K Number
K132701
Manufacturer
Date Cleared
2013-09-26

(28 days)

Product Code
Regulation Number
870.1250
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Primi™ Support Catheter is intended to be used in conjunction with guidewires in order to access discreet regions of the peripheral vasculature. It may be used to facilitate placement and exchange of guidewires and other interventional devices. The support catheter also may be used to selectively infuse/deliver physician specified diagnostic and/or therapeutic agents, such as sterile saline, contrast agents, vaso-dilators, anti-blastics, lytics, and anti-thrombotics, with or without a guidewire in position.

Device Description

The ThermopeutiX Support Catheter that is the submission is a single lumen catheter with a radiopaque marker that aids in positioning the catheter and a tapered tip to facilitate entry. The catheter utilizes an integral spine wire. The catheter is available in several diameters and lengths which are 0.014", 0.018", or 0.035" guidewire compatible. The working lengths are 90 cm thru 170 cm. Reference Table 1 for model number availability. The distal portion of the catheter shaft has a hydrophilic coating to facilitate insertion. The device has a luer hub connector to allow infusion of diagnostic and/or therapeutic agents. The device is supplied sterile and is intended for single use only. Reference Table 2 below for physical and performance characteristics.

AI/ML Overview

The provided text describes a 510(k) premarket notification for the ThermopeutiX Primi™ Support Catheter with an advanced hydrophilic coating. This submission aims to demonstrate substantial equivalence to a previously cleared version of the same catheter (K130850) by ThermopeutiX, Inc., with the primary difference being the advanced hydrophilic coating.

Here's an analysis of the acceptance criteria and study information:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria are implicitly defined by the physical and performance characteristics of the predicate device and the new device. The study demonstrates that the new device's performance is either identical or comparable, and meets internal protocol requirements and product specifications.

Element/FeatureAcceptance Criteria (from Predicate Device)Reported Device Performance (New Device)
Working Length (s)90 cm – 170 cm; polymer shaft with integral spine wire90 cm – 170 cm; polymer shaft with integral spine wire
Outside Diameter (s)Distal:0.045" – 0.064"Proximal: 0.060" – 0.080"Distal:0.045" – 0.064"Proximal: 0.060" – 0.080"
LumenOneOne
Guide wire Compatibility0.014"-0.035"0.014"-0.035"
Guide wire UseOver the WireOver the Wire
Distal End ConfigurationTapered tip for ease of crossingTapered tip for ease of crossing
Fluid connectionLuer hub connectorLuer hub connector
Flow rate @ 150 psi w/Saline0.014" wire models > 0.80 cc/sec0.035" wire models > 0.50 cc/sec0.014" wire models > 0.80 cc/sec0.035" wire models > 0.50 cc/sec
Flow rate @ 150 psi w/Contrast0.014" wire models > 0.30 cc/sec0.035" wire models > 0.15 cc/sec0.014" wire models > 0.30 cc/sec0.035" wire models > 0.15 cc/sec
Maximum Injection PressureDo not exceed 300 psiDo not exceed 300 psi
CoatingYes; hydrophilic, distal portionYes; advanced hydrophilic, distal portion (provides comparable lubricity with a reduction in particulates)
Coating Length65 cm, 115 cm or 140 cm65 cm, 115 cm or 140 cm
Coating ThicknessProprietarySame as predicate (considered equivalent)
Radiopaque MarkersOneOne
Sheath Compatibility5 Fr and 6 Fr5 Fr and 6 Fr
Materials UsedIdentical list of materials as the new device, except for the coating.Hub-PebaxOuter Shaft-PebaxInner Shaft-HDPEDistal Shaft-Pebax w/radiopaque fillerCore Wire-Stainless Steel tapered wireCoating-Advanced hydrophilic proprietary coatingMarker Band Platinum/Iridium

Study Proving Device Meets Acceptance Criteria:

The study conducted is a non-clinical performance and safety study comparing the new device with the advanced hydrophilic coating to the predicate device.

2. Sample Size Used for the Test Set and Data Provenance

The document does not specify an exact numerical sample size for each individual test. It states "The non-clinical tests performed on sample devices are listed below," and "All performance testing was repeated on samples with a minimum of 10 month aging." This indicates that multiple samples were used for testing, including aged samples, but the exact number isn't quantified.

