K Number
K131686
Date Cleared
2013-11-29

(172 days)

Product Code
Regulation Number
868.5165
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The INOmax® DS delivery system delivers INOMAX® (nitric oxide for inhalation) therapy gas into the inspiratory limb of the patient breathing circuit in a way that provides a constant concentration of nitric oxide (NO), as set by the user, to the patient throughout the inspired breath. It uses a specially designed injector module, which enables tracking of the ventilator waveforms and the delivery of a synchronized and proportional dose of NO. It may be used with most ventilators.

The INOmax® DS provides continuous integrated monitoring of inspired O₂, NO₂, and NO, and a comprehensive alarm system.

The INOmax® DS incorporates a battery that provides up to 6 hours of uninterrupted NO delivery in the absence of an external power source.

The INOmax® DS includes a backup NO delivery capability that provides a fixed flow of 250 mL/min of NO which along with user supplied 10 L/min of oxygen provides 20 ppm in the gas flow to a patients breathing circuit. It may also use the INOblender® for backup.

The target patient population is controlled by the drug labeling for INOMAX® and is currently neonates. The primary targeted clinical setting is the Neonatal Intensive Care Unit (NICU) and secondary targeted clinical setting is the transport of neonates.

Device Description

The INOmax DSIR® uses a "dual-channel" design to ensure the safe delivery of INOMAX®. The first channel has the delivery CPU, the flow controller and the injector module to ensure the accurate delivery of NO. The second channel is the monitoring system, which includes a separate monitor CPU, the gas cells (NO, NO₂, and O₂ cells) and the user interface including the display and alarms. The dual-channel approach to delivery and monitoring permits INOMAX® delivery independent of monitoring but also allows the monitoring system to shutdown INOMAX® delivery if it detects a fault in the delivery system such that the NO concentration could become greater than 100 ppm. The delivery system can also shut down delivery if it detects certain serious problems with the monitoring system.

AI/ML Overview

The provided document describes the INOmax DSIR® (Delivery System), a device for delivering nitric oxide to patients. The submission is a 510(k) for a software update (version 3.0) and compatibility with two additional respiratory care devices (Hamilton C1 and T1 Ventilators).

Here's an analysis of the acceptance criteria and the study conducted:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria / RequirementReported Device Performance
02 dilutionTested and confirmed compatibility with the new respiratory care devices. (Implied that 02 dilution was within acceptable limits, as the "Testing Conclusion" states the device performed within published specifications.)
Effect on delivered pressuresTested and confirmed compatibility with the new respiratory care devices. (Implied that delivered pressures were within acceptable limits, as the "Testing Conclusion" states the device performed within published specifications.)
INOmax DSIR® delivery accuracyTested and confirmed compatibility with the new respiratory care devices. (Implied that delivery accuracy was maintained, as the "Testing Conclusion" states the device performed within published specifications.)
NO2 generationTested and confirmed compatibility with the new respiratory care devices. (Implied that NO2 generation was within acceptable limits, as the "Testing Conclusion" states the device performed within published specifications.)
Software Functionality (new/modified user convenience features and alarms)Software verification confirmed the INOmax DSIR® is compliant with its system level requirements and that the new/modified user convenience features and alarms function as specified.
Electrical Safety (IEC 60601-1:2005)Support for substantial equivalence was provided as a result of risk management and testing which included electrical safety tests. (Implied compliance with the standard).
Electromagnetic Compatibility (IEC 60601-1-2:2007)Support for substantial equivalence was provided as a result of risk management and testing which included performance tests. (Implied compliance with the standard).
Alarm Systems (IEC 60601-1-8:2006)Support for substantial equivalence was provided as a result of risk management and testing which included safety tests. (Implied compliance with the standard, and new alarm functionality was verified for the "Low Calibration Failed" alarm).
Usability (Human Factors Engineering) (ANSI/AAMI HE75)A formative usability study was conducted to aid in the development of requirements. (While a formative study is mentioned, the text doesn't explicitly state the acceptance criteria derived from this or the outcome in terms of meeting specific usability performance metrics, only that it aided requirement development.)

2. Sample Size Used for the Test Set and Data Provenance

The document describes testing for the new software version and compatibility with two new ventilators. The "test set" consisted of:

  • Two additional respiratory care devices: Hamilton C1 Ventilator (K120574) and Hamilton T1 Ventilator (K120670).
  • INOmax DSIR® settings: Five settings were used: 0 (baseline), 1, 5, 20, 40, and 80 ppm, for each setting and mode of ventilation, as well as the Backup mode.

