(28 days)
The MectaLIF implants in combination with supplemental fixation are indicated for use with autogenous bone graft in patients with degenerative disc disease (DDD) at one or two contiguous spinal levels from L2 - S1 whose condition requires the use of interbody fusion. Degenerative disc disease is defined as discogenic pain with degeneration of the disc confirmed by patient history and radiographic studies. These patients may have had a previous non-fusion spinal surgery at the involved spinal level(s). Palients must be skeletally mature. Patients should have received 6 months of non-operative treatment prior to treatment with the devices.
MectaLIF Extension is a design modification to MectaLIF Posterior and Oblique (K110927) and MectaLIF Transforaminal (K120024). The bullet nose modification to the implant tip geometry creates a self-distracting feature which facilitates insertion of the implant into the disc space. MectaLIF Extension also includes intermediate sizes of MectaLIF Posterior and Oblique as well as new sizes of MectaLIF Oblique. The materials and indications for use are identical to the predicate devices. The materials are PEEK-OPTIMA LT1: Implant Grade Polyetheretherketone (ASTM F 2026) and Tantalum (ISO 13782 / ASTM F 560). MectaLIF Transforaminal also has a Gear made of Titanium: Ti6Al4V ELI (ISO 5832-3/ASTM F 136). The devices are intended to be used in combination with posterior fixation (e.g. Pedicle Screw System) as well as an autogenous bone graft.
The provided document describes the MectaLIF Extension, an intervertebral body fusion device. The submission primarily focuses on demonstrating substantial equivalence to predicate devices rather than presenting a standalone study with acceptance criteria and performance data for a new AI-powered device. Therefore, much of the requested information regarding AI device performance, sample sizes for test and training sets, expert involvement, and ground truth establishment is not applicable or cannot be extracted from this document.
However, I can extract information related to the performance testing conducted to support the substantial equivalence claim.
1. Table of acceptance criteria and the reported device performance
| Test | Acceptance Criteria | Reported Device Performance |
|---|---|---|
| Static Axial Compression (ASTM F2077) | Performance equivalent to or better than predicate devices. | Similar performance to predicates; MectaLIF Extension determined not to be the worst case. |
| Dynamic Axial Compression (ASTM F2077) | Performance equivalent to or better than predicate devices. | Similar performance to predicates; MectaLIF Extension determined not to be the worst case. |
| Static Compression/Shear (ASTM F2077) | Performance equivalent to or better than predicate devices. | Similar performance to predicates; MectaLIF Extension determined not to be the worst case. |
| Dynamic Compression/Shear (ASTM F2077) | Performance equivalent to or better than predicate devices. | Similar performance to predicates; MectaLIF Extension determined not to be the worst case. |
| Subsidence Resistance (ASTM F2267) | Performance equivalent to or better than predicate devices. | Similar performance to predicates; MectaLIF Extension determined not to be the worst case. |
2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
The document does not specify the exact sample sizes (number of devices tested) for each performance test. It states that "The testing was conducted on the worst case component size and option/design based on engineering analysis." This implies a limited selection of devices representing the most challenging configurations. Data provenance (country of origin, retrospective/prospective) is not provided as this is a device performance test, not a clinical data study.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
Not applicable. This is a mechanical performance study for an intervertebral body fusion device, not an AI device requiring expert-established ground truth from clinical data.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable. This is not a study involving human readers or clinical data adjudication.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This document does not describe a multi-reader multi-case (MRMC) comparative effectiveness study or any study involving human readers with or without AI assistance. This is a 510(k) submission for a physical medical device.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. This document is not for an AI algorithm.
7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)
For the mechanical performance tests, the "ground truth" (or reference for comparison) was established by the pre-defined acceptance criteria based on standards (like ASTM F2077, ASTM F2267), FDA guidance, and comparison to predicate device systems.
8. The sample size for the training set
Not applicable. This is not an AI device or a study involving a training set.
9. How the ground truth for the training set was established
Not applicable. This is not an AI device or a study involving a training set.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for Medacta International. The logo consists of two triangles, followed by the word "Medacta" in a sans-serif font. Below the word "Medacta" is a dashed line, and below that is the word "International" in a smaller font. To the right of the word "Medacta" is a small square with a plus sign inside.
