K Number
K130804
Date Cleared
2013-05-22

(58 days)

Product Code
Regulation Number
876.5470
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Uromax Ultra Balloon Catheters are recommended for dilatation of the urinary tract.

Device Description

The Balloon Dilatation Catheter, styled after the Gruntzig technique, is a multiple lumen catheter with a dilatation balloon mounted at the distal tip. Dilatation balloon catheters are used to exert radial force to dilate narrow ureteral segments.

AI/ML Overview

The Uromax Ultra™ Balloon Dilatation Catheter did not involve a study of software/AI performance, thus many of the requested categories are not applicable. The device is a physical medical device, not a software or AI product.

Here’s a breakdown of the available information regarding its acceptance criteria and the study that proves it meets those criteria, given the context of a physical device submission:

1. Table of Acceptance Criteria and Reported Device Performance

The submission does not explicitly list acceptance criteria in a table format with corresponding performance results. Instead, it describes "performance testing (bench evaluation)" conducted to evaluate the effects of a design change. The conclusion is that the device demonstrates "equivalence of the Uromax Ultra Balloon Catheter to the predicate balloon dilatation catheter (K980795)."

Below is a table summarizing the tested characteristics:

Acceptance Criteria (Characteristic Evaluated)Reported Device Performance
Effective Working LengthDemonstrated equivalence to predicate device.
Catheter Tip LengthDemonstrated equivalence to predicate device.
Deflated ProfileDemonstrated equivalence to predicate device.
CoatingDemonstrated equivalence to predicate device.
Deflation TimeDemonstrated equivalence to predicate device.
Balloon Diameter at RBPDemonstrated equivalence to predicate device.
Balloon Dilatation Length at RBPDemonstrated equivalence to predicate device.
Radiopaque Markerband to Balloon AlignmentDemonstrated equivalence to predicate device.
Balloon ComplianceDemonstrated equivalence to predicate device.
Multiple InflationDemonstrated equivalence to predicate device.
Balloon BurstDemonstrated equivalence to predicate device.
Proximal Balloon BondDemonstrated equivalence to predicate device.
Surface FinishDemonstrated equivalence to predicate device.
Scope CompatibilityDemonstrated equivalence to predicate device.
Guidewire Compatibility/PassabilityDemonstrated equivalence to predicate device.
Balloon Protector (Wingtool) Removal ForceDemonstrated equivalence to predicate device.
Biocompatibility (supported by K121614)Demonstrated equivalence to predicate device.

2. Sample Size Used for the Test Set and Data Provenance

The document states that "Boston Scientific has conducted performance testing with samples aged at T=0 and T=7 months accelerated aging." However, it does not specify the exact sample size used for these tests. The data provenance is internal to Boston Scientific, as these are bench evaluations of a physical device. No country of origin for the data is explicitly mentioned beyond the company's location (Marlborough, MA, USA), and it is a prospective bench study (evaluating new design changes).

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

This information is not applicable. The device is a physical catheter, and the "ground truth" for its performance is established through objective engineering measurements and tests against predefined specifications and comparisons to a predicate device, not through expert consensus on interpretations.

4. Adjudication Method for the Test Set

This information is not applicable as the "test set" consists of physical device samples undergoing engineering performance evaluations, not human interpretations requiring adjudication.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, an MRMC comparative effectiveness study was not done. This type of study is relevant for diagnostic devices that rely on human interpretation, often with AI assistance. The Uromax Ultra™ Balloon Dilatation Catheter is a therapeutic device, and its performance is evaluated through physical characteristics in bench testing.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

No, this is not applicable. The device is a physical catheter, not an algorithm.

7. The Type of Ground Truth Used

The ground truth used for evaluating the Uromax Ultra™ Balloon Dilatation Catheter is based on objective engineering and performance specifications derived from the predicate device (K980795) and industry standards for balloon dilatation catheters. The goal was to demonstrate "substantial equivalence" regarding intended use, technological characteristics, and performance characteristics to the predicate device. Biocompatibility was supported by data from K121614 for a similar product.

8. The Sample Size for the Training Set

This information is not applicable as this is a physical device submission, not an AI/software submission that would involve a training set.

9. How the Ground Truth for the Training Set Was Established

This information is not applicable for the reasons stated in point 8.

