(106 days)
Not Found
No
The document describes a standard computed radiography system with image processing, but there is no mention of AI, ML, or related concepts. The performance studies focus on image quality comparisons with a predicate device using phantom images, not on evaluating AI/ML algorithm performance.
No
The device is described as an x-ray imaging system used to capture diagnostic quality radiographic images, not to treat or alleviate a medical condition.
Yes
The "Intended Use / Indications for Use" section explicitly states that the device is "indicated for use... to capture for display diagnostic quality radiographic images of human anatomy."
No
The device description explicitly states it is a "solid state x-ray imaging device" and lists hardware components like "Phosphor coated imaging plates and cassettes" and a "Laser digitizer". While it includes software (NX workstation), it is not solely software.
No, this device is not an IVD (In Vitro Diagnostic).
Here's why:
- IVD Definition: In Vitro Diagnostic devices are used to examine specimens taken from the human body, such as blood, urine, or tissue, to provide information about a person's health.
- Device Function: The description clearly states that this device is a Computed Radiography (CR) System used for capturing radiographic images of human anatomy using X-rays. This is an in vivo imaging technique, meaning it examines the body directly, not through a sample taken from the body.
- Intended Use: The intended use is for "general projection radiographic applications to capture for display diagnostic quality radiographic images of human anatomy." This aligns with medical imaging, not laboratory testing of samples.
The information provided describes a medical imaging device, not an IVD.
N/A
Intended Use / Indications for Use
Agfa's Computed Radiography (CR) System with CR 12-X Digitizer is indicated for use in general projection radiographic applications to capture for display diagnostic quality radiographic images of human anatomy. The system may be used wherever conventional screen-film systems are used.
Agfa's Computed Radiography (CR) System With CR 12-X Digitizer is not indicated for use in mammography.
Product codes
MOB
Device Description
Agfa's Computed Radiography (CR) Systems with CR 12-X Digitizer is a solid state x-ray imaging device. Principles of operation and technological characteristics of the new and predicate devices are largely the same as other computed radiography systems:
Phosphor coated imaging plates and cassettes for image capture. Laser digitizer for generating the electronic image. NX workstation for image previewing, processing and routing.
Mentions image processing
Yes
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
X-Ray
Anatomical Site
human anatomy
Indicated Patient Age Range
Not Found
Intended User / Care Setting
Not Found
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Tests included image quality tests with internal and experts comparing the CR 12-X to the CR 10-X predicate. Clinical Images were unnecessary to evaluate modifications from the predicate device; therefore, phantom images were used to evaluate substantial equivalence with the CR 10-X predicate device (K121948).
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
The device has completed verification and validation testing to confirm it meets specifications and operates as planned. Tests included image quality tests with internal and experts comparing the CR 12-X to the CR 10-X predicate. Clinical Images were unnecessary to evaluate modifications from the predicate device; therefore, phantom images were used to evaluate substantial equivalence with the CR 10-X predicate device (K121948).
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s)
Reference Device(s)
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information
Not Found
§ 892.1680 Stationary x-ray system.
(a)
Identification. A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
0
K 131408
Page 1 of 5
Computed Radiography (CR) Systems with CR 12-X Digitizer
510(K) SUMMARY Agfa Computed Radiography (CR) Systems with CR 12-X Digitizer
Common Name: Computed Radiography System Classification Name: Solid State X-Ray System Regulatory Classification: 21 CFR 892.1680 Product Code: MOB Proprietary Name: Computed Radiography (CR) Systems with CR 12-X Digitizer Agfa HealthCare N.V. Septestraat 27 B-2640 Mortsel Belgium Contact: Koen Vervoort. Prepared: May 15, 2013 Telephone: +32-34444-7368
AUG 2 9 2013
LEGALLY MARKETED PREDICATE DEVICES A.
This is a 510(k) for Agfa's Computed Radiography (CR) System with CR 12-X Digitizer, a solid state x-ray imaging device. It is substantially equivalent to systems with Agfa's CR 10-X Digitizer (K121948) and 3Disc America's Fire CR+ Digitizer (K102619).
DEVICE DESCRIPTION B.
Agfa's Computed Radiography (CR) Systems with CR 12-X Digitizer is a solid state x-ray imaging device. Principles of operation and technological characteristics of the new and predicate devices are largely the same as other computed radiography systems:
Phosphor coated imaging plates and cassettes for image capture. Laser digitizer for generating the electronic image. NX workstation for image previewing, processing and routing.
C. INTENDED USE
Agfa's Computed Radiography (CR) System with CR 12-X Digitizer is indicated for use in general projection radiographic applications to capture for display diagnostic quality radiographic images of human anatomy. The system may be used wherever conventional screen-film systems are used.
