(77 days)
The Illumina MiSeqDx(TM) Cystic Fibrosis Clinical Sequencing Assay is a targeted sequencing in vitro diagnostic system that re-sequences the protein coding regions and intrones of the Cystic Fibrosis Transmembrane Conductance Regulator (CFTR) gene in genomic DNA isolated from human periphers collected in K2EDTA. The test detects single nucleotide variants, and small InDels within the region sequenced, and additionally reports on two deep intronic mutations and two large deletions. The test is intended to be used on the Illumina MiSeqDx Instrument.
The test is intended to be used as an aid in the diagnosis of individuals with suspected cystic fibrosis (CF). The test is most appropriate when the patient has an atypical or non-classic presentation of CF or when other mutation panels have failed to identify both causative mutations. The results of the test are intented by a board-certified clinical molecular geneticist or equivalent and should be used in conjunction with other available information including clinical symptoms, other diagnostic tess, and family history. This test is not indicated for use for stand-alone diagnostic testing, for pre-implantation testing, carrier screening, newborn screening, or population screening.
The Illumina MiSeqDx Cystic Fibrosis Clinical Sequencing Assay consists of library preparation and sample indexing reagents, sequencing reagents and consumables, MiSeqDx instrument and data analysis software. Testing begins with genomic DNA from a peripheral whole blood sample. The genomic DNA is processed through the library preparation steps, which specifically amplifies the intended genomic regions of each sample while also adding the indexes for sample identification. Flow cell capture sequences are also added to the amplified products. The resulting sample libraries are then transferred into a MiSeqDx reagent cartridge which contains all of the reagents required for cluster generation and sequencing (Sequencing By Synthesis). The MiSeqDx Cartridge, MiSeqDx Flow Cell, and MiSeqDx SBS Solution (PR2) are then inserted into the MiSeqDx instrument, which performs cluster generation, sequencing and data analysis.
Here's a summary of the acceptance criteria and study details for the Illumina MiSeqDx™ Cystic Fibrosis Clinical Sequencing Assay based on the provided text:
Acceptance Criteria and Device Performance
| Acceptance Criteria | Reported Device Performance |
|---|---|
| Accuracy (Overall Agreement) | >99.99% |
| Positive Agreement (PA) for Variants (excluding PolyTG/PolyT) | 100% |
| Positive Agreement (PA) for Variants (including PolyTG/PolyT) | 99.66% |
| Positive Agreement (PA) for PolyTG/PolyT variants alone | 98.44% |
| Negative Agreement (NA) for Wild Type positions | >99.99% |
| Reproducibility (Overall Agreement, All Sites) | 99.70% |
| Reproducibility (PA) for Variants (excluding PolyTG/PolyT) | 99.60% |
| Reproducibility (PA) for Variants (including PolyTG/PolyT) | 99.22% |
| Reproducibility (PA) for PolyTG/PolyT variants alone | 97.83% |
| Reproducibility (NA) for Wild Type | 99.70% |
| DNA Extraction (Call Rate & Accuracy) | >99.99% (for all 3 methods tested) |
| DNA Extraction (Sample First Pass Rate) | 100% (for all 3 methods tested) |
| DNA Input (Accuracy & Call Rate) | 100% (for 25 ng to 1250 ng range) |
| Interfering Substances (Call Rate & Reproducibility) | 100% |
Study Details
2. Sample Size Used for the Test Set and Data Provenance
-
Accuracy Study Test Set:
- Total Samples: 500
- Clinical Samples: 355 (archived, anonymized gDNA from human blood) - Country of origin not specified, but the review is by the US FDA.
- Commercial Cell Line Samples: 11 (from clinical accuracy study) + 68 (from reproducibility study) = 79
- Clinical Samples (from extraction method evaluation): 14
- Synthetic Plasmid Samples: 52
- Data Provenance: Retrospective for archived clinical samples, prospective for cell line and synthetic samples generated for the study.
-
Reproducibility Study Test Set:
- Total Samples: 46 distinct samples (implied from the HGVS tables, with many tested in multiple replicates). The tables show analysis for 68 samples, each tested across 3 sites, with 2 operators each doing 3 runs per site, and with some samples having multiple replicates within those runs. This results in a large number of individual tests. The text explicitly states 68 cell line samples were evaluated in the reproducibility study for accuracy data.
- Data Provenance: Prospective (generated specifically for the study). The operator details suggest this was done as part of the validation study.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
- The ground truth reference methods were Sanger bi-directional sequencing and validated PCR based assays.
- The text does not specify the number of experts used to establish the ground truth for Sanger sequencing or the PCR assays, nor their specific qualifications. It only states these were "reference methods."
4. Adjudication Method for the Test Set
- The text mentions that "All results are based on initial testing. No repeat testing was done for this study" for the accuracy study. For reproducibility, some samples had "no calls" or "miscalls" initially, and further investigation was done (e.g., misidentified variant, low level contamination, improperly prepared synthetic specimen, switched samples). However, a formal adjudication method (e.g., 2+1, 3+1) involving a specific number of experts to resolve discrepancies against the reference is not explicitly described. The determination of miscalls and no calls seems to stem from comparison with the reference methods.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done
- No, an MRMC comparative effectiveness study involving human readers with and without AI assistance was not done. This study focuses on the standalone analytical performance (accuracy, reproducibility) of the device (an in vitro diagnostic system for genetic sequencing), not on its impact on human reader performance.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
- Yes, a standalone performance study was done. The entire accuracy and reproducibility analysis describes the performance of the Illumina MiSeqDx Cystic Fibrosis Clinical Sequencing Assay (which includes the instrument and data analysis software) in detecting genetic variants. The results (PA, NA, OA, call rates) are indicative of the algorithm's performance compared to established reference methods (Sanger sequencing and PCR assays). The "human-in-the-loop" aspect for interpretation is mentioned in the intended use, where results are to be interpreted by a board-certified clinical molecular geneticist, but the performance metrics provided are for the device's ability to 'call' the variants.
7. The Type of Ground Truth Used
- Expert Consensus (implied via Gold Standard Methods): The primary ground truth for SNVs and small InDels was Sanger bi-directional sequencing. For large deletions, two validated PCR-based assays were used, with accuracy confirmed by Sanger Sequencing. These are considered gold-standard methods in molecular diagnostics, implying an expert-established and validated ground truth.
8. The Sample Size for the Training Set
- The document does not specify a training set size. As this is a performance validation document for a sequencing assay, it describes the testing of the device rather than the development or training of an AI model in the conventional sense. The "MiSeqDx instrument and data analysis software" are components, but details on their development/training are not provided in this regulatory summary.
9. How the Ground Truth for the Training Set Was Established
- Since a training set is not explicitly mentioned or detailed for the device's software/algorithm development, the method for establishing its ground truth is not provided in this document. The focus is on validating the final product's performance against reference standards.
{0}------------------------------------------------
: --
510(k) Summary 1.3
The following 510(k) summary was prepared in accordance with 21 CFR 807.92.
NOV 1 9 2013
{1}------------------------------------------------
510(k) Summary
GENERAL INFORMATION
Illumina Inc. Submitted by: 5200 Illumina Way San Diego, CA 92122 858-202-4500 (phone) 858-202-4600 (fax)
Company Contact: Bryan Schneider Associate Director, Regulatory Affairs 858-255-5228 (phone) bschneider@illumina.com . :
:
: ,
Date Prepared: November 18, 2013
DEVICE IDENTIFICATION
Assay: ..... ... ... ..
Trade or Proprietary Name:
| Illumina MiSeqDx™ Cystic Fibrosis ClinicalSequencing Assay | |
|---|---|
| Common Name: | Sequencing by synthesis cystic fibrosis test |
| Classification Name: | CFTR (cystic fibrosis transmembrane conductanceregulatory) gene mutation detection (21 CFR 866.5900,Product Code PFS) |
| Predicate Device: | x-TAG Cystic Fibrosis 60 Kit v2 (K083845) |
{2}------------------------------------------------
DEVICE DESCRIPTION
The Illumina MiSeqDx Cystic Fibrosis Clinical Sequencing Assay consists of library preparation and sample indexing reagents, sequencing reagents and consumables, MiSeqDx instrument and data analysis software. Testing begins with genomic DNA from a peripheral whole blood sample. The genomic DNA is processed through the library preparation steps, which specifically amplifies the intended genomic regions of each sample while also adding the indexes for sample identification. Flow cell capture sequences are also added to the amplified products. The resulting sample libraries are then transferred into a MiSeqDx reagent cartridge which contains all of the reagents required for cluster generation and sequencing (Sequencing By Synthesis). The MiSeqDx Cartridge, MiSeqDx Flow Cell, and MiSeqDx SBS Solution (PR2) are then inserted into the MiSeqDx instrument, which performs cluster generation, sequencing and data analysis.
INTENDED USE
Illumina MiSeqDx™ Cystic Fibrosis Clinical Sequencing Assay
The Illumina MiSeqDx(TM) Cystic Fibrosis Clinical Sequencing Assay is a targeted sequencing in vitro diagnostic system that re-sequences the protein coding regions and intron/exon boundaries of the Cystic Fibrosis Transmembrane Conductance Regulator (CFTR) gene in genomic DNA isolated from human peripheral whole blood specimens collected in K2EDTA. The test detects single nucleotide variants, and small InDels within the region sequenced, and additionally reports on two deep intronic mutations and two large deletions. The test is intended to be used on the Illumina MiSeqDx Instrument.
The test is intended to be used as an aid in the diagnosis of individuals with suspected cystic fibrosis (CF). The test is most appropriate when the patient has an atypical or non-classic presentation of CF or when other mutation panels have failed to identify both causative mutations. The results of the test are intended to be interpreted by a board-certified clinical molecular geneticist or equivalent and should be used in conjunction with other available information including clinical symptoms, other diagnostic tests, and family history. This test is not indicated for use for stand-alone diagnostic purposes, fetal diagnostic testing, for preimplantation testing, carrier screening, newborn screening, or population screening. i : . . . . . . . . .
{3}------------------------------------------------
SUBSTANTIAL EQUIVALENCE
・・・・
| Characteristic | Illumina | Luminex (K083845) |
|---|---|---|
| Assay Name | Illumina MiSeqDx Cystic FibrosisClinical Sequencing Assay | Luminex xTAG®CysticFibrosis 60 Kit v2 |
| IntendedUse | The Illumina MiSeqDx CysticFibrosis Clinical SequencingAssay is a targeted sequencing invitro diagnostic system that re-sequences the protein codingregions and intron/exonboundaries of the Cystic FibrosisTransmembrane ConductanceRegulator (CFTR) gene ingenomic DNA isolated fromhuman peripheral whole bloodspecimens collected in K2EDTA.The test detects single nucleotidevariants, and small indels withinthe region sequenced, andadditionally reports on two deepintronic mutations and two largedeletions. The test is intended tobe used on the Illumina MiSeqDxInstrument.The test is intended to be used asan aid in the diagnosis ofindividuals with suspected cysticfibrosis (CF).This assay is mostappropriate when the patient has | The xTAG® Cystic Fibrosis 60kit v2 is a device used tosimultaneously detect andidentify a panel of mutationsand variants in the cysticfibrosis transmembraneconductance regulator(CFTR) gene in human bloodspecimens. The panelincludes mutations andvariants currentlyrecommended by theAmerican College of MedicalGenetics and AmericanCollege of Obstetricians andGynecologists(ACMG/ACOG) plus someof the world's most commonand North Americanprevalent mutations. ThexTAG Cystic Fibrosis 60 kitv2 is a qualitativegenotyping test whichprovides informationintended to be used forcarrier testing in adults of |
| Confidential 4 | ||
| Characteristic | Illumina | Luminex (K083845) |
| an atypical or non-classicpresentation of CF or when othermutation panels have failed toidentify both causative mutations.The results of the test areintended to be interpreted by aboard-certified clinical moleculargeneticist or equivalent andshould be used in conjunctionwith other available informationincluding clinical symptoms,other diagnostic tests, and familyhistory. This test is not indicatedfor use for fetal diagnostic testing,for pre-implantation testing,carrier screening, newbornscreening, or populationscreening.The test is intended to be used onthe Illumina MiSeqDx™Instrument. | reproductive age, as an aidin newborn screening, and inconfirmatory diagnostictesting in newborns andchildren.The kit is not indicated foruse in fetal diagnostic or pre-implantation testing. The kitis also not indicated forstand-alone diagnosticpurposes. | |
| Assay type | Sequencing by synthesis test | Qualitative nucleic acidmultiplex test |
| VariantsDetected | Mutations and variants from theprotein coding regions andintron/exon boundaries of theCFTR gene, including two deepintronic mutations and two largedeletions. | 60 CFTR mutations and 4variants (benignpolymorphisms) |
| Characteristic | Illumina | Luminex (K083845) |
| Technology | PCR-based amplification ofregions of interest that are thenhybridized to a flow cell to allowsequencing by synthesis | Multiplex PCR followed bymultiplex allele specificprimer extension forgenotyping, hybridized tomultiplex fluorescentmicroparticles, detected byflow cytometry. |
| Sample Type | Nucleic acid from K2EDTAanticoagulated blood | Nucleic acid from wholeblood anticoagulated witheither EDTA or citrate. |
| SamplePreparation | DNA extraction using validatedlaboratory method | Same |
| Contra-indications | This test is not indicated for usefor fetal diagnostic testing, forpre-implantation testing, carrierscreening, newborn screening, orpopulation screening. | Not indicated for fetaldiagnostic testing, for pre-implantation testing, or forstand-alone diagnosticpurposes. |
| AssayControls | Positive and negative controlsrequired, not supplied | Negative controls required,not supplied. Positivecontrols recommended, notsupplied. |
| InstrumentSystem | MiSeqDx | Luminex 100 or 200 IS |
{4}------------------------------------------------
MiSeqDx Cystic Fibrosis Clinical Sequencing Assay
.
