K Number
K130211
Date Cleared
2013-07-05

(157 days)

Product Code
Regulation Number
880.2800
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The SPSmedical AirView™ Bowie Dick Test Pack and AirView™ Bowie Dick Indicator Sheets are designed to detect the presence of residual air in pre-vacuum steam sterilizers. When tested in a prevacuum sterilizer operating at 134°C the indicator will demonstrate a uniform color change from cream or blue to dark brown/black when proper sterilization conditions have been met and no air is detected. It is designed to be used for daily Bowie Dick testing of steam sterilizers as described in ANSVAAMI ST79.

Device Description

The SPSmedical AirView™ Bowie Dick Test Pack is a disposable device which is intended for use as a replacement for the ANSI/AAMI ST79 Bowie Dick Test Pack. The SPSmedical AirView™ Bowie Dick Indicator Sheets are designed to be used within the ANSI/AAMI ST79 Bowie Dick test pack. The layers of paper within the SPSmedical AirView™ Bowie Dick Test Pack provide resistance to steam penetration and also trap air between them which is difficult for marginal performing prevacuum sterilizers to remove. Both products are comprised of paper sheets, foam, a containment box or wrap, external chemical indicator label and printed indicator sheet.

AI/ML Overview

The provided text describes a submission for a 510(k) premarket notification for the "AirView™ Bowie Dick Test Pack and AirView™ Bowie Dick Indicator Sheets". This device is a chemical indicator used to detect residual air in pre-vacuum steam sterilizers.

Here's an analysis of the acceptance criteria and study information based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance:

The document doesn't explicitly state "acceptance criteria" in a quantified manner typical of a medical device study (e.g., "Sensitivity must be >90%"). Instead, it presents performance claims and criteria for substantial equivalence to a predicate device. The core performance criterion is the ability to detect residual air by displaying a non-uniform color change.

Criterion TypeAcceptance Criterion (Implied/Claimed)Reported Device Performance
Intended UseDetect the presence of residual air in pre-vacuum steam sterilizers operating at 134°C.Same as intended use. Provides visual indication if residual air was left.
Visual IndicationUniform dark brown/black signal color under proper sterilization and air removal conditions. Non-uniform color change for failure.Achieved; uniform dark brown/black with proper conditions, non-uniform with failure.
Air DetectionSensitive enough to detect when enough air is left within the sterilizer chamber to create a 2°C or greater temperature difference in the test pack.Demonstrated with testing, resulting in a non-uniform color change.
Shelf LifeThree (3) years from date of manufacture.Verified three (3) years.
BiocompatibilityNon-toxic (as the device is intended to be run in an empty chamber, but addressed for safety).Concluded to be non-toxic.
StorageStore in a cool, dry place (15-30°C).Confirmed.
Interfering SubstancesNot sensitive to acidic or basic environments (both unprocessed and processed forms).Verified not sensitive to acidic or basic environments.
Equivalence to Predicate"AirView™ Bowie Dick Test Pack & AirView™ Bowie Dick Indicator Sheets (K041017)" for intended use, functional characteristics, design, endpoint color, indicator agent, sterilization method, device materials, and performance.Demonstrated and claimed substantial equivalence.

2. Sample Size Used for the Test Set and Data Provenance:

The document broadly states: "We have demonstrated with testing that the SPSmedical AirView™ Bowie Dick Test Pack and AirView™ Bowie Dick Indicator Sheets perform consistently with results which indicate that the indicator is sensitive enough to detect when enough air is left within the sterilizer chamber to create a 2°C or greater temperature difference in the test pack as compared to the sterilizer chamber when running pre-vacuum sterilization at 134°C."

  • Sample Size for Test Set: Not specified. The document mentions "test data obtained during validation studies" and "testing verified," but does not provide specific numbers of test packs/sheets used or the number of tests performed.
  • Data Provenance: The document does not explicitly state the country of origin. Given the submitter (SPSmedical Supply Corp. in Rush, NY, U.S.A.) and the submission to the FDA, it's highly probable the testing was conducted in the USA. The data is retrospective in the sense that it's presented as already completed validation studies for the 510(k) submission, not as a prospective clinical trial.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

Not applicable. This device is a chemical indicator with an objective visual color change. The ground truth for its performance is established by physical measurements (e.g., temperature difference in the sterilizer chamber) and the objective observation of color change, not by expert interpretation of complex images or clinical outcomes.