The data provenance is non-clinical testing, performed in a controlled laboratory setting (likely within the company or a certified testing facility). There is no indication of country of origin of the data, but the company is based in Madison, Wisconsin, USA. The testing is prospective in the sense that it was conducted specifically to support this 510(k) submission.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

There is no mention of "experts" being used to establish ground truth in the context of the device's technical performance. The "ground truth" for this engineering validation would be the physical measurements taken by calibrated instruments, and the functionality verification against predetermined specifications. The text does not refer to human experts evaluating the "ground truth" of the device's performance characteristics.

4. Adjudication Method for the Test Set

Not applicable. This was a technical performance verification study, not a study involving human interpretation where adjudication would typically be used to resolve discrepancies in expert opinions.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. The document explicitly states: "The subject of this premarket submission, INOmax DSix®, with updated software and interfaced to each of the selected respiratory care devices, did not require clinical studies to support substantial equivalence."

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

Yes, this was effectively a standalone performance evaluation of the device. The testing described assesses the device's ability to accurately deliver nitric oxide, monitor gases, and function with specific ventilators, without human intervention being part of the performance measurement itself (though human users operate the system). The tests focused on the device's intrinsic mechanical and software performance.

7. The Type of Ground Truth Used

The ground truth used for the performance testing was instrumental measurements of various parameters (e.g., O2 dilution, delivered pressures, NO delivery accuracy, NO2 generation) and functional verification against predetermined specifications for software features and alarms.

8. The Sample Size for the Training Set

Not applicable. This device is a hardware/software system, not an AI/ML algorithm that requires a "training set" in the conventional sense. The "software update" refers to deterministic code changes and feature enhancements, not a learned model.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as there is no "training set" in the context of AI/ML. The "ground truth" for the device's design and functionality would be established through engineering specifications, previous predicate device performance, and compliance with recognized standards.

{0}------------------------------------------------

ડ. 510(K) SUMMARY

510(k) Summary

In accordance with 21 CFR 807.92 the following summary of information is provided:

Submitter Information

Date:July 8, 2013
Company:INO Therapeutics doing business as Ikaria2902 Dairy DriveMadison, Wisconsin 53718
Contact Person:Robert BovyAssociate Director, Regulatory Affairs, Devices
Email:Robert.Bovy@Ikaria.com
Telephone:608-226-3450
FAX:608-395-3905
Secondary Contact Person:David TruebloodNOV 29 2013
Email:David.Trueblood@Ikaria.com
Telephone:608-395-3910

Identification of the Device

Device Trade Name:INOmax DSIR® (Delivery System)
Common Name:Nitric Oxide Administration Apparatus (primary)Nitric Oxide Administration Apparatus, Back-up SystemNitric Oxide AnalyzerNitrogen Dioxide Analyzer
Classification Name:Apparatus, Nitric Oxide Delivery, or Apparatus, NitricOxide Backup Delivery
Device Classification:Class II – 21 CFR 868.5165
Product Code:MRN (Primary), MRO, MRP, MRQ
Predicate Device(s)K061901, K070867, K071516, K080484, K081691,K090958, K092545, K093922, K110344, K110635,K113272, K121021, K130605
Description of DeviceThe INOmax DSIR® uses a "dual-channel" design to ensure the safedelivery of INOMAX®. The first channel has the delivery CPU, theflow controller and the injector module to ensure the accurate deliveryof NO. The second channel is the monitoring system, which includes
a separate monitor CPU, the gas cells (NO, NO₂, and O₂ cells) and theuser interface including the display and alarms. The dual-channelapproach to delivery and monitoring permits INOMAX® deliveryindependent of monitoring but also allows the monitoring system toshutdown INOMAX® delivery if it detects a fault in the deliverysystem such that the NO concentration could become greater than 100ppm. The delivery system can also shut down delivery if it detectscertain serious problems with the monitoring system.
Intended UseThe INOmax® DS delivery system delivers INOMAX® (nitric oxidefor inhalation) therapy gas into the inspiratory limb of the patientbreathing circuit in a way that provides a constant concentration ofnitric oxide (NO), as set by the user, to the patient throughout theinspired breath. It uses a specially designed injector module, whichenables tracking of the ventilator waveforms and the delivery of asynchronized and proportional dose of NO. It may be used with mostventilators.The INOmax® DS provides continuous integrated monitoring ofinspired O₂, NO₂, and NO, and a comprehensive alarm system.The INOmax® DS incorporates a battery that provides up to 6 hours ofuninterrupted NO delivery in the absence of an external power source.The INOmax® DS includes a backup NO delivery capability thatprovides a fixed flow of 250 mL/min of NO which along with usersupplied 10 L/min of oxygen provides 20 ppm in the gas flow to apatients breathing circuit. It may also use the INOblender® forbackup.The target patient population is controlled by the drug labeling forINOMAX® and is currently neonates. The primary targeted clinicalsetting is the Neonatal Intensive Care Unit (NICU) and secondarytargeted clinical setting is the transport of neonates.
TechnologyAll revisions of INOmax DSIR® utilize component technology todeliver Nitric Oxide gas to the patient. The components consist of theDelivery System unit, the blender, a stand/cart and the NO gas tanks.In this revision of the INOmax DSIR®, the software and labeling havebeen updated.