{0}------------------------------------------------

K 130804

Boston Scientific page 1 of 3

SECTION 5

510K SUMMARY

510(k) Summary for Uromax Ultra™ Balloon Dilatation Catheter

MAY 2 2 2013

A. Sponsor

Boston Scientific Corporation Urology and Women's Health Division 100 Boston Scientific Way Marlborough, MA 01756

B. Contact

Lauren Russo Principal Specialist, Regulatory Affairs 508-683-4707 lauren.russo@bsci.com or

Donna Gardner Director, Regulatory Affairs 508-683-4398 gardnerd@bsci.com

C. Device Name

Trade name: Uromax Ultra . High Pressure Balloon Dilatation Catheter Device Type: Dilator, Catheter, Ureteral Classification: 21 CFR 876.5470, 21CFR 876.5520 Product Code: EZN, KOE

D. Predicate Devices

Trade name: Urological Balloon Dilatation Catheter Device Type: Dilator, Catheter, Ureteral Classification: 21 CFR 876.5470, 21CFR 876.5520 Product Code: EZN, KOE

Premarket Notification: Boston Scientific, K980795

Nephromax™ High Pressure Balloon Dilatation Catheter Trade name: Common usual/name: Catheter, Nephrostomy Classification: LJE- Catheter, Nephrostomy Pre-Amendment

Premarket Notification: Boston Scientific, K121614

{1}------------------------------------------------

K130 8.04

Boston Scientific page 2 of 3

SECTION 5510K SUMMARY
-------------------------

E. Device Description

The Balloon Dilatation Catheter, styled after the Gruntzig technique, is a multiple lumen catheter with a dilatation balloon mounted at the distal tip. Dilatation balloon catheters are used to exert radial force to dilate narrow ureteral segments.

F. Intended Use

The Uromax Ultra Balloon Catheters are recommended for dilatation of the urinary tract.

G. Technological Characteristics

The proposed Uromax Ultra Balloon Catheter has the same technological characteristics and fundamental multi-lumen balloon dilatation catheter design as the predicate device (K980795). The proposed Uromax Ultra Balloon Catheter is . available in the same existing sixteen single and kit configurations.

H. Substantial Equivalence

A direct comparison of key characteristics demonstrates that the proposed balloon dilatation catheter is substantially equivalent to the predicate devices in terms of intended use, technological characteristics, and performance characteristics. The Uromax Ultra Balloon Catheter is as safe, as effective, and performs as well as the predicate devices.

I. Performance Testing (Bench Evaluation)

Boston Scientific has conducted performance testing with samples aged at T=0 and T=7 months accelerated aging in support of the balloon design change. The following testing was completed to evaluate the effects of the design change:

  • Effective Working Length �
  • Catheter Tip Length �
  • Deflated Profile .
  • Coating .
  • . Deflation Time
  • Balloon Diameter at RBP .
  • Balloon Dilatation Length at RBP .
  • Radiopaque Markerband to Balloon Alignment �
  • . Balloon Compliance
  • Multiple Inflation .
  • Balloon Burst ♥
  • Proximal Balloon Bond .
  • . Surface Finish
  • Scope Compatibility �
  • 9 Guidewire Compatibility/Passability
  • Balloon Protector (Wingtool) Removal Force ●

Page 26 of 219

{2}------------------------------------------------

K130804

page 3 of 3

Boston Scientific

SECTION 5

510K SUMMARY

ي والي والي ويوري

Page 27 of 219

Performance data for biocompatibility provided in K121614 for the Nephromax High Pressure Balloon Dilatation Catheter supports the Uromax Ultra Balloon Dilatation Catheter. The results of the performance testing demonstrate equivalence of the Uromax Ultra Balloon Catheter to the predicate balloon dilatation catheter (K980795).

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three tail feathers. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

May 22, 2013

· Boston Scientific Corporation Urology/Women's Health % Ms. Lauren B. Russo Principal Regulatory Affairs Specialist 100 Boston Scientific Way MARLBOROUGH MA 01752

Re: K130804

Trade/Device Name: Uromax Ultra™ Balloon Dilatation Catheter Regulation Number: 21 CFR& 876.5470 Regulation Name: Ureteral dilator Regulatory Class: II Product Code: EZN, KOE Dated: March 22, 2013 Received: March 25, 2013

Dear Ms. Russo:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Glenn B. Bell -S

Acting Director for: Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Boston Scientific

SECTION 4 INDICATIONS FOR USE

Indications for Use Statement

510(k) Number

Uromax Ultra™ Balloon Dilatation Catheter

Indications For Use

Device Name

Uromax Ultra™ Balloon Dilatation Catheters are recommended for dilatation of the urinary tract.

Prescription Use X (21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

To be determined.

Glenn B. Bell -S

for BRF

(Division Sign-Off) Division of Reproductive, Gastro-Renal, and Urological Devices 510(k) Number K130804

Proprietary and Confidential Information of Boston Scientific Corporation

Page 23 of 219

§ 876.5470 Ureteral dilator.

(a)
Identification. A ureteral dilator is a device that consists of a specially shaped catheter or bougie and is used to dilate the ureter at the place where a stone has become lodged or to dilate a ureteral stricture.(b)
Classification. Class II (performance standards).