  • Sample Size for Test Set: Not explicitly stated as a number, but "sample devices" were used, including aged samples.
  • Data Provenance: The data is from in-vitro (bench) testing performed by the manufacturer, ThermopeutiX, Inc., in the United States. This is prospective testing performed for the purpose of this 510(k) submission.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

This type of submission (510(k) for a medical device with minor modification) typically relies on engineering and biological testing against established standards and product specifications, rather than expert interpretation of a test set in the way AI/ML algorithms use ground truth. Therefore, the concept of "experts establishing ground truth for the test set" is not applicable here. The "ground truth" is the objective physical and performance measurements determined by standardized testing protocols.

4. Adjudication Method for the Test Set

Not applicable. As noted above, this study is based on objective physical and performance measurements via bench testing, not subjective assessment requiring adjudication.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size

No, an MRMC comparative effectiveness study was not done. This type of study is relevant for diagnostic imaging AI, not for performance testing of a physical medical device like a catheter. The aim here is to show "substantial equivalence" of device characteristics and safety, not to measure human reader improvement with AI.

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

Not applicable. This is a physical medical device, not an algorithm or AI system. The "standalone" performance is the device's measured performance during bench testing without human intervention in the device's function.

7. The Type of Ground Truth Used

The ground truth used for this study is based on objective physical and performance measurements derived from standardized engineering and biocompatibility tests (e.g., flow rate, tensile strength, lubricity, cytotoxicity, etc.) and comparison to the known characteristics of the predicate device. It's essentially test results against predetermined specifications and accepted standards (e.g., ISO, ASTM).

8. The Sample Size for the Training Set

Not applicable. This is a physical medical device, not an AI/ML algorithm that requires a training set. The device itself is the product, and its performance is being evaluated, not an algorithm's learning capability.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as there is no training set for a physical medical device.

{0}------------------------------------------------

510 (k) SUMMARY ·

Applicant

ThermopeutiX, Inc. 9951B Businesspark Avenue San Diego, California 92131 Phone: (858) 549-1760 Fax: (858) 549-1717

Manufacturer

ThermopeutiX, Inc. 9951B Businesspark Avenue San Diego, California 92131 Phone: (858) 549-1760 Fax: (858) 549-1717

2 6 2013

Contact Person

Jason Ford, Quality Assurance

Common Names: Support Catheter

Classification Name: Devices of this type are classified as Class II under 21 CFR Part CFR Part 870.1250, Percutaneous Catheter (Product Code DQY).

Proprietary Name: Primi™ Support Catheter

Predicate Devices

The ThermopeutiX Primi™ Support Catheter with the advanced hydrophilic coating is substantially equivalent in indications, design, construction and features to the ThermopeutiX Primi™ Support Catheter cleared under 510(k) K130850 with the exception of the hydrophilic coating.

{1}------------------------------------------------

Indications for Use

There is no change in the indications for use for the Primi™ Support Catheter with the advanced hydrophilic coating. The existing Primi™ Support Catheter indications statement remains unchanged. The indications for use are as follows:

The Primi™ Support Catheter is intended to be used in conjunction with guidewires in order to access discreet regions of the peripheral vasculature. It may be used to facilitate placement and exchange of guidewires and other interventional devices. The support catheter also may be used to selectively infuse/deliver physician specified diagnostic and/or therapeutic agents, such as sterile saline, contrast agents, vaso-dilators, anti-blastics, lytics, and anti-thrombotics, with or without a guidewire in position.

Device Description

The ThermopeutiX Support Catheter that is the submission is a single lumen catheter with a radiopaque marker that aids in positioning the catheter and a tapered tip to facilitate entry. The catheter utilizes an integral spine wire. The catheter is available in several diameters and lengths which are 0.014", 0.018", or 0.035" guidewire compatible. The working lengths are 90 cm thru 170 cm. Reference Table 1 for model number availability. The distal portion of the catheter shaft has a hydrophilic coating to facilitate insertion. The device has a luer hub connector to allow infusion of diagnostic and/or therapeutic agents. The device is supplied sterile and is intended for single use only. Reference Table 2 below for physical and performance characteristics.

