(22 days)
Orbit Infusion Sets are intended for the subcutaneous delivery of fluids and medication, such as insulin, from an external infusion pump.
Orbit Infusion sets are comprised of tubing that is connected on one end to the medication reservoir of the infusion pump using a luer lock connection and on the other end to the patient, attached to the skin by an adhesive base, anchoring a catheter that is inserted into the subcutaneous tissue. These sets have a patented design feature which allows the tubing to freely rotate 360° at the adhesive attachment and to disconnect the tubing set from the infusion base.
The Orbit Infusion Set is intended for the subcutaneous delivery of fluids and medication from an external infusion pump. The device aims to demonstrate substantial equivalence to the predicate device, Talon Infusion Set (K033290), by meeting performance and safety criteria.
1. Table of Acceptance Criteria and Reported Device Performance
| Test | Acceptance Criteria | Reported Device Performance (Compliance) |
|---|---|---|
| Material strength of steel cannula | Per ISO 9626 | Met relevant sections of standards |
| Activated pressure leak | No leak when subjected to pumping pressures up to 20psi under normal delivery conditions and occluded fluid path conditions | Met relevant sections of standards |
| Penetration force | Needle and catheter shall penetrate a 0.025 inch thick membrane with a speed of 50mm/min. and a force of less than 0.8N | Met relevant sections of standards |
| Needle retention | No separation of the needle from the cap when subjected to a minimum force of 10N (ISO 10555-1, Annex B) | Met relevant sections of standards |
| Catheter retention | No separation of the catheter from the base when subjected to a minimum force of 3N (ISO 8536-8) | Met relevant sections of standards |
| Bond strength of tubing/fittings | No separation of the tubing assembly when subjected to a static tensile force of 15N for 15 sec. | Met relevant sections of standards |
| Bond strength of tape/base | No separation of the tape from the base when subjected to a minimum force of 18N | Met relevant sections of standards |
| Engagement force tubing cap/base | The cap locks on the base with a force less than 13N | Met relevant sections of standards |
| Disengagement force tubing cap/base | The force to remove the tubing cap from the base is more than 13N | Met relevant sections of standards |
| Occlusion test | No occlusion of the device when tested with a water flow at a hydrostatic pressure of 0.1 bar | Met relevant sections of standards |
| Tape adhesion | Removal of adhesive from a stainless steel plate with a 90 degree peel force of minimum 2.5N (0.56lbs) | Met relevant sections of standards |
The study concluded that "The verifications have shown evidence that the Orbit Infusion Sets meet the acceptance criteria of these standards. Based on the results it can be concluded that the device performance and safety are acceptable for the product."
2. Sample size used for the test set and the data provenance
The document does not explicitly state the sample size used for each specific test. The data provenance is described as "relevant assessments specified in the following international and internal standards and protocols." This suggests that the tests were conducted in a controlled laboratory environment according to established engineering and medical device standards. There is no mention of country of origin of the data, or if it was retrospective or prospective human subject data, as these are technical performance tests on the device itself.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable. The ground truth for these performance tests is based on the specified engineering and medical device standards, which are objective measurements. Expert consensus or clinical interpretation of data is not required to establish ground truth for these types of device performance criteria.
4. Adjudication method for the test set
Not applicable. Adjudication methods like 2+1 or 3+1 are typically used in studies involving human interpretation or clinical outcomes where ambiguity might exist. For objective performance tests with clear pass/fail criteria based on standards, an adjudication method is not relevant.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This submission is for a medical device (infusion set) and not an AI or imaging diagnostic tool. Therefore, an MRMC comparative effectiveness study involving human readers and AI assistance is not relevant to this device.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. This is a physical medical device, not an algorithm, so standalone algorithm performance is not relevant.
7. The type of ground truth used
The ground truth for the performance tests is based on objective measurements defined by international and internal engineering/medical device standards. For example, "Material strength per ISO 9626," "No leak when subjected to pumping pressures up to 20psi," and "No separation...when subjected to a minimum force of 10N (ISO 10555-1, Annex B)."
8. The sample size for the training set
Not applicable. This is a physical medical device being evaluated against engineering and safety standards, not a machine learning model that requires a training set.
9. How the ground truth for the training set was established
Not applicable. As this is not a machine learning model, there is no training set and therefore no ground truth establishment for a training set.