Agfa's Computed Radiography (CR) System With CR 12-X Digitizer is not indicated for use in mammography.
The intended use has not changed as a result of any labeling modifications.
1
D. SUBSTANTIAL EQUIVALENCE SUMMARY
Agfa's Computed Radiography (CR) System with CR 12-X Digitizer has an Indications For Use statement identical to the statement for the predicate devices, K121948 and K102619. Intended uses are the same. The devices have the same technological characteristics. Descriptive characteristics and performance data are adequate to ensure equivalence.
Differences in devices do not alter the intended therapeutic/diagnostic effect.
PRODUCT COMPARISON TABLE | |||
---|---|---|---|
Systems with CR 12-X | |||
Digitizer | |||
(New Device) | Systems with CR 10-X Digitizer | ||
(PREDICATE-K121948) | 3Disc: FireCR+ Digitizer | ||
(PREDICATE-K102619) | |||
Communications | Same as both predicates | DICOM | DICOM |
Cassette and Image Plates | |||
Cassettes and Image | |||
Plate Sizes | Same as CR 10-X | 35x43 cm | 35x43 cm & |
24x30 cm | |||
Image Plate | |||
Phosphors | Same as both predicates | BaSrFBrl:Eu | BaSrFBrl: Eu |
Digitizer | |||
Scanning technology | Same as both predicates | Point at-a-time | Point at-a-time |
Light collection | Same as CR 10-X | Solid, high efficiency | Unavailable |
Scanning resolution | 100µm, 150µm, 200µm | 100µm | 100µm and 200µm |
Active matrix | Same as CR 10-X | 3420 x 4218 | Unavailable |
Dynamic range | |||
(acquisition) | Same as both predicates | 16 bit, sq. root compressed | 16 bit, sq. root compressed |
Throughput | |||
(35x43 cm plates/hr.) | 60 @ 200µm | ||
72 @ 150µm | |||
90 @ 100µm | 34 @ 100µm | 20 @ 200µm | |
70 @ 100µm | |||
Power Supply | Same as CR 10-X | 100-240V 50/60 Hz | 110-240V 50/60 Hz |
Mobile installations | Same as CR 10-X | Mobile mounting option | Solid Transport case |
Image stitching | Same as CR 10-X | Full Leg Full Spine | N/A |
Workstation | |||
Image processing | Same as CR 10-X | MUSICA, MUSICA2 | QuantorMed+ Imaging |
Software | |||
Dynamic range | |||
(display) | Same as CR 10-X | 12 bit | 32 bit & 64 bit |
Operating system | Same as both predicates | Windows 7 | Windows 7 |
Power Supply | Same as CR 10-X | 100 - 240V 50/60 Hz | 110-240V 50/60 Hz |
Display System | Same as CR 10-X | Standard PC display or separately | |
cleared medical display | 1,280x900 pixel, DVI | ||
interface |
2
TECHNOLOGICAL CHARACTERISTICS E.
Agfa's Computed Radiography (CR) Systems with CR 12-X Digitizer is a solid state x-ray imaging device. It uses phosphor coated imaging plates to capture an x-ray exposure. Upon scanning of the image plates by the CR-12-X digitizer, the x-ray image is read and transmitted to the NX workstation where the image can be previewed, processed and transmitted to other devices for viewing, printing or storage.
The CR 12-X digitizer is identical to the CR 10-X with the exception of the following:
60 plates/hr with scanning resolution of 200um
72 plates/hr with added scanning resolution of 150um
Increase to 90 plates/hr with scanning resolution of 100um
Agfa CR 12-X and 3Disc America's FireCR+(K102619) both enable selection of resolution options. The remaining differences would not be expected to impact safety and effectiveness.
F. TESTING
The device has completed verification and validation testing to confirm it meets specifications and operates as planned. Tests included image quality tests with internal and experts comparing the CR 12-X to the CR 10-X predicate. Clinical Images were unnecessary to evaluate modifications from the predicate device; therefore, phantom images were used to evaluate substantial equivalence with the CR 10-X predicate device (K121948).
The change in the design of the CR 10-X to include the image processing was verified and validated per the following:
Exhibit 3-2, Software Design Input Output Verification (DIOV) Report - Arcus Digitizer ..
Exhibit 5-1, Verification Plan - Consolidated Requirement & Testing (incl. SW), Arcus Solution.
Exhibit 5-2, Verification Report - Consolidated Requirement & Testing (incl. SW), Arcus Solution.
Exhibit 5-3, Validation Test Plan - Arcus Solution
Exhibit 5-4, Design Validation Report - Arcus Solution
Exhibit 5-5, Validation Test Report - Arcus Solution
Exhibit 5-6, Final Design Review for the CR 12-X Digitizer.
