K Number
K130990
Device Name
VARIANT(TM) II TURBO HBA1C KIT - 2.0
Date Cleared
2013-05-09

(29 days)

Product Code
Regulation Number
864.7470
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
The Bio-Rad VARIANT™ II TURBO HbA Ic Kit - 2.0 is intended for the quantitative determination of hemoglobin Alc in human whole blood using ion-exchange high performance liquid chromatography (HPLC) on the VARIANT II TURBO Hemoglobin Testing System. Measurement of hemoglobin Alc is effective in monitoring long term glycemic control in individuals with diabetes mellitus. The Bio-Rad VARIANT II TURBO HbAlc Kit - 2.0 is intended for professional Use.
Device Description
The VARIANT II TURBO Hemoglobin Testing System is the next generation HPLC system with higher volume capability when compared to the VARIANT II testing system. The VARIANT II TURBO Hemoglobin Testing System provides an integrated method for sample preparation, separation, and determination of specific hemoglobin in whole blood. It is a fully automated, high-throughput system. It consists of 2 modules: the VARIANT II TURBO Sampling Station (VSS) and the VARIANT II TURBO Chromatographic Station (VCS). A personal computer (PC) is used to control the VARIANT II TURBO System using Clinical Data Management (CDM™) software. The CDM software supports import of sample information from and export of patient results to a Laboratory Information System (LIS). Control results are displayed on Levy-Jennings Charts and are exportable to Unity Real Time™.
More Information

Not Found

No
The description focuses on the hardware (HPLC system, sampling station, chromatographic station) and software (CDM™) for automated sample processing, separation, and data management. There is no mention of AI or ML algorithms for data analysis, interpretation, or decision-making. The software appears to be for system control, data import/export, and basic quality control charting.

No.
The device is an in-vitro diagnostic system used for the quantitative determination of hemoglobin A1c to monitor long-term glycemic control, not for direct therapeutic intervention.

Yes.

This device quantifies hemoglobin A1c in human whole blood to monitor long-term glycemic control in individuals with diabetes mellitus, which is a diagnostic purpose.

No

The device description clearly states it is a "Hemoglobin Testing System" consisting of hardware modules (Sampling Station and Chromatographic Station) controlled by software. It is not solely software.

Yes, this device is an IVD (In Vitro Diagnostic).

Here's why:

  • Intended Use: The intended use explicitly states it's for the "quantitative determination of hemoglobin Alc in human whole blood." This involves testing a sample taken from the human body in vitro (outside the body).
  • Device Description: The description details a system that performs "sample preparation, separation, and determination of specific hemoglobin in whole blood." This is a typical process for in vitro diagnostic testing.
  • Professional Use: The intended user is "Professional Use Only," which is common for IVD devices used in clinical laboratories.

The core function of the device is to analyze a biological sample (whole blood) to provide diagnostic information (hemoglobin Alc levels for monitoring glycemic control). This aligns perfectly with the definition of an In Vitro Diagnostic device.

N/A

Intended Use / Indications for Use

The Bio-Rad VARIANT™ II TURBO HbA Ic Kit - 2.0 is intended for the quantitative determination of hemoglobin Alc in human whole blood using ion-exchange high performance liquid chromatography (HPLC) on the VARIANT II TURBO Hemoglobin Testing System. Measurement of hemoglobin Alc is effective in monitoring long term glycemic control in individuals with diabetes mellitus. The Bio-Rad VARIANT II TURBO HbAlc Kit - 2.0 is intended for professional Use.

Product codes (comma separated list FDA assigned to the subject device)

LCP

Device Description

The VARIANT II TURBO Hemoglobin Testing System is the next generation Description of HPLC system with higher volume capability when compared to the VARIANT II Instrument testing system. The VARIANT II TURBO Hemoglobin Testing System provides an integrated method for sample preparation, separation, and determination of specific hemoglobin in whole blood. It is a fully automated, high-throughput system. It consists of 2 modules: the VARIANT II TURBO Sampling Station (VSS) and the VARIANT II TURBO Chromatographic Station (VCS).

