 1.21 pg/mL3.6% CV @ 11.5 pg/mL3.5% CV @ 48.5pg/mL4.0% CV @ 482 pg/mL3.4% CV @ 1910 pg/mL0.26 SDPCV1 @ 8.88 pg/mL2.6% CVPCV2 @ 97.7 pg/mL
Analytical Sensitivity2 pg/mLLimit of Blank (LoB): 0.3 pg/mLLimit of Detection (LoD): 0.5 pg/mLLimit of Quantitation (LoQ): 1 pg/mL
Hook EffectNo high-dose hook effect up to 25,000 pg/mL.No high-dose hook effect up to 1,000,000 pg/mL.
Limitations (Interference)Hemoglobin < 550 mg/dL, Bilirubin < 200 mg/L, Lipemia < 3000 mg/dL.Hemoglobin < 200 mg/dL, Bilirubin < 66 mg/dL, Lipemia < 2000 mg/dL, Biotin < 40 ng/mL, Rheumatoid factors < 1,200 IU/mL, IgG < 4 g/dL, IgM < 0.7 g/dL, IgA < 1.6 g/dL. No interference from 17 commonly used and 12 special pharmaceuticals. (Minimizes interference from extremely high titers of antibodies).
Method Comparison (Correlation)n = 150 (Min = 2.04 pg/mL, Max = 1779 pg/mL)Passing/Bablok: Slope 1.01, Intercept -0.404, Tau/r 0.925Deming: Slope 0.998, Intercept -0.305, Tau/r 0.991Not explicitly stated as "acceptance criteria", but results are provided for comparison to predicate.
Clinical Reference Range (95th/97.5th percentile)95th percentile:Female 5.0 pg/mLMale 8.4 pg/mL97.5th percentile:Female 7.63 pg/mLMale 14.3 pg/mL

2. Sample Size Used for the Test Set and Data Provenance

  • Method Comparison: n = 150 patient samples were used for the method comparison study against the predicate device.
  • Data Provenance: The document does not explicitly state the country of origin or whether the data was retrospective or prospective for the method comparison or other performance studies. However, the manufacturer is Roche Diagnostics, with establishment registrations in Germany and the United States.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

  • This information is not provided in the given document. For an in vitro diagnostic device measuring a quantitative analyte like Calcitonin, "ground truth" is typically established by the reference method or by the predicate device's measured values, rather than by human expert review of images or clinical cases.

4. Adjudication Method for the Test Set

  • This information is not applicable and not provided for this type of in vitro diagnostic device. Adjudication methods (e.g., 2+1, 3+1) are typically used in studies involving subjective interpretation, such as imaging studies, where multiple experts review and reconcile findings.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size

  • A Multi-Reader Multi-Case (MRMC) comparative effectiveness study is not applicable and not mentioned for this in vitro diagnostic device. These studies are relevant for medical imaging AI where human readers interpret diagnostic images. The performance of this device is measured through analytical parameters like precision, sensitivity, and method comparison with a predicate device.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

  • This device is an in vitro diagnostic immunoassay, meaning its performance is inherently "standalone" in the sense that it analytically measures calcitonin levels. Its performance characteristics (precision, analytical sensitivity, measuring range, etc.) are determined directly by the assay and instrument without a human "in-the-loop" during the measurement process. The results are then interpreted by clinicians. The provided performance metrics (precision, analytical sensitivity, hook effect, method comparison, and interference studies) represent the standalone performance of the assay.

7. The Type of Ground Truth Used

  • For the method comparison study, the predicate device's measurements (Immulite 1000 Calcitonin) served as the "ground truth" or reference for comparison. The new device's measurements were correlated against these values.
  • For standardization and traceability, both the predicate and the new device were standardized against the WHO 2nd IRP 89/620 (International Reference Preparation) for calcitonin. This international standard effectively acts as the ultimate "ground truth" for calibrating and standardizing calcitonin measurements.

