K Number
K132180
Manufacturer
Date Cleared
2014-06-25

(345 days)

Product Code
Regulation Number
862.1345
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The IDEAL LIFE GlucoManager™ Blood Glucose Monitoring System, Model GMM0002 is intended to be used for the quantitative measurement of glucose in fresh capillary whole blood samples drawn from the fingertip or alternative sites (palm, forearm, upper arm, calf, and thigh). The system is intended to be used by a single patient and should not be shared.

The IDEAL LIFE GlucoManager™ Blood Glucose Monitoring System, Model GMM0002 is intended for self-testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The system is not to be used for the diagnosis of or screening for diabetes or for neonatal use. Alternative site testing should only be done during steady state times when blood glucose is not changing rapidly.

The IDEAL LIFE Blood Glucose Test Strips, Model AGS-1112 are for use with the IDEAL LIFE GlucoManager™ to quantitatively measure glucose in fresh capillary whole blood drawn from the fingertips, palm, forearm, upper arm, calf, and thigh.

The IDEAL LIFE Control Solution, Model GCS-0104 is intended for use with IDEAL LIFE GlucoManager™ and IDEAL LIFE Blood Glucose Test Strips, Model AGS-1112. The control solution can be used to check that the qlucose meter and test strips are working properly and that the measurement is accurate.

The IDEAL LIFE GlucoManager™ can wirelessly communicate with a communication gateway such as the IDEAL LIFE Pod™ or the IDEAL LIFE Gateway Application™. The IDEAL LIFE Gateway Application receives data wirelessly from IDEAL LIFE devices to transmit over the Internet from the user's mobile device. The Gateway Application is intended to aid people at home and health care professionals to review and evaluate historical blood glucose results, to support effective health care management.

The IDEAL LIFE Gateway Application™ makes no interpretation, evaluation, medical iudgment or recommendations for treatment. This device is not intended as a substitute for medical care.

Device Description

The IDEAL LIFE Gluco-Manager™ Blood Glucose Monitoring System, Model GMM0002 consists of a blood glucose meter, single use test strips and control solution. The new device IDEAL LIFE Gluco-Manager™, Model GMM0002 is based on an electrochemical biosensor technology (electrochemical) and the principle of capillary action. Capillary action at the end of the test strip draws the blood into the action chamber and the blood glucose result is displayed in 5 seconds. The control solution available is used to test the performance of the device. It uses the same technological characteristics for testing with its predicate device. The IDEAL LIFE Gluco-Manager™ can be used alone to measure your blood glucose. If the user would like to transmit blood glucose information for display on a personal computer, the IDEAL LIFE Gluco-Manager™ can wirelessly communicate with a communication gateway (the optional IDEAL LIFE Pod™) or by using the Gateway Application, a mobile medical application.

AI/ML Overview

The IDEAL LIFE Gluco-Manager™ Blood Glucose Monitoring System, Model GMM0002, demonstrates its performance through compliance with ISO 15197: In vitro diagnostic test systems- Requirements for blood-glucose monitoring systems for self-testing in managing diabetes mellitus. While the document asserts compliance with this standard, it does not explicitly state specific acceptance criteria values or detailed performance results in the provided text.

Therefore, I cannot directly complete Table 1 with specific acceptance criteria and reported device performance values from the provided text. The document refers to the ISO standard as the basis for performance.

Here's what can be inferred and what is missing:

