K Number
K132567
Device Name
ARGYLE PERIPHERALLY INSERTED CENTRAL CATHETER SINGLE LUMEN, ARGYLE PERIPHERALLY INSERTED CENTRAL CATHETER DUAL LUMEN
Manufacturer
Date Cleared
2013-11-05

(82 days)

Product Code
Regulation Number
880.5970
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdparty
Intended Use
The catheter is designed for cases in which venous catheterization or long term I.V. administration is necessary. Placement is routinely achieved from peripheral venous site, but the catheter may be placed via subclavian cutdown as well. The catheter may be used to administer fluids for hydration and parenteral nutrition, as well as other commonly used intravenous medications.
Device Description
The Argyle™ Peripherally Inserted Central Catheter is a sterile, single use indwelling catheter inserted into a venous access site for the infusion of fluids, medications and/or nutritional products.
More Information

Not Found

No
The summary describes a standard medical catheter and does not mention any AI or ML capabilities.

No.
The device is used to administer fluids, medications, and nutritional products, but it does not directly treat a disease or condition itself.

No
The device is described as a catheter for administering fluids, medications, and nutritional products, which is a therapeutic function, not a diagnostic one.

No

The device description clearly states it is a "sterile, single use indwelling catheter," which is a physical hardware component, not software.

Based on the provided information, this device is not an IVD (In Vitro Diagnostic).

Here's why:

  • IVD Definition: In vitro diagnostics are tests performed on samples taken from the human body, such as blood, urine, or tissue, to detect diseases, conditions, or infections. They are used to provide information for diagnosis, monitoring, or screening.
  • Device Function: The description clearly states that this device is a catheter designed for infusion of fluids, medications, and nutritional products directly into the venous system. It is an indwelling catheter used for administering substances into the body, not for analyzing samples taken from the body.
  • Intended Use: The intended use is for venous catheterization and long-term IV administration, which are therapeutic and supportive procedures, not diagnostic testing.

Therefore, the Argyle™ Peripherally Inserted Central Catheter is a medical device used for treatment and support, not an in vitro diagnostic device.

N/A

Intended Use / Indications for Use

The Argyle Peripherally Inserted Central Catheter Single Lumen The Argyle Peripherally Inserted Central Catheter Dual Lumen

Indications for Use:
The catheter is designed for cases in which venous catheterization or long term I.V. administration is necessary. Placement is routinely achieved from peripheral venous site, but the catheter may be placed via subclavian cutdown as well. The catheter may be used to administer fluids for hydration and parenteral nutrition, as well as other commonly used intravenous medications.

Product codes (comma separated list FDA assigned to the subject device)

FOZ, LJS

Device Description

The Argyle Peripherally Inserted Central Catheter is a sterile, single use indwelling catheter inserted into a venous access site for the infusion of fluids, medications and/or nutritional products.

Argyle™ Peripherally Inserted Central Catheter Single Lumen
The Argyle Peripherally Inserted Central Catheter is a single-lumen catheter made of a radiopague polyurethane catheter tube that is 1.9 Fr. at the distal end. The catheter transitions into a 5 Fr. diameter clear polyurethane tube (pigtail) with an insert molded polypropylene hub through a stabilizing wing junction that is used for attaching the Peripherally Inserted Central Catheter to the patient.

Argyle™ Peripherally Inserted Central Catheter Dual Lumen
The Argyle Peripherally Inserted Central Catheter is a dual-Lumen catheter made of a radiopaque polyurethane catheter tube that is 1.9 Fr. at the distal end. The catheter transitions into two 5 Fr. diameter clear polyurethane tubes (pigtails) with an insert molded polypropylene hub through a stabilizing wing junction that is used for attaching the Peripherally Inserted Central Catheter to the patient. The primary lumen pigtail is longer than the secondary lumen pigtail.

The insertion tip of the Argyle Peripherally Inserted Central Catheter is manufactured in a straight or blunt cut configuration. Placement of the device is facilitated by depth markings, which are printed on the catheter at 1 cm intervals, beginning at 3 cm from the insertion tip and terminating at 30 cm at the proximal end near the stabilizing wing. The 5 cm marks are identified in bold with the 20 cm and 30 cm marks identified with two and three bold marks respectively.