、
・・・・・・・・・・・・・・・・・・・
Confidential 5
: :
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image contains the word "illumina" in a bold, sans-serif font. The letters are closely spaced, and the word appears to be a logo or brand name. The overall impression is clean and modern.
PERFORMANCE.CHARACTERISTICS
Accuracy
Accuracy of the Illumina MiSeqDx Cystic Fibrosis Clinical Sequencing Assay was assessed by evaluating 500 samples representing a wide variety of CFTR variants from four separate sources. The primary source of accuracy data was a clinical accuracy study conducted using a panel of 366 samples. The majority (n = 355) of
{6}------------------------------------------------
samples consisted of archived, anonymized clinical gDNA specimens isolated from human blood, the remaining 11 samples were obtained from commercially available cell line specimens.
Data from this study was supplemented with accuracy data from 68 cell line samples evaluated in the reproducibility study, 14 clinical samples from the extraction method evaluation analytical study, and 52 synthetic plasmid samples. The synthetic plasmids were designed to include the genomic context of rare variants, and contained anywhere from 1 to 10 variants within the same construct. They were linearized, diluted to genomic DNA equivalent copy numbers, and blended with human genomic DNA samples of wild type genotype at equivalent copy numbers to mimic a heterozygous sample.
For the MiSeqDx Cystic Fibrosis Clinical Sequencing Assay, a total of 5,206 positions were compared to the reference methods of Sanger bi-directional sequencing and PCR testing. The genotyping results for SNV and small InDel sites, including the PolyTG/PolyT region, were compared to Sanger bi-directional sequence analysis.
Two validated PCR based assays were used as the reference method for the two large deletions in the panel. Each duplex PCR assay made use of 2 primer sets to discriminate between wild type, heterozygous; and homozygous genotypes. One of the primer sets was designed to flank the deletion breakpoints, whereas the other amplified a region internal to the deletion. The two products were detected by size separation on an agarose gel. The PCR assays were validated using a panel of 28 samples in all (22 samples for each deletion) consisting of cell line and blood derived genomic DNA samples, and synthetic plasmids which encompassed the WT, HET and HOM genotypes for each large deletion. The PCR assays were confirmed to have 100% specificity and reproducibility for all samples tested, by evaluation of PCR products on an agarose gel. The accuracy of the PCR assays was confirmed using Sanger Sequencing and found to be 100% for all samples.
Accuracy was determined for each genotype through three statistical measures. Positive Agreement (PA) was calculated for each variant genotype by dividing the number of samples with agreeing variant calls by the total number of samples with that variant as identified by the reference methods. Negative Agreement (NA) was calculated across all wild type (WT) positions by dividing the number of concordant WT positions by the total number of WT positions as
{7}------------------------------------------------
MiSeqDx Cystic Fibrosis Clinical Sequencing Assay
illumına
defined by the reference methods. Overall Agreement (OA) was calculated across all reported positions by dividing the number of concordant WT and variant positions by the total number of reported positions as determined by the reference methods.
The MiSeqDx Cystic Fibrosis Clinical Sequencing Assay had a genotype-level PA of 99.66% including the PolyTG/PolyT variants and 100%, excluding PolyTG/PolyT variants (Table 1). The NA for all wild types was >99.99% and the OA for all WT and variants was >99.99%. The PolyTG/PolyT variant PA was 98.44%. All results are based on initial testing. No repeat testing was done for this study.
Accuracy of the PolyTG/PolyT variants is demonstrated in Table 2.
(),,一个人
" " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " "
・・・・・・・・・・・・・・・・・・・
and the country of the capital of the county of
、『『『』『
... .. . . . . . . . . . . . . . . . . . . . . . . . . . . .
{8}------------------------------------------------
| Company of the was and | ||
|---|---|---|
| A | ||
| All - Alle - A deller a more - comment - comments | ||
| .- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - | ||
| 1 | ||
| . | ||
| 1 | ||
| A September 2017 A 1 A 2 A 2 A 2 A 2 2 A 2 1 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 | ||
| 0011136 | ||
| 1 | ||
| the us dife which and | ||
| A M A S A T W S M A M & P W B B B B B B B B B B B B B B B B B B B B B B B B B B B B B B B B B B B B B B B B B B B B B B B B B B B B B B B B B B B B B B B B B B B B B B B B B | ||
| 1210 1 | ||
| Genotype(CommonName) | cDNA namecoordinate | cDNA name | VariantType | CFTRgeneregion(hg19) | Positive calls (Variants) | NoCalls | Miscalls | PositiveAgreement | ||
|---|---|---|---|---|---|---|---|---|---|---|
| ClinicalSamples | CellLineSamples | SyntheticSamples | ||||||||
| 117120141 | N/A | SNV | Exon1 | 25 | 3 | 0 | 0 | 0 | 100 | |
| 117120145 | N/A | SNV | Exon1 | 3 | 2 | 0 | 0 | 0 | 100 | |
| M1V | c.1A>G | SNV | Exon1 | 0 | 0 | 1 | 0 | 0 | 100 | |
| CFTR dele2, 3 | c.54-5940_273+10250del21kb | Del | Intron1 | 4 | 1 | 0 | 0 | 0 | 100 | |
| R31C | c.91C>T | SNV | Exon2 | 3 | 1 | 0 | 0 | 0 | 100 | |
| Q39X | c.115C>T | SNV | Exon2 | 0 | 0 | 1 | 0 | 0 | 100 | |
| E60X | c.178G>T | SNV | Exon3 | 6 | 1 | 0 | 0 | 0 | 100 | |
| P67L | c.200C>T | SNV | Exon3 | 1 | 0 | 1 | 0 | 0 | 100 | |
| R74W | c.220C>T | SNV | Exon3 | 0 | 2 | 0 | 0 | 0 | 100 | |
| R74Q | c.221G>A | SNV | Exon3 | 2 | 0 | 0 | 0 | 0 | 100 | |
| R75X | c.223C>T | SNV | Exon3 | 3 | 1 | 0 | 0 | 0 | 100 | |
| R75Q | c.224G>A | SNV | Exon3 | 20 | 1 | 0 | 0 | 0 | 100 | |
| G85E | c.254G>A | SNV | Exon3 | 6 | 2 | 0 | 0 | 0 | 100 | |
| Genotype(CommonNamecDNA namecoordinate) | cDNA name | VariantType | CFTRgeneregion(hg19) | Positive calls (Variants) | PositiveAgreement | |||||
| ClinicalSamples | CellLineSamples | SyntheticSamples | NoCalls | Miscalls | ||||||
| 394delTT | c.262_263delTT | DIV | Exon3 | 3 | 1 | 0 | 0 | 0 | 100 | |
| 405+1G>A | c.273+1G>A | SNV | Intron3 | 0 | 0 | 1 | 0 | 0 | 100 | |
| 406-1G>A | c.274-1G>A | SNV | Exon4 | 4 | 0 | 0 | 0 | 0 | 100 | |
| E92K | c.274G>A | SNV | Exon4 | 0 | 0 | 1 | 0 | 0 | 100 | |
| E92X | c.274G>T | SNV | Exon4 | 0 | 1 | 1 | 0 | 0 | 100 | |
| Q98X | c.292C>T | SNV | Exon4 | 0 | 0 | 2 | 0 | 0 | 100 | |
| 444delA | c.312delA | DIV | Exon4 | 0 | 2 | 0 | 0 | 0 | 100 | |
| 457TAT>G | c.325_327delTATinsG | DIV | Exon4 | 0 | 0 | 1 | 0 | 0 | 100 | |
| D110H | c.328G>C | SNV | Exon4 | 1 | 0 | 1 | 0 | 0 | 100 | |
| R117C | c.349C>T | SNV | Exon4 | 4 | 0 | 0 | 0 | 0 | 100 | |
| R117H | c.350G>A | SNV | Exon4 | 17 | 2 | 0 | 0 | 0 | 100 | |
| Y122X | c.366T>A | SNV | Exon4 | 0 | 1 | 0 | 0 | 0 | 100 | |
| F143LfsX10 | c.425delT | DIV | Exon4 | 0 | 1 | 0 | 0 | 0 | 100 | |
| 574delA | c.442delA | DIV | Exon4 | 0 | 0 | 2 | 0 | 0 | 100 | |
| Q151K | c.451C>A | SNV | Exon4 | 1 | 0 | 0 | 0 | 0 | 100 |
{9}------------------------------------------------
{10}------------------------------------------------
| કુﻟﻤﺴﺎnS0 | |
|---|---|
| ﻣﻨ | 0œœœ |
| 1 | S1 |
| Genotype(CommonNamecDNA namecoordinate) | cDNA name | VariantType | CFTRgeneregion(hg19) | Positive calls (Variants) | NoCalls | Miscalls | PositiveAgreement | ||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| 621+1G>T | c.489+1G>T | SNV | Intron4 | ClinicalSamples | CellLineSamples | SyntheticSamples | 0 | 0 | 100 | ||
| 621+3A>G | c.489+3A>G | SNV | Intron4 | 1 | 0 | 0 | 0 | 0 | 100 | ||
| 663delT | c.531delT | DIV | Exon5 | 1 | 0 | 1 | 0 | 0 | 100 | ||
| G178R | c.532G>A | SNV | Exon5 | 1 | 1 | 0 | 0 | 0 | 100 | ||
| 711+1G>T | c.579+1G>T | SNV | Intron5 | 3 | 1 | 0 | 0 | 0 | 100 | ||
| 711+3A>G | c.579+3A>G | SNV | Intron5 | 0 | 0 | 1 | 0 | 0 | 100 | ||
| 711+5 G->A | c.579+5G>A | SNV | Intron5 | 0 | 0 | 1 | 0 | 0 | 100 | ||
| 712-1 G->T | c.580-1G>T | SNV | Exon6 | 0 | 0 | 1 | 0 | 0 | 100 | ||
| H199Y | c.595C>T | SNV | Exon6 | 0 | 0 | 1 | 0 | 0 | 100 | ||
| P205S | c.613C>T | SNV | Exon6 | 1 | 0 | 1 | 0 | 0 | 100 | ||
| L206W | c.617T>G | SNV | Exon6 | 8 | 1 | 0 | 0 | 0 | 100 | ||
| A209S | c.625G>T | SNV | Exon6 | 0 | 1 | 0 | 0 | 0 | 100 | ||
| Q220X | c.658C>T | SNV | Exon6 | 0 | 0 | 1 | 0 | 0 | 100 | ||
| L227R | c.680T>G | SNV | Exon6 | 0 | 0 | 1 | 0 | 0 | 100 | ||
| 852del22 | c.720_741delAGGGAGAATCATCATCACTAC | DIV | Exon6 | 0 | 0 | 1 | 0 | 0 | 100 | ||
| Genotype(CommonNamecDNA name,coordinate) | cDNA name | VariantType | CFTRgeneregion(hg19) | Positive calls (Variants) | ClinicalSamples | CellLineSamples | SyntheticSamples | NoCalls" | Miscalls | PositiveAgreement | |
| E279D | c.837A>T | SNV | Exon7 | 1 | 0 | 0 | 0 | 0 | 100 | ||
| R297Q | c.890G>A | SNV | Exon8 | 2 | 0 | 0 | 0 | 0 | 100 | ||
| 1078delT | c.948delT | DIV | Exon8 | 1 | 1 | 0 | 0 | 0 | 100 | ||
| L320V | c.958T>G | SNV | Exon8 | 1 | 0 | 0 | 0 | 0 | 100 | ||
| G330X | c.988G>T | SNV | Exon8 | 1 | 1 | 0 | 0 | 0 | 100 | ||
| R334W | c.1000C>T | SNV | Exon8 | 6 | 1 | 0 | 0 | 0 | 100 | ||
| I336K | c.1007T>A | SNV | Exon8 | 0 | 1 | 0 | 0 | 0 | 100 | ||
| T338I | c.1013C>T | SNV | Exon8 | 0 | 0 | 1 | 0 | 0 | 100 | ||
| 1154insTC | c.1022_1023insTC | DIV | Exon8 | 0 | 1 | 0 | 0 | 0 | 100 | ||
| S341P | c.1021T>C | SNV | Exon8 | 0 | 0 | 1 | 0 | 0 | 100 | ||
| R347H | c.1040G>A | SNV | Exon8 | 6 | 1 | 1 | 0 | 0 | 100 | ||
| R347P | c.1040G>C | SNV | Exon8 | 3 | 2 | 0 | 0 | 0 | 100 | ||
| R352Q | c.1055G>A | SNV | Exon8 | 5 | 0 | 0 | 0 | 0 | 100 | ||
| Q359K/T360K | c.[1075C>A;1079C>A] | SNV | Exon8 | 0 | 0 | 1 | 0 | 0 | 100 | ||
| 1213delT | c.1081delT | DIV | Exon8 | 0 | 0 | 1 | 0 | 0 | 100 | ||
| Genotype(CommonName | cDNA namecoordinate) | cDNA name | VariantType | CFTRgeneregion(hg19) | Positive calls (Variants) | ClinicalSamples | CellLineSamples | SyntheticSamples | NoCalls | Miscalls | PositiveAgreement |
| 1248+1G>A | c.1116+1G>A | SNV | Intron8 | 0 | 0 | 1 | 0 | 0 | 100 | ||
| 1259insA | c.1127_1128insA | DIV | Exon9 | 0 | 0 | 2 | 0 | 0 | 100 | ||
| W401X(c.1202G>A) | c.1202G>A | SNV | Exon9 | 0 | 0 | 1 | 0 | 0 | 100 | ||
| W401X(c.1203G>A) | c.1203G>A | SNV | Exon9 | 0 | 0 | 1 | 0 | 0 | 100 | ||
| 1341+1G->A | c.1209+1G>A | SNV | Intron9 | 0 | 0 | 2 | 0 | 0 | 100 | ||
| PolyTGPolyT | N/A | PolyTGPolyT | Intron9 | 369 | 79 | 52 | 3 | 4 | 98.60 | ||
| 1461ins4 | c.1329_1330insAGAT | DIV | Exon10 | 0 | 0 | 1 | 0 | 0 | 100 | ||
| A455E | c.1364C>A | SNV | Exon10 | 4 | 2 | 0 | 0 | 0 | 100 | ||
| 1525-1G->A | c.1393-1G>A | SNV | Exon11 | 0 | 0 | 1 | 0 | 0 | 100 | ||
| S466X (C->A) | c.1397C>A | SNV | Exon11 | 0 | 0 | 1 | 0 | 0 | 100 | ||
| S466X (C->G) | c.1397C>G | SNV | Exon11 | 1 | 0 | 1 | 0 | 0 | 100 | ||
| L467P | c.1400T>C | SNV | Exon11 | 0 | 0 | 1 | 0 | 0 | 100 | ||
| V470M | c.1408G>A | SNV | Exon11 | 311 | 71 | 0 | 0 | 0 | 100 |
.
