K Number
K130065
Date Cleared
2014-04-30

(475 days)

Product Code
Regulation Number
882.5890
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Beautiful Image Model 900 Facial Toning Device uses microcurrent electrical energy to stimulate facial tissues for aestheric purposes.
Biosonic Technologies Model 900 Facial Toning Device is intended for facial stimulation and is indicated for prescription cosmetic use. The anatomical site for application of the Model 900 is the face.

Device Description

Beautiful Image Model 900 Facial Toning Device is intended for facial stimulation and is indicated for prescription cosmetic use. The anatomical site for application of the Model 900 is the face.
To enhance safety, the device is operated only from battery voltage and will not operate while connected to line power. A proper recharge should be performed prior to using the machine.
The Model 900 is microprocessor controlled and housed in a white plastic case.

AI/ML Overview

Here's an analysis of the provided text regarding the acceptance criteria and study for the Beautiful Image Model 900 Facial Toning Device:

The provided document describes a 510(k) premarket notification for a medical device. In this type of submission, the primary goal is to demonstrate substantial equivalence to a legally marketed predicate device, rather than proving efficacy through a de novo clinical study with specific acceptance criteria in the traditional sense. Therefore, the information you're asking for, particularly the detailed breakdown of acceptance criteria and a study proving the device meets those criteria in the way a clinical trial would, is largely not present in this type of regulatory document.

Instead, the submission focuses on comparing the subject device's technical specifications and safety features to a predicate device and conducting non-clinical tests to ensure safety and effectiveness.

Here's an attempt to answer your questions based on the provided text, highlighting what's available and what's not:


1. A table of acceptance criteria and the reported device performance

As mentioned, this is a 510(k) submission, not a de novo clinical trial report. Therefore, there isn't a table of explicit "acceptance criteria" for a primary efficacy endpoint and corresponding "reported device performance" in terms of clinical outcomes.

The comparison is made against the predicate device's specifications and safety profile to demonstrate substantial equivalence. The "acceptance criteria" here are implicitly that the subject device's characteristics do not raise new questions of safety or effectiveness compared to the predicate.

Here's a table summarizing the comparison of parameters between the subject device and the predicate, which serves as the basis for demonstrating substantial equivalence:

ParameterSubject Device (Model 900)Predicate Device (NuFace Original)Compliance/Difference
Intended UseFacial stimulation for aesthetic purposes (prescription cosmetic use)Facial stimulation for aesthetic purposesSubstantially equivalent
Power SourceOne 6V batteryOne 9V batteryDifference: Lower voltage battery (6V vs 9V). Justification: Lower voltage is a lower risk factor.
Software/Firmware/Microprocessor Control?YesNoDifference: Subject device has microprocessor control. Justification: Adds safety features.
Automatic Overload Trip?YesNoDifference: Subject device has automatic overload trip. Justification: Adds safety features.
Automatic No-Load Trip?YesNoDifference: Subject device has automatic no-load trip. Justification: Adds safety features.
Automatic Shut Off?YesNoDifference: Subject device has automatic shut off. Justification: Adds safety features.
Low Battery Indicator Display Status?YesNoDifference: Subject device has low battery indicator. Justification: Adds safety features.
Voltage/Current Level Display Status?YesNoDifference: Subject device has voltage/current level indicator. Justification: Adds safety features.
WaveformBiphasicPulsed MonophasicDifference: Subject device is Biphasic vs Pulsed Monophasic. Justification: Not explicitly justified as a safety/effectiveness difference in the text, but the overall output parameters are compared.
Max Output Voltage (@500Ω)0.347 V0.16 VDifference: Subject device has higher maximum output voltage. Justification: Implicitly acceptable as overall safety is maintained.
Max Output Current (@500Ω)0.647 mA0.324 mADifference: Subject device has higher maximum output current. Justification: Implicitly acceptable as overall safety is maintained.
Max Current Density (@500Ω)1.486 mA/cm² r.m.s.0.341 mA/cm² r.m.s.Difference: Subject device has higher current density. Justification: Attributed to "more precise surface contact area," and stated to be "as safe as the predicate, and more effective at the point of contact than the predicate."
Max Average Power Density (@500Ω)366E-6 W/cm²3.02 µW/cm²Difference: Subject device has higher average power density. Justification: Attributed to "more precise surface contact area," and stated to be "as safe as the predicate, and more effective at the point of contact than the predicate." It is also explicitly stated that this remains "less than 0.25 watts per square centimeter of electrode conductive surface area to reduce the risk of thermal burns."
Weight10 lbs0.5 lbsDifference: Subject device is much heavier. Justification: "Rrobust Model 900, which will not as easily move about and is designed in this fashion in direct response to user design inputs."
Dimensions5.5x15.3x11.3 in.2.25x7x0.75 in.Difference: Subject device is larger. Justification: "Robust Model 900, which will not as easily move about and is designed in this fashion in direct response to user design inputs," and "designed to support the handheld operation of the contacts from a desktop device by means of cables."
Intended UserLicensed practitionersPublicDifference: Subject device intended for licensed practitioners. Justification: No specific justification about this difference related to safety/effectiveness for the device itself, but implies different use environment and user expertise.
Compliance with Voluntary StandardsIEC 60601-1, IEC 60601-1-2, ISO 10993, 21 CFR 898IEC 60601-1, 21 CFR 898Subject device passed additional/more recent versions of standards (IEC 60601-1-2 and ISO 10993 explicitly mentioned for subject device, not for predicate) and compliance to stated standards for both devices is listed. This effectively acts as an "acceptance criterion" for safety.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Sample Size for Test Set: Not applicable. This is a non-clinical submission. The "test set" in this context refers to the device itself and its components undergoing engineering and electrical safety tests, not a set of subjects for clinical evaluation.
  • Data Provenance: Not applicable for clinical data. The non-clinical test data would be generated in a lab setting by the manufacturer or contracted testing facilities (e.g., for IEC/ISO standards).

