K Number
K131858
Device Name
DIGITAL ELECTROCARDIOGRAPH
Date Cleared
2014-06-11

(352 days)

Product Code
Regulation Number
870.2340
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
Digital Electrocardiographs are intended to acquire ECG signals from adult and pediatric patients through body surface ECG electrodes. The obtained ECG records can help users to analyze and diagnose heart disease. Digital Electrocardiographs shall be used in healthcare facilities by doctors and/or trained healthcare professionals.
Device Description
Digital Electrocardiographs, E30, E65, E70 and E80, are intended to acquire ECG signals from adult and pediatric patients through body surface ECG electrodes. The obtained ECG records can help users to analyze and diagnose heart disease. There is no specific contraindication identified for the proposed devices. The proposed devices acquire ECG electrical signals from patient body surface by ECG electrodes. After been amplified, filtered and transferred, the ECG signal waveforms are displayed on the LCD and recorded on the paper through thermal printer. The components of the proposed devices, Digital Electrocardiograph, include: (1) chest electrodes (2) limb electrodes, (3) cable and (4) Batteries;
More Information

No
The description focuses on standard ECG signal acquisition, amplification, filtering, display, and printing. There is no mention of AI, ML, or advanced analytical capabilities beyond basic signal processing.

No
The device is intended for diagnosis and analysis of heart disease based on ECG signals, not for treatment.

Yes

The device, a Digital Electrocardiograph, is intended to acquire ECG signals which "can help users to analyze and diagnose heart disease." This explicitly states its role in diagnosis.

No

The device description explicitly lists hardware components such as chest electrodes, limb electrodes, cable, and batteries, indicating it is not solely software.

Based on the provided information, this device is not an IVD (In Vitro Diagnostic).

Here's why:

  • IVD Definition: In vitro diagnostics are tests performed on samples taken from the human body, such as blood, urine, or tissue, to detect diseases, conditions, or infections.
  • Device Function: The description clearly states that the device acquires ECG signals from the body surface using electrodes. This is a non-invasive measurement of electrical activity within the body, not a test performed on a sample taken from the body.
  • Intended Use: The intended use is to acquire ECG signals to help analyze and diagnose heart disease. This is a direct measurement of physiological function, not an analysis of a biological sample.

Therefore, the Digital Electrocardiograph described is a medical device used for physiological monitoring and diagnosis, but it does not fit the definition of an In Vitro Diagnostic device.

N/A

Intended Use / Indications for Use

Digital Electrocardiographs are intended to acquire ECG signals from adult and pediatric patients through body surface ECG electrodes. The obtained ECG records can help users to analyze and diagnose heart disease. Digital Electrocardiographs shall be used in healthcare facilities by doctors and/or trained healthcare professionals.

Product codes

DPS

Device Description

Digital Electrocardiographs, E30, E65, E70 and E80, are intended to acquire ECG signals from adult and pediatric patients through body surface ECG electrodes. The obtained ECG records can help users to analyze and diagnose heart disease. There is no specific contraindication identified for the proposed devices.

The proposed devices acquire ECG electrical signals from patient body surface by ECG electrodes. After been amplified, filtered and transferred, the ECG signal waveforms are displayed on the LCD and recorded on the paper through thermal printer.

The components of the proposed devices, Digital Electrocardiograph, include: (1) chest electrodes (2) limb electrodes, (3) cable and (4) Batteries;

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

Not Found

Indicated Patient Age Range

adult and pediatric patients

Intended User / Care Setting

healthcare facilities by doctors and/or trained healthcare professionals.