ﺗ
i
{11}------------------------------------------------
T
1
1
T
T
{12}------------------------------------------------
{13}------------------------------------------------
.
| VariantType | CFTRgeneregion(hg19) | ClinicalSamples | CellLineSamples | SyntheticSamples | NoCalls | Miscalls | PositiveAgreement |
|---|---|---|---|---|---|---|---|
| DIV | Exon11 | 1 | 0 | 1 | 0 | 0 | 100 |
| SNV | Exon11 | 0 | 1 | 0 | 0 | 0 | 100 |
| SNV | Exon11 | 1 | 0 | 0 | 0 | 0 | 100 |
| SNV | Exon11 | 0 | 0 | 2 | 0 | 0 | 100 |
| SNV | Exon11 | 0 | 0 | 1 | 0 | 0 | 100 |
| SNV | Exon11 | 4 | 2 | 0 | 0 | 0 | 100 |
| SNV | Exon11 | 7 | 0 | 0 | 0 | 0 | 100 |
| DIV | Exon11 | 4 | 2 | 0 | 0 | 0 | 100 |
| DIV | Exon11 | 84 | 29 | 0 | 0 | 0 | 100 |
| SNV | Exon11 | 0 | 1 | 0 | 0 | 0 | 100 |
| SNV | Exon11 | 1 | 1 | 0 | 0 | 0 | 100 |
| DIV | Exon11 | 1 | 0 | 0 | 0 | 0 | 100 |
| SNV | Exon11 | 2 | 0 | 0 | 0 | 0 | 100 |
| SNV | Exon11 | 0 | 0 | 1 | 0 | 0 | 100 |
| SNV | Exon11 | 3 | 2 | 0 | 0 | 0 | 100 |
'r
{14}------------------------------------------------
| greemerPositive | 100 | 100 | 00 I | 00 T | 00 T | 100 | 00 I | 100 | 100 | 100 | 100 | 100 | 100 | 100 | |
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Aiscall | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | |
| ાદ) તાલુકાઓ પૈકીના એક એવા ગુજરાત રાજ્યના | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | |
| yntheti< Samples | 0 | L | 0 | 0 | I | L | 0 | 0 | L | 0 | 0 | I | I | I | |
| ositive calls (Variants | ាជាតិ (ទី) | ਟ | 0 | I | દ | 0 | ਨ | I | € | 0 | ਟ | 0 | 0 | 0 | 0 |
| amplesClinica | 9 | 0 | P | ਹ | 0 | ਨ | દ | 8 | 0 | 8 | I | 0 | র ব | 0 | |
| (618) (1784) അമ്ലിക്CFTR | Exon11 | ntron11 | Exon12 | Exon12 | Exon12 | Exon12 | Exon12 | Exon12 | Exon12 | Exon12 | Exon12 | Exon12 | Exon12 | Exon12 | |
| Varian Type | SNV | SNV | SNV | SNV | SNV | SNV | SNV | SNV | SNV | SNV | SNV | SNV | SNV | SNV | |
| DNA nam | .1584G>I | .1585-8G> | .1585-1G>/ | .1624G> | :.1645A> | .1646G> | :1647T>G | :1652G>/ | .1654C> | .1657C>" | :.1666A>( | c.1673T>< | .1675G>A | .1679G> | |
| DNA name្រុកប្រចេះ សារបោះបាន សុវណ្ណបក្សត្រ សុំសុប្រជុំសូលប្រជុំស្oordinat | E528E | 717-8G-> | 717-1G>/ | G542X | S549R c.1645A>C | 5549N | S549R c.1647T>G | G551D | Q552X | R553X | 556V | resss | A559T | R560K | |
| Positive calls (Variants) | |||||||||||||||
| Genotype(CommonNamecDNA namecoordinate) | cDNA name | VariantType | CFTRgeneregion(hg19) | ClinicalSamples | CellLineSamples | SyntheticSamples | NoCalls | Miscalls | PositiveAgreement | ||||||
| R560T | c.1679G>C | SNV | Exon12 | 6 | 1 | 0 | 0 | 0 | 100 | ||||||
| 811+1.6kb A->G | c.1679+1.6kbA>G | SNV | Intron12 | 0 | 0 | 1 | 0 | 0 | 100 | ||||||
| 1812-1 G->A | c.1680-1G>A | SNV | Exon13 | 1 | 2 | 0 | 0 | 0 | 100 | ||||||
| A561T | c.1681G>A | SNV | Exon13 | 1 | 0 | 0 | 0 | 0 | 100 | ||||||
| V562I | c.1684G>A | SNV | Exon13 | 1 | 0 | 0 | 0 | 0 | 100 | ||||||
| Y569D | c.1705T>G | SNV | Exon13 | 0 | 0 | 1 | 0 | 0 | 100 | ||||||
| P574H | c.1721C>A | SNV | Exon13 | 0 | 1 | 0 | 0 | 0 | 100 | ||||||
| G576A | c.1727G>C | SNV | Exon13 | 4 | 1 | 0 | 0 | 0 | 100 | ||||||
| D579G | c.1736A>G | SNV | Exon13 | 0 | 1 | 0 | 0 | 0 | 100 | ||||||
| E585X | c.1753G>T | SNV | Exon13 | 0 | 1 | 0 | 0 | 0 | 100 | ||||||
| 1898+1G>A | c.1766+1G>A | SNV | Intron13 | 2 | 1 | 0 | 0 | 0 | 100 | ||||||
| 1898+3A>G | c.1766+3A>G | SNV | Intron13 | 0 | 1 | 0 | 0 | 0 | 100 | ||||||
| H609R | c.1826A>G | SNV | Exon14 | 0 | 1 | 0 | 0 | 0 | 100 | ||||||
| D614G | c.1841A>G | SNV | Exon14 | 0 | 0 | 2 | 0 | 0 | 100 | ||||||
| R668C | c.2002C>T | SNV | Exon14 | 5 | 2 | 0 | 0 | 0 | 100 |
{15}------------------------------------------------
{16}------------------------------------------------
| Genotype(Common Name:cDNA name coordinate) | cDNA name | Variant Type | CFTR gene region (hg19) | Positive calls (Variants)Clinical Samples | Positive calls (Variants)Cell Line Samples | Positive calls (Variants)Synthetic Samples | No Calls | Miscalls | Positive Agreement |
|---|---|---|---|---|---|---|---|---|---|
| R668H | c.2003G>A | SNV | Exon14 | 1 | 0 | 0 | 0 | 0 | 100 |
| 2143delT | c.2012delT | DIV | Exon14 | 2 | 1 | 0 | 0 | 0 | 100 |
| K684TfsX4 | c.2046_2047delAA | DIV | Exon14 | 0 | 0 | 1 | 0 | 0 | 100 |
| 2183AA>G | c.2051_2052delAAinsG | DIV | Exon14 | 3 | 1 | 0 | 0 | 0 | 100 |
| 2184delA | c.2052delA | DIV | Exon14 | 1 | 1 | 0 | 0 | 0 | 100 |
| 2184insA | c.2052_2053insA | DIV | Exon14 | 3 | 0 | 1 | 0 | 0 | 100 |
| S686Y | c.2057C>A | SNV | Exon14 | 0 | 1 | 0 | 0 | 0 | 100 |
| R709X | c.2125C>T | SNV | Exon14 | 1 | 0 | 2 | 0 | 0 | 100 |
| K710X | c.2128A>T | SNV | Exon14 | 3 | 0 | 0 | 0 | 0 | 100 |
| E725K | c.2173G>A | SNV | Exon14 | 2 | 0 | 0 | 0 | 0 | 100 |
| 2307insA | c.2175_2176insA | DIV | Exon14 | 3 | 0 | 2 | 0 | 0 | 100 |
| L732X | c.2195T>G | SNV | Exon14 | 0 | 0 | 2 | 0 | 0 | 100 |
| 2347delG | c.2215delG | DIV | Exon14 | 0 | 0 | 2 | 0 | 0 | 100 |
| P750L | c.2249C>T | SNV | Exon14 | 1 | 0 | 0 | 0 | 0 | 100 |
| V754M | c.2260G>A | SNV | Exon14 | 2 | 1 | 0 | 0 | 0 | 100 |
| Genotype(CommonNamecDNA namecoordinate) | cDNA name | VariantType | CFTRgeneregion(hg19) | Positive calls (Variants) | PositiveAgreement | ||||
| ClinicalSamples | CellLineSamples | SyntheticSamples | NoCalls | Miscalls | |||||
| R764X | c.2290C>T | SNV | Exon14 | 1 | 0 | 2 | 0 | 0 | 100 |
| 2585delT | c.2453delT | DIV | Exon14 | 0 | 0 | 2 | 0 | 0 | 100 |
| E822X | c.2464G>T | SNV | Exon14 | 0 | 0 | 2 | 0 | 0 | 100 |
| 2622+1G>A | c.2490+1G>T | SNV | Intron14 | 0 | 0 | 2 | 0 | 0 | 100 |
| E831X | c.2491G>T | SNV | Exon15 | 0 | 0 | 1 | 0 | 0 | 100 |
| D836Y | c.2506G>T | SNV | Exon15 | 0 | 1 | 0 | 0 | 0 | 100 |
| W846X | c.2537G>A | SNV | Exon15 | 0 | 1 | 0 | 0 | 0 | 100 |
| R851X | c.2551C>T | SNV | Exon15 | 0 | 0 | 1 | 0 | 0 | 100 |
| T854T | c.2562T>G | SNV | Exon15 | 212 | 44 | 0 | 0 | 0 | 100 |
| 2711delT | c.2583delT | DIV | Exon15 | 0 | 0 | 1 | 0 | 0 | 100 |
| V868V | c.2604A>G | SNV | Exon15 | 2 | 0 | 0 | 0 | 0 | 100 |
| 2657+2_2657+3insA | c.2657+2_2657+3insA | DIV | Intron16 | 0 | 0 | 1 | 0 | 0 | 100 |
| 2789+5G>A | c.2657+5G>A | SNV | Intron16 | 9 | 1 | 0 | 0 | 0 | 100 |
| Q890X | c.2668C>T | SNV | Exon17 | 1 | 0 | 0 | 0 | 0 | 100 |
| A923A | c.2769C>T | SNV | Exon17 | 1 | 0 | 0 | 0 | 0 | 100 |
| Genotype(CommonNamecDNA namecoordinate) | cDNA name | VariantType | CFTRgeneregion(hg19) | Positive calls (Variants) | NoMiscalls | PositiveAgreement | |||
| ClinicalSamples | CellLineSamples | SyntheticSamples | |||||||
| L927P | c.2780T>C | SNV | Exon17 | 0 | 0 | 1 | 0 | 100 | |
| S945L | c.2834C>T | SNV | Exon17 | 0 | 0 | 1 | 0 | 100 | |
| M952T | c.2855T>C | SNV | Exon17 | 1 | 0 | 0 | 0 | 100 | |
| 3007delG | c.2875delG | DIV | Exon17 | 0 | 0 | 1 | 0 | 100 | |
| T966T | c.2898G>A | SNV | Exon17 | 5 | 0 | 0 | 0 | 100 | |
| G970R | c.2908G>C | SNV | Exon17 | 0 | 0 | 1 | 0 | 100 | |
| S977F | c.2930C>T | SNV | Exon18 | 0 | 0 | 1 | 0 | 100 | |
| 3120G>A | c.2988G>A | SNV | Exon18 | 1 | 0 | 0 | 0 | 100 | |