The Argyle Peripherally Inserted Central Catheter will continue to be available in both the single lumen and the dual lumen 1.9 Fr. size. Each device will continue to be packaged in a Tyvek pouch: 10 pouches are packaged in a carton. The Argyle Peripherally Inserted Central Catheter does not contain DEHP and is Ethylene Oxide sterilized. The product and packaging is not made of natural rubber latex.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

venous access site

Indicated Patient Age Range

Not Found

Intended User / Care Setting

Not Found

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

Laboratory testing was completed to support substantial equivalence between the modified devices and the current devices. The modified devices were evaluated to show compliance to the applicable sections of the standards requirements as well as performance characteristics related to the modification of the devices. The results of the testing show that the modified devices continue to meet the requirements of the product specifications and support the determination of substantial equivalence.

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

Not Found

Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.

K974015, K042461, K061936

Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.

Not Found

Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).

Not Found

§ 880.5970 Percutaneous, implanted, long-term intravascular catheter.

(a)
Identification. A percutaneous, implanted, long-term intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings, such as luer hubs, and accessories that facilitate the placement of the device. The device allows for repeated access to the vascular system for long-term use of 30 days or more, and it is intended for administration of fluids, medications, and nutrients; the sampling of blood; and monitoring blood pressure and temperature. The device may be constructed of metal, rubber, plastic, composite materials, or any combination of these materials and may be of single or multiple lumen design.(b)
Classification. Class II (special controls) Guidance Document: “Guidance on Premarket Notification [510(k)] Submission for Short-Term and Long-Term Intravascular Catheters.”

0

K132567

NOV - 5 2013

Image /page/0/Picture/2 description: The image shows the word COVIDIEN in bold, black letters. To the left of the word is a black and white symbol that looks like a plus sign. Below the word COVIDIEN is the phrase "positive results for life" in a smaller, italicized font.

510(k) Summary

1. 510(k) Owner:

Covidien 15 Hampshire Street Mansfield, MA 02048 Telephone: 508-452-1646 508-261-8461 Fax:

Contact: Dolly Mistry Title: Senior Regulatory Affairs Associate Date Prepared: August 14, 2013

2. Device:

Argyle™ Peripherally Inserted Central Catheter Single Lumen

Trade or proprietary name: The Argyle™ 1.9 Fr Single Lumen Neonatal/Pediatric Peripherally Inserted Central Catheter Common or usual name: Peripherally Inserted Central Catheter Device Description: Short-Term Less Than 30 Days, Therapeutic Intravascular Catheter Regulation Description: Intravascular Catheter Product Code: FOZ Regulation Number: 880.5200 Device Class: 2 Predicate Device: K974015 - Argyle® Quick-PICC™ Neonatal/Pediatric Peripherally Inserted Central Catheter and Accessories* * Please note, the predicate device 510(k) for the Argyle® Quick-PICC™ Neonatal/Pediatric Peripherally Inserted Central Catheter and Accessories, K974015, was submitted by Sherwood Medical Co , this company was acquired by Covidien, formally Tyco Healthcare/Kendall on October 1, 1998.

Argyle™ Peripherally Inserted Central Catheter Dual Lumen

Trade or proprietary name: The Argyle™ 1.9 Fr Dual Lumen Neonatal/Pediatric Peripherally Inserted Central Catheter Common or usual name: Peripherally Inserted Central Catheter Device Description: Long-Term Greater Than 30 Days, Therapeutic Intravascular Catheter

Regulation Description: Percutaneous, implanted, Long-Term Intravascular Catheter Product Code: LJS Regulation Number: 880.5970 Device Class: 2 Predicate Device: K042461 and K061936 - Kendall Argyle® 1.9 Fr. Dual Lumen Neonatal/Pediatric Peripherally Inserted Central Catheter

3. Device Description:

The Argyle™ Peripherally Inserted Central Catheter is a sterile, single use indwelling catheter inserted into a venous access site for the infusion of fluids, medications and/or nutritional products.

1

Arqule™ Peripherally Inserted Central Catheter Single Lumen

The Argyle Peripherally Inserted Central Catheter is a single-lumen catheter made of a radiopague polyurethane catheter tube that is 1.9 Fr. at the distal end. The catheter transitions into a 5 Fr. diameter clear polyurethane tube (pigtail) with an insert molded polypropylene hub through a stabilizing wing junction that is used for attaching the Peripherally Inserted Central Catheter to the patient.

Argyle™ Peripherally Inserted Central Catheter Dual Lumen

The Argyle Peripherally Inserted Central Catheter is a dual-Lumen catheter made of a radiopaque polyurethane catheter tube that is 1.9 Fr. at the distal end. The catheter transitions into two 5 Fr. diameter clear polyurethane tubes (pigtails) with an insert molded polypropylene hub through a stabilizing wing junction that is used for attaching the Peripherally Inserted Central Catheter to the patient. The primary lumen pigtail is longer than the secondary lumen pigtail.

