K Number
K131325

Validate with FDA (Live)

Date Cleared
2013-08-16

(100 days)

Product Code
Regulation Number
888.3040
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Cayenne Medical, Inc. Quattro® GL Suture Anchor is intended for the reattachment of soft tissue to bone for the following indications:

Hip

  • Hip capsule repair
  • Acetabular labrum reattachment

Shoulder

  • Capsular stabilization
  • Bankart repair
  • Anterior shoulder instability
  • SLAP lesion repairs
  • Capsular shift or capsulolabral reconstructions
  • Acromioclavicular separation repairs
  • Deltoid repairs
  • Rotator cuff repairs
  • Biceps tenodesis

Elbow, Wrist, and Hand

  • Biceps tendon reattachment
  • Ulnar or radial collateral ligament reconstruction
  • Lateral epicondylitis repair

Knee

  • Extra-capsular repairs
  • Medial collateral ligament
  • Lateral collateral ligament
  • Posterior oblique ligament
  • Patellar realignment and tendon repairs
  • Vastus medials obliquous advancement
  • Illiotibial band tenodesis

Foot and Ankle

  • Hallux valgus repairs
  • Medial or Lateral instability repairs/reconstructions
  • Achilles tendon repairs/reconstructions
  • Midfoot reconstructions
  • Metatarsal ligament/tendon repairs/reconstructions
  • Bunionectomy
Device Description

This traditional 501(k) premarket notification is to expand the indications for use for Cayenne Medical Quattro GL Suture Anchor. The Quattro GL (LabraLink) Suture Anchor was cleared per premarket notification K112960. Cayenne seeks to expand the existing indications for use for the subject device to include the indications listed above.

The Quattro® GL Suture Anchor is a sterile, manually operated, single procedure suture anchor. The anchor has two suture eyelets allowing for one or two sutures to be loaded through the eyelets. The suture anchor is mounted on an inserter. The Quattro GL Suture Anchor incorporates design features that facilitate suture anchor placement under arthroscopic, open, or limited access conditions in soft tissue to bone reattachment procedures. The Quattro GL Suture Anchor is only offered in one size, 2.9mm with four suture color options. The anchors are offered in two configurations, single loaded or double loaded sutures. Suture(s) used on the anchor are size # 2 non-absorbable surgical sutures. The Quattro GL inserter has a working length of 25.8 cm with an outer shaft diameter of 3.2 mm. Since the market clearance of this device, the length of the Suture Anchor was decreased from 15mm to 11.4mm. The technological characteristics of the Quattro GL Suture Anchor have not changed.

AI/ML Overview

Here's an analysis of the provided text regarding the acceptance criteria and study for the Quattro® GL Suture Anchor:

1. A table of acceptance criteria and the reported device performance

The provided 510(k) summary does not explicitly state specific acceptance criteria in a quantitative or qualitative manner that would be typically found for performance metrics like sensitivity, specificity, accuracy, etc., for a diagnostic device.

Instead, the submission focuses on demonstrating substantial equivalence to existing predicate devices based on:

Acceptance Criteria CategoryReported Device Performance (Summary)
Mechanical Testing (Pull-Out Strength)"Testing showed that the Quattro GL Suture Anchor ultimate pull-out strength was comparable to that of the predicate device."
Intended UseThe expanded indications for use for the Quattro GL Suture Anchor align with its intended purpose of reattaching soft tissue to bone across various anatomical locations (Hip, Shoulder, Elbow/Wrist/Hand, Knee, Foot/Ankle). This comparison demonstrates equivalence to its own previously cleared version and other predicate devices with similar indications.
Design Features"The subject device... has the same intended use, design, technology, and materials... as the Cayenne Medical Quattro GL (LabraLink) Suture Anchor with Inserter (K112960)."
Materials"The subject device... has the same... materials... as the Cayenne Medical Quattro GL (LabraLink) Suture Anchor with Inserter (K112960)."
Manufacturing Processes"The subject device... has the same... manufacturing processes... as the Cayenne Medical Quattro GL (LabraLink) Suture Anchor with Inserter (K112960)."
Sterilization Method"The subject device... has the same... sterilization method... as the Cayenne Medical Quattro GL (LabraLink) Suture Anchor with Inserter (K112960)."
Shelf Life"The subject device... has the same... shelf life... as the Cayenne Medical Quattro GL (LabraLink) Suture Anchor with Inserter (K112960)."

Key takeaway: The primary "acceptance criterion" for this 510(k) appears to be demonstrating that the device is substantially equivalent to legally marketed predicate devices, particularly in terms of its mechanical properties and that the expanded indications for use do not introduce new safety or effectiveness concerns. The specific numerical values for "pull-out strength" that define "comparable" are not provided in this summary.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document mentions "Non-clinical testing data submitted, referenced, or relied upon" including "Mechanical testing (pull-out strength)."

