K Number
K130714
Device Name
PROGIGY SUPPORT CATHETER
Date Cleared
2013-08-22

(160 days)

Product Code
Regulation Number
870.1250
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
The Prodigy Support Catheter is intended to be used in conjunction with steerable guidewires in order to access discrete regions of the arterial and or coronary vasculature. It may be used to facilitate placement and exchange of guidewires and other interventional devices.
Device Description
The Prodigy Support Catheter is an over-the-wire (OTW) 2-lumen catheter with a 5 mm clastomeric balloon located near the distal tip. One lumen of the catheter is used to inflate and deflate the balloon with contrast fluid. The second lumen provides access for a guidewire. Two side legs are located at the proximal catheter end to provide access to the balloon lumen and the guidewire lumcu. A pressure relief valve is attached at the proximal end of the balloon side leg and allows for connection with an inflation device. There are two radiopaque marker bands located at each end of the balloon. The inflated balloon diameter is variable with increased pressure up to 5 mm. The Prodigy Support Catheter is packaged in a mylar/Tyvek pouch and ETO sterilized to SAL 10-6.
More Information

K082337, P860019/183, P860019/241

Not Found

No
The device description and performance studies focus on the physical characteristics and in vitro performance of a support catheter, with no mention of AI or ML capabilities.

No
The device is described as a support catheter intended to facilitate the placement and exchange of guidewires and other interventional devices, not to directly treat a condition. While it has a balloon, its primary use is for support and access, not therapeutic dilatation or other direct treatment.

No

The device is a support catheter used to facilitate the placement and exchange of guidewires and other interventional devices, rather than to diagnose medical conditions.

No

The device description clearly details a physical catheter with lumens, a balloon, side legs, a pressure relief valve, and radiopaque marker bands. This indicates a hardware-based medical device, not a software-only one.

No, this device is not an IVD (In Vitro Diagnostic).

Here's why:

  • IVD Definition: In Vitro Diagnostics are medical devices used to perform tests on samples taken from the human body (like blood, urine, tissue) to provide information about a person's health.
  • Device Function: The Prodigy Support Catheter is a physical device used within the body (in the arterial and coronary vasculature) to facilitate the placement and exchange of other devices like guidewires. It does not perform any tests on bodily samples.
  • Intended Use: The intended use clearly describes a procedural function within the circulatory system, not a diagnostic test on a sample.

The description of the device and its intended use aligns with an interventional medical device used for accessing and navigating within blood vessels.

N/A

Intended Use / Indications for Use

The Prodigy Support Catheter is intended to be used in conjunction with steerable guidewires in order to access discrete regions of the arterial and or coronary vasculature. It may be used to facilitate placement and exchange of guidewires and other interventional devices.

Product codes (comma separated list FDA assigned to the subject device)

DQY

Device Description

The Prodigy Support Catheter is an over-the-wire (OTW) 2-lumen catheter with a 5 mm clastomeric balloon located near the distal tip. One lumen of the catheter is used to inflate and deflate the balloon with contrast fluid. The second lumen provides access for a guidewire. Two side legs are located at the proximal catheter end to provide access to the balloon lumen and the guidewire lumcu. A pressure relief valve is attached at the proximal end of the balloon side leg and allows for connection with an inflation device. There are two radiopaque marker bands located at each end of the balloon. The inflated balloon diameter is variable with increased pressure up to 5 mm.

The Prodigy Support Catheter is packaged in a mylar/Tyvek pouch and ETO sterilized to SAL 10-6.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

arterial and or coronary vasculature

Indicated Patient Age Range

Not Found

Intended User / Care Setting

Not Found

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

Biocompatibility tests were conducted on the Prodigy Support Catheter according to the requirements of Guidance for Industry and FDA Staff-Class II Special Controls for Certain Percutaneous Transluminal Coronary Angioplasty (PTCA) Catheters, September 8, 20102010 and ISO10993-1:2009, Biological Evaluation of Medial Devices Part-1: Evaluation and Testing. The following test were conducted:

