K Number
K132281

Validate with FDA (Live)

Date Cleared
2013-08-30

(38 days)

Product Code
Regulation Number
870.5150
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ReVive™ PV Thrombectomy Device is indicated for:

  • · The non-surgical removal of emboli and thrombi from peripheral blood vessels,
  • · The non-surgical removal of thrombi from synthetic grafts,
  • · Temporary use in peripheral vessel/graft occlusion,
  • Use with aspiration and with the injection or infusion of contrast media and other . fluids
Device Description

The ReVive" PV (Peripheral Vasculature) Thrombectomy Device is a sterile, disposable, thrombectomy device that is intended to restore blood flow via non-surgical removal of emboli and thrombi from peripheral blood vessels, non-surgical removal of thrombi from synthetic grafts and also for temporary use in peripheral vessel/graft occlusion. The ReVive " PV Thrombectomy Device is radiopaque so that fluoroscopy can be used to guide insertion and verify position as necessary.

The ReVive™ PV Thrombectomy Device consists of a self-expanding Nitinol basket attached to a core pusher wire. Platinum markers at the proximal and distal ends of the basket facilitate visualization under fluoroscopy. A flexible radiopaque tip is attached to the distal aspect of the basket.

The ReVive™ PV Thrombectomy Device is provided as a sterile device and is intended for single patient use only. It is not intended to be resterilized and/or reused. No accessories are provided with the ReVive™ PV Thrombectomy Device. The system is used in conjunction with the appropriately sized microcatheter, as indicated in the proposed Instructions for Use.

AI/ML Overview

The provided document is a 510(k) summary for the ReVive™ PV Thrombectomy Device. It primarily focuses on demonstrating substantial equivalence to a predicate device (F.A.S.T. System SED) rather than providing detailed acceptance criteria and a study proving performance against those criteria.

The document explicitly states: "No new technological characteristics are being introduced with the proposed ReVive™ PV Thrombectomy Device and the design, materials, mechanism of action, performance specifications, and clinical utility of the device are the same as the predicate device (K092623). Codman performed non-clinical testing necessary to demonstrate substantial equivalence to the predicate device. Bench testing was conducted using the same test methods as the predicate device. Resulting data demonstrated that the ReVive™ PV Thrombectomy Device performed according to the established performance characteristics of the predicate device."

And, importantly: "No clinical studies were required as appropriate verification and validation of device modifications were achieved based on the similarities of the proposed device to the predicate device, and from results of bench testing."

Therefore, the study supporting this submission is a non-clinical bench testing study designed to demonstrate that the ReVive™ PV Thrombectomy Device performs equivalently to its predicate device across various physical and functional characteristics, and that changes (sterilization method, shelf life, labeling, packaging) do not negatively impact safety or effectiveness.

Here’s a breakdown based on the information provided, keeping in mind that the specific acceptance criteria values are not detailed, but inferred to be "met" based on equivalence to the predicate device's established performance:

1. Table of Acceptance Criteria and Reported Device Performance

CharacteristicAcceptance Criteria (Implied)Reported Device Performance
Radial ForcePerformance equivalent to predicate device (F.A.S.T. System SED)All acceptance criteria were met, demonstrating equivalence to predicate's established performance.
Tip Deflection ForcePerformance equivalent to predicate deviceAll acceptance criteria were met.
Rapid Flow RestorationPerformance equivalent to predicate deviceAll acceptance criteria were met.
CorrosionNo unacceptable corrosion; equivalent to predicate deviceAll acceptance criteria were met.
Tensile TestingStrength equivalent to predicate deviceAll acceptance criteria were met.
Track TestTracking performance equivalent to predicate deviceAll acceptance criteria were met.
Cyclic FatigueDurability equivalent to predicate deviceAll acceptance criteria were met.
Deploy and RecaptureAbility to deploy and recapture reliably, equivalent to predicate deviceAll acceptance criteria were met.
TorsionTorsional stability/resistance equivalent to predicate deviceAll acceptance criteria were met.
Force for RecaptureForce required for recapture within acceptable limits, equivalent to predicate deviceAll acceptance criteria were met.
BiocompatibilityMeets ISO 10993-1:2009 standards for Cytotoxicity, Sensitization, Pyrogenicity, Irritation, Acute Systemic Toxicity, Hemocompatibility (hemolysis, thrombogenicity, complement activation, PTT)All acceptance criteria were met (demonstrated compliance with ISO 10993-1:2009).
SterilizationValidation according to ISO 11137:2006 (e.g., bioburden, verification/confirmatory dose, B/F, LAL)All acceptance criteria were met (demonstrated compliance with ISO 11137:2006 for eBeam irradiation).
PackagingValidation according to ISO 11607-1:2006 and ISO 11607-2:2006 (e.g., integrity, dye leak, seal strength)All acceptance criteria were met (demonstrated compliance with ISO 11607 standards).
Shelf Life3 years (increased from 1 year for predicate)Achieved the 3-year shelf life.