| 3120+1G>A | c.2988+1G>A | SNV | Intron18 | 7 | 1 | 0 | 0 | 100 | |
| 3121-1G->A | c.2989-1G>A | SNV | CFSyn_Ex19 | 0 | 0 | 1 | 0 | 100 | |
| L997F | c.2991G>C | SNV | Exon19 | 2 | 1 | 0 | 0 | 100 | |
| I1027T | c.3080T>C | SNV | Exon19 | 1 | 2 | 0 | 0 | 100 | |
| 3272-26A>G | c.3140-26A>G | SNV | Intron19 | 0 | 1 | 0 | 0 | 100 | |
| F1052V | c.3154T>G | SNV | Exon20 | 0 | 1 | 0 | 0 | 100 | |
| L1065P | c.3194T>C | SNV | Exon20 | 0 | 0 | 1 | 0 | 100 | |
| Genotype(CommonNamecDNA namecoordinate) | cDNA name | VariantType | CFTRgeneregion(hg19) | Positive calls (Variants) | PositiveAgreement | ||||
| ClinicalSamples | CellLineSamples | SyntheticSamples | No"Calls" | Miscalls | |||||
| R1066C | c.3196C>T | SNV | Exon20 | 6 | 0 | 0 | 0 | 0 | 100 |
| R1066H | c.3197G>A | SNV | Exon20 | 1 | 0 | 1 | 0 | 0 | 100 |
| G1069R | c.3205G>A | SNV | Exon20 | 0 | 1 | 0 | 0 | 0 | 100 |
| R1070W | c.3208C>T | SNV | Exon20 | 0 | 2 | 0 | 0 | 0 | 100 |
| R1070Q | c.3209G>A | SNV | Exon20 | 0 | 1 | 0 | 0 | 0 | 100 |
| L1077P | c.3230T>C | SNV | Exon20 | 0 | 0 | 1 | 0 | 0^ | 100 |
| W1089X | c.3266G>A | SNV | Exon20 | 4 | 0 | 0 | 0 | 0 | 100 |
| Y1092X (C>A) | c.3276C>A | SNV | Exon20 | 3 | 1 | 0 | 0 | 0 | 100 |
| Y1092X (C>G) | c.3276C>G | SNV | Exon20 | 0 | 0 | 1 | 0 | 0 | 100 |
| T1095T | c.3285A>T | SNV | Exon20 | 7 | 0 | 0 | 0 | 0 | 100 |
| M1101K | c.3302T>A | SNV | Exon20 | 2 | 2 | 0 | 0 | 0 | 100 |
| E1104X | c.3310G>T | SNV | Exon20 | 0 | 0 | 1 | 0 | 0 | 100 |
| c.3368-2A>T | c.3368-2A>T | SNV | Intron20 | 0 | 1 | 0 | 0 | 0 | 100 |
| D1152H | c.3454G>C | SNV | Exon21 | 10 | 1 | 0 | 0 | 0 | 100 |
| V1153E | c.3458T>A | SNV | Exon21 | 1 | 0 | 0 | 0 | 0 | 100 |
| Genotype(CommonNamecDNA namecoordinate) | cDNA name | VariantType | CFTRgeneregion(hg19) | Positive calls (Variants) | PositiveAgreement | ||||
| ClinicalSamples | CellLineSamples | SyntheticSamples | NoCalls | Miscalls | |||||
| R1158X | c.3472C>T | SNV | Exon22 | 7 | 1 | 0 | 0 | 0 | 100 |
| R1162X | c.3484C>T | SNV | Exon22 | 5 | 1 | 0 | 0 | 0 | 100 |
| R1162L | c.3485G>T | SNV | Exon22 | 0 | 2 | 0 | 0 | 0 | 100 |
| 3659delC | c.3528delC | DIV | Exon22 | 4 | 1 | 0 | 0 | 0 | 100 |
| S1196X | c.3587C>G | SNV | Exon22 | 1 | 0 | 0 | 0 | 0 | 100 |
| W1204X(c.3611G>A) | c.3611G>A | SNV | Exon22 | 0 | 0 | 1 | 0 | 0 | 100 |
| W1204X(c.3612G>A) | c.3612G>A | SNV | Exon22 | 0 | 0 | 1 | 0 | 0 | 100 |
| 3791delC | c.3659delC | DIV | Exon22 | 2 | 0 | 0 | 0 | 0 | 100 |
| I1234V | c.3700A>G | SNV | Exon22 | 1 | 0 | 1 | 0 | 0 | 100 |
| S1235R | c.3705T>G | SNV | Exon22 | 9 | 1 | 0 | 0 | 0 | 100 |
| 3849+10kbC>T | c.3717+12191C>T | SNV | Intron22 | 11 | 2 | 0 | 0 | 0 | 100 |
| G1244E | c.3731G>A | SNV | Exon23 | 0 | 0 | 1 | 0 | 0 | 100 |
| 3876delA | c.3744delA | DIV | Exon23 | 6 | 1 | 0 | 0 | 0 | 100 |
| S1251N | c.3752G>A | SNV | Exon23 | 1 | 0 | 1 | 0 | 0 | 100 |
| Genotype(CommonNamecDNA namecoordinate) | cDNA name | VariantType | CFTRgeneregion(hg19) | Positive calls (Variants) | |||||
| ClinicalSamples | CellLineSamples | SyntheticSamples | NoCalls | Miscalls | PositiveAgreement | ||||
| 3905insT | c.3773_3774insT | DIV | Exon23 | 3 | 1 | 0 | 0 | 0 | 100 |
| D1270N | c.3808G>A | SNV | Exon23 | 0 | 2 | 0 | 0 | 0 | 100 |
| W1282X | c.3846G>A | SNV | Exon23 | 9 | 1 | 0 | 0 | 0 | 100 |
| P1290P | c.3870A>G | SNV | Exon23 | 10 | 3 | 0 | 0 | 0 | 100 |
| 4005+1G->A | c.3873+1G>A | SNV | Intron23 | 0 | 0 | 1 | 0 | 0 | 100 |
| 4016insT | c.3884_3885insT | DIV | Exon24 | 0 | 0 | 1 | 0 | 0 | 100 |
| T1299T | c.3897A>G | SNV | Exon24 | 3 | 0 | 0 | 0 | 0 | 100 |
| N1303K | c.3909C>G | SNV | Exon24 | 9 | 1 | 0 | 0 | 0 | 100 |
| Q1313X | c.3937C>T | SNV | Exon24 | 0 | 0 | 1 | 0 | 0 | 100 |
| G1349D | c.4046G>A | SNV | Exon25 | 0 | 1 | 0 | 0 | 0 | 100 |
| 4209TGTT>AA | c.4077_4080delTGTTinsAA | DIV | Exon25 | 0 | 0 | 1 | 0 | 0 | 100 |
| CFTRdele22,23 | c.3964-78_4242+577del | Del | Intron24 | 1 | 0 | 1 | 0 | 0 | 100 |
| 4382delA | c.4251delA | DIV | Exon27 | 0 | 0 | 1 | 0 | 0 | 100 |
| Y1424Y | c.4272C>T | SNV | Exon27 | 6 | 2 | 0 | 0 | 0 | 100 |
| Genotype(CommonNamecoordinate) | cDNA name | VariantType | CFTRgeneregion(hg19) | Positive calls (Variants) | NoSyntheticSamples | Miscalls | PositiveAgreement | ||
| Q1463Q | cDNA namec.4389G>A | SNV | Exon27 | ClinicalSamples150 | CellLineSamples32 | 0 | 0 | 100 | |
| Total All Variants (PA)† | 2072 | 3 | 4 | 99.66 | |||||
| Total All WT (NA) | 2600928 | 1 | 2** | >99.99 | |||||
| Total All WT and Variants (OA) | 2603000 | 4 | 6 | >99.99 |
{17}------------------------------------------------
illumina*
:
..
::
.…"
{18}------------------------------------------------
・・
{19}------------------------------------------------
•
20 Confidential
. + T
1
{20}------------------------------------------------
{21}------------------------------------------------
{22}------------------------------------------------
amples were not retested
nowever indicated that the variant was in fact homozygous and incorrectly reported. MiSeqDx reported the variant as The Sanger report listed the P205S variant as heterozygous for the clinical sample. A review of the Sanger trac omozygous.
One of the discordant results was from the reproducibility study. The PolyTG/PolyT result for the sampl ncordant across all 18 replicates, but discordant with Sanger bi-directional sequencing.
The original synthetic heterozygous specimen was determined to be improperly prepared. When it was subsequen ested after it was re-prepared, using the same plasmid, it would be detected
ole, Sanger synthetic sample heterozygous for exon 8 was reported as heterozygous for the variant CFTR dele22, 23. Fur nvestigation revealed that this result was likely from low level contamination. Additionally, for a second samply ners could not fully detect the variant Q1463Q due to indels both upstream and downstream of the variant PA excluding PolyTG/PolyT calls was 100%
{23}------------------------------------------------
| PolyTGPolyT Genotype | # Clinical Samples | # Cell Line Samples | # Synthetic Samples | # Miscalls | # No Calls^ | % Accuracy |
|---|---|---|---|---|---|---|
| (TG)9(T)7/(TG)11(T)7 | 2 | 0 | 0 | 0 | 1 | 50.00 |
| (TG)9(T)9/(TG)10(T)7 | 1 | 0 | 0 | 0 | 0 | 100 |
| (TG)9(T)9/(TG)11(T)7 | 5 | 1 | 0 | 0 | 0 | 100 |
| (TG)9(T)9/(TG)11(T)9 | 1 | 0 | 0 | 0 | 0 | 100 |
| (TG)10(T)7/(TG)10(T)7 | 25 | 8 | 0 | 0 | 0 | 100 |
| (TG)10(T)7/(TG)10(T)9 | 39 | 16 | 0 | 0 | 0 | 100 |
| GPolyT Genotype | # Clinical Samples | # Cell Line Samples | # Synthetic Samples | # Miscalls | # No Calls^ | % Accuracy |
| 10(T)9/(TG)12(T)7 | 13 | 0 | 0 | 0 | 1 | 92.31 |
| 11(T)5/(TG)11(T)7 | 6 | 0 | 0 | 1 | 0 | 83.33 |
| 11(T)7/(TG)11(T)7 | 52 | 8 | 0 | 0 | 0 | 100 |
| 11(T)7/(TG)11(T)9* | 2 | 1 | 0 | 3 | 0 | 0 |
| 11(T)7/(TG)12(T)5 | 2 | 0 | 0 | 0 | 0 | 100 |
| 11(T)7/(TG)12(T)7 | 37 | 3 | 0 | 0 | 0 | 100 |
| 11(T)9/(TG)12(T)7 | 3 | 0 | 0 | 0 | 0 | 100 |
| 12(T)7/(TG)12(T)7 | 2 | 2 | 0 | 0 | 0 | 100 |
| Total | 448 | 4 | 3 | 98.44 |
.
-.
, "
. ..
: .
