(206 days)
1064 nm wavelength: Tattoo Removal (Dark Ink, Black & Blue ) Ota's nevus
532 nm wavelength: Tattoo Removal (Light Ink, Red, Sky Blue and Green) Treatment of Pigmented Lesions Solar Lentiginos Senile Lentiginos
The proposed device includes three models as YILIYA- 1064QCH, YILIYA- 1064QCL and Aeslight-1064QEH, the differences between three models described in the subsection 5 specification. The proposed device is the laser system with modularization and multi-wavelength, includes the waveform as 1064 nm and 532 nm. The proposed device consisted by control system, power supplier, cooling system, laser generator, articulated arm and foot switch.
The provided text is a 510(k) Summary for a Q-Switched ND: YAG Laser Device. This type of submission is for demonstrating substantial equivalence to a predicate device, not typically for a clinical study with acceptance criteria and performance reporting in the same way a new, high-risk device might be.
Therefore, many of the requested categories for a clinical study will not be explicitly present or applicable in this document. The submission focuses on bench tests and technological characteristics comparison to show equivalence.
Here's the information extracted and organized based on your request, with notes on what is not available in this type of submission:
1. Table of Acceptance Criteria and Reported Device Performance
Note: This document does not present acceptance criteria and performance in the "acceptance criteria" vs. "reported performance" format typically found in clinical studies validating device effectiveness. Instead, it demonstrates Substantial Equivalence (SE) to a predicate device by comparing technical specifications and ensuring compliance with recognized standards.
The table below summarizes the key technical specifications compared to the predicate device to demonstrate SE. The implicit "acceptance criteria" here is that the proposed device's specifications are within an acceptable range or are equivalent to the predicate device, especially regarding safety and intended use.
| Feature / Acceptance Criteria (Implicit for SE) | Predicate Device (RevLite™ Q-Switched Nd: YAG Laser System - K063834) | Proposed Device (YILIYA-, Aeslight- models) |
|---|---|---|
| Laser Medium | Nd: YAG | Nd: YAG |
| Wavelength | 1064 nm, 532 nm | 1064 nm, 532 nm |
| Output Energy | 1600 mJ @ 1064 nm, 500 mJ @ 532 nm | 800 mJ at max. (Wavelength not specified, implying for both) |
| Max. Energy Density | 22.64 J/cm² | 25.47 J/cm² at max. |
| Spot Size | 3, 4, 6 & 8 mm (1064 nm); 2, 3, 4 & 6 mm (532 nm) | 2-7 mm |
| Pulse Width | 5-20 ns | 6-10 ns |
| Repetition Rate | 1, 2, 5, 10 Hz | 1-10 Hz |
| Power Calibration | --- (Not explicitly detailed, likely internal) | Calibrated by laser energy/power meter no less than 1 time each year |
| Laser Class | Class 4 | Class 4 |
| Aiming Beam | Red Aiming Beam | A focusing device (viewfinder) or a red aiming beam for 1064 QEH |
| Compliance with Safety Standards | (Implicit for predicate) | IEC 60601-1, IEC 60601-1-2, IEC 60601-2-22, IEC 60825-1 |
| Intended Use | Tattoo Removal (Dark Ink, Light Ink), Treatment of Pigmented Lesions, Solar Lentiginos, Senile Lentiginos, Ota's nevus | Tattoo Removal (Dark Ink, Black & Blue, Light Ink, Red, Sky Blue and Green), Ota's nevus, Treatment of Pigmented Lesions, Solar Lentiginos, Senile Lentiginos |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size: Not applicable. This submission relies on non-clinical bench tests and comparison of technical specifications rather than a clinical human subject test set. The "test set" would be the device itself undergoing various engineering and electrical tests.
- Data Provenance: The data comes from bench tests conducted by the manufacturer, Beijing Honkon Technologies Co., Ltd. (P.R.China). This is a prospective evaluation of the newly designed device's physical and electrical characteristics against established standards and the predicate device's specifications.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
- Not applicable. Ground truth in the context of clinical expert consensus or diagnostic accuracy is not relevant here. The "ground truth" for non-clinical bench tests is defined by international safety and performance standards (e.g., IEC standards) and the specifications of the legally marketed predicate device.
