(49 days)
The CardioSleeve System consisting of the CardioSleeve Front-End stethoscope with integrated ECG device. mobile heart sound and ECG recording application and the remote diagnostic heart murmur software is a decision support device intended to be used on a single patient to assist the qualified clinician in analyzing cardiac sounds and electrical signals for the identification and classification including of suspected murmurs. It is used to distinguish between normal/physiological and pathological heart murmurs by recording the acoustic signal of the heart and the ECG signal simultaneously and analyzing these signals. The acoustic heart signal is analyzed to identify specific heart sounds that may be present. Identified sounds include S1, S2, and suspected murmurs.
CardioSleeve indicates whether or not a recorded heart sound contains a suspected heart murmur. The device must be used in a clinical setting by trained personnel with the prescribed accessories and all relevant patient information must be taken into consideration before a diagnosis is made.
The interpretations of heart sounds offered by the CardioSleeve device are only significant when used in conjunction with physician over-read as well as consideration of all other relevant patient data.
CardioSleeve is not intended to be a diagnostic device. It does not supersede the judgment of the clinician. The device is intended to aid the physician in the evaluation of heart sounds. The clinicians are responsible for reviewing and interpreting the results, along with the auscultatory findings and medical history, when making a referral decision.
Caution: Federal (USA) law restricts this device to sale by or on the order of a clinician
The CardioSleeve with Diagnostic Heart Murmur Application, (application identical to the FDA cleared Sensi with Diagnostic Heart Murmur Software-K110704), is a decision support device intended to acquire, record, and analyze heart sounds. It is used to distinguish between normal and pathological heart murmurs by recording the acoustic signal of the heart and the ECG signal simultaneously and analyzing these signals, and to acquire, record and display 3 Lead ECG signals for diagnostic support.
The complete system comprises of: a) Diagnostic Heart Murmur Algorithm (algorithm identical to the FDA cleared Diagnostic Heart Murmur Algorithm in K110704), that runs on a hosted server environment with Linux operating system, b) a mobile device software application that captures patient data, displays patient data and instructions for use and c) a stethoscope front-end recorder device with integrated ECG to acquire the acoustic and electrical heart signals. The CardioSleeve Front-end recorder device will interface via Bluetooth with the clinician's hand held mobile device.
Here's a breakdown of the acceptance criteria and the study information based on the provided 510(k) summary:
Acceptance Criteria and Device Performance
| Acceptance Criteria | Reported Device Performance |
|---|---|
| Biocompatibility | CardioSleeve material presented no issues of biocompatibility |
| Electrical Safety | Meets the requirements of IEC 60601-1: 2005 + CORR. 1 (2006) + CORR. 2 (2007) |
| Feasibility & Usability | On average, all users scored the usability of the CardioSleeve Software more than 4 out of 5 |
| Comparative accuracy (vs. Sensi system) | CardioSleeve and Sensi achieve comparable accuracy of 70.8% |
| Diagnostic Heart Murmur Algorithm Specificity | 94% |
| Diagnostic Heart Murmur Algorithm Sensitivity | 91% |
| Acoustic Performance (frequency response) | Responds to heart sound frequency of 20hz to 500hz |
Study Details
1. Sample sized used for the test set and the data provenance:
- Test Set Size: 1568 heart sounds (for the algorithm validation study to distinguish between functional and pathological heart murmurs in the pediatric population).
- Data Provenance: Not explicitly stated regarding country of origin or whether it was retrospective/prospective. The description mentions "pediatric population," implying clinical data, but no further specifics are given.
2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not explicitly stated. The document mentions that the 1568 heart sounds "were accepted to meet the criteria of good quality and match the recorded pathological conditions," which implies an expert assessment, but the number and qualifications of those experts are not provided.
3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:
- Not explicitly stated.