8. The Sample Size for the Training Set

  • This document describes performance verification studies, not the development or "training" of an algorithm in the machine learning sense. Therefore, information on a "training set" sample size is not provided and not directly applicable to the described analytical performance studies of a chemical immunoassay.

9. How the Ground Truth for the Training Set Was Established

  • As mentioned in point 8, the concept of a "training set" with ground truth in the context of an algorithm or AI is not directly applicable here. For the standardization of the Elecsys Calcitonin Immunoassay, it is stated that the device and its calibrators are standardized to the IRP WHO Reference Standard 89/620. This international reference standard defines the "ground truth" for calcitonin concentration for the purpose of accurate measurement and calibration.

{0}------------------------------------------------

Introduction According to the requirements of 21 CFR 807.92, the following information provides sufficient detail to understand the basis for a determination of substantial equivalence. Submitter Roche Diagnostics 9115 Hague Road Name, Address, P.O. Box 50416 DEC 0 4 2013 Indianapolis, IN 46250-0415 Contact Contact Person: Kelli Turner (Regulatory Affairs Principal) . Phone: (317) 521-4515 Fax: (317) 521-2324 o Email: kelli.turner@roche.com o Date Prepared: November 25, 2013 Device Name Proprietary name: (1) Elecsys Calcitonin Immunoassay (2) Elecsys Calcitonin CalSet (3) Elecsys Calcitonin CalCheck 5 (4) Elecsys PreciControl Varia Common name: (1) Calcitonin assay (2) Calcitonin CalSet (3) Calcitonin CalCheck 5 (4) PreciControl Varia Classification name: 1) Radioimmunoassay, Calcitonin 2) Calibrator, Secondary 3) Quality control material 4) Quality control material

{1}------------------------------------------------

Device Name,continuedProduct Code:1) JKR2) JIT3) JJY4) JJY
Predicate Device:1) Assay: Siemens Immulite 1000 Calcitonin,Model LKCL (K023304)2) Calibrator: Elecsys Vitamin D CalSet (K113546)3) Control: Elecsys T4 CalCheck 5 (K112528)4) Control: Elecsys PreciControl Varia (K111506)
EstablishmentRegistrationFor the Elecsys Calcitonin test system, the establishment registrationnumber for Roche Diagnostics GmbH in Mannheim, Germany, is 9610126and for Penzberg, Germany, is 9610529. The establishment registrationnumber for Roche Diagnostics in the United States is 1823260
ClassificationThe FDA has classified the Radiometry, Calcitonin test system and theCalibrator, Secondary (Calcitonin CalSet) as a Class II device. The FDA hasclassified the multi-analyte controls, all kinds (assayed and unassayed) as aClass I device (Calcitonin CalCheck 5 and PreciControl Varia).
PanelProductCodeClassification NameRegulation Citation
Clinical ChemistryJKRRadioimmunoassay,Calcitonin21 CFR 862.1140
Clinical ChemistryJITCalibrator, secondary21 CFR 862.1150
Clinical ChemistryJJYQuality Control material21 CFR 862.1660
Clinical ChemistryJJYQuality Control material21 CFR 862.1660

Analyzer Platform

For the analytical and clinical studies, the cobas e 411 was used as the master analyzer.

{2}------------------------------------------------

continued

The Calcitonin Assay employs monoclonal antibodies specifically directed Device against hCT. The antibodies labeled with a ruthenium complex and biotin, Description respectively consist of mouse-specific components.

Results are determined using a calibration curve that is generated specifically on each instrument by a 2 point calibration and a master curve (5-point-calibration) provided with the reagent bar code.

The Elecsys Calcitonin CalSet is a lyophilized product based on buffered equine serum. It has been standardized to IRP WHO Reference Standard 89/620.

The Elecsys Calcitonin CalCheck 5 is a lyophilized product based on buffered equine serum. It has been standardized to IRP WHO Reference Standard 89/620.

