K Number
K131965
Device Name
XVI R5.0
Manufacturer
Date Cleared
2013-09-25

(90 days)

Product Code
Regulation Number
892.5050
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Elekta X-Ray Volume Imaging system, XVI, is an electronic imaging device (EID), designed to be used with the Elekta range of medical linear accelerators and intended to be used as part of the radiation therapy treatment process for single or multiple fractions, of static and/or dynamic treatments, in gated or un-gated deliveries, in all areas of the body where such image guidance is determined by a licensed physician.

It provides real-time 2D, 3D & 4D image guidance before, during and after treatment delivery, and is intended to support confirmation of patient positioning, monitoring and management of internal motion, and decision making in response to target position, size. shape and displacement resulting from organ deformation and anatomical movement in relation to surrounding critical structures.

X VI facilitates precise and accurate dose placement, and patient set-up correction, through visualization of internal anatomy including target, critical structures and soft tissue with or without the use of implanted markers.

Symmetry™ is a software option within XVI that can be used to acquire and display volumetric images of sequential phases of the breathing cycle for the evaluation of respiration induced motion, to assist in identification of appropriate target locations within anatomical structures in motion.

The Elekta digital linear accelerator can be used for treatment that includes but is not limited to malignant and benign brain tumors, brain metastases, spine lesions treated using SRS, squamous cell carcinoma of the head and neck, lung, breast, pancreatic, hepatic malignancies treated using SBRT, prostate, and bone metastases.

Device Description

This Traditional 510/k) describes modifications made to the XVI kilo-voltage imaging accessory of the Elekta range of digital linear accelerators. The primary reason for these modifications is to further enhance the imaging tools to support the monitoring and management of motion, by allowing visualization of tumor and surrounding critical structures during treatment delivery.

Improvements have also been made to the operator interface connectivity with other systems to support distributed workflow and in the provision of licensable options to tailor individual features.

The XVI system consists of a kV radiation source mounted onto the linac gantry drum and a kV radiation image detector. Incorporation of the kV imaging system onto the same structure as the treatment system allows high quality images of the patient anatomy to be acquired at the point of treatment and their content to be spatially related to the planned MV treatment, as previously cleared under Control Number (K100115)

There has been no change made to the underlying technological characteristics of the product.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the Elekta XVI R5.0, based on the provided document:

This 510(k) summary (K131965) for Elekta XVI R5.0 does not contain specific acceptance criteria with numerical targets and reported device performance values nor details of a formal clinical study to prove these criteria. It focuses on demonstrating substantial equivalence to predicate devices and describes non-clinical performance testing.

Therefore, many of your requested points cannot be fully answered from this document. I will fill in what can be inferred or is explicitly stated and note where information is absent.


Acceptance Criteria and Device Performance

This document does not specify quantitative acceptance criteria (e.g., in terms of sensitivity, specificity, accuracy, precision, or detection rates) for the XVI R5.0. Instead, it relies on demonstrating that the device's functionality and performance are substantially equivalent to its predicates and that it meets internal requirement specifications and voluntary consensus safety standards.

Acceptance Criteria (Inferred from functionality claims)Reported Device Performance (as stated or inferred)
Visualize bony anatomy on imagePerformance is "No Change" (to predicate behavior)
Visualize soft tissue on imagePerformance is "No Change" (to predicate behavior)
Visualize implanted markers in patientPerformance is "No Change" (to predicate behavior)
Image patient in treatment positionPerformance is "No Change" (to predicate behavior)
2D image acquisitionPerformance is "No Change" (to predicate behavior)
Stereoscopic ImagingPerformance is "Enhanced"
3D image acquisitionPerformance is "No Change" (to predicate behavior)
4D image acquisitionPerformance is "No Change" (to predicate behavior)
Low (kV) radiation dose requiredPerformance is "Low (kV)"
3D Volumetric RegistrationPerformance is "No Change" (to predicate behavior)
Dual RegistrationPerformance is "No Change" (to predicate behavior)
4D Volumetric RegistrationPerformance is "No Change" (to predicate behavior)
Intra-fraction ImagingPerformance is "Yes" (new capability)
Conformance to applicable technical design specificationMet
Safety & EffectivenessAchieved
Compliance to voluntary consensus safety standardsAchieved (e.g., IEC 60601 series, IEC 62304, etc.)

Study Details

  1. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective):

