(50 days)
The Trinity Acetabular System is intended for use in total hip arthroplasty in skeletally mature patients, to provide increased mobility and reduce pain by replacing the damaged hip joint articulation where there is evidence of sufficient sound bone to seat and support the components.
The indications for the Trinity Acetabular System as a total hip arthroplasty include:
- Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis
- Rheumatoid arthritis
- Correction of functional deformity
- Developmental dysplasia of the hip (DDH) or congenital dysplasia of the hip (CDH)
The Trinity Acetabular System is indicated for cementless, single use only.
The Trinity Acetabular System is a modular acetabular cup system consisting of a press fit, titanium alloy shell with, ultra high molecular weight polyethylene (UHMWPE), highly cross-linked polyethylene (HXLPE) and vitamin E highly cross-linked polyethylene (ECiMa) acetabular liners in neutral offset, +4mm offset, +4mm oblique, neutral 4mm EPW. The Trinity system also includes, BIOLOX delta™ ceramic and CoCr modular heads which are intended for use with Corin titanium femoral stems. The acetabular shell comes in two variants, a standard Trinity (K093472 and K110087) and a Trinity-i (K122305) shell, both coated with a rough titanium plasma spray with an additional top layer of electrochemically deposited calcium phosphate (BONIT™). The Trinity and Trinity-i acetabular shells are provided with screw holes and come assembled with titanium occluders in place. Dedicated titanium screws are also provided for additional fixation if required.
The purpose of this submission is to add two more acetabular shell variants to the Trinity Acetabular System, the standard Trinity No Hole acetabular shell and the Trinityi No Hole acetabular shell. Both of these variants, subject of this submission, are similar to the standard Trinity Acetabular shells previously cleared (K093472, K110087 and K122305) with the only difference being the shells subject of this submission do not have screw holes in them.
Here's an analysis of the provided text regarding the acceptance criteria and supporting study for the Corin Trinity Acetabular System (K130128):
1. Table of Acceptance Criteria and Reported Device Performance & 6. Standalone Performance:
No explicit acceptance criteria in terms of numerical performance metrics (e.g., sensitivity, specificity, accuracy, precision) or a standalone performance study of the device's diagnostic or therapeutic efficacy are presented in this 510(k) summary.
Instead, the acceptance criteria are met through demonstrating:
| Acceptance Criteria Type | Reported Device Performance |
|---|---|
| Similarity in: | The Trinity No Hole acetabular shells are identical to the predicate Trinity Acetabular Shell cleared in K093472 and K110087 in terms of: |
| Intended Use | Identical |
| Indications for Use | Identical |
| Substrate Materials | Identical |
| Size Range | Identical |
| Design | Similar (The only difference is the absence of screw holes) |
| Performance | Similar |
| Similarity in: | The Trinity-i No Hole acetabular shells are identical to the predicate Trinity-i Acetabular Shell cleared in K122305 in terms of: |
| Intended Use | Identical |
| Indications for Use | Identical |
| Substrate Materials | Identical |
| Size Range | Identical |
| Design | Similar (The only difference is the absence of screw holes) |
| Performance | Similar |
| Similarity to DePuy Pinnacle 100 Acetabular Shell (K001534): | The intended use, indications for use, substrate material, size range, and design of the Trinity No Hole shells are similar. |
The study underpinning this submission focuses on non-clinical testing to demonstrate substantial equivalence to predicate devices, rather than a standalone clinical performance study verifying new performance metrics for the modified device.
2. Sample Size Used for the Test Set and Data Provenance:
No specific "test set" in the context of clinical trials or data analysis is mentioned because this submission relies on demonstrating substantial equivalence to previously cleared devices. The testing conducted was "non-clinical testing" including comparisons of designs, materials, intended use, indications, and dimensions. Therefore, there is no sample size for a test set or data provenance from a patient population.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications:
Not applicable. As described above, there was no clinical test set or ground truth established by experts in the context of diagnosing or categorizing patient data.
4. Adjudication Method for the Test Set:
Not applicable. There was no clinical test set or adjudication process for diagnostic outcomes.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:
No. An MRMC study was not performed. The submission relies on establishing substantial equivalence through non-clinical testing and comparison to predicate devices, not on proving improved human reader performance with AI assistance. The device in question is a physical medical device (hip prosthesis components), not an AI-powered diagnostic tool.
7. Type of Ground Truth Used:
For the purpose of this 510(k) submission, the "ground truth" is effectively the established safety and effectiveness of the predicate devices (Corin Trinity Acetabular System K093472, K110087, K122305, and DePuy Pinnacle 100 Acetabular Shell K001534). The current device's "no hole" variants are considered substantially equivalent if they match the predicates in key aspects.
8. Sample Size for the Training Set:
Not applicable. This device is a modified physical implant, not an algorithm that requires a training set of data.
9. How the Ground Truth for the Training Set Was Established:
Not applicable. As stated above, there is no training set for an algorithm.
{0}------------------------------------------------
K130128 (pg 1/3)
MAR 0 8 2013 510(K) SUMMARY 3. 1. Applicant/Sponsor: Corin USA 5670 W. Cypress St. Suite C Tampa, Florida 33607 Establishment Registration No. : 1056629 2. Contact Person: Lucinda Gerber Regulatory Affairs Associate Corin USA 813-977-4469 Lucinda.gerber@coringroup.com 3. Date: January 16, 2013 4. Proprietary Name: Trinity No Hole Acetabular Shells
-
Common Name: Hip Prosthesis
-
Product Codes: LZO, MEH
7. Classification Name:
Hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis (21CFR 888.3353)
Legally Marketed Devices to which Substantial Equivalence is claimed: 8.