{1}------------------------------------------------

.

{2}------------------------------------------------

Determination of Substantial Equivalence

The INOmax DS18 modifications included an update of the software to version 3.0 and modified labeling to identify compatibility with two additional respiratory care devices. The INOmax DSIR® with software version 3.0 and additional respiratory care devices has the same intended use and uses the same device hardware as the cleared INOmax DSIR® predicate device. All features are identical except those described in the table below.

Feature /SpecificationINOmax DSIR® - K130605INOmax DSIR® with software version 3.0and additional respiratory care devices
Low-range GasSensor CalibrationThis calibration is initiatedmanually by the user by pressingthe "Low Cal" button on thecalibration screen.This calibration can still be initiated manuallybut the calibration is initiated automaticallywhen the device is turned on and is initiatedautomatically every 12 hours while incontinuous operation at a constant dose.
Purge During Pre-Use CheckoutThe purge step is accomplishedby instructing the user to set a NOdose of 40 ppm using the controlwheel on the device.After the first step of the Pre-Use Checkout, theuser presses the "Next" button on the Pre-UseCheckout wizard screen and this automaticallyinitiates the purge step.
Other UserConvenienceFeaturesThe steps required for pre-usecheckout, gas cell calibration, useof backup NO delivery mode andtroubleshooting of alarms are alldescribed in the device labeling.In addition to the instructions in the devicelabeling, wizards are available on the userinterface to guide the user through the requiredsteps for these functions.
Alarms/AlertsDelivery Failure for NO > 100ppmThis alarm has been modified so that the devicewill stop NO delivery if the NO concentrationexceeds 100 ppm for 12 seconds but willautomatically resume NO delivery if the NOconcentration drops back below 100 ppm for12 seconds. Previously this condition caused adevice shutdown and required the user torestart the device to resume delivery.
Delivery FailureOne system voltage condition during whichthis alarm was triggered has been eliminated.
Not ApplicableA new alarm was added for a failure during thenewly added automatic low-range calibrationand is indicated as "Low Calibration Failed".This alarm was added because the automaticlow-range calibration can occur while thedevice is not attended by the user, so it isimportant to notify the user if this calibration isnot successful.

Comparison to Predicate Device

{3}------------------------------------------------

Feature /SpecificationINOmax DSIR® - K130605INOmax DSIR® with software version 3.0and additional respiratory care devices
Labeling forcompatibility withrespiratory caredevicesA variety of transport, neonatal,adult/ped, high frequency andanesthesia ventilators, nasalCPAP and nasal high flowcannulas.Additional respiratory care devices include:Hamilton C1 Ventilator (K120574)Hamilton T1 Ventilator (K120670)

Summary of Nonclinical Tests

The following quality assurance measures were applied to the modification of the system:

  • . Risk analysis
  • . Formative usability study
  • Requirements reviews .
  • . Design reviews
  • . Testing on unit level (module verification)
  • . Integration testing (system verification)
  • Performance testing (verification) .
  • . Safety testing (verification)

Support for the substantial equivalence of the INOmax DSIR® was provided as a result of risk management and testing which included electrical safety, performance and software tests. This testing includes conformity to the FDA recognized consensus standards and voluntary standards as follows:

  • IEC 60601-1:2005: Medical electrical equipment Part 1: General requirements for basic . safety and essential performance
  • IEC 60601-1-2:2007 General requirements for basic safety and essential performance -. Collateral standard: Electromagnetic compatibility - Requirements and tests
  • IEC 60601-1-8:2006 General requirements for basic safety and essential performance -. Collateral Standard: General requirements, tests and guidance for alarm systems in medical electrical equipment and medical electrical systems.

A formative usability study was conducted per ANSI/AAMI HE75 Human Factors Engineering, Design of Medical Devices, to aid in the development of requirements and software verification confirmed the INOmax DSJR® is compliant with its system level requirements and that the new/modified user convenience features and alarms function as specified.