1. Table of Acceptance Criteria and Reported Device Performance

Performance CharacteristicAcceptance Criteria (from ISO 15197)Reported Device Performance (from K132180)
System AccuracyISO 15197 (2003) Requirements: • For glucose concentrations < 75 mg/dL (4.2 mmol/L): ≥ 95% of results must be within ±15 mg/dL (±0.83 mmol/L) of the reference method. • For glucose concentrations ≥ 75 mg/dL (4.2 mmol/L): ≥ 95% of results must be within ±20% of the reference method.Not explicitly stated with numerical values in the provided document, but "conforms to... ISO 15197." This implies the device met the ISO 15197 criteria through its studies.
Precision (Repeatability/Reproducibility)ISO 15197 requirements (e.g., CV% or SD for different glucose levels)Not explicitly stated with numerical values in the provided document.
Interferences (Hematocrit, temperature, known interfering substances)ISO 15197 requirements for acceptable ranges and impact on resultsHematocrit Range: 20-60% (Stated as a characteristic, implying it meets the ISO standard for this range). Operating Temperature Range: 10°C to 40°C (50°-104°F) (Stated as a characteristic, implying it meets the ISO standard for this range). Other interferences not explicitly detailed.
Usability/User PerformanceISO 15197 requirements for user error rates and ease of useNot explicitly stated with numerical values in the provided document.

2. Sample size used for the test set and the data provenance:

  • Sample Size: Not explicitly stated for the test set or any performance study.
  • Data Provenance: Not explicitly stated (e.g., country of origin, retrospective/prospective). The document mentions it's for self-testing at home and the submitter is in Toronto, ON, Canada, but this doesn't clarify the study's data origin.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Not stated. For blood glucose monitoring systems, ground truth is typically established using a laboratory reference method (e.g., YSI analyzer) rather than expert consensus.

4. Adjudication method for the test set:

  • Not applicable/Not stated. Given the nature of blood glucose measurements, an adjudication method like 2+1 or 3+1 for image interpretation is not relevant. The ground truth would be based on a laboratory reference method.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • Not applicable. This device is a standalone blood glucose monitoring system, not an AI-assisted diagnostic tool for human readers.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

  • Yes. The document describes the IDEAL LIFE Gluco-Manager™ as a blood glucose monitoring system that measures glucose in fresh capillary whole blood and displays the result. It states it "can be used alone to measure your blood glucose." The wireless communication and Gateway Application are for transmitting and reviewing data, not for making interpretations, evaluations, or medical judgments. The core measurement is standalone.

7. The type of ground truth used:

  • Not explicitly stated, but for a blood glucose monitoring system claiming conformance to ISO 15197, the ground truth would invariably be established using a laboratory reference method (e.g., a YSI Glucose Analyzer) which is considered the gold standard for glucose measurement in blood.

8. The sample size for the training set:

  • Not applicable/Not stated. As a biosensor-based system, it likely does not involve a "training set" in the machine learning sense. Its performance is based on its electrochemical design and calibration, not a trained algorithm.

9. How the ground truth for the training set was established:

  • Not applicable. (See point 8).

In summary: The provided 510(k) summary for the IDEAL LIFE Gluco-Manager™ states that the device conforms to ISO 15197, which serves as the primary evidence of meeting acceptance criteria. However, explicit numerical performance data (e.g., accuracy percentages, precision metrics) and details about the study design, sample sizes, and ground truth methodology are not provided within this specific document, beyond the general statement of ISO compliance.

{0}------------------------------------------------

K132180

JUN 2 5 2014

510(k) Summary

. .

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirement of the SMDA and 21 CFR 807.92.

1. Submitter's information

ldeal Life Inc. Name: 2200 Yonge Street, Suite 1300 Address: Toronto ON M4S 2C6

.

Phone number:(416) 489-1494
Fax number:(416) 489-3009
Contact:Jason Goldberg
Date of Summary:May 16, 2014

2. Device Information

Trade name:IDEAL LIFE Gluco-Manager™ Blood Glucose Monitoring System,
Model GMM0002
Common name:Blood Glucose Monitoring System
Classification name:Blood Glucose Monitoring System

3. Classification

Product code:NBW, Blood Glucose Monitoring SystemCGA, Glucose Test SystemJQP, Calculator/data processing module for clinical useJJX, Quality Control Material (assayed and unassayed)
Regulation number:862.1345
Class:2
Panel:Clinical Chemistry

4. Predicate Device Information

Manufacturer:Andon Medical Co., Ltd.
Device:AG-608N Single Blood Glucose Monitoring System
510(k) Number:K110017

5. Device Description

The IDEAL LIFE Gluco-Manager™ Blood Glucose Monitoring System, Model GMM0002 consists of a blood glucose meter, single use test strips and control solution.