4. If a multi-reader, multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No, a multi-reader, multi-case (MRMC) comparative effectiveness study with human readers assisted by AI was not explicitly described for this device. The documentation focuses on the device's performance in distinguishing murmurs and its equivalence to a predicate device. The device is described as a "decision support device" that should be used "in conjunction with physician over-read," but no specific study evaluating human performance improvement with or without AI assistance is detailed.
5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
- Yes, a standalone performance evaluation of the algorithm was conducted. The "Design verification of a CAA algorithm" (Clinical Audit Algorithm, presumably related to the diagnostic heart murmur algorithm) reported a Specificity of 94% and Sensitivity of 91%. Additionally, the study validating algorithms for distinguishing functional/pathological murmurs in the pediatric population (which used 1568 heart sounds) appears to be a standalone evaluation of the algorithm's performance against established ground truth.
6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- For the algorithm validating the distinction between functional and pathological heart murmurs, the ground truth was based on "recorded pathological conditions" and "criteria of good quality." This implies a clinical diagnosis established by experts, likely based on a combination of auscultatory findings, medical history, and potentially other diagnostic tests, leading to a categorization of the heart sound as normal/physiological or pathological. It is most likely expert diagnosis/consensus, although not explicitly named.
7. The sample size for the training set:
- Not explicitly stated in the provided document. The 1568 heart sounds are referred to as being "accepted to meet the criteria... and match the recorded pathological conditions," which sounds more like a test/validation set. The training set size for the underlying algorithm is not disclosed.
8. How the ground truth for the training set was established:
- Not explicitly stated, as the training set details are not provided. However, given that the algorithm is "identical to the FDA cleared Diagnostic Heart Murmur Algorithm in K110704" (the predicate device), it can be inferred that the ground truth for the training set would have been established similarly to the test set, likely through expert clinical assessment of heart sounds against "recorded pathological conditions" by qualified medical professionals.
{0}------------------------------------------------
510 (k) Summary 807.92(c)
JUN 2 4 2013
SPONSOR
Company Name: Company Address:
807.92(a)(1) Rijuven Corporation 624 Whispering Pines Drive Pittsburgh PA, 15238 +1 (301) 335-9163 +1 (412) 967-9393 Evens Augustin March 2, 2013
Telephone:
Fax:
Contact Person Summary Preparation Date:
DEVICE NAME
807.92(a)(2)
Trade Name: Common/Usual Name:
Classification Name: Regulation Number: Product Code: Device Class:
CardioSleeve Electronic Stethoscope / Heart Sound Analyzer Electronic Stethoscope; Phonocardiograph 21 CFR 870.1875, 870.2390 DQD, DQC Class II
PREDICATE DEVICE
807.92(a)(3)
Legally Marketed Equivalent Device
| Company | Product | 510(k) # |
|---|---|---|
| Diacoustic MedicalDevices (Pty) Ltd | Sensi with DiagnosticHeart Murmur Software | K110704 |
{1}------------------------------------------------
DEVICE DESCRIPTION 807.92(a)(4)
The CardioSleeve with Diagnostic Heart Murmur Application, (application identical to the FDA cleared Sensi with Diagnostic Heart Murmur Software-K110704), is a decision support device intended to acquire, record, and analyze heart sounds. It is used to distinguish between normal and pathological heart murmurs by recording the acoustic signal of the heart and the ECG signal simultaneously and analyzing these signals, and to acquire, record and display 3 Lead ECG signals for diagnostic support.
The complete system comprises of: a) Diagnostic Heart Murmur Algorithm (algorithm identical to the FDA cleared Diagnostic Heart Murmur Algorithm in K110704), that runs on a hosted server environment with Linux operating system, b) a mobile device software application that captures patient data, displays patient data and instructions for use and c) a stethoscope front-end recorder device with integrated ECG to acquire the acoustic and electrical heart signals. The CardioSleeve Front-end recorder device will interface via Bluetooth with the clinician's hand held mobile device.
