K Number
K131763
Device Name
TRUSCOPE SERIES PATIENT MONITOR
Date Cleared
2013-07-15

(28 days)

Regulation Number
870.1025
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
Truscope Elite A8, Truscope Elite A6, Truscope Elite A5 and Truscope Elite A3: The Truscope series patient monitors (Truscope Elite A8, Truscope Elite A6, Truscope Elite A5 and Truscope Elite A3) is intended to be used for monitoring, displaying, reviewing, storing and alarming of multiple physiological parameters including ECG (3-lead or 5-lead or 12-lead selectable), arrhythmia detection, ST segment analysis, Heart Rate (HR), Respiration Rate (RESP), Temperature (TEMP), Pulse Oxygen Saturation (SpO2), Pulse Rate (PR), Non-invasive Blood Pressure (NIBP), Invasive Blood Pressure (IBP), Carbon dioxide (CO2), Anesthetic Gas (AG), Impedance Cardiograph (ICG) and Cerebral State Index (CSI). The arrhythmia detection, ST segment analysis only applied to a single adult patient. The monitors are to be used in healthcare facilities by clinical physicians or appropriate medical staff under the direction of physician. It is not intended for helicopter transport, hospital ambulance, or home use. Truscope II and MINI: Truescope Series Patient Monitor (Truscope II and MINI) is intended to be used for monitoring. displaying, reviewing, storing and alarming of multiple physiological parameters including ECG Heart Rate (HR), Respiration Rate (RESP), Pulse Oxygen Saturation (SpO2), Pulse Rate (PR), Non-invasive Blood Pressure (NIBP), Invasive Blood Pressure (IBP), carbon dioxide (CO2), and Temperature (TEMP) of adult, pediatric and neonatal patient. Truescope Series Patient Monitor (Truscope II and MINI) is to be used in healthcare facilities by clinical physicians or appropriate medical staff under the direction of physicians. It is not intended for helicopter transport or hospital ambulance. Note: Truscope MINI does not have functions of Invasive Blood Pressure and Carbon Dioxide Truscope Classic: Truescope Series Patient Monitor (Truscope Classic) is intended to be used for monitoring, displaying, reviewing, storing and alarming of multiple physiological parameters including ECG Heart Rate (HR), Respiration Rate (RESP), Pulse Oxygen Saturation (SpO2). Pulse Rate (PR), Non-invasive Blood Pressure (NIBP), Invasive Blood Pressure (IBP), carbon dioxide (CO2), and Temperature (TEMP) of adult, pediatric and neonatal patient. The monitors are to be used in healthcare facilities by clinical physicians or appropriate medical staff under the direction of physicians. They are not intended for helicopter transport or hospital ambulance.
Device Description
The proposed device Truscope Patient Monitor is change from the Paitent Monitor manufactured by Guangdong Biolight Meditech Co., Ltd, all models included in this submission has exactly same design, materials, and manufacture process with their original model, only different is the model name and device name.
More Information

Not Found

No
The summary describes a standard patient monitor that measures and displays physiological parameters. There is no mention of AI, ML, or any advanced analytical capabilities beyond basic signal processing for parameters like arrhythmia detection and ST segment analysis, which are common in patient monitors and do not inherently require AI/ML. The "Mentions AI, DNN, or ML" section explicitly states "Not Found".

No
The device is a patient monitor intended for monitoring, displaying, reviewing, storing, and alarming of physiological parameters. It does not provide any therapeutic intervention.

No

The device is described as a "patient monitor" which is intended for "monitoring, displaying, reviewing, storing and alarming of multiple physiological parameters." While patient monitors provide data that can inform a diagnosis, they do not, by themselves, perform a diagnostic function. Their purpose is to observe and report physiological states, not to identify or determine a disease or condition.

No

The device description explicitly states "The proposed device Truscope Patient Monitor is change from the Paitent Monitor manufactured by Guangdong Biolight Meditech Co., Ltd, all models included in this submission has exactly same design, materials, and manufacture process with their original model". This indicates a physical hardware device, not a software-only solution. The performance studies also reference standards for physical medical devices (IEC 60601-1, etc.).

Based on the provided text, this device is not an IVD (In Vitro Diagnostic).

