(238 days)
InterV Kyphoplasty Catheter is intended to be used for reduction and fixation of fractures and/or creation of a void in cancellous bone in the spine during balloon kyphoplasty (for use with cleared spinal polymethymethacrylate (PMMA) bone cements).
InterV Kyphoplasty Catheter is designed for use in balloon kyphoplasty; it comes as a single-use double lumen catheter with a low profile balloon mounted on the distal tip. The balloon is designed to compress cancellous bone and/or move cortical bone as it inflates. The balloon is designed to withstand bone resistance leading to an a internal pressure of up to 400 psi (27 ATM). The key components are the balloon, shaft, Y-connector and two radiopaque marker bands positioned on the inner tubing/lumen at the proximal and distal ends of the inflatable component
The provided text describes the InterV Kyphoplasty Catheter and its premarket notification [510(k)] submission. It outlines the device's technical characteristics and the non-clinical tests conducted to demonstrate its substantial equivalence to predicate devices.
1. A table of acceptance criteria and the reported device performance
| Characteristic | Acceptance Criteria (Implied) | Reported Device Performance |
|---|---|---|
| Inflation Pressure | Exceed minimum burst pressure in a constrained environment | The balloon catheter exceeded the requirements for the minimum burst pressure in a constrained environment. |
| Inflation Volume | Exceed minimum burst volume in a constrained environment | The balloon catheter exceeded the requirements for the minimum burst volume in a constrained environment. |
| Inflated Balloon Dimensions | Substantially equivalent to predicate devices | The inflated balloon dimensions were substantially equivalent to those of the predicate devices. |
| Balloon Double Wall Thickness | Substantially equivalent to predicate devices | The double wall thickness of the balloons was substantially equivalent to those of the predicate devices. |
Note: The document does not explicitly state "acceptance criteria" values but rather reports that the device "exceeded the requirements" or was "substantially equivalent" to predicate devices, implying these as the criteria.
2. Sample size used for the test set and the data provenance
The document does not provide details on the specific sample sizes for the non-clinical tests (e.g., number of catheters tested for each characteristic). It also does not specify the data provenance (e.g., country of origin, retrospective or prospective) for these non-clinical tests. As these were non-clinical tests, the concept of "data provenance" in the context of human subjects is not applicable.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
This information is not applicable. The study involved non-clinical, in-vitro testing of device characteristics, not an evaluation requiring human expert interpretation of data or images.
4. Adjudication method for the test set
This information is not applicable. Adjudication methods (like 2+1, 3+1) are typically used in clinical studies or studies involving human readers, which this submission explicitly states were not conducted ("N/A- No clinical tests were conducted for this submission"). The tests described are engineering performance tests.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
No MRMC comparative effectiveness study was done. The document explicitly states: "N/A- No clinical tests were conducted for this submission." This product is a physical medical device (catheter), not an AI diagnostic tool.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done
A standalone performance evaluation was done in the sense that the device's physical characteristics were tested independently. However, this is not an "algorithm only" or "AI without human-in-the-loop" study as the device is not an AI algorithm. The performance evaluation was focused on the mechanical and physical properties of the catheter itself.
7. The type of ground truth used
The ground truth for the non-clinical tests was established by direct measurement and observation against predefined engineering specifications and criteria, or by comparison to the performance of legally marketed predicate devices. For example:
- Inflation Pressure: Measured burst pressure.
- Inflation Volume: Measured burst volume.
- Inflated Balloon Dimensions: Measured dimensions compared against predicate devices.
- Balloon Double Wall Thickness: Calibrated measurement.
8. The sample size for the training set
This information is not applicable. This submission describes a physical medical device, not an AI algorithm that requires a training set.
9. How the ground truth for the training set was established
This information is not applicable as there was no training set for an AI algorithm.
{0}------------------------------------------------
APR 1 6 2014
K132620 - Page 1 of 3
Image /page/0/Picture/2 description: The image shows the logo for "pan medical". The word "pan" is written in white letters inside of a black circle. The word "medical" is written in black letters to the right of the circle.
InterV Kyphoplasty Catheter Premarket Notification [510(k)] Submission
Premarket Notification [510(k)] Summary
.
SUBMITTER INFORMATION
| Manufacturer's Name: | Pan Medical Ltd |
|---|---|
| Manufacturer's Address: | Barnett Way, Barnwood, Gloucester, GL4 3RT UK |
| Telephone (DDI): | +44 1452 621621 |
| Fax: | +44 1452 372140 |
| Establishment Registration: | 3005146147 |
| Contact Person: | · Jennie Budding (Director of R&D/Production) |
| Date Prepared: | 16-April-2014 |
DEVICE INFORMATION
| Trade Name: | InterV Kyphoplasty Catheter |
|---|---|
| Common Name: | Inflatable Bone Tamp |
| Device Class: | II |
| Classification Name: | Polymethylmethacrylate (PMMA) Bone CementArthroscope |
| Classification Panel: | Orthopedic Devices |
| Classification Regulation: | 21 CFR 888.302721 CFR 888.1100 |
| Product Code(s): | NDNHRX |
| Predicate Device: | Kyphon Inflatable Bone Tamp, K981251Kyphx® Xpander Inflatable Bone Tamps, K041454 |
| Reason for 510(k) submission: | New Device |
{1}------------------------------------------------
Pan Medical Ltd.
