(53 days)
No
The description details a standard wet chemistry analyzer and reagent cartridges, with no mention of AI or ML in the device operation, data analysis, or intended use. The performance studies are based on traditional analytical methods.
No.
This device is an in vitro diagnostic (IVD) test system designed for quantitative determination of total protein and albumin in blood samples, which aids in diagnosis and monitoring but does not provide direct therapy to a patient.
Yes
The document explicitly states that the S TEST Reagent Cartridge TP and ALB are "For in vitro diagnostic use only." It also details their use in measuring total protein and albumin in biological samples, which "are used in the diagnosis and treatment of a variety of diseases."
No
The device described is a reagent cartridge used with a physical clinical analyzer, which is a hardware system. The submission is for the reagent cartridges themselves, which are physical components containing chemical reagents.
Yes, this device is an IVD (In Vitro Diagnostic).
Here's why:
- Explicit Statement: The "Intended Use / Indications for Use" section clearly states: "For in vitro diagnostic use only."
- Intended Use: The device is intended for the quantitative determination of Total Protein (TP) and Albumin (ALB) in human biological samples (serum and plasma). These measurements are used in the diagnosis and treatment of various diseases.
- Device Description: The device is a wet chemistry system that analyzes samples and reagents to determine the concentration of analytes. This is a typical function of an IVD.
- Chemistry Reactions: The description of the chemical reactions used to measure TP and ALB confirms that the device is performing in vitro tests on biological samples.
- Performance Studies: The document details performance studies (analytical sensitivity, linearity, precision, interference testing, method comparisons, matrices comparisons) which are standard for demonstrating the analytical performance of an IVD.
- Predicate Devices: The mention of predicate devices (K100853 Cobas c systems TP2 and ALB2) which are also IVDs further supports that this device falls under the IVD category.
All these points align with the definition and characteristics of an In Vitro Diagnostic device.
N/A
Intended Use / Indications for Use
The S TEST Reagent Cartridge Total Protein (TP) is intended for the quantitative determination of TP in serum, lithium heparinized plasma, K3 EDTA plasma and sodium citrate plasma using the HITACHI Clinical Analyzer E40. The S TEST Reagent Cartridge TP is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only.
Total protein measurements are used in the diagnosis and treatment of a variety of diseases involving the liver, kidney, or bone marrow as well as other metabolic or nutritional disorders.
The S TEST Reagent Cartridge Albumin (ALB) is intended for the quantitative determination of ALB in serum, lithium heparinized plasma, K3 EDTA plasma and sodium citrate plasma using the HITACHI Clinical Analyzer E40. The S TEST Reagent Cartridge ALB is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only.
Albumin measurements are used in the diagnosis and treatment of numerous diseases involving primarily the liver or kidneys.
Product codes (comma separated list FDA assigned to the subject device)
CIX, JGQ
Device Description
The Hitachi Clinical Analyzer is an automatic, bench-top, wet chemistry system intended for use in clinical laboratories or physician office laboratories. The instrument consists of a desktop analyzer unit, an operations screen that prompts the user for operation input and displays data, a printer, and a unit cover. The analyzer unit includes a single probe, an incubation rotor, carousels for sample cups and reagent cartridges, and a multi-wavelength photometer. The single-use reagent cartridges may be placed in any configuration on the carousel, allowing the user to develop any test panel where the reagent cartridges are available.
The S TEST reagent cartridges are made of plastic and include two small reservoirs capable of holding two separate reagents (R1 and R2), separated by a reaction cell/photometric cuvette. The cartridges also include a dot code label that contains all chemistry parameters. calibration factors, and other production-related information, e.g., expiration dating. The dimensions of the reagent cartridges are: 13.5 mm (W) × 28 mm (D) × 20.2 mm (H).
System operation: After the sample cup is placed into the carousel, the analyzer pipettes the sample, pipettes the reagent, and mixes (stirs) the sample and reagent together. After the sample and reagent react in the incubator bath, the analyzer measures the absorbance of the sample, and based on the absorbance of the reactions, it calculates the concentration of analyte in the sample. The test system can measure analytes in serum or plasma and results are available in approximately 15 minutes per test. This submission is for Reagent Cartridges TP and ALB.