Here's why:

  • Intended Use: The intended use clearly states that the device is for "monitoring, displaying, reviewing, storing and alarming of multiple physiological parameters" directly from the patient (ECG, respiration, temperature, SpO2, blood pressure, etc.).
  • IVD Definition: In vitro diagnostics are tests performed on samples taken from the human body, such as blood, urine, or tissue, to detect diseases, conditions, or infections. This device does not perform such tests on samples.
  • Device Description: The description focuses on the device being a patient monitor with specific physiological parameter monitoring capabilities.
  • Lack of IVD-related information: There is no mention of analyzing biological samples, reagents, or any other components typically associated with IVD devices.

The device is a patient monitor that directly measures physiological signals from the patient's body.

N/A

Intended Use / Indications for Use

Truscope Elite A8, Truscope Elite A6, Truscope Elite A5 and Truscope Elite A3

The Truscope series patient monitors (Truscope Elite A8, Truscope Elite A6, Truscope Elite A5 and Truscope Elite A3) is intended to be used for monitoring, displaying, reviewing, storing and alarming of multiple physiological parameters including ECG (3-lead or 5-lead or 12-lead selectable), arrhythmia detection, ST segment analysis, Heart Rate (HR), Respiration Rate (RESP), Temperature (TEMP), Pulse Oxygen Saturation (SpO2), Pulse Rate (PR), Non-invasive Blood Pressure (NIBP), Invasive Blood Pressure (IBP), Carbon dioxide (CO2), Anesthetic Gas (AG), Impedance Cardiograph (ICG) and Cerebral State Index (CSI).

The arrhythmia detection, ST segment analysis only applied to a single adult patient.

The monitors are to be used in healthcare facilities by clinical physicians or appropriate medical staff under the direction of physician. It is not intended for helicopter transport, hospital ambulance, or home use.

Truescope Series Patient Monitor (Truscope II and MINI) is intended to be used for monitoring. displaying, reviewing, storing and alarming of multiple physiological parameters including ECG Heart Rate (HR), Respiration Rate (RESP), Pulse Oxygen Saturation (SpO2), Pulse Rate (PR), Non-invasive Blood Pressure (NIBP), Invasive Blood Pressure (IBP), carbon dioxide (CO2), and Temperature (TEMP) of adult, pediatric and neonatal patient.

Truescope Series Patient Monitor (Truscope II and MINI) is to be used in healthcare facilities by clinical physicians or appropriate medical staff under the direction of physicians. It is not intended for helicopter transport or hospital ambulance.

Note: Truscope MINI does not have functions of Invasive Blood Pressure and Carbon Dioxide

Truescope Series Patient Monitor (Truscope Classic) is intended to be used for monitoring, displaying, reviewing, storing and alarming of multiple physiological parameters including ECG Heart Rate (HR), Respiration Rate (RESP), Pulse Oxygen Saturation (SpO2). Pulse Rate (PR), Non-invasive Blood Pressure (NIBP), Invasive Blood Pressure (IBP), carbon dioxide (CO2), and Temperature (TEMP) of adult, pediatric and neonatal patient.

The monitors are to be used in healthcare facilities by clinical physicians or appropriate medical staff under the direction of physicians. They are not intended for helicopter transport or hospital ambulance.

Product codes

DSI, MLD, DRT, DXN, DSK, FLL, DQA, CCK, CBQ, CBS, CCL, GWQ, DSB

Device Description

The proposed device Truscope Patient Monitor is change from the Paitent Monitor manufactured by Guangdong Biolight Meditech Co., Ltd, all models included in this submission has exactly same design, materials, and manufacture process with their original model, only different is the model name and device name.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

Not Found

Indicated Patient Age Range

adult, pediatric and neonatal patient.

Intended User / Care Setting

healthcare facilities by clinical physicians or appropriate medical staff under the direction of physician.

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

Bench tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:

  • IEC 60601-1, Medical Electrical Equipment Part 1: General Requirements for Safety, 1988; a) Amendment 1, 1991-11, Amendment 2, 1995.
  • b) IEC 60601-1-2:2007, Medical Electrical Equipment - Part 1-2: General Requirements for Safety - Collateral Standard: Electromagnetic Compatibility -- Requirements and Tests.
  • AAMI EC13:2002, Cardiac monitors, heart rate meters, and alarms; c)
  • AAMI EC57:1998/(R) 2008, Testing and reporting performance results of cardiac rhythm and d) ST-segment measurement algorithms.
  • AAMI SP10:2002/ A1:2003 (R) 2008, Manual, electronic or automated sphygmomanometers; e)
  • () ISO 9919:2005, Medical electrical equipment - Particular requirements for the basic safety and essential performance of pulse oximeter equipment for medical use;

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

Not Found

Predicate Device(s)

K120193, K102040, K100046

Reference Device(s)

Not Found

Predetermined Change Control Plan (PCCP) - All Relevant Information

Not Found

§ 870.1025 Arrhythmia detector and alarm (including ST-segment measurement and alarm).