: : : :
・・
e 2: PolyTG/PolyT variant accuracy for the MiSeqDx Cystic Fibrosis Clinical Sequencing A
←
onfidentia
ব্য
90.91
I
0
0
I
OL
G)10(T)7/(TG)12(T)
100
0
0
0
9
ש
G)10(T)9/(TG)10(T)
100
0
0
0
0
ഹ
G)10(T)9/(TG)11(T
100
0
0
0
07
92
G)10(T)9/(TG)11(T)
100
0
0
0
0
ಳ
G)10(T)9/(TG)11(T)
100
0
0
0
ਟ
ಕ
G)10(T)9/(TG)12(T
100
0
0
0
0
โ
"G)10(T)7/(TG)12(T)
100
0
0
0
0
ਟ
S)10(T)7/(TG)11(T
100
0
0
0
II
と
G)10(T)7/(TG)11(T
{24}------------------------------------------------
:
amples were not reteste
*One of the discordant results was from the reproducibility study. The PolyT result for the sample
was concordant across all 18 replicates, but discordant with Sanger bi-dire
{25}------------------------------------------------
Reproducibili
o oreators at each it for a total one one on the parel contand a mix of genomic DNA
iphoblastoid cell lines with the CFTT gene a well as ome leukory on energer on and order he reproductibility of the MiSeqDx Cystic Fibrosis Clinical Sequencing Assay was determined through a blir
sing 3 trial sites and 2 operators at each site. Two well characte
nple pass rate, defined as the number of samples passing QC metrics on the first attempt, was 99.7%. All result ed on initial testir he genotype-level PA for all variants including the PolyTwas 99.22% and excluding the PolyTG/PolyT v
ras 99.60%. The NA for all WT was 99.70% and the OA for all reported pos A was 97.83%
{26}------------------------------------------------
MiSeqDx Cystic Fibrosis Clinical Sequencing Assay
illumina:
| Sample | HGVS Name (orLocation if no HGVS) | MutationName | Total Results | Agreeing Calls | Total #(All Sites) | %Agreement | |||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| Per Site | AllSites | Site 1 | Site 2 | Site 3 | NoCalls | Miscalls | |||||
| 1 | c.1408G>A | V470M | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 | |
| 1 | c.1646G>A | S549N | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 | |
| 1 | c.2562T>G | T854T | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 | |
| 2 | c.1408G>A | V470M | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 | |
| 2 | c.1581A>G | E527E | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 | |
| 2 | c.1680-1G>A | 1812-1 G>A | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 | |
| 2 | c.2562T>G | T854T | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 | |
| 2 | c.312delA | 444delA | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 | |
| 2 | c.3870A>G | P1290P | 6 | 18 | 6 | 5 | 6 | 0 | 1 | 94.44 | |
| Sample | HGVS Name (or Location if no HGVS) | Mutation Name | Total ResultsPer Site | Total ResultsAll Sites | Agreeing Calls | Total #(All Sites)No Calls" | Total #(All Sites)Miscalls | %Agreement | |||
| Site 1 | Site 2 | Site 3 | |||||||||
| 2 | c.4389G>A | Q1463Q | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 | |
| 3 | c.1408G>A | V470M | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 | |
| 3 | c.1477C>T | Q493X | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 | |
| 3 | c.1521_1523delCTT | F508del | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 | |
| 3 | c.2562T>G | T854T | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 | |
| 3 | c.4389G>A | Q1463Q | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 | |
| 4 | c.1408G>A | V470M | 6 | 18 | 5 | 6 | 6 | 1 | 0 | 94.44 | |
| 4 | c.1521_1523delCTT | F508del | 6 | 18 | 5 | 6 | 6 | 1 | 0 | 94.44 | |
| 4 | c.2052delA | 2184delA | 6 | 18 | 5 | 6 | 6 | 1 | 0 | 94.44 | |
| 5 | c.1408G>A | V470M | 6 | 18 | 6 | 5 | 6 | 1" | 0 | 94.44 | |
| Sample | HGVS Name (orLocation if no HGVS) | MutationName | Total ResultsPer Site | Total ResultsAllSites | Agreeing Calls | Total #(All Sites)NoCalls | Total #(All Sites)Miscalls | Total #(All Sites)%Agreement | |||
| Site 1 | Site 2 | Site 3 | |||||||||
| 5 | c.224G>A | R75Q | 6 | 18 | 6 | 5 | 6 | 11 | 0 | 94.44 | |
| 5 | c.2562T>G | T854T | 6 | 18 | 6 | 5 | 6 | 11 | 0 | 94.44 | |
| 5 | c.3472C>T | R1158X | 6 | 18 | 6 | 5 | 6 | 11 | 0 | 94.44 | |
| 5 | c.366T>A | Y122X | 6 | 18 | 6 | 5 | 6 | 11 | 0 | 94.44 | |
| 5 | c.625G>T | A209S | 6 | 18 | 6 | 5 | 6 | 11 | 0 | 94.44 | |
| 6 | c.1408G>A | V470M | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 | |
| 6 | c.1521_1523delCTT | F508del | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 | |
| 6 | c.2051_2052delAAinsG | 2183AA>G | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 | |
| 7 | c.1408G>A | V470M | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 | |
| 7 | c.223C>T | R75X | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 | |
| % | greemer | 100 | 100 | 100 | 100 | 100 | 100 | 100 | 100 | 100 | 94.44 |
| (All SitesTotal # | Miscalls | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | i I |
| CallsªNo | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | |
| Site 3 | 9 | 9 | 9 | 9 | 9 | 9 | 9 | 9 | 9 | 9 | |
| greeing Call | Site 2 | 9- | 9 | 9 | 9 | 9 | 9 | 9 | 9 | 9 | g |
| Site 1 | 9 | 9 | 9 | 9 | 9 | 9 | 9 | 9 | 9 | 9 | |
| otal Results | SitesAll | 18 | 81 | 18 | 81 | 81 | 81 | 81 | 81 | 18 | 81 |
| Per Site | !9 | 9 | 9 | 9 | 9 | 9 | 9 | 9 | 9 | 9 | |
| Iutation | Name | 1854T | V470M | 507del | F508de | T854T | Q1463Q | V470M | F508de | T854T | V1282X |
| GVS Name | cation if no HGV: | 2562T>G | c.1408G> | 519_1521delAT | 1521_1523delCT | .2562T>G | .4389G>/ | c.1408G> | 1521_1523delC | 2562T>G | 3846G> |
| Sample | 1 | 8 | 8 | 8 | 8 | 8 | 6 | б | 6 | б | |
| HGVS Name (orLocation if no HGVS) | MutationName | Total Results | Agreeing Calls | Total #(All Sites) | %Agreement | ||||||
| ample | Per Site | AllSites | Site 1 | Site 2 | Site 3 | NoCalls | Miscalls | ||||
| 9 | c.4389G>A | Q1463Q | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 | |
| 10 | c.1408G>A | V470M | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 | |
| 10 | c.1521_1523delCTT | F508del | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 | |
| 10 | c.2562T>G | T854T | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 | |
| 10 | c.3140-26A>G | 3272-26A>G | 6 | 18 | 6 | 5 | 6 | 0 | 1' | 94.44 | |
| 10 | c.4389G>A | Q1463Q | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 | |
| 11, 39 | c.1408G>A | V470M | 12 | 36 | 12 | 12 | 12 | 0 | 0 | 100 | |
| 11, 39 | c.1521_1523delCTT | F508del | 12 | 36 | 12 | 12 | 12 | 0 | 0 | 100 | |
| 11, 39 | c.2002C>T | R668C | 12 | 36 | 12 | 12 | 12 | 0 | 0 | 100 | |
| 11, 39 | c.2562T>G | T854T | 12 | 36 | 12 | 12 | 12 | 0 | 0 | 100 | |
| Sample | HGVS Name (orLocation if no HGVS) | MutationName | Total Results | Agreeing Calls | Total #(All Sites) | %Agreement | |||||
| Per Site | AllSites | Site 1 | Site 2 | Site 3 | NoCalls | Miscalls | |||||
| 11, 39 | c.3717+12191C>T | 3849+10kbC>T | 12 | 36 | 12 | 12 | 12 | 0 | 0 | 100 | |
| 11, 39 | c.4389G>A | Q1463Q | 12 | 36 | 12 | 12 | 12 | 0 | 0 | 100 | |
| 12, 40 | c.1408G>A | V470M | 12 | 36 | 12 | 12 | 12 | 0 | 0 | 100 | |
| 12, 40 | c.2562T>G | T854T | 12 | 36 | 12 | 12 | 12 | 0 | 0 | 100 | |
| 12, 40 | c.2988+1G>A | 3120+1G>A | 12 | 36 | 12 | 12 | 12 | 0 | 0 | 100 | |
| 12, 40 | c.4389G>A | Q1463Q | 12 | 36 | 12 | 12 | 12 | 0 | 0 | 100 | |
| 12, 40 | c.489+1G>T | 621+1G>T | 12 | 36 | 12 | 12 | 12 | 0 | 0 | 100 | |
| 13 | c.1408G>A | V470M | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 | |
| 13 | c.1521_1523delCTT | F508del | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 | |
| 13 | c.178G>T | E60X | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 | |
| Sample | HGVS Name (orLocation if no HGVS) | MutationName | Total Results | Agreeing Calls | Total #(All Sites) | %Agreement | |||||
| Per Site | AllSites | Site 1 | Site 2 | Site 3 | NoCalls" | Miscalls | |||||
| 13 | c.4389G>A | Q1463Q | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 | |
| 14 | c.1408G>A | V470M | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 | |
| 14 | c.1584G>A | E528E | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 | |
| 14 | c.2562T>G | T854T | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 | |
| 14 | c.3302T>A | M1101K | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 | |
| 15 | c.1408G>A | V470M | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 | |
| 15 | c.1584G>A | E528E | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 | |
| 15 | c.2562T>G | T854T | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 | |
| 15 | c.3302T>A | M1101K | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 | |
| 16 | c.1408G>A | V470M | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 | |
| Sample | HGVS Name (or Location if no HGVS) | Mutation Name | Total ResultsPer Site | Total ResultsAll Sites | Agreeing Calls | Total #(All Sites)No Calls | Total #(All Sites)Miscalls | %Agreement | |||
| Site 1 | Site 2 | Site 3 | |||||||||
| 16 | c.1521_1523delCTT | F508del | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 | |
| 16 | c.3080T>C | I1027T | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 | |
| 17, 41 | c.1408G>A | V470M | 12 | 36 | 12 | 12 | 12 | 0 | 0 | 100 | |
| 17, 41 | c.1521_1523delCTT | F508del | 12 | 36 | 12 | 12 | 12 | 0 | 0 | 100 | |
| 17, 41 | c.3528delC | 3659delC | 12 | 36 | 12 | 12 | 12 | 0 | 0 | 100 | |
| 18, 42 | 117120145 | 117120145 | 12 | 36 | 12 | 12 | 12 | 0 | 0 | 100 | |
| 18, 42 | c.1408G>A | V470M | 12 | 36 | 12 | 12 | 12 | 0 | 0 | 100 | |
| 18, 42 | c.1521_1523delCTT | F508del | 12 | 36 | 12 | 12 | 12 | 0 | 0 | 100 | |
| 18, 42 | c.350G>A | R117H | 12 | 36 | 12 | 12 | 12 | 0 | 0 | 100 | |
| 19 | c.1408G>A | V470M | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 | |
| Sample | HGVS Name (orLocation if no HGVS) | MutationName | Total Results | Agreeing Calls | Total #(All Sites) | %Agreement | |||||
| Per Site | AllSites | Site 1 | Site 2 | Site 3 | NoCalls | Miscalls | |||||
| 19 | c.489+1G>T | 621+1G>T | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 | |
| 19 | c.579+1G>T | 711+1G>T | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 | |
| 20, 43 | c.1408G>A | V470M | 12 | 36 | 12 | 12 | 12 | 0 | 0 | 100 | |
| 20, 43 | c.254G>A | G85E | 12 | 36 | 12 | 12 | 12 | 0 | 0 | 100 | |
| 20, 43 | c.489+1G>T | 621+1G>T | 12 | 36 | 12 | 12 | 12 | 0 | 0 | 100 | |
| 21, 44 | c.1364C>A | A455E | 12 | 36 | 12 | 12 | 12 | 0 | 0 | 100 | |
| 21, 44 | c.1408G>A | V470M | 12 | 36 | 12 | 12 | 12 | 0 | 0 | 100 | |
| 21, 44 | c.1521_1523delCTT | F508del | 12 | 36 | 12 | 12 | 12 | 0 | 0 | 100 | |
| 22 | c.1408G>A | V470M | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 | |
| 22 | c.1521_1523delCTT | F508del | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 | |
| Sample | HGVS Name (or Location if no HGVS) | Mutation Name | Total ResultsPer Site | Total ResultsAll Sites | Agreeing CallsSite 1 | Agreeing CallsSite 2 | Agreeing CallsSite 3 | Total #No Calls | Total #Miscalls | Total #(All Sites)% Agreement | |
| 22 | c.1679G>C | R560T | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 | |
| 22 | c.2562T>G | T854T | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 | |
| 22 | c.4389G>A | Q1463Q | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 | |
| 23 | c.1408G>A | V470M | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 | |
| 23 | c.1521_1523delCTT | F508del | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 | |
| 23 | c.3276C>A | Y1092X (C>A) | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 | |
| 24, 45 | c.1408G>A | V470M | 12 | 36 | 12 | 12 | 12 | 0 | 0 | 100 | |
| 24, 45 | c.3909C>G | N1303K | 12 | 36 | 12 | 12 | 12 | 0 | 0 | 100 | |
| 24, 45 | c.4046G>A | G1349D | 12 | 36 | 12 | 12 | 12 | 0 | 0 | 100 | |
| 25 | c.1408G>A | V470M | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 | |
| Total # | |||||||||||
| Sample | IGVS Name | lutation | otal Result | greeing Call | (All Sites | 90 | |||||
| ation if no HGV | Name | Per Site | SitesAll | Site 1 | Site 2 | Site 3 | CallsªNo | Miscalls | greeme | ||
| ਤੀ | .1624G> | G542X | 9 | 18 | 9 | 9 | 9 | 0 | 0 | 001 | |
| 97 | 1712014 | 17120141 | 9 | 18 | 9 | 9 | 9 | 0 | 0 | 100 | |
| 97 | c.1408G>A | /470M | 9 | 18 | 9 | 9 | 9 | 0 | 0 | 100 | |
| 97 | c.1624G> | G542X | 9 | 81 | 9 | 9 | 9 | 0 | 0 | 100 | |
| 27, 46 | .1408G>r | /470M | टा | 98 | ਟਾ | टा | 12 | 0 | 0 | 100 | |
| 27, 46 | c.1652G> | G551D | ਟ I | 9€ | ਟਾ | 12 | 12 | 0 | 0 | 100 | |
| 27, 46 | c.1657C>1 | R553X | ਟ I | 98 | ਟ I | ਟਾ | 12 | 0 | 0 | 100 | |
| 27, 46 | 2562T>G | T854T | ਟ I | 96 | ਟ। | ਟ L | 12 | 0 | 0 | 100 | |
| 27, 46 | :4389G>A | Q1463Q | ਟ I | 9€ | ਟਾ | ਟਾ | 12 | 0 | 0 | 100 | |
| 87 | .1408G> | /470M | 9 | 81 | 9 | 9 | 9 | 0 | 0 | 100 | |
| Sample | HGVS Name (or Location if no HGVS) | Mutation Name | Total ResultsPer Site | Total ResultsAll Sites | Agreeing Calls | Total #(All Sites)No Calls | Total #(All Sites)Miscalls | %Agreement | |||
| 28 | c.2562T>G | T854T | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 | |
| 28 | c.3717+12191C>T | 3849+10kbC>T | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 | |
| 28 | c.4389G>A | Q1463Q | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 | |
| 29 | c.1408G>A | V470M | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 | |
| 29 | c.2562T>G | T854T | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 | |
| 29 | c.91C>T | R31C | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 | |
| 30 | c.1408G>A | V470M | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 | |
| 30 | c.1521_1523delCTT | F508del | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 | |
| 30 | c.2562T>G | T854T | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 | |
| 30 | c.3485G>T | R1162L | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 | |
| Sample | HGVS Name (or Location if no HGVS) | Mutation Name | Total ResultsPer Site | Total ResultsAll Sites | Agreeing Calls | Total #(All Sites)No Calls | Total #(All Sites)Miscalls | % Agreement | |||
| Site 1 | Site 2 | Site 3 | |||||||||
| 30 | c.4389G>A | Q1463Q | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 | |
| 31 | c.1408G>A | V470M | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 | |
| 31 | c.1585-1G>A | 1717-1G>A | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 | |
| 31 | c.2562T>G | T854T | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 | |
| 31 | c.4389G>A | Q1463Q | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 | |
| 32 | c.1408G>A | V470M | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 | |
| 32 | c.2562T>G | T854T | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 | |
| 32 | c.3484C>T | R1162X | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 | |
| 32 | c.4389G>A | Q1463Q | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 | |
| 33 | c.1040G>C | R347P | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 |
27 Confidential
...