4. Adjudication Method for the Test Set:

Not applicable. As noted above, the "failure" or "success" of the test pack is determined by a visual color change and the physical conditions it's designed to detect (presence of residual air creating a temperature differential). There is no mention of a human adjudication process for interpreting results.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done:

No. This device is a chemical indicator, not a diagnostic imaging or AI-driven decision support system. MRMC studies are used to assess the impact of a device (often AI) on human reader performance. This device has an objective output (color change).

6. If a Standalone Study (Algorithm Only Without Human-in-the-Loop Performance) Was Done:

Yes, in essence. The "study" described is the standalone performance of the Bowie Dick test pack itself. Its function is to autonomously detect residual air and produce a visual signal. The described testing verifies the indicator's sensitivity to physical conditions (2°C temperature difference). There is no "human-in-the-loop" component to the detection mechanism of this device; humans merely observe and interpret the final color change.

7. The Type of Ground Truth Used:

The ground truth used is based on physical conditions and outcomes in a sterilizer environment:

  • The actual presence of residual air within the sterilizer chamber.
  • The resulting temperature difference (specifically, a 2°C or greater difference) in the test pack compared to the sterilizer chamber due to this residual air.
  • The absence of proper sterilization and air removal conditions, leading to a non-uniform color change.
  • The presence of proper sterilization and air removal conditions, leading to a uniform dark brown/black color change.

8. The Sample Size for the Training Set:

Not applicable. This device is a passive chemical indicator, not an AI or machine learning algorithm that requires a training set. Its functionality is based on chemical reactions and physical design, not learned patterns from data.

9. How the Ground Truth for the Training Set Was Established:

Not applicable, as there is no training set for this type of device.

{0}------------------------------------------------

KISOZIT

510(k) Summary

SUBMITTER INFORMATION:

JUL 5 2013

SPSmedical Supply Corp. a division of Crosstex International 6789 West Henrietta Road Rush, NY 14543 U.S.A. Phone: (800) 722-1529 Fax: (585) 359-0167

Contact Person:Michael NolanResearch and Development Coordinator
Date of Summary:July 5, 2013
DEVICE INFORMATION:
Device Trade Name:AirView™ Bowie Dick Test Pack and AirView™ Bowie Dick IndicatorSheets
Common Name:Bowie Dick Test Pack and Bowie Dick Indicator Sheets
Device Classification:Indicator, Physical/Chemical Sterilization Process21 CFR § 880.2800
Device Class:2
Product Code:JOJ

PREDICATE DEVICE:

AirView™ Bowie Dick Test Pack and AirView™ Bowie Dick Indicator Sheets K041017

DEVICE DESCRIPTION:

The SPSmedical AirView™ Bowie Dick Test Pack is a disposable device which is intended for use as a replacement for the ANSI/AAMI ST79 Bowie Dick Test Pack. The SPSmedical AirView™ Bowie Dick Indicator Sheets are designed to be used within the ANSI/AAMI ST79 Bowie Dick test pack

INTENDED USE:

The SPSmedical AirView™ Bowie Dick Test Pack and AirView™ Bowie Dick Indicator Sheets are designed to detect the presence of residual air in pre-vacuum steam sterilizers operating at 134°C. The Bowie Dick test pack and indicator sheets are reliable tools used for the monitoring of air removal in steam pre-vacuum processes and provides a visual indication if residual air was left in the chamber during sterilization. Its internal indicator sheet changes to a uniform dark brown/black signal color under proper sterilization and air removal conditions. A failure would result in a nonuniform color change on the indicator sheet.

TECHNICAL CHARACTERISTICS:

The layers of paper within the SPSmedical AirView™ Bowie Dick Test Pack provide resistance to steam penetration and also trap air between them which is difficult for marginal performing prevacuum sterilizers to remove.