  • Cytotoxicity
  • Sensitization
  • Intracutaneous Reactivity
  • Acute Systemic Toxicity
  • Hemolysis
  • Material Mediated Pyrogenicity
  • Thombogenicity
  • Complement Activation

In vitro performance tests were performed on the Prodigy Support Catheter according to the requirements of Guidance for Industry and FDA Staff-Class II Special Controls for Certain Percutaneous Transluminal Coronary Angioplasty (PTCA) Catheters, September 8, 2010. In-vitro performance testing conducted on the Prodigy Support Catheter included:

  • Dimensional Verification
  • Balloon Preparation, Deployment and Retraction
  • Balloon Burst Pressure
  • Balloon Fatigue
  • Balloon Inflation/Deflation Time
  • Catheter Bond Strength
  • Flexibility and Kink Test
  • Torque Strength
  • Radiopacity
  • Balloon Holding Force

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

Not Found

Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.

K082337, P860019/183, P860019/241

Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.

Not Found

Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).

Not Found

§ 870.1250 Percutaneous catheter.

(a)
Identification. A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.(b)
Classification. Class II (performance standards).

0

K130714

510(K) SUMMARY

The 510(k) Summary is submitted in accordance with 21 CFR 807.92 and the requirements of the Safe Medical Device ACT (SMDA) of 1990.

    1. Sponsor Name Submitter's Name: Radius Medical, LLC Address: 577 Main St, Suitc 360 Hudson, MA 01749 Phone: (978) 263-4466 Fax: (978) 263-4465 Contact Person: Richard DeMello Date of Preparation: March 7, 2013
      AUG 2 2 2013
    1. Device Information
Trade Name:Prodigy Support Catheter
Common Name:Percutaneous catheter
Class:II
Classification Name:Percutaneous catheter (21 CFR 870.1250, Product Code DQY)
    1. Predicate Devices
      Minnie™ Support Catheter (K082337) manufactured by Vascular Solutions, Inc.

Maverick XL PTCA Catheter (P860019/183) manufactured by Boston Scicntific.

NC Quantum Apex OTW PTCA Dilatation Catheter (P860019/241) manufactured by Boston Scientific.

4. Device Description

The Prodigy Support Catheter is an over-the-wire (OTW) 2-lumen catheter with a 5 mm clastomeric balloon located near the distal tip. One lumen of the catheter is used to inflate and deflate the balloon with contrast fluid. The second lumen provides access for a guidewire. Two side legs are located at the proximal catheter

1

end to provide access to the balloon lumen and the guidewire lumcu. A pressure relief valve is attached at the proximal end of the balloon side leg and allows for connection with an inflation device. There are two radiopaque marker bands located at each end of the balloon. The inflated balloon diameter is variable with increased pressure up to 5 mm.

The Prodigy Support Catheter is packaged in a mylar/Tyvek pouch and ETO sterilized to SAL 10-6.

  • న. Intended Use
    The Prodigy Support Catheter is intended to be used in conjunction with steerable guidewires in order to access discrete regions of the arterial and or coronary vasculature. It may be used to facilitate placement and exchange of guidewires and other interventional devices.

    1. Comparison of Technological Characteristics
      The Prodigy Support Catheter is similar in matcrials, function and design to predicate NC Quantum Apex OTW PTCA Dilatation Catheter (P860019/241) and the Maverick XL PTCA Catheter (P860019/183). All devices have two-lumen polymer shafts with a polymer balloon located near the tip of the catheter. One lumcn of the shaft allows for inflation of the balloon with contrast medium while the other lumen permits the guidewire to facilitate advancement of the catheter. The balloon sizes on the predicate devices range from 1.5 mm to 6 mm while the balloon on the Prodigy Support Catheter expands up to 6 mm. The Prodigy and NC Quantum Apex catheters both include side leg adaptors to provide separate access to the balloon inflation lumcn and the guidewire lumen. All catheters include two radiopaque marker bands at cither end of the balloon to aid in positioning the balloon within the patient. The NC Quantum Apcx OTW and Maverick XL catheters are used for dilating a stenotic portion of a coronary artery. The Prodigy Support Catheter is for accessing discrete regions of the arterial and or coronary vasculature and to facilitate placement and exchange of guidewires and other interventional devices. All are introduced into the body through a guidecatheter or outer sheath and are directed through the artery over a guidcwire. All devices have a balloon that is inflated at the distal end of the catheter into contact with the vessel wall. The NC Quantum Apex and Maverick balloon is used to dilate a narrowing in an artery using high pressures. The Prodigy balloon is used to anchor the catheter within the artery using low pressures. Guidewires can be placed and exchanged through the Prodigv and NC Quantum Apex OTW catheters.

