(106 days)
Intended for the reduction and fixation of fractures and/or creation of a void in cancellous bone in the spine for kyphoplasty (for use with CareFusion Radiopaque Bone Cement).
The Inflatable Bone Tamp (IBT) was designed for use in balloon kyphoplasty. The balloon serves to create a cavity in the vertebral body, thereby reducing the fracture and preventing cement leakage, while still allowing for cement interdigitation. The balloon catheter is the functional part of the device that creates a cavity and reduces the fracture. The balloon catheter provides a conduit through which the physician can inflate the balloon at the distal end of the catheter.
The provided document describes the acceptance criteria and study for the CareFusion AVAmax Vertebral Balloon.
Acceptance Criteria and Reported Device Performance
| Characteristic | Acceptance Criteria (Standard/Test/FDA Guidance) | Reported Device Performance (Results Summary) |
|---|---|---|
| Inflation pressure | Constrained burst test | The balloon catheters exceeded the requirements for the minimum burst pressure in a constrained environment. |
| Inflation volume | Unconstrained burst test | The balloon catheters exceeded the requirements for the minimum burst volume in an unconstrained environment. |
| Balloon double wall thickness | Calibrated measurement | The double wall thickness of the balloons was substantially equivalent to that of the predicate devices. |
Study Details
-
Sample size used for the test set and the data provenance: Not applicable. This submission relies on non-clinical performance testing of the device itself, rather than clinical data from a "test set" of patients.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. The ground truth for the performance tests was based on engineering measurements and established test standards, not expert consensus.
-
Adjudication method for the test set: Not applicable.
-
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: No MRMC study was conducted. This device is a medical instrument, not an AI or imaging diagnostic tool.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This device is a physical medical instrument.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc.): The ground truth for the non-clinical performance tests was based on quantitative measurements against predefined engineering standards and requirements.
-
The sample size for the training set: Not applicable. No training set was used as this involved non-clinical performance testing of a physical device.
-
How the ground truth for the training set was established: Not applicable.
{0}------------------------------------------------
K131820 Page 1 of 2
OCT 0 4 2013
510(k) SUMMARY
A summary of 510(k) safety and effectiveness information in accordance with 21 CFR 807.92.
| SUBMITTER INFORMATION | |
|---|---|
| Name | CareFusion |
| Address | 1500 Waukegan Road MPWM, McGaw Park, IL 60085 USA |
| Phone number | (847) 473-7404 |
| Fax number | (312) 949-0583 |
| EstablishmentRegistration Number | 1423507 |
| Name of contact person | Joy Greidanus |
| Date prepared | June 1 , 2013 |
| NAME OF DEVICE | |
| Trade or proprietaryname | AVAmax Vertebral Balloon |
| Common or usual name | Inflatable Bone Tamp |
| Classification name | Arthroscope |
| Classification panel | Orthopedic |
| Regulation | Class Il per 21CFR §888.1100, Procode HRX:Class II per 21CFR §888.3027, Procode NDN: |
| Product Code(s) | VBT0810, VBT0815, VBT0820, VBT1010, VBT1015, VBT1020, VBT1115,VBT1120 |
| Legally marketeddevice(s) to whichequivalence is claimed | CareFusion Inflatable Bone Tamps, K103064, K093463, K090211,Radiopaque Bone Cement, K043518 |
| Reason for 510(k)submission | Modifications to Device |
| Device description | The Inflatable Bone Tamp (IBT) was designed for use in balloon kyphoplasty.The balloon serves to create a cavity in the vertebral body, thereby reducingthe fracture and preventing cement leakage, while still allowing for cementinterdigitation. The balloon catheter is the functional part of the device thatcreates a cavity and reduces the fracture. The balloon catheter provides aconduit through which the physician can inflate the balloon at the distal end ofthe catheter. |
| Intended use of thedevice | Intended for the reduction and fixation of fractures and/or creation of a void incancellous bone in the spine for kyphoplasty (for use with CareFusionRadiopaque Bone Cement) |
.