InterV Kyphoplasty Catheter Premarket Notification [510(k)] Submission
Device Description:
InterV Kyphoplasty Catheter is designed for use in balloon kyphoplasty; it comes as a single-use double lumen catheter with a low profile balloon mounted on the distal tip. The balloon is designed to compress cancellous bone and/or move cortical bone as it inflates. The balloon is designed to withstand bone resistance leading to an a internal pressure of up to 400 psi (27 ATM). The key components are the balloon, shaft, Y-connector and two radiopaque marker bands positioned on the inner tubing/lumen at the proximal and distal ends of the inflatable component
Intended use:
InterV Kyphoplasty Catheter is intended to be used for reduction and fixation of fractures and/or creation of a void in cancellous bone in the spine during balloon kyphoplasty (for use with cleared spinal polymethymethacrylate (PMMA) bone cements).
SUMMARY OF TECHNOLOGICAL CHARACTERISTICS OF THE DEVICE COMPARED TO THE PREDICATE DEVICES
| Characteristic | New device | Predicate Device |
|---|---|---|
| Trade name | InterV Kyphoplastycatheter | Kyphon Inflatable Bone Tamp,K981251Kyphx® Xpander Inflatable BoneTamps, K041454 |
| Compatible Cannula Size | 8 G / 4.2 mm | 8 G / 4.2 mm |
| Balloon Inflation Medium | 60% Contrast | 60% Contrast |
| Balloon Material | Polyurethane | Polyurethane |
| Balloon Size (Deflated Length) | 10, 15 and 20 mm | 10, 15 and 20 mm |
| Guide wire (Stylet) | Stainless Steel | Stainless Steel |
| Balloon Shape | Cylindrical | Cylindrical |
| Maximum Recommended InflationPressure | 400 psi (27 ATM) | 400 psi (27 ATM) |
| Maximum Recommended InflationVolume (10 mm and 15 mm) | 4 ml | 4 ml |
| Maximum Recommended InflationVolume (20 mm) | 6 ml | 6 ml |
{2}------------------------------------------------
InterV Kyphoplasty Catheter Premarket Notification [510(k)] Submission
| Characteristic | Test Method | Results Summary |
|---|---|---|
| Inflationpressure | Constrained BurstTest | The balloon catheter exceeded the requirements for theminimum burst pressure in a constrained environment |
| InflationVolume | Unconstrained BurstTest | The balloon catheter exceeded the requirements for theminimum burst volume in a constrained environment |
| Inflated BalloonDimensions | Balloon InflationTest | The inflated balloon dimensions were substantiallyequivalent to those of the predicate devices |
| Balloon DoubleWall Thickness | CalibratedMeasurement | The double wall thickness of the balloons was substantiallyequivalent to those of the predicate devices |
SUMMARY OF NON-CLINICAL TESTS
SUMMARY OF CLINICAL TESTS
N/A- No clinical tests were conducted for this submission
CONCLUSION DRAWN FROM NON-CLINICAL DATA
The results of the non-clinical tests show that the InterV Kyphoplasty Catheter meets or exceeds all performance requirements, and are substantially equivalent to the predicate devices.
.
{3}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal is circular and contains the department's name around the perimeter. In the center of the seal is a stylized caduceus, a symbol often associated with medicine and healthcare. The caduceus consists of a staff with two snakes coiled around it.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 16, 2014
Pan Medical Ltd. Ms. Jennie Budding Director of R&D/Production Barnett Way, Barnwood Gloucester GL4 3RT United Kingdom
Re: K132620
Trade/Device Name: InterV Kyphoplasty Catheter Regulation Number: 21 CFR 888.3027 Regulation Name: Polymethylmethacrylate (PMMA) bone cement Regulatory Class: Class II Product Code: NDN, HRX Dated: February 28, 2014 Received: March 4, 2014
Dear Ms. Budding:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing
{4}------------------------------------------------
Page 2 - Ms. Jennie Budding
(21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Lori A. Wiggins
for
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Pan Medical Ltd.
InterV Kyphoplasty Catheter Premarket Notification [510(k)] Submission
Indications for Use
510(k) Number (if known): K132620
Device Name: InterV Kyphoplasty Catheter
Indications for Use:
InterV Kyphoplasty Catheter is intended to be used for reduction and fixation of fractures and/or creation of a void in cancellous bone in the spine during balloon kyphoplasty (for use with cleared spinal polymethymethacrylate (PMMA) bone cements).
Prescription Use __ X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use ________ (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Laurence D. Coyne -S
(Division Sign-Off) Division of Orthopedic Devices 510(k) Number: K132620
Indications For Use Statement
§ 888.3027 Polymethylmethacrylate (PMMA) bone cement.
(a)
Identification. Polymethylmethacrylate (PMMA) bone cement is a device intended to be implanted that is made from methylmethacrylate, polymethylmethacrylate, esters of methacrylic acid, or copolymers containing polymethylmethacrylate and polystyrene. The device is intended for use in arthroplastic procedures of the hip, knee, and other joints for the fixation of polymer or metallic prosthetic implants to living bone.(b)
Classification. Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Polymethylmethacrylate (PMMA) Bone Cement.”