RECOMMENDED STORAGE CONDITIONS:

Store in a cool, dry place (15-30°C)

{1}------------------------------------------------

510(k) Summary (Continued)

INTERFERING SUBSTANCES OR CONDITIONS:

Testing verified that the indicators in their unprocessed form are not sensitive to an acidic or basic environment. Testing verified that the indicators in their processed form are not sensitive to an acidic or basic environment.

SHELF LIFE:

The Shelf Life of the SPSmedical AirView™ Bowie Dick Test Pack and AirView™ Bowie Dick Indicator Sheets shall be three (3) years from the date of manufacture, when stored in a cool, dry place (15-30°C)

BIOCOMPATIBILITY:

The purpose of a biocompatibility test is to demonstrate that the active ingredients of the chemical indicator do not release any substance known to be toxic onto surgical instruments. Although the SPSmedical AirView™ Bowie Dick Test Pack and AirView™ Bowie Dick Indicator Sheets are intended to be run in an empty chamber it was necessary to address this concern. Biocompatibility testing on the indicators concluded that the SPSmedical AirView™M Bowie Dick Test Pack and AirView™ Bowie Dick Indicator Sheets were non-toxic.

SUBSTANTIAL EQUIVALENCE ANALYSIS

SPSmedical has identified the AirView™ Bowje Dick Test Pack & AirView™ Bowie Dick Indicator Sheets (K041017) as the (primary) predicate device. We believe the predicate device to be substantially equivalent to the AirView™ Bowie Dick Test Pack & AirView™ Bowie Dick Indicator Sheets which is the subject of this submission in terms of their intended use and functional characteristics in determining the effectiveness of air removal in pre-vacuum steam sterilizers.

FUNCTIONAL CHARACTERISTICS:

The SPSmedical AirView™ Bowie Dick Test Pack and Air View™ Bowie Dick Indicator Sheets are designed to detect the presence of residual air in pre-vacuum steam sterilizers operating at 134°C. The layers of paper within the pack provide resistance to steam penetration and trap residual air which is difficult for marginal performing pre-vacuum sterilizers to remove. The pack is a reliable tool used for the monitoring of air removal in pre-vacuum steam sterilizers and provides a visual indication if residual air was left in the chamber during sterilization. The indicator sheet changes from its initial color to a uniform dark brown/black signal color under proper sterilization and air removal conditions. A failure would result in a non-uniform color change on the indicator sheet.

DISCUSSION:

SPSmedical is claiming substantial equivalence for its SPSmedical AirView™ Bowie Dick Test Pack and AirView™ Bowie Dick Indicator Sheets to the original AirView™ Bowie Dick Test Pack and AirView™ Bowie Dick Indicator Sheets (K041017) based on test data obtained during validation studies. We have demonstrated with testing that the SPSmedical AirView™ Bowie Dick Test Pack and AirView™ Bowie Dick Indicator Sheets perform consistently with results which indicate that the indicator is sensitive enough to detect when enough air is left within the sterilizer chamber to create a 2°C or greater temperature difference in the test pack as compared to the sterilizer chamber when running pre-vacuum sterilization at 134°C. Under these conditions, the indicator sheet would demonstrate a non-uniform color change. The SPSmedical AirView™ Bowie Dick Test Pack and AirView™ Bowie Dick Indicator Sheets are also comparable to other commercially available Bowie Dick test packs cleared by the FDA. See examples of similarly cleared devices in Tab 8.

{2}------------------------------------------------

SUBSTANTIAL EQUIVALENCE CONCLUSIONS:

The SPSmedical AirView™ Bowie Dick Test Pack and AirView™ Bowie Dick Indicator Sheets have the same intended use and characteristics as the primary predicate device (K041017). They both provide a visual indication that proper air removal and steam penetration conditions have been met within the sterilizer's chamber. Both products are comprised of paper sheets, foam, a containment box or wrap, external chemical indicator label and printed indicator sheet.

SPSmedical believes that the SPSmedical AirView™ Bowie Dick Test Pack and AirView™ Bowie Dick Indicator Sheets are substantially equivalent to the predicate device because it has the same intended use, technical characteristics and performance. Because the ability to perform its intended function as been shown through validated testing, the SPSmedical AirView™ Bowie Dick Test Pack and AirView™ Bowie Dick Indicator Sheets raise no issues related to safety or effectiveness and therefore, the device should be allowed for market in the United states. See Table 1 below for a substantial equivalence comparison.