DEVICE INDICATIONS FOR USE 807.92(a)(5)
The CardioSleeve System consisting of the CardioSleeve Front-End stethoscope with integrated ECG device, mobile heart sound and ECG recording application and the remote diagnostic heart murmur software is a decision support device intended to be used on a single patient to assist the qualified clinician in analyzing cardiac sounds and electrical signals for the identification and classification including of suspected murmurs. It is used to distinguish between normal/physiological and pathological heart murmurs by recording the acoustic signal of the heart and the ECG signal simultaneously and analyzing these signals. The acoustic heart signal is analyzed to identify specific heart sounds that mav be present. Identified sounds include S1, S2, and suspected murmurs.
CardioSleeve indicates whether or not a recorded heart sound contains a suspected heart murmur. The device must be used in a clinical setting by trained personnel with the prescribed accessories and all relevant patient information must be taken into consideration before a diagnosis is made.
The interpretations of heart sounds offered by the CardioSleeve device are only significant when used in conjunction with physician over-read as well as consideration of all other relevant patient data.
CardioSleeve is not intended to be a diagnostic device. It does not supersede the judgment of the clinician. The device is intended to aid the physician in the evaluation of heart sounds. The clinicians are responsible for reviewing and interpreting the results, along with the auscultatory findings and medical history. when making a referral decision.
Caution: Federal (USA) law restricts this device to sale by or on the order of a clinician
{2}------------------------------------------------
COMPARISON OF TECHNICAL CHARACTERISTICS 807.92(a)(6)
The CardioSleeve device features were directly compared with the FDA cleared Sensi with Diagnostic Heart Murmur Software, WelchAllyn Master Elite Stethoscope and the WelchAllyn Meditron Analyzer (K110704).
Synopsis of the comparison analysis:
-
. Intended Use:
CardioSleeve and Sensi with Diagnostic Heart Murmur Software do have an equivalent intended use. -
. Indications for Use:
CardioSleeve and Sensi with Diagnostic Heart Murmur Software do have identical indications for use. -
. Composition
CardioSleeve and Sensi with Diagnostic Heart Murmur Software compromises equivalent functions system composition. -
. Physical Properties
CardioSleeve with Front-end unit and Diagnostic Heart Murmur Software, and Sensi with Diagnostic Heart Murmur Software, the WelchAllyn Master Elite Stethoscope and the WelchAllyn Meditron Analyzer share equivalent physical characteristics (K110704). -
. Technology Characteristics
CardioSleeve consists of a front-end recorder device with a Bluetooth interface connected to a standard acoustic stethoscope recording both the acoustic heart sound and/or ECG signal. The Sensi consists of an electronic stethoscope and ECG recorder connecting to a PC by means of a USB interface for recording of acoustical body sounds.
CardioSleeve and Sensi with Diagnostic Heart Murmur Software are using identical heart murmur analysis algorithm.
CardioSleeve and Sensi with Diagnostic Heart Murmur Software are using equivalent user interface properties.
SAFETY AND EFFECTIVENESS
807.92(b)
A comprehensive list of verification and validation testing was performed in accordance to Rijuven's Design Control procedures.
Software validation was performed for all aspects of the CardioSleeve System and Software. The graphical user interface and usability were compared to the predicate devices.
Validation of the CardioSleeve was performed to ensure that the CardioSleeve system consistently fulfills its intended use and the needs of the user. A clinical software validation was performed to insure the performance of the software algorithm.