The PreciControl Varia is a multi-composite lyophilized 3 level control set which has been previously cleared (K111506), with added calcitonin analyte.

IntendedUse/Indicationsfor UseElecsys Calcitonin Reagent:• The Calcitonin immunoassay is intended for the in vitro quantitativedetermination of human calcitonin (thyrocalcitonin) in serum andplasma. The calcitonin determination is intended to be used as an aidin the diagnosis and treatment of diseases involving the thyroid andparathyroid glands, including carcinoma and hyperparathyroidism inconjunction with other clinical and laboratory findings.The electrochemiluminescence immunoassay “ECLIA” is intendedfor use on the indicated Elecsys and cobas e immunoassay analyzers.
Elecsys Calcitonin CalSet:• Calcitonin CalSet is used for calibrating the quantitative ElecsysCalcitonin assay on the Elecsys and cobas e immunoassay analyzers
Elecsys Calcitonin CalCheck 5:• The Elecsys Calcitonin CalCheck 5 is an assayed control for use incalcitonin calibration verification and for use in the verification ofthe calcitonin assay range established by the Elecsys and cobas eimmunoassay analyzers.
Elecsys PreciControl Varia:• The Elecsys PreciControl Varia is used for quality control of theElecsys immunoassays on Elecsys and cobas e immunoassayanalyzers.

{3}------------------------------------------------

510(k) Summary for Calcitonin test system, continued

Substantial The Elecsys Calcitonin Test System is substantially equivalent to other devices legally marketed in the United States. Equivalence

The Elecsys Calcitonin immunoassay is equivalent to Immulite 1000 Calcitonin (K023304).

Elecsys Calcitonin CalSet is equivalent to Elecsys Vitamin D CalSet (K113546).

Elecsys Calcitonin CalCheck 5 is equivalent to the Elecsys T4 CalCheck 5 (K112528).

Elecsys PreciControl Varia is equivalent to Elecsys PreciControl Varia (K111506).

Substantial Equivalence -Comparison

The following tables compare:

  • . The Elecsys Calcitonin Immunoassay with its predicate device (K023304).
  • . The Calcitonin CalSet with its predicate device (K113546).
  • The Calcitonin CalCheck 5 its the predicate device (K112528). .
  • PreciControl Varia to its predicate device (K111506). .

{4}------------------------------------------------

Comparison of Assays, Similarities and Differences

Table 1

Assay Comparison
FeaturePredicate Device: Immulite 1000Calcitonin, Model LKCL (K023304)Elecsys Calcitonin Immunoassay
IntendedUse/Indicationsfor UseIntended to measure the thyroidhormone calcitonin (thryocalcitonin)levels in serum. Calcitoninmeasurements are used in thediagnosis and treatment of diseasesinvolving the thyroid and parathyroidglands, including carcinoma andhyperparathyroidism (excessiveactivity of the parathyroid gland).The Calcitonin immunoassay isintended for the in vitro quantitativedetermination of human calcitonin(thyrocalcitonin) in serum and plasma.The calcitonin determination isintended to be used as an aid in thediagnosis and treatment of diseasesinvolving the thyroid and parathyroidglands, including carcinoma andhyperparathyroidism in conjunctionwith other clinical and laboratoryfindings.The electrochemiluminescenceimmunoassay "ECLIA" is intendedfor use on the indicated Elecsys andcobas e immunoassay analyzers.
AssayProtocolA solid-phase, enzyme-labeled, two-site chemiluminescent immunometricassay. The solid-phase, a polystyrenebead enclosed within an IMMULITE1000 Test Unit, is coated with amonoclonal murine antibody specificfor calcitonin.Incubation cycles: 2 x 30 minutesTotal duration of assay: 60 minutes.Assay employs murine monoclonalantibodies specifically directed againsthCT. One antibody is labeled withruthenium complex; the secondantibody is labeled with Biotin.Streptavidin coated microparticles areused as solid phase.Sandwich principle. Total duration ofassay: 18 minutes
DetectionProtocolChemiluminescent ImmunometricassayElectrochemiluminescentImmunoassay
Applications60-minute application18-minute application
Assay Comparison
FeaturePredicate Device: Immulite 1000Calcitonin, Model LKCL (K023304)Elecsys Calcitonin Immunoassay
General Assay Features
InstrumentPlatformImmulite 1000cobas e 411
SampleVolume75 µL50 µL
SampleTypeHuman serum or heparinized plasmaHuman serum and plasma treated withK2-EDTA, K3-EDTA, lithium heparinand lithium heparin plasma gelseparation tubes.
ReagentsA solid-phase, enzyme-labeled, two-site chemiluminescent immunometricassay. The solid-phase, a polystyrenebead enclosed within an IMMULITE1000 Test Unit, is coated with amonoclonal murine antibody specificfor calcitonin.Incubation cycles: 2 x 30 minutesTotal duration of assay: 60 minutes.Sandwich principle. Total duration ofassay: 18 minutesAssay employs murine monoclonalantibodies specifically directed againsthCT. One antibody is labeled withruthenium complex; the secondantibody is labeled with Biotin.Streptavidin coated microparticles areused as solid phase.
CalibratorSiemens Calcitonin Adjustors (LCLL,LCLH)Elecsys Calcitonin CalSet

