(125 days)
ExSpiron 1xi is indicated for use by healthcare professionals in healthcare facilities, such as post-operative care and critical care units, to monitor breathing in adult (at least 21 years old) patients.
ExSpiron 1xi is a non-invasive system that graphically displays lung volume against time and reports an approximate value of:
Tidal volume,
Respiratory rate, and
Minute ventilation
ExSpiron 1xi measurements are used as an adjunct to other clinical information sources.
The ExSpiron 1xi system consists of:
Bioimpedance measurement system: A stabilized high frequency current generator is connected to two outer electrodes. The inner four electrodes are connected to an adaptive circuit that conditions the resulting voltage signal and converts it to digital form. Firmware performs signal acquisition and relays data to the panel PC.
Computer: A Windows 7 PC performs signal processing and calibration, and runs the graphical user interface (GUI). The PC takes user input from a touch screen. The GUI is used for recording patient data and displaying the respiratory trace as well as scalar values and trends for minute volume, tidal volume, and respiratory rate.
Single Patient Use ExSpiron™ 1xi Electrode PadSet: An electrode assembly containing six electrodes to be placed on the torso. It delivers current and records impedance measurements. The electrode PadSet is also used to perform subsystem checks prior to patient measurements.
The ExSpiron™ 1xi respiratory monitor is indicated for use by healthcare professionals in healthcare facilities to monitor breathing in adult (at least 21 years old) patients. It is a non-invasive system that graphically displays lung volume against time and reports approximate values for Tidal Volume, Respiratory Rate, and Minute Ventilation. The measurements are used as an adjunct to other clinical information sources.
The provided documentation does not contain a table of acceptance criteria or reported device performance for this specific 510(k) submission (K130170). The submission is a Special 510(k) where the manufacturer states that "No clinical testing was performed for this 510(k). The measurement algorithm and software are unchanged from the predicate. Measurement performance statistics were determined by clinical trials reported in the predicate 510(k), K120087." This means that the device's performance relies on the data submitted for its predecessor, the ExSpiron Respiratory Monitor (K120087).
However, based on the information provided for K130170:
1. Table of Acceptance Criteria and Reported Device Performance:
As stated above, no specific acceptance criteria or reported performance for the ExSpiron™ 1xi is provided in this document as it references the predicate device. For the predicate device, K120087, the performance statistics would have been determined through clinical trials. Without access to K120087, these specific details cannot be provided here.
The document states that changes for K130170 were related to:
- Software Verification & Validation
- Safety Testing - IEC 60601-1 2nd Edition
- Electromagnetic Compatibility - IEC 60601-1-2
- ExSpiron 1Xi System Test Plan
- Patient Cable Design Verification (AAMI / ANSI EC53:1995/(R) 2008)
- Electrode PadSet Design Verification (AAMI / ANSI EC12:2000/(R)2010, AAMI / ANSI EC53:1995/(R) 2008)
- Electrode - Biocompatibility - ANSI/AAMI/ISO 10993-1
All performance testing for these modifications was reported as successful, raising no new safety or efficacy concerns and demonstrating substantial equivalence to the predicate device.
2. Sample Size for Test Set and Data Provenance:
- Sample Size for Test Set: Not applicable for this specific 510(k) (K130170) as no new clinical testing was performed. The measurement performance statistics were determined by clinical trials reported in the predicate 510(k), K120087.
- Data Provenance: Not applicable for this specific 510(k). For the predicate device (K120087), this information would be detailed in its submission. It's likely that the data would be prospective, given the nature of clinical trials for device performance. Country of origin not specified in the current document.
3. Number of Experts and Qualifications for Ground Truth - Test Set:
- Not applicable for this specific 510(k) (K130170) as no new clinical testing was performed. The ground truth for the predicate device's clinical trials (K120087) would have been established through a reference standard method, but the number and qualifications of experts are not described in this document.
4. Adjudication Method for the Test Set:
- Not applicable for this specific 510(k) (K130170) as no new clinical testing was performed.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:
- No MRMC comparative effectiveness study was done for this specific 510(k) (K130170). The device is a monitor that reports values, not an AI for interpretation requiring human readers.