(a)
Identification. The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.(b)
Classification. Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.

0

K131763

b 1/5

JUL 1 5 2013

Section 3 510(k) Summary

This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and 21 CFR 807.92.

The assigned 510(k) Number:

    1. Date of Prepared: May. 15, 2013
  • Sponsor 2.

Guangdong Biolight Meditech Co., Ltd Innovation First Road, Technology Innovation Coast Zhuhai, Guangdong, 519085, China

Establishment Registration Number: 3007305624

Contact Person: Mr. Tianbao Li, Chief Engineer T: +86-756-3399963 F:+86-756-3399989 E:li_tb@blt.com.cn

    1. Submission Correspondent
      Diana Hong & Tarzan. Wang Mid-Link Consulting Co., Ltd P.O. Box 237-023, Shanghai, 200237, China Tel: +86-21-22815850 Fax: 240-238-7587 Email: info@mid-link.net

1

K131763 P 2/5

Proposed Device Identification 4.

Proposed Device Name: Truscope Patient Monitors

Device Common Name: Patient Monitors

Proposed Device Model: Truscope Elite A8, Truscope Elite A6, Truscope Elite A5, Truscope Elite A3, Truscope Classic, Truscope II, and Truscope MINI;

Classification:

Regulation No.Classification Name.Product CodeDevice Class
Main Code
21 CFR 870.1025Arrhythmia detector and alarm (including ST-segment measurement and alarm)DSIClass II
Subsequent Product Codes
21 CFR 870.1025Detector and Alarm, ArrhythmiaDSIClass II
21 CFR 870.1025Monitor, ST Segment with AlarmMLDClass II
21 CFR 870.2300Cardiac monitor (including cardiotachometer and rate alarm)DRTClass II
21 CFR 870.1130Non-invasive blood pressure measurement systemDXNClass II
21 CFR 870.1110Blood pressure computerDSKClass II
21 CFR 880.2910Clinical Electronic Thermometers - Temperature Monitor with ProbeFLLClass II
21 CFR 870.2700Oximeter, PulseDQAClass II
21 CFR 868.1400Carbon Dioxide Gas AnalyzerCCKClass II
21 CFR 868.1500Enflurane gas analyzerCBQClass II
21 CFR 868.1620Halothane gas analyzerCBSClass II
21 CFR 868.1700Nitrous Oxide gas analyzerCBRClass II
21 CFR 868.1720Oxygen gas analyszerCCLClass II
21 CFR 882.1400ElectroencephalographGWQClass II
21 CFR 870.2770Impedance plethysmographDSBClass II

Table I-1 Classification of Truscope Patient Monitors

Intended Use Statement:

Truscope Elite A8, Truscope Elite A6, Truscope Elite A5 and Truscope Elite A3

The Truscope series patient monitors (Truscope Elite A8, Truscope Elite A6, Truscope Elite A5 and Truscope Elite A3) is intended to be used for monitoring, displaying, reviewing, storing and alarming of multiple physiological parameters including ECG (3-lead or 5-lead or 12-lead

2

K131763 P 3/5

selectable), arrhythmia detection, ST segment analysis, Heart Rate (HR), Respiration Rate (RESP), Temperature (TEMP), Pulse Oxygen Saturation (SpO2), Pulse Rate (PR), Non-invasive Blood Pressure (NIBP), Invasive Blood Pressure (IBP), Carbon dioxide (CO2), Anesthetic Gas (AG), Impedance Cardiograph (ICG) and Cerebral State Index (CSI).

The arrhythmia detection, ST segment analysis only applied to a single adult patient.

The monitors are to be used in healthcare facilities by clinical physicians or appropriate medical staff under the direction of physician. It is not intended for helicopter transport, hospital ambulance, or home use.

Truscope II and MINI

Truescope Series Patient Monitor (Truscope II and MINI) is intended to be used for monitoring. displaying, reviewing, storing and alarming of multiple physiological parameters including ECG Heart Rate (HR), Respiration Rate (RESP), Pulse Oxygen Saturation (SpO2), Pulse Rate (PR), Non-invasive Blood Pressure (NIBP), Invasive Blood Pressure (IBP), carbon dioxide (CO2), and Temperature (TEMP) of adult, pediatric and neonatal patient.