{5}------------------------------------------------

Comparison of Assays-Similarities and Differences, continued

Table 1 continued

{6}------------------------------------------------

510(k) Summary for Elecsys Calcitonin, continued

Comparison of Assays-Similarities and Differences, continued

Table 1 continued

Assay Comparison
FeaturePredicate Device: Immulite 1000Calcitonin, Model LKCL(K023304)Elecsys Calcitonin Immunoassay
General Assay Features
CalibrationIntervalRecommended Adjustment interval:4 weeksCalibration must be performed onceper reagent lot using fresh reagent (i.e.not more than 24 hours since thereagent kit was registered on theanalyzer). Renewed calibration isrecommended as follows:cobas e 411 analyzers:After 2 months (56 days) whenusing the same reagent lot.After 7 days (when using thesame reagent kit on theanalyzer).As required: e.g. quality controlfindings outside the specified limits
ControlsUse controls or sample pools with atleast two levels (low and high) ofcalcitonin.Elecsys PreciControl Varia
Traceability /StandardizationStandardized against WHO 2nd IRP89/620Same
ReagentStabilityStable at 2-8°C until expiration dateUnopened:2-8°C - Up to the stated expirationdateOpened 2-8°C - 12 weeksOn Analyzers - 4 weeks

{7}------------------------------------------------

510(k) Summary for Elecsys Calcitonin, continued

Comparison of Assays-Similarities and Differences, continued

Table 1 continued

Assay Comparison
FeaturePredicate Device: Immulite 1000Calcitonin, Model LKCL (K023304)Elecsys Calcitonin Immunoassay
Labeled Performance Characteristics
MeasuringRange0.6 - 2000 pg/mL1.0 - 2000 pg/mL
PrecisionWithin-runSampleMeanSDCVcobas e 411:Within-run (will be labeled Repeatability)
1220.863.9%0.03 SD @ 1.21 pg/mL
2441.33.0%3.0% CV @ 11.5 pg/mL
31453.32.3%2.5% CV @ 48.5pg/mL
462113.82.2%2.9% CV @ 482 pg/mL
5102924.82.4%2.2% CV @ 1910 pg/mL
6120727.42.3%0.19 SDPCV1 @ 8.88 pg/mL1.5% CVPCV2 @ 97.7 pg/mL
TotalSampleMeanSDCVTotal (will be labeled Intermediate)
1222.511.4%0.04 SD @ 1.21 pg/mL
2443.37.5%3.6% CV @ 11.5 pg/mL
314510.97.5%3.5% CV @ 48.5pg/mL
462165.610.6%4.0% CV @ 482 pg/mL
5102987.38.5%3.4% CV @ 1910 pg/mL
612071159.5%0.26 SDPCV1 @ 8.88 pg/mL2.6% CVPCV2 @ 97.7 pg/mL

