(112 days)
Caregiver professional Clinical Thermometer is an infrared thermometer intended for the measurement of human body temperature in people of all ages without contact to the body and may be used by medical professionals or by consumers in a home environment.
Caregiver Professional Clinical Thermometer (Model PRO-TF300) is characterized by measuring human body temperature from the surface of human skin. It utilizes infrared technology to measure infrared energy emitted from the skin surface when making a temperature measurement.
The provided text describes the 510(k) summary for the Caregiver Professional Clinical Thermometer (Model PRO-TF300). Here's a breakdown of the acceptance criteria and study information:
Acceptance Criteria and Reported Device Performance
The device's performance was validated according to ASTM E1965-98. The specific acceptance criteria are not explicitly detailed as numerical thresholds in this document, but the tests performed and their outputs are listed.
| Item | Standard Complied | Data Presentation |
|---|---|---|
| Laboratory accuracy | ASTM E1965-98 | Measurement error |
| Clinical accuracy | ASTM E1965-98 | Clinical bias |
| Clinical repeatability | ||
| Storage stability | ASTM E1965-98 | Measurement error |
| Shock | ASTM E1965-98 | Measurement error |
| Cleaning procedure | ASTM E1965-98 | Measurement error |
| Safety | IEC 60601-1 | Evaluated by SGS |
| Electromagnetic | IEC 60601-1-2 | Evaluated by SGS |
| compatibility (EMC) |
Study Details
Due to the nature of the device (a clinical thermometer), many typical AI/ML study components (like expert consensus, adjudication, MRMC studies, or training/test set provenance) are not directly applicable or documented in this type of 510(k) submission.
-
Sample size used for the test set and the data provenance:
- Specific sample sizes for the clinical accuracy and laboratory accuracy tests are not provided in this summary.
- The data provenance (country of origin, retrospective/prospective) is not specified in this summary.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- For a clinical thermometer, "ground truth" would typically refer to a reference temperature measurement (e.g., from a rectal thermometer or an accredited blackbody calibrator). The document does not specify the number or qualifications of experts involved in establishing this reference.
-
Adjudication method (e.g., 2+1, 3+1, none) for the test set:
- Adjudication methods are generally used for subjective assessments (like image interpretation). For objective measurements like temperature, adjudication as typically understood for AI models is not applicable and therefore not mentioned.
-
If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- This is a standalone diagnostic device, not an AI-assisted interpretation tool for human readers. Therefore, an MRMC comparative effectiveness study is not applicable and was not performed.
-
If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
- Yes, the performance tests (laboratory accuracy, clinical accuracy) described were for the device in a standalone capacity, meaning its performance was evaluated inherently. The device itself is a measurement tool, not an AI algorithm in the contemporary sense.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- For laboratory accuracy, the ground truth would typically be an accredited blackbody calibrator or similar highly accurate reference thermometer.
- For clinical accuracy, the ground truth would typically be core body temperature measured by a reference method (e.g., rectal thermometer). The document does not explicitly state the reference method used.
-
The sample size for the training set:
- This device is a traditional infrared thermometer and does not involve AI/ML requiring a training set. Therefore, this concept is not applicable.
-
How the ground truth for the training set was established:
- As there is no training set for an AI/ML algorithm involved, this question is not applicable.
{0}------------------------------------------------
ОСТ 0 7 2013
B2. 510(k) Summary
510(K) SUMMARY
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92.
The Assigned 510(k) number is:
- l. Submitter Information: Application Correspondence: Contact Person: Paul Liu Address: 6F, No. 127, Wugong 2nd Rd., Wugu District, New Taipei City, 24888, Taiwan Phone: +886-2-6625-8188 #6134 Fax: +886-2-6625-0868 Email: paul@taidoc.com.tw
Date of submission: June 10, 2013
Applicant: Company Name: Thermomedics Inc. Contact Person: Gary J O'Hara Address: 18310 Calle La Serra Rancho Santa Fe, CA 92091 Phone: 858 779 1060 Fax: 305 433 5129 E-mail: gohara@thermomedics.com
-
- Device name:
·
- Device name:
Proprietary name: Caregiver Professional Clinical Thermometer Model no: PRO-TF Series (Model PRO-TF300) Regulatory information:
Regulation section: 21 CFR 880.2910 A.
- B. Classification: Class II
- Product Code: FLL, Clinical electronic thermometer C.