PU14 -20-90PU14-40-90PU14 -60-90PU14 -80-90PU14 -100-90PU14 -120-90
PU14 -20-90HFPU14 -40-90HFPU14 -60-90HFPU14 -80-90HFPU14 -100-90HFPU14 -120-90HF
PU18-20-90PU18-40-90PU18-60-90PU18-80-90PU18-100-90PU18-120-90
PU35-20-90PU35-40-90PU35-60-90PU35-80-90PU35-100-90PU35-120-90
PU14 -20-130PU14 -40-130PU14 -60-130PU14 -80-130PU14 -100-130PU14 -120-130
PU14 -20-130HFPU14-40-130HFPU14 -60-130HFPU14 -80-130HFPU14 -100-130HFPU14 -120-130HF
PU18-20-130PU18-40-130PU18-60-130PU18-80-130PU18-100-130PU18-120-130
PU35-20-130PU35-40-130PU35-60-130PU35-80-130PU35-100-130PU35-120-130
PU14 -20-150PU14 -40-150PU14 -60-150PU14 -80-150PU14 -100-150PU14 -120-150
PU14 -20-150HFPU14 -40-150HFPU14 -60-150HFPU14 -80-150HFPU14 -100-150HFPU14 -120-150HF
PU18-20-150PU18-40-150PU18-60-150PU18-80-150PU18-100-150PU18-120-150
PU35-20-150PU35-40-150PU35-60-150PU35-80-150PU35-100-150PU35-120-150
PU14 -20-170PU14 -40-170PU14 -60-170PU14 -80-170PU14 -100-170PU14 -120-170
PU14 -20-170HFPU14 -40-170HFPU14 -60-170HFPU14 -80-170HFPU14 -100-170HFPU14 -120-170HF
PU18-20-170PU18-40-170PU18-60-170PU18-80-170PU18-100-170PU18-120-170
PU35-20-170PU35-40-170PU35-60-170PU35-80-170PU35-100-170PU35-120-170

Table 1- Model Numbers for the Primi™ Support Catheter with the advanced hydrophilic coating

  • The HF suffix indicates the high flow version

PUxx-yy-zzz, xx=guide wire size (0.014", 0.018", or 0.035")

yy=distal shaft length (20mm, 40mm, 60mm, 100mm, or 120mm) zzz=Overall catheter length (150cm or 170cm)

{2}------------------------------------------------

Element/FeatureThermopeutiX Primi Support Catheters
Working Length (s)90 cm - 170 cm; polymer shaft with integral spine wire
Distal Shaft Length (mm)20, 40, 60, 80, 100, 120
Outside Diameter (s)Distal:0.045" - 0.064"Proximal: 0.060" - 0.080"
LumenOne
Guide wire Compatibility0.014" - 0.035"
Distal End ConfigurationTapered tip for ease of crossing
Fluid connectionLuer hub connector
Flow rate @ 150 psi0.014" wire models > 0.80 cc/sec, Saline0.035" wire models > 0.50 cc/sec, Saline0.014" wire models > 0.30 cc/sec, Contrast0.035 wire models > 0.15 cc/sec, Contrast
Maximum Injection Pressure300 psi
Hydrophilic CoatingYes; distal portion, for lubricity
Radiopaque MarkersOne
Sheath Compatibility5 Fr and 6 Fr
Catheter MaterialsHub-PebaxOuter Shaft-PebaxInner Shaft-HDPEDistal Shaft-Pebax w/radiopaque fillerCore Wire-Stainless Steel tapered wireCoating-Hydrophilic proprietary coating

Table 2-Physical and Performance Characteristics

Technological Characteristics Comparison

The Primi™ Support Catheter with the advanced hydrophilic coating is identical in design and construction to the currently marketed Primi™ Support Catheter with the exception of the hydrophilic coating. The advanced hydrophilic coating provides comparable lubricity with a reduction in particulates. The coating is applied to the same portion of the candidate device as the predicate device. The coating thickness is considered equivalent between the candidate and predicate device.

