K Number
K131093
Date Cleared
2013-05-21

(33 days)

Product Code
Regulation Number
888.3030
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Wright's ORTHOLOC® 3Di Ankle Plating System is intended for fixation of fractures, osteotomies, and non-unions of the distal tibia and fibula such as:

  • Lateral Malleolar Fractures
  • Syndesmosis injuries
  • Medial Malleolar Fractures
  • Bi-Malleolar Fractures
  • Tri-Malleolar Fractures
  • Posterior Malleolar Fractures
  • Distal Anterior Tibia Fractures
  • Vertical Shear Fractures of the Medial Malleolous
  • Pilon Fractures
  • Distal Tibia Shaft Fractures
  • Distal Fibula Shaft Fractures
  • Distal Tibia Periarticular Fractures
  • Medial Malleolar Avulsion Fractures
  • Lateral Malleolar Avulsion Fractures

ORTHOLOC® 3Di Locking Screws:
The ORTHOLOC® 3Di locking screws are intended for use with Wright's ORTHOLOC® 3Di Plating Systems of the same base material.

ORTHOLOC® Bone Screws:
ORTHOLOC® Screws are indicated for use in bone reconstruction, osteotomy, arthrodesis, joint fusion, fracture repair, and fracture fixation, appropriate for the size of the device.

Washer
Wright's washers are intended to prevent a screw head from breaking through the cortex of the bone by distributing the forces/load over a large area when used for fracture fixation of bone fragments.

Device Description

The ORTHOLOC® 3Di .Ankle Plating System contains plates belonging to 1 of 2 general categories-distal tibia and fibula-based on the contouring of each plate: All plates feature polyaxial locking screw holes. Some plates have k-wire holes, compression slots, or syndesmosis slots. The plates are made from titanium alloy and accept 2.7mm ORTHOLOC® 3Di locking screws. 2.7mm. 3.5mm and 4.0mm ORTHOLOC® Bone Screws, and 4.0mm Wright™ Compression Screws (cleared under K082320, now branded DART-FIRE®). Washers are also available for use with the ORTHOLOC® Bone Screws.

AI/ML Overview

Here's an analysis of the provided text regarding the Ortholoc® 3Di Ankle Plating System, focusing on acceptance criteria and study details:

1. A table of acceptance criteria and the reported device performance

The provided document (K131093) is a 510(k) summary for a medical device. For a 510(k) submission, the primary acceptance criteria for a new device are typically Substantial Equivalence to a legally marketed predicate device. This means the new device must be as safe and effective as the predicate.

The document does not outline specific, quantitative acceptance criteria in terms of performance metrics (e.g., bone healing rates, complication rates, specific mechanical load tolerances against a statistically defined threshold). Instead, it relies on a comparison to the predicate device.

Acceptance Criteria (Implicit from 510(k) Process)Reported Device Performance (as stated in document)
Substantial Equivalence in Design Features"The design features of the ORTHOLOC® 3Di Ankle Plating System - Line Addition are substantially equivalent to the design features of the predicate ORTHOLOC® 3Di Ankle Plating System"
Substantial Equivalence in Indications for Use"The ORTHOLOC® 3Di Ankle Plating System - Line Addition and the legally marketed predicate ORTHOLOC® 3Di Ankle Plating System have identical indications..."
Substantial Equivalence in Material, Sterilization, Selection, and Instrumentation"...utilize the same instrumentation, and are identical in material, sterilization methods, and selection."
Expected Performance at least as well as Predicate"From the evidence submitted in this 510(k), the subject line addition can be expected to perform at least as well as the predicate systems."
No New Worst-Case Scenario (Mechanical)"Through mechanical analysis the new plates do not represent a new worst-case."
No New Types of Safety or Effectiveness Questions"The design characteristics of the subject system do not raise any new types of questions of safety or effectiveness."