... : : . : •
{27}------------------------------------------------
.
{28}------------------------------------------------
:
{29}------------------------------------------------
. .
{30}------------------------------------------------
:
{31}------------------------------------------------
. Li
،
{32}------------------------------------------------
Confidential
{33}------------------------------------------------
Confidential
{34}------------------------------------------------
{35}------------------------------------------------
,;'
{36}------------------------------------------------
{37}------------------------------------------------
38 Confidential
:『
{38}------------------------------------------------
{39}------------------------------------------------
| HGVS Name (orLocation if no HGVS) | MutationName | Total Results | Agreeing Calls | Total #(All Sites) | %Agreement | ||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| ample | Per Site | Site 1 | Site 2 | Site 3 | NoCalls | Miscalls | |||||
| 33 | c.1408G>A | V470M | 6 | 6 | 6 | 6 | 0 | 0 | 100 | ||
| 33 | c.1652G>A | G551D | 6 | 6 | 6 | 6 | 0 | 0 | 100 | ||
| 33 | c.2562T>G | T854T | 6 | 6 | 6 | 6 | 0 | 0 | 100 | ||
| 33 | c.4272C>T | Y1424Y | 6 | 6 | 6 | 6 | 0 | 0 | 100 | ||
| 33 | c.4389G>A | Q1463Q | 6 | 6 | 6 | 6 | 0 | 0 | 100 | ||
| 34 | c.1000C>T | R334W | 6 | 6 | 6 | 6 | 0 | 0 | 100 | ||
| 34 | c.3368-2A>T | c.3368-2A>T | 6 | 6 | 6 | 6 | 0 | 0 | 100 | ||
| 35 | c.1523T>G | F508C | 6 | 6 | 6 | 6 | 0 | 0 | 100 | ||
| 36 | c.254G>A | G85E | 6 | 6 | 6 | 6 | 0 | 0 | 100 | ||
| 36 | c.3454G>C | D1152H | 6 | 6 | 6 | 6 | 0 | 0 | 100 | ||
| Sample | HGVS Name (orLocation if no HGVS) | MutationName | Total ResultsPer Site | Total ResultsAll Sites | Agreeing Calls | Total #(All Sites)No Calls | Total #(All Sites)Miscalls | %Agreement | |||
| Site 1 | Site 2 | Site 3 | |||||||||
| 37 | c.1007T>A | I336K | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 | |
| 37 | c.1408G>A | V470M | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 | |
| 37 | c.2562T>G | T854T | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 | |
| 37 | c.3705T>G | S1235R | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 | |
| 38 | c.1408G>A | V470M | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 | |
| 38 | c.1727G>C | G576A | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 | |
| 38 | c.2002C>T | R668C | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 | |
| 38 | c.2057C>A | S686Y | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 | |
| 38 | c.2562T>G | T854T | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 | |
| 38 | c.4389G>A | Q1463Q | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 | |
| Sample | HGVS Name (orLocation if no HGVS) | MutationName | Total ResultsPer Site | Total ResultsAllSites | Agreeing CallsSite 1 | Agreeing CallsSite 2 | Agreeing CallsSite 3 | Total #(All Sites)NoCalls | Total #(All Sites)Miscalls | %Agreement | |
| 47, 85 | c.1408G>A | V470M | 12 | 36 | 12 | 12 | 12 | 0 | 0 | 100 | |
| 47, 85 | c.2562T>G | T854T | 12 | 36 | 12 | 12 | 12 | 0 | 0 | 100 | |
| 47, 85 | c.2657+5G>A | 2789+5G>A | 12 | 36 | 12 | 12 | 12 | 0 | 0 | 100 | |
| 47, 85 | c.4389G>A | Q1463Q | 12 | 36 | 12 | 12 | 12 | 0 | 0 | 100 | |
| 48, 86 | 5940_273+10250del21kbc.54- | CFTRdele2,3 | 12 | 36 | 12 | 11 | 12 | 1 | 0 | 97.22 | |
| 48, 86 | c.1408G>A | V470M | 12 | 36 | 12 | 11 | 12 | 1 | 0 | 97.22 | |
| 48, 86 | c.1521_1523delCTT | F508del | 12 | 36 | 12 | 11 | 12 | 1 | 0 | 97.22 | |
| 49, 87 | c.1408G>A | V470M | 12 | 36 | 12 | 12 | 12 | 0 | 0 | 100 | |
| 49, 87 | c.1521_1523delCTT | F508del | 12 | 36 | 12 | 12 | 12 | 0 | 0 | 100 | |
| Sample | HGVS Name (or Location if no HGVS) | Mutation Name | Total ResultsPer Site | Total ResultsAll Sites | Agreeing Calls | Total #(All Sites) | %Agreement | ||||
| Site 1 | Site 2 | Site 3 | No Calls | Miscalls | |||||||
| 49, 87 | c.1766+1G>A | 1898+1G>A | 12 | 36 | 12 | 12 | 12 | 0 | 0 | 100 | |
| 50, 88 | c.1408G>A | V470M | 12 | 36 | 12 | 12 | 12 | 0 | 0 | 100 | |
| 50, 88 | c.220C>T | R74W | 12 | 36 | 12 | 12 | 12 | 0 | 0 | 100 | |
| 50, 88 | c.2562T>G | T854T | 12 | 36 | 12 | 12 | 12 | 0 | 0 | 100 | |
| 50, 88 | c.3808G>A | D1270N | 12 | 36 | 12 | 12 | 12 | 0 | 0 | 100 | |
| 51, 89 | c.1408G>A | V470M | 12 | 36 | 12 | 12 | 12 | 0 | 0 | 100 | |
| 51, 89 | c.1521_1523delCTT | F508del | 12 | 36 | 12 | 12 | 12 | 0 | 0 | 100 | |
| 51, 89 | c.2012delT | 2143delT | 12 | 36 | 12 | 12 | 12 | 0 | 0 | 100 | |
| 52 | c.3744delA | 3876delA | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 | |
| 53, 90 | c.3773_3774insT | 3905insT | 12 | 36 | 12 | 12 | 12 | 0 | 0 | 100 | |
| Sample | HGVS Name (orLocation if no HGVS) | MutationName | Total Results | Agreeing Calls | Total #(All Sites) | %Agreement | |||||
| Per Site | AllSites | Site 1 | Site 2 | Site 3 | NoCalls" | Miscalls | |||||
| 54, 91 | c.1408G>A | V470M | 12 | 36 | 12 | 12 | 12 | 0 | 0 | 100 | |
| 54, 91 | c.262_263delTT | 394delTT | 12 | 36 | 12 | 12 | 12 | 0 | 0 | 100 | |
| 55,92 | c.1408G>A | V470M | 12 | 36 | 12 | 12 | 12 | 0 | 0 | 100 | |
| 55,92 | c.1519A>G | I507V | 12 | 36 | 12 | 12 | 12 | 0 | 0 | 100 | |
| 55,92 | c.1521_1523delCTT | F508del | 12 | 36 | 12 | 12 | 12 | 0 | 0 | 100 | |
| 55,92 | c.2562T>G | T854T | 12 | 36 | 12 | 12 | 12 | 0 | 0 | 100 | |
| 55,92 | c.3080T>C | I1027T | 12 | 36 | 12 | 12 | 12 | 0 | 0 | 100 | |
| 55,92 | c.4389G>A | Q1463Q | 12 | 36 | 12 | 12 | 12 | 0 | 0 | 100 | |
| 56 | c.1408G>A | V470M | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 | |
| 56 | c.2562T>G | T854T | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 |
Confidential
{40}------------------------------------------------
.
..
{41}------------------------------------------------
:
:
{42}------------------------------------------------
T
Confidential
{43}------------------------------------------------
{44}------------------------------------------------
| HGVS Name (orLocation if no HGVS) | MutationName | Total Results | Agreeing Calls | Total #(All Sites) | %Agreement | |||||
|---|---|---|---|---|---|---|---|---|---|---|
| Per Site | AllSites | Site 1 | Site 2 | Site 3 | NoCalls | Miscalls | ||||
| c.3154T>G | F1052V | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 | |
| c.4389G>A | Q1463Q | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 | |
| 117120141 | 117120141 | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 | |
| c.1408G>A | V470M | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 | |
| c.2562T>G | T854T | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 | |
| c.3209G>A | R1070Q | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 | |
| c.1408G>A | V470M | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 | |
| c.1521_1523delCTT | F508del | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 | |
| c.2991G>C | L997F | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 | |
| c.1408G>A | V470M | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 | |
| HGVS Name (orLocation if no HGVS) | MutationName | Total ResultsPer Site | Total ResultsAll Sites | Agreeing Calls | Total #(All Sites)No Calls | Total #(All Sites)Miscalls | %Agreement | |||
| Sample | Site 1 | Site 2 | Site 3 | |||||||
| 59 | c.2562T>G | T854T | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 |
| 59 | c.3205G>A | G1069R | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 |
| 60 | c.1408G>A | V470M | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 |
| 60 | c.2562T>G | T854T | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 |
| 60 | c.4389G>A | Q1463Q | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 |
| 60 | c.617T>G | L206W | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 |
| 61 | c.1408G>A | V470M | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 |
| 61 | c.2260G>A | V754M | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 |
| 61 | c.4389G>A | Q1463Q | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 |
| 62 | c.1408G>A | V470M | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 |
{45}------------------------------------------------
{46}------------------------------------------------
| Sample | HGVS Name (or Location if no HGVS) | Mutation Name | Total Results Per Site | Total Results All Sites | Agreeing Calls | Total # (All Sites) No Calls | Total # (All Sites) Miscalls | % Agreement | ||
|---|---|---|---|---|---|---|---|---|---|---|
| Site 1 | Site 2 | Site 3 | ||||||||
| 62 | c.2562T>G | T854T | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 |
| 62 | c.988G>T | G330X | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 |
| 64 | c.1040G>A | R347H | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 |
| 64 | c.1408G>A | V470M | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 |
| 64 | c.2562T>G | T854T | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 |
| 64 | c.4389G>A | Q1463Q | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 |
| 65 | c.948delT | 1078delT | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 |
| 66 | c.1408G>A | V470M | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 |
| 66 | c.1521_1523delCTT | F508del | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 |
| 66 | c.532G>A | G178R | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 |
{47}------------------------------------------------
| Sample | HGVS Name (or Location if no HGVS) | Mutation Name | Total Results Per Site | Total Results All Sites | Agreeing Calls Site 1 | Agreeing Calls Site 2 | Agreeing Calls Site 3 | Total # (All Sites) No Calls" | Total # (All Sites) Miscalls | % Agreement |
|---|---|---|---|---|---|---|---|---|---|---|
| 67 | c.1408G>A | V470M | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 |
| 67 | c.1647T>G | S549R (c.1647T>G) | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 |
| 68 | c.1408G>A | V470M | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 |
| 68 | c.1646G>A | S549N | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 |
| 68 | c.2562T>G | T854T | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 |
| 68 | c.4389G>A | Q1463Q | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 |
| 69 | c.2506G>T | D836Y | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 |
| 69 | c.2537G>A | W846X | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 |
| 70 | c.1408G>A | V470M | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 |
{48}------------------------------------------------
| Sample | HGVS Name (or Location if no HGVS) | Mutation Name | Total ResultsPer Site | Total ResultsAll Sites | Agreeing Calls | Total #(All Sites)No Calls | Total #(All Sites)Miscalls | %Agreement | ||
|---|---|---|---|---|---|---|---|---|---|---|
| 70 | c.2562T>G | T854T | 6 | 18 | Site 16 | Site 26 | Site 36 | 0 | 0 | 100 |
| 70 | c.3485G>T | R1162L | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 |
| 70 | c.4389G>A | Q1463Q | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 |
| 71 | c.1408G>A | V470M | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 |
| 71 | c.1521_1523delCTT | F508del | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 |
| 71 | c.2562T>G | T854T | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 |
| 71 | c.274G>T | E92X | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 |
| 71 | c.4389G>A | Q1463Q | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 |
| 72 | c.1022_1023insTC | 1154insTC | 6 | 18 | 6 | 6 | 5 | 1 | 0 | 94.44 |
| 72 | c.1408G>A | V470M | 6 | 18 | 6 | 6 | 5 | 1 | 0 | 94.44 |
Confidential
.