4. Adjudication Method for the Test Set
- Not applicable. Adjudication is a process for resolving discrepancies among expert readers in clinical studies. This document describes non-clinical bench testing.
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance
- No. This is not an AI-assisted diagnostic device, nor is it a multi-reader, multi-case study. It's a laser device for dermatological and aesthetic procedures.
6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done
- Not applicable. This device does not involve an algorithm working in standalone mode. Its performance is evaluated through its physical and electrical outputs.
7. The Type of Ground Truth Used (Expert Consensus, Pathology, Outcomes Data, etc.)
- The "ground truth" for this submission are the international safety and performance standards (e.g., IEC 60601-1, IEC 60601-1-2, IEC 60601-2-22, IEC 60825-1) and the established technical specifications and Intended Use of the predicate device (RevLite™ Q-Switched Nd: YAG Laser System, K063834). Compliance with these standards and equivalence to the predicate's technical characteristics serves as the basis for demonstrating safety and effectiveness.
8. The Sample Size for the Training Set
- Not applicable. There is no "training set" in the context of machine learning or clinical trials for this type of device submission.
9. How the Ground Truth for the Training Set Was Established
- Not applicable L. As there is no training set, this question is not relevant.
{0}------------------------------------------------
Exhibit #4 510(k) Summary
Project #:M0272013Bd
1 6 2014 JAN
Exhibit #4 510(k) Summary
This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and 21 CFR 807.92.
The assigned 510(k) Number: __________________________________________________________________________________________________________________________________________________
- Date of Prepared: 01/11/2014 1.
-
- Sponsor
Beijing Honkon Technologies Co., Ltd. No.3 Building, No.11 Yard, Kangding Street, BDA, 100176,Beijing, P.R.China
Establishment Registration Number: Not yet registered
Contact Person: Li Zhao Position: Management Representative Tel: +86 10 56370050 ext. 508 Fax: +86 10 56370067 ext. 610 Email: small@honkonlaser.com
- Submission Correspondent 3.
Diana Hong & Tarzan. Wang Mid-Link Consulting Co., Ltd P.O. Box 237-023, Shanghai, 200237, China Tel: +86-21-22815850 Fax: 240-238-7587 Email: info@mid-link.net
{1}------------------------------------------------
- नं Proposed Device Identification
Proposed Device Name: Q-Switched ND: YAG Laser Device Common Name: Laser Device Proposed Device Model: YILIYA- 1064QCH, YILIYA- 1064QCL, Aeslight- 1064QEH
Classification: Classification Name: powered laser surgical instrument Classification: 11; Product Code: GEX: Regulation Number: 21 CFR 878.4810; Review Panel: General & Plastic Surgery;
Intended Use Statement:
1064 nm wavelength:
Tattoo Removal (Dark Ink, Black & Blue )
Ota's nevus
532 nm wavelength:
Tattoo Removal (Light Ink, Red, Sky Blue and Green)
Treatment of Pigmented Lesions
Solar Lentiginos
Senile Lentiginos
- Predicate Device Identification
510(k) Number: K063834 Product Name: RevLite™ Q-Switched Nd: YAG Laser System Manufacturer: Hoya ConBio, Inc.
Device Description 6.
.
The proposed device includes three models as YILIYA- 1064QCH, YILIYA- 1064QCL and Aeslight-1064QEH, the differences between three models described in the subsection 5 specification. The proposed device is the laser system with modularization and multi-wavelength, includes the
2/4
{2}------------------------------------------------
waveform as 1064 nm and 532 nm.
The proposed device consisted by control system, power supplier, cooling system, laser generator, articulated arm and foot switch.
-
- Non-Clinical Test Conclusion
Bench tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:
- Non-Clinical Test Conclusion
-
IEC 60601-1, Medical Electrical Equipment Part 1: General Requirements for Safety, 1988; a) Amendment 1, 1991-11, Amendment 2, 1995.
-
IEC 60601-1-2:2007, Medical Electrical Equipment Part 1-2: General Requirements for b) Safety - Collateral Standard: Electromagnetic Compatibility -- Requirements and Tests.
