K Number
K132250
Date Cleared
2014-03-24

(248 days)

Product Code
Regulation Number
888.3800
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The INTEGRA® Freedom Wrist Arthroplasty System is indicated for intractable pain resulting from traumatic arthritis, osteoarthritis, rheumatoid arthritis, trauma-induced osteoarthritis of the radial/carpal joint and is intended to replace functionality of the joint due to deformity or elements stated above. The INTEGRA® Freedom Wrist Arthroplasty System is intended for cemented use.

Device Description

The INTEGRA® Freedom Wrist Arthroplasty (IFW) System three part semi-constrained implant system designed to replace the radiocarpal joint (distal radius and proximal row of carpal bones) and is intended to alleviate pain while restoring functionality and mobility of the joint. IFW is intended for patients with intractable pain and loss function due to traumatic arthritis, osteoarthritis, rheumatoid arthritis, and trauma-induced osteoarthritis. The system consists of three major components:

  • Radial implant made from Cobalt Chromium Molybdenum Alloy (CrCoMo) .
  • . Carpal plate - (includes two variable angle screws with locking caps) - made from Titanium Alloy
  • . Carpal Poly bearing - made from Ultra-High-Molecular-Weight Polyethylene (UHMWPe)
    Portions of the radial implant and carpal plate have a titanium plasma sprayed coating. The implant system is provided with instrumentation necessary to complete the procedures for which the system is indicated. The components are intended to be implanted together as a system, not individually as hemi-arthroplasty components and are intended for cemented fixation.
AI/ML Overview

The Integra® Freedom Wrist Arthroplasty System is a medical device, and the provided document is a 510(k) summary for its clearance. This summary details non-clinical tests performed to demonstrate substantial equivalence to previously cleared devices. It does not describe a study involving human subjects, AI systems, or the specific criteria typically associated with clinical performance evaluation of an AI/ML device.

Therefore, the requested information components related to AI/ML device studies (such as sample sizes for test/training sets, data provenance, number/qualifications of experts, adjudication methods, MRMC studies, standalone performance, and ground truth establishment for training data) are not applicable to this document. This document focuses on the mechanical and material performance of the implant itself.

Here's a breakdown of the available information based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The document states that "The results of these performance tests met their respective acceptance criteria." However, it does not provide the specific numerical acceptance criteria for each test, only that the device met them.

TestReported Device Performance
Fatigue Life TestMet acceptance criteria
Carpal Poly Bearing RemovalMet acceptance criteria
Movement to InstabilityMet acceptance criteria
Axial Screw Pull-out (ASTM F543)Met acceptance criteria
Insertion Torque (ASTM F543)Met acceptance criteria
Breaking Torque (ASTM F543)Met acceptance criteria
Breaking Angle (ASTM F543)Met acceptance criteria
Shear Fatigue Strength (ASTM F1160)Met acceptance criteria
Static Shear Strength (ASTM F1044)Met acceptance criteria
Static Tensile (ASTM F1147)Met acceptance criteria
Abrasion (ASTM F1978)Met acceptance criteria

2. Sample Size Used for the Test Set and Data Provenance

Not applicable. The tests mentioned are non-clinical (mechanical/material performance) rather than studies on a "test set" of patient data.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

Not applicable. Ground truth, in the context of AI/ML, refers to clinically validated diagnoses or outcomes. These are non-clinical engineering tests.

4. Adjudication Method for the Test Set

Not applicable. Adjudication methods are used in clinical trials or expert review processes for diagnostic accuracy, not physical device testing.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done

No, an MRMC comparative effectiveness study was not done. This is a non-clinical device clearance for a physical implant, not an AI/ML diagnostic or assistive tool.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

No, a standalone algorithm performance study was not done. This refers to the performance of an AI/ML algorithm without human intervention, which is not relevant to this physical implant.

7. The Type of Ground Truth Used

Not applicable in the AI/ML context. For these non-clinical tests, "ground truth" would be the objectively measured physical properties/performance of the device under specific test conditions (e.g., maximum load bearing before failure, precise wear rates). The standards referenced (ASTM F543, F1160, F1044, F1147, F1978) define the methods for establishing these objective measurements.