Truescope Series Patient Monitor (Truscope II and MINI) is to be used in healthcare facilities by clinical physicians or appropriate medical staff under the direction of physicians. It is not intended for helicopter transport or hospital ambulance.

Note: Truscope MINI does not have functions of Invasive Blood Pressure and Carbon Dioxide

Truscope Classic

Truescope Series Patient Monitor (Truscope Classic) is intended to be used for monitoring, displaying, reviewing, storing and alarming of multiple physiological parameters including ECG Heart Rate (HR), Respiration Rate (RESP), Pulse Oxygen Saturation (SpO2). Pulse Rate (PR), Non-invasive Blood Pressure (NIBP), Invasive Blood Pressure (IBP), carbon dioxide (CO2), and Temperature (TEMP) of adult, pediatric and neonatal patient.

The monitors are to be used in healthcare facilities by clinical physicians or appropriate medical staff under the direction of physicians. They are not intended for helicopter transport or hospital ambulance.

  • Predicate Device Identification 5.
    510(k) Number: K120193

3

K 131763 P 4/5

Predicate Device Name: ANY VIEW PATIENT MONITORS ANY VIEW A6, ANY VIEW A8, ANYVIEW A3, ANY VIEW A5

Manufacturer: Guangdong Biolight Meditech Co., Ltd

510(k) Number: K102040

Predicate Device Name: PATIENT MONITOR ANY VIEW A6, ANY VIEW A8, M9500 Manufacturer: Guangdong Biolight Meditech Co., Ltd

510(k) Number: K100046

Predicate Device Name: M SEIRES PATIENT MONITOR MODEL M66, M69, M8000, M9000, M7000

Manufacturer: Guangdong Biolight Meditech Co., Ltd

Device Description 6.

The proposed device Truscope Patient Monitor is change from the Paitent Monitor manufactured by Guangdong Biolight Meditech Co., Ltd, all models included in this submission has exactly same design, materials, and manufacture process with their original model, only different is the model name and device name.

The relationship between with the Truscope Patient Monitor and their original model matrix table list as following:

Truscope Patient Monitor ModelOriginal Patient Monitor Model
Truscope Elite A8ANY VIEW A8 (K120193)
Truscope Elite A5ANY VIEW A6 (K120193)
Truscope Elite A6ANY VIEW A5 (K120193)
Truscope Elite A3ANY VIEW A3 (K120193)
Truscope ClassicM9500 (K102040)
Truscope IIM69 (K100046)
Truscope MINIM7000 (K100046)

7. Non-Clinical Test Conclusion

Bench tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:

4

K131763 P 5/5

  • IEC 60601-1, Medical Electrical Equipment Part 1: General Requirements for Safety, 1988; a) Amendment 1, 1991-11, Amendment 2, 1995.
  • b) IEC 60601-1-2:2007, Medical Electrical Equipment - Part 1-2: General Requirements for Safety - Collateral Standard: Electromagnetic Compatibility -- Requirements and Tests.
  • AAMI EC13:2002, Cardiac monitors, heart rate meters, and alarms; c)
  • AAMI EC57:1998/(R) 2008, Testing and reporting performance results of cardiac rhythm and d) ST-segment measurement algorithms.
  • AAMI SP10:2002/ A1:2003 (R) 2008, Manual, electronic or automated sphygmomanometers; e)
  • () ISO 9919:2005, Medical electrical equipment - Particular requirements for the basic safety and essential performance of pulse oximeter equipment for medical use;
    1. Technological Characteristics Comparison There are any differences between proposed device and predicate except model name.
    1. Substantially Equivalent Conclusion

The proposed device, Truscope Patient Monitors, are determined to be Substantially Equivalent (SE) to the predicate device, ANY VIEW PATIENT MONITORS (K120193), PATIENT MONITOR ( K102040) and M SEIRES PATIENT MONITOR (K100046) in respect of safety and effectiveness.

