(30 days)
The CD HORIZON® Spinal System with or without SEXTANT® instrumentation is intended for posterior, non-cervical fixation as an adjunct to fusion for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis and/or lordosis); tumor; pseudarthrosis; and/or failed previous fusion.
Except for hooks, when used as an anterolateral thoracic/lumbar system, the CD HORIZON® Spinal System may also be used for the same indications as an adjunct to fusion.
With the exception of degenerative disc disease, the CD HORIZON® LEGACY™ 3.5mm rods and the CD HORIZON® Spinal System PEEK rods and associated components may be used for the aforementioned indications in skeletally mature patients as an adjunct to fusion. The 3.5mm rods may be used for the specific pediatric indications noted below.
When used for posterior non-cervical pedicle screw fixation in pediatric patients, the CD HORIZON® Spinal System implants are indicated as an adjunct to fusion to treat adolescent idiopathic scoliosis. Additionally, the CD HORIZON® Spinal System is intended to treat pediatric patients diagnosed with the following conditions: spondylolisthesis/spondylolysis and fracture caused by turnor and/or trauma. These devices are to be used with autograft and/or allograft. Pediatric pedicle screw fixation is limited to a posterior approach.
The CD HORIZON SPIRE™ Plate is a posterior, single level, non-pedicle supplemental fixation device intended for use in the non-cervical spine (TI-SI) as an adjunct to fusion in skeletally mature patients. It is intended for plate fixation/attachment to spinous processes for the purpose of achieving supplemental fixation in the following conditions: degenerative disc disease (as previously defined); spondylolisthesis, trauma; and/or tumor.
In order to achieve additional levels of fixation, the CD HORIZON® Spinal System rods may be connected to the VERTEX® Reconstruction System with the VERTEX® rod connector. Refer to the VERTEX® Reconstruction System Package Insert for a list of the VERTEX® indications of use.
The CD HORIZON® Spinal System consists of a variety of shapes and sizes of rods, hooks, screws, CROSSLINK® Plates, staples and connecting components, as well as implant components from other Medtronic spinal systems, which can be rigidly locked into a variety of configurations, with each construct being tailor-made for the individual case.
A subset of CD HORIZON® Spinal System components may be used for posterior pedicle screw fixation in pediatric cases. These constructs may be comprised of a variety of shapes and sizes of rods (ranging in diameter from 3.5mm to 6.35mm), hooks, screws, CROSSLINK® Plates, and connecting components. Similarly to the CD HORIZON® implants used in adult cases, these components can be rigidly locked into a variety of configurations, with each construct being tailor-made for the individual case.
Certain components within the CD HORIZON® Spinal System are specifically excluded for use in pediatric patients. These include PEEK rods, Shape Memory Alloy Staples, SPIRE™ Plates and DYNALOK® bolts. All screws used in pediatric cases are only cleared for use via a posterior approach. All of the components used in pediatric cases are fabricated from medical grade stainless steel, medical grade titanium, titanium alloy, and medical grade cobalt-chromium-molybdenum alloy.
Certain implant components from other Medtronic spinal systems can be used with the CD HORIZON® Spinal System in non-pediatric cases. These components include TSRH® rods, hooks, screws, plates, CROSSLINK® plates, connectors, staples, washers, GDLH® rods, hooks, connectors and CROSSLINK® bar and connectors; LIBERTY® rods and screws; DYNALOK® PLUS and DYNALOK CLASSIC® bolts along with rod/bolt connectors; and Medtronic Multi-Axial rods and screws. Please note that certain components are specifically designed to connect to specific rod diameters, while other components can connect to multiple rod diameters. Care should be taken so that the correct components are used in the spinal construct.
CD HORIZON® hooks are intended for posterior use only. CD HORIZON® staples and CD HORIZON® ECLIPSE® rods and associated screws are intended for anterior use only. However, for patients of smaller stature and pediatric patients, CD HORIZON® 4.5mm rods and associated components may be used posteriorly.