PCV2=PreciControl Varia level 2

{8}------------------------------------------------

Comparison of Assays-Similarities and Differences, continued

Table 1 continued

Assay Comparison
FeaturePredicate Device: Immulite 1000Calcitonin, Model LKCL (K023304)Elecsys Calcitonin Immunoassay
Labeled Performance Characteristics
AnalyticalSensitivity2 pg/mLLimit of Blank (LoB): = 0.3 pg/mLLimit of Detection (LoD): = 0.5pg/mLLimit of Quantitation (LoQ): = 1pg/mL
Hook EffectThere is no high-dose hook effect atconcentrations up to 25,000 pg/mL.There is no high-dose hook effect atconcentrations up to 1,000,000 pg/mL.
LimitationsThe assay is unaffected by the presence of:● Hemoglobin < 550 mg/dL.● Bilirubin < 200 mg/L● Lipemia < 3000 mg/dLThe assay is unaffected by:● Hemoglobin < 200 mg/dL.● Bilirubin < 66 mg/dL● Lipemia < 2000 mg/dL● Biotin < 40 ng/mL● Rheumatoid factors < 1,200 IU/mL● IgG < 4 g/dL● IgM < 0.7 g/dL● IgA < 1.6 g/dLIn vitro tests were performed on 17commonly used and 12 specialpharmaceuticals. No interference withthe assay was found.In rare cases, interference due toextremely high titers of antibodies toanalyte-specific antibodies, streptavidinor ruthenium can occur. These effectsare minimized by suitable test design.

{9}------------------------------------------------

Comparison of Assays-Similarities and Differences, continued

Table 1 continued

Immunoassay Comparison
FeaturePredicate Device: Immulite 1000Calcitonin, Model LKCL (K023304)Elecsys Calcitonin Immunoassay
Labeled Performance Characteristics
MethodComparisonn = 150Min = 2.04 pg/mLMax = 1779 pg/mL
Passing/BablokDeming
SlopeInterceptTau/r1.01-0.4040.9250.998-0.3050.991
ClinicalStudy/referencerange95th percentile:Female 5.0 pg/mLMale 8.4 pg/mL97.5th percentile:Female 7.63 pg/mLMale 14.3 pg/mL

{10}------------------------------------------------

Comparison of Assays-Similarities and Differences, continued

Table 1 continued

Elecsys Calcitonin Immunoassay on cobas e 411(master analyzer)
Analytic Specificity
Cross reactantMax.concentration tested(ng/mL)Highestcross-reactivityobserved (%)
Salmon Calcitonin2000.017
Porcine Calcitonin10000.007
Chicken Calcitonin10000.005
ACTH (1-39)human2000.037
C-Peptide800000.000
Calcitonin GeneRelated Peptide20000.002
PTH (1-84)3000.013
TSH2000 $ \mu $ IU/mL0.009
Insulin670000.000
Prolactin20000.001
Gastrin I40000.001
Elcatonin2000000.000
Katacalcin800000.000