8. The Sample Size for the Training Set

Not applicable. There is no AI/ML "training set" for this physical device.

9. How the Ground Truth for the Training Set was Established

Not applicable. There is no AI/ML "training set" for this physical device.

{0}------------------------------------------------

INTEGRA® Freedom Wrist Arthroplasty System

MAR 2 4 2014

Section 5: 510(k) Summary

510(k) Summary

807.92(a)(1) - Submitter Information
NameIntegra LifeSciences Corporation
Address311 Enterprise DrivePlainsboro NJ 08536
Name of Contact PersonSally K. Wixson, VMD, RAC
Phone Number(609) 936-2454
Fax Number(609) 750 4277
EstablishmentRegistration Number3004608878
Date PreparedJuly 8, 2013
807.92(a)(2) - Name of device
Trade or Propriety NameINTEGRA® Freedom Wrist Arthroplasty System
Common or Usual NameProsthesis, wrist, 3 part metal-plastic-metal articulation,semi-constrained
Classification NameWrist joint metal/polymer semi-constrained cementedprosthesis (21CFR §888.3800)
Classification PanelOrthopedic
RegulationClass II, under 21CFR §888.3800
Product CodeJWJ
807.92(a)(3) - Legally marketed device(s) to which equivalence is claimed

Universal Total Wrist System (K030037), cleared February 05, 2003 ●

. Maestro™ Total Wrist (K042032), cleared October 7, 2004

807.92(a)(4) - Device description

The INTEGRA® Freedom Wrist Arthroplasty (IFW) System three part semi-constrained implant system designed to replace the radiocarpal joint (distal radius and proximal row of carpal bones) and is intended to alleviate pain while restoring functionality and mobility of the joint. IFW is intended for patients with intractable pain and loss function due to traumatic arthritis, osteoarthritis, rheumatoid arthritis, and trauma-induced osteoarthritis. The system consists of three major components:

  • Radial implant made from Cobalt Chromium Molybdenum Alloy (CrCoMo) .
  • . Carpal plate - (includes two variable angle screws with locking caps) - made from Titanium Alloy
  • . Carpal Poly bearing - made from Ultra-High-Molecular-Weight Polyethylene (UHMWPe)

Portions of the radial implant and carpal plate have a titanium plasma sprayed coating. The implant system is provided with instrumentation necessary to complete the procedures for which the system is indicated. The components are intended to be

{1}------------------------------------------------

Integra LifeSciences Corporation-Traditional 510(k)

INTEGRA® Freedom Wrist Arthroplasty System

Section 5: 510(k) Summary

implanted together as a system, not individually as hemi-arthroplasty components and are intended for cemented fixation.

807.92(a)(5) - Intended Use of the device
Indications for UseThe INTEGRA® Freedom Wrist Arthroplasty System is indicatedfor intractable pain resulting from traumatic arthritis, osteoarthritis,rheumatoid arthritis, trauma-induced osteoarthritis of theradial/carpal joint and is intended to replace functionality of thejoint due to deformity or elements stated above. The INTEGRA®Freedom Wrist Arthroplasty System is intended for cemented use.
807.92(a)(6) Summary of the technological characteristics of the device compared tothe predicate
The proposed INTEGRA® Freedom Wrist Arthroplasty (IFW) system and the predicatesystems (UTW and MTW) are designed to replace the radiocarpal joint and are intendedto alleviate pain while restoring functionality and mobility of the joint. The overall designand materials of the proposed device and the predicate devices are similar. The systems

consist of titanium, cobalt chrome, and Ultra-High-Molecular-Weight Polyethylene (UHMWPe) components which replace the articulation of the distal radius and proximal row of carpal bones of the wrist joint. The devices share three basic design elements in that they employ a radial component, a carpal bearing, and a carpal component. The carpal and radial components of the proposed and predicated devices have porous coated stems that are intended to be fixated into the bone with cement and utilize two screws to secure the carpal component into the carpal bones. The implant devices are provided

The INTEGRA® Freedom Wrist Arthroplasty System has similar indication for use, intended use, and fundamental scientific technology as the predicate devices (K030037 and K042032). The proposed device utilizes similar materials and design features. The differences between the predicates and proposed device do not raise any new issues regarding safety and effectiveness; therefore, The INTEGRA® Freedom Wrist Arthroplasty System is considered substantially equivalent to the predicate devices.