Exhibit 5-7, Design Complete Acceptance for the CR 12-X Digitizer
Exhibit 5-8, CRB (Change Review Board) Arcus Software Version ARC 1404 (This is the maintenance release for digitizer software version ARC 1404).
3
The product, manufacturing and development processes have been shown to conform to product safety, radiology, and imaging standards including:
PRODUCT STANDARDS
IEC 60601-1: 2005 Medical Electrical Equipment - Part 1: General Requirements for Safety
IEC 60601-1-2: 2007 Medical Electrical Equipment - Part 1-2: General Requirements for Safety - Collateral Standard: Electromagnetic Compatibility - Requirements And Tests
ACR/NEMA PS3.1-3.20: 2011 Digital Imaging and Communications in Medicine (DICOM)
MANAGEMENT STANDARDS
ISO 14971:2007 Application of Risk Management to Medical Devices
ISO 13485:2003 Medical Devices - Quality Management Systems - Requirements For Regulatory purposes
G. RISK ASSESSMENT AND MANAGEMENT SUMMARY
During the final risk analysis meeting, the risk management team concluded that the medical risk is not higher than previous released system to the field.
After review of these risks with the risk analysis team. it was decided that this risk could not be further mitigated and that for this risk, the benefits of the system outweigh the risks.
There are a total of 22 risks in the broadly acceptable region and 2 riskd in the ALARP region. Zero risks were identified in the not acceptable region.
The residual risk of the device is acceptable.
- · the risk control measures are effective
- · usability acceptance criteria for safety-related functions have been met (IEC 62366)
- · no residual risk is situated in the intolerable region
- · the residual risks are controlled as far as reasonably practicable
Of the 72 identified medical risks, 7 are in the ALARP region. The residual risks cannot be reduced any further with reasonable effort. Therefore is assumed to be safe, the benefits of the device are assumed to outweigh the residual risk.
The term "Level of Concern" means the level of risk that the software device is determined to be if the software were to fail. The Level of Concern for the device has been determined to be moderate.
Refer to Section 16 Software, for more information regarding risk management.
4
Image /page/4/Picture/1 description: The image shows a handwritten document with the text "K131408 Page 5 of 5". The text appears to be a document identifier and page number. The handwriting is legible, but the style is casual and not formal.
H. CONCLUSIONS
Agfa's Computed Radiography (CR) System with CR 12-X Digitizer has an Indications For Use statement is identical to the statement for the predicate devices. K121948 and K 102619. Intended uses are the same. The devices have the same technological characteristics in that they use the same imaging phosphor and the same laser diode for stimulation light operating with the same spot size as both predicates. Descriptive characteristics and performance data are adequate to ensure equivalence.
Agfa's Computed Radiography (CR) System with CR 12-X Digitizer is substantially equivalent to its CR 10-X System (K121948) in that it uses the same technology to capture and transmit images, use the NX user workstation, use the same MD1.0 image plate and cassette with manual cassette insertion, includes the full leg full spine function, provides a mobile mounting option, and uses 100-240V 50/60 Hz power supply.
Agfa CR 12-X and 3Disc America's FireCR+(K102619) both enable selection of resolution options from 100um to 200um.
The CR 12-X is equivalent to the predicate devices in regards to the plate speed.
This 510(k) has demonstrated Substantial Equivalence as defined and understood in the Federal Food Drug and Cosmetic Act and various guidance documents issued by the Center for Devices and Radiological Health.
5
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles three overlapping human figures or abstract shapes, with flowing lines suggesting movement or connection.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 29, 2013
AGFA HealthCare N.V. % Ms. ShaeAnn Cavanagh Regulatory Affairs Specialist NA AGFA HealthCare 10 South Academy Street GREENVILLE SC 29601
Re: K131408
Trade/Device Name: CR 12-X Digitizer Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: II Product Code: MQB Dated: August 1, 2013 Received: August 7, 2013
Dear Ms. Cavanagh:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
6
Page 2-Ms. Cavanagh
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resources/or You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/RcportaProblem/dcfault.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
for
Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
7
Indications for Use
510(k) Number (if known): K131408
Device Name: CR 12-X Digitizer
Indications For Use:
Agfa's Computed Radiography (CR) System with CR 12-X Digitizer is indicated for use in general projection radiographic applications to capture for display diagnostic quality radiographic images of human anatomy. The system may be used wherever conventional screen-film systems are used.
Agfa's Computed Radiography (CR) System With CR 12-X Digitizer is not indicated for use in mammography.
Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of Center for Devices and Radiological Health (CDRH)
Smh7)
Page 1 of ____________________________________________________________________________________________________________________________________________________________________