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

  • Number of Experts: Not applicable. This is not a study requiring expert-established ground truth for a test set in a clinical context.
  • Qualifications of Experts: The testing would be conducted by engineers and technicians qualified in electrical safety, electromagnetic compatibility, and biocompatibility testing according to the relevant international standards (IEC, ISO).

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • Adjudication Method: Not applicable. This refers to consensus among human readers/experts, which is not relevant for the non-clinical testing performed.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • MRMC Study: No. This is not a clinical study, and the device is not an AI-driven diagnostic tool. It's a physical device for facial stimulation.
  • Effect Size: Not applicable.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Standalone Performance: Not applicable. This is a physical device, not an algorithm. The device's performance is inherently "standalone" in its operation as a facial toning device, but not in the context of an AI algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

  • Type of Ground Truth: For this 510(k) submission, the "ground truth" for the device's acceptable performance is its compliance with specific engineering and safety standards (IEC 60601-1, IEC 60601-1-2, ISO 10993, 21 CFR 898) and its electrical output specifications meeting predefined ranges that are deemed safe and effective based on the predicate device. In essence, the benchmark is regulatory compliance and substantial equivalence to a device already deemed safe and effective.

8. The sample size for the training set

  • Sample Size for Training Set: Not applicable. This device does not involve a "training set" in the machine learning sense.

9. How the ground truth for the training set was established

  • Ground Truth for Training Set: Not applicable.

Summary of the study conducted (as described in the 510(k) submission):

The "study" in this context refers to non-clinical testing to demonstrate the safety and effectiveness of the Beautiful Image Model 900 Facial Toning Device and its substantial equivalence to the NūFace Facial Toning Device (predicate).

  • Tests Performed:
    • Product Safety testing conforming to IEC 60601-1.
    • Electromagnetic compatibility (EMC) testing conforming to IEC 60601-1-2.
    • Biocompatibility testing conforming to ISO 10993.
    • Comparison of electrical and physical parameters against the predicate device.
  • Purpose: To confirm that the Model 900 is safe, effective, and substantially equivalent to the predicate device, and that the differences between the two do not adversely affect safety and effectiveness.
  • Conclusion: The tests confirmed that the Model 900 meets the safety standards and its performance characteristics (including electrical outputs and maximum current/power density, while having different physical attributes and added safety features) did not raise new issues of safety or effectiveness compared to the predicate. The differences were justified as either enhancing safety (e.g., microprocessor control, automatic trips, indicators) or being a result of user design inputs (e.g., larger size, desktop operation by licensed practitioners) without compromising safety or effectiveness. The increased current and power density were explained by a more precise surface contact area, remaining within safe limits.

{0}------------------------------------------------

Section 005 - Traditional 510(k) Summary

As required in 21 CFR807.92, we hereby submit this Traditional 510(k) Summary.

Traditional 510(k) owners name, address, phone, fax, contact person & preparation date

The Traditional 510(k) owner is Biosonic Technologies, LLC.

  • Address: 333 Las Olas Way, Suite 1906
    Fort Lauderdale, FL 33311

Phone: (954) 300 - 2511

(866) 830 - 8759 Fax:

The contact person is Mr. Ray Baker, President, Biosonic Technologies, LLC.

