AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The CD HORIZON® Spinal System is intended for posterior, non-cervical fixation for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kynhonis and/or lordosis); tumor; pseudarthrosis; and/or failed previous fusion.

When used in a percutaneous, non-cervical, posterior approach with the SEXTANT instrumentation, the CD HORIZON® screws are intended for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis and/or lordosis); tumor; pseudoarthrosis; and/or failed previous fusion.

Except for hooks, when used as an anterolateral thoracic/lumbar system, CD HORIZON® components such as ECLIPSE® components are intended for the following indications: (1) degenerative disc disease (as defined by back pain of discogenic origin with degencration of the disc confirmed by patient history and radiographic studies), (2) spinal stenosis, (3) spondylolisthesis, (4) spinal deformities (i.e., scoliosis, kyphosis, and/or lordosis), (5) fracture, (6) pseudarthrosis, (7) tumor resection, and/or (8) failed previous fusion.

The CD HORIZON® SPIRE Plate is posterior, non-pedicle supplemental fixation device, intended for use in the non-cervical spine (T1 - S1). It is intended for plate fixation/attachment to spinous process for the purpose of achieving supplemental fusion in the following cuaditions: degenerative disc disease - defined as back pain of disougane origin with regeneration of the disc confirmed by history and radiographic studies; spondylolisthesis; trauma (i.e., fracture ro dislocation); and/or tumor.

In order to achieve additional levels of fixation, the CD HORIZON® Spinal System rods may be connected to the VERTEX™ Reconstruction System with the VERTEXTM rod connector. Refer to the VERTEX™ Reconstruction System Package Insert for a list of the VERTEXTM indications of use.

Device Description

The CD HORIZON® Spinal System consists of a variety of rods, hooks, screws, CROSSLINK® plates, staples, and other connecting components used to build a spinal construct. Instrumentation is also available to facilitate implantation of the device components.

The CD HORIZON® Spinal System is intended to help provide immobilization and stabilization of spinal segments as an adjunct to fusion of the thoracic, lumbar, and/or sacral spine. The CD HORIZON® Spinal System implant components can be rigidly locked into a variety of configurations, with each construct being tailor-made for the individual case. If necessary, the CD HORIZON® Spinal System can be connected to the VERTEX™ Reconstruction System through a rod connector.

Certain implant components from other Medtronic Sofamor Danek spinal systems can be used with the CD HORIZON® Spinal System. These components include TSRH® rods, hooks, screws, plates, CROSSLINK® plates, connectors, staples and washers; GDLH® rods, hooks, connectors and CROSSLINK® bar and connectors; LIBERTY® rods and screws; DYNALOK PLUS® bolts; and Medtronic Sofamor Danek Multi-Axial rods and screws.

CD HORIZON® hooks are intended for posterior use only. CD HORIZON® staples and CD HORIZON® ECLIPSE® rods and screws are intended for anterior use only. However, for patients of smaller stature, CD HORIZON® 4.5mm rods and associated components may be used posteriorly.

The purpose of this 510(k) submission is to include additional hooks, multi axial screws, fixed angle screws and sagittal adjusting screws to the CD HORIZON® Spinal System.

AI/ML Overview

The provided text describes the CD HORIZON® Spinal System and its indications for use, as well as the FDA's clearance of the device. However, it does not contain any information regarding acceptance criteria, device performance, or details of a study that proves the device meets specific performance criteria.

The document is a 510(k) premarket notification summary, which focuses on demonstrating substantial equivalence to a legally marketed predicate device, rather than providing detailed performance study results against specific acceptance criteria. This type of submission relies on the existing safety and effectiveness of the predicate device and the fact that the new device does not raise new questions of safety or effectiveness.

Therefore, I cannot populate the table or answer most of the questions you've posed based on the provided text. The document confirms:

  • Device: CD HORIZON® Spinal System
  • Manufacturer: Medtronic Sofamor Danek, Inc. USA
  • Regulatory Clearance: FDA 510(k) clearance (K042025) based on substantial equivalence to previously cleared CD HORIZON® Spinal System components.