{49}------------------------------------------------
| Sample | HGVS Name (orLocation if no HGVS) | MutationName | Total ResultsPer Site | Total ResultsAll Sites | Agreeing Calls | Total #(All Sites) | %Agreement | |||
|---|---|---|---|---|---|---|---|---|---|---|
| Site 1 | Site 2 | Site 3 | NoCalls | Miscalls | ||||||
| 72 | c.2562T>G | T854T | 6 | 18 | 6 | 6 | 5 | 1 | 0 | 94.44 |
| 72 | c.4389G>A | Q1463Q | 6 | 18 | 6 | 6 | 5 | 1 | 0 | 94.44 |
| 72 | c.489+1G>T | 621+1G>T | 6 | 18 | 6 | 6 | 5 | 1 | 0 | 94.44 |
| 73 | c.1408G>A | V470M | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 |
| 73 | c.1624G>T | G542X | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 |
| 73 | c.1826A>G | H609R | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 |
| 74 | c.1408G>A | V470M | 6 | 18 | 6 | 6 | 5 | 0 | 1 | 94.44 |
| 74 | c.1429C>T | P477S | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 |
| 74 | c.1521_1523delCTT | F508del | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 |
| 75 | c.1408G>A | V470M | 6 | 18 | 6 | 5 | 6 | 11 | 0 | 94.44 |
Confidential
{50}------------------------------------------------
| Sample | HGVS Name (orLocation if no HGVS) | MutationName | Total ResultsPer Site | Total ResultsAll Sites | Agreeing Calls | Total #(All Sites)NoCalls | Total #(All Sites)Miscalls | %Agreement | ||
|---|---|---|---|---|---|---|---|---|---|---|
| Site 1 | Site 2 | Site 3 | ||||||||
| 75 | c.1521_1523delCTT | F508del | 6 | 18 | 6 | 5 | 6 | 1a | 0 | 94.44 |
| 75 | c.1721C>A | P574H | 6 | 18 | 6 | 5 | 6 | 1a | 0 | 94.44 |
| 76 | c.1408G>A | V470M | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 |
| 76 | c.1521_1523delCTT | F508del | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 |
| 76 | c.2562T>G | T854T | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 |
| 76 | c.425delT | F143LfsX10 | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 |
| 76 | c.4389G>A | Q1463Q | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 |
| 77 | c.1364C>A | A455E | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 |
| 77 | c.1408G>A | V470M | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 |
| 77 | c.489+1G>T | 621+1G>T | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 |
Confidential
{51}------------------------------------------------
| Sample | HGVS Name (orLocation if no HGVS) | MutationName | Total Results | Agreeing Calls | Total #(All Sites) | %Agreement | ||||
|---|---|---|---|---|---|---|---|---|---|---|
| Per Site | Site 1 | Site 2 | Site 3 | AllSites | NoCalls | Miscalls | ||||
| 78 | c.1408G>A | V470M | 6 | 6 | 6 | 6 | 18 | 0 | 0 | 100 |
| 78 | c.1581A>G | E527E | 6 | 6 | 6 | 6 | 18 | 0 | 0 | 100 |
| 78 | c.1680-1G>A | 1812-1 G>A | 6 | 6 | 6 | 6 | 18 | 0 | 0 | 100 |
| 78 | c.2562T>G | T854T | 6 | 6 | 6 | 6 | 18 | 0 | 0 | 100 |
| 78 | c.312delA | 444delA | 6 | 6 | 6 | 6 | 18 | 0 | 0 | 100 |
| 78 | c.3870A>G | P1290P | 6 | 6 | 6 | 6 | 18 | 0 | 0 | 100 |
| 78 | c.4389G>A | Q1463Q | 6 | 6 | 6 | 6 | 18 | 0 | 0 | 100 |
| 79 | c.1408G>A | V470M | 6 | 6 | 6 | 6 | 18 | 0 | 0 | 100 |
| 79 | c.220C>T | R74W | 6 | 6 | 6 | 6 | 18 | 0 | 0 | 100 |
| 79 | c.2562T>G | T854T | 6 | 6 | 6 | 6 | 18 | 0 | 0 | 100 |
Confidential
{52}------------------------------------------------
| Sample | HGVS Name (orLocation if no HGVS) | MutationName | Total Results | Agreeing Calls | Total #(All Sites) | %Agreement | ||||
|---|---|---|---|---|---|---|---|---|---|---|
| Per Site | AllSites | Site 1 | Site 2 | Site 3 | NoCalls" | Miscalls | ||||
| 79 | c.3808G>A | D1270N | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 |
| 80 | 117120141 | 117120141 | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 |
| 80 | c.1408G>A | V470M | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 |
| 80 | c.1521_1523delCTT | F508del | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 |
| 80 | c.1657C>T | R553X | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 |
| 80 | c.2562T>G | T854T | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 |
| 81 | c.1408G>A | V470M | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 |
| 81 | c.1521_1523delCTT | F508del | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 |
| 81 | c.1652G>A | G551D | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 |
| 81 | c.2562T>G | T854T | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 |
{53}------------------------------------------------
| Sample | HGVS Name (or Location if no HGVS) | Mutation Name | Total ResultsPer Site | Total ResultsAll Sites | Agreeing Calls | Total #(All Sites)No Calls" | Total #(All Sites)Miscalls | Total #(All Sites)% Agreement | ||
|---|---|---|---|---|---|---|---|---|---|---|
| Site 1 | Site 2 | Site 3 | ||||||||
| 81 | c.4389G>A | Q1463Q | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 |
| 82 | c.1040G>C | R347P | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 |
| 82 | c.1408G>A | V470M | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 |
| 82 | c.1521_1523delCTT | F508del | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 |
| 82 | c.4272C>T | Y1424Y | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 |
| 83 | 117120145 | 117120145 | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 |
| 83 | c.1408G>A | V470M | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 |
| 83 | c.1521_1523delCTT | F508del | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 |
| 83 | c.350G>A | R117H | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 |
| 84 | c.1408G>A | V470M | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 |
{54}------------------------------------------------
| HGVS Name (orLocation if no HGVS) | MutationName | Total ResultsPer Site | Total ResultsAll Sites | Agreeing Calls | Total #(All Sites)NoCalls | Total #(All Sites)Miscalls | Total #(All Sites)%Agreement | |||
|---|---|---|---|---|---|---|---|---|---|---|
| Sample | Site 1 | Site 2 | Site 3 | |||||||
| 84 | c.1519_1521delATC | I507del | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 |
| 84 | c.2562T>G | T854T | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 |
| 84 | c.4389G>A | Q1463Q | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 |
| Total All Variants (PA)* (including thePolyTGPolyT data in Table 4) | 2580 | 7740 | 2562 | 2553 | 2565 | 37 | 23 | 99.22 | ||
| Total All WT (NA) | 2871132 | 8613396 | 2865930 | 2855526 | 2865932 | 26006 | 2 | 99.70 | ||
| Total All WT and variants (OA) | 2873712 | 8621136 | 2868492 | 2858079 | 2868497 | 26043 | 25 | 99.70 |
Samples were not retest
¶ One replicate each of samples 5 and 75 had a 0% call rate. Further investigation indicated that the samples had likely no een added to the sample plate prior to library preparation
Upon review, samples 9 and 10 were likely switched by the operator prior to library preparati
{55}------------------------------------------------
.
Excluding PolyTG/PolyT variants, the PA was 99.60%
...
:
.
—
{56}------------------------------------------------
| SSAV |
|---|
| 41 |
| Seatlencing |
| t |
| Clinical Sequencing194 |
| TOTOSIS : |
| 11 9 8 8 8 0 8 0 8 0 8 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 |
| 99.89ﺑ1 |
| UC |
| l- |
| ﭘ11 |
| M1400. |
| CHE ITELUCULA |
| . |
| ﻟﻠﻘﺎﻧﻮﻥ ﺍﻟﻤﺴﺎﺣﺔ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘ |
| Kontod1101h111711111111" In VIVUHUUUIII11to the |
| Parcel1. In course for a |
| ﻟTODA |
| ﻟﻤﺴﺎﺣﺔ ﺍﻟﻤﺴﺎﺣﺔ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮ1 |
| ﻟStﺎ ﺍﻟﻤﺴﺎﺣﺔ ﺍﻟﻤﺴﺎﺣﺔ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤ |
| 4.4: |
| Table |
.
.
·
.
.
. ·
:
···
.
| Panel | Sample | Genotype | TotalResults | Agreeing Calls | Total #(All Sites) | %Agreement | ||||
|---|---|---|---|---|---|---|---|---|---|---|
| PerSite | Sites | Site1 | Site2 | Site3 | NoCalls | Miscalls | ||||
| A | 1 | (TG)12(T)7/(TG)12(T)7 | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 |
| A | 2 | (TG)10(T)9/(TG)10(T)7 | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 |
| A | 3 | (TG)10(T)7/(TG)10(T)9 | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 |
| A | 4 | (TG)10(T)9/(TG)11(T)7 | 6 | 18 | 5 | 6 | 6 | 1 | 0 | 94.44 |
| A | 5 | (TG)10(T)7/(TG)11(T)7 | 6 | 18 | 6 | 5 | 6 | 1 | 0 | 94.44 |
| A | 6 | (TG)10(T)9/(TG)10(T)7 | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 |
| A | 7 | (TG)10(T)9/(TG)11(T)7 | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 |
| A | 8 | (TG)10(T)7/(TG)10(T)9 | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 |
| A | 9 | (TG)10(T)9/(TG)10(T)7 | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 |
| A | 10 | (TG)10(T)9/(TG)10(T)7 | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 |
| TotalResults | Agreeing Calls | Total #(All Sites) | ||||||||
| A | 11, 39 | (TG)10(T)9/(TG)10(T)7 | 12 | 36 | 12 | 12 | 12 | 0 | 0 | 100 |
| A | 12, 40 | (TG)10(T)9/(TG)11(T)7 | 12 | 36 | 12 | 12 | 12 | 0 | 0 | 100 |
| A | 13 | (TG)10(T)9/(TG)11(T)7 | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 |
| A | 14 | (TG)10(T)7/(TG)11(T)7 | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 |
| A | 15 | (TG)10(T)7/(TG)11(T)7 | 6 | 18 | 6 | 5 | 6 | 1 | 0 | 94.44 |
| A | 16 | (TG)10(T)9/(TG)10(T)9 | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 |
| A | 17, 41 | (TG)10(T)9/(TG)11(T)7 | 12 | 36 | 12 | 12 | 12 | 0 | 0 | 100 |
| A | 18, 42 | (TG)10(T)9/(TG)12(T)5 | 12 | 36 | 12 | 12 | 12 | 0 | 0 | 100 |
| A | 19 | (TG)10(T)9/(TG)11(T)7 | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 |
| A | 20, 43 | (TG)10(T)9/(TG)11(T)7 | 12 | 36 | 12 | 12 | 12 | 0 | 0 | 100 |
| A | 21, 44 | (TG)10(T)9/(TG)10(T)9 | 12 | 36 | 12 | 12 | 12 | 0 | 0 | 100 |
| A | 22 | (TG)10(T)9/(TG)10(T)7 | 6 | 18 | 6 | 6 | 6 | 0 | 0 | 100 |
{57}------------------------------------------------
ﺮ ﺗ
··
{58}------------------------------------------------
| TotalResults | Agreeing Calls | Total #(All Sites) | ||||
|---|---|---|---|---|---|---|
| A | 23 | (TG)10(T)9/(TG)11(T)7 | 6 | 6 | 0 | 100 |
| A | 24, 45 | (TG)10(T)9/(TG)11(T)7 | 12 | 12 | 0 | 100 |
| A | 25 | (TG)10(T)9/(TG)10(T)9 | 6 | 6 | 0 | 100 |
| A | 26 | (TG)10(T)9/(TG)11(T)7 | 6 | 6 | 0 | 100 |
| A | 27, 46 | (TG)10(T)7/(TG)11(T)7 | 11 | 12 | 1 | 97.22 |
| A | 28 | (TG)10(T)7/(TG)10(T)7 | 6 | 6 | 0 | 100 |
| A | 29 | (TG)10(T)7/(TG)12(T)7 | 4 | 4 | 0 | 77.77 |
| A | 30 | :(TG)10(T)9/(TG)10(T)7 | 6 | 6 | 0 | 100 |
| A | 31 | (TG)10(T)7/(TG)11(T)7 | 6 | 6 | 0 | 100 |
| A | 32 | (TG)10(T)7/(TG)10(T)7 | 6 | 6 | 0 | 100 |
| A | 33 | (TG)10(T)7/(TG)11(T)7 | 5 | 6 | 0 | 94.44 |
| A | 34 | (TG)11(T)7/(TG)12(T)7 | 6 | 6 | 0 | 100 |
{59}------------------------------------------------
. .