    • Sample Size: Not specified. The document mentions "clinically representative conditions" and "testing... at hospital sites," but no specific number of patients, cases, or images is provided for any test set.
    • Data Provenance: Not specified. The company is based in the UK, but where the "hospital sites" are located for testing is not mentioned. It is unclear if data was retrospective or prospective.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Number of experts: Not specified.
    • Qualifications: "Competent and professionally qualified personnel" performed validation, but their specific qualifications (e.g., radiologists, oncologists, number of years experience) are not detailed.
  3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • Not specified. The document does not describe any expert adjudication process for establishing ground truth for a test set.
  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not performed/Not applicable. This document describes a device that provides image guidance for radiation therapy, not an AI-assisted diagnostic tool for human readers in the traditional sense. It's an imaging accessory for a linear accelerator. There is no mention of an MRMC study comparing human reader performance with and without this device's assistance.
  5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • Partially. The product itself is an "electronic imaging device (EID)" and its functions (image acquisition, 3D/4D registration, etc.) operate algorithmically. Non-clinical performance testing validated the "performance and functionality of the new and existing features against the requirement specification," which implies testing the algorithm's standalone capabilities. However, clinical performance in terms of patient outcomes or diagnostic accuracy without human intervention is not described, as the device is intended for human use "as part of the radiation therapy treatment process." The term "standalone" in this context might refer to testing the imaging and processing capabilities independently of patient treatment.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    • Not explicitly stated. Given the context of medical device testing for functionality and safety, ground truth for validation likely refers to:
      • Technical specifications/phantoms/known values: For verifying image quality, geometric accuracy, registration capabilities.
      • Clinical observation/expert verification during "clinically representative conditions": For assessing whether the visualization and motion management features operate as intended for patient positioning and internal motion.
    • There is no mention of pathology, post-treatment outcomes data, or a formal expert consensus process as "ground truth" for a performance study.
  7. The sample size for the training set:

    • Not applicable. This document describes a medical imaging device that acquires and processes images, not a machine learning model that requires a training set in the conventional sense (e.g., for classification or detection tasks). While the device's algorithms would have been developed and refined, the document does not refer to a "training set" for the purpose of a regulatory submission.
  8. How the ground truth for the training set was established:

    • Not applicable (as per point 7).

Summary of Limitations in the Provided Document:

The provided 510(k) summary is typical for a device demonstrating substantial equivalence, especially for modifications to an existing device (XVI R4.5) and expanding functionalities already present in predicate devices (Varian TrueBeam and Brainlab ExacTrac Vero). It emphasizes:

  • Technological characteristics: Demonstrating that changes are minor or equivalent to predicates.
  • Non-clinical performance testing: Verifying that the device meets its own functional and safety specifications.
  • Compliance with standards: Adhering to relevant industry and safety standards.

It does not include the detailed clinical study data, quantitative acceptance criteria, or robust ground truth methodologies that would be expected for a novel device or an AI/CADe (Computer-Aided Detection) device undergoing a de novo or PMA pathway, or for an AI device where improved diagnostic performance is the primary claim.

{0}------------------------------------------------

K131965
Page 1 of 3

510(k) SUMMARY

Date of preparation of summary: June 2010 2013

Submitted by:Elekta LimitedLinac House, Fleming Way, Crawley, West SussexRH10 9RR, United Kingdom
Telephone: +44 (0) 1293 654201Fax: +44 (0) 1293 658445
Contact name:Mr Andrew Hedges
Trade Name:XVI R5.0
Common Name:X-Ray Volume Imaging system, XVI
Classification Name:Medical Linear Accelerator Accessory, 21 CFR 892.5050
Product Code:IYE
Predicate Device:Elekta XVI (K100115), Varian TrueBeam (K123291),Brainlab ExacTrac Vero (K122451)

Product Description:

This Traditional 510/k) describes modifications made to the XVI kilo-voltage imaging accessory of the Elekta range of digital linear accelerators. The primary reason for these modifications is to further enhance the imaging tools to support the monitoring and management of motion, by allowing visualization of tumor and surrounding critical structures during treatment delivery.

Improvements have also been made to the operator interface connectivity with other systems to support distributed workflow and in the provision of licensable options to tailor individual features.

Intended Use and Indications for use statement:

The Elekta X-Ray Volume Imaging system, XVI, is an electronic imaging device (EID), designed to be used with the Elekta range of medical linear accelerators and intended to be used as part of the radiation therapy treatment process for single or multiple fractions, of static and/or dynamic treatments, in gated or un-gated deliveries, in all areas of the body where such image guidance is determined by a licensed physician.

It provides real-time 2D, 3D & 4D image guidance before, during and after treatment delivery, and is intended to support confirmation of patient oositioning and management of internal motion, and decision making in response to target position, size, shape and displacement resulting from organ deformation and anatomical movement in relation to surrounding critical structures.

XVI facilitates precise and accurate dose placement, and patient set-up correction, through visualization of internal anatomy including target, critical structures and soft tissue with or without the use of implanted markers.

Symmetry™ is a software option within XVI that can be used to acquire and display volumetric images of sequential phases of the breathing cvcle for the evaluation of respiration induced motion, to assist in identification of appropriate target locations within anatomical structures in motion.

The Elekta digital linear accelerator can be used for treatments that includes but is not limited to malignant and benign brain tumors, brain metastases, spine lesions treated using SRS, squamous cell carcinoma of the head and neck, lung, breast, pancreatic, hepalic malignancies treated using SBRT, prostate, and bone metastases.

{1}------------------------------------------------

Summary of Technological Characteristics:

The XVI system consists of a kV radiation source mounted onto the linac gantry drum and a kV radiation image detector. Incorporation of the kV imaging system onto the same structure as the treatment system allows high quality images of the patient anatomy to be acquired at the point of treatment and their content to be spatially related to the planned MV treatment, as previously cleared under Control Number (K100115)

There has been no change made to the underlying technological characteristics of the product.

