K Number
K110344

Validate with FDA (Live)

Manufacturer
Date Cleared
2011-06-24

(140 days)

Product Code
Regulation Number
868.5165
Age Range
All
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The INOmax DS delivery system delivers INOmax® (nitric oxide for inhalation) therapy gas into the inspiratory limb of the patient breathing circuit in a way that provides a constant concentration of nitric oxide (NO), as set by the user, to the patient throughout the inspired breath. It uses a specially designed injector module, which enables tracking of the ventilator waveforms and the delivery of a synchronized and proportional dose of NO. It may be used with most ventilators.

The INOmax DS provides continuous integrated monitoring of inspired O2, NO2 and NO, and a comprehensive alarm system.

The INOmax DS incorporates a battery that provides up to 6 hours of uninterrupted NO delivery in the absence of an external power source.

The INOmax DS includes a backup NO delivery capability that provides a fixed flow of 250 mL/min of NO which along with user supplied 10 L/min of oxygen provides 20 ppm in the gas flow to a patients breathing circuit. It may also use the INOblender for backup.

The target patient population is controlled by the drug labeling for INOmax and is currently neonates. The primary targeted clinical setting is the Neonatal Intensive Care Unit (NICU) and secondary targeted clinical setting is the transport of neonates.

Device Description

The INOmax DSIR uses a "dual-channel" design to ensure the safe delivery of INOmax. The first channel has the delivery CPU the flow controller and the injector module to ensure the accurate delivery of NO. The second channel is the monitoring system, which includes a separate monitor CPU, the gas cells (NO, NO2, and O2 cells) and the user interface including the display and alarms. The dual-channel approach to delivery and monitoring permits INOmax delivery independent of monitoring but also allows the monitoring system to shutdown INOmax delivery if it detects a fault in the delivery system such that the NO concentration could become greater than 100 ppm.

AI/ML Overview

The provided text describes the INOmax DS delivery system and its acceptance criteria, primarily focusing on its compatibility with the Vapotherm Precision Flow system. It does not contain information about a study proving the device meets the acceptance criteria in the typical sense of a clinical statistical study for AI/machine learning devices. Instead, it describes non-clinical engineering tests.

Here's a breakdown of the requested information based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

Performance MetricAcceptance CriteriaReported Device Performance
O2 Dilution (with Vapotherm)Measured O2 within ± 4% v/v of the expected value (does not include Precision Flow tolerance)The measured O2 was within ± 4% v/v of the expected value.
INOmax DS Delivery AccuracyMeasured values +15% of setting or 1.3 ppm, whichever is greater (within INOmax DSIR specification)The measured values were +15% of setting or 1.3 ppm whichever is greater, which is within specification of the INOmax DSIR.
NO2 Generation (Max at 80 ppm NO, 100% O2)Expected behavior (no specific numerical limit stated, but within label limits)The maximum generated NO2 was at 80 ppm and 100% oxygen, as would be expected.
NO2 Levels (general, with Vapotherm)Did not exceed 1.3 ppm on any flow settings (per INOmax label < 3 ppm on 100% FiO2 and any dose setting)NO2 levels did not exceed 1.3 ppm on any of the flow settings.
NO2 Levels (at 40 ppm NO, 60% O2)< 1 ppm (per FDA guideline)The maximum generated NO2 at 40 ppm and 60% oxygen was 0.4 ppm.
Backup Delivery (NO Accuracy)Within 5% of the INOmax DSIR specification at flow rates above 5 L/minThe delivered NO during backup delivery was within 5% of the INOmax DSIR specification at flow rates above 5 L/min.
Backup Delivery (NO2 Levels)Did not exceed 0.2 ppm on any settings tested and 100% oxygen (per INOmax label < 3 ppm on 100% FiO2 and any dose setting)Measured NO2 levels did not exceed 0.2 ppm on any settings tested and 100% oxygen.
Backup Delivery (O2 Accuracy)Measured O2 within + 3% v/v of the expected valueThe measured O2 was within + 3% v/v of the expected value.

2. Sample Size Used for the Test Set and Data Provenance
The document describes non-clinical laboratory testing ("The testing concluded four requirements necessary for the operation of the INOmax DSIR and Vapotherm Precision Flow system to be compatible"). It does not specify a "test set" in terms of patient data or a sample size of cases. The tests relate to the device's compatibility and performance characteristics. Therefore, there is no mention of data provenance (e.g., country of origin, retrospective/prospective).

3. Number of Experts Used to Establish Ground Truth and Qualifications
Not applicable. This was a non-clinical engineering validation, not a study requiring expert readers to establish ground truth from patient data.

4. Adjudication Method for the Test Set
Not applicable. There was no "test set" in the context of human-reviewed data, and thus no adjudication method as described.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. The submission explicitly states: "The subject of this premarket submission, INOmax DSJR interfaced to Vapotherm Precision Flow did not require clinical studies to support substantial equivalence."