Traditional 510(k) preparation date: 19 August, 2011

SUBJECT DEVICE:

Name of the device, trade name, proprietary name, and classification name:

Trade name: Beautiful Image Model 900 Facial Toning Device

Common name: Facial Toning Device

Classification: Stimulator, Transcutaneous Electrical, Aesthetic Purposes.

CFR Reference: 21 CFR 882.5890

Product Code: NFO

Device Description:

Beautiful Image Model 900 Facial Toning Device is intended for facial stimulation and is indicated for prescription cosmetic use. The anatomical site for application of the Model 900 is the face.

{1}------------------------------------------------

To enhance safety, the device is operated only from battery voltage and will not operate while connected to line power. A proper recharge should be performed prior to using the machine.

The Model 900 is microprocessor controlled and housed in a white plastic case.

Dimensions & Appearance:
Height (rear)5.5 inches (13.97 cm)
Height (front)3.3 inches (8.38 cm)
Width15.3 inches (38.86 cm)
Depth11.3 inches (28.7 cm)
Weight10.0 pounds (4.5 kgs)
ColorWhite case with blue front panel
Electrical Specifications
Device powerModel 900 will operate only on battery power; it willnot operate on line voltage
Device Output800µA maximum, from constant current source
Complex wave output frequency3 Hz to 300 Hz
Maximum voltage30Vpp on any output with respect to common; 5Vpptypical
Battery Charger9vdc @ 1 amp. Use of the charger disables thepatient circuits
Battery6 volt, 7 amp hours
Power Requirements CAC
Adapter
Input Voltage100 to 120 VAC, 60 Hz
Input Current0.15 amps
ConfigurationStandard U.S. 2-Prong Plug
Device output Voltage9v DC
Output Current1 amp
Output PlugDe coas: 5.5mm OD, 2.1mm pin receptacle
Cable length6 ft (240 cm)
Environmental Specifications
Operating Temperature+ 50 to 104° F (+10 to +40°C)
Storage Temperature- 29 to + 167° F (-34 to +76° C)
Humidity0 to 95% - non-condensing

Specifications:

5-2

{2}------------------------------------------------

FUNCTIONAL DESCRIPTION

OPERATING CYCLE:

The Model 900 operating cycle is initiated by momentarily pressing the POWER button located in the POWER section of the front panel. All indicator lights (LEDs) on the panel will illuminate for a short period of time except for the CHARGING and the BATTERY LOW or BATTERY FULL indicators (depending on the charge level of the battery).

The INTERVAL PACE lights (located in the WAVEFORM section of the panel) will step from left to right providing a visual indication of the approximate delay time remaining prior to the application of energy to the probes. The INTERVAL DELAY adjustment sets the amount of time for this delay.

The SEQUENCING lights (also located in the WAVEFORM section of the panel) indicate the application of each of the different waveform outputs from the equipment, seven in total. The SEQUENCE DURATION adjustment sets the amount of time for the total sequenced set.

OPERATING RANGES:

Adjustment of the output energy intensity is set by the INTENSITY ADJUSTMENT button located in the PROCEDURE section of the panel. Each press of the button will advance the output energy from NORMAL, to MODERATE, to HIGH, and then the sequence will repeat.

The INTERVAL DELAY provides an adjustment range of approximately 1 to 6 seconds.

The SEQUENCE DURATION provides an adjustment range of approximately 10 to 30 seconds.

Waveform frequencies: automatically sequencing from 3 Hz to 300 Hz

OPERATING MODE:

An indicator shows that the wand set is in use. The indicator will be illuminated when the wand is connected to the machine.

CALIBRATION:

The Model 900 does not require calibration.

{3}------------------------------------------------

INDICATION FOR USE:

The Beautiful Image Model 900 Facial Toning Device uses microcurrent electrical energy to stimulate facial tissues for aesthetic purposes. INTENDED USE:

Biosonic Technologies Model 900 Facial Toning Device is intended for facial stimulation and is indicated for prescription cosmetic use. The anatomical site for application of the Model 900 is the face.

COMPARISON WITH THE PREDICATE DEVICE:

The predicate device for this Traditional 510(k) is the NūFace Facial Toning Device. The subject device is Biosonic Technologies Beautiful Image Model 900 Facial Toning Device, that is substantially equivalent to the predicate device, differing only in that the unit is larger and designed to support the handheld operation of the contacts from a desktop device by means of cables, and is intended for the use of licensed practitioners rather than the public.

