(37 days)
This device automatically measures human's Systolic, Diastolic blood pressure and heart rate by using the oscillometric method. All values can be read out in one LCD panel. Measurement position is at human being's upper arm. The intended use of this over-the-counter device is for adults aged 18 years and older with arm circumference ranging from 9 inches to 17 inches (approx.23 cm to 43 cm) and for home use.
When the device detects the appearance of irregular heartbeats during measurement, an indicated symbol will appear with measuring readings. And this device can let the memory data be transferred to the connected personal computer (PC) via USB cable.
HL858CA automatically measures human's Systolic, Diastolic blood pressure and heart rate by using the oscillometric method. All values can be read out in one LCD panel. Measurement position is at human being's upper arm. The intended use of this over-the-counter device is for adults aged 18 years and older with arm circumference ranging from 9 inches to 17 inches (approx.23 cm to 43 cm) and for home use.
The device will display a symbol "orman, to indicate the detection of irregular heartbeat rhythm as defined as a rhythm is more than or less than 25% from the average heartbeat intervals during the measurement. Additionally, after measurement, the Risk Category Indicator function will show the information with the readings on the screen for the user tracking their blood pressure level: Furthermore, the user can save and manage the measurement data by transferring the measured readings of blood pressure to the connected personal computer (PC) via USB cable.
Besides, When Triple Check mode is turned on by user, the symbol ( 3 mm ) will display on the LCD. Then press Start/Stop button the device will take three consecutive measurements automatically at 1 minute intervals. After measurements are completed, LCD will display the average values of the three measurements. And this device is designed with Rest Assure function as a countdown timer to help user in relax state for 5 minutes before taking measurement.
Here's a breakdown of the acceptance criteria and study information for the Full Automatic (NIBP) Blood Pressure Monitor, Model HL858CA, based on the provided text:
Acceptance Criteria and Reported Device Performance
The device's performance is primarily evaluated against the ANSI/AAMI SP10 standard for manual, electronic, or automated sphygmomanometers. The reported performance for the subject device is stated to be the same as the predicate device.
| Metric | Acceptance Criteria (Predicate Device) | Reported Device Performance (Subject Device) |
|---|---|---|
| Pressure Accuracy | +/- 3mmHg | +/- 3mmHg |
| Pulse Accuracy | +/- 5% | +/- 5% |
Note: The document states "All the relevant activities were performed by designate individual(s) and the results demonstrated that the predetermined acceptance criteria were fully met" in section 7, "Discussion of Clinical Tests Performed." This implies the subject device met the same accuracy criteria as the predicate.
Study Information
2. Sample size used for the test set and the data provenance:
- Sample Size for Test Set: The document does not explicitly state the sample size used for the test set. It only mentions compliance with the ANSI/AAMI SP10 standard.
- Data Provenance: The document does not specify the country of origin of the data or whether it was retrospective or prospective.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- This information is not provided in the document. The ANSI/AAMI SP10 standard typically involves a comparison to a reference measurement method (e.g., auscultation by trained observers), but the specifics of the expert panel or their qualifications are not detailed here.
4. Adjudication method for the test set:
- This information is not explicitly stated in the document.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- This is not applicable to this device. The device is a blood pressure monitor, not an AI-assisted diagnostic tool that would involve human readers interpreting data with or without AI assistance.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:
- Yes, the performance evaluation described in the document for accuracy (Pressure +/- 3mmHg, Pulse +/- 5%) is a standalone performance test of the device's ability to measure blood pressure and heart rate. It operates as an "algorithm only" in the sense that its measurements are automatically generated by the device itself, without a human actively interpreting data to arrive at the measurement.
7. The type of ground truth used:
- The ground truth for blood pressure monitor accuracy studies, as typically conducted under the ANSI/AAMI SP10 standard, involves simultaneous measurements using a recognized reference standard, often referred to as "auscultatory reference pressures" obtained by trained observers. The document states compliance with ANSI/AAMI SP10:2002/(R) 2008 & ANSVAAMI SP10:2002 /A1:2003/(R) 2008 & ANSI/AAMI SP10:2002/ A2:2006/(R) 2008, which implies that this type of ground truth was used.
8. The sample size for the training set:
- This information is not applicable as the device is not described as utilizing a machine learning algorithm that requires a "training set" in the conventional sense. The device uses an oscillometric method, which is a deterministic algorithm.
9. How the ground truth for the training set was established:
- This information is not applicable for the reason stated above.