  • Sample Size: The specific sample size used for the pull-out strength testing is not specified in the provided summary.
  • Data Provenance: The document does not provide details on the country of origin of the data or whether the testing was retrospective or prospective. Given it's mechanical testing of physical anchors, it's inherently prospective in nature for a new batch of tested devices.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This section is not applicable to the provided document. The Quattro® GL Suture Anchor is a medical device (suture anchor) for surgical reattachment, not a diagnostic device that requires expert interpretation for establishing ground truth on a test set (e.g., for image analysis). The "ground truth" for this device lies in its physical and mechanical properties (e.g., strength, biocompatibility), which are evaluated through non-clinical laboratory testing.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This section is not applicable for the same reasons as point 3. Adjudication methods like 2+1 (two readers agree, third is tie-breaker) are used in diagnostic studies to resolve discrepancies in expert interpretation of medical images or data. The testing described for the suture anchor is mechanical in nature and does not involve human interpretation requiring adjudication.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This section is not applicable. The Quattro® GL Suture Anchor is a surgical implant, not an AI-powered diagnostic tool or system designed to assist human readers. Therefore, an MRMC comparative effectiveness study involving human readers and AI assistance is not relevant to this device.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This section is not applicable. The device is a physical surgical implant. There is no algorithm or AI component to this device that would have standalone performance.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

For the mechanical testing (pull-out strength), the "ground truth" would be the objective, empirically measured physical properties of the suture anchors under controlled laboratory conditions, typically using standardized testing methods (e.g., ASTM standards) on a material testing machine. This is not "expert consensus" or "pathology" in the diagnostic sense, but rather a direct measurement of physical performance.

8. The sample size for the training set

This section is not applicable. This device is a physical medical implant, not an AI or machine learning model that requires a "training set" of data.

9. How the ground truth for the training set was established

This section is not applicable for the same reason as point 8.

{0}------------------------------------------------

K131325 (1/3)

Quattro® GL Suture Anchor

510(k) Summary

Cayenne Medical, Inc.

Quattro® GL Suture Anchor

ADMINISTRATIVE INFORMATION

510(k) number:

Date of summary:04-22-2013
Manufacturer Name:Cayenne Medical, Inc.16597 N. 92nd St., Suite 101Scottsdale, AZ 85260Telephone (480) 502-3661FAX (480) 502-3670
Official Contact:Kereshmeh Shahriari16597 N. 92nd St., Suite 101Scottsdale, AZ 85260kshahriari@cayennemedical.comTelephone (480) 502-3661FAX (480) 502-3670

DEVICE NAME

Classification Name:Smooth or threaded metallic bone fixation fastener
Trade/Proprietary Name:Quattro ® GL Suture Anchor

AUG 1 6 2013Common Name:

Suture Anchor

DEVICE CLASSIFICATION

FDA has classified bone screws as Class II devices (21 CFR 888.3040). The product code for this suture anchor is MBI -- Fastener, Fixation, Nondegradable, Soft Tissue. These devices are reviewed by the Orthopedic Joint Devices Branch.

INTENDED USE

The Cayenne Medical, Inc. Quattro® GL Suture Anchor is intended for the reattachment of soft tissue to bone for the following indications:

{1}------------------------------------------------

Hip

  • । Hip capsule repair
    • o Acetabular labrum reattachment

Shoulder

  • Capsular stabilization
    • o Bankart repair
    • Anterior shoulder instability o
    • o SLAP lesion repairs
    • Capsular shift or capsulolabral reconstructions o
  • Acromioclavicular separation repairs l
  • Deltoid repairs -
  • Rotator cuff repairs l
  • -Biceps tenodesis

Elbow, Wrist, and Hand

  • Biceps tendon reattachment —
  • l Ulnar or radial collateral ligament reconstruction
  • -Lateral epicondylitis repair

Knee

  • -Extra-capsular repairs
    • o Medial collateral ligament
    • o Lateral collateral ligament
    • o Posterior oblique ligament
  • Patellar realignment and tendon repairs |
    • o Vastus medials obliquous advancement
  • Illiotibial band tenodesis

Foot and Ankle

  • l Hallux valgus repairs
  • -Medial or Lateral instability repairs/reconstructions
  • Achilles tendon repairs/reconstructions l
  • -Midfoot reconstructions
  • । Metatarsal ligament/tendon repairs/reconstructions
  • । Bunionectomy

DEVICE DESCRIPTION

This traditional 501(k) premarket notification is to expand the indications for use for Cayenne Medical Quattro GL Suture Anchor. The Quattro GL (LabraLink) Suture Anchor was cleared per premarket notification K112960. Cayenne seeks to expand the existing indications for use for the subject device to include the indications listed above.