Chemistry reactions:
(TP) Proteins in samples react with the biuret reagent to form a purplered complex. The concentration of total protein can be determined by measuring the absorbance of the purple-red substance.
(ALB) Albumin in the sample combines with bromcresol green to form a blue-green dye conjugate. The albumin concentration is directly proportional to the color intensity and can be determined photometrically by measuring the absorbance of this resulting blue-green color.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
Not Found
Indicated Patient Age Range
Not Found
Intended User / Care Setting
clinical laboratories or physician office laboratories.
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
Nonclinical data:
Analytical Sensitivity (Limits of Detection)- TP: The limit of detection was found to be 0.2 g/dL. The quantitation limit was found to be 0.2 g/dL.
Analytical Sensitivity (Limits of Detection)- ALB: The limit of detection was found to be 0.1 g/dL. The quantitation limit was found to be 0.5 g/dL.
Linearity- TP: The range of linearity was 0.2 to 11.0 g/dL. The reportable range is 0.2 to 11.0 g/dL.
Linearity- ALB: The range of linearity was 0.1 to 8.0 g/dL. The reportable range is 0.5 to 8.0 g/dL.
20-day In-house Precision- TP: %CVs between 1.8% to 2.5%.
20-day In-house Precision- ALB: %CVs between 1.6% to 4.8%.
Interference Testing: TP test system was not affected by high levels of unconjugated bilirubin (up to 50 mg/dL), lipemia (up to 500 mg/dL), ascorbic acid (up to 50 mg/dL), and hemoglobin (up to 1,000 mg/dL). ALB test system was not affected by high levels of hemoglobin (up to 250 mg/dL), unconjugated bilirubin (up to 12.5 mg/dL), lipemia (up to 500 mg/dL), and ascorbic acid (up to 50 mg/dL).
Method Comparison - TP: A total of 115 clinical specimens (0.8 g/dL to 10.9 g/dL) were assayed. Deming regression: r = 0.989, Slope = 1.02 (1.01 to 1.04), y-intercept = 0.01 (-0.13 to 0.15).
Method Comparison - ALB: A total of 118 clinical specimens (0.5 g/dL to 6.4 g/dL) were assayed. Deming regression: r = 0.975, Slope = 1.01 (0.96 to 1.06), y-intercept = 0.24 (0.06 to 0.41).
Matrices Comparisons- TP: 45 matched serum/plasma samples (0.5 to 10.5 g/dL). Linear regression results for Heparinized Plasma (Slope=1.00, r=0.989), EDTA Plasma (Slope=1.00, r=0.992), Na Citrate Plasma (Slope=0.98, r=0.987).
Matrices Comparisons- ALB: 41 matched serum/plasma samples (1.0 to 7.1 g/dL). Linear regression results for Heparinized Plasma (Slope=0.99, r=0.992), EDTA Plasma (Slope=0.95, r=0.995), Na Citrate Plasma (Slope=1.00, r=0.986).
Clinical Data:
External site precision study (POL-type sites):
TP: 3 sites, each with 3 blinded serum samples (low, middle, high concentrations), assayed 6 times/day for 5 days (30 results/level). Total Precision %CVs ranged from 0.7% to 4.4%.
ALB: 3 sites, each with 3 blinded serum samples, assayed 6 times/day for 5 days. Total Precision %CVs ranged from 0.0% to 4.8%.
External method comparison studies (POL-type sites):
TP: Approximately 50 - 80 serum specimens (0.6 to 10.9 g/dL) across 3 sites. Deming regression results: Site 1 (n=52, r=0.996, y=0.98x+0.14), Site 2 (n=52, r=0.994, y=1.00x-0.07), Site 3 (n=53, r=0.996, y=0.96x+0.03).
ALB: Approximately 50 - 80 serum specimens (0.5 to 6.7 g/dL) across 3 sites. Deming regression results: Site 1 (n=87, r=0.982, y=0.99x+0.24), Site 2 (n=81, r=0.979, y=0.95x+0.30), Site 3 (n=81, r=0.985, y=0.91x+0.35).
Conclusion: The test systems were shown to be safe and effective for their intended uses.