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document explicitly states: "(b)(2). Substantial Equivalence - Clinical Evidence N/A". This indicates that no clinical test set (human patient data) was used in this 510(k) submission. The comparison for substantial equivalence was based on non-clinical evidence (mechanical analysis and comparison of design/materials/indications to the predicate). Therefore, there is no sample size for a clinical test set, no data provenance (country, retrospective/prospective) to report for human studies.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

Since no clinical test set was used, and specifically no human data requiring expert assessment for ground truth, this information is not applicable to this submission. The "ground truth" for the submission was established by comparing the device's design and mechanical properties to those of the predicate device, likely by Wright Medical Technology's internal engineering and regulatory teams, and subsequently reviewed by the FDA.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

As no clinical test set was utilized requiring expert adjudication, this information is not applicable.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No MRMC study was done. This device is a bone plating system, not an imaging or diagnostic device that would typically involve human readers or AI assistance in interpretation. The nature of the device (implantable hardware) means this type of study is irrelevant.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This question is not applicable as the device is a bone plating system and does not involve algorithms or AI for diagnostic or interpretive purposes. Performance is related to its mechanical properties and biocompatibility as an implant.

7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

For this 510(k) submission, the "ground truth" for the device's safety and effectiveness was established by:

  • Comparison to the Predicate Device: The predicate device, ORTHOLOC® 3Di Ankle Plating System (K102429), serves as the primary "ground truth" reference. The new device must demonstrate it is as safe and effective as this previously cleared device.
  • Non-Clinical Mechanical Analysis: This involves engineering evaluations (e.g., stress testing, fatigue testing) to ensure the new plates do not represent a "new worst-case" scenario compared to the predicate. The "ground truth" for these tests would be established by validated engineering standards and methods.

8. The sample size for the training set

Since no clinical studies were performed, and no AI/algorithm components are part of this device, there is no training set in the context of machine learning or clinical trials. The "development" of the device would involve engineering design and validation, not a training set of data.

9. How the ground truth for the training set was established

As there is no training set for this device type within the context of the 510(k) submission, this question is not applicable.

{0}------------------------------------------------

K131093 page 1 of 2

Create Motion.® WRIC

510(K) SUMMARY OF SAFETY AND EFFECTIVENESS

In accordance with the Food and Drug Administration Rule to implement provisions of the Safe Medical Devices Act of 1990 and in conformance with 21 CFR 807.92, this information serves as a Summary of Safety and Effectiveness for the use of the ORTHOLOC® 3Di Ankle Plating System.

Wright Medical Technology, Inc. (a)(1). Submitted By: 5677 Airline Road Arlington, TN 38002 April 17, 2013 Date: Ryan Bormann Contact Person: Regulatory Affairs Specialist II Office - (901) 867-4409 Fax - (901) 867-4190 (a)(2). Proprietary Name: ORTHOLOC® 3Di Ankle Plating System Bone Plate System Common Name: 21 CFR 888.3030 - Class II Classification Name and Reference: HRS: Plate, Fixation Bone Device Product Code, Device Panel: (a)(3). Predicate Device: K 102429 ORTHOLOC® 3Di Ankle Plating System

(a)(4). Device Description .

4

The ORTHOLOC® 3Di .Ankle Plating System contains plates belonging to 1 of 2 general categories-distal tibia and fibula-based on the contouring of each plate: All plates feature polyaxial locking screw holes. Some plates have k-wire holes, compression slots, or syndesmosis slots. The plates are made from titanium alloy and accept 2.7mm ORTHOLOC® 3Di locking screws. 2.7mm. 3.5mm and 4.0mm ORTHOLOC® Bone Screws, and 4.0mm Wright™ Compression Screws (cleared under K082320, now branded DART-FIRE®). Washers are also available for use with the ORTHOLOC® Bone Screws.

The design features of the ORTHOLOC® 3Di Ankle Plating System - Line Addition are substantially equivalent to the design features of the predicate ORTHOLOC® 3Di Ankle Plating System

(a)(5). INTENDED USE

Ankle Plates:

Wright's ORTHOLOC® 3Di Ankle Plating System is intended for fixation of fractures, osteotomies, and non-unions of the distal tibia and fibula such as:

headquarters
Wright Medical Technology, Inc.5677 Airline Roadwmt.com

で、この上、このことになっていました。 この日には、このため、この時には、このため、この時には、この時には、この時には、いつでも、このと、あり、このことになるので、

A | - |

{1}------------------------------------------------

K131093 page 2 of 2

  • Lateral Malleolar Fractures
  • Syndesmosis injuries .
  • Medial Malleolar Fractures ●
  • Bi-Malleolar Fractures .
  • Tri-Malleolar Fractures ●
  • Posterior Malleolar Fractures �
  • Distal Anterior Tibia Fractures ●
  • Vertical Shear Fractures of the Medial Malleolous .
  • Pilon Fractures .
  • Distal Tibia Shaft Fractures �
  • Distal Fibula Shaft Fractures .
  • Distal Tibia Periarticular Fractures .
  • Medial Malleolar Avulsion Fractures
  • Lateral Malleolar Avulsion Fractures ●

ORTHOLOC® 3Di Locking Screws:

The ORTHOLOC® 3Di locking screws are intended for use with Wright's ORTHOLOC® 3Di Plating Systems of the same base material.

