(165 days)
The Sonicaid™ Freedom Wireless Fetal Monitoring System ('Freedom') is a wireless transducer system for monitoring fetal heart movement and maternal contractions during intrapartum and antepartum periods of pregnancy.
It is an optional accessory for use with Huntleigh Healthcare Limited's Sonicaid™ FM820 and FM830 Encore Fetal Monitors ('FM800E Monitors') as an alternative to their wired transducers. When connected to an FM800E Monitor, the system monitors:
• Uterine activity by using an external, pressure-sensitive TOCO transducer, and
• Fetal Heart Rate (FHR) by pulsed Doppler ultrasound using an external Ultrasound transducer.
Freedom is suitable for use with pregnant women in clinical and hospital facilities. The transducers are water tight allowing pregnant women to be monitored while they are mobile, stationary or in a bath or shower environment.
This system should only be used by, or under the supervision of, a licensed physician or other health practitioner who is trained in the use of FHR monitors.
The Sonicaid™ Freedom is intended to monitor fetal heart movement and uterine activity by using wireless, non-invasive transducers on pregnant women during antepartum and intrapartum phases of pregnancy. This system is an optional accessory that must be connected to, and used with, a Sonicaid™ FM800E Monitor when a woman wishes to be mobile or in a water environment when being monitored or giving birth.
The Freedom has three major components; the Receiver and Ultrasound ('US') and Tocograph ('TOCO') transducers. The wireless transducers are powered by internal batteries that are recharged when docked on the Receiver.
The Sonicaid ™Freedom operates in the WMTS frequency bandwidth. There are 100 selectable channels available for use. If there is interference when installing the system, then the channel may be changed to achieve clear transmission.
The Huntleigh Healthcare Ltd. Sonicaid™ Freedom Wireless Fetal Monitoring System (K132650) did not report specific acceptance criteria or a dedicated clinical study to demonstrate its performance relative to specific metrics. Instead, the submission relies on demonstrating substantially equivalent performance to a predicate device through bench tests.
Here’s a breakdown of the information based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
As noted, specific quantitative acceptance criteria (e.g., sensitivity, specificity, accuracy) are not explicitly stated for the device's main function of fetal monitoring. The performance is indirectly demonstrated by showing comparability to the predicate device through bench tests.
| Acceptance Criteria Category | Specific Metric (Not Explicitly Stated) | Reported Device Performance (as demonstrated by bench tests) |
|---|---|---|
| Acoustic Output | Compliance with IEC 60601-2-37 | Achieved compliance. Ultrasound transducer elements have similar structure and acoustic output levels are comparable to predicate, within safety limits. |
| Environmental Requirements | Compliance with ISO 60068-2 | Achieved compliance. |
| Electrical Safety | Compliance with EN 60601-1 | Achieved compliance. |
| EMC | Compliance with EN 60601-1-2 / FCC | Achieved compliance. |
| Software | Compliance with EN 62304 | Achieved compliance. |
| Material Biocompatibility | Compliance with ISO 10993 | Achieved compliance. |
| FHR and UA Detection | Not explicitly quantified. Performance of Ultrasound and TOCO transducers is comparable to Sonicaid™ FM800E transducers (which are wired, not the predicate). This implies similar accuracy and reliability in detecting FHR and UA. | |
| Wireless Communication | Range, signal strength, interference handling, battery life. | Comparable to predicate device. WMTS frequency band used, 100 selectable channels. Indicators for interference, low battery, out-of-range provided. |
| Water Ingress Rating | Water tightness for use in water environments. | Same as predicate device. |
2. Sample Size Used for the Test Set and Data Provenance
- Test Set Sample Size: Not applicable/not reported. The submission did not involve a clinical test set with patient data for performance evaluation. Performance was assessed through simulated bench tests.
- Data Provenance: Not applicable, as there were no clinical test sets or patient data involved. The tests were simulated bench tests.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
- Not applicable. No expert-established ground truth was reported for a clinical test set, as no clinical tests were performed.
4. Adjudication Method for the Test Set
- Not applicable. No clinical test set or adjudication method was reported.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done
- No, an MRMC comparative effectiveness study was not done. The submission explicitly states: "No clinical tests were required to determine substantial equivalence between the Freedom and the predicate device."
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done
- The device itself is a fetal monitoring system that collects data, not an algorithm that interprets data. While the performance of the transducers (the "standalone" components for data acquisition) was assessed through bench tests, it's not a "standalone algorithm" in the typical sense of AI. The system is designed to be used by trained health practitioners.
7. The Type of Ground Truth Used
- For the technical and safety aspects, the "ground truth" was established by compliance with recognized standards (e.g., IEC, ISO, EN, FCC) and by comparison to the established performance of existing Sonicaid™ FM800E transducers (for FHR and UA detection) and the predicate device (Avalon CTS Cordless Fetal Transducer System) for overall system functionality and design. There was no clinical ground truth (e.g., pathology, outcomes data) based on patient results provided.
8. The Sample Size for the Training Set
- Not applicable. This device is a hardware system for signal acquisition, not an AI/machine learning algorithm requiring a "training set" for model development.
