K Number
K131210

Validate with FDA (Live)

Date Cleared
2014-03-18

(323 days)

Product Code
Regulation Number
880.2910
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

It is intended for use in measuring temperature in the human body (Armpit or Oral).

Device Description

The body temperature is converted into electronic signal by the temperature sensor, and then the electronic signal is converted into LCD digital display. The Electronic Thermometers TF3100, TF3101 and TF3102 have the same basic principles, main function, performance and intended use, and they are consistent in product structure and material.

AI/ML Overview

The provided PDF outlines the acceptance criteria and a study demonstrating that the Electronic Thermometer (Models TF3100, TF3101, TF3102) meets these criteria.

1. Table of Acceptance Criteria and Reported Device Performance:

The primary performance characteristics addressed in the submission are accuracy and response time. The acceptance criteria are based on recognized international and national standards for clinical electronic thermometers.

Acceptance Criteria (Standard)Reported Device Performance (Electronic Thermometer TF3100, TF3101, TF3102)
Accuracy (ASTM E1112-00)
$0.05°C$ for temperatures between $35.30°C$ and $39.00°C$$0.05°C$ (for $35.30°C ~ 39.00°C$)
$0.1°C$ for temperatures below $35.30°C$ or above $39.00°C$$0.1°C$ (for $< 35.30°C$ or $> 39.00°C$)
Response Time (Not explicitly stated as an acceptance criterion but matched)5 minutes
Precision and Repeatability4 numerical digits, display in 0.01 degree increments
Other Standards Compliance:Compliance demonstrated for:
Electrical Safety (IEC 60601-1)Yes
Home-used medical equipment requirements and Environmental test (IEC 60601-1-11)Yes
Electromagnetic Compatibility Requirements (EN 60601-1-2)Yes
Biocompatibility Evaluation (ISO 10993-5, ISO 10993-10)Yes (for parts contacted to patient)
Temperature Range$32.00°C ~ 42.00°C$
Ambient Temperature Environment$+5°C~+40°C$
Water-proofIP22

2. Sample size used for the test set and the data provenance:

The document states that "Clinical evaluation of Electronic Thermometers has been conducted by SHENZHEN PUMP MEDICAL SYSTEM CO., LTD." However, the specific sample size used for the test set and the data provenance (e.g., country of origin, retrospective or prospective) for this clinical evaluation are not provided in the given text.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

This information is not provided in the document. For a clinical thermometer, the ground truth would typically be established by highly accurate reference thermometers and not by "experts" in the sense of medical professionals adjudicating images or cases.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

This information is not applicable and therefore not provided in the context of evaluating a clinical electronic thermometer. Adjudication methods like 2+1 or 3+1 are typically used in studies involving expert review of medical images or diagnoses, not for the direct measurement of physical parameters like temperature.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

This information is not applicable and therefore not provided. The device is an electronic thermometer, not an AI-assisted diagnostic tool that would involve human readers.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

The device itself is a standalone electronic thermometer, designed to provide direct temperature readings. The "clinical evaluation" conducted would have assessed the device's accuracy in measuring human body temperature without human subjective interpretation. The document, however, does not explicitly describe the methodology or results of this standalone performance assessment beyond stating that "clinical evaluation... has been conducted."

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

For a clinical electronic thermometer, the ground truth would most likely be established by a highly accurate reference thermometer (e.g., a laboratory-grade mercury-in-glass thermometer or a highly calibrated electronic reference thermometer) under controlled conditions. The document does not explicitly state the type of ground truth used for the clinical evaluation, but it is inferable given the nature of the device.

8. The sample size for the training set:

This information is not applicable as the device is not an AI/ML product that would require a "training set." The submission describes a traditional electronic thermometer.

9. How the ground truth for the training set was established:

This information is not applicable as the device is not an AI/ML product that would require a "training set."

{0}------------------------------------------------

K131210

510(k) Summary

1. Applicant

Applicant Name:SHENZHEN PUMP MEDICAL SYSTEM CO., LTD.
Address:2/F West, M-7 Sinosteel Building, Maqueling Estate,Hi-Tech Industrial Park, Nanshan District,Shenzhen 518057, China

Contact person:

Name:Xie Qiongyu
Phone numbers:86-0755-26710795
Fax numbers:86-0755-26012025
E-mail:xieqy@bpump.com.cn
Date Prepared:2013-04-20

2. Device information

  • Trade name: Electronic Thermometer .
  • Model No.: TF3100, TF3101, TF3102 .
  • Classification name: Thermometer, electronic, clinical (per CER 880.2910) .
  • . Class: 2
  • . Requlation Medical Specialty : General Hospital
  • . Review Panel: General Hospital
  • Product code: FLL- Clinical Electronic Thermometer .
  • Regulation Description: Clinical electronic thermometer .
  • Regulation Number: 880.2910

Indications for Use: It is intended for use in measuring temperature in the human body (Armpit or Oral).

Predicates

Digital thermometer Model MT Series K-number: K062784 Product Code: FLL

Indications for Use:

The devices Model MT series (MT-201, MT-301, MT-402, and MT-403) are intended to measure the human body temperature in regular mode orally, rectally or under the arm, and the devices are reusable for clinical or home use on people of all ages.

Manufacture: SEJOY ELECTRONICS & INSTRUMENTS CO., LTD.

{1}------------------------------------------------

4. Description of the device

The body temperature is converted into electronic signal by the temperature sensor, and then the electronic signal is converted into LCD digital display.