To answer your questions, information from a dedicated performance study or clinical trial report would be required, which is not present in this 510(k) summary.

{0}------------------------------------------------

KOH2025

CD HORIZON® Spinal System Summary of Safety and Effectiveness July 2004

  • I. Medtronic Sofamor Danek, Inc. USA Company: 1800 Pyramid Place Memphis, TN 38132 (901) 396-3133 Contact: Richard W. Treharne, PhD
  • Proposed Proprictary Trade Name: CD HORIZON® Spinal System II.

Sr. Vice President, Regulatory Affairs

  • III. Classification Name: Spinal Interlaminal Fixation and Spinal Intervertebral Fixation Orthosis and/or l'edicle Screw Spinal System (per 21 CFR Section 888.3050, 888.3060 and/or 888.3070)

IV. Product Description

AUG 2 5 2004

The CD HORIZON® Spinal System consists of a variety of rods, hooks, screws, CROSSLINK® plates, staples, and other connecting components used to build a spinal construct. Instrumentation is also available to facilitate implantation of the device components.

The CD HORIZON® Spinal System is intended to help provide immobilization and stabilization of spinal segments as an adjunct to fusion of the thoracic, lumbar, and/or sacral spine. The CD HORIZON® Spinal System implant components can be rigidly locked into a variety of configurations, with each construct being tailor-made for the individual case. If necessary, the CD HORIZON® Spinal System can be connected to the VERTEX™ Reconstruction System through a rod connector.

Certain implant components from other Medtronic Sofamor Danek spinal systems can be used with the CD HORIZON® Spinal System. These components include TSRH® rods, hooks, screws, plates, CROSSLINK® plates, connectors, staples and washers; GDLH® rods, hooks, connectors and CROSSLINK® bar and connectors; LIBERTY® rods and screws; DYNALOK PLUS® bolts; and Medtronic Sofamor Danek Multi-Axial rods and screws.

CD HORIZON® hooks are intended for posterior use only. CD HORIZON® staples and CD HORIZON® ECLIPSE® rods and screws are intended for anterior use only. However, for patients of smaller stature, CD HORIZON® 4.5mm rods and associated components may be used posteriorly.

The purpose of this 510(k) submission is to include additional hooks, multi axial screws, fixed angle screws and sagittal adjusting screws to the CD HORIZON® Spinal System.

Page 1 of 2

000037

{1}------------------------------------------------

K042025

V. Indications

V. Co Indications
The CD HORIZON® Spinal System is intended for posterior, non-cervical fixation for the The CD HOROZON® Spinal System is intenate (defined as back pain of discogenic origin with following indications: degenerative discuse (uschies); spondylolisthesis; trauma
degeneration of the disc confirmed by history and radiographic studies); degeneration of the disc contifuled by instoly and tuckeraphic of the many of the materials and/or lordosis); tumor; pseudarthrosis; and/or failed previous fusion.

When used in a percutaneous, non-cervical, posterior approach with the SEXTANT when used in a perculancous, non oct rical, possessionended for the following indications: instrumentation, the CD HONZON® serews are intessogenic origin with degeneration of the disc
degenerative disc disease (defined as back pain of discogenic origin with of matu degeneralive discuse (demice as backed); spondylolisthesis; trauma (i.e., frauma (i.e., fracture or contrimed by instory and radiographio station, of 1977, scoliosis and/or lordosis); tumor, pseudoarthrosis; and/or failed previous fusion.

Except for hooks, when used as an anterolateral thoracic/lumbar system, CD HORIZON® Except for llocks, when used as an allercrates are intended for the following indications: (1) )
components such as ECLIPSE® components are intended for the following indica components such as ECLI'SE® components and interest origin with degeneration of the discussion of the disc degenerative disc disease (as delined by oater paint of answer (3) spondylolisthesis, (4)
confirmed by patient history and radiographic studies), (2) spondylolishesis, (7) confirmed by patient mistory and faciographie stazios), (5) fracture, (6) pseudarthrosis, (7) tumor resection, and/or (8) failed previous fusion.