9. How the Ground Truth for the Training Set was Established
- Not applicable, as no training set was used.
{0}------------------------------------------------
Huntleigh Healthcare Ltd
FEB - 7 2014
HUNTLEIGH DIAGNOSTIC PRODUCTS DIVISION
Section 5: 510(k) Summary as required by Section 807.92(c)
APPLICANT INFORMATION 5.1
| Huntleigh Healthcare Limited, |
|---|
| Diagnostic Products Division |
| 35 Portmanmoor Road |
| CARDIFF, CF24 5HN |
| United Kingdom |
| Laura Garcia Deacon |
| Regulatory Affairs Engineer |
| +44 2920 447703 |
| +(44) 2920 492520 |
| Laura.deacon@huntleigh-diagnostics.co.uk |
| January 9, 2014 |
5.2 DEVICE INFORMATION
| Proprietary Name: | Sonicaid™ FreedomWireless Fetal Monitoring System | |
|---|---|---|
| Common/Classification Name: | Perinatal Monitoring Device & Accessories | |
| Classification | Class II | |
| Product Code | HGM | |
| Classification Regulation | 884.2740 |
5.3 PREDICATE DEVICE
| Predicate Device: | Avalon CTS Cordless Fetal Transducer System(K023931) |
|---|---|
| Owned by: | Philips Medical Systems,Nederland B.V. Post bus 10.000,5680 DA Best,Netherlands |
| Manufactured by: | Philips Medizin Systeme Boeblingen GmbH,Cardiac and Monitoring Systems,Hewlett-Packard Str.2,71034 Boeblingen,Germany |
{1}------------------------------------------------
5.4 INDICATIONS FOR USE
Sonicaid™ Freedom Wireless Fetal Monitoring System
The Sonicaid™ Freedom Wireless Fetal Monitoring System ('Freedom') is a wireless transducer system for monitoring fetal heart movement and maternal contractions during intrapartum and antepartum periods of pregnancy.
It is an optional accessory for use with Huntleigh Healthcare Limited's Sonicaid™ FM820 and FM830 Encore Fetal Monitors ('FM800E Monitors') as an alternative to their wired transducers. When connected to an FM800E Monitor, the system monitors:
• Uterine activity by using an external, pressure-sensitive TOCO transducer, and
• Fetal Heart Rate (FHR) by pulsed Doppler ultrasound using an external Ultrasound transducer.
Freedom is suitable for use with pregnant women in clinical and hospital facilities. The transducers are water tight allowing pregnant women to be monitored while they are mobile, stationary or in a bath or shower environment.
This system should only be used by, or under the supervision of, a licensed physician or other health practitioner who is trained in the use of FHR monitors.
DEVICE DESCRIPTION 5.5
The Sonicaid™ Freedom is intended to monitor fetal heart movement and uterine activity by using wireless, non-invasive transducers on pregnant women during antepartum and intrapartum phases of pregnancy. This system is an optional accessory that must be connected to, and used with, a Sonicaid™ FM800E Monitor when a woman wishes to be mobile or in a water environment when being monitored or giving birth.
The Freedom has three major components; the Receiver and Ultrasound ('US') and Tocograph ('TOCO') transducers. The wireless transducers are powered by internal batteries that are recharged when docked on the Receiver.
The Sonicaid ™Freedom operates in the WMTS frequency bandwidth. There are 100 selectable channels available for use. If there is interference when installing the system, then the channel may be changed to achieve clear transmission.
Technology Overview
Receiver
The Receiver must be connected to a Sonicaid™ FM800E Monitor to operate the wireless transducers. The front panel has two docking stations for charging the TOCO and US transducers and for storing them and three sets of visual indicators, which inform an operator about the status of the transducers when docked or in use, include:
- . The charging indicators ensure that transducers are recognised by the receiver, docked properly and recharging until the battery is fully recharged.
- . The transmission indicators confirm that the transducers are in range and there is no interference during use.
- The battery indicator provides the operator with the relative level of battery capacity. ●
These indicators also inform the operator when there is interference, low battery capacity, when a patient is out of transmission range or if there is a fault.
{2}------------------------------------------------
The back of the Receiver contains the connecting points for the receiving antenna, RF-channel selector, interface to the Sonicaid FM800E Monitor and the power supply input.
The Receiver's internal components drive the system through the microcontroller, selectable radio frequency channels, an antenna and a near field communication ('NFC') transmitter.
Transducers
The US and TOCO transducers are both wireless, water tight and may be used when patients are stationary, mobile or in a water environment.
Ultrasound Transducer
The US transducer detects fetal heart movement by pulsed Doppier ultrasound. This is driven by a master clock that produces the pulse streams necessary for the detection of FHR. Other components include the battery power supply and an RF transmitter.
TOCO Transducer
The TOCO transducer is a pressure sensitive device that detects contractions or uterine activity. The key components within the TOCO transducer are the NFC transmitter, pressure sensor and the battery power supply.
5.6 SUBSTANTIAL EQUIVALENCE
The Sonicaid™ Freedom is substantially equivalent to the cleared predicate device, the Avalon CTS Cordless Fetal Transducer System (K023931).