{2}------------------------------------------------

The Quattro® GL Suture Anchor is a sterile, manually operated, single procedure suture anchor. The anchor has two suture eyelets allowing for one or two sutures to be loaded through the eyelets. The suture anchor is mounted on an inserter. The Quattro GL Suture Anchor incorporates design features that facilitate suture anchor placement under arthroscopic, open, or limited access conditions in soft tissue to bone reattachment procedures. The Quattro GL Suture Anchor is only offered in one size, 2.9mm with four suture color options. The anchors are offered in two configurations, single loaded or double loaded sutures. Suture(s) used on the anchor are size # 2 non-absorbable surgical sutures. The Quattro GL inserter has a working length of 25.8 cm with an outer shaft diameter of 3.2 mm. Since the market clearance of this device, the length of the Suture Anchor was decreased from 15mm to 11.4mm. The technological characteristics of the Quattro GL Suture Anchor have not changed.

NON-CLINICAL TESTING

Non-clinical testing data submitted, referenced, or relied upon to demonstrate substantial equivalence is included. Mechanical testing (pull-out strength) was performed on the Quattro GL Suture Anchor and the predicate device. Testing showed that the Quattro GL Suture Anchor ultimate pull-out strength was comparable to that of the predicate device.

EQUIVALENCE TO MARKETED PRODUCT

Cayenne Medical, Inc. demonstrated that, for the purposes of FDA's regulation of medical devices, the Quattro® GL Suture Anchor is substantially equivalent in indication and design principles to predicate devices, each of which has been determined by FDA to be substantially equivalent to preamendment devices: Smith & Nephew BIORAPTOR 2.3 PK Suture Anchor (K071586), Quattro® GL (LabraLink) Suture Anchor (K112960), Force Fiber Blue Co-Braid Polyethylene non-absorbable surgical suture (K040472), Force Fiber Black Co-Braid Polyethylene non-absorbable surgical suture (K070673), Force Fiber Green Co-Braid Polyethylene non-absorbable surgical suture (K100506), and Force Fiber Blue Polyethylene non-absorbable surgical suture (K092533). The substantial equivalence of Quattro GL Suture Anchor is based on similarities in indications for use. design features, technology, and materials to the predicate device, Bioraptor 2.3 PK Suture Anchor.

The subject device, Quattro GL Suture Anchor with Inserter, has the same intended use, design, technology, materials, manufacturing processes, packaging, sterilization method, and shelf life as the Cayenne Medical Quattro GL (LabraLink) Suture Anchor with Inserter (K112960) and the expansion of indications for use does not introduce new issues of safety and effectiveness.

{3}------------------------------------------------

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

August 16, 2013

Mr. Kereshmeh Shahriari Senior Director of Regulatory Affairs, Quality Assurance & Compliance Cayenne Medical, Incorporated 16597 North 92nd Street, Suite 101 Scottsdale, Arizona 85260

Re: K131325

Trade/Device Name: Quattro® GL Suture Anchor Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: Class II Product Code: MBI Dated: May 17, 2013 Received: May 20, 2013

Dear Mr. Shahriari:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{4}------------------------------------------------

Page 2 - Mr. Kereshmeh Shahriari

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the clectronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (2)CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (2.1 CFR Part 803). please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincercly yours,

Erin I. Keith

For

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Quattro® GL Suture Anchor

Indications for Use

510(k) Number (if known): K131325

Device Name: Quattro® GL Suture Anchor

Indications for Use:

The Cayenne Medical, Inc. Quattro® GL Suture Anchor is intended for the reattachment of soft tissue to bone for the following indications:

Hip

  • -Hip capsule repair
    • o Acetabular labrum reattachment

Shoulder

  • Capsular stabilization
    • o Bankart repair
    • o Anterior shoulder instability
    • o SLAP lesion repairs
    • 0 Capsular shift or capsulolabral reconstructions
  • Acromioclavicular separation repairs -
  • Deltoid repairs l
  • Rotator cuff repairs ー
  • । Biceps tenodesis

Elbow, Wrist, and Hand

  • ー Biceps tendon reattachment
  • । Ulnar or radial collateral ligament reconstruction
  • Lateral epicondylitis repair t

Knee

  • ー Extra-capsular repairs
    • o Medial collateral ligament
    • o Lateral collateral ligament
    • o Posterior oblique ligament
  • । Patellar realignment and tendon repairs
    • o Vastus medials obliquous advancement
  • Illiotibial band tenodesis

{6}------------------------------------------------

510(k) Premarket Notification

Quattro® GL Suture Anchor

Foot and Ankle

  • Hallux valgus repairs
  • । Medial or Lateral instability repairs/reconstructions
  • Achilles tendon repairs/reconstructions l
  • r Midfoot reconstructions
  • ເ Metatarsal ligament/tendon repairs/reconstructions
  • Bunionectomy ।

Prescription Use _____________________________________________________________________________________________________________________________________________________________ x (Part 21 CFR 801 Subpart D)

Over-The-Counter Use AND/OR Over-The-Counter Use

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEFLOED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

············································································································································································· ・・・・・ Casey L. Hanley, Ph.D.

Division of Orthopedic Devices

Page 15 of 197

§ 888.3040 Smooth or threaded metallic bone fixation fastener.

(a)
Identification. A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.(b)
Classification. Class II.