{0}------------------------------------------------
MAR 1 9 2013
Summary of Safety and Effectiveness
"510(K) SUMMARY" -
Submitted By/ Contact Person:
Dr. Benjamin Reinmann Vice President Quality Management & Regulatory Affairs Ypsomed AG Brunnmattstrasse 6 CH - 3401 Burgdorf Switzerland Tel. +41 34 424 3117 Fax. +41 34 424 3460 E-mail: benjamin.reinmann@ypsomed.com
Date Prepared:
February 19, 2013
Orbit Infusion Set
Subcutaneous Infusion Set
Set, Administration, Intravascular
1.1. Trade/Proprietary Name:
1.2. Common/Usual Name:
1.3. Classification Name:
1.4. Classification:
l .5. Purpose of Submission:
1.6. Substantial Equivalence:
1.7. Device Description
Class: I I Panel: 80 Product Code: FPA 21 CFR 880.5440 Cite:
Yosomed acquired the Orbit infusion set business of ICU Medical Inc., USA in November 2011, including 510(k) K033290 under which the devices were cleared for marketing as the Talon Infusion Set. The purpose of this submission is to update the premarket notification file to reflect the product as manufactured. The only additional change is to modify the labeling to reflect Ypsomed, which previously had been a distributor of the product manufactured by ICU Medical.
The Orbit Infusion Sets are substantially equivalent to the Talon Infusion Set (K033290). The equivalence is supported by the attached documentation.
The Orbit Infusion Set is used for the subcutaneous delivery of fluids and medication from an external infusion pump.
{1}------------------------------------------------
Orbit Infusion sets are comprised of tubing that is connected on one end to the medication reservoir of the infusion pump using a luer lock connection and on the other end to the patient, attached to the skin by an adhesive base, anchoring a catheter that is inserted into the subcutaneous tissue. These sets have a patented design feature which allows the tubing to freely rotate 360° at the adhesive attachment and to disconnect the tubing set from the infusion base.
1.8. Intended Use:
The intended use of the modified device remains the same as the predicate device (Talon Infusion Set. K033290): Orbit Infusion Sets are intended for the subcutaneous delivery of fluids and medication, such as insulin, from an external infusion pump.
- 1.9. Technological Characteristics:
Orbit Infusion Sets are considered substantial equivalent to K033290. Talon Infusion Set, in intended use and in the device's operating principles.
1.10. Performance and Safety Data: Yosomed has performed the relevant assessments specified in the following international and internal standards and protocols and confirmed compliance of the modified devices and equivalence to the predicate devices. The Orbit Infusion Sets have met the requirements of the relevant sections of the following standards:
| Test | Specification |
|---|---|
| Material strength of steel cannula | Material strength per ISO 9626 |
| Activated pressure leak | No leak when subjected to pumping pressures up to20psi under normal delivery conditions andoccluded fluid path conditions |
| Penetration force | Needle and catheter shall penetrate a 0.025 inchthick membrane with a speed of 50mm/min. and aforce of less than 0.8N |
| Needle retention | No separation of the needle from the cap whensubjected to a minimum force of 10N (ISO 10555-1,Annex B) |
| Catheter retention | No separation of the catheter from the base whensubjected to a minimum force of 3N (ISO 8536-8) |
| Bond strength of tubing/fittings | No separation of the tubing assembly whensubjected to a static tensile force of 15N for 15 sec. |
| Bond strength of tape/base | No separation of the tape from the base whensubjected to a minimum force of 18N |
| Engagement force tubing cap/base | The cap locks on the base with a force less than 13N |
| Disengagement force tubingcap/base | The force to remove the tubing cap from the base ismore than 13N |
{2}------------------------------------------------
| Occlusion test | No occlusion of the device when tested with a waterflow at a hydrostatic pressure of 0.1 bar |
|---|---|
| Tape adhesion | Removal of adhesive from a stainless steel platewith a 90 degree peel force of minimum 2.5N(0.56lbs) |
The verifications have shown evidence that the Orbit Infusion Sets meet the acceptance criteria of these standards. Based on the results it can be concluded that the device performance and safety are acceptable for the product.
1.11. Conclusion Ypsomed AG concludes based on the information presented that the modified product is substantially equivalent to the current product legally approved in the USA.
{3}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized graphic of an eagle or bird-like figure, with three curved lines representing its wings or body.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Dr. Beniamin Reinmann · Vice President Ypsomed AG Quality Management & Regulatory Affairs Brunnmattstrasse 6 Burgdorf Switzerland CH-3401
Re: K130468
Trade/Device Name: Orbit Infusion Sets Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: II Product Code: FPA Dated: February 22, 2013 Received: February 25, 2013
Dear Dr. Reinmann:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
March 19, 2013
{4}------------------------------------------------
Page 2 - Dr. Reinmann
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Indications for Use
510(k) Number (if known):
K 130468
Device Name:
Indications For Use:
Orbit Infusion Sets
Orbit Infusion Sets are intended for the subcutaneous delivery of fluids and medication, such as insulin, from an external infusion pump.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
FDA
Richard C. Chapman 2013.03.18 15:38:51 -04'00'
:
| (Division Sign-Off) |
|---|
| --------------------- |
| Division of Anesthesiology, General Hospital |
|---|
| ---------------------------------------------- |
| Infection Control, Dental Devices |
|---|
| ----------------------------------- |
K130468 510(k) Number:
page 2 of 2
§ 880.5440 Intravascular administration set.
(a)
Identification. An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.(b)
Classification. Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.