{11}------------------------------------------------

Elecsys CalSet comparison
FeaturePredicate DeviceElecsys Vitamin D CalSetElecsys Calcitonin CalSet
Intended UseElecsys Vitamin D CalSet isused for calibrating thequantitative Elecsys Vitamin Dassay on the Elecsys and cobase immunoassay analyzers.Calcitonin CalSet is used forcalibrating the quantitative ElecsysCalcitonin assay on the Elecsysand cobas e immunoassayanalyzers.
LevelsTwoTwo
MatrixHuman serumBuffered (50 mmol/L HEPES)equine serum
FormatLyophilizedLyophilized
StabilityUnopened: up to the statedexpiration dateAfter reconstituting:At 2-8°C - 120 hoursAt -20°C - 90 days (freeze onlyonce)On Elecsys 2010/cobas e 411 at20-25°C: up to 5 hoursOn MODULAR ANALYTICSE170/cobas e 601: Use onlyonceUnopened: up to the statedexpiration dateAfter reconstituting:At 2-8°C – 72 hoursAt -20°C – 84 days (freeze onlyonce)At 20-25°C: up to 5 hours
CompositionBuffer: HEPES 50mMPreservative: Bronidox L 0.5%Buffer: HEPES 50mMHuman Calcitonin (synthetic)

Table 2

{12}------------------------------------------------

Table 3
-------------

ନ୍ଦର

: 上一篇:

Elecsys CalCheck comparison
CharacteristicElecsys T4 CalCheck 5(Predicate Device)Elecsys Calcitonin CalCheck 5
Intended UseThe Elecsys T4 CalCheck 5is an assayed control foruse in calibrationverification and for use inthe verification of the assayrange established by theElecsys T4 quantitativeassay reagent on theindicated Elecsys andcobas e immunoassayanalyzers.The Elecsys Calcitonin CalCheck 5 isan assayed control for use incalcitonin calibration verification andfor use in the verification of thecalcitonin assay range established bythe Elecsys and cobas e immunoassayanalyzers.
AnalyteT4Calcitonin
LevelsFiveFive
AssayMeasuringRange0.300 - 10.0 nmol/L1.0 - 2000 pg/mL
Check TargetValuesCheck 1: ≤ 10 nmol/LCheck 2: 100 nmol/LCheck 3: 160 nmol/LCheck 4: 250 nmol/LCheck 5: 320 nmol/LCheck 1: ≤ 1.00 pg/mLCheck 2: 50 pg/mLCheck 3: 500 pg/mLCheck 4: 1000 pg/mLCheck 5: 2000 pg/mL
FormatLyophilizedSame
HandlingReconstitute Check 1,Check 2, Check 3, Check 4,and Check 5 with exactly1.0 mL distilled ordeionized water. Allow tostand closed for 15 minutes,then mix gently byinversion.Same
StabilityUnopened:Store at 2-8°C untilexpiration dateReconstituted:15-25°C: 4 hoursUnopened:Store at 2-8°C until expirationdateReconstituted:20-25°C: 4 hours
MatrixLevel 1: BSA/Buffer matrixLevel 2-5: Human serumBuffered (50 mmol/L HEPES) equineserum