{1}------------------------------------------------

The new device IDEAL LIFE Gluco-Manager™, Model GMM0002 is based on an electrochemical biosensor technology (electrochemical) and the principle of capillary action. Capillary action at the end of the test strip draws the blood into the action chamber and the blood glucose result is displayed in 5 seconds. The control solution available is used to test the performance of the device. It uses the same technological characteristics for testing with its predicate device.

The IDEAL LIFE Gluco-Manager™ can be used alone to measure your blood glucose. If the user would like to transmit blood glucose information for display on a personal computer, the IDEAL LIFE Gluco-Manager™ can wirelessly communicate with a communication gateway (the optional IDEAL LIFE Pod™) or by using the Gateway Application, a mobile medical application.

6. Intended Use

The IDEAL LIFE GlucoManager™ Blood Glucose Monitoring System, Model GMM0002 is intended to be used for the quantitative measurement of qlucose in fresh capillary whole blood samples drawn from the fingertip or alternative sites (palm, forearm, upper arm, calf, and thigh). The system is intended to be used by a single patient and should not be shared.

The IDEAL LIFE GlucoManager™ Blood Glucose Monitoring System, Model GMM0002 is intended for self-testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The system is not to be used for the diagnosis of or screening for diabetes or for neonatal use. Alternative site testing should only be done during steady state times when blood glucose is not changing rapidly.

The IDEAL LIFE Blood Glucose Test Strips, Model AGS-1112 are for use with the IDEAL LIFE GlucoManager™ to quantitatively measure glucose in fresh capillary whole blood drawn from the fingertips, palm, forearm, upper arm, calf, and thigh.

The IDEAL LIFE Control Solution, Model GCS-0104 is intended for use with IDEAL LIFE GlucoManager™ and IDEAL LIFE Blood Glucose Test Strips, Model AGS-1112. The control solution can be used to check that the qlucose meter and test strips are working properly and that the measurement is accurate.

The IDEAL LIFE GlucoManager™ can wirelessly communicate with a communication gateway such as the IDEAL LIFE Pod™ or the IDEAL LIFE Gateway Application™. The IDEAL LIFE Gateway Application receives data wirelessly from IDEAL LIFE devices to transmit over the Internet from the user's mobile device. The Gateway Application is intended to aid people at home and health care professionals to review and evaluate historical blood glucose results, to support effective health care management.

The IDEAL LIFE Gateway Application™ makes no interpretation, evaluation, medical iudgment or recommendations for treatment. This device is not intended as a substitute for medical care.

{2}------------------------------------------------

CHARACTERISTICNEW DEVICE: IDEAL LIFEGluco-Manager™ ModelGMM0002PREDICATE DEVICE: AG-608N Single GBMS(K110017)
Detection MethodAmperometryAmperometry
EnzymeGlucose OxidaseGlucose Oxidase
Type of MeterBiosensor (Electrode)Biosensor (Electrode)
Sample SourceCapillary whole blood fromAST (Alternative site testing)and fingerCapillary whole blood fromAST (Alternative site testing)and finger
Sample ApplicationBlood sample is placeddirectly to the test strip afterfinger or AST is lancedBlood sample is placeddirectly to the test strip afterfinger or AST is lanced
Hematocrit Range20-60%20-60%
Operating Temperature Range10°C to 40°C (50°-104°F)10°C to 40°C (50°-104°F)
Dimensions77.3mm × 60.2mm ×25mm87mmx 53mmx 9.9mm
DisplayLCDLCD
Result Presentationmg/dL or mmol/Lmg/dL or mmol/L
Memory Capabilities575 times with time and datedisplay500 times with time and datedisplay
Test StartAutomaticAutomatic
Test Time5 seconds5 seconds
Power SourceDC 3V (2*AAA batteries)DC 3V (CR2032)
Battery LifeOver 200 determinantsApprox. 500 normal tests
Measurement Range20mg/dL-600mg/dL20mg/dL-600mg/dL
(1.1mmol/L~33.3mmol/L)(1.1mmol/L~33.3mmol/L)
Qualified Test StripAGS-1112 Test StripAGS-1000N Test Strip
Sample VolumeMinimum 0.7 micro literMinimum 0.7 micro liter
Other functionalityBluetooth function to uploaddata to IDEAL LIFE websiteUSB function