The CD HORIZON® Spinal System implant components are fabricated from medical grade stainless steel, medical grade titanium, titanium alloy, medical grade cobaltchromium-molybdenum alloy, or medical grade PEEK OPTIMA-LT1. Certain CD HORIZON® Spinal System components may be coated with hydroxyapatite. No warranties expressed or implied are made. Implied warranties of merchantability and fitness for a particular purpose or use are specifically excluded.
Never use stainless steel and titanium implant components in the same construct. Medical grade titanium, titanium alloy, and/or medical grade cobalt-chromiummolybdenum alloy may be used together. Never use titanium, titanium alloy, and/or medical grade cobalt-chromium-molybdenum alloy with stainless steel in the same construct.
The CD HORIZON® Spinal System also includes anterior staples made of Shape Memory Alloy (Nitinol - NiTi). Shape Memory Alloy is compatible with titanium, titanium alloy, and cobalt-chromium-molybdenum alloy. Do not use with stainless steel. These staples are not to be used in pediatric patients.
PEEK implants may be used with stainless steel, titanium, or cobalt-chromiummolybdenum alloy implants. CD HORIZON® PEEK Rods are not to be used with CROSSLINK® Plates or in pediatric patients.
The purpose of this 510(k) submission includes the addition of uni-axial screws into the CD HORIZON® Spinal System. The subject uni-axial screws are a four-piece subcomponent assembly like the predicate multi-axial screws consisting of a bone screw, ring, crown, and head; however, instead of the screw assembly rotating in a multi-axial rotation, the screw pivots in a uni-axial direction.
Here's an analysis of the provided text regarding the CD HORIZON® Spinal System, focusing on acceptance criteria and supporting studies, formatted as requested:
Acceptance Criteria and Device Performance
| Acceptance Criteria (Implied by Equivalence) | Reported Device Performance |
|---|---|
| Substantial equivalence to predicate device (CD HORIZON® SOLERA™ 5.5/6.0 Multi-axial Screws (K113174) and CD HORIZON® LEGACY 4.5 Multi-axial Screws (K042025)) in: | The subject CD HORIZON® SOLERA™ 5.5/6.0 Uni-axial Screws are a modification of the predicate device, differing only in the pivot direction (uni-axial vs. multi-axial). |
| * Indications for Use | Identical to predicate device. |
| * Intended Use | Identical to predicate device. |
| * Performance Specifications | "The previously submitted testing for the CD HORIZON® LEGACY 4.5 multi-axial screw still applies." This implies the new device meets the same performance specifications (e.g., mechanical strength, fatigue resistance) through equivalence, without requiring new tests. |
| * Technological Characteristics | Very similar, with the key difference described above. Composed of the same materials (commercially pure titanium, titanium alloy, and cobalt-chromium-molybdenum). Available in solid and cannulated versions. |
| Risk Analysis Completion | A risk analysis was completed. |
| Appropriate Design Control Processes | An "assessment of the device modifications was completed in accordance with Medtronic design control processes." |
Study Information
-
Sample Size used for the test set and the data provenance:
- Test Set Sample Size: Not applicable. No clinical testing was performed for this 510(k) submission. The non-clinical testing refers to the previously submitted data for predicate devices.
- Data Provenance: Not applicable. The submission relies on non-clinical (mechanical/material) testing data from previously cleared predicate devices (K113174 and K042025). The origin of that original testing data is not specified in this summary, but would typically be from the manufacturer's internal testing facilities.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. No human expert-based ground truth was established for this submission, as it relied on non-clinical equivalence to predicate devices.
-
Adjudication method for the test set: Not applicable. No test set requiring adjudication was used.
-
If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is a spinal implant device, not an AI or diagnostic imaging device.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This is a spinal implant device, not an algorithm.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Not applicable to this 510(k) submission. The "ground truth" for demonstrating safety and effectiveness was based on the performance of previously cleared, substantially equivalent predicate devices (mechanical performance, material properties).
-
The sample size for the training set: Not applicable. This is a spinal implant device, not a machine learning model requiring a training set.