2. Sample Size Used for the Test Set and Data Provenance

The document does not specify exact sample sizes for each non-clinical bench test. It generally states that "verification and validation testing was identified as appropriate to support device modifications."

Data Provenance: The data is from non-clinical bench testing conducted by Codman & Shurtleff, Inc. The country of origin for the data is not explicitly stated but would likely be the location of the manufacturing and R&D facilities (e.g., USA, as the company contact is in Massachusetts). The tests are prospective in the sense that they were designed and executed to validate the modified device and demonstrate equivalence.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This type of information is not applicable to this submission. The "ground truth" for the non-clinical tests is established by engineering specifications, material standards (e.g., Nitinol properties), and international standards (ISO). There are no human experts "establishing ground truth" in the way a clinical study uses expert consensus readings. The "experts" would be the engineers and quality control personnel performing and evaluating the bench tests against established physical and performance metrics.

4. Adjudication Method for the Test Set

Not applicable. Bench testing results are typically objective measurements against predefined specifications or comparative measurements against a predicate device. There is no ambiguous "case" requiring adjudication by multiple readers or experts.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

No. A MRMC comparative effectiveness study is a type of clinical study, commonly used for diagnostic imaging devices, comparing diagnostic performance among multiple readers and cases. This submission explicitly states "No clinical studies were required."

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. The ReVive™ PV Thrombectomy Device is a physical medical device (catheter with a basket) for mechanical thrombectomy, not an algorithm or AI. Its performance is inherent in its physical characteristics and how it interacts with the vascular system and thrombus, not in any computational standalone analysis.

7. The Type of Ground Truth Used

The ground truth for the non-clinical tests is based on:

  • Engineering Specifications: Designed tolerances, material properties, and functional parameters of the device.
  • International Standards: Adherence to ISO standards like ISO 10993-1 (biocompatibility), ISO 11137 (sterilization), and ISO 11607 (packaging).
  • Predicate Device Performance: The established, cleared performance characteristics and functionality of the F.A.S.T. System SED serve as the benchmark for demonstrating substantial equivalence.

8. The Sample Size for the Training Set

Not applicable. This is a physical medical device, not a machine learning model or algorithm, so there is no "training set."

9. How the Ground Truth for the Training Set Was Established

Not applicable. As there is no training set for a physical device, this question is not relevant.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the agency's name, "U.S. Food & Drug Administration," in blue text.

July 27, 2020

Codman & Shurtleff, Inc. Amarilys Machado Sr. Manager, Regulatory Affairs 325 Paramount Dr. Raynham, Massachusetts 02767-0350

Re: K132281

Trade/Device Name: ReVive PV (Peripheral Vasculature) Thrombectomy Device Regulation Number: 21 CFR 870.5150 Regulation Name: Embolectomy catheter Regulatory Class: Class II Product Code: QEW

Dear Amarilys Machado:

The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated August 30, 2013. Specifically, FDA is updating this SE Letter because FDA has created a new product code to better categorize your device technology.

Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Gregory O'Connell, OHT2: Office of Cardiovascular Devices, (301) 796-6075, Gregory. Oconnell@FDA.HHS.gov.

Sincerely,

Digitally signed by Gregory W. O'connell -S Date: 2020.07.27 08:19:38 -04'00

Gregory O'Connell Assistant Director Plaque Modification Devices Team DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image contains the logos of the U.S. Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, featuring a stylized symbol. To the right is the FDA logo, with the acronym "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

April 29, 2019

Codman & Shurtleff, Inc. Amarilys Machado Sr. Manager, Regulatory Affairs 325 Paramount Dr. Raynham, MA 02767

Re: K132281

Trade/Device Name: Revasc Thrombectomy Device Regulation Number: 21 CFR 870.5150 Regulation Name: Embolectomy Catheter Regulatory Class: Class II Product Code: OEW Dated: July 22, 2013 Received: July 23, 2013

Dear Ms. Amarilys Machado:

This letter corrects our substantially equivalent letter of August 30, 2013.