{13}------------------------------------------------

Premarket Notification, 510(k) for the Elecsys Calcitonin test system, continued

Elecsys PreciControl Varia comparison
FeaturePredicate DeviceElecsys PreciControl Varia(K111506)Elecsys PreciControl Varia
Intended UseElecsys PreciControl Varia is usedfor quality control of specifiedElecsys immunoassays on theElecsys and cobas e immunoassayanalyzers.Same with the addition of calcitoninanalyte.
Analyzer systemElecsys and cobas eimmunoassay analyzers.Same
FormatLyophilizedSame
LevelsThreeSame
MatrixHuman serumSame
FormatLyophilizedSame
Traceability• β-CrossLaps/serum (β-CTx)-Gravimetry• Osteocalcin-In-housereference system (commerciallyavailable osteocalcinimmuno/radio-binding assay• Parathyroid Hormone (PTHand PTH STAT)-In-housereference system (commerciallyavailable PTH radiobindingassay)• Vitamin B12-Commerciallyavailable radio-binding assay• Ferritin-NIBSC Standard80/602• Folate-Elecsys Folate IIAssay• Vitamin D -LC-MS/MS(which in turn has beenstandardized to the NISTstandard)• β-CrossLaps/serum (β-CTx)-Gravimetry• Osteocalcin-In-house referencesystem (commercially availableosteocalcin immuno/radio-bindingassay• Parathyroid Hormone (PTH andPTH STAT)-In-house referencesystem (commercially availablePTH radiobindingassay)• Vitamin B12-Commerciallyavailable radio-binding assay• Ferritin-NIBSC Standard80/602• Folate-Elecsys Folate II Assay• Vitamin D -LC-MS/MS (whichin turn has been standardized tothe NIST standard)• Calcitonin- standardized to IRPWHO Reference Standard 89/620
Elecsys PreciControl Varia comparison, continued
FeaturePredicate DeviceElecsys PreciControl VariaElecsys PreciControl Varia
Analyte· Ferritin (ng/mL)· Ferritin (ng/mL)
ConcentrationLevel 0 = 14Level 0 = 14
Level 1 = 150Level 1 = 150
Level 2 = 1000Level 2 = 1000
· Folate (ng/mL)· Folate (ng/mL)
Level 0 = N/ALevel 0 = N/A
Level 1 = 3.9Level 1 = 3.9
Level 2 = 12Level 2 = 12
· Vitamin B12 (pg/mL)· Vitamin B12 (pg/mL)
Level 0 = 230Level 0 = 230
Level 1 = 500Level 1 = 500
Level 2 = 1000Level 2 = 1000
· B-CTx (pg/mL)· B-CTx (pg/mL)
Level 0 = N/ALevel 0 = N/A
Level 1 = 320Level 1 = 320
Level 2 = 750Level 2 = 750
· Osteocalcin (ng/mL)· Osteocalcin (ng/mL)
Level 0 = N/ALevel 0 = N/A
Level 1 = 20Level 1 = 20
Level 2 = 100Level 2 = 100
· PTH and PTH STAT (pg/mL)· PTH and PTH STAT (pg/mL)
Level 0 = 25Level 0 = 25
Level 1 = 60Level 1 = 60
Level 2 = 200Level 2 = 200
· Vitamin D (ng/mL)· Vitamin D (ng/mL)
Level 0 = 12.8Level 0 = 12.8
Level 1 = 17Level 1 = 17
Level 2 = 32Level 2 = 32
· Calcitonin (pg/mL)
Level 0 = N/A
Level 1 = 10
Level 2 = 100

{14}------------------------------------------------

Premarket Notification, 510(k) for the Elecsys Calcitonin test system, continued

Table 4, continued

. •

{15}------------------------------------------------

Premarket Notification, 510(k) for the Elecsys Calcitonin test system, continued

Elecsys PreciControl Varia comparison, continued
FeaturePredicate DeviceElecsys PreciControl VariaElecsys PreciControl Varia
HandlingDissolve carefully the contentsof one bottle by adding exactly3.0 mL of distilled or deionizedwater and allow standing closedfor 30 minutes to reconstitute.Mix carefully, avoiding theformation of foam.Same
Volume(reconstituted)3.0 mLSame
StabilityUnopened at 2 – 8 °C:up to the stated expiration dateReconstituted/thawed serum:• at – 20 °C: 31 days (freezeonlyonce)• at 2 – 8 °C: 72 hours• at 20 – 25 °C on-board theanalyzers: up to 5 hoursSame

Table 4, continued

{16}------------------------------------------------

and the comments of the comments of the comments of the comments of the comments of the comments of the comments of the comments of the comments of the contraction of the con

Standard/GuidanceDocumentReferenceIn addition to FDA guidance regarding 510(k) submissions, the followingstandards were used for the performance studies.
• Evaluation of Precision Performance of QuantitativeMeasurement Methods; Approved Guideline - Second Edition.CLSI document EP5-A2, Volume 24, No. 25, August 2004.
• Protocols for Determination of Limits of Detection and Limits ofQuantitation; Approved Guideline. CLSI document EP 17-A,Volume 24, No. 34, October 2004.
• Evaluation of Detection Capability for Clinical LaboratoryMeasurement Procedures; Approved Guideline- Second Edition.CLSI document EP17-A2, Volume 32, No. 8, June 2012
• Evaluation of the Linearity of Quantitative MeasurementProcedures: A Statistical Approach; Approved Guideline. CLSIdocument EP6-A, Volume 23, No. 16, April 2003.
• Method Comparison and Bias Estimation Using Patient Samples;Approved Guideline; Approved Guideline. CLSI document EP-09-A2-IR, Volume 22, No. 19, September 2002

Conclusion Statement

.