The Electronic Thermometers TF3100, TF3101 and TF3102 have the same basic principles, main function, performance and intended use, and they are consistent in product structure and material.

ELEMENT OFCOMPARISONElectronic ThermometerModel No.: TF3100, TF3101,TF3102Manufacture: SHENZHEN PUMPMEDICAL SYSTEM CO., LTD.Digital thermometer MT Series(K062784)Manufacture: SEJOY ELECTRONICS& INSTRUMENTS CO., LTD.
thermometer typeTF3100, TF3101, TF3102MT-301
intended use(s)It is intended for use in measuringtemperature in the human body(Armpit or Oral).The devices Model MT series(MT-201, MT-301, MT-402, andMT-403) are intended to measure thehuman body temperature in regularmode orally, rectally or under thearm, and the devices are reusable forclinical or home use on people of allages.
componentsTemperature sensor, liquid crystal display, battery and circuit of motherboard
sensorThermistor
signal processing anddisplayUsing the resistance change of thermal resistor to detect body temperature,and displayed through the LCD.
power requirementsDC3.7V Rechargeable LithiumBatteryOne 1.5 V Button Battery model LR 41
MaterialsABS plastic and stainless steel
temperature range$32.00°C ~ 42.00°C$$32.00°C - 42.90°C$
ambient temperatureenvironment$+5°C~+40°C$$10°C ~ 35°C$
accuracy$0.05°C(35.30°C ~ 39.00°C)$$0.1°C( < 35.30°C or > 39.00°C )$$±0.1°C,35.5°C - 42.0°C$$±0.2°C under 35.5°C or over 42.0°C$
precision andrepeatability4 numerical digits, display in 0.01degree increments4 numerical digits, display in 0.01degree increments
Response time5 minutes5 minutes
Water-proofIP22IP22

5. Comparison to Predicate Devices

{2}------------------------------------------------

Comparing to the predicate, the subject devices used same the Measuring Principle, same electrical measurement technologic and same measurement site (Measuring on Armpit or Oral). The subject devices do not constitute a new intended use.

As a result, the subject device is Substantially Equivalent (SE) to the predicate which is US legally market device.

6. Discussion of Non-Clinical Tests Performed for Determination of Substantial

Equivalence

Laboratory testing was conducted to validate and verify that Electronic Thermometer met all requirements of related international standards, including electrical safety, EMC, biocompatibility, specification. Results of these tests demonstrate compliance to the requirements of the bellow consensus standards.

Standard:
Electrical Safety and performance requirements:
IEC 60601-1
ASTM E1112-00
Home-used medical equipment requirements and Environmental test:
IEC 60601-1-11
Electromagnetic Compatibility Requirements:
EN 60601-1-2
Biocompatibility Evaluation for the part contacted to patient
ISO 10993-5, ISO 10993-10

7. Clinical Evaluation

Clinical evaluation of Electronic Thermometers has been conducted by SHENZHEN PUMP MEDICAL SYSTEM CO., LTD.

8. Conclusion

As stated above, the Electronic Thermometer (Models: TF3100, TF3101 and TF3102) have the same intended use and similar technological characteristics as the cleared devices of the digital thermometer MT series (K062784).

{3}------------------------------------------------

Moreover, the Electronic Thermometer (Models: TF3100, TF3101 and TF3102) comply with the appropriate medical device standards. Verification and validation test reports contained in this submission demonstrate that any differences in their characteristics do not raise any new questions of safety or effectiveness.

Furthermore, those engineering differences do not affect the intended use or alter the fundamental scientific technology of the cleared devices of the digital thermometer MT series (K062784).

As a result, Electronic Thermometers (Models: TF3101 and TF3102) are safety and effective, and substantially equivalent to the earlier identified predicate devices.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the text "DEPARTMENT OF HEALTH & HUMAN SERVICES" in all caps. The text is in a bold, sans-serif font. The words are arranged on a single line and centered.

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular fashion around the caduceus symbol.

Public Health Service

Food and Drug Administration 0903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

March 18, 2014

SHENZHEN PUMP MEDICAL SYSTEM Company, Limited C/O Ivy Chen Shenzhen Huatongwei International Inspection Co., Ltd. Keji Nan No.12 Road, Hi-tech Park Shenzhen, Guangdong 518057 CHINA

Re: K131210

Trade/Device Name: Electronic Thermometer, Models TF3100, TF3101, TF3102 Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: Class II Product Code: FLL Dated: November 27, 2013 Received: March 13, 2014

Dear Ms. Chen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

|

{5}------------------------------------------------

Page 2 - Ms. Chen

Please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also. please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803). please go to

http://www.fda.gov/Medical/Devices/Safety/Reportal?roblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely vours.

Mary S. Runner -S

Erin 1. Keith. M.S. Acting Division Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K131210

Device Name

Electronic Thermometer, Models TF3100, TF3101, TF3102

Indications for Use (Describe)

(Armpit or Oral) . It is intended for use in measuring human body temperature

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

Image /page/6/Figure/11 description: The image shows a document with the header "FOR FDA USE ONLY". It indicates the concurrence of the Center for Devices and Radiological Health (CDRH) and includes a digital signature by Richard C. Chapman. The signature date is specified as 2014.03.18 at 09:36:56 with a time zone offset of -04'00'.

{7}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

§ 880.2910 Clinical electronic thermometer.

(a)
Identification. A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.(b)
Classification. Class II (performance standards). The device is exempt from the premarket notification procedures in part 807, subpart E of this chapter, subject to the limitations in § 880.9 and the following conditions for exemption:(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.