{0}------------------------------------------------
K131121
PREMARKET NOTIFICATION
MAY 2 9 2013
510(k) SUMMARY
(As Required By 21 CFR 807.92)
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.
Date: The assigned 510(k) number is: _
APR 17 2013
1. Submitter:
Health & Life Co., Ltd. 9F. No.186, Jian Yi Road, Zhonghe District, New Taipei City, Taiwan, R.O.C TEL: +886-2-8227-1300 FAX: +886-2-8227-1301
Contact person: Sarah Su/ Regulatory Affairs Dept. E-mail: sarah.su@himt.com.tw Tel: 886-2-8227-1300 ext.1201 Fax: 886-2-8227-1301
2. Name of the Device:
Trade Name: Full Automatic (NIBP) Blood Pressure Monitor, Model HL858CA Common Name: Blood Pressure Monitor Classification Name: Non-invasive Blood Pressure Measurement System Classification: Class II
Regulation Number: 21 CFR 870.1130
Product Code: DXN
Panel: 74 Cardiovascular
3. Information for the 510(k) Cleared Device (Predicate Device):
Full Automatic (NIBP) Blood Pressure Monitor, Model HL868RT, K093831
4. Device Description:
HL858CA automatically measures human's Systolic, Diastolic blood pressure and heart
7-1
{1}------------------------------------------------
K131121
rate by using the oscillometric method. All values can be read out in one LCD panel. Measurement position is at human being's upper arm. The intended use of this over-the-counter device is for adults aged 18 years and older with arm circumference ranging from 9 inches to 17 inches (approx.23 cm to 43 cm) and for home use.
The device will display a symbol " orman, to indicate the detection of irregular heartbeat rhythm as defined as a rhythm is more than or less than 25% from the average heartbeat intervals during the measurement. Additionally, after measurement, the Risk Category Indicator function will show the information with the readings on the screen for the user tracking their blood pressure level: Furthermore, the user can save and manage the measurement data by transferring the measured readings of blood pressure to the connected personal computer (PC) via USB cable.
Besides, When Triple Check mode is turned on by user, the symbol ( 3 mm ) will display on the LCD. Then press Start/Stop button the device will take three consecutive measurements automatically at 1 minute intervals. After measurements are completed, LCD will display the average values of the three measurements. And this device is designed with Rest Assure function as a countdown timer to help user in relax state for 5 minutes before taking measurement.
5. Intended Use .
This device automatically measures human's Systolic, Diastolic blood pressure and heart rate by using the oscillometric method. All values can be read out in one LCD panel. Measurement position is at human being's upper arm. The intended use of this over-the-counter device is for adults aged 18 years and older with arm circumference ranging from 9 inches to 17 inches (approx.23 cm to 43 cm) and for home use.
When the device detects the appearance of irregular heartbeats during measurement, an indicated symbol will appear with measuring readings. And this device can let the memory data be transferred to the connected personal computer (PC) via USB cable.
6. Comparison of device to predicate device:
Product Specification Comparison Table of HL858CA and HL868RT (K093831)
| Item | Predicate DeviceHL868RT (K093831) | Subject DeviceHL858CA |
|---|---|---|
| Method ofmeasurement | Oscillimetric | Same as left |
| Range ofmeasurement | Pressure 0-300mmHg,Pulse 40-199 Beats/minute | Same as left |
| Accuracy | Pressure +/- 3mmHgPulse +/- 5% | Same as left |
| Inflation | Automatic inflation(Air pump) | Same as left |
| Deflation | Automatic air releasecontrol valve | Same as left |
| Exhaust | Automatic exhaust valve | Same as left |
| Display | Liquid Crystal DigitalDisplay | Same as left |
| Power Supply | AA (1.5V) Alkaline batteries×4 or6V 1A AC adapter | Same as left |
| Storage/TransportationEnvironment | -20℃ ~ +70℃(-4℉~+158℉),≤ 90% R.H. | -25℃ ~ +70℃(-13℉~+158℉),≤ 93% R.H. |
| OperatingEnvironment | 10℃~40℃(50℉~104℉),15%~90%R.H. | 5℃~40℃(41℉~104℉),15%~93% R.H. |
| Material | ABS housing andrubber keys | ABS housing andABS keys |
| Sets of memory | 2*60, total 120 | 2*120, total 240 |
| Number ofPush Button | 7 | 7+2 switch control(Triple check, Rest assure) |
| Storage pouch | Yes | Same as left |
| Cuff size | Arm circumference approx.23-43 cm (9~17 inches) | Same as left |
| Unit Weight | Approx. 293 ± 5g(Excluding Batteries) | Approx. 393 ± 10g(Excluding cuff and Batteries) |
Page 2 of 5
{2}------------------------------------------------
Changes from the predicate devices HL868RT (K093831):
*Changing of exterior casing design and the material of keys
:
- Additional product features of Triple check function, and Rest Assure Function.