6. Standalone (Algorithm Only) Performance Study:
- Yes, implicitly. The statement that "The measurement algorithm and software are unchanged from the predicate" implies that the core algorithm's performance was evaluated in a standalone manner during the clinical trials for the predicate device (K120087). This 510(k) (K130170) focuses on modifications to hardware components and software updates that do not alter the fundamental measurement algorithm.
7. Type of Ground Truth Used:
- For the performance statistics referenced from the predicate K120087, the ground truth for respiratory measurements (Tidal Volume, Respiratory Rate, Minute Ventilation) would typically be established using a gold standard reference measurement system (e.g., highly accurate spirometry, capnography, or volumetric respirators) rather than expert consensus on images or pathology. The document does not specify the exact ground truth method used in the predicate study.
8. Sample Size for the Training Set:
- Not applicable for this 510(k) (K130170). The clinical trials for the predicate device (K120087) would have involved a patient sample. The document does not provide this information.
9. How Ground Truth for the Training Set Was Established:
- Not applicable for this 510(k) (K130170). For the predicate device's clinical trials, the ground truth would have been established using a gold standard reference measurement system, as mentioned in point 7.
{0}------------------------------------------------
510(k) Summary ..
MAY 2 9 2013
: : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : :
í
| 510(k) Owner | Respiratory Motion, Inc. | ||||
|---|---|---|---|---|---|
| Address | 305 Second Avenue, Suite B | ||||
| Waltham, Massachusetts 02451 | |||||
| Phone | 781-373-1636 | ||||
| Fax | 781-373-1653 | ||||
| Contact person | Jenny Freeman, MD | ||||
| Date 510(k) | January 13, 2013 | ||||
| Summary | |||||
| prepared | |||||
| Trade name | ExSpiron™ 1xi | ||||
| Common name | Respiratory Monitor | ||||
| Classifications | Monitoring spirometer | ||||
| Product code: BZK | |||||
| Regulation: 21 CFR 868.1850 | |||||
| Breathing frequency monitor | |||||
| Product code BZQ | |||||
| Regulation: 21 CFR 868.2375 | |||||
| Predicate | ExSpiron Respiratory Monitor, cleared in 510(k) K120087. | ||||
| devices | |||||
| Device | The ExSpiron 1xi system consists of: | ||||
| Description | Bioimpedance measurement system: A stabilized high frequency currentgenerator is connected to two outer electrodes. The inner four electrodes areconnected to an adaptive circuit that conditions the resulting voltage signaland converts it to digital form. Firmware performs signal acquisition andrelays data to the panel PC.Computer: A Windows 7 PC performs signal processing and calibration,and runs the graphical user interface (GUI). The PC takes user input from atouch screen. The GUI is used for recording patient data and displaying therespiratory trace as well as scalar values and trends for minute volume, tidalvolume, and respiratory rate.Single Patient Use ExSpiron™ 1xi Electrode PadSet: An electrodeassembly containing six electrodes to be placed on the torso. It deliverscurrent and records impedance measurements. The electrode PadSet is alsoused to perform subsystem checks prior to patient measurements. | ||||
| Intended use | ExSpiron 1xi is indicated for use by healthcare professionals in healthcare facilities,such as post-operative care and critical care units, to monitor breathing in adult (atleast 21 years old) patients.ExSpiron 1xi is a non-invasive system that graphically displays lung volume againsttime and reports an approximate value of:Tidal volume,●Respiratory rate, andMinute ventilation� | ||||
| ExSpiron 1xi measurements are used as an adjunct to other clinical informationsources. |
·
.
:
.
.