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Linearity, Precision, Limits of Detection, Regression (slope, y-intercept, r), %CV.
Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.
Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.
Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).
Not Found
§ 862.1635 Total protein test system.
(a)
Identification. A total protein test system is a device intended to measure total protein(s) in serum or plasma. Measurements obtained by this device are used in the diagnosis and treatment of a variety of diseases involving the liver, kidney, or bone marrow as well as other metabolic or nutritional disorders.(b)
Classification. Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 862.9.
0
JUN 0 7 2013
SECTION 8 510(k) SUMMARY
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. The assigned 510(k) number is K131051.
| 807.92 (a)(1): Name: | Address: | Hitachi Chemical Diagnostics
630 Clyde Court
Mountain View, CA 94043 |
|----------------------|----------|----------------------------------------------------------------------------|
| | Phone: | (650) 961 5501 |
| | FAX: | (650) 969 2745 |
| | Contact: | Mr. Charles Tsou |
807.92 (a)(2): Device name- trade name and common name, and classification
Trade name:
S TEST Reagent Cartridge Total Protein (TP) S TEST Reagent Cartridge Albumin (ALB)
Common Name: Routine chemistry analyzer for TP Routine chemistry analyzer for ALB
Classifications: 21 CFR § 862.1635 Total Protein (TP) 21 CFR § 862.1035 Albumin (ALB)
807.92 (a)(3): Identification of the legally marketed predicate devices
Cobas c systems TP2 (Roche Diagnostics, Inc., Indianapolis, IN)- K100853 Cobas c systems ALB2 (Roche Diagnostics, Inc., Indianapolis, IN)- K100853
807.92 (a)(4): Device Description
The Hitachi Clinical Analyzer is an automatic, bench-top, wet chemistry system intended for use in clinical laboratories or physician office laboratories. The instrument consists of a desktop analyzer unit, an operations screen that prompts the user for operation input and displays data, a printer, and a unit cover. The analyzer unit includes a single probe, an incubation rotor, carousels for sample cups and reagent cartridges, and a multi-wavelength photometer. The single-use reagent cartridges may be placed in any configuration on the carousel, allowing the user to develop any test panel where the reagent cartridges are available.
The S TEST reagent cartridges are made of plastic and include two small reservoirs capable of holding two separate reagents (R1 and R2), separated by a reaction cell/photometric
Page 1 of 9
Hitachi Chemical Diagnostics, Inc.
630 Clyde Court. Mountain View. CA 94043-2239 Tel: 800 233 6278
のです。 このため、アイテム、アイデント、アイデント、アイデン、アイデン、アイテリアのアイデリアのアイデリアのアイデリアのアイデリアのアイデリアのアイデリアのアイデリアのアイデリアのアイデリアのアイデリアのアニメーションの | |
---|---|
www.hcqiaonostics.com | |
-15 Section |
(1) | |
1
cuvette. The cartridges also include a dot code label that contains all chemistry parameters. calibration factors, and other production-related information, e.g., expiration dating. The dimensions of the reagent cartridges are: 13.5 mm (W) × 28 mm (D) × 20.2 mm (H).
System operation: After the sample cup is placed into the carousel, the analyzer pipettes the sample, pipettes the reagent, and mixes (stirs) the sample and reagent together. After the sample and reagent react in the incubator bath, the analyzer measures the absorbance of the sample, and based on the absorbance of the reactions, it calculates the concentration of analyte in the sample. The test system can measure analytes in serum or plasma and results are available in approximately 15 minutes per test. This submission is for Reagent Cartridges TP and ALB.
Chemistry reactions: (TP) Proteins in samples react with the biuret reagent to form a purplered complex. The concentration of total protein can be determined by measuring the absorbance of the purple-red substance.
(ALB) Albumin in the sample combines with bromcresol green to form a blue-green dye conjugate. The albumin concentration is directly proportional to the color intensity and can be determined photometrically by measuring the absorbance of this resulting blue-green color.
807.92 (a)(5): Intended Use
The S TEST Reagent Cartridge Total Protein (TP) is intended for the quantitative determination of TP in serum, lithium heparinized plasma, K3 EDTA plasma and sodium citrate plasma using the HITACHI Clinical Analyzer E40. The S TEST Reagent Cartridge TP is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only.