We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{2}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

GregoryO'Connell2019.04.2912:59:52 -04'00'
For
Bram D. Zuckerman, M.D.
Director
Division of Cardiovascular Devices
Office of Device Evaluation
Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

SECTION 4.

INDICATIONS FOR USE STATEMENT

510(k) Number: K132281

Device Name: ReVive™ PV Thrombectomy Device

The ReVive™ PV Thrombectomy Device is indicated for:

  • · The non-surgical removal of emboli and thrombi from peripheral blood vessels,
  • · The non-surgical removal of thrombi from synthetic grafts,
  • · Temporary use in peripheral vessel/graft occlusion,
  • Use with aspiration and with the injection or infusion of contrast media and other . fluids

Prescription Use AND/OR Over-The-Counter Use

(Part 21 CFR 801 Subpart D)

(21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Bram D. Zuckerman -S 2013.08.30 12:47:01 -04'00'

{4}------------------------------------------------

SECTION 5. 510(k) SUMMARY

(As Required By 21 CFR 807.92(a))

A. Company Information

Company Name:Codman & Shurtleff, Inc.
Address:325 Paramount DriveRaynham, MA 02767
Telephone:(305) 265-6869
Fax:(305) 265-6889
Contact Person:Amarilys Machado
Date of Submission:July 22, 2013
B. Trade/Device Name:The ReVive™ PV Thrombectomy Device
Common Name:Embolectomy Catheter
Classification Name:Embolectomy Catheter
Regulation Number:21 CFR 870.5150
Product Code:DXE

C. Predicate Device Information:

Table 1. Prior 510(k) Clearances
510(k)NumberDateClearedNameManufacturer
K092623Predicate10/30/09F.A.S.T. System SED (Self ExpandingDevice) and F.A.S.T CXD (ControlledExpansion Device)Genesis MedicalInterventional Inc.*
*On 01/18/10, certain assets of Genesis Medical Inc. were acquired by MicrusEndovascular Corporation. Effective 09/27/10, Micrus Endovascular Corporation wasacquired by Johnson & Johnson and now operates as a wholly-owned subsidiary of Codman& Shurtleff within the Johnson & Johnson family of companies.

{5}------------------------------------------------

D. Device Description:

The ReVive" PV (Peripheral Vasculature) Thrombectomy Device is a sterile, disposable, thrombectomy device that is intended to restore blood flow via non-surgical removal of emboli and thrombi from peripheral blood vessels, non-surgical removal of thrombi from synthetic grafts and also for temporary use in peripheral vessel/graft occlusion. The ReVive " PV Thrombectomy Device is radiopaque so that fluoroscopy can be used to guide insertion and verify position as necessary.

The ReVive™ PV Thrombectomy Device consists of a self-expanding Nitinol basket attached to a core pusher wire. Platinum markers at the proximal and distal ends of the basket facilitate visualization under fluoroscopy. A flexible radiopaque tip is attached to the distal aspect of the basket.

The ReVive™ PV Thrombectomy Device is provided as a sterile device and is intended for single patient use only. It is not intended to be resterilized and/or reused. No accessories are provided with the ReVive™ PV Thrombectomy Device. The system is used in conjunction with the appropriately sized microcatheter, as indicated in the proposed Instructions for Use.

An overview of the main components comprising the ReVive™ PV Thrombectomy Device is presented in Table 2, followed by illustrative representation in Figure 1.