The insertion tip of the Argyle Peripherally Inserted Central Catheter is manufactured in a straight or blunt cut configuration. Placement of the device is facilitated by depth markings, which are printed on the catheter at 1 cm intervals, beginning at 3 cm from the insertion tip and terminating at 30 cm at the proximal end near the stabilizing wing. The 5 cm marks are identified in bold with the 20 cm and 30 cm marks identified with two and three bold marks respectively.

The Argyle Peripherally Inserted Central Catheter will continue to be available in both the single lumen and the dual lumen 1.9 Fr. size. Each device will continue to be packaged in a Tyvek pouch: 10 pouches are packaged in a carton. The Argyle Peripherally Inserted Central Catheter does not contain DEHP and is Ethylene Oxide sterilized. The product and packaging is not made of natural rubber latex.

4. Intended Use:

The Argyle™ Peripherally Inserted Central Catheter is designed for cases in which venous catheterization or long term I.V. administration is necessary. Placement is routinely achieved from peripheral venous site, but the catheter may be placed via subclavian cutdown as well. The catheter may be used to administer fluids for hydration and parenteral nutrition, as well as other commonly used intravenous medications.

5. Technological Characteristics:

The modified devices have the same fundamental technological characteristics as compared to the predicate devices: however, there have been non-fundamental technological changes made to the devices since the product clearance by the Agency. In April 2011, the Centers for Disease Control and Prevention (CDC) uporaded their recommendation regarding the use of alcohol to minimize contamination risk by scrubbing the access port. As a result, Covidien is modifying the product labeling and the device design to address the changes in clinical practice for the cleaning of access ports of intravascular catheters.

6. Non-Clinical Performance Data:

Laboratory testing was completed to support substantial equivalence between the modified devices and the current devices. The modified devices were evaluated to show compliance to the applicable sections of the standards requirements as well as performance characteristics

2

related to the modification of the devices. The results of the testing show that the modified devices continue to meet the requirements of the product specifications and support the determination of substantial equivalence.

7. Clinical Data:

No clinical testing was required to be performed for determination of substantial equivalence.

8. Conclusion:

Based on the nonclinical tests performed on the proposed devices, the modified Argyle Peripherally Inserted Central Catheters are as safe and effective as its legally marketed respective predicate devices. The information provided within this 510(k) demonstrates that the modified Argyle Peripherally Inserted Central Catheters are substantially equivalent to the predicate devices.

3

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus, which is a symbol often associated with healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

November 5, 2013

Covidien Dolly Mistry Senior Regulatory Affairs Associate 15 Hampshire Street Mansfield, MA 02048

Re: K132567

Trade/Device Name: Argyle Peripherally Inserted Central Catheter Single Lumen and Argyle Peripherally Inserted Central Catheter Dual Lumen Regulation Number: 21 CFR 880.5970 Regulation Name: Catheter, Intravascular, Therapeutic, Long-Term Regulatory Class: II Product Code: LJS/FOZ Dated: October 10, 2013 Received: October 11, 2013

Dear Ms. Mistry:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

4

Page 2 - Ms. Mistry

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH0ffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Image /page/4/Picture/6 description: The image shows the text "MaryStBunner-S". The text is in a bold, sans-serif font. The letters "StB" are stylized with a geometric pattern. The text is black against a white background.

Kwame Ulmer M.S. Acting Division Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

5

Indications for Use

510(k) Number (if known): K132567

Device Name:

The Argyle™ Peripherally Inserted Central Catheter Single Lumen

The Argyle™ Peripherally Inserted Central Catheter Dual Lumen

Indications for Use:

The catheter is designed for cases in which venous catheterization or long term I.V. administration is necessary. Placement is routinely achieved from peripheral venous site, but the catheter may be placed via subclavian cutdown as well. The catheter may be used to administer fluids for hydration and parenteral nutrition, as well as other commonly used intravenous medications.

Prescription Use XAND/OROver-The-Counter Use
(Part 21 CFR 801 Subpart D) --------------------------------------------------------------------------------------------------------------------------------------------------(21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Image /page/5/Picture/10 description: The image shows the FDA logo. The logo is made up of the letters F, D, and A. The letters are all capitalized and are in a bold, sans-serif font. The letters are arranged in a horizontal line, with the F on the left, the D in the middle, and the A on the right. The logo is black and white.

Richard C. Chapman 2013.11.04 16:38:13 -05,00'