5.7 TECHNOLOGY COMPARISONS
The Sonicaid™ Freedom and the predicate device are equivalent in the following ways:
- . Each system has a Receiver unit with docking stations for each transducer. Both systems have a least one Ultrasound transducer and one TOCO transducer.
- · Both wireless transducer systems have been designed as accessories for use with compatible fetal monitoring systems with wired transducers.
- Neither unit can be used as a stand-alone system; both must be used only when they are . connected to specified fetal monitoring systems.
- · Both systems have TOCO and Ultrasound transducers that are wireless, powered by rechargeable lithium batteries and have the same water ingress rating; these features enable patients to be monitored when they are stationary, mobile or in a water environment.
- · Communication between the transducers and the Receiver is via a WMTS frequency band.
- . Both systems automatically recharge the transducers' batteries when transducers are docked on their respective Receivers.
- · Both systems have similar indicators to monitor the battery capacity and the operating status of the transducers.
- · The intended use, indications for use, patient population, clinical settings and operating conditions for monitoring fetal heart movement and uterine contractions are the same.
- Both systems use the same or similar methods and technologies for determining FHR and uterine activity; FHR by Doppler ultrasound and UA by pressure detection.
{3}------------------------------------------------
- • The Ultrasound transducer elements from both systems have a similar structure and their acoustic output levels are comparable and within safety limits.
5.8 BENCH TESTS
Verification and validation tests were conducted to demonstrate performance, safety and substantial equivalence to the predicate device. Compliance to the following was achieved:
- · Ultrasound Transducer Acoustic Output (IEC 60601-2-37)
- · Environmental Requirements (ISO 60068-2)
- · Electrical Safety (EN 60601-1)
- · EMC (EN 60601-1-2 /FCC)
- · Software (EN 62304)
- · Material Biocompatibility (ISO 10993)
Simulated bench tests were used to demonstrate that the performance of the Sonicaid™ Freedom Ultrasound and TOCO transducers is comparable to the Sonicaid™ FM800E transducers.
CLINICAL TESTS 5.9
No clinical tests were required to determine substantial equivalence between the Freedom and the predicate device.
5.10 CONCLUSIONS
The conclusions drawn from the data detailed within this submission demonstrate that the Sonicaid™ Freedom is as safe, as effective, and performs as well as the legally marketed Avalon CTS Cordless Fetal Transducer System (K023931).
The Sonicaid™ Freedom does not raise any new concerns regarding safety and effectiveness.
Project Number: HD 778
{4}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle or bird-like symbol with three curved lines representing its body and wings. The logo is surrounded by text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 7, 2014
Huntleigh Healthcare Limited Laura Garcia Deacon Regulatory Affairs Engineer 35 Portmanmoor Road Cardiff, Wales CF24 5HN United Kingdom
Re: K132650
Trade/Device Name: Sonicaid™ Freedom Wireless Fetal Monitoring System Regulation Number: 21 CFR§ 884.2740 Regulation Name: Perinatal monitoring system and accessories Regulatory Class: II Product Code: HGM Dated: January 9, 2014 Received: January 16, 2014
Dear Laura Garcia Deacon,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{5}------------------------------------------------
Page 2 - Laura Garcia Deacon
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
Herbert P. Lerner -S
for
Beniamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Indications for Use
| 510(k) Number: | K132650 |
|---|---|
| ---------------- | --------- |
Sonicaid™ Freedom Wireless Fetal Monitoring System Device Name:
Indications for Use:
The Sonicaid™ Freedom Wireless Fetal Monitoring System ('Freedom') is a wireless transducer system for monitoring fetal heart movement and maternal contractions during intrapartum and antepartum periods of pregnancy.
It is an optional accessory for use with Huntleigh Healthcare Limited's Sonicaid™ FM820 and FM830 Encore Fetal Monitors ('FM800E Monitors') as an alternative to their wired transducers. When connected to an FM800E Monitor, the system monitors:
- · Uterine activity by using an external, pressure-sensitive TOCO transducer, and
- Fetal Heart Rate (FHR) by pulsed Doppler ultrasound using an external Ultrasound transducer.
Freedom is suitable for use with pregnant women in clinical and hospital facilities. The transducers are water tight allowing pregnant women to be monitored while they are mobile, stationary or in a bath or shower environment.
This system should only be used by, or under the supervision of, a licensed physician or other health practitioner who is trained in the use of FHR monitors.
Prescription Use YES (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use NO (Part 21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Herbert P. Lerner -S 2014.02.07 12:04:08 -05'00'
§ 884.2740 Perinatal monitoring system and accessories.
(a)
Identification. A perinatal monitoring system is a device used to show graphically the relationship between maternal labor and the fetal heart rate by means of combining and coordinating uterine contraction and fetal heart monitors with appropriate displays of the well-being of the fetus during pregnancy, labor, and delivery. This generic type of device may include any of the devices subject to §§ 884.2600, 884.2640, 884.2660, 884.2675, 884.2700, and 884.2720. This generic type of device may include the following accessories: Central monitoring system and remote repeaters, signal analysis and display equipment, patient and equipment supports, and component parts.(b)
Classification. Class II (performance standards).