-
How the ground truth for the training set was established: Not applicable.
{0}------------------------------------------------
K130646 Page 1 of 5
CD HORIZON® Spinal System 510(k) Summary
APR 1 0 2013
. ----
March 08, 2013
Medtronic Sofamor Danek USA, Inc. l, Company: 1800 Pyramid Place Memphis, Tennessee 38132 Telephone: (901) 396-3133 (901) 346-9738 Fax:
Contact: Mr. Brad Sheals, MS Senior Regulatory Affairs Specialist
- CD HORIZON® Spinal System II. Trade Name:
- III. Orthosis, Spinal pedicle fixation, for degenerative disc disease Common Name:
- IV. Classification Name: Spinal Interlaminal Fixation Orthosis Spinal Intervertebral Body Fixation Orthosis Pedicle Screw Spinal System
- V. III (Preamendment) Class:
- VI. Product Code: KWP, KWQ, MNH, MNI, NKB, OSH
VII. Regulation Numbers: 888.3050, 888.3060 and 888.3070
VIII. Product Description:
The CD HORIZON® Spinal System consists of a variety of shapes and sizes of rods, hooks, screws, CROSSLINK® Plates, staples and connecting components, as well as implant components from other Medtronic spinal systems, which can be rigidly locked into a variety of configurations, with each construct being tailor-made for the individual case.
A subset of CD HORIZON® Spinal System components may be used for posterior pedicle screw fixation in pediatric cases. These constructs may be comprised of a variety of shapes and sizes of rods (ranging in diameter from 3.5mm to 6.35mm), hooks, screws, CROSSLINK® Plates, and connecting components. Similarly to the CD HORIZON®
{1}------------------------------------------------
implants used in adult cases, these components can be rigidly locked into a variety of configurations, with each construct being tailor-made for the individual case.
Certain components within the CD HORIZON® Spinal System are specifically excluded for use in pediatric patients. These include PEEK rods, Shape Memory Alloy Staples, SPIRE™ Plates and DYNALOK® bolts. All screws used in pediatric cases are only cleared for use via a posterior approach. All of the components used in pediatric cases are fabricated from medical grade stainless steel, medical grade titanium, titanium alloy, and medical grade cobalt-chromium-molybdenum alloy.
Certain implant components from other Medtronic spinal systems can be used with the CD HORIZON® Spinal System in non-pediatric cases. These components include TSRH® rods, hooks, screws, plates, CROSSLINK® plates, connectors, staples, washers, GDLH® rods, hooks, connectors and CROSSLINK® bar and connectors; LIBERTY® rods and screws; DYNALOK® PLUS and DYNALOK CLASSIC® bolts along with rod/bolt connectors; and Medtronic Multi-Axial rods and screws. Please note that certain components are specifically designed to connect to specific rod diameters, while other components can connect to multiple rod diameters. Care should be taken so that the correct components are used in the spinal construct.
CD HORIZON® hooks are intended for posterior use only. CD HORIZON® staples and CD HORIZON® ECLIPSE® rods and associated screws are intended for anterior use only. However, for patients of smaller stature and pediatric patients, CD HORIZON® 4.5mm rods and associated components may be used posteriorly.
The CD HORIZON® Spinal System implant components are fabricated from medical grade stainless steel, medical grade titanium, titanium alloy, medical grade cobaltchromium-molybdenum alloy, or medical grade PEEK OPTIMA-LT1. Certain CD HORIZON® Spinal System components may be coated with hydroxyapatite. No warranties expressed or implied are made. Implied warranties of merchantability and fitness for a particular purpose or use are specifically excluded.
Never use stainless steel and titanium implant components in the same construct.
{2}------------------------------------------------
Medical grade titanium, titanium alloy, and/or medical grade cobalt-chromiummolybdenum alloy may be used together. Never use titanium, titanium alloy, and/or medical grade cobalt-chromium-molybdenum alloy with stainless steel in the same construct.