A personal computer (PC) is used to control the VARIANT II TURBO System using Clinical Data Management (CDM™) software. The CDM software supports import of sample information from and export of patient results to a Laboratory Information System (LIS). Control results are displayed on Levy-Jennings Charts and are exportable to Unity Real Time™.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

Not Found

Indicated Patient Age Range

Not Found

Intended User / Care Setting

professional Use

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

Design verification/validation tests met established acceptance criteria.

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

Not Found

Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.

K122472

Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.

Not Found

Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).

Not Found

§ 864.7470 Glycosylated hemoglobin assay.

(a)
Identification. A glycosylated hemoglobin assay is a device used to measure the glycosylated hemoglobins (A1a , A1b , and A1c ) in a patient's blood by a column chromatographic procedure. Measurement of glycosylated hemoglobin is used to assess the level of control of a patient's diabetes and to determine the proper insulin dosage for a patient. Elevated levels of glycosylated hemoglobin indicate uncontrolled diabetes in a patient.(b)
Classification. Class II (performance standards).

0

Special 510(k) Summary – Device Modification

| Introduction | This 510(k) summary is being submitted in accordance with the requirements of
21 CFR 807.92 and the Safe Medical Device Act of 1990. | | | | |
|----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------|--------------------------------|--|--|
| Submitter | Bio-Rad Laboratories, Inc.
Clinical Systems Division
4000 Alfred Nobel Drive
Hercules, CA 94545 | | | | |
| Contact
Person | Ebony McKinnies
Regulatory Affairs Representative | | | | |
| Date
Submitted | April 5, 2013 | | | | |
| Device Name | VARIANT™ II TURBO HbA Ic Kit - 2.0, Catalog No.: 270-2455 | | | | |
| Classification | Glycosylated hemoglobin assay, 21 CFR 864.7470 [LCP] | | | | |
| Predicate | Table 1: Predicate Device | | | | |
| Device | Device Name | 510(k)
Number | Product Regulation and
Code | | |
| | VARIANT™ II TURBO HbAlc Kit - 2.0 | K122472 | 21 CFR 864.7470 [LCP] | | |
| Intended and
Indications
for Use | The Bio-Rad VARIANT™ II TURBO HbA Ic Kit - 2.0 is intended for the
quantitative determination of hemoglobin Alc in human whole blood using ion-
exchange high performance liquid chromatography (HPLC) on the VARIANT II
TURBO Hemoglobin Testing System. Measurement of hemoglobin Alc is
effective in monitoring long term glycemic control in individuals with diabetes
mellitus. The Bio-Rad VARIANT II TURBO HbAlc Kit - 2.0 is intended for
professional Use. | | | | |
| Description of
Change | The software updates include customer requested features, whereas both software
and firmware include specific defect fixes. When compared to the predicate
device, there are no changes to the performance specifications, intended or
indications for use, or operating principles. Moreover, Risk Analysis and
Verification/Validation testing results demonstrate that the changes do not affect
product safety, effectiveness, and substantial equivalency claims. | | | | |

Bio-Rad Laboratories, Inc. VARIANT™ II TURBO HbAlc Kit – 2.0 Special 510(k) - Device Modification

।-E-S

1

The VARIANT II TURBO Hemoglobin Testing System is the next generation Description of HPLC system with higher volume capability when compared to the VARIANT II Instrument testing system. The VARIANT II TURBO Hemoglobin Testing System provides an integrated method for sample preparation, separation, and determination of specific hemoglobin in whole blood. It is a fully automated, high-throughput system. It consists of 2 modules: the VARIANT II TURBO Sampling Station (VSS) and the VARIANT II TURBO Chromatographic Station (VCS).