{3}------------------------------------------------
| 1 | Biocompatibility Testing | CardioSleeve material presented noissues of biocompatibility |
|---|---|---|
| 2 | Electrical Safety Testing | The CardioSleeve meets therequirements of IEC 60601-1: 2005 +CORR. 1 (2006) + CORR. 2 (2007). |
| 3 | Feasibility & Usability Study | Intended users performed usabilityvalidation within real life clinicalsettings.On average all users scored theusability of the CardioSleeve Softwaremore than 4 out of 5. |
| 4 | Comparative study betweenFDA approved CardioSleeveand Sensi system | CardioSleeve and Sensi achievecomparable accuracy of 70.8% |
| 5 | Design verification of a CAAalgorithm | Specificity of 94% and Sensitivity of91% |
| 6 | Validate algorithms used todistinguish between functionaland pathological heart murmursin the pediatric population. | 1568 heart sounds were accepted tomeet the criteria of good quality andmatch the recorded pathologicalconditions |
| 7 | Verification of the acousticperformance of theCardioSleeve front-end | The CardioSleeve Front-End performfavorably, in responding to heartsound frequency of 20hz to 500hz |
CONCLUSION
807.92(b)(3)
Based upon the indications for use, technology characteristics and safety and performance testing, it is the conclusion of Rijuven that the CardioSleeve device consisting of the CardioSleeve Front-End stethoscope and ECG device, mobile heart sound recording application and the remote diagnostic heart murmur software is as safe and effective as the predicate devices and raises no new issues of safety and effectiveness.
{4}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is a stylized symbol that resembles a person with outstretched arms.
June 24, 2013
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
Rijuven Corp Mr. E.J. Smith Regulatory Consultant 1468 Harwell Ave. Crofton, MD 21114 US
Re: K131287
Trade/Device Name: Cardiosleeve Regulation Number: 21 CFR 870.1875 Regulation Name: Electronic Stethoscope / Heart Sound Analyzer Regulatory Class: Class II Product Code: DQD, DQC Dated: May 9, 2013 Received: May 9, 2013
Dear Mr. E.J. Smith:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
{5}------------------------------------------------
Page 2 - Mr. E.J. Smith
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fdagov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, Misbranding by reference to premarket notification (21CFR Part 807.97), For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htmfor the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Bram D. Zuckerman -S
Bram D. Zuckerman, Ph.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
{6}------------------------------------------------
Indications for Use
510(k) Number (if known):
Device Name: CardioSleeve
Indications For Use:
The CardioSleeve System consisting of the CardioSleeve Front-End stethoscope with integrated ECG device. mobile heart sound and ECG recording application and the remote diagnostic heart murmur software is a decision support device intended to be used on a single patient to assist the qualified clinician in analyzing cardiac sounds and electrical signals for the identification and classification including of suspected murmurs. It is used to distinguish between normal/physiological and pathological heart murmurs by recording the acoustic signal of the heart and the ECG signal simultaneously and analyzing these signals. The acoustic heart signal is analyzed to identify specific heart sounds that may be present. Identified sounds include S1, S2, and suspected murmurs.
CardioSleeve indicates whether or not a recorded heart sound contains a suspected heart murmur. The device must be used in a clinical setting by trained personnel with the prescribed accessories and all relevant patient information must be taken into consideration before a diagnosis is made.
The interpretations of heart sounds offered by the CardioSleeve device are only significant when used in conjunction with physician over-read as well as consideration of all other relevant patient data.
CardioSleeve is not intended to be a diagnostic device. It does not supersede the judgment of the clinician. The device is intended to aid the physician in the evaluation of heart sounds. The clinicians are responsible for reviewing and interpreting the results, along with the auscultatory findings and medical history, when making a referral decision.
Caution: Federal (USA) law restricts this device to sale by or on the order of a clinician
Prescription Use X
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(Part 21 CFR 801 Subpart D)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Bram D. Zuckerman 2013.06.24 16:58:04 -
§ 870.1875 Stethoscope.
(a)
Manual stethoscope —(1)Identification. A manual stethoscope is a mechanical device used to project the sounds associated with the heart, arteries, and veins and other internal organs.(2)
Classification. Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.(b)
Electronic stethoscope —(1)Identification. An electronic stethoscope is an electrically amplified device used to project the sounds associated with the heart, arteries, and veins and other internal organs.(2)
Classification. Class II (special controls). The device, when it is a lung sound monitor, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.