6. Standalone Performance Study
Yes, in a sense. The described tests ("Summary of Non-Clinical Tests") represent standalone performance evaluations of the INOmax DSIR's functional parameters and its compatibility when interfaced with the Vapotherm Precision Flow system. It's not "algorithm only without human-in-the-loop performance" in the AI context, but rather device-only performance in an engineering context. The results validated that "the additional interface to the Vapotherm Precision Flow system did not alter the safety or effectiveness of the INOmax DS operation."

7. Type of Ground Truth Used
The "ground truth" was established based on engineering specifications, expected values, and regulatory guidelines (e.g., "within specification of the INOmax DSIR," "per INOmax label," "per FDA guideline"). It was not derived from expert consensus, pathology, or outcomes data.

8. Sample Size for the Training Set
Not applicable. This is not an AI/machine learning device; therefore, there is no concept of a training set as used in that context. The device's operation is based on predefined mechanical and electronic controls and sensors.

9. How Ground Truth for the Training Set Was Established
Not applicable, as there is no training set.

{0}------------------------------------------------

Appendix A – K110344 Section 5; Summary Additional Information Response.

::

·

. :

. .

::

510(k) Summary

JUN 2 4 2011

In accordance with 21 CFR 807.92 the following summary of information is provided:
Date:December 16, 2010
Submitter:INO Therapeutics/Ikaria
2902 Dairy Drive
Madison, Wisconsin 53718
Primary Contact Person:Larry Lepley
Associate Director, Regulatory Affairs
INO Therapeutics/Ikaria
T: 608-226-3415
F: 608-226-3402
Secondary Contact Person:David Trueblood
Director, Regulatory Affairs
INO Therapeutics/Ikaria
T: 608-395-3910
F: 608-226-3402
Device: Trade Name:INOmax DS (Delivery System)
Common/Usual Name:Nitric Oxide Administration Apparatus (primary)
Nitric Oxide Administration Apparatus, Back-up System
Nitric Oxide Analyzer
Nitrogen Dioxide Analyzer
Classification Names:Apparatus, Nitric Oxide Delivery, or Apparatus, Nitric OxideBackup Delivery, Class II - 21 CFR 868.5165
Product Code:MRN (Primary), MRQ, MRP
Predicate Device(s):INOmax DSIR K061901, K070867, K071516, K080484,K081691, K090958, K092545, K093922
Device Description:The INOmax DSIR uses a "dual-channel" design to ensure thesafe delivery of INOmax. The first channel has the delivery CPUthe flow controller and the injector module to ensure the accuratedelivery of NO. The second channel is the monitoring system,which includes a separate monitor CPU, the gas cells (NO, NO2,and O2 cells) and the user interface including the display andalarms. The dual-channel approach to delivery and monitoringpermits INOmax delivery independent of monitoring but alsoallows the monitoring system to shutdown INOmax delivery if itdetects a fault in the delivery system such that the NOconcentration could become greater than 100 ppm.

{1}------------------------------------------------

Appendix A - K110344 Section 5; Summary Additional Information Response.

Intended Use: The INOmax DS delivery system delivers INOmax® (nitric oxide of inhalation) therapy gas into the inspiratory limb of the patient breathing circuit in a way that provides a constant concentration of nitric oxide (NO), as set by the user, to the patient throughout the inspired breath. It uses a specially designed injector module, which enables tracking of the ventilator waveforms and the delivery of a synchronized and proportional dose of NO. It may be used with most ventilators. The INOmax DS provides continuous integrated monitoring of inspired O2, NO2, and NO, and a comprehensive alarm system. The INOmax DS incorporates a battery that provides up to 6 hours of uninterrupted NO delivery in the absence of an external power source. The INOmax DS includes a backup NO delivery capability that provides a fixed flow of 250 mL/min of NO which along with user supplied 10 L/min of oxygen provides 20 ppm in the gas flow to a patients breathing circuit. It may also use the INOblender for backup. The target patient population is controlled by the drug labeling for INOmax and is currently neonates. The primary targeted clinical setting is the Neonatal Intensive Care Unit (NICU) and secondary targeted clinical setting is the transport of neonates. Technology: All revisions of INOmax DSIR utilize component technology to deliver Nitric Oxide gas to the patient. The components consist of the Delivery System unit, the blender, a stand/cart and the NO gas tanks. Determination of Summary of Non-Clinical Tests: Substantial Equivalence: The testing concluded four requirements necessary for the operation of the INOmax DSrg and Vapotherm Precision Flow system to be compatible: O2 dilution The measured O2 was within ± 4% v/v of the expected value (see Pass/Fail Table on page 8; does not include tolerance of the Precision Flow). INOmax DSm delivery accuracy The measured values were + 15% of setting or 1.3 ppm whichever is greater which is within specification of the INOmax DSIR (see Pass/Fail Table on page 8). NO2 generation The maximum generated NO2 was at 80 ppm and 100% oxygen, as would be expected. NO2 levels did not exceed 1.3 ppm on any of the flow settings (per INOmax label < 3 ppm on 100% FiO2 and any dose setting). The maximum generated NO2 at 40 ppm

{2}------------------------------------------------

Appendix A - K110344 Section 5; Summary Additional Information Response,

and 60% oxygen was 0.4 ppm (worse case per FDA guideline of < 1 ppm on 60% FiO2 at 40 ppm dose setting).