To confirm compatibility with the new respiratory care devices, both devices were set up and calibrated according to the manufacturer's recommendations and tested using the settings established for each respiratory care device test. The INOmax DSR was set up and calibrated according to the manufacturer's recommendations.

Five INOmax DSig settings were used [0 (baseline), 1, 5, 20, 40, and 80 ppm] for each setting and mode of ventilation, as well as the Backup mode.

{4}------------------------------------------------

The measured values on the INOmax DSR were also recorded along with any anomalies found.

The testing concluded four requirements necessary for the INOmax DSp and the two respiratory care devices to be compatible:

  • 02 dilution
  • Effect on delivered pressures
  • INOmax DSIR® delivery accuracy
  • NO2 generation

Testing Conclusion:

The INOmax DSm performed within published specifications when used with each of the ventilators in both primary and backup delivery.

Summary of Clinical Tests

The subject of this premarket submission, INOmax DSix®, with updated software and interfaced to each of the selected respiratory care devices, did not require clinical studies to support substantial equivalence.

Conclusion

INO Therapeutics/Ikaria considers the INOmax DSJR® to be as safe and as effective as the predicate device, with performance substantially equivalent to the predicate device.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus or a bird-like figure, with three wave-like lines extending from the bottom.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G60 Silver Spring MD 20993-0002

November 29, 2013

Mr. Robert Bovy Associate Director, Regulatory Affairs INO Therapeutics 2902 Dairy Drive MADISON WI 53718

Re: K131686

Trade/Device Name: INOmax DSm® (Delivery System) Regulation Number: 21 CFR 868.5165 Regulation Name: Nitric Oxide Administration Apparatus Regulatory Class: II Product Code: MRN, MRQ, MRP Dated: October 28, 2013 Received: October 30, 2013

Dear Mr. Bovy:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{6}------------------------------------------------

Page 2 - Mr. Bovy

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Image /page/6/Picture/8 description: The image shows the text "Kwame Ulmer's" in a bold, sans-serif font. The text is arranged in two lines, with "Kwame" on the top line and "Ulmer's" on the bottom line. To the right of the text is a logo that appears to be a stylized design, possibly incorporating geometric shapes and letters.

Erin I. Keith, M.S. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation · Center for Devices and Radiological Health

for

Enclosure

{7}------------------------------------------------

K131686

Ikaria 510(k) Premarket Notification

INOmax DSIR with Version 3.0 Software

INDICATIONS FOR USE STATEMENT 4.

Indications for Use

510(k) Number (if known): INOmax® DS Device Name: Indications for Use:

The INOmax® DS delivery system delivers INOMAX® (nitric oxide for inhalation) therapy gas into the inspiratory limb of the patient breathing circuit in a way that provides a constant concentration of nitric oxide (NO), as set by the user, to the patient throughout the inspired breath. It uses a specially designed injector module, which enables tracking of the ventilator waveforms and the delivery of a synchronized and proportional dose of NO. It may be used with most ventilators.

The INOmax® DS provides continuous integrated monitoring of inspired 02, NO2, and NO, and a comprehensive alarm system.

The INOmax® DS incorporates a battery that provides up to 6 hours of uninterrupted NO delivery in the absence of an external power source.

The INOmax® DS includes a backup NO delivery capability that provides a fixed flow of 250 ml/min of NO which along with user supplied 10 L/min of oxygen provides 20 ppm in the gas flow to a patients breathing circuit. It may also use the INOblender® for backup.

The target patient population is controlled by the drug labeling for INOMAX® and is currently neonates. The primary targeted clinical setting is the Neonatal Intensive Care Unit (NICU) and secondary targeted clinical setting is the transport of neonates.

Prescription Use_X_ And/Or (21 CFR Part 801 Subpart D)

Over the Counter Use_ (21 CFR Part 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Image /page/7/Picture/15 description: The image shows the name "Nayan J. Patel -S" at the top. Below the name, the year "2013" is printed, followed by the date and time "11-27 14:32:27". The time zone offset "-05'00" is at the bottom left. There is a circular stamp in the middle of the image.

1

§ 868.5165 Nitric oxide administration apparatus.

(a)
Identification. The nitric oxide administration apparatus is a device used to add nitric oxide to gases that are to be breathed by a patient. The nitric oxide administration apparatus is to be used in conjunction with a ventilator or other breathing gas administration system.(b)
Classification. Class II. The special control for this device is FDA's “Guidance Document for Premarket Notification Submissions for Nitric Oxide Administration Apparatus, Nitric Oxide Analyzer, and Nitrogen Dioxide Analyzer.”