-
IEC 60601-2-22 (1995) Medical Electrical Equipment Part 2: Particular requirements for c) the safety of diagnostic and therapeutic.
-
IEC 60825-: 2007 Safety of laser products Part 1: Equipment classification and requirements d)
-
- Technological Characteristics Comparison
The proposed device has the same technological characteristics with the predicate device, such as Laser Medium, wavelength, control method and intended use.
-
- Substantially Equivalent Conclusion
Based on the comparison and analysis above, the proposed device is determined to be Substantially Equivalent (SE) to the predicate device.
- Substantially Equivalent Conclusion
| ITEM | Proposed Device | Predicate Device K063834 | ||||
|---|---|---|---|---|---|---|
| Product Name | YILIYA- | YILIYA- | Aeslight- | RevLite™ Q-Switched Nd: YAG | ||
| 1064QCH | 1064QCL | 1064QEH | Laser System | |||
| Laser Medium | Nd: YAG | Nd: YAG | ||||
| wavelength | 1064 nm | 1064 nm | ||||
| 532 nm | 532 nm | |||||
| Output energy | 800 mJ at max. | 1600 mJ @ 1064 nm | ||||
| 500 mJ @ 532 nm | ||||||
| EnergyMax.Density | 25.47J/cm2 at max. | 22.64J/cm2 | ||||
| Spot Size | 2-7 mm | 3. 4. 6& 8 mm (a) 1064 nm2, 3, 4 & 6 mm (a,532 nm | ||||
| Pulse Width | 6-10 ns | 5-20 ns |
| Table III-1 | Comparison Table | ||
|---|---|---|---|
3/4
{3}------------------------------------------------
·
:
.
"
| Repetition Rate | 1-10 Hz | 1, 2, 5, 10 Hz |
|---|---|---|
| Powercalibration | The device is calibrated by laserenergy/power meter no less than 1time each year | --- |
| Laser Class | Class 4 | Class 4 |
| Aiming Beam | A focusing device (viewfinder) fittedon the end instead of aiming beamAndA red aiming beam for 1064 QEH | Red Aiming Beam |
.
.
. . . . . . . . . . . .
:
.
.
:
:
4/4
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circle with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle or other bird with outstretched wings.
DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Beijing Honkon Technologies Company, Ltd. % Ms. Diana Hong Mid-Link Consulting Company, Ltd P.O. Box 120-119 Shanghai, 200120 CHINA
January 16, 2014
Re: K131857
· Trade/Device Name: YILIYA-1064QCH, YILIYA-1064QCL, and Aeslight-1064QEH Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: GEX Dated: December 2, 2013 Received: December 6, 2013
Dear Ms. Hong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability . warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{5}------------------------------------------------
Page 2 - Ms. Diana Hong
CFR Part 807): labeling (21 CFR Part 801): medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Joshua C. Nipper -S
For
Binita S. Ashar, M.D., M.B.A., F.A.C.S. Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
510(k) Number (if known) K131857
Device Name YILIYA- 1064QCH, YILIYA- 1064QCL, Aeslight- 1064QEH
Indications for Use (Describe) 1064 nm wavelength: Tattoo Removal (Dark Ink, Black & Blue ) .Ota's nevus
532 nm wavelength: Tattoo Removal (Light Ink. Red, Sky Blue and Green) Treatment of Pigmented Lesions Solar Lentiginos Senile Lentiginos
Type of Use (Select one or both, as applicable)
[x] Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FOR FOR FOR FOR FOR FOR FOR USE ONLY - CHE CHARLES - CHEAR - CHEAR
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
Image /page/6/Picture/12 description: The image shows the name "Neil R Ogden -S" in bold black font on the top line. The second line shows the date and time "2014.01.14 16:15" also in bold black font. To the right of the name is a faded logo that is difficult to make out.
FORM FDA 3881 (1/14)
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page.
PSC Publishing Services (101) 443-6740
§ 878.4810 Laser surgical instrument for use in general and plastic surgery and in dermatology.
(a)
Identification. (1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
(b)
Classification. (1) Class II.(2) Class I for special laser gas mixtures used as a lasing medium for this class of lasers. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 878.9.