Total protein measurements are used in the diagnosis and treatment of a variety of diseases involving the liver, kidney, or bone marrow as well as other metabolic or nutritional disorders.
The S TEST Reagent Cartridge Albumin (ALB) is intended for the quantitative determination of ALB in serum, lithium heparinized plasma, K3 EDTA plasma and sodium citrate plasma using the HITACHI Clinical Analyzer E40. The S TEST Reagent Cartridge ALB is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only.
Albumin measurements are used in the diagnosis and treatment of numerous diseases involving primarily the liver or kidneys.
Page 2 of 9
(2) Hitachi Chemical Diagnostics, Inc. 630 Clyde Court, Mountain View, CA 94043-2239 Tel: 800 233 5278
2
807.92 (a)(6): Technological Similarities and Differences to the Predicate
The following chart describes similarities and differences between the TP test systems.
Characteristic | Hitachi S TEST Systems | PREDICATE |
---|---|---|
Instrument Platform | Hitachi Clinical Analyzer | |
(originally cleared under K111753) | Roche cobas c systems - K100853 | |
Total Protein (TP) | K number- 131051 | Roche K number- K100853 |
Device Class, Regulation Code | Class II, Exempt, Reserved, 21 CFR | |
862.1635 | Same, except Exempt (not POC) | |
Classification Product Code | JGQ | Same |
Intended Use | Quantitative determination of TP | Same |
Testing Environment | Physician office or clinical lab | Clinical lab |
Test Principle | Proteins in samples react with the | |
biuret reagent to form a purple-red | ||
complex. The concentration of total | ||
protein can be determined by | ||
measuring the absorbance of the | ||
purple-red substance. | Divalent copper reacts with protein | |
peptide bonds to form the | ||
characteristic purple-colored biuret | ||
complex. The color intensity is | ||
directly proportional to the protein | ||
concentration which is determined | ||
photometrically | ||
Specimen Type | Human serum or plasma | Same |
Reportable Range | 0.2 to 11.0 g/dL | 0.2 to 12.0 g/dL |
Detection Wavelength | 546/700 nm | Same |
Detection Limit | 0.2 g/dL | Same |
Linearity | 0.2 to 11.0 g/dL | 0.2 to 12.0 g/dL |
Precision | %CVs range from 1.8% to 2.5% | %CVs range from 0.9% to 2.5% |
Page 3 of 9
( Hitachi Chemical Diagnostics, Inc.
630 Clyde Court, Mountain View, CA 94043-2239 Tel: 800 233 6278 Fax: 650 969 2745
-------www.hcdiagnostics.com ﺐ ﺑﺎﻟﻤ ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
3
Characteristic | Hitachi S TEST Systems | PREDICATE |
---|---|---|
Instrument Platform | Hitachi Clinical Analyzer | |
(originally cleared under K111753) | Roche cobas c systems - K100853 | |
Albumin (ALB) | ||
Device Class, Regulation Code | K number- K131051 | Roche K number- K100853 |
Classification Product Code | Class II, 21 CFR 862.1035 | 'Same' |
Intended Use | CIX | 'Same' |
Testing Environment | Quantitative determination of ALB | 'Same' |
Test Principle | Physician office or clinical lab | Clinical lab |
Albumin in the sample combines | ||
with bromcresol green to form a | ||
blue-green dye conjugate. The | ||
albumin concentration is directly | ||
proportional to the color intensity | ||
and can be determined | ||
photometrically by measuring the | ||
absorbance of this resulting blue- | ||
green color. | 'Same' | |
Specimen Type | Human serum or plasma | Human serum, plasma, or urine |
Reportable Range | 0.5 to 7.1 g/dL | 0.2 to 6.0 g/dL |
Detection Wavelength | 660/700 nm | 570/505 nm |
Detection Limit | 0.1 g/dL | 0.2 g/dL |
Linearity | 0.1 to 8.0 g/dL | 0.2 to 6.0 g/dL |
Precision | %CVs range from 1.6% to 4.8% | %CVs range from 0.7% to 1.2% |
The following chart describes similarities and differences between the ALB test systems.
