K Number
K131951

Validate with FDA (Live)

Date Cleared
2013-08-14

(48 days)

Product Code
Regulation Number
888.3080
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Irix-C Cervical Integrated Fusion System is a stand-alone cervical intervertebral fusion device intended for spinal fusion procedures at one level (C3 - T1 inclusive) in skeletally mature patients for treatment of degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies). Implants are to be implanted via an open, anterior approach and packed with autogenous bone graft. Patients should receive at least six (6) weeks of nonoperative treatment prior to treatment with a cervical intervertebral fusion device.

Device Description

The Irix-C Cervical Integrated Fusion System is a stand-alone intervertebral fusion device to restore biomechanical height and act as an aid in fusion of the cervical spine in anterior discectomy procedures. The device is generally boxed shaped with teeth on the superior and inferior faces of the Irix-C implant is manufactured from both titanium alloy (Ti6Al4V) in accordance with ASTM F136 and Invibio PEEK Optima LT1 in accordance with ASTM F2026, or from Ti6Al4V titanium alloy alone. The device will be supplied with the option of having the superior and inferior surfaces of the device plasma coated with medical-grade commercially pure titanium (CP Ti) per ASTM F1580. The device is then secured in location through the use of bone screws, also manufactured from titanium alloy (Ti6Al4V) per ASTM F136. The devices are provided in various sizes and screws are offered in multiple lengths to adjust for variations in patient anatomy.

AI/ML Overview

Here's an analysis of the provided text regarding the acceptance criteria and study for the Irix-C™ Cervical Integrated Fusion System:

Summary of Acceptance Criteria and Device Performance

The provided document describes the mechanical testing performed on the Irix-C Cervical Integrated Fusion System to demonstrate its substantial equivalence to predicate devices and its ability to safely and effectively perform its intended use. The acceptance criteria are implicitly defined by the standards and types of tests conducted, with the reported performance being "substantially equivalent" to predicate devices. Specific quantitative acceptance criteria or detailed numerical performance results beyond this qualitative statement are not provided in the excerpt.

1. Table of Acceptance Criteria and Reported Device Performance

TestAcceptance Criteria (Implicit)Reported Device Performance
ASTM F2077:- Static and Dynamic Compression: Withstand specified axial loads and cyclic loading without failure or excessive deformation, demonstrating structural integrity under physiological spinal forces.Substantially equivalent to predicate device performance. Capable of safely and effectively performing in accordance with its intended use.
- Static and Dynamic Torsion: Withstand specified torsional loads and cyclic torsional loading without failure or excessive deformation, simulating rotational stresses in the spine.
ASTM F2267:- Measuring Load Induced Subsidence of an Intervertebral Body Fusion Device under Static Axial Compression: Maintain construct height and integration with bone under static load, preventing subsidence into vertebrae.Substantially equivalent to predicate device performance. Capable of safely and effectively performing in accordance with its intended use.
Expulsion TestResist dislodgement or migration from the surgical site under simulated physiological forces.Tests performed and results indicate substantial equivalence.
Screw Back Out TestMaintain secure screw fixation within the device and bone under simulated physiological forces, preventing screw migration.Tests performed and results indicate substantial equivalence.
Dissociation TestMaintain integrity of the integrated fixation mechanism (device and screws) under simulated physiological forces, preventing separation of components.Tests performed and results indicate substantial equivalence.

2. Sample Size and Data Provenance for the Test Set

  • Sample Size: The document does not specify the exact number of devices tested for each mechanical test. However, it indicates that "The implant components were tested." This typically implies a statistically sufficient sample for medical device mechanical testing, often guided by standards like ASTM, but the precise number is not given.
  • Data Provenance: The data is from mechanical testing performed by the manufacturer, X-spine Systems, Inc. This is prospective testing, where the devices are manufactured and then subjected to specific test protocols in a lab setting. The country of origin of the data is implicitly the USA, where X-spine Systems, Inc. is located.

3. Number of Experts and Qualifications for Ground Truth

  • This section is not applicable to the provided document. The study described is a mechanical performance study of a medical device (spinal implant), not a study involving human interpretation of data where expert consensus for ground truth would be relevant. Ground truth in this context refers to the physical properties and performance of the device under mechanical load, measured directly by testing equipment.

4. Adjudication Method for the Test Set

  • This section is not applicable to the provided document for the same reasons as above. Adjudication methods like 2+1 or 3+1 are used in studies where human readers are making interpretations (e.g., medical image diagnosis) and disagreements need to be resolved to establish ground truth.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • No, a MRMC comparative effectiveness study was not done. The provided text describes mechanical performance testing of a physical implant device, not a study evaluating human readers' performance with or without AI assistance.

6. Standalone (Algorithm Only) Performance Study

  • No, a standalone (algorithm only) performance study was not done. This type of study would be relevant for software or AI-driven devices. The Irix-C Cervical Integrated Fusion System is a physical medical implant.

7. Type of Ground Truth Used

  • The ground truth used for this study is based on direct physical measurement and observation of the device's mechanical behavior when subjected to defined loads and conditions according to established ASTM standards (F2077, F2267) and specific proprietary tests (expulsion, screw back out, dissociation). The "ground truth" is that the device either passes or fails these pre-defined mechanical criteria, and its performance is quantifiable (e.g., load capacity, deformation, subsidence). The reported outcome is its "substantial equivalence" to predicate devices, implying its performance metrics fell within an acceptable range comparable to already approved devices.

8. Sample Size for the Training Set

  • This section is not applicable to the provided document. The study described is mechanical testing of a physical device. There is no "training set" in the context of machine learning or AI that would be relevant here.