COMPARISON OF THE SUBJECT DEVICE AND PREDICATE DEVICE PARAMETERS:

ParameterYour DevicePredicateDevice
510(k) NumberK130065K072260
Device Name and ModelModel 900NuFaceOriginal
ManufacturerBiosonicTechnologies,LLCCarol ColeCompany
Power Source(s) †One 6V batteryOne 9V battery
- Method of Line Current IsolationN/AN/A
- Patient Leakage Current††N/AN/A
- Normal Condition (μΑ)N/AN/A
- Single Fault Condition (μΑ)N/AN/A
Average DC current through electrodes whendevice is on but no pulses are being applied (µA)NoneNone
Number of Output Modes †††OneOne
Number of Output Channels††††OneOne
Synchronous or Alternating?N/AN/A

{4}------------------------------------------------

ParameterYour DevicePredicate Device
Method of Channel IsolationN/AN/A
Regulated Current or Regulated Voltage?BothBoth
Software/Firmware/Microprocessor Control?YesNo
Automatic Overload Trip?YesNo
Automatic No-Load Trip?YesNo
Automatic Shut Off?YesNo
User Override Control?YesYes
Indicator Display Status
On/Off Display Status?YesYes
Low Battery?YesNo
Voltage/Current Level?YesNo
Timer Range (minutes)NoneNone
Compliance with Voluntary Standards?IEC 60601-1IEC 60601-1
Compliance with 21 CFR 898?YesYes
Weight (lbs., oz.)10lbs0.5lbs
Dimensions (in.) [W x H x D]5.5x15.3x11.32.25x7x0.75
Housing Materials and ConstructionThermoplasticThermoplastic

† For AC line-powered devices, we recommend that you specify the line voltage and frequency, the method of line current isolation, and the measured patient leakage current. For battery-powered devices, we recommend that you specify the number, size and type of batteries.

++ We recommend that you follow IEC 60601-1 "Medical Electrical Equipment - Part 1: General Requirements for Safety" or an equivalent method to show that the levels of patient leakage current, measured under both normal and single fault conditions, are acceptable.

ttt For devices with more than one output mode, we recommend that you provide the information in Output Waveforms (see Section 7A above) and Output Specifications (see Section 7C below) for each output mode.

tttt For devices with more than one output channel, we recommend that you describe whether the outputs are delivered in a synchronous and/or alternating fashion and the method of achieving channel isolation.

.Parameter- March Made de문.------------------------------------------------------------------------------------------------------------------------------------------------------------------------------.JevicePredicate
Device
. Bof the first and a program of the company and any come and a
ParameterYour DevicePredicateDevice
Mode or Program NameBeautifulImageNuFace
Waveform (e.g., pulsed monophasic, biphasic)BiphasicPulsedMonophasic
Shape (e.g., rectangular, spike, rectified sinusoidal)RectangularRectangular
Maximum Output Voltage (volts) (+/- 5%)0.347 @500Ω0.16 @500Ω
1.242 @ 2 kΩ0.78 @ 2 kΩ
5.780 @10 kΩ2.6 @10 kΩ
Maximum Output Current (mA) (+/- 5%)0.647 @500Ω0.324 @500Ω
0.625 @ 2 kΩ0.358 @ 2 kΩ
0.584 @10 kΩ0.263 @10 kΩ
Duration of primary (depolarizing) phase† (msec) SeeNote 30.648 - 322---
Pulse Duration† (msec)3.24 - 1610----
Frequency† (Hz) [or Rate† (pps)]0.621 - 308.6----
For multiphasicSymmetrical phases?YesYes / No
waveforms only:Phase Duration† (msec),(state range, if applicable),(both phases, if asymmetrical0.324-161pulsed
Net Charge (micro coulombs (µC) per pulse) (If zero,state method of achieving zero net charge.)0μC @500ΩN/A @500Ω
Maximum Phase Charge, (µC)190 @500Ω18.13 @500Ω
Maximum Current Density, †† (mA/cm², r.m.s.)1.486 @500Ω0.341 @500Ω
Maximum Average Current (average absolute value),mA0.493 @500Ω@500Ω
Maximum Average Power Density, †† (W/cm²), (usingsmallest electrode conductive surface area)366E-6 @500Ω3.02μW@500Ω
Burst Mode(a) Pulses per burstN/A2.1
(i.e., pulse trains):(b) Bursts per secondN/A9.1
(c) Burst duration (seconds)N/A2.3
(d) Duty Cycle: Line (b) x Line(c)N/A20.9
ON Time (seconds)10-30Constant
OFF Time (seconds)1-6None
Additional Features (specify, if applicable)NoneNone

{5}------------------------------------------------

† For continuously variable parameters, we recommend that you specify the full range; for parameters with discrete settings, we recommend that you specify all available selections.

tt We recommend that you calculate the maximum current density and maximum average power density values by using the conductive surface area of the smallest electrodes intended for use with the unit, and include sample calculations in your 510(k). We also recommend that you calculate the maximum power density by using the maximum duty cycle and by averaging over output duration of one second. The maximum average power density should be less than 0.25 watts per square centimeter of electrode conductive surface area to reduce the risk of thermal burns.