7. Comparison of Technological Characteristics with Predicate Device

8. Performance Summary

IDEAL LIFE Gluco-Manager™ GMM0002 Blood Glucose Monitoring System conforms to the following standards:

  • . ISO 15197: In vitro diagnostic test systems- Requirements for blood-glucose monitoring systems for self-testing in managing diabetes mellitus.
  • . IEC 61010-1: 2001, Safety requirements for electrical equipment for measurement, control, and laboratory use Part 1:General requirements.
  • . IEC 61010-2-101: 2002,Safety requirements for electrical equipment for measurement, control and laboratory use - Part 2-101: Particular requirements for in vitro diagnostic (IVD) medical equipment.
  • . EN 61326-1:2006, Electrical equipment for measurement, control and laboratory use - EMC requirements part 1: General requirements.

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles a person embracing another person, or a symbol representing care and support.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

IDEAL LIFE INC. C/O DIANE HORWITZ MANDELL HORWITZ CONSULTANTS LLC 2995 STEVEN MARTIN DRIVE FAIRFAX VA 22031

June 25, 2014

Re: K132180

Trade/Device Name: IDEAL LIFE GlucoManager™ Blood Glucose Monitoring System, Model GMM0002

Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose test system Regulatory Class: II Product Code: NBW, CGA, JQP, JJX Dated: June 16, 2014 Received: June 16, 2014

Dear Ms. Diane Horwitz:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations; Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{4}------------------------------------------------

Page 2-Mr. Horwitz

If you desire specific advice for your device on our labeling regulations (2) CFR Parts 801 and 809). please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely vours.

Katherine Serrano -S

For : Courtney H. Lias. Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K132180

Device Name

IDEAL LIFE GlucoManager™ Blood Glucose Monitoring System, Model GMM0002

Indications for Use (Describe)

The IDEAL LIFE GlucoManager™ Blood Glucose Monitoring System. Model GMM0002 is intended to be used for the quantitative measurement of glucose in fresh capillary whole blood samples drawn from the sites (nalm, forearm, upper arm, calf, and thigh). The system is intended to be used by a single patient and should not be shared.

The IDEAL LIFE GlucoManager™ Blood Glucose Monitoring System, Model GMM0002 is intended for self-itsting outside the body (in vitro diagnostic use) by people with diabetes an aid to monitor the effectiveness of diabetes control. The system is not to be used for the diagnosis of or screening for diabetes or for neonatal use. Alternative site testing should only be done during steady state times when blood glucose is not changing rapidly.

The IDEAL LIFE Blood Glucose Test Strips, Model AGS-1 1 12 are for use with the IDEAL LIFE GlucoManager™ to quantitatively measure glucose in fresh capillary whole blood drawn from the fingertips, palm, calf, and thigh.

The IDEAL LIFE Control Solution, Model GCS-0104 is intended for use with IDEAL LIFE GlucoManager™ and IDEAL LIFE Blood Glucose Test Strips, Model AGS-1112. The control solution can be used to check that the glucose meter and test strips are working properly and that the measurement is accurate.

The IDEAL LIFE GlucoManager™ can wirelessly communication gateway such as the IDEAL LIFE Pod™ or the IDEAL LIFE Gateway Application™. The IDEAL LIFE Gateway Application receives data wirelessly from IDEAL LIFE devices to transmit over the Internet from the user's mobile device. The Gateway Application is intended to aid beople at home and health care professionals to review and evaluate historical blood glucose results, to support effective health care management.

The IDEAL LIFE Gateway Application makes no interpretation, medical judement or recommendations for treatment. This device is not intended as a substitute for medical care.

Type of Use (Select one or both, as applicable).

_ Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

Stayce Beck -S

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

§ 862.1345 Glucose test system.

(a)
Identification. A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.(b)
Classification. Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.