The CD HORIZON® Spinal System also includes anterior staples made of Shape Memory Alloy (Nitinol - NiTi). Shape Memory Alloy is compatible with titanium, titanium alloy, and cobalt-chromium-molybdenum alloy. Do not use with stainless steel. These staples are not to be used in pediatric patients.
PEEK implants may be used with stainless steel, titanium, or cobalt-chromiummolybdenum alloy implants. CD HORIZON® PEEK Rods are not to be used with CROSSLINK® Plates or in pediatric patients.
The purpose of this 510(k) submission includes the addition of uni-axial screws into the CD HORIZON® Spinal System. The subject uni-axial screws are a four-piece subcomponent assembly like the predicate multi-axial screws consisting of a bone screw, ring, crown, and head; however, instead of the screw assembly rotating in a multi-axial rotation, the screw pivots in a uni-axial direction.
IX. Indications:
The CD HORIZON® Spinal System with or without SEXTANT® instrumentation is intended for posterior, non-cervical fixation as an adjunct to fusion for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis and/or lordosis); tumor; pseudarthrosis; and/or failed previous fusion.
Except for hooks, when used as an anterolateral thoracic/lumbar system, the CD HORIZON® Spinal System may also be used for the same indications as an adjunct to fusion.
With the exception of degenerative disc disease, the CD HORIZON® LEGACY™ 3.5mm rods and the CD HORIZON® Spinal System PEEK rods and associated
{3}------------------------------------------------
components may be used for the aforementioned indications in skeletally mature patients as an adjunct to fusion. The 3.5mm rods may be used for the specific pediatric indications noted below.
When used for posterior non-cervical pedicle screw fixation in pediatric patients, the CD HORIZON® Spinal System implants are indicated as an adjunct to fusion to treat adolescent idiopathic scoliosis. Additionally, the CD HORIZON® Spinal System is intended to treat pediatric patients diagnosed with the following conditions: spondylolisthesis/spondylolysis and fracture caused by turnor and/or trauma. These devices are to be used with autograft and/or allograft. Pediatric pedicle screw fixation is limited to a posterior approach.
The CD HORIZON SPIRE™ Plate is a posterior, single level, non-pedicle supplemental fixation device intended for use in the non-cervical spine (TI-SI) as an adjunct to fusion in skeletally mature patients. It is intended for plate fixation/attachment to spinous processes for the purpose of achieving supplemental fixation in the following conditions: degenerative disc disease (as previously defined); spondylolisthesis, trauma; and/or tumor.
In order to achieve additional levels of fixation, the CD HORIZON® Spinal System rods may be connected to the VERTEX® Reconstruction System with the VERTEX® rod connector. Refer to the VERTEX® Reconstruction System Package Insert for a list of the VERTEX® indications of use.
Identification of Legally Marketed Predicate Devices Used to Claim Substantial X. Equivalence:
The design features, product sizes, and materials of the subject CD HORIZON® Spinal System are the same as the predicate CD HORIZON® Spinal System components previously cleared in K113174 (S.E. 11/21/2011). An additional predicate such as the CD HORIZON® LEGACY 4.5 Multi-axial screws (K042025, S.E. 08/25/2004) was also utilized to support the substantial equivalence of the subject device. The safety and effectiveness of the subject pedicle screws are adequately supported by documentation within this premarket notification.
{4}------------------------------------------------
XI. Summary of the Technological Characteristics:
The subject CD HORIZON® SOLERA™ 5.5/6.0 Uni-axial Screws are a modification of the predicate CD HORIZON® SOLERA™ 5.5/6.0 Multi-axial Screws (KI 13174, S.E. 11/21/2011). The subject uni-axial screws are a four-piece subcomponent assembly like the predicate multi-axial screws consisting of a bone screw, ring, crown, and head; however, instead of the screw assembly rotating in a multi-axial rotation, the screw pivots in a uni-axial direction. The subject device will be available non-sterile in solid and cannulated versions. The subject and predicate pedicle screws are identical in terms of indications for use, intended use, performance specifications and technological characteristics. Like the predicate device, the subject CD HORIZON® SOLERATM 5.5/6.0 Uni-axial Screws are composed of commercially pure titanium, titanium alloy and cobalt-chromium-molybdenum.
