K Number
K131837
Date Cleared
2014-01-22

(216 days)

Product Code
Regulation Number
878.4810
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The CO2 Laser (10600nm) is indicated for use in dermatological procedures requiring ablation, resurfacing and coagulation of soft tissue.

Device Description

The CO2 Laser consists of semiconductor solid-state lasers, switching power supply, laser power, treatment handle, key switch, cooling system and accessories. The CO2 Laser produces a beam of coherent infrared light with 10600nm wavelength. The CO2 laser comprises five models: YILIYA-10600AH, YILIYA-10600AL, YILIYA-10600CH, Aeslight-10600EH, Aeslight-Ultra peel; their technical parameters and functions are the same, the only difference is their appearances and sizes.

AI/ML Overview

The provided 510(k) Premarket Notification Submission (K131837) for the HONKON CO2 Laser focuses on establishing substantial equivalence to a predicate device (K080496). This type of submission generally relies on demonstrating that the new device has the same intended use and similar technological characteristics as a legally marketed predicate device, and that any differences do not raise new questions of safety or effectiveness.

Therefore, the "acceptance criteria" and "study that proves the device meets the acceptance criteria" in this context are primarily related to demonstrating substantial equivalence through technical comparison with the predicate device and adherence to relevant safety standards, rather than a clinical performance study with specific metrics like sensitivity, specificity, or reader improvement.

Here's a breakdown of the requested information based on the provided document:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (based on predicate equivalence and safety standards)Reported Device Performance (Proposed CO2 Laser)
Intended Use: For dermatological procedures requiring ablation, resurfacing, and coagulation of soft tissue.Met: "The CO2 Laser is indicated for use in dermatological procedures requiring ablation (removal), resurfacing and coagulation of soft tissue." (Identical to predicate)
Energy Output: Comparable to predicate (2-240mj)Met: 2-200mj (Within acceptable range or minor difference not raising new safety/effectiveness questions)
Laser Transfer Method: Articulated Arm with HandpieceMet: Articulated Arm with Handpiece (Identical to predicate)
CO₂ RF Module Maximum Power: Maximum 30w at continuous waveMet: Maximum 30w at continuous wave (Identical to predicate)
Wavelength (nm): 10600Met: 10600 (Identical to predicate)
Pulse Duration (ms): Comparable to predicate (5-100)Met: 5-100 (Identical to predicate)
Frequency (Hz): Comparable to predicate (10-200Hz)Met: 10-200Hz (Identical to predicate)
Patient Contacting Materials: AluminumMet: Aluminum (Identical to predicate)
Compatibility with environment and other devices: Comply with IEC 60601-1-2Met: Comply with the IEC 60601-1-2
Electrical Safety: Comply with IEC 60601-1Met: Comply with the IEC 60601-1
Radiation Safety: Comply with IEC 60825-1; IEC 60601-2-22Met: Comply with the IEC 60825-1; IEC 60601-2-22
Biocompatibility: Comply with ISO 10993-5; ISO 10993-10Met: Comply with the ISO 10993-5; ISO 10993-10
Other Device Specific Guidance Requirements: Adherence to "Guidance on the Content and Organization of a Premarket Notification for a Medical Laser; 21 CFR 1040.10"Met: The submission states: "The CO2 Laser and its application comply with standards as detailed in section 9, 11 and 17 of this premarket notification. Therefore, Beijing Honkon states that the non-clinical tests determined that the CO2 Laser to be as safe, as effective and performance is substantially equivalent to the predicate device(s)."

2. Sample size used for the test set and the data provenance

This is a 510(k) submission for a medical laser device, not an AI or diagnostic device that typically uses "test sets" of data for performance evaluation in the way you might describe for software.