Backup Delivery

The delivered NO during backup delivery was within 5% (see Pass/Fail table, NO Measured section) of the INOmax DSIR specification at flow rates above 5 L/min. Measured NO2 levels did not exceed 0.2 ppm on any settings tested and 100% oxygen (per INOmax label < 3 ppm on 100% FiO2 and any dose setting). The measured O2 was within + 3% v/v of the expected value (see Pass/Fail table 100% Oxygen section of test report).

The INOmax DS revision complies to "Guidance Document for Premarket Notification Submissions for Nitric Oxide Delivery Apparatus, Nitric Oxide Analyzer and Nitrogen Dioxide Analyzer" (January 24, 2000) FDA-1157

The results of testing validated that the additional interface to the Vapotherm Precision Flow system did not alter the safety or effectiveness of the INOmax DS operation.

Summary of Clinical Tests:

The subject of this premarket submission, INOmax DSJR interfaced to Vapotherm Precision Flow did not require clinical studies to support substantial equivalence.

Comparison of the Similarities and Differences with the Predicate Device.

Other than the item listed below, all functional/technological characteristics remain the same between the reviewed INOmax DSIR and the predicate INOmax DSJR.

Similarities:

No functional or technological change was made to INOmax DSJR for this 510(k) submission. The INOmax DS18 submitted for this review retains the same as the predicate device(s).

Differences:

Labeling change; listing and connection criteria for Vapotherm Precision Flow as an accepted interface for INOmax DSIR to be added in the INOmax DSIR Instructions for Use (IFU).

Conclusion:

INO Therapeutics/Ikaria considers the INOmax DSB to be as safe, as effective, and performance is substantially equivalent to the predicate device.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is an abstract image of three human profiles facing to the right, with a wavy line underneath them.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Mr. Larry Lepley Associate Director, Regulatory Affairs INO Therapeutics 2902 Dairy Drive Madison, Wisconsin 53718

JUN 2 4 2011

Re: K110344

Trade/Device Name: INOmax DS Regulation Number: 21 CFR 868.5165 Regulation Name: Nitric Oxide Administration Apparatus Regulatory Class: II Product Code: MRN Dated: June 2, 2011 Received: June 3, 2011

Dear Mr. Lepley:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2 - Mr. Lepley

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Anthony D. Watson

Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital. Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

510(k) Number (if known):

INOmax DS Device Name:

Indications for Use:

The INOmax DS delivery system delivers INOmax® (nitric oxide for inhalation) therapy gas into the inspiratory limb of the patient breathing circuit in a way that provides a constant concentration of nitric oxide (NO), as set by the user, to the patient throughout the inspired breath. It uses a specially designed injector module, which enables tracking of the ventilator waveforms and the delivery of a synchronized and proportional dose of NO. It may be used with most ventilators.

The INOmax DS provides continuous integrated monitoring of inspired 02, NO2 and NO, and a comprehensive alarm system.

The INOmax DS incorporates a battery that provides up to 6 hours of uninterrupted NO delivery in the absence of an external power source.

The INOmax DS includes a backup NO delivery capability that provides a fixed flow of 250 mL/min of NO which along with user supplied 10 L/min of oxygen provides 20 ppm in the gas flow to a patients breathing circuit. It may also use the INOblender for backup.

The target patient population is controlled by the drug labeling for INOmax and is currently neonates. The primary targeted clinical setting is the Neonatal Intensive Care Unit (NICU) and secondary targeted clinical setting is the transport of neonates.

Prescription Use_X____________________________________________________________________________________________________________________________________________________________ AND/OR Use (Part 21 CFR 801 Subpart D)

Over-The-Counter

(Part 21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

L. Schulther

(Division Sign-Uff) (Division Sign-On)
Division of Anesthesiology, General Hospital
Division of Anesthesion of Boxicos ാസ്ട്രിംസ് 6.
Infection Control, Dental Devices

510(k) Number: K110344

§ 868.5165 Nitric oxide administration apparatus.

(a)
Identification. The nitric oxide administration apparatus is a device used to add nitric oxide to gases that are to be breathed by a patient. The nitric oxide administration apparatus is to be used in conjunction with a ventilator or other breathing gas administration system.(b)
Classification. Class II. The special control for this device is FDA's “Guidance Document for Premarket Notification Submissions for Nitric Oxide Administration Apparatus, Nitric Oxide Analyzer, and Nitrogen Dioxide Analyzer.”