ElementModified DevicePredicate(K041017)
Intended UseAir Removal IndicatorAir Removal Indicator
Device DesignBowie Dick Test Pack or SheetBowie Dick Test Pack or Sheet
Endpoint ColorDark Brown/BlackDark Brown/Black
Indicator AgentIndicator InkIndicator Ink
Sterilization MethodSteam Pre-vacuumSteam Pre-vacuum
Device MaterialsEquivalentEquivalent
Air PorosityEquivalentEquivalent
Performance understandard fault conditionEquivalentEquivalent
Performance understandard Bowie Dick cycleEquivalentEquivalent
Shelf LifeThree (3) yearsThree (3) years

TABLE 1-COMPARISON OF THE MODIFIED DEVICE TO THE PREDICATE

CONCLUSION:

Supportive data has demonstrated that the SPSmedical AirView™ Bowie Dick Test Pack and AirView™M Bowie Dick Indicator Sheets are equivalent to the predicate device. The SPSmedical AirView™ Bowie Dick Test pack and AirView™ Bowie Dick Indicator Sheet raise no issues related to safety or effectiveness and therefore should be allowed for market in the United States.

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows a circular logo for the U.S. Department of Health & Human Services. The logo features the department's emblem, which consists of a stylized caduceus-like symbol with three wavy lines extending from a central point. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the upper portion of the circle, following its curvature.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

July 5, 2013

SPS Medical Supply Corporation C/O Mr. Michael Nolan Research and Development Coordinator 6789 West Henrietta Road RUSH, NY 14543

Re: K130211

Trade/Device Name: AirView™ Bowie Dick Test Pack and AirView™ Bowie Dick Indicator Sheets Regulation Number: 21 CFR 880.2800 Regulation Name: Sterilization Process Indicator Regulatory Class: II Product Code: JOJ Dated: May 30, 2013 Received: June 7, 2013

Dear Mr. Nolan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2 - Mr. Nolan

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely vours.

Image /page/4/Picture/8 description: The image shows a signature and some text. The text includes the name "Tefashri Purohit-Sheth, M.D." and the title "Clinical Deputy Director". The image also contains the acronym "DAGRID".

Kwame Ulmer, M.S. Acting Division Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health'

Enclosure

{5}------------------------------------------------

Statement of Indications for Use

5 10(k) Number (if known): K130211

Device Trade Name:

AirView™ Bowie Dick Test Pack and AirView™ Bowie Dick Indicator Sheets

INDICATIONS FOR USE:

The SPSmedical AirView™ Bowie Dick Test Pack and AirView™ Bowie Dick Indicator Sheets are designed to detect the presence of residual air in pre-vacuum steam sterilizers. When tested in a prevacuum sterilizer operating at 134°C the indicator will demonstrate a uniform color change from cream or blue to dark brown/black when proper sterilization conditions have been met and no air is detected. It is designed to be used for daily Bowie Dick testing of steam sterilizers as described in ANSVAAMI ST79.

Prescription Use __ (Part 21 CFR 801 Subpart D) and/or

Over the Counter Use X (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation

Elizabeth F. Clavette
2013.06.28 21:36:53 04'00'

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

510(k) Number;

§ 880.2800 Sterilization process indicator.

(a)
Biological sterilization process indicator —(1)Identification. A biological sterilization process indicator is a device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor adequacy of sterilization. The device consists of a known number of microorganisms, of known resistance to the mode of sterilization, in or on a carrier and enclosed in a protective package. Subsequent growth or failure of the microorganisms to grow under suitable conditions indicates the adequacy of sterilization.(2)
Classification. Class II (performance standards).(b)
Physical/chemical sterilization process indicator —(1)Identification. A physical/chemical sterilization process indicator is a device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor one or more parameters of the sterilization process. The adequacy of the sterilization conditions as measured by these parameters is indicated by a visible change in the device.(2)
Classification. Class II (performance standards).