·
{1}------------------------------------------------
| Comparison of | Characteristic | ExSpiron 1xi | ExSpiron | Comment | |||
|---|---|---|---|---|---|---|---|
| technological | Intended Use | See above | Same | The Indications for Use are not | |||
| characteristics | changed in this Special 510(k). | ||||||
| Technology | Measurement | Same | The fundamental technology is | ||||
| is by thoracic | not changed | ||||||
| bioimpedance. | |||||||
| Volume | Tidal volume | Same | No change | ||||
| Measurements | |||||||
| Minute | |||||||
| volume | |||||||
| Volume vs. | |||||||
| time chart | |||||||
| Rate | Respiratory | Same | No change | ||||
| Measurements | rate | ||||||
| (breaths/min) | |||||||
| Safety | IEC 60601-1, | Same | No change | ||||
| second edition | |||||||
| Nonclinical | Performance testing addressed specific modifications to the Monitor, Patient Cable, | ||||||
| performance | and Electrode PadSet as follows: | ||||||
| testing: | Software Verification & Validation | ||||||
| Safety Testing - IEC 60601-1 2nd Edition | |||||||
| Electromagnetic Compatibility - IEC 60601-1-2 | |||||||
| ExSpiron 1Xi System Test Plan | |||||||
| Patient Cable Design Verification - Included testing to all the relevant | |||||||
| sections of AAMI / ANSI EC53:1995/(R) 2008, ECG Cables and Leadwires | |||||||
| Electrode PadSet Design Verification - Included testing to all the relevant | |||||||
| section of AAMI / ANSI EC12:2000/(R)2010, Disposable ECG Electrodes | |||||||
| & AAMI / ANSI EC53:1995/(R) 2008, ECG Cables and Leadwires | |||||||
| Electrode - Biocompatibilty - ANSI/AAMI/ISO 10993-1 (Skin, prolonged | |||||||
| duration) | |||||||
| All performance testing conducted on modifications to the Monitor, Patient | |||||||
| Cable, and Electrode Padset were completed with successful | |||||||
| results. Performance testing raised no new safety or efficacy concerns and | |||||||
| demonstrated substantial equivalence to the predicate device. | |||||||
| Clinical | No clinical testing was performed for this 510(k). The measurement algorithm and | ||||||
| performance | software are unchanged from the predicate. Measurement performance statistics | ||||||
| testing: | were determined by clinical trials reported in the predicate 510(k), K120087. | ||||||
| Conclusions | Based on the comparisons of intended use and on the results of testing, the ExSpironI xi is substantially equivalent in intended use, safety, and effectiveness to the | ||||||
| regarding safety | ExSpiron respiratory monitor. | ||||||
| and | |||||||
| effectiveness: |
: . .
.
:
100000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000
. .
:
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services (HHS). The logo consists of a stylized caduceus, which is a symbol often associated with healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular fashion around the caduceus.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 29, 2013
Jenny Freeman, M.D. President Respiratory Motion 305 Second Avenue, Suite B WALTHAM, M.A., 02451
Re: K130170
Trade/Device Name: ExSpiron™ 1xi Regulation Number: 21 CFR 868.1850 Regulation Name: Monitoring Spirometer Regulatory Class: II Product Code: BZK, BZQ Dated: April 26, 2013 Received: April 29, 2013
Dear Ms. Freeman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Ms. Freeman
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part
807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Kwame O. Ulmer -S for
Anthony D. Watson, B.S., M.S., M.B.A. Director
Division of Anesthesiology, General Hospital, Respiratory, Infection Control and
Dental Devices
Office of Device Evaluation
Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Indications For Use
510(k) Number (if known): K 130170 Device Name: ExSpiron 1xi Indications for Use:
ExSpiron 1xi is indicated for use by healthcare professionals in healthcare facilities, such as postoperative care and critical care units, to monitor breathing in adult (at least 21 years of age) patients.
ExSpiron 1xi is a non-invasive system that graphically displays lung volume against time and reports an approximate value of:
- Tidal volume, .
- . Respiratory rate, and
- Minute ventilation. *
ExSpiron 1xi measurements are used as an adjunct to other clinical information sources.
Lester Wª 2013.05 29 10
(Division Sign-Off) Division of Anesthesiology, General Hospital Intection Control, Dental Devices
K130170 510 (k) Number:
Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (Part 21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
§ 868.1850 Monitoring spirometer.
(a)
Identification. A monitoring spirometer is a device used to measure continuously a patient's tidal volume (volume of gas inhaled by the patient during each respiration cycle) or minute volume (the tidal volume multiplied by the rate of respiration for 1 minute) for the evaluation of the patient's ventilatory status.(b)
Classification. Class II (performance standards).