  • Sample Size for Test Set: Not applicable in the context of clinical data. The "test set" here refers to the physical device being tested against engineering standards and compared to the predicate device's specifications. The document does not specify a "sample size" of devices tested, but rather refers to conformance of "the CO2 Laser" (implying a representative unit/design) with established standards.
  • Data Provenance: The data provenance is primarily from non-clinical testing (bench testing, engineering verification, and compliance with recognized standards) conducted by the manufacturer (Beijing Honkon Technologies Co., Ltd.) to demonstrate safety and performance according to the cited IEC and ISO standards. The country of origin for the manufacturing and testing would be China. This is a prospective evaluation in the sense that the device was manufactured and then tested to meet these criteria.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Not applicable. For a laser device like this, "ground truth" is not established by medical experts for diagnostic accuracy. Instead, safety and performance are assessed against predefined engineering specifications, international safety standards (IEC, ISO), and the performance characteristics of the predicate device. The "experts" involved would be engineers, quality control specialists, and regulatory affairs personnel who conducted and reviewed the technical testing.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable. Adjudication methods like 2+1 or 3+1 are typically used in clinical studies for diagnostic accuracy to resolve discrepancies between human readers or between human readers and AI. This submission is based on technical specifications and compliance with standards.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. An MRMC study is relevant for diagnostic devices (often AI-powered) where human readers evaluate cases. This document describes a medical laser, not a diagnostic imaging or AI system.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This is a physical medical device, not a standalone algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

The "ground truth" for this device's acceptance is based on:

  • Engineering Specifications: Conformance to the stated technical parameters (e.g., wavelength, energy output, pulse duration, frequency).
  • International Safety and Performance Standards: Compliance with recognized standards like IEC 60601-1, IEC 60601-1-2, IEC 60825-1, IEC 60601-2-22, ISO 10993-5, and ISO 10993-10. These standards define the acceptable limits and test methods for safety and performance of such devices.
  • Predicate Device Equivalence: The key "ground truth" is showing that the device's technical characteristics and intended use are substantially equivalent to a legally marketed predicate device, implying it is as safe and effective.

8. The sample size for the training set

Not applicable. This is a physical medical device, not an AI model that requires a training set.

9. How the ground truth for the training set was established

Not applicable. As there is no training set for an AI model, there is no ground truth established for it in this context.

{0}------------------------------------------------

K13183

HONKON
北京宏强富瑞技术有限公司

510(k) Premarket Notification Submission

JAN 2 2 2014

:

.

:

.

:

. .

:

: :

: . ·

510(k) Summary

In accordance with 21 CFR 807.92 the following summary of information is provided:

Date:18 June, 2013
Submitter:Beijing Honkon Technologies Co., Ltd.Address: No.3 Building, No.11 Yard, Kangding Street, BDA,100176, Beijing, P.R.China
Primary Contact Person:Mike GuAddress: 7th floor, Jingui Business Building, 982 Congyun Road,Baiyun District, 510420, Guangzhou, ChinaOSMUNDA Medical Device Consulting Co., Ltd.Tel: (+86) 20 6232 1333Fax:(+86) 208633 0253
Secondary Contact Person:Zhao LiManagement representativeBeijing Honkon Technologies Co., Ltd.Tel: (+86) 10-56370050Fax: (+86) 10-56370076-610
Trade Name:CO2 Laser (10600nm)
Common/Usual Name:CO2 Laser
Regulatory Number:878.4810
Regulation Name:Laser surgical instrument for use in general and plastic surgeryand in dermatology
Product Code:GEX
Predicate Device(s):K080496
Device Description:The CO2 Laser consists of semiconductor solid-state lasers,switching power supply, laser power, treatment handle, keyswitch, cooling system and accessories. The CO2 Laser producesa beam of coherent infrared light with 10600nm wavelength.
The CO2 laser comprises five models: YILIYA-10600AH, YILIYA-10600AL, YILIYA-10600CH, Aeslight-10600EH, Aeslight-Ultrapeel; their technical parameters and functions are the same, theonly difference is their appearances and sizes.
Intended Use:The CO2 Laser (10600nm) is indicated for use in dermatologicalprocedures requiring ablation, resurfacing and coagulation ofsoft tissue.

Confidential and Privileged. This document contains contidential and privileged trade secrets and other information of Beijing Contridential and i Whileged. "This document contribution the provinced by Beijing Honkon Technologies Co., Ltd. All rights reserved

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows a logo with the word "HONKON" in bold, black letters on the top line. Below the word, there is a line of Chinese characters that reads "北京宏强富瑞技术有限公司". The text is enclosed in a rectangular box.