807.92(b)(1-2) - Nonclinical Tests Submitted

Testing to verify the performance for the The The INTEGRA® Freedom Wrist Arthroplasty System included the following:

  • Fatigue Life Test .
  • . Carpal Poly Bearing Removal

with corresponding instrumentation.

  • Movement to Instability .
  • . Screw Thread Verification:
    • O Axial Screw Pull-out, Insertion Torque, Breaking Torque and Breaking Angle per ASTM F543
  • Porous Coating Testing: . Shear Fatigue Strength per ASTM F1160 o

{2}------------------------------------------------

Integra LifeSciences Corporation-Traditional 510(k)

INTEGRA® Freedom Wrist Arthroplasty System

Section 5: 510(k) Summary

  • Static Shear Strength per ASTM F1044 o
  • Static Tensile per ASTM F1147 o
  • Abrasion per ASTM F1978 o

The results of these performance tests met their respective acceptance criteria and demonstrate that the Integra® Freedom Wrist Arthroplasty System is safe for the intended use, and is substantially equivalent to the predicate device identified.

807.92(b)(3) - Conclusions drawn from non-clinical data

The design features, materials, intended use, and overall fundamental scientific technology of the Integra® Freedom Wrist Arthroplasty System are substantially equivalent to the predicate device. The safety and effectiveness of the Integra® Freedom Wrist Arthroplasty System is adequately supported by the substantial equivalence information, materials information, and performance data provided within this Premarket Notification submission.

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized human figure with outstretched arms, representing care and protection. The figure is positioned to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA", which is arranged in a circular fashion around the figure.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

March 24, 2014

Integra LifeSciences Corporation Mr. Frederic Testa Director. Regulatory Affairs 311 Enterprise Drive Plainsboro, New Jersey 08536

Re: K132250

Trade/Device Name: INTEGRA® Freedom Wrist Arthroplasty System Regulation Number: 21 CFR 888.3800 Regulation Name: Wrist joint metal/polymer semi-constrained cemented prosthesis Regulatory Class: Class II Product Code: JWJ Dated: February 20, 2014 Received: February 21, 2014

Dear Mr. Testa:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{4}------------------------------------------------

Page 2 - Mr. Frederic Testa

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Ronald P. Jean -S for

  • Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
    Enclosure

{5}------------------------------------------------

Integra LifeSciences Corporation-Traditional 510(k) INTEGRA® Freedom Wrist Arthroplasty System

Section 4: Indications for Use

Indications for Use

K132250 510(k) Number (if known):

Device Name:

INTEGRA® Freedom Wrist Arthroplasty System

Indications For Use:

The INTEGRA® Freedom Wrist Arthroplasty System is indicated for intractable pain resulting from traumatic arthritis, osteoarthritis, rheumatoid arthritis, trauma-induced osteoarthritis of the radial/carpal joint and is intended to replace functionality of the joint due to deformity or elements stated above. The INTEGRA® Freedom Wrist Arthroplasty System is intended for cemented use.

Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Per 21 CFR 801 Subpart D) Over-The Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

AND/OR

Concurrence of CDRH, Office of Device Evaluation (ODE)

Elizabeth₪»§rank -S

Page 1 of 1

Division of Orthopedic Devices

§ 888.3800 Wrist joint metal/polymer semi-constrained cemented prosthesis.

(a)
Identification. A wrist joint metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a wrist joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have either a one-part radial component made of alloys, such as cobalt-chromium-molybdenum, with an ultra-high molecular weight polyethylene bearing surface, or a two-part radial component made of alloys and an ultra-high molecular weight polyethylene ball that is mounted on the radial component with a trunnion bearing. The metallic portion of the two-part radial component is inserted into the radius. These devices have a metacarpal component(s) made of alloys, such as cobalt-chromium-molybdenum. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).(b)
Classification. Class II.