A personal computer (PC) is used to control the VARIANT II TURBO System using Clinical Data Management (CDM™) software. The CDM software supports import of sample information from and export of patient results to a Laboratory Information System (LIS). Control results are displayed on Levy-Jennings Charts and are exportable to Unity Real Time™.

| VARIANT II
TURBO Assay | Assay Part
No. | Component Names and Part Nos. | Explanation of Test |
|-------------------------------------------|-------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| VARIANT II
TURBO
HbA1c Kit –
2.0 | 270-2455 | The assay contains the following
components – Whole Blood Primer, 270-0350,
270-0351, 270-0352 Elution Buffer A, 270-2456 Elution Buffer B, 270-2457 Calibrator/Diluent Set, 270-2458 CD-ROM, 270-2461 Analytical Cartridge, 270-2462 Sample Vials, 270-2149 Additional Required/Available
components: Wash/Diluent Solution Set, 270-
2730 Cartridge Holder Installation Kit,
270-2463 Prefilters, 270-2464 Stainless Steel Prefilter Adapters,
270-2465 Microvial Adapters, 270-2016-10 | The VARIANT II
TURBO HbA1c Kit
– 2.0 is a well
established method of
measuring the level
of Hemoglobin A1c
in red blood cells.
Therapy for diabetes
requires the long-
term maintenance of
a blood glucose level
as close as possible to
normal levels to
minimize the risk of
long-term vascular
consequences. |

Table 2: FDA-cleared assays for use on the VARIANT II TURBO Hemoglobin Testing System with CDM Software

Comparison to Predicate Device

The following table shows the similarities and differences between the predicate and modified device.

Bio-Rad Laboratories, Inc. VARIANT™ II TURBO HbAlc Kit - 2.0 Special 510(k) -- Device Modification

l-E-6

2

Table 3: VARIANT II TURBO HbA1c Kit – 2.0

| Feature | Predicate:
Bio-Rad VARIANT™ II TURBO
HbA1c Kit -2.0, 510(k) 122472 | Modified device:
Bio-Rad VARIANT™ II TURBO HbA1c
Kit -2.0 |
|-------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------|
| Similarities | | |
| Technology | Ion-exchange high performance liquid chromatography | |
| Sample type | Anticoagulated whole blood (EDTA) | |
| Calibrator | Human anticoagulated whole blood treated with EDTA | |
| Calibration
frequency | Once every 500 injections/ 2500 injections total column life | |
| Certification | Certified by the NGSP as traceable to the Diabetes Control and Complications
Trial (DCCT) Reference method. | |
| Certification | Certified by the IFCC as traceable to the IFCC Reference Measurement
Procedure. | |
| Instrument Control | Windows Operating System with Proprietary Assay Software | |
| Kit configuration | 2500 Tests: Whole Blood Primer (2 each), Elution Buffer A (5 each),
Elution B (1 each), Calibrator/Diluent Set (1 each), CD-ROM (1 each),
Analytical Cartridge (1 each), Sample Vials – package of 100 (1 each). | |
| Chemistry | Cation Exchange Matrix | |
| Safety Standards for
Electrical Equipment
for IVD Use | BS EN 61010 Certified | |
| Electromagnetic
Compatibility | BS EN 61326 Certified | |
| Reporting units | % HbA1c (NGSP), mmol/mol HbA1c (IFCC), or %HbA1c (JDS) | |
| Intended Use | Intended for the quantitative determination of HbA1c in human whole blood
using ion-exchange HPLC on the VARIANT II TURBO Hemoglobin Testing
System. Measurement of percent HbA1c is effective in monitoring long-term
glucose control in individuals with diabetes mellitus. | |
| Performance Claims | No change, claims transferred from predicate device. | |
| Differences | | |
| CDM Software | CDM Software version 5.1.1 | CDM Software version 5.2 |
| | EPROM | FLASH |
| VARIANT II
TURBO Testing
System Firmware | VCS 41.507
VSS 51.505
VSS PUMP 4.50 | VCS 42.507
VSS 52.523
VSS PUMP 5.00 |
| Historical Database
Review | N/A | Archive Viewer - this tool does not
allow transmission to an LIS, and is not
intended for repeat reporting. |

