K Number
K130567

Validate with FDA (Live)

Date Cleared
2013-06-11

(99 days)

Product Code
Regulation Number
892.1680
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The software's intended use is to allow post-capture positioning and joining of individual images to produce an additional composite image. The software is sold only for use on Carestream Health system solutions and is available as an integrated feature within the DirectView software.

Device Description

The DR Long Length Imaging Software is a software component for use on a diagnostic x-ray system utilizing digital radiography (DR) technology. The software option is comprised of image stitching software and a manual stitch editor that allows user adjustment to automatic stitching results. The software allows up to 5 separately acquired DR images to be stitched together to create a "composite" long length image.

AI/ML Overview

The acceptance criteria and study details for the DR Long Length Imaging Software (K130567) are as follows:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria CategoryAcceptance Criteria (Implicit)Reported Device Performance
Diagnostic AcceptabilityComposite images must be clinically acceptable for diagnostic use."Predefined acceptance criteria were met and demonstrated that the device is as safe, as effective, and performs as well as the predicate device." (Implies diagnostic acceptability was achieved).
Design Input/Output VerificationDesign output must meet design input requirements."Performance testing was conducted to verify the design output met the design input requirements..."
User Needs/Intended Use ValidationDevice must conform to defined user needs and intended uses."...and to validate the device conformed to the defined user needs and intended uses."
Safety and Effectiveness EquivalenceDevice must be as safe, as effective, and perform as well as the predicate device."Predefined acceptance criteria were met and demonstrated that the device is as safe, as effective, and performs as well as the predicate device."

2. Sample Size Used for the Test Set and Data Provenance

The document does not explicitly state the numerical sample size used for the clinical testing (test set). It only mentions that "Clinical testing was conducted to evaluate the acceptability of the composite images for diagnostic use."

The data provenance (country of origin, retrospective/prospective) is not specified.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

The document does not provide information on the number of experts used or their qualifications for establishing ground truth in the clinical testing.

4. Adjudication Method for the Test Set

The document does not specify any adjudication method (e.g., 2+1, 3+1, none) for the test set.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Its Effect Size

The document does not mention a multi-reader multi-case (MRMC) comparative effectiveness study, nor does it provide any effect size for human readers improving with or without AI assistance. The study described focuses on the device's performance in creating diagnostically acceptable images.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Study Was Done

The testing description includes "Nonclinical testing was conducted under simulated use conditions," which could imply standalone algorithm performance was assessed for technical aspects, but the primary clinical evaluation was described as "acceptability of the composite images for diagnostic use," suggesting human review. The software includes a "manual stitch editor that allows user adjustment to automatic stitching results," indicating a human-in-the-loop component is integral to its intended use. Therefore, a purely standalone clinical performance study for diagnostic interpretation does not appear to be the focus.

7. The Type of Ground Truth Used

The ground truth for the clinical testing was based on the "acceptability of the composite images for diagnostic use." This implies a qualitative assessment by medical professionals, likely radiologists, but the specific method (e.g., expert consensus) is not detailed.

8. The Sample Size for the Training Set

The document does not provide information about a training set or its sample size. The focus is on validation testing of the software.

9. How the Ground Truth for the Training Set Was Established

Since no training set is mentioned, there is no information on how its ground truth was established.

{0}------------------------------------------------

Traditional 510(k) for DR LLI Software

K130567
page 1 of 2

Carestream

JUN 1 1 2013

"510(k) Summary"

585-627-6543

585-323-7643

510(k) Owner Name: 510(k) Owner Address: Carestream Health, Inc. 150 Verona Street Rochester, New York 14608

510(k) Owner Phone: 510(k) Owner Fax:

Contact Person & Info:

Carolyn Wagner Regulatory Affairs Manager, Medical Imaging carolyn.wagner@carestreamhealth.com 585-627-6588

Date Summary Prepared:

June 5, 2013

Device Trade Name: Device Common Name: Classification Name:

Device Class: Device Code: Regulation Number:

Predicate Devices:

DR Long Length Imaging Software DR digital imager Solid state x-ray imager (flat panel/digital imager)

Class II MQB 21 CFR 892.1650

Cedara I-Softview™ (Accustitch"M component) Manufactured by Cedara Software Corp. 510(k) No. - K022881 (October 22, 2002)

Kodak Eclipse Image Processing Software Manufactured by Carestream Health, Inc. 510(k) No. - K060137 (May 16, 2006)

Device Description:

The DR Long Length Imaging Software is a software component for use on a diagnostic x-ray system utilizing digital radiography (DR) technology. The software option is comprised of image stitching software and a manual stitch editor that allows user adjustment to automatic stitching results. The software allows up to 5 separately acquired DR images to be stitched together to create a "composite" long length image.

{1}------------------------------------------------

Traditional 510(k) for DR LLI Softwar

K130561
Page 2 of 2

Indications for Use / Intended Use:

The Indications for Use for the device, as described in its labeling, are: "The software's intended use is to allow post-capture positioning and joining of individual images to produce an additional composite image. The software is sold only for use on Carestream Health system solutions and is available as an integrated feature within the DirectView software."

The intended use for this device, as determined by descriptions and the proposed labeling contained in this submission, is similar to the Indications for Use statement provided above. The DR Long Length Imaging software component provides the ability to perform operations related to image manipulation and enhancement of medical images. We believe that the DR Long Length Imaging software and the predicate devices have the same intended use.

The Indications for Use for the subject device is different than the predicate device, but these differences do not alter the intended diagnostic use of the software component. Differences are appropriately characterized as descriptive, and the intended use remains unchanged. Any variation in features or technical specifications have been identified and addressed through testing (described below) to support a substantial equivalence determination.

Comparison of Technological Characteristics:

The DR Long Length Imaging Software has the same technological characteristics as the predicate device, the Cedara I-Softview "Accustitch" component. The DR Long Length Imaging Software is an optional software component that allows a composite image to be formed from up to five separately acquired general radiography images. It is available as an integrated optional feature within the predicate device Kodak Eclipse Image Processing Software (DirectView software). The software device raises no new issues of safety or effectiveness.

Discussion of Testing

Performance testing was conducted to verify the design output met the design input requirements and to validate the device conformed to the defined user needs and intended uses. Nonclinical testing was conducted under simulated use conditions. Clinical testing was conducted to evaluate the acceptability of the composite images for diagnostic use. Predefined acceptance criteria were met and demonstrated that the device is as safe, as effective, and performs as well as the predicate device.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo features a stylized image of an eagle with three lines forming its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G Silver Spring. MD 20993-0002

June 11. 2013

Carestream Health, Inc. % Ms. Carolyn L. Wagner Regulatory Affairs Manager 150 Verona Street ROCHESTER NY 14608

Re: K130567

Trade/Device Name: DR Long Length Imaging Software Regulation Number: 21 CFR 892.1650 Regulation Name: Solid State X-Ray Imager (Flat Panel/Digital Imager) Regulatory Class: Class II Product Code: MQB Dated: May 7, 2013 Received: May 10, 2013

Dear Ms. Wagner:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). Yourmay therefore market-the device subject to the general controls provisions of the Act .- The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic

{3}------------------------------------------------

Page 2 - Ms. Wagner

product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.lda.gov/MedicalDevices/Resourcesfor.You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safetv/ReportalProblem/default.htm for the CDRF/'s Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industrv/default.htm.

Sincerely yours,

for

Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use

510(k) Number (if known): K130567

Device Name: DR Long Length Imaging Software

Indications for Use:

The software's intended use is to allow post-capture positioning and joining of individual images to produce an additional composite image. The software is sold only for use on Carestream Health system solutions and is available as an integrated feature within the DirectView software.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

Sm. P)

(Division Sign-Off) Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health

K130567 510(k)

Page 1 of 1

N/A