807.92 (b)(1): Brief Description of Nonclinical Data
A series of studies were performed that evaluated the following nonclinical performance characteristics for each analytical sensitivity (limits of detection), linearity, 20-day in-house precision, interference testing, in-house method comparisons, and matrices comparison between serum and various plasma types.
Analytical Sensitivity (Limits of Detection)- TP
The study followed CLSI EP17-A, and the limit of detection was found to be 0.2 g/dL. The quantitation limit was found to be 0.2 g/dL
Analytical Sensitivity (Limits of Detection)- ALB
The study followed CLSI EP17-A, and the limit of detection was found to be 0.1 g/dL. The quantitation limit was found to be 0.5 g/dL.
Linearity- TP
The study followed CLSI EP-6A, and the range of linearity was 0.2 to 11.0 g/dL. The reportable range is 0.2 to 11.0 g/dL.
Page 4 of 9
O Hitachi Chemical Diagnostics
630 Clyde Court, Mountain View, CA 94043-2239 Tel: 800 233 6278 Fax: 650 969 2745
4
Linearity- ALB
The study followed CLSI EP-6A, and the range of linearity was 0.1 to 8.0 g/dL. The reportable range is 0.5 to 8.0 g/dL.
20-day In-house Precision- TP
The studies followed CLSI EP5-A2, where three levels of samples were each tested in two runs, twice a day, for 20 days. The results were as follows:
Precision Summary:
TP- Low, Level 1, Summary
TP | Within-Run | Total |
---|---|---|
Mean (g/dL) | 4.19 | 4.19 |
SD (g/dL) | 0.05 | 0.09 |
%CV | 1.1% | 2.1% |
TP- Middle, Level 2, Summary
TP | Within-Run | Total |
---|---|---|
Mean (g/dL) | 5.51 | 5.51 |
SD (g/dL) | 0.08 | 0.14 |
%CV | 1.4% | 2.5% |
TP- High, Level 3, Summary
TP | Within-Run | Total |
---|---|---|
Mean (g/dL) | 7.19 | 7.19 |
SD (g/dL) | 0.07 | 0.13 |
%CV | 1.0% | 1.8% |
20-day In-house Precision- ALB
The studies followed CLSI EP5-A2, where three levels of samples were each tested in two runs, twice a day, for 20 days. The results were as follows:
Precision Summary:
ALB- Low, Level 1, Summary
ALB | Within-Run | Total |
---|---|---|
Mean (g/dL) | 2.28 | 2.28 |
SD (g/dL) | 0.04 | 0.11 |
%CV | 1.5% | 4.8% |
Page 5 of 9
O Hitachi Chemical Diagnostics, Inc.
5
ALB | Within-Run | Total |
---|---|---|
Mean (g/dL) | 4.72 | 4.72 |
SD (g/dL) | 0.07 | 0.13 |
%CV | 1.4% | 2.8% |
ALB- Middle, Level 2, Summary
ALB- High, Level 3, Summary
ALB | Within-Run | Total |
---|---|---|
Mean (g/dL) | 6.01 | 6.01 |
SD (g/dL) | 0.07 | 0.09 |
%CV | 1.2% | 1.6% |
Interference Testing (per CLSI EP7-A2)
The data demonstrated that the TP test system was not affected by high levels of the following substances at the levels noted:
Unconjugated bilirubin: no interference up to 50 mg/dL Lipemia: no interference up to 500 mg/dL Ascorbic acid: no interference up to 50 mg/dL Hemoglobin: no interference up to 1,000 mg/dL
Lack of interference was defined as recoveries between 90% and 110% of the neat value, and assay performance claims were established on the HITACHI Clinical Analyzer by testing two serum pools containing approximately 4.0 and 6.5 g/dL TP.