ORTHOLOC® Bone Screws:

ORTHOLOC® Screws are indicated for use in bone reconstruction, osteotomy, arthrodesis, joint fusion, fracture repair, and fracture fixation, appropriate for the size of the device.

Washer

Wright's washers are intended to prevent a screw head from breaking through the cortex of the bone by distributing the forces/load over a large area when used for fracture fixation of bone fragments.

(a)(6). Technological Characteristics Comparison

The ORTHOLOC® 3Di Ankle Plating System - Line Addition and the legally marketed predicate ORTHOLOC® 3Di Ankle Plating System have identical indications, utilize the same instrumentation, and are identical in material, sterilization methods, and selection.

(b)(1). Substantial Equivalence - Non-Clinical Evidence

Through mechanical analysis the new plates do not represent a new worst-case. ●

(b)(2). Substantial Equivalence - Clinical Evidence

N/A

(b)(3). Substantial Equivalence - Conclusions

The design characteristics of the subject system do not raise any new types of questions of safety or effectiveness. From the evidence submitted in this 510(k), the subject line addition can be expected to perform at least as well as the predicate systems.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or a stylized human figure with outstretched arms. The logo is printed in black and white.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

May 21, 2013

Wright Medical Technology, Incorporated % Mr. Ryan Bormann Regulatory Affairs Specialist II 5677 Airline Road Arlington, Tennessee 38002

Re: K131093

Trade/Device Name: ORTHOLOC® 3Di APS - Straight Tubular Plates Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: HRS Dated: April 17, 2013 Received: April 23, 2013

Dear Mr. Bormann:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent ffor the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21

{3}------------------------------------------------

Page 2 - Mr. Ryan Bormann

CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Image /page/3/Picture/8 description: The image shows the name "Erin D. Keith" in a stylized font. The first name, "Erin," is in a simple, bold typeface. The middle initial "D" is enclosed in a decorative, geometric shape. The last name, "Keith," is also in a bold font, with some decorative elements.

For

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use

510(k) Number (if known): K131093

page 1 of 2 ·

A3-1

Device Name: ORTHOLOC® 3Di APS - Straight Tubular Plates

Indications For Use:

Wright's ORTHOLOC® 3Di Ankle Plating System is intended for fixation of fractures, osteotomies, and non-unions of the distal tibia and fibula such as:

  • . Lateral Malleolar Fractures
  • Syndesmosis injuries .
  • Medial Malleolar Fractures �
  • Bi-Malleolar Fractures .
  • Tri-Malleolar Fractures .
  • Posterior Malleolar Fractures .
  • Distal Anterior Tibia Fractures ●
  • Vertical Shear Fractures of the Medial Malleolous .
  • . Pilon Fractures
  • Distal Tibia Shaft Fractures .
  • Distal Fibula Shaft Fractures .
  • Distal Tibia Periarticular Fractures .
  • Medial Malleolar Avulsion Fractures .
  • . Lateral Malleolar Avusion Fractures

ORTHOLOC® 3Di Locking Screws:

The ORTHOLOC® 3Di locking screws are intended for use with Wright's ORTHOLOC® 3Di Plating Systems of the same base material.

ORTHOLOC® Bone Screws:

ORTHOLOC® Screws are indicated for use in bone reconstruction, osteotomy, arthrodesis, joint fusion, fracture repair, and fracture fixation, appropriate for the size of the device.

Washer

Wright's washers are intended to prevent a screw head from breaking through the cortex of the bone by distributing the forces/load over a large area when used for fracture fixation of bone fragments.

{5}------------------------------------------------

K131093 page 2 of 2

Prescription Use _ xxx (Part 21 CFR 801 Subpart D)

AND/OR .

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) . . · ·

Elizabet标订%Erank -S

Division of Orthopedic Devices

Concurrence of CDRH, Office of Device Evaluation (ODE) 1 of 1

A3-2

§ 888.3030 Single/multiple component metallic bone fixation appliances and accessories.

(a)
Identification. Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.(b)
Classification. Class II.