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies

Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:

  • IEC 60601-1, Medical Electrical Equipment - Part 1: General Requirements for Safety, 1988; Amendment 1, 1991-11, Amendment 2, 1995;
  • IEC 60601-1-2 Edition 3:2007-03, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility -Requirements and tests.
  • AAMI / ANSI EC11:1991/(R)2007, Diagnostic electrocardiographic devices.
  • ISO 10993-5:2009, Biological Evaluation of Medical Devices - Part 5: Test for in vitro cytotoxicity;
  • ISO 10993-10:2002/A2006, Biological Evaluation of Medical Devices - Part 10: Tests for irritation and delayed-type hypersensitivity;

Key Metrics

Not Found

Predicate Device(s)

K101876

Reference Device(s)

Not Found

Predetermined Change Control Plan (PCCP) - All Relevant Information

Not Found

§ 870.2340 Electrocardiograph.

(a)
Identification. An electrocardiograph is a device used to process the electrical signal transmitted through two or more electrocardiograph electrodes and to produce a visual display of the electrical signal produced by the heart.(b)
Classification. Class II (performance standards).

0

Section 3 510 Summary

Project #: M0202013 A

Section 3 510(k) Summary

This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and Title 21, CFR Section 807.92.

.'

The assigned 510(k) Number:

    1. Date of Submission: 6/11/2013
    1. Sponsor Identification

Guangdong Biolight Meditech Co., Ltd Innovation First Road, Technology Innovation Coast Zhuhai, Guangdong, 519085, China

Establishment Registration Number: 3007305624

Contact Person: Ms. Liang, Jing Position: R&D Director Tel: +86-756-3399967 Fax: +86-756-3399989 Email: liang_j@blt.com.cn

    1. Submission Correspondent
      Ms. Diana Hong & Mr. Lee Fu Mid-Link Consulting Co., Ltd P.O. Box 120-119 Shanghai, 200120, China Tel: +86-21-22815850 Fax: 240-238-7587 Email: info@mid-link.net

1

Premarket Notification 510(k) Submission

Project #: M0202013A

4. Proposed Device Identification

Proposed Device Name: Digital Electrocardiograph Models: E30, E35, E40, E65, E70 and E80 Proposed Device Common Name: Electrocardiograph

Regulatory Information: Classification Name: Electrocardiograph Classification: II; Product Code: DPS; Regulation Number: 21 CFR part 870.2340; Review Panel: Cardiovascular;

Intended Use Statement:

Digital Electrocardiographs are intended to acquire ECG signals from adult and pediatric patients through body surface ECG electrodes. The obtained ECG records can help users to analyze and diagnose heart disease. Digital Electrocardiographs shall be used in healthcare facilities by doctors and/or trained healthcare professionals.

    1. Predicate Device Identification
      510(k) Number: K101876 Predicate Device Name: Digital Electrocardiographs Manufacturer: Shenzhen Biocare Electronics Co., Ltd

Device Description 6.

Digital Electrocardiographs, E30, E65, E70 and E80, are intended to acquire ECG signals from adult and pediatric patients through body surface ECG electrodes. The obtained ECG records can help users to analyze and diagnose heart disease. There is no specific contraindication identified for the proposed devices.

The proposed devices acquire ECG electrical signals from patient body surface by ECG electrodes. After been amplified, filtered and transferred, the ECG signal waveforms are displayed on the LCD and recorded on the paper through thermal printer.

The components of the proposed devices, Digital Electrocardiograph, include: (1) chest electrodes (2) limb electrodes, (3) cable and (4) Batteries;

3-2

2

7. Non-Clinical Test Conclusion

Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:

  • ア IEC 60601-1, Medical Electrical Equipment - Part 1: General Requirements for Safety, 1988; Amendment 1, 1991-11, Amendment 2, 1995;
  • IEC 60601-1-2 Edition 3:2007-03, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility -Requirements and tests.

  • AAMI / ANSI EC11:1991/(R)2007, Diagnostic electrocardiographic devices.

  • A ISO 10993-5:2009, Biological Evaluation of Medical Devices - Part 5: Test for in vitro cytotoxicity;
  • ISO 10993-10:2002/A2006, Biological Evaluation of Medical Devices - Part 10: Tests for irritation and delayed-type hypersensitivity;

  • Substantially Equivalent (SE) Conclusion 8.