{3}------------------------------------------------

Table 3-Predicate Device Characteristics Comparison
Table 3-Predicate Device Characteristics Comparison
Element/FeatureThermopeutiX Primi Support Catheter (This 510k)ThermopeutiX Primi Support Catheter 510(k) K130850
Working Length (s)90 cm – 170 cm; polymer shaft with integral spine wire90 cm – 170 cm; polymer shaft with integral spine wire
Outside Diameter (s)Distal:0.045" – 0.064"Proximal: 0.060" – 0.080"Distal:0.045" – 0.064"Proximal: 0.060" – 0.080"
LumenOneOne
Guide wire Compatibility0.014"-0.035"0.014"-0.035"
Guide wire UseOver the WireOver the Wire
Distal End ConfigurationTapered tip for ease of crossingTapered tip for ease of crossing
Fluid connectionLuer hub connectorLuer hub connector
Flow rate @ 150 psi w/Saline0.014" wire models > 0.80 cc/sec0.035" wire models > 0.50 cc/sec0.014" wire models > 0.80 cc/sec0.035" wire models > 0.50 cc/sec
Flow rate @ 150 psi w/Contrast0.014" wire models > 0.30 cc/sec0.035" wire models > 0.15 cc/sec0.014" wire models > 0.30 cc/sec0.035" wire models > 0.15 cc/sec
Maximum Injection PressureDo not exceed 300 psiDo not exceed 300 psi
CoatingYes; advanced hydrophilic, distal portionYes; hydrophilic, distal portion
Coating Length65 cm, 115 cm or 140 cm(lengths are the same as the predicate across all models)65 cm, 115 cm or 140 cm
Coating ThicknessSame as predicateProprietary
Radiopaque MarkersOneOne
Sheath Compatibility5 Fr and 6 Fr5 Fr and 6 Fr
Materials UsedHub-PebaxOuter Shaft-PebaxInner Shaft-HDPEDistal Shaft-Pebax w/radiopaque fillerCore Wire-Stainless Steel tapered wireCoating-Advanced hydrophilic proprietary coatingMarker Band Platinum/IridiumHub-PebaxOuter Shaft-PebaxInner Shaft-HDPEDistal Shaft-Pebax w/radiopaque fillerCore Wire-Stainless Steel tapered wireCoating-Hydrophilic proprietary coatingMarker Band Platinum/Iridium

Non-Clinical Tests

The non-clinical tests performed on sample devices are listed below. All testing passed internal protocol requirements and met product specifications. All of the requirements in the standards utilized were met.

{4}------------------------------------------------

-Biocompatibility Testing (ISO10993-1)

-Cytotoxicity (ISO 10993-5)

-Sensitization (Guinea Pig Maximization ISO 10993-10)

-Irritation (ISO 10993-10)

-Systemic Toxicity (ISO 10993-11)

-Complement Activation (C3a and SC5b-9 ISO10993-4)

-Pyrogen (ISO 10993-11)

-Hemocompatibility (Hemolysis, Thromboresistance, Partial Thromboplastin Time (ISO

10993-4 and ASTM F2382)

-Device Functional Testing

-Dimensional

-Wire Movement/Compatibility

-Catheter Torque

-Catheter Kink

-Hub Leakage

-Peripheral Model Passage

-Coating Particulate

-Coating Durability/Integrity

-Radiopacity

-Maximum Injection Pressure

-Flow Rate Testing (wire in/wire out)

-Tensile Test (distal tip, distal tip to wire, distal shaft to proximal shaft, hub to shaft)

-Pouch Seal Integrity

-All performance testing was repeated on samples with a minimum of 10 month aging.

Conclusion

Based on the non-clinical testing results, the Primi™ Support Catheters with the advanced hydrophilic coating are substantially equivalent to the legally marketed device identified in this submission.

{5}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines.

September 26, 2013

Public Health Service

Food and Drug Administration 0903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

Thermopeutix, Inc. Mr. Jason Ford Quality Engineer 9951B Businesspark Avenue San Diego, CA 92131

Re: K132701

Trade/Device Name: Primi™ Support Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous catheter Regulatory Class: II Product Code: DOY Dated: July 30, 2013 Received: August 29, 2013

Dear Mr. Ford:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{6}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Bram D. Zuckerman -S

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

{7}------------------------------------------------

ThermopeutiX, Inc. Special 510(k) Premarket Notification July 30, 2013

INDICATIONS FOR USE STATEMENT

510(k) Number (If Known):

Device Name: Primi™ Support Catheter

Indications For Use:

The Primi™ Support Catheter is intended to be used in conjunction with guidewires in order to access discreet regions of the peripheral vasculature. It may be used to facilitate placement and exchange of guidewires and other interventional devices. The support catheter also may be used to selectively infuse/deliver physician specified diagnostic and/or therapeutic agents, such as sterile saline, contrast agents, vaso-dilators, anti-blastics, lytics, and anti-thrombotics, with or without a guidewire in position.

Prescription Use: (Part 21 CFR 801 Subpart D) AND/OR

Over-the-counter Use: (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Bram D. Zuckerman -S 2013.09.26 20:13:14 -04

§ 870.1250 Percutaneous catheter.

(a)
Identification. A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.(b)
Classification. Class II (performance standards).