The data demonstrated that the ALB test system was not affected by high levels of the following substances at the levels noted:
Hemoglobin: no interference up to 250 mg/dL Unconjugated bilirubin: no interference up to 12.5 mg/dL Lipemia: no interference up to 500 mg/dL Ascorbic acid: no interference up to 50 mg/dL
Lack of interference was defined as recoveries between 90% and 110% of the neat value, and assay performance claims were established on the HITACHI Clinical Analyzer by testing two serum pools containing approximately 2.5 and 4.0 g/dL ALB
Method Comparison - TP
A total of 115 clinical specimens spanning the dynamic range (0.8 g/dL to 10.9 g/dL), were assayed in singleton and in a blinded fashion by both the Hitachi E40 system and a standard laboratory system. The comparative data were analyzed by Deming regression and are shown below. (CI = confidence interval)
Page 6 of 9
Hitachi Chemical Diagnostics, Inc.
| 1 ============================================================================================================================================================================
And and every and the comments of any and on the really and in the first | |
---|---|
aruuuu | |
Section of Schools And | |
Therefore the fine " very Called on annual commend on the perform a compares of the came of the came of the comments of | |
In answer, conventure and the backs befor has be a den at a commend of the comments of the comments of the mail | |
March A |
6
TP Regression Statistics:
| n | r | Slope
(95% CI) | y-intercept
(95% CI) | X mean | Y mean |
|-----|-------|------------------------|-------------------------|--------|--------|
| 115 | 0.989 | 1.02
(1.01 to 1.04) | 0.01
(-0.13 to 0.15) | 7.0 | 7.2 |
Method Comparison - ALB
A total of 118 clinical specimens spanning the dynamic range (0.5 g/dL to 6.4 g/dL), were assayed in singleton and in a blinded fashion by both the Hitachi E40 system and a standard laboratory system. The comparative data were analyzed by Deming regression and are shown below. (CI = confidence interval)
ALB Regression Statistics:
| n | r | Slope
(95% CI) | y-intercept
(95% CI) | X mean | Y mean |
|-----|-------|------------------------|-------------------------|----------|-----------|
| 118 | 0.975 | 1.01
(0.96 to 1.06) | 0.24
(0.06 to 0.41) | 4.0 g/dL | 4.3 g/dL. |
Matrices Comparisons- TP
A study was performed to validate the use of three plasma types as an alternative to serum for the Hitachi Clinical Analyzer with S TEST Reagent Cartridge TP. The plasma types were sodium citrate, K3 EDTA, and lithium heparin. Forty-five (45) matched serum/plasma samples that spanned the dynamic range (0.5 to 10.5 g/dL) were assayed in singleton and the results were compared using linear regression (plasma = y-axis, each type). The performance characteristics were as follows.
n = 45, range (serum) = 0.5 to 10.5 g/dL TP
Heparinized Plasma | EDTA Plasma | Na Citrate Plasma | |
---|---|---|---|
Slope (95% CIs) | 1.00 (0.96 to 1.04) | 1.00 (0.96 to 1.04) | 0.98 (0.93 to 1.03) |
y-intercept (95% CIs) | -0.11(-0.43 to -0.21) | -0.06 (-0.33 to 0.22) | -0.09 (-0.45 to 0.26) |
r | 0.989 | 0.992 | 0.987 |
Matrices Comparisons- ALB
A study was performed to validate the use of three plasma types as an alternative to serum for the Hitachi Clinical Analyzer with S TEST Reagent Cartridge ALB. The plasma types were sodium citrate, K3 EDTA, and lithium heparin. Forty-one (41) matched serum/plasma samples that spanned the dynamic range (1.0 to 7.1 g/dL) were assayed in singleton and the results were compared using linear regression (plasma = y-axis, each type). The performance characteristics were as follows.
N = 41, Range (serum) = 1.0 to 7.1 g/dL ALB
Heparinized Plasma | EDTA Plasma | Na Citrate Plasma | |
---|---|---|---|
Slope (95% CIs) | 0.99 (0.95 to 1.03) | 0.95 (0.92 to 0.98) | 1.00 (0.94 to 1.05) |
y-intercept (95% CIs) | -0.01 (-0.20 to 0.18) | 0.22 (0.08 to 0.36) | -0.22 (-0.48 to -0.03) |
r | 0.992 | 0.995 | 0.986 |
Page 7 of 9
Hitachi Chemical Diagnostics, Inc.