{6}------------------------------------------------

Table 2. ReVive™ PV Thrombectomy Device Components
Component*Description
Introducer SheathThe device is packaged such that the basket is constrained within a polymerintroducer sheath. The device is designed to be front-loaded via the introducersheath into the proximal hub of a 0.027 inch inner diameter microcatheter.
Soft Distal Tip (A)A short atraumatic guidewire type element extends distal to the distal termination ofthe basket. The Soft Distal Tip also has a platinum maker affixed for visualizationpurposes.
Distal (B)/ Proximal (F)Radiopaque MarkerAs stated above, the Distal Tip serves as the distal radiopaque marker using aplatinum coil. The proximal radiopaque marker is also provided via a platinumcoil, covering the tapered portion of the core pusher wire from the proximaljunction of the basket and extending 13cm proximally.
Distal End / DistalMesh (C)The basket terminates distally to the Soft Distal Tip radiopaque marker as a meshhaving a higher strut density than the proximal end.
Basket (D)A permanently affixed self-expanding basket element. The basket is attached nearthe distal end of the pusher wire. The basket is self-expanding when deployed out ofthe microcatheter and reduces into the microcatheter when reconstrained.
Proximal End /Proximal Mesh (E)The basket terminates proximally to the core pusher wire as a mesh having a lowerstrut density than the distal end.
Core Pusher Wire (G)The proximal portion of the core pusher wire is similar to a standard core pusherwire. The remaining distal portion of the core pusher wire is tapered to a smallerdiameter and is attached to the "Basket".
*The letters following the component name refer to the letters used in Figure 1.

Image /page/6/Figure/3 description: This image is titled "Figure 1. Basket Design and Distal Tip". The image shows a basket design with labels A through G. The labels correspond to the following items: Soft Distal tip, Distal radiopaque marker, Distal end / Distal mesh, Basket, Proximal end / Proximal mesh, Proximal radiopaque marker, and Core pusher wire.

{7}------------------------------------------------

E. Intended Use:

The ReVive™ PV Thrombectomy Device is indicated for:

  • · The non-surgical removal of emboli and thrombi from peripheral blood vessels,
  • · The non-surgical removal of thrombi from synthetic grafts,
  • · Temporary use in peripheral vessel/graft occlusion,
  • Use with aspiration and with the injection or infusion of contrast media and other ● fluids

F. Summary of technological characteristics of the proposed device to the predicate device:

The ReVive™ PV Thrombectomy Device, (the basket, core pusher wire, radiopaque markings and introducer sheath), is identical to the predicate F.A.S.T System SED with regard to design, material, function, mechanism of action, clinical utility and manufacturing processes.

Compared to the predicate device, device modifications for the ReVive™ PV Thrombectomy Device include a change of sterilization method and product shelf life, and modifications to product labeling and packaging. Labeling changes are focused on rebranding the product with updated directions for use, including clarification and simplification of the product intended use and directions for use.

Table 3 provides a comparison of the regulatory information and intended use for the predicate and subject devices.

No new technological characteristics are being introduced with the proposed device. A summary table including characteristics of the proposed device compared with those of the predicate device is provided in Table 4.

{8}------------------------------------------------

CharacteristicF.A.S.T. System SEDReVive PV Thrombectomy Device
(Predicate Device)(Subject Device)
510(k) NumberK092623To Be Assigned
Clearance Date10/30/09To Be Assigned
Classification NameEmbolectomy CatheterEmbolectomy Catheter
Product CodeDXEDXE
FDA Regulation870.5150870.5150
Indication• The non-surgical removal of emboli and thrombi fromblood vessels,• The non-surgical removal of thrombi from synthetic grafts,• Temporary use in vessel/graft occlusion,• Injection, infusion, and/or aspiration of contrast media andother fluids into or from a vessel/graft, and• Catheter placement over a guidewire• The non-surgical removal of emboli and thrombifrom peripheral blood vessels,• The non-surgical removal of thrombi from syntheticgrafts,• Temporary use in peripheral vessel/graft occlusion,• Use with aspiration and with the injection or infusionof contrast media and other fluids