2

The design and function of the Prodigy Support Catheter is similar to the Minnie™ Support Catheter (K082337). Both devices utilize a polymer shaft, are introduced into the body through a guidccatheter or outer sheath and are directed . through the artery over a guidewire. Both catheters provide support for guidewires during interventional procedures and allow for exchanging of guidewires and other interventional devices. The Minnie catheter has a single lumen that accepts a guidewire. The Prodigy catheter has two lumens, one for a guidewire and one to inflate the distal balloon. Both catheters have radiopaque markers at the distal end to aid in location of the catheter tip.

    1. Performance Data

Biocompatibility Testing

Biocompatibility tests were conducted on the Prodigy Support Catheter according to the requirements of Guidance for Industry and FDA Staff-Class II Special Controls for Certain Percutaneous Transluminal Coronary Angioplasty (PTCA) Catheters, September 8, 20102010 and ISO10993-1:2009, Biological Evaluation of Medial Devices Part-1: Evaluation and Testing. The following test were conducted:

  • . Cytotoxicity
  • . Sensitization
  • . Intracutaneous Reactivity
  • . Acute Systemic Toxicity
  • . Hemolysis
  • . Material Mediated Pyrogenicity
  • . Thombogenicity
  • . Complement Activation

In Vitro Performance Testing

In vitro performance tests were performed on the Prodigy Support Catheter according to the requirements of Guidance for Industry and FDA Staff-Class II Special Controls for Certain Percutaneous Transluminal Coronary Angioplasty (PTCA) Catheters, September 8, 2010. In-vitro performance testing conducted on the Prodigy Support Catheter included:

  • . Dimensional Verification
  • . Balloon Preparation, Deployment and Retraction
  • . Balloon Burst Pressure
  • . Balloon Fatigue

3

  • Balloon Inflation/Deflation Time
  • . Catheter Bond Strength
  • . Flexibility and Kink Test
  • Torque Strength
  • . Radiopacity
  • . Balloon Holding Force

Conclusion

Based upon these biocompatibility and in vitro performance tests, the Prodigy Support Catheter has been shown to be substantially equivalent to the currently marketed Minnie Support Catheter, Maverick XL PTCA Catheter and NC Quantum Apex OTW PTCA Dilatation Catheter.

4

Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is a stylized symbol that resembles an abstract caduceus or a bird-like figure with three wavy lines extending from its body.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

August 22, 2013

Radius Medical, LLC c/o Mr. Richard DeMello Vice President 577 Main St, Suite 360 Hudson, MA 01749

Re: K130714

Trade/Device Name: Prodigy Support Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous catheter Regulatory Class: Class II Product Code: DOY Dated: July 12, 2013 Received: July 18, 2013

Dear Mr. DeMello:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading,

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act

5

Page 2 - Mr. Richard DeMello

or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note

the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

M.A. Hillerman

for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

6

510(k) Number (if known):

Device Name: Prodigy Support Catheter

Indications For Use:

The Prodigy Support Catheter is intended to be used in conjunction with steerable guidewires in order to access discrete regions of the arterial and or coronary vasculature. It may be used to facilitate placement and exchange of guidewires and other interventional devices.

X Prescription Use (Per 21 CFR 801.109)

OR

Over-The-Counter Use

(PLEASE DO NOT WRITE BELOW THIS LINE. CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

M.A. Zillerman