510(k) Premarket Notification Submission

Technology: The CO2 laser therapy device emits a highly focused laser beam. Absorbed by water of the soft tissue, the CO2 laser incorporating the advanced technologies can immediate ablate discrete columns of tissue without charring. This device can be used in multiple specialties, such as Oral Surgery, Plastic Surgery, Dermatology.

Determinationof SubstantialEquivalence:SpecificationPredicate DeviceK080496ProposedCO2 Laser
ManufacturerLutronic CorporationHonkon
Intended UseIt is indicated for use in dermatological procedures requiring ablation (removal), resurfacing and coagulation of soft tissue.The CO2 Laser is indicated for use in dermatological procedures requiring ablation (removal), resurfacing and coagulation of soft tissue.
Energy output2-240mj2-200mj
Laser transfer methodArticulated Arm with HandpieceArticulated Arm with Handpiece
CO₂ RF ModuleMaximum PowerMaximum 30w at continuous waveMaximum 30w at continuous wave
Wavelength (nm)1060010600
Pulse duration (ms)5-1005-100
Frequency (Hz)10-200Hz10-200Hz
Patient contacting materialsAluminumAluminum
Compatibility with environment and other devicesComply with the IEC 60601-1-2Comply with the IEC 60601-1-2
Electrical SafetyComply with the IEC 60601-1Comply with the IEC 60601-1
Radiation SafetyIEC 60825-1;IEC 60601-2-22IEC 60825-1;IEC 60601-2-22
BiocompatibilityISO 10993-5;ISO 10993-10ISO 10993-5;ISO 10993-10

Confidential and Privileged. This document contidential and privileged trade secrets and other information of Heijing Honkon Technologies Co., Ltd. and as such may not be disclosed to others not employed by Beijing Honkon Technologies Co., Ltd. All rights reserved

19

{2}------------------------------------------------

510(k) Premarket Notification Submission
Other: DeviceSpecific GuidanceRequirements forComparisonGuidance on the Content andOrganization of a PremarketNotification for a MedicalLaser;21 CFR 1040.10Guidance on the Content andOrganization of a PremarketNotification for a MedicalLaser;21 CFR 1040.10
  • The CO2 Laser and its application comply with standards as detailed in section 9, 11 Conclusion: and 17 of this premarket notification. Therefore, Beijing Honkon states that the non-clinical tests determined that the CO2 Laser to be as safe, as effective and performance is substantially equivalent to the predicate device(s).
    Confidential and Privileged. This document confidential and privileged trade secrets and other information of Beijing Honkon Technologies Co., Ltd. and as such may not be disclosed to others not employed by Beijing Honkon Technologies Co., Ltd. All rights reserved

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

January 22, 2014

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

Osmunda Medical Device Consulting Co., Ltd. Mike Gu Regulatory Affairs Manager 7th floor. Jingui Business Building, 982 Congyun Road, Baiyun District, 510420, Guangzhou, China

Re: K131837

Trade/Device Name: CO2 Laser Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: II Product Code: GEX Dated: January 15, 2014 Received: January 17, 2014

Dear Mr. Gu:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{4}------------------------------------------------

Page 2 - Mr. Mike Gu

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours.

Joshua CDNhipper-S

For

Dr. Binita Ashar Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

510(k) Premarket Notification Submission

510(k) Number: K131837

Device Name: CO2 Laser (10600nm)

.

Indications for Use:

The CO2 Laser (10600nm) is indicated for use in dermatology procedures requiring ablation,
resurfacing and coagulation of soft tissue.

Prescription Use X AND/OR (Part 21 CFR 801 Subpart D)

Over-The-Counter Use_ . (Part 21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Neil R Ogden -S 2014.01.21 13:10:40
-05'00' -05'00' -05'00' -05'00' -05'00' -05'00' -05'00' -05'00' -05'00' -0

(Division Sign-Off) for BSA Division of Surgical Devices K131837 S10(k) Number __

§ 878.4810 Laser surgical instrument for use in general and plastic surgery and in dermatology.

(a)
Identification. (1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
(b)
Classification. (1) Class II.(2) Class I for special laser gas mixtures used as a lasing medium for this class of lasers. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 878.9.