:
{3}------------------------------------------------
These additional feature has been verified and validated and do not affect the safety and effectiveness of subject device HL858CA.
7. Discussion of Clinical Tests Performed:
HL858CA is compliant to the ANSI/AAMI SP10:2002/(R) 2008 & ANSVAAMI SP10:2002 /A1:2003/(R) 2008 & ANSI/AAMI SP10:2002/ A2:2006/(R) 2008 Manual, electronic or automated sphygmomanometers. All the relevant activities were performed by designate individual(s) and the results demonstrated that the predetermined acceptance criteria were fully met.
8. Discussion of Non-Clinical Tests Performed for Determination of Substantial
Equivalence are as follows:
The subject device was tested to evaluate its safety and effectiveness, including the followings:
a. Safety Test:
-
-IEC 60601-1:2005 Medical electrical equipment Part 1: General requirements for basic safety and essential performance
-IEC 60601-1-11:2010, Medical electrical equipment-Part 1-11: General Requirement for basic safety and essential performance - Collateral Standard: Requirements for medical electrical systems used in the home healthcare environment. -
b. EMC Test: IEC 60601-1-2 Edition 3:2007-03 Medical Electrical Equipment Part 1-2: General requirements for safety - collateral standard: Electromagnetic compatibility -Requirements and Tests.
-
c. Reliability Test: ANSI/AAMI SP10:2002/(R) 2008 & ANSI/AAMI SP10:2002 /A1:2003/(R) 2008 & ANSI/AAMI SP10:2002/ A2:2006/(R) 2008 Manual, electronic or automated sphygmomanometers.
-
d. Risk Assessment: ISO 14971:2007 Second Edition, Medical devices Application of risk management to medical devices
-
e. Software Verification and Validation: IEC 62304 Ed.1.0(2006), Medical device software - Software life cycle processes, and IEC 60601-1-4 Medical electrical equipment - Part 1-4: General requirements for safety - Collateral standard: Programmable electrical medical systems, edition 1.1
-
f. Usability Validation: IEC 62366:2007 Medical devices Application of usability engineering to medical devices
9. Conclusions:
The subject device was tested and fulfilled the requirements of those standards mentioned
{4}------------------------------------------------
.' ,
.
:
:
:
.
:
.
.
above, and it's concluded that the subject device is substantially equivalent to the predicate device.
November 1978
.
:
the country of the county of the county of
.
. . . .
7-5
:
:
. . .
1
.
.
.
.
. . . . . .
.
Page 5 of 5
: . . . . .
:
{5}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol that resembles an abstract human figure or a caduceus, rendered in black.
Public Health Service
Food and Drug A'dministration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 29, 2013
Health & Life Co., Ltd. c/o Ms. Sarah Su, Manager Regulatory Affairs Dept. 9F, No. 186, Jian Yi Road Zhonghe District, New Taipei Citv Taiwan 23553
Re: K131121
Trade/Device Names: Full Automatic (NIBP) Blood Pressure Monitor, Model HL858CA Regulatory Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II (Two) Product Code: DXN Dated: April 25, 2013 Received: April 29, 2013
Dear Ms. Su:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
Page 2 -- Ms: Sarah Su, Manager
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Owen 22%Earis -S
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
Indications for Use
510(k) Number (if known): K131121
Device Name: Full Automatic (NIBP) Blood Pressure Monitor, Model HL858CA
Indications for Use:
This device automatically measures human's Systolic, Diastolic blood pressure and heart rate by using the oscillometric method. All values can be read out in one LCD panel. Measurement position is at human being's upper arm. The intended use of this over-the-counter device is for adults aged 18 years and older with arm circumference ranging from 9 inches to 17 inches (approx.23 cm to 43 cm) and for home use.
When the device detects the appearance of irregular heartbeats during measurement, an indicated symbol will appear with measuring readings. And this device can let the memory data be transferred to the connected personal computer (PC) via USB cable.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR '
Over-The-Counter Use V (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Devices Evaluation (ODE)
Page 1 of 1
Owen P. Faris -S
2013.05.29
'09:34:35 -04'00'
§ 870.1130 Noninvasive blood pressure measurement system.
(a)
Identification. A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.(b)
Classification. Class II (performance standards).