{6}------------------------------------------------

DISCUSSION OF WHY THE DIFFERENCES BETWEEN THE SUBJECT DEVICE AND THE PREDICATE DEVICE DO NOT ADVERSELY AFFECT SAFETY AND EFFECTIVENESS:

The predicate device has a 9v battery, while the subject device has a 6v battery. Generally, a lower voltage battery represents a lower risk factor, and this advantage counts in favor of the Model 900 while providing the device ample power to perform in its intended use. Furthermore, the predicate device does not have any software/firmware/microprocessor control, automatic overload trip, automatic no-load trip, automatic shutoff, low battery indicator display status, nor voltage/current level indicator display status. Each of these safety features exists within the subject device, and therefore provides a safer device than the predicate device. Additionally, the predicate device is much smaller and lighter than the robust Model 900, which will not as easily move about and is designed in this fashion in direct response to user design inputs. Regarding the current, the maximum current density of the Model 900 measures 1.486 mA/cm2 r.m.s. at 500 Ohms of resistance, versus the reported 0.341 mA/cm2 r.m.s. at 500 Ohms resistance of the NūFace, and although this appears to represent a difference within the overall current, it reflects simply a more precise surface contact area that applies the microcurrent as compared to the predicate. The overall current delivered from the Model 900 is unchanged. Finally, the maximum average power density is 366E-6 W at 500 Ohms versus 3.02 µW at 500 Ohms, which represents a larger energy output than the predicate, given the more precise surface contact area, while remaining as safe as the predicate, and more effective at the point of contact than the predicate. Thus, given the foregoing, the Model 900 is at least as safe and as effective as the predicate NūFace device.

All other parameters including the intended use and safety profile remain unchanged.

NONCLINICAL TEST DATA:

Product Safety testing and electromagnetic compatibility testing conforming to IEC 60601-1 and IEC 60601-1-2, and biocompatibility testing conforming to ISO 10993 were performed in support of the overall safety and effectiveness characteristics of the device, and have confirmed that the Model 900 is safe, effective and substantially equivalent regarding the outputs of the predicate device.

SUMMARY:

No new issues of safety or effectiveness are raised regarding the Model 900 Facial Toning Device.

End of Traditional 510(k) Summary Section

{7}------------------------------------------------

Image /page/7/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or a stylized human figure with outstretched arms.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

April 30, 2014

Biosonic Technologies, LLC c/o Shepard G. Bentley Bentley Biomedical Consulting, LLC 28241 Crown Valley Parkway, Suite 510(K) Laguna Niguel, CA 92677

Re: K130065

Trade/Device Name: Beautiful Image Model 900 Facial Toning Device Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous electrical nerve stimulator for pain relief Regulatory Class: Class II Product Code: NFO Dated: March 30, 2014 Received: April 1, 2014

Dear Mr. Bentley:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it your do rise to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean I that FDA has made a determination that your device complies with other requirements of the Act that I Dri has made a dolly and regulations administered by other Federal agencies. You must of any I vith all the Act's requirements, including, but not limited to: registration and listing (21

{8}------------------------------------------------

Page 2 - Mr. Shepard G. Bentley

CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Felipe Aquel -S

Carlos L. Peña, PhD, MS for Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{9}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K130065

Device Name

Beautiful Image Model 900 Facial Toning Device

Indications for Use (Describe)

The Beautiful Image Model 900 Facial Toning Device uses microcurrent electrical energy to stimulate facial tissues for aestheric purposes.

Type of Use (Select one or both, as applicable)

[ Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

Form Approved: OMB No. 0910-0120

Expiration Date: January 31, 2017

See PRA Statement on last page.

PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

.

ം Date: 2014.04.30

-0 16:54:05 -04'00'

{10}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

§ 882.5890 Transcutaneous electrical nerve stimulator for pain relief.

(a)
Identification. A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.(b)
Classification. Class II (performance standards).