Page 1 of 3
{1}------------------------------------------------
- D. Panel: General Hospital (80)
-
- Intended Use:
Caregiver professional Clinical Thermometer is an infrared thermometer intended for the measurement of human body temperature in people of all ages without contact to the body and may be used by medical professionals or by consumers in a home environment.
-
- Device Description:
Caregiver Professional Clinical Thermometer (Model PRO-TF300) is characterized by measuring human body temperature from the surface of human skin. It utilizes infrared technology to measure infrared energy emitted from the skin surface when making a temperature measurement.
- Device Description:
Substantial Equivalence Information:
- Predicate device name: A. U-RIGHT TD-1240 Thermometer
- B. Predicate K number: K113159
- C. Comparison with predicate:
Caregiver Professional Clinical Thermometer (Model PRO-TF300) has the following similarities to the predicate device:
- same operating principle, 그
- same fundamental scientific technology,
- 베 incorporate the same basic circuit design,
- 문 incorporate the same materials,
- 그 same shelf life
- manufactured by the same process 1
The modifications encompass:
- displayed temperature range 그
{2}------------------------------------------------
- memory storage capacity
- modification in the physical appearance
- 글 a modification in the user interface
- 트 labeling change due to the modifications
-
- Test Principle:
The thermometer measures temperature by reading infrared radiation emitting from the skin and converts it into a temperature value.
-
- Performance Characteristics:
Caregiver Professional Clinical Thermometer (Model PRO-TF300) was validated by the tests according to ASTM E1965-98.
- Performance Characteristics:
A brief description for each test was given in this section. Table 1 lists items of tests related standard complied and format of data presentation.
| Item | Standard complied | Data presentation | Attachment |
|---|---|---|---|
| Laboratory accuracy | ASTM E1965-98 | Measurement error | A3.1 |
| Clinical accuracy | ASTM E1965-98 | Clinical biasClinical repeatability | A3.2 |
| Storage stability | ASTM E1965-98 | Measurement error | A3.3 |
| Shock | ASTM E1965-98 | Measurement error | A3.4 |
| Cleaning procedure | ASTM E1965-98 | Measurement error | A3.5 |
| Safety | IEC 60601-1 | Evaluated by SGS | A3.6 |
| Electromagnetic compatibility (EMC) | IEC 60601-1-2 | Evaluated by SGS | A3.7 |
Table 1. Summary of test
7. Conclusion:
Based on the information provided in this submission, the Caregiver Professional Clinical Thermometer (Model PRO-TF300) is substantially equivalent to the predicate U-RIGHT TD-1240 Thermometer.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three tail feathers, representing the department's commitment to health, human services, and well-being. The eagle is encircled by the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" in a circular arrangement.
Public Health Service
Food and Drue Administration 0903 New Hampshire Avenue Document Control Center - WObb-G Iver Spring, MD 20993-0002
October 7, 2013
Thermomedics, Incorporated Mr. Paul Liu 6F, NO. 127, Wugong 2nd Road, Wugu District New Taipei City TAIWAN 24888
Re: K131771
Trade/Device Name: Caregiver Professional Clinical Thermometer, PRO-TF Series (Model PRO-TF300) Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical electronic thermometer Regulatory Class: II Product Code: FLL Dated: September 6, 2013 Received: September 9, 2013
Dear Mr. Liu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Mr. Liu
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers. International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office. of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Mary's Runner-S
Kwame Ulmer M.S. Acting Division Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
B1. Indications for Use
Indications for Use
510(k) Number: K131771
Caregiver Professional Clinical Thermometer, PRO-TF Series (Model PRO-TF300)
Indications for Use:
Caregiver professional Clinical Thermometer is an infrared thermometer intended for the measurement of human body temperature in people of all ages without contact to the body and may be used by medical professionals or by consumers in a home environment.
And/Or Over the Counter Use __ X ____________________________________________________________________________________________________________________________________________________ Prescription Use _____________________________________________________________________________________________________________________________________________________________ (21 CFR Part 801 Subpart D) (21 CFR Part 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Saijad H. Sye
Division Sign-Off Office of Device Evaluation (ODE) 510(k)________________________________________________________________________________________________________________________________________________________________________
Page I of ____________________________________________________________________________________________________________________________________________________________________
§ 880.2910 Clinical electronic thermometer.
(a)
Identification. A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.(b)
Classification. Class II (performance standards). The device is exempt from the premarket notification procedures in part 807, subpart E of this chapter, subject to the limitations in § 880.9 and the following conditions for exemption:(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.