The following table compares the DEVICE to the predicate device with respect to intended use, technological characteristics and principles of operation, etc.

ItemProposed Device(s)Predicate Device(s)
Product CodeDPSDPS
Regulation Number870.2340870.2340
Intended UseDigital Electrocardiographs are
intended to acquire ECG signals
from adult and pediatric patients
through body surface ECG
electrodes. The obtained ECG
records can help users to analyze
and diagnose heart disease. Digital
Electrocardiographs shall be used
in healthcare facilities by doctors
and/or trained healthcare
professionals.Digital Electrocardiographs,
ECG-1210 / ECG-1230 / ECG-3010
/ECG-6010, are intended to acquire
ECG signals from adult and
pediatric patients through body
surface ECG electrodes. The
obtained ECG records can help
users to analyze and diagnose heart
disease. Digital Electrocardiographs
shall be used in healthcare facilities
by doctors and/or trained healthcare
professionals.
Patient Contact MaterialABS / Cooper AlloyABS / Tin Alloy
Table 3-1 Comparison of Technology Characteristics

3-3

3

:

Section 3 510 Summary

Project #: M0202013 A

SterileNoNo
Single UseNoNo
Energy SourceAC Power / DC PowerAC Power / DC Power
SafetyIEC 60601-1 / IEC 60601-1-2IEC 60601-1 / IEC 60601-1-2
BiocompatibilityNon-CytotoxicityNon-Cytotoxicity
Non-IrritationNon-Irritation
Non-SensitizationNon-Sensitization
PerformanceAAMI EC11AAMI EC11
AccessoriesLimb ElectrodeLimb Electrode
Chest ElectrodeChest Electrode
BatteryBattery

The proposed devices, Digital Electrocardiographs, E30, E35, E70 and E80, are determined to be Substantially Equivalent (SE) to the predicate devices, Digital Electrocardiographs, ECG-120 / ECG-1230 / ECG-3010 /ECG-6010. .

.

.

.

4

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/4/Picture/1 description: The image shows the seal for the Department of Health & Human Services - USA. The seal is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the seal is an abstract symbol that resembles a stylized caduceus or a representation of human figures. The image is black and white.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

June 11, 2014

Guangdong Biolight Meditech Co., Ltd. % Diana Hong Mid-Link Consulting Co., Ltd P.o. Box 120-119 Shanghai, 200120 CH

Re: K131858

Trade/Device Name: Digital Electrocardiograph (E30, E35, E40, E65, E70, E80) Regulation Number: 21 CFR 870.2340 Regulation Name: Electrocardiograph Regulatory Class: Class II Product Code: DPS Dated: April 30, 2014 Received: May 5, 2014

Dear Diana Hong:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act

5

Page 2 - Diana Hong

or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYowIndustry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Image /page/5/Picture/8 description: The image shows a logo with the letters FDA. The letters are stylized and have a bold, geometric design. There are lines drawn through the logo, possibly indicating that it is not valid or has been rejected. The logo is in black and white.

for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

6

K131858 Page 1 of 1

Section 2 Indications for Use

510(k) Number: Device Name: Digital Electrocardiograph Models: E30, E35, E40, E65, E70 and E80

Indications for Use:

Digital Electrocardiographs are intended to acquire ECG signals from adult and pediatric through body surface ECG electrodes. The obtained ECG records can help users to analyze and disease. Digital Electrocardiographs shall be used in healthcare facilities by doctors and/or trained healthcare professionals.

| ☑PRESCRIPTION USE
(Part 21 CFR 801 Subpart D) | OR | ☐OVER-THE-COUNTER USE
(21 CFR 801 Subpart C) |

-------------------------------------------------------------------------------------------------------

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Date
20.04.05.11
09.44.03 -04'00'

Page 1 of 1

2-1