{9}------------------------------------------------

Table 4. Comparison of Subject and Predicate Devices: Device Characteristics
CharacteristicF.A.S.T. System SEDReVive PV Thrombectomy Device
(Predicate Device)(Subject Device)
GENERAL DESIGN
Thrombectomy MechanismCompressed basket deploymentSame as predicate device
Microcatheter compatibility0.021 inch (0.533 mm) Inner Diameter microcatheters0.027 inch (0.686 mm) Inner Diameter microcatheters
Guidewire compatibility0.014 inch systemsSame as predicate device
Introducer MechanismIntroducer sheath and core pusher wireSame as predicate device
RadiopacityProximal and distal end markersSame as predicate device
Sterilization methodEthylene OxideeBeam Irradiation
Shelf Life1 year3 years
MATERIALS
Distal TipNitinol, platinumSame as predicate device
Catheter hubNvlonSame as predicate device
Core pusher pusher WireNitinolSame as predicate device
DIMENSIONS
Basket length22 mm - 28 mm; full expansion into sheath/microcatheter(engaging portion)Same as predicate device
Basket Diameter0.6 mm (in sheath/microcatheter)Same as predicate device
Basket Working Length22 mm to 28 mm for 1.5 mm to 5.0mm vessel range.Same as predicate device
Expanded Basket Diameter4.5 mmSame as predicate device
Distal Tip Length6 mmSame as predicate device
PACKAGING
Configuration1, 5 or 10 pouched devices and one (1) IFU into one (1) cartonOne (1) device and one (1) IFU placed into one (1) carton
Protective enclosureDevice placed into PETG formed trayDevice placed into a medical grade polyethylene dispenser hoopwith a thermoplastic clip to hold the wire
Sterility barrierTyvek/polyethylene film thermo sealed pouchPolyester Film/LD Polyethylene thermo sealed pouch
LabelsInner (pouch) and outer (carton)Same as predicate device
CartonChipboard boxReverse tuck end style carton with thumb notch, 24 pointchinboard

{10}------------------------------------------------

G. Summary of Nonclinical Testing:

No new technological characteristics are being introduced with the proposed ReVive" PV Thrombectomy Device and the design, materials, mechanism of action, performance specifications, and clinical utility of the device are the same as the predicate device (K092623). Codman performed non-clinical testing necessary to demonstrate substantial equivalence to the predicate device. Bench testing was conducted using the same test methods as the predicate device. Resulting data demonstrated that the ReVive" PV Thrombectomy Device performed according to the established performance characteristics of the predicate device.

Bench Testing

In alignment with the Design Risk Assessment, verification and validation testing was identified as appropriate to support device modifications. This testing demonstrated equivalence between the proposed and predicate device, and also that modifications made had no impact to device performance.

The following verification and validation testing was conducted:

  • Radial Force Test
  • Tip Deflection Force ●
  • Rapid Flow Restoration ●
  • Corrosion ●
  • Tensile Testing
  • Track Test
  • Cyclic Fatigue
  • Deploy and Recapture
  • Torsion ●
  • Force for Recapture
  • Biocompatibility testing in accordance with ISO 10993-1: 2009 ●
    • I Cytotoxicity
    • I Sensitization
    • I Pyrogenicity
    • Irritation: Intracutaneous Reactivity
    • l Acute Systemic Toxicity
    • I Hemocompatibility: hemolysis

{11}------------------------------------------------

  • Thrombogenicity
  • I Complement Activation
  • Partial Thromboplastin Time (PTT)
  • · Sterilization method validation in accordance with ISO 11137: 2006
    • Bioburden measurement
    • Identify top three organisms
    • I Bioburden recovery study (Extraction efficiency)
    • Verification dose experiment
    • Bacteriostasis/Fungistasis (B/F) ■
    • 트 Confirmatory dose experiment (if Sterility Test fails)
    • LAL Validation I
  • Packaging Validation in accordance with ISO 11607-1: 2006 and ISO 11607-2: ● 2006
    • Package Integrity Testing
    • Dye Leak
    • Seal Strength Testing

Test results demonstrated that all acceptance criteria were met, and, therefore, the ReVive PV Thrombectomy Device conforms to expected device performance and intended use.

H. Summary of Clinical Testing:

No clinical studies were required as appropriate verification and validation of device modifications were achieved based on the similarities of the proposed device to the predicate device, and from results of bench testing.

Conclusion:

Based upon the design, materials, function, intended use, and the non-clinical testing performed by Codman & Shurtleff, Inc., it is concluded that the ReVive™ PV Thrombectomy Device is substantially equivalent to the F.A.S.T. System SED, and therefore, does not raise any new questions of safety and effectiveness.

§ 870.5150 Embolectomy catheter.

(a)
Identification. An embolectomy catheter is a balloon-tipped catheter that is used to remove thromboemboli, i.e., blood clots which have migrated in blood vessels from one site in the vascular tree to another.(b)
Classification. Class II (performance standards).