Risk Management Process for Device Modifications

In accordance with ISO 14971:2012, and internal risk management processes and procedures a defined risk analysis was used to identify, mitigate, or eliminate potential risks associated with the device modifications. For each identified risk, a Failure Mode and Effects Analysis (FMEA) was conducted. This was performed in a systematic manner by a trained risk assessment team until consensus was reached that an adequate analysis had been performed.

Bio-Rad Laboratories, Inc. VARIANT™ II TURBO HbAlc Kit - 2.0 Special 510(k) - Device Modification

1-E-7

3

The risk evaluation for the device software and firmware modifications included the following tasks:

  • Reviewed modifications and design inputs to identify potential risks and . hazards:
  • Reviewed existing product risk tables and customer complaints to 그 identify potential risks and hazards;
  • Considered requirements of IEC 62304:2009, Software Design and . Development processes and plan to identify potential risks and hazards;
  • Identified and implemented risk mitigations and hazard controls through . software, hardware, and labeling for misuse and use scenarios;
  • . Updated existing FMEA and Hazard Analysis tables with newly identified risks, software defects, residual risks, mitigations and hazard controls:
  • . Evaluated modified product using established verification and validation processes, plans and protocols with appropriate acceptance criteria that determined whether risk mitigations, hazard controls, and residual risks were as safe and effective as the predicate device;
  • . Conducted a comprehensive risk management review and wrote a Risk Management Report that summarized all risk activities and deemed the modified product safe, effective, and comparable to the predicate device.

Design verification/validation tests met established acceptance criteria.

Conclusion

When considering the similarities of the intended use, general features and characteristics of the assay, and use of the same technology, it can be concluded that the VARIANT II TURBO HbA1c Kit - 2.0 is substantially equivalent to the cleared and currently marketed predicate device.

Bio-Rad Laboratories, Inc. VARIANT™ II TURBO HbA1c Kit – 2.0 Special 510(k) - Device Modification

!-E-8

4

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular pattern around the symbol. The caduceus is depicted in black, and the text is also in black against a white background. The overall design is simple and recognizable, representing the department's focus on health and human services.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

May 9, 2013

Bio-Rad Laboratories, Inc. C/O Ms. Ebony McKinnies 4000 Alfred Nobel Drive HERCULES CA 94547-1803

Re: K130990

Trade/Device Name: VARIANT™ II TURBO HbA1c Kit - 2.0 Regulation Number: 21 CFR 864.7470 Regulation Name: Glycosylated hemoglobin assay Regulatory Class: II Product Code: LCP Dated: April 05, 2013 Received: April 10, 2013

Dear Ms. McKinnies:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

5

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Carol G. Benson - S for

Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics - and Radiological Health Center for Devices and Radiological Health

Enclosure

6

Indications for Use Form

510(k) Number (if known): K130990

Device Name: VARIANT™ II TURBO HbA1c Kit - 2.0

Indications for Use:

The Bio-Rad VARIANT II TURBO HbA1c Kit – 2.0 is intended for the quantitative determination of hemoglobin A1c in human whole blood using ionexchange high performance liquid chromatography (HPLC) on the VARIANT II TURBO Hemoglobin Testing System. Measurement of hemoglobin A1c is effective in monitoring long-term glycemic control in individuals with diabetes mellitus. The Bio-Rad VARIANT II TURBO HbA1c Kit - 2.0 is intended for Professional Use Only.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiologic Health (OIR)

Ruth A. Chesler S.

Division Sign-Off Office of In Vitro Devices and Radiologic Health

510(k) K130990

Page 1 of 1