- Corin Trinity Acetabular System (K093472) .
- Corin Trinity Acetabular System with HXLPE (K110087) .
- Corin Trinity Acetabular System with Trinity-i Acetabular Shells (K122305) .
- DePuy Pinnacle 100 Acetabular Shell (K001534) .
9. Device Description:
The Trinity Acetabular System is a modular acetabular cup system consisting of a press fit, titanium alloy shell with, ultra high molecular weight polyethylene (UHMWPE), highly cross-linked polyethylene (HXLPE) and vitamin E highly cross-linked polyethylene (ECiMa) acetabular liners in neutral offset, +4mm offset, +4mm oblique,
{1}------------------------------------------------
neutral 4mm EPW. The Trinity system also includes, BIOLOX delta™ ceramic and CoCr modular heads which are intended for use with Corin titanium femoral stems. The acetabular shell comes in two variants, a standard Trinity (K093472 and K110087) and a Trinity-i (K122305) shell, both coated with a rough titanium plasma spray with an additional top layer of electrochemically deposited calcium phosphate (BONIT™). The Trinity and Trinity-i acetabular shells are provided with screw holes and come assembled with titanium occluders in place. Dedicated titanium screws are also provided for additional fixation if required.
The purpose of this submission is to add two more acetabular shell variants to the Trinity Acetabular System, the standard Trinity No Hole acetabular shell and the Trinityi No Hole acetabular shell. Both of these variants, subject of this submission, are similar to the standard Trinity Acetabular shells previously cleared (K093472, K110087 and K122305) with the only difference being the shells subject of this submission do not have screw holes in them.
10. Intended Use / Indications:
The Trinity Acetabular System is intended for use in total hip arthroplasty in skeletally mature patients, to provide increased mobility and reduce pain by replacing the damaged hip joint articulation where there is evidence of sufficient sound bone to seat and support the components.
The indications for the Trinity Acetabular System as a total hip arthroplasty include:
- Non-inflammatory degenerative joint disease including osteoarthritis and avascular . necrosis
- Rheumatoid arthritis .
- Correction of functional deformity ●
- . Developmental dysplasia of the hip (DDH) or congenital dysplasia of the hip (CDH)
The Trinity Acetabular System is indicated for cementless, single use only.
11. Summary of Technologies/Substantial Equivalence:
The Trinity No Hole acetabular shells are identical to the standard Trinity Acetabular Shell cleared in K093472 and K110087 in terms of intended use, indications for use, substrate materials, and size range and are similar in design and performance. Likewise, the Trinity-i No Hole acetabular shells are identical to the Trinity-i Acetabular Shell cleared in K122305
{2}------------------------------------------------
K130128 (pg 3/3)
in terms of intended use, indications for use, substrate materials, and size range and are similar in design and performance. The intended use, indications for use, substrate material, size range and design of the Trinity No Hole shells are similar to the DePuy Pinnacle 100 acetabular shell (K001534). Based on these similarities, the additional variants subject of this submission, are determined to be substantially equivalent to the predicate devices.
12. Non-Clinical Testing:
Non-clinical testing conducted to demonstrate substantial equivalence includes a comparison of the designs, materials, intended use, indications and dimensions of the additional Trinity Acetabular shells to the predicate devices. The results of this testing show that the Trinity and Trinity-i No Hole variants are expected to be safe and effective for the proposed indications and are substantially equivalent to the predicate devices.
13. Clinical Testing:
Clinical testing was not necessary to determine substantial equivalence between the Trinity Acetabular System and the predicate devices.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circle around the eagle. The logo is black and white.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 8, 2013
Corin, Medical % Ms. Lucinda Gerber Regulatory Affairs Associate 5670 West Cypress, Suite C Tampa, Florida 33607
Re: K130128
Trade/Device Name: Trinity Acetabular System Regulation Number: 21 CFR 888.3353 Regulation Name: Hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis Regulatory Class: II Product Code: LZO, MEH Dated: February 06, 2013 Received: February 11, 2013
Dear Ms. Gerber:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{4}------------------------------------------------
Page 2 - Ms. Lucinda Gerber
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Image /page/4/Picture/6 description: The image shows the name "Erin Keith" in a stylized font. The letters "F.D." are embedded within the name, with the "F" appearing within the "i" of "Erin" and the "D" appearing within the "K" of "Keith." The font is bold and black, and the letters are slightly distorted.
Mark N. Melkerson Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
2. INDICATIONS FOR USE
510(k) Number (if known): __ K130128_(pg 1/1)
Device Name: Trinity Acetabular System
Indications for Use:
The indications for the Trinity Acetabular System as a total hip arthroplasty include:
- Non-inflammatory degenerative joint disease including osteoarthritis and avascular . necrosis
- Rheumatoid arthritis
- Correction of functional deformity .
- Developmental dysplasia of the hip (DDH) or congenital dysplasia of the hip (CDH) ●
The Trinity Acetabular System is indicated for cementless use only.
Prescription Use _ X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use _ (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Elizabetព្យ័ត្រ រួចជាច្រើនrank -S
Page _ 1 of _1
Division of Orthopedic Devices
§ 888.3353 Hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis.
(a)
Identification. A hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis is a device intended to be implanted to replace a hip joint. This device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. The two-part femoral component consists of a femoral stem made of alloys to be fixed in the intramedullary canal of the femur by impaction with or without use of bone cement. The proximal end of the femoral stem is tapered with a surface that ensures positive locking with the spherical ceramic (aluminium oxide, A12 03 ) head of the femoral component. The acetabular component is made of ultra-high molecular weight polyethylene or ultra-high molecular weight polyethylene reinforced with nonporous metal alloys, and used with or without bone cement.(b)
Classification. Class II.