: .

י

The submitted information in this premarket notification supports a substantial equivalence decision.

{17}------------------------------------------------

Image /page/17/Picture/0 description: The image shows the text "DEPARTMENT OF HEALTH & HUMAN SERVICES" in all caps. The text is in a bold, sans-serif font. The words are arranged on a single line and are centered horizontally. The text appears to be part of a document or letterhead.

Image /page/17/Picture/1 description: The image shows a circular logo for the U.S. Department of Health & Human Services. The logo features the department's name encircling a stylized symbol. The symbol is a white bird-like figure with three curved lines representing its body and wings, set against a dark background.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

December 4, 2013

ROCHE DIAGNOSTICS KELLI TURNER REGULATORY AFFAIRS PRINCIPAL 9115 HAGUE ROAD INDIANAPOLIS IN 46250-0416

Re: K132828 Trade/Device Name: Elecsys Calcitonin Immunoassay Elecsys Calcitonin CalSet Elecsys Calcitonin CalCheck 5 Elecsys PreciControl Varia Regulation Number: 21 CFR 862.1140 Regulation Name: Calcitonin test system Regulatory Class: II Product Codc: JKR, JIT, JJY Dated: September 17, 2013 Received: September 18, 2013

Dear Ms. Turner:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28. 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may. therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice. labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803): good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the

{18}------------------------------------------------

Page 2-Ms. Turner

electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulations (2) CFR Parts 801 and 809). please contact the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRF' s Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincercly yours.

Carol C. Benson -S for

Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{19}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known)

K132828

Device Name

Elecsys Calcitonin Immunoassay, Elecsys Calcitonin CalSet, Elecsys Calcitonin CalCheck 5, Elecsys PreciControl Varia

Indications for Use (Describe)

Elecsys Calcitonin Immunoassay:

The Calcitonin Immunoasay is intended for the in vitro quantitative determination of human calcitonin (thyrocalcitonin) in serum and plasma. The calcitonin determination is intended to be used as an aid in the diagnosis and treatment of diseases involving the thyroid and parathyroid glands, including carcinoma and hyperparathyroidism in conjunction with other clinical and laboratory findings.

Elecsys Calcitonin CalSet:

Calcitonin CalSet is used for calibrating the quantitative Elecsys Calcitonin assay on the Elecsys and cobas e immunoassay analyzers.

Elecsys Calcitonin CalCheck 5:

The Elecsys CalCheck 5 is an assayed control for use in calibration verification and for use in the verification of the calcitonin assay range established by the Elecsys and cobas e immunoassay analvzers.

Elecsys PreciControl Varia:

The Elecsys PreciControl Varia is used for quality control of the Elecsys immunoassays on Elecsys and cobas e immunoassay analyzers.

Type of Use (Select one or both, as applicable)

x Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FINA LISE
$ 410 W. Banim 4mAnd Annual Property of Children Come of Children Come of Children Come of Children Come of Children Come of Children Come of Children Come of Children Come of Children Come oיישואלי י
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

FORM FDA 3881 (9/13)

§ 862.1140 Calcitonin test system.

(a)
Identification. A calcitonin test system is a device intended to measure the thyroid hormone calcitonin (thyrocalcitonin) levels in plasma and serum. Calcitonin measurements are used in the diagnosis and treatment of diseases involving the thyroid and parathyroid glands, including carcinoma and hyperparathyroidism (excessive activity of the parathyroid gland).(b)
Classification. Class II.