The CD HORIZON® SPIRE Plate is posterior, non-pedicle supplemental fixation device, intended The CD HOKIZON® SI INE Tate is posterior, non prace fixation/attachment to spinous for use in the non-cervical spille (11) - 17). The intenses as the following conditions: degenerative process for the purpose of acine in a supprements. Incomith degeneration of the disconfirmed by dise disease - delined as back pan of disougane origin was are or dislocation); and/or tumor.

In order to achieve additional levels of fixation, the CD HORIZON® Spinal System rods may be In order to achieve addition levels of fixanon, and OD Th the VERTEX™ rod connector. Refer to connected to the VERTEX - Reconstruction Dystem with and into VERTEX™ indications of
the VERTEX™ Reconstruction System Package Insert for a list of the VERTEX™ indications of use.

Substantial Equivalence VI.

VI. Substantial Equirinchee
Documentation was provided which demonstrated the CD HORIZON® Spinal shorest Documentally equivalent to CD HORIZON® Spinal System components previously cleared in substantially equivalent to CD HOREON® Opinal BJScon 94, K034056, K981676 and K001532.
K040583, K020709, K031655, K041118, K032265, K000094, K034056, K981676 and K001532.

00000008

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three stylized, curved lines.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

AUG 2 5 2004

Richard W. Treharne, Ph.D. Sr. Vice President, Regulatory Affairs Medtronic Sofamor Danek, Inc. USA 1800 Pyramid Place Memphis. Tennessee 38132

Re: K042025

Trade/Device Name: CD HORIZON® Spinal System Regulation Number: 21 CFR 888.3050, 888.3060, 888.3070 Regulation Name: Spinal interlaminal fixation orthosis, Spinal intervertebral body fixation orthosis, Pedicle screw spinal system Regulatory Class: Class II Product Code: KWQ, KWP, MNH, MNI Dated: July 23, 2004 Received: July 28, 2004

Dear Dr. Treharne:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Mark A. Milburn

Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Page 1 of 1 July 2004

Device Name: CD HORIZON® Spinal System

Indications for Use:

The CD HORIZON® Spinal System is intended for posterior, non-cervical fixation for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kynhonis and/or lordosis); tumor; pseudarthrosis; and/or failed previous fusion.

When used in a percutaneous, non-cervical, posterior approach with the SEXTANT instrumentation, the CD HORIZON® screws are intended for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis and/or lordosis); tumor; pseudoarthrosis; and/or failed previous fusion.

Except for hooks, when used as an anterolateral thoracic/lumbar system, CD HORIZON® components such as ECLIPSE® components are intended for the following indications: (1) degenerative disc disease (as defined by back pain of discogenic origin with degencration of the disc confirmed by patient history and radiographic studies), (2) spinal stenosis, (3) spondylolisthesis, (4) spinal deformities (i.e., scoliosis, kyphosis, and/or lordosis), (5) fracture, (6) pseudarthrosis, (7) tumor resection, and/or (8) failed previous fusion.

The CD HORIZON® SPIRE Plate is posterior, non-pedicle supplemental fixation device, intended for use in the non-cervical spine (T1 - S1). It is intended for plate fixation/attachment to spinous process for the purpose of achieving supplemental fusion in the following cuaditions: degenerative disc disease - defined as back pain of discogenic origin with regeneration of the disc confirmed by history and radiographic studies; spondylolisthesis; trauma (i.e., fracture ro dislocation); and/or tumor.

In order to achieve additional levels of fixation, the CD HORIZON® Spinal System rods may be connected to the VERTEX™ Reconstruction System with the VERTEXTM rod connector. Refer to the VERTEX™ Reconstruction System Package Insert for a list of the VERTEXTM indications of use.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

(Division Sign-Off)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Division of General, Restorative, and Neurological Devices

510(k) Number K042035

000320

N/A