{1}------------------------------------------------
| SUMMARY OF THE TECHNOLOGICAL CHARACTERISTICS OF THE DEVICE COMPAREDTO THE PREDICATE DEVICE | ||
|---|---|---|
| Characteristic | New Device | PredicateCareFusion Vertebral Balloons (K103064, K093463, K090211) |
| Compatible cannula size | 8G, 10G, 11G | 8G, 10G, 11G |
| Balloon inflation medium | 60% contrast recommended | 60% contrast recommended |
| Balloon and catheter materials | Polyurethane | Polyurethane |
| Wire mandrel material | Stainless steel | Stainless steel |
| Balloon shape | Cylindrical | Cylindrical |
| Maximum recommended inflation pressure | 400 psi (27 ATM) | 400 psi (27 ATM) |
| Maximum recommended inflation volume | 4-6 mL | 4-6 mL |
| PERFORMANCE DATA | ||
| SUMMARY OF NON-CLINICAL TESTS CONDUCTED FOR DETERMINATION OFSUBSTANTIAL EQUIVALENCE | ||
| Performance Test Summary-New Device | ||
| Characteristic | Standard/Test/FDA Guidance | Results Summary |
| Inflation pressure | Constrained burst test | The balloon catheters exceeded therequirements for the minimum burst pressure in aconstrained environment |
| Inflation volume | Unconstrained burst test | The balloon catheters exceeded therequirements for the minimum burst volume in anunconstrained environment |
| Balloon double wall thickness | Calibrated measurement | The double wall thickness of the balloons wassubstantially equivalent to that of the predicatedevices |
| SUMMARY OF CLINICAL TESTS CONDUCTED FOR DETERMINATION OF SUBSTANTIALEQUIVALENCE AND/OR OF CLINICAL INFORMATION | ||
| N/A - No clinical tests were conducted for this submission | ||
| CONCLUSIONS DRAWN FROM NON-CLINICAL AND CLINICAL DATA | ||
| The results of the non-clinical tests show that the CareFusion balloon catheters meet or exceed allperformance requirements, and are substantially equivalent to the predicate devices. |
SUMMARY OF THE TECHNOLOGICAL CHARACTERISTICS OF THE DEVICE COMPARED
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of an eagle or bird-like figure with three wavy lines extending from its body, representing the flow of services to people. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the emblem.
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public I leath Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WD66-G609 Silver Spring, MID 20993-0002
October 4, 2013
CareFusion Ms. Joy Greidanus Manager, Regulatory Affairs 75 North Fairway Drive Vernon Hills. Illinois 6006 l
Re: K131820
Trade/Device Name: AVAmax Vertebral Balloon Regulation Number: 21 CFR 888.3027 Regulation Name: Polymethylmethacrylate (PMMA) bone cement Regulatory Class: Class II Product Code: NDN, HRX Dated: August 26, 2013 Received: August 28, 2013
Dear Ms. Greidanus:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28. 1976. the enactment date of the Medical Device Amendments, or 10 devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Ilisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warrantics. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{3}------------------------------------------------
Page 2 - Ms. Joy Greidanus
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 331-542 of the Act): 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers. International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803). please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/McdicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Erin Keith
lior
Mark N. Melkerson IDirector Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
CareFusion - June 2013 - Traditional 510(k): Inflatable Bone Tamps
Image /page/4/Picture/1 description: The image contains the logo for CareFusion. The logo consists of a circular graphic on the left and the word "CareFusion" on the right. The circular graphic appears to have a stylized design with concentric circles and a shape resembling a heart or shield in the center.
1500 Waukegan Road McGaw Park, Illinois 60085-6787 PH: (847) 473-7404 FAX: (312) 949-0583
510(k) Number (if known): K131820
Device Name:
AVAmax Vertebral Balloon
Indications For Use:
Intended for the reduction and fixation of fractures and/or creation of a void in cancellous bone in the spine for kyphoplasty (for use with CareFusion Radiopaque Bone Cement).
Prescription Use_X_ (Per 21 CFR 801 Subpart D) __ (Per 21 CFR 807 Subpart C) And/Or Over-The Counter Use_ (Please Do not WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Laurence D. Coyne -S
(Division Sign-Off) Division of Orthopedic Devices 510(k) Number: K131820
§ 888.1100 Arthroscope.
(a)
Identification. An arthroscope is an electrically powered endoscope intended to make visible the interior of a joint. The arthroscope and accessories also is intended to perform surgery within a joint.(b)
Classification. (1) Class II (performance standards).(2) Class I for the following manual arthroscopic instruments: cannulas, currettes, drill guides, forceps, gouges, graspers, knives, obturators, osteotomes, probes, punches, rasps, retractors, rongeurs, suture passers, suture knotpushers, suture punches, switching rods, and trocars. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 888.9.