. | |||
---|---|---|---|
: | and the control of the country of the country of the county of the county of the county of | ||
www.hcdiagnostics.com | |||
1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 |
7
807.92 (b)(2): Brief Description of Clinical Data
Studies for precision and method comparison (accuracy) were performed at three external POL-type sites to evaluate the Hitachi E40 Clinical Analyzer with S TEST Reagent Cartridges TP and ALB in one of its targeted intended use environments, the physician's office laboratory.
For the external site precision study, each site received three blinded serum samples (the Precision Panel, labeled A, B, and C) that were chosen to represent low, middle, and high concentrations of TP or ALB. Each sample was assayed six times per day for five days, reporting 30 results per level. Precision estimates for total precision were as follows:
| Site # | Sample | Mean
(g/dL) | Within-run Precision | | Total Precision | |
|--------|--------|----------------|----------------------|------|-----------------|------|
| | | | SD (g/dL) | %CV | SD (g/dL) | %CV |
| 1 | A | 4.1 | 0.05 | 1.2% | 0.07 | 1.6% |
| 2 | A | 4.1 | 0.05 | 1.1% | 0.07 | 1.6% |
| 3 | A | 3.8 | 0.05 | 1.4% | 0.13 | 3.5% |
| 1 | B | 5.5 | 0.05 | 0.8% | 0.06 | 1.1% |
| 2 | B | 5.5 | 0.06 | 1.2% | 0.07 | 1.2% |
| 3 | B | 5.0 | 0.04 | 0.9% | 0.20 | 4.0% |
| 1 | C | 7.1 | 0.05 | 0.7% | 0.05 | 0.7% |
| 2 | C | 7.1 | 0.06 | 0.8% | 0.07 | 0.9% |
| 3 | C | 6.5 | 0.07 | 1.1% | 0.28 | 4.4% |
TP (g/dL) n = 30 renlicates ner sample per site
ALB (g/dL) n = 30 replicates per sample per site
Site # | Sample | Mean | Within-run Precision | Total Precision | ||
---|---|---|---|---|---|---|
(g/dL) | SD (g/dL) | %CV | SD (g/dL) | %CV | ||
1 | A | 0.88 | 0.03 | 3.9% | 0.04 | 4.8% |
2 | A | 0.80 | 0.00 | 0.0% | 0.00 | 0.0% |
3 | A | 0.81 | 0.02 | 2.8% | 0.04 | 4.5% |
1 | B | 4.67 | 0.05 | 1.0% | 0.04 | 1.3% |
2 | B | 4.60 | 0.05 | 1.6% | 0.06 | 1.7% |
3 | B | 4.47 | 0.12 | 2.7% | 0.13 | 2.8% |
1 | C | 7.03 | 0.05 | 0.8% | 0.18 | 2.5% |
2 | C | 6.93 | 0.06 | 0.8% | 0.11 | 1.6% |
3 | C | 6.72 | 0.08 | 1.2% | 0.16 | 2.3% |
Page 8 of 9
C Hitachl Chemical Diagnostics, Inc.
ского и се вести и и полнение приведение пристепения пристепения пристепения в пристемента в пристеме се пристеме се пристем се пристеме се пристеме се пристеме се пристическ | ||
---|---|---|
www.hodiannostics.com | ||
1 1 44 . 1 1 . 1 . 1 . 1 . 1 . 1 . 1 . 1 . 1 . 1 . 1 . 1 . 1 . 1 . 1 . 1 . 1 . 1 . 1 . 1 . 1 . 1 . 1 . 1 . 1 . 1 . 1 . 1 . 1 . 1 . 1 . 1 . 1 . 1 . 1 . 1 . 1 . 1 . 1 . 1 . 1 . |
8
For the external method comparison studies, a series of approximately 50 - 80 serum specimens with TP values ranging from 0.6 to 10.9 g/dL, and ALB values ranging from 0.5 to 6.7 g/dL, were assayed on the Hitachi E40 Clinical Analyzer at three sites using S TEST Reagent Cartridges TP and ALB (y) and a comparative method as the reference method (x). Deming regression analyses yielded the following results:
| Site # | n | Range
(g/dL) | Regression
Equation | "r" | CI*
Slope | CI Intercept |
|--------|----|-----------------|------------------------|-------|--------------|---------------|
| 1 | 52 | 0.8 to 10.8 | y = 0.98x + 0.14 | 0.996 | 0.96 to 1.01 | 0.00 to 0.29 |
| 2 | 52 | 0.8 to 10.9 | y = 1.00x - 0.07 | 0.994 | 0.97 to 1.03 | -0.31 to 0.16 |
| 3 | 53 | 0.6 to 10.5 | y = 0.96x + 0.03 | 0.996 | 0.93 to 0.98 | -0.14 to 0.19 |
POL ACCURACY DATA SUMMARY- TP (g/dL)
*95% Confidence Interval
Site # | n | Range (g/dL) | Regression Equation | "r" | CI* Slope | CI Intercept |
---|---|---|---|---|---|---|