Substantial Equivalence

The release of XVI R5.0 provides the ability to acquire KV images during treatment delivery, expanding the indications for use of its predicate device XVI R4.5 (K100115). The capability for monitoring and managing target motion and patient position during treatment is substantially equivalent to the predicate devices Varian TrueBeam (K123291) and ExacTrac Vero (K122451). Both devices provide localization to assist with patient positioning before, during and affer treatment and facilitate patient setup, through visualization of internal anatomy.

The differences in technological characteristics between XVI and the predicate devices do not raise questions of safety and effectiveness.

FunctionalityXVI R5.0XVI R4.5TrueBeam(Varian)ExacTracVero(BrainLab)
Control NumberK - - - - -K 100115K123291K122451
Visualize bonyanatomy on imageNo ChangeYesYesYes
Visualize soft tissueon imageNo ChangeYesYesYes
Visualize implantedmarkers in patientNo ChangeYesYesYes
Image patient intreatment positionNo ChangeYesYesYes
2D image acquisitionNo ChangeYesYesYes
Stereoscopic ImagingEnhancedYesYesYes
3D image acquisitionNo ChangeYesYesYes
4D image acquisitionNo ChangeYesYesNo
Radiation doserequiredLow (kV)Low (kV)Low (kV)Low(kV)
3D VolumetricRegistrationNo ChangeYesYesYes
Dual RegistrationNo ChangeYesNoNo
4D VolumetricRegistrationNo ChangeYesUnknownNo
Intra-fraction ImagingYesNoYesYes

{2}------------------------------------------------

K131965
Page 3 of 3

Summary of non clinical performance testing

Testing in the form of module, integration and system level verification was performed to evaluate the performance and functionality of the new and existing features against the requirement specification.

Regression testing has been performed successfully to verify the integrity of any changes.

Validation of the system under clinically representative conditions has been performed by competent and professionally qualified personnel. Results from verification and validation testing demonstrate that conformance to applicable technical design specification have been met and safety & effectiveness have been achieved.

Testing has been undertaken on both production equivalent systems at Elekta and at hospital sites.

The system is subject to compliance testing to voluntary consensus safety standards. Details of the standards employed in the design are specified in the Standard Data Report in section 9 which includes but not limited to IEC 60601-1, IEC 60601-2-1, IEC 60601-1-3, IEC 60601-2-44, IEC 60601-2-54, IEC 62304, IEC 62366 and ISO 14971.

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows a black and white seal of the U.S. Department of Health and Human Services. The seal features the department's symbol, which consists of a stylized caduceus-like design with three parallel lines curving upwards and ending in a flourish. The symbol is encircled by the words "U.S. DEPARTMENT OF HEALTH & HUMAN SERVICES" in a circular arrangement.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-Citifie Stivet Spring, MD 20093-0002

September 25. 2013

Elekta Limited % Mr. Andrew Hedges Regulatory Affairs Engineer Linac House, Fleming Way Crawley, West Sussex RH10 9RR UNITED KINGDOM

Rc: K131965

Trade/Device Name: XVI R5.0 Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: II Product Code: IYE Dated: June 20, 2013 Received: June 27, 2013

Dear Mr. Hedges:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments. or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you. however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820): and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{4}------------------------------------------------

Page 2-Mr. Hedges

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.lida.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Michael D. O'Hara for

Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications for Use Form

510(k) Number (if known): K131962

Device Name: XVI R5.0

Indications for Use:

The Elekta X-Ray Volume Imaging system, XVI, is an electronic imaging device (EID), designed to be used with the Elekta range of medical linear accelerators and intended to be used as part of the radiation therapy treatment process for single or multiple fractions, of static and/or dynamic treatments, in gated or un-gated deliveries, in all areas of the body where such image guidance is determined by a licensed physician.

It provides real-time 2D, 3D & 4D image guidance before, during and after treatment delivery, and is intended to support confirmation of patient positioning, monitoring and management of internal motion, and decision making in response to target position, size. shape and displacement resulting from organ deformation and anatomical movement in relation to surrounding critical structures.

X VI facilitates precise and accurate dose placement, and patient set-up correction, through visualization of internal anatomy including target, critical structures and soft tissue with or without the use of implanted markers.

Symmetry™ is a software option within XVI that can be used to acquire and display volumetric images of sequential phases of the breathing cycle for the evaluation of respiration induced motion, to assist in identification of appropriate target locations within anatomical structures in motion.

The Elekta digital linear accelerator can be used for treatment that includes but is not limited to malignant and benign brain tumors, brain metastases, spine lesions treated using SRS, squamous cell carcinoma of the head and neck, lung, breast, pancreatic, hepatic malignancies treated using SBRT, prostate, and bone metastases.

Prescription Use YES (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use NO (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Michal D. O'Hara

an Sign-Off

Page 1 of 1

(Division Sign-Off) Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health 510(k)

§ 892.5050 Medical charged-particle radiation therapy system.

(a)
Identification. A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.(b)
Classification. Class II. When intended for use as a quality control system, the film dosimetry system (film scanning system) included as an accessory to the device described in paragraph (a) of this section, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.