XII. Discussion of Non-Clinical Testing:
An assessment of the device modifications was completed in accordance with Medtronic design control processes.
The previously submitted testing for the CD HORIZON® LEGACY 4.5 multi-axial screw still applies.
Discussion of Clinical Testing: XIII.
No clinical testing was performed.
Conclusions Drawn from the Non-Clinical Tests: XIV.
A risk analysis was completed. Based on the risk analysis and additional supporting documentation provided in this premarket notification, the subject device demonstrates substantial equivalence to the predicate CD HORIZON® SOLERA™ 5.5/6.0 Multi-axial Screws (K113174, S.E. 11/21/2011) and predicate CD HORIZON® LEGACY 4.5 Multiaxial Screws (K042025, S.E. 08/25/2004).
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the caduceus symbol. The logo is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 27, 2013
Medtronic Sofamor Danek USA, Incorporated % Brad Sheals, MS Senior Regulatory Affairs Specialist 1800 Pyramid Place Memphis, Tennessee 38132
Re: K130646
Trade/Device Name: CD HORIZON® Spinal System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: Class III Product Code: NKB, OSH, MNH, MNI, KWP, KWO Dated: March 8, 2013 Received: March 11, 2013
Dear Mr. Sheals:
This letter corrects our substantially equivalent letter of April 10, 2013.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements. including, but not limited to: registration and listing (21
{6}------------------------------------------------
Page 2 - Brad Sheals. MS
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
Erin Foreith
For
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
510(k) Number (if known): · K130646
Device Name: CD HORIZON® Spinal System
Indications for Use:
The CD HORIZON® Spinal System with or without SEXTANT® instrumentation is intended for posterior, non-cervical fixation as an adjunct to fusion for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis and/or lordosis); tumor; pseudarthrosis; and/or failed previous fusion.
Except for hooks, when used as an anterolateral thoracic/lumbar system, the CD HORIZON® Spinal System may also be used for the same indications as an adjunct to fusion.
With the exception of degenerative disc disease, the CD HORIZON® LEGACY™ 3.5mm rods and the CD HORIZON® Spinal System PEEK rods and associated components may be used for the aforementioned indications in skeletally mature patients as an adjunct to fusion. The 3.5mm rods may be used for the specific pediatric indications noted below.
When used for posterior non-cervical pedicle screw fixation in pediatric patients, the CD HORIZON® Spinal System implants are indicated as an adjunct to treat adolescent idiopathic scoliosis. Additionally, the CD HORIZON® Spinal System is intended to treat pediatic patients diagnosed with the following conditions: spondylolisthesis/spondylolysis, and fracture caused by tumor and/or trauma. These devices are to be used with autograft and/or allograft. Pedicle screw fixation is limited to a posterior approach.
· The CD HORIZON SPIRE™ Plate is a posterior, single level, non-pedicle supplemental fixation device intended for use in the non-cervical spine (T1-S1) as an adjunct to fusion in skeletally mature patients. It is intended for plate fixation/attachment to spinous processes for the purpose of achieving supplemental fixation in the following conditions: degenerative disc disease (as previously defined); spondylolisthesis, trauma; and/or tumor.
In order to achieve additional levels of fixation, the CD HORIZON® Spinal System rods may be connected to the VERTEX® Reconstruction System with the VERTEX® rod connector. Refer to the VERTEX® Reconstruction System Package Insert for a list of the VERTEX® indications of use.
Prescription Use X AND/OR (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH. Office of Device Evaluation (ODE) Ronald P. Jean -S
(Division Sign-Off) Division of Orthopedic Devices 510(k) Number: K130646
§ 888.3070 Thoracolumbosacral pedicle screw system.
(a)
Identification. (1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.
(b)
Classification. (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
“Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.