9. How the Ground Truth for the Training Set Was Established

  • This section is not applicable to the provided document, as no training set was used.

{0}------------------------------------------------

510(k) Summary

ADMINISTRATIVE INFORMATION

Manufacturer Name:

X-spine Systems, Inc. 452 Alexandersville Rd. Miamisburg, OH 45342

Telephone (937) 847-8400 FAX (937) 847-8410

Official Contact:

David Kirschman, M.D. Chief Medical Officer

June 26, 2013

AUG 1 4 2013

Date Prepared:

DEVICE NAME

Trade/Proprietary Name: Common Name: Device Class: Regulation Number: Product Code: Classification Name:

Irix-C™ Cervical Integrated Fusion System Intervertebral body fusion device Class II $888.3080 OVE Intervertebral Fusion Device with Integrated Fixation, Cervical

ESTABLISHMENT REGISTRATION NUMBER

The X-spine Systems, Inc. establishment registration number is 3005031160. The owner/operator number for X-spine Systems, Inc. is 9063903.

INDICATIONS FOR USE

The Irix-C Cervical Integrated Fusion System is a stand-alone cervical intervertebral fusion device intended for spinal fusion procedures at one level (C3 - T1 inclusive) in skeletally mature patients for treatment of degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies). Implants are to be implanted via an open, anterior approach and packed with autogenous bone graft. Patients should receive at least six (6) weeks of nonoperative treatment prior to treatment with a cervical intervertebral fusion device.

DEVICE DESCRIPTION

The Irix-C Cervical Integrated Fusion System is a stand-alone intervertebral fusion device to restore biomechanical height and act as an aid in fusion of the cervical spine in anterior discectomy procedures. The device is generally boxed shaped with teeth on the superior and inferior faces of the Irix-C implant is manufactured from both titanium alloy (Ti6Al4V) in accordance with ASTM F136 and Invibio PEEK Optima LT1 in accordance with ASTM F2026, or from Ti6Al4V titanium alloy alone. The device will be supplied with the option of having the superior and inferior surfaces of the device plasma coated with medical-grade commercially pure titanium (CP Ti) per ASTM F1580. The device is then secured in location through the use of bone screws, also manufactured from titanium alloy (Ti6Al4V) per ASTM F136. The devices are provided in various sizes and screws are offered in multiple lengths to adjust for variations in patient anatomy.

{1}------------------------------------------------

The implant components are provided clean and non-sterile. These devices are sterilized by a healthcare professional using a steam autoclave in accordance with the instructions for use provided by X-spine Systems Inc., as well as the instructions provided by the manufacturer of the autoclave.

EQUIVALENCE TO MARKETED PRODUCT

X-spine Systems, Inc. has submitted information to demonstrate that, for the purposes of FDA's regulation of medical devices, the Irix-C Cervical Integrated Fusion System is substantially equivalent to predicate devices based on a comparison including the following characteristics:

  • . FDA Product Code
  • Indications for Use .
  • . Surgical Approach
  • . Anatomical Region
  • . Implant Materials
  • . Product Dimensions
  • . Mechanical Performance

PREDICATE DEVICES

  • . Globus Medical, Inc. - COALITION Spacer (K083389)
  • . Centinel Spine. Inc. - STALIF-C (K120819)
  • . X-spine Systems, Inc. - Calix PC Spinal Implant System (K112036)

PERFORMANCE DATA

The implant components were tested using the following standards:

ASTM F2077 - Test Methods for Intervertebral Body Fusion Devices

  • Static and Dynamic Compression .
  • Static and Dynamic Torsion .

ASTM F2267 - Standard Test Method for Measuring Load Induced Subsidence of an Intervertebral Body Fusion Device under Static Axial Compression

In addition to the above standard testing, expulsion, screw back out and dissociation tests were performed as part of this submission. There are no cited standard for these tests.

The mechanical testing results indicate that the Irix-C Cervical Integrated Fusion System is substantially equivalent to predicate device performance and is capable of safely and effectively performing in accordance with its intended use.

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center -- WO66-G609 Silver Spring, MD 20993-0002

August 14, 2013

X-Spine Systems, Incorporated David Kirschman, M.D. Chief Medical Officer 452 Alexandersville Road Miamisburg, Ohio 45342

Re: K131951

Trade/Device Name: Irix-C Cervical Integrated Fusion System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: OVE Dated: June 26, 2013 Received: June 27, 2013

Dear Dr. Kirschman:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

Image /page/2/Picture/11 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three wing-like shapes, representing health, human services, and the USA. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle symbol.

{3}------------------------------------------------

Page 2 - David Kirschman, M.D.

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/dcfault.htm.

Sincerely yours,

Erinil. Keith

For

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

510(k) Premarket Notification

Irix-C Cervical Integrated Fusion System

Indications for Use

510(k) Number (if known): K131951

Device Name: Irix-C™ Cervical Integrated Fusion System

Indications for Use:

The Irix-C Cervical Integrated Fusion System is a stand-alone cervical intervertebral fusion device intended for spinal fusion procedures at one level (C3 - T1 inclusive) in skeletally mature patients for treatment of degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies). Implants are to be implanted via an open, anterior approach and packed with autogenous bone graft. Patients should receive at least six (6) weeks of nonoperative treatment prior to treatment with a cervical intervertebral fusion device.

Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Anton E. Dmitriev, PhD Division of Orthopedic Devices

§ 888.3080 Intervertebral body fusion device.

(a)
Identification. An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.(b)
Classification. (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
Date premarket approval application (PMA) or notice of product development protocol (PDP) is required. Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.