JUL 0 5 2013
510(k) Summary
Applicant/Sponsor: Medacta International SA Strada Regina 6874 Castel San Pietro (CH) Switzerland Phone (+41) 91 696 60 60 Fax (+41) 91 696 60 66
- Contact Person: Mr. Adam Gross Director of Regulatory, Quality and Compliance Medacta USA 4725 Calle Quetzal, Unit B Camarillo, CA, 93012 Phone: (805) 322-3289 Fax: (805) 437-7553 Email: AGross@medacta.us.com
Date Prepared: June 7, 2013
DEVICE INFORMATION
Trade/Proprietary Name: MectaLIF Extension Common Name: Intervertebral Body Fusion Device Classification Name: Intervertebral Fusion Device with Bone Graft, Lumbar
21 CFR 888.3080 Class II Device Product Codes: MAX
Predicate Devices:
| 510(k) | Product | 510(k) Holder | Clearance Date |
|---|---|---|---|
| K110927 | MectaLIF Posterior and Oblique | Medacta International | 6/13/2011 |
| K120024 | MectaLIF Transforaminal | Medacta International | 2/28/2012 |
| K072791 | OPAL Spacer | Synthes | 12/26/2007 |
| K040536 | Boomerang | Medtronic | 5/5/2004 |
| K073291 | Capstone | Medtronic | 04/24/2008 |
| K081917 | Devex/Leopard | Depuy Spine | 5/22/2009 |
| P960025 | Lumbar IF Cage | Depuy Acromed | 2/2/1999 |
MectaLIF Extension 510(k)
Section 5 - Page 2 of 4
{1}------------------------------------------------
Product Description
MectaLIF Extension is a design modification to MectaLIF Posterior and Oblique (K110927) and MectaLIF Transforaminal (K120024). The bullet nose modification to the implant tip geometry creates a self-distracting feature which facilitates insertion of the implant into the disc space. MectaLIF Extension also includes intermediate sizes of MectaLIF Posterior and Oblique as well as new sizes of MectaLIF Oblique. The materials and indications for use are identical to the predicate devices. The materials are PEEK-OPTIMA LT1: Implant Grade Polyetheretherketone (ASTM F 2026) and Tantalum (ISO 13782 / ASTM F 560). MectaLIF Transforaminal also has a Gear made of Titanium: Ti6Al4V ELI (ISO 5832-3/ASTM F 136). The devices are intended to be used in combination with posterior fixation (e.g. Pedicle Screw System) as well as an autogenous bone graft.
Indications for Use
The MectaLIF implants in combination with supplemental fixation are indicated for use with autogenous bone graft in patients with degenerative disc disease (DDD) at one or two contiquous spinal levels from L2 -- S1 whose condition requires the use of interbody fusion. Degenerative disc disease is defined as discogenic pain with degeneration of the disc confirmed by patient history and radiographic studies. These patients may have had a previous non-fusion spinal surgery at the involved spinal level(s). Patients must be skeletally mature. Patients should have received 6 months of non-operative treatment prior to treatment with the devices.
Comparison to Predicate Devices
The indications for use and materials of the MectaLIF Extension are identical to the previously cleared predicate devices. The design features, geometries, and sizes of the subject devices are substantially equivalent to those of the predicate devices. The fundamental scientific technology of the modified devices has not changed relative to the predicate devices. The safety and effectiveness of the MectaLIF Extension are adequately supported by the substantial equivalence information, materials information, and analysis data provided within this Premarket Notification.
Performance Testing
The modification to the device system to include the addition of MectaLIF Extension was evaluated by risk analysis to identify any new risks associated with the change. Based on the risk analysis, design verification was conducted to written protocols with pre-defined acceptance criteria. The protocols and pre-defined acceptance criteria were based on the standards, FDA guidance, and comparison to the predicate device system. The testing was conducted on the worst case component size and
{2}------------------------------------------------
option/design based on engineering analysis. The MectaLIF Extension was compared to the worst case of the predicate devices and it was determined that the MectaLIF Extension are not worst case.
MectaLIF Extension has similar performance testing as the predicates in terms of: Static Axial Compression - ASTM F2077 Dynamic Axial Compression - ASTM F2077 Static Compression/Shear - ASTM F2077 Dynamic Compression/Shear - ASTM F2077 Subsidence Resistance - ASTM F2267
Conclusion:
Based on the above information, the MectaLIF Extension can be considered as substantially equivalent to its predicate devices.
Section 5 - Page 4 of 4
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. Inside the circle is a stylized image of an eagle.
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Medacta International SA % Mr. Adam Gross Director of Regulatory, Quality and Compliance 4725 Calle Quetzal, Unit B Camarillo, California 93012
Re: K131671
Trade/Device Name: MectaLIF Extension Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: MAX Dated: June 7, 2013 Received: June 7, 2013
Dear Mr. Gross:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading,
July 5, 2013
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{4}------------------------------------------------
Page 2 - Mr. Adam Gross
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803). please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/dcfault.htm for the CDRH3's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Mark N. Melkerson -S
- Mark Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Indications for Use
510(k) Number (if known): K131671
Device Name: MectaLIF Extension
Indications for Use:
The MectaLIF implants in combination with supplemental fixation are indicated for use with autogenous bone graft in patients with degenerative disc disease (DDD) at one or two contiguous spinal levels from L2 - S1 whose condition requires the use of interbody fusion. Degenerative disc disease is defined as discogenic pain with degeneration of the disc confirmed by patient history and radiographic studies. These patients may have had a previous non-fusion spinal surgery at the involved spinal level(s). Palients must be skeletally mature. Patients should have received 6 months of non-operative treatment prior to treatment with the devices.
Prescription Use × (21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Anton E. Dmitriev, PhD Division of Orthopedic Devices
MectaLIF Extension 510(k) June 6, 2013
Section 4 - Page 2 of 2
§ 888.3080 Intervertebral body fusion device.
(a)
Identification. An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.(b)
Classification. (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
Date premarket approval application (PMA) or notice of product development protocol (PDP) is required. Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.