1 | 87 | 0.5 to 6.7 | y = 0.99x + 0.24 | 0.982 | 0.92 to 1.06 | -0.06 to 0.53 |
2 | 81 | 0.5 to 6.6 | y = 0.95x + 0.30 | 0.979 | 0.88 to 1.02 | 0.00 to 0.61 |
3 | 81 | 0.9 to 6.1 | y = 0.91x + 0.35 | 0.985 | 0.85 to 0.97 | 0.10 to 0.60 |
POL ACCURACY DATA SUMMARY- ALB (g/dL)
*95% Confidence Interval
807.92 (b)(3): Conclusions from Nonclinical and Clinical Testing
Nonclinical and clinical testing was performed for the Hitachi E40 Clinical Analyzer with Reagent Cartridges TP and ALB. The test systems were shown to be safe and effective for their intended uses.
Page 9 of 9
O Hitachi Chemical Diagnostics, inc.
and the control control of the contrôleant of the many of the many of the comments of the comments of the comments of | ||||
---|---|---|---|---|
www.hcdiagnostics.com | and the maint of the mail of the mail of the mail of | and the state of the state of the will be and the works and the may be | the submit of the country of the world be the | |
. | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | with the state of the problem on one of a comments of a comments of a comments of a comments of a comments of a comments of a comments of a comments of a comments of a commen |
9
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/9/Picture/1 description: The image shows the logo for the Department of Health & Human Services (HHS). The logo features the department's name encircling a stylized symbol. The symbol consists of three ascending lines that curve and resemble a stylized human figure.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 7, 2013
Hitachi Chemical Diagnostics, Inc C/O Erika Ammirati 575 Shirlynn Court LOS ALTOS CA 94022
Re: K131051
Trade/Device Name: Hitachi S TEST Reagent Cartridge Total Protein (TP) Hitachi S TEST Reagent Cartridge Albumin (ALB)
Regulation Number: 21 CFR 862.1035
Regulation Name: Albumin test system
Regulatory Class: II
Product Code: CIX, JGQ Dated: April 12, 2013 Received: April 17, 2013
Dear Ms. Ammirati:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions, against misbranding and adulteration.-- Please note:- GDRH-does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
10
Page 2-Ms. Ammirati
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Carol Benson -S for
Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health . Center for Devices and Radiological Health
Enclosure
11
Indications for Use
510(k) Number (if known):
Device Name: S TEST Reagent Cartridge Total Protein (TP) S TEST Reagent Cartridge Albumin (ALB)
Indications For Use:
The S TEST Reagent Cartridge Total Protein (TP) is intended for the quantitative determination of TP in serum, lithium heparinized plasma, K3 EDTA plasma and sodium citrate plasma using the HITACHI Clinical Analyzer E40. The S TEST Reagent Cartridge TP is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only.
Total protein measurements are used in the diagnosis and treatment of a variety of diseases involving the liver, kidney, or bone marrow as well as other metabolic or nutritional disorders.
The S TEST Reagent Cartridge Albumin (ALB) is intended for the quantitative determination of ALB in serum, lithium heparinized plasma, K3 EDTA plasma and sodium citrate plasma using the HITACHI Clinical Analyzer E40. The S TEST Reagent Cartridge ALB is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only.
Albumin measurements are used in the diagnosis and treatment of numerous diseases involving primarily the liver or kidneys.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
YungW.Chan-S
Division Sign-Off Office of In Vitro Diagnostics and Radiological Health
510(k): K131051
Page 1 of 1