5

Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure with outstretched wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA" are arranged in a circular pattern around the bird.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002

July 15, 2013

Guangdong Biolight Meditech Co., Ltd. c/o Ms. Diana Hong General Manager P.O. Box 120-119 Shanghai. 200120 CH

Re: K131763

Trade/Device Name: Truscope Series Patient Monitors Regulation Number: 21 CFR 870.1025 Regulation Name: Patient Physiological Monitor (With Arrhythmia Detection Or Alarms) Regulatory Class: Il (two) Product Code: MHX, DSI, MLD, DRT, DXN, DSK, FLL. DQA, CCK, CBQ, CBR. CBS, CCL, CWQ. DSB Dated: June 9, 2013 Received: June 17, 2013

Dear Ms. Diana Hong:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments. or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

6

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

OwenPFAaris-S

for Bram D. Zuckerman, Ph.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

7

K131763

b 1/3

Indication for Use Statement

510(k) Number:

Device Name:Truscope Patient Monitor/Model: Truscope Elite A8, Truscope Elite A6, Truscope Elite A5 and Truscope Elite A3

Indications for Use:

Truscope Elite A8, Truscope Elite A6, Truscope Elite A5 and Truscope Elite A3

The Truscope series patient monitors (Truscope Elite A6, Truscope Elite A6, Truscope Elite A5 and Truscope Elite A3) is intended to be used for monitoring, displaying, storing and alarming of multiple physiological parameters including ECG (3-lead or 12-lead selectable), arthythmia detection, ST segment analysis, Heart Rate (HR), Respiration Rate (RESP), Temperature (TEMP), Pulse Oxygen Saturation (SpO2), Pulse Rate (PR), Non-invasive Blood Pressure (NIBP), Invasive Blood Pressure (IBP), Carbon dioxide (CO2), Anesthetic Gas (AG), Impedance Cardiograph (ICG) and Cerebral State Index (CSI).

The arrhythmia detection, ST segment analysis only applied to a single adult patient.

The monitors are to be used in healthcare facilities by clinical physicians or appropriate medical staff under the direction of physician. It is not intended for helicopter transport, hospital ambulance, or home use.

XPRESCRIPTION USE (Part 21 CFR 801 Subpart D) OOVER-THE-COUNTER USE (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)
Digitally signed by Owen P.
Date:2013.07.15 10:39:59
-04'00'

II-1

8

Project #:M0162012B

Page I of 3

K131763 P 2/3

510(k) Number:

Device Name: Truscope Patient Monitor/Model: Truscope II and MINI

Indications for Use:

Truscope II and MINI

Truescope Series Patient Monitor (Truscope II and MINI) is intended to be used for monitoring, displaying, reviewing, storing and alarming of multiple physiological parameters including ECG, Heart Rate (HR), Respiration Rate (RESP), Pulse Oxygen Saturation (SpO2), Pulse Rate (PR), Non-invasive Blood Pressure (NIBP), Invasive Blood Pressure (IBP), carbon dioxide (CO2), and Temperature (TEMP) of adult, pediatric and neonatal patient.

Truescope Series Patient Monitor (Truscope II and MINI) is to be used in healthcare facilities by clinical physicians or appropriate medical staff under the direction of physicians. It is not intended for helicopter transport or hospital ambulance.

Note: Truscope MINI does not have functions of Invasive Blood Pressure and Carbon Dioxide

ZPRESCRIPTION USE (Part 21 CFR 801 Subpart D)

OVER-THE-COUNTER USE (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Digitally signed by Owen
Paris
Date: 2013.07.15 10:40:28
-04'00'

Page 2 of 3

II-2

9

K 131763 P 3/3

510(k) Number: Device Name:Truscope Patient Monitor/Model: Truscope Classic

Indications for Use:

Truscope Classic

Truescope Series Patient Monitor (Truscope Classic) is intended to be used for monitoring, displaying, reviewing, storing and alarming of multiple physiological parameters including ECG, Heart Rate (HR), Respiration Rate (RESP), Pulse Oxygen Saturation (SpO2), Pulse Rate (PR), Non-invasive Blood Pressure (NIBP), Invasive Blood Pressure (IBP), carbon dioxide (CO2), and Temperature (TEMP) of adult, pediatric and neonatal patient.

The monitors are to be used in healthcare facilities by clinical physicians or appropriate medical staff under the direction of physicians. They are not intended for helicopter transport or hospital ambulance.

OPRESCRIPTION USE (Part 21 CFR 801 Subpart D)

OVER-THE-COUNTER USE (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Digitally signed by Owen .
R. Faris: S
Date: 2013.07.15 10:40:49
-04'00'

Page 3 of 3