| TotalResults | Agreeing Calls | Total #(All Sites) | |
|---|---|---|---|
| 35(TG)11(T)7/(TG)11(T)7 | 18 | 6 | 100 |
| 36(TG)11(T)7/(TG)11(T)7 | 18 | 6 | 100 |
| 37(TG)11(T)7/(TG)12(T)7 | 18 | 6 | 100 |
| 38(TG)10(T)7/(TG)11(T)7 | 18 | 6 | 100 |
| 47, 85(TG)10(T)7/(TG)10(T)7 | 36 | 12 | 100 |
| 48, 86(TG)10(T)9/(TG)11(T)7 | 36 | 11 | 94.44 |
| 49, 87(TG)10(T)9/(TG)11(T)7 | 36 | 12 | 100 |
| 50, 88(TG)10(T)9/(TG)11(T)7 | 36 | 12 | 100 |
| 51, 89(TG)10(T)9/(TG)10(T)9 | 36 | 12 | 100 |
| 52(TG)11(T)7/(TG)11(T)7 | 18 | 6 | 100 |
| 53, 90(TG)11(T)7/(TG)11(T)7 | 36 | 12 | 100 |
| 91, 54(TG)10(T)9/(TG)11(T)7 | 36 | 12 | 100 |
t
{60}------------------------------------------------
| TotalResults | Agreeing Calls | Total #(All Sites) | ||||||
|---|---|---|---|---|---|---|---|---|
| B | 92, 55 | (TG)10(T)9/(TG)10(T)7 | 12 | 12 | 12 | 0 | 0 | 100 |
| B | 56 | (TG)10(T)7/(TG)10(T)9 | 6 | 6 | 6 | 0 | 0 | 100 |
| B | 57 | (TG)12(T)7/(TG)12(T)7 | 6 | 6 | 6 | 0 | 0 | 100 |
| B | 58 | (TG)10(T)9/(TG)10(T)9 | 6 | 6 | 6 | 0 | 0 | 100 |
| B | 59 | (TG)11(T)7/(TG)12(T)7 | 6 | 5 | 6 | 1 | 0 | 94.44 |
| B | 60 | (TG)9(T)9/(TG)11(T)7 | 6 | 6 | 6 | 0 | 0 | 100 |
| B | 61 | (TG)10(T)9/(TG)11(T)7 | 6 | 6 | 6 | 0 | 0 | 100 |
| B | 62 | (TG)10(T)7/(TG)11(T)7 | 6 | 5 | 6 | 1 | 0 | 94.44 |
| B | 63 | (TG)11(T)7/(TG)11(T)7 | 6 | 6 | 6 | 0 | 0 | 100 |
| B | 64 | (TG)10(T)7/(TG)11(T)7 | 6 | 5 | 6 | 1 | 0 | 94.44 |
| B | 65 | (TG)11(T)7/(TG)11(T)7 | 6 | 6 | 6 | 0 | 0 | 100 |
| B | 66 | (TG)10(T)9/(TG)11(T)7 | 6 | 6 | 6 | 0 | 0 | 100 |
| TotalResults | Agreeing Calls | Total #(All Sites) | ||||||
| B | 67 | (TG)11(T)7/(TG)11(T)7 | 6 | 6 | 6 | 0 | 0 | 100 |
| B | 68 | (TG)10(T)7/(TG)11(T)7 | 6 | 6 | 6 | 0 | 0 | 100 |
| B | 69 | (TG)11(T)7/(TG)11(T)7 | 6 | 6 | 6 | 0 | 0 | 100 |
| B | 70 | (TG)10(T)7/(TG)10(T)7 | 6 | 6 | 6 | 0 | 0 | 100 |
| B | 71 | (TG)10(T)9/(TG)11(T)7 | 6 | 6 | 6 | 0 | 0 | 100 |
| B | 72 | (TG)10(T)7/(TG)10(T)9 | 6 | 5 | 6 | 5 | 2 | 88.88 |
| B | 73 | (TG)10(T)9/(TG)11(T)7 | 6 | 6 | 6 | 6 | 0 | 100 |
| B | 74 | (TG)10(T)9/(TG)11(T)7 | 6 | 6 | 6 | 6 | 0 | 100 |
| B | 75 | (TG)10(T)7/(TG)10(T)9 | 6 | 6 | 5 | 6 | 1 | 94.44 |
| B | 76 | (TG)10(T)7/(TG)10(T)9 | 6 | 6 | 6 | 6 | 0 | 100 |
| B | 77 | (TG)10(T)9/(TG)10(T)9 | 6 | 6 | 6 | 6 | 0 | 100 |
| B | 78 | (TG)10(T)7/(TG)10(T)9 | 6 | 5 | 6 | 6 | 1 | 94.44 |
| TotalResults | Agreeing Calls | Total #(All Sites) | ||||||
| (TG)10(T)7/(TG)11(T)7 | 18 | 6 | 0 | 100 | ||||
| (TG)11(T)7/(TG)11(T)9 | 18 | 0 | 18 | 0 | ||||
| (TG)10(T)7/(TG)10(T)9 | 18 | 6 | 0 | 100 | ||||
| (TG)10(T)9/(TG)11(T)7 | 18 | 6 | 0 | 100 | ||||
| (TG)10(T)9/(TG)12(T)5 | 18 | 6 | 0 | 100 | ||||
| (TG)10(T)7/(TG)10(T)7 | 18 | 6 | 0 | 100 | ||||
| olyT Variants (PA) | 1656 | 537 | 17 | 97.83 |
... |
: .
{61}------------------------------------------------
{62}------------------------------------------------
·
.
, ·
..
.. -
.
·
··· . ...
・
ﺘ
·
.
{63}------------------------------------------------
DNA Extraction
Three commonly used, commercially available extraction methods representing magnetic bead extraction, alcohol precipitation and silica filter column isolation methods, were evaluated using K2EDTA anti-coagulated whole blood. A total of 14 blood samples were used during the study; two were wild type, while the remaining samples carried unique genotypes representing 9 different variants, including both common and rare variants. For the polyTG/polyT variation, samples with (T)5-9 and (TG)10-12 were included. The three DNA extraction methods were tested independently by 2 different operators who each performed 3 runs per extraction method. Each extraction was performed by each operator on different days. The DNA concentration and A260/A280 ratio of the extracted gDNA samples was determined using spectrophotometry. The total sample size for each extraction method in this study was 168 (14 samples x 2 operators/extraction method x 3 runs/operator x 2 replicates/extracted gDNA sample).
| ExtractionMethod | Number ofsamples tested | CallRate | Accuracy | Sample FirstPass Rate* |
|---|---|---|---|---|
| AlcoholPrecipitation | 168 | >99.99% | >99.99% | 100% |
| Silica FilterColumn Isolation | 168 | >99.99% | >99.99% | 100% |
| Magnetic BeadExtraction | 168 | >99.99% | >99.99% | 100% |
*Percent of samples having call rate of >99% in first run.
DNA input
The DNA input range of the Illumina MiSeqDx Cystic Fibrosis Clinical Sequencing Assay was evaluated by performing a serial dilution study using 14 representative DNA samples containing 16 unique CF variants. Each sample was tested in duplicate at 9 DNA input levels ranging from 1250 ng to 1 ng (1250 ng, 500 ng, 250 ng, 100 ng, 50 ng, 25 ng, 10 ng, 5 ng, and 1 ng). For determination of accuracy, sample genotypes were compared to bidirectional Sanger sequencing
{64}------------------------------------------------
data and the deletions were compared to PCR assay. 1250 ng and 25 ng were identified as the upper and lower bound for DNA input respectively as they had ≥95% sample first pass rate with no incorrect calls (100% accuracy and call rate).
DNA inputs of 1250 ng, 250 ng, and 100 ng were further tested with 4 representative DNA samples and at least 20 replicates per DNA input level for each sample (n= 4 x 20=80 samples), while the lower bound of 25 ng was tested with 14 samples, 20 replicates for each sample (n=14 x 20=280 samples). The accuracy and sample first pass rate was 100% at all DNA input levels.
Interfering Substances
To assess the impact of interfering substances on the Illumina MiSeqDx Cystic Fibrosis System, the performance of the assay was evaluated in the presence and absence of potential interferents. Sixteen whole blood specimens having unique genotypes were included in the study. Four endogenous interfering substances (bilirubin, cholesterol, hemoglobin, and triglycerides) were tested by spiking them into blood specimens prior to DNA extraction. The concentration limits for each substance is shown in the table below. Additionally, to assess interference resulting from blood collection (short draw), EDTA was spiked into blood samples, and to assess interference resulting from sample preparation, the final wash buffer from a silica filter column isolation method was added to purified genomic DNA.
The MiSeqDx Cystic Fibrosis Clinical Sequencing Assay achieved 100% call rate for all samples tested, and 100% reproducibility in genotype calls between samples in the presence and absence of interfering substances. No interference was observed for any of the potential interferents.
To assess the impact of multiplexing index primer interference, a cross contamination study using two samples, each with unique homozygous genotypes at 4 different genomic positions, and two respective index primers was performed. NO change in variant calling was observed with contamination levels <40%. The sample genotype became heterozygous when contamination levels were ≥40%. ***
Confidential રેર
{65}------------------------------------------------
:
.
| Test Substance | Concentrationtested in blood(upper limit) | Concentration tested inblood (lower limit) |
|---|---|---|
| Bilirubin | 684 µmol/L | 137 µmol/L |
| Cholesterol | 13 mmol/L | 2.6mmol/L |
| Hemoglobin | 2g/L | 0.4 g/L |
| EDTA | 7.0 mg/mL | 2.8mg/mL |
| Triglycerides | 37 mmol/L | 7.4 mmol/L |
:
、
100 - 100 - 100 -
{66}------------------------------------------------
MiSeqDx Cystic Fibrosis Clinical Sequencing Assay
:
.
: : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : :
.
... . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ,:
. :
.
.
:
. . . . . . .
Confidential 67
.
{67}------------------------------------------------
Image /page/67/Picture/1 description: The image shows a circular seal or logo. The seal contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter of the circle. In the center of the seal is a stylized emblem featuring three curved lines or shapes, possibly representing a symbol or abstract design.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-Go(b) Silver Spring, MD 20993-00002
November 19, 2013
ILLUMINA, INC. BRYAN SCHNEIDER ASSOCIATE DIRECTOR, REGULATORY AFFAIRS 5200 ILLUMINA WAY SAN DIEGO. CA 92122
Re: K132750
Trade/Device Name: Illumina MiSeqDx Cystic Fibrosis Clinical Sequencing Assay Regulation Number: 21 CFR 866.5900 Regulation Name: Cystic fibrosis transmembrane conductance regulator (CFTR) gene mutation detection system Regulatory Class: II Product Code: PFS Dated: August 30, 2013 Received: September 3, 2013
Dear Mr. Schneider:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or 10 devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into cither class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act s requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable. the
{68}------------------------------------------------
Page 2-Mr. Schneider
electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/RcsourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Reena Philip -S
for
Maria M. Chan, Ph.D. Director Division of Immunology and Hematology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{69}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: December 31, 2013 See PRA Statement on last page.
510(k) Number (if known)
Device Name
Illumina MiSeqDx(TM) Cystic Fibrosis Clinical Sequencing Assay
Indications for Use (Describe)
The Illumina MiSeqDx(TM) Cystic Fibrosis Clinical Sequencing Assay is a targeted sequencing in vitro diagnostic system that re-sequences the protein coding regions and intrones of the Cystic Fibrosis Transmembrane Conductance Regulator (CFTR) gene in genomic DNA isolated from human periphers collected in K2EDTA. The test detects single nucleotide variants, and small InDels within the region sequenced, and additionally reports on two deep intronic mutations and two large deletions. The test is intended to be used on the Illumina MiSeqDx Instrument.
The test is intended to be used as an aid in the diagnosis of individuals with suspected cystic fibrosis (CF). The test is most appropriate when the patient has an atypical or non-classic presentation of CF or when other mutation panels have failed to identify both causative mutations. The results of the test are intented by a board-certified clinical molecular geneticist or equivalent and should be used in conjunction with other available information including clinical symptoms, other diagnostic tess, and family history. This test is not indicated for use for stand-alone diagnostic testing, for pre-implantation testing, carrier screening, newborn screening, or population screening.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
- November 19, 2013
Image /page/69/Picture/17 description: The image shows the name "Donna M. Roscoe" in bold, black font. The text is slightly distorted, with some pixelation and blurring, particularly around the letters. The background is plain white, providing a clear contrast to the text.
§ 866.5900 Cystic fibrosis transmembrane conductance regulator (CFTR) gene mutation detection system.
(a)
Identification. The CFTR gene mutation detection system is a device used to simultaneously detect and identify a panel of mutations and variants in the CFTR gene. It is intended as an aid in confirmatory diagnostic testing of individuals with suspected cystic fibrosis (CF), carrier identification, and newborn screening. This device is not intended for stand-alone diagnostic purposes, prenatal diagnostic, pre-implantation, or population screening.(b)
Classification. Class II (special controls). The special control is FDA's guidance document entitled “Class II Special Controls Guidance Document: CFTR Gene Mutation Detection System.” See § 866.1(e) for the availability of this guidance document.