K Number
K130056
Device Name
PATIENT MONITOR
Manufacturer
Date Cleared
2013-04-11

(92 days)

Product Code
Regulation Number
870.2300
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The TM-Oxi device indicated for use in non-invasively measuring and displaying functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate (PR), Non-invasive measurement of blood pressure (NIBP) of adult patients. The TM-Oxi software uploads the data of the device and displays the data into a computer for enhanced data management. The TM-Oxi software is intended for use in clinical settings as an aid for health care professionals to review, analyze and evaluate the historical tests results. TheTM-Oxi device is intended for spot-checking of patients. The TM-Oxi system does not report any diagnosis but provides values. It is the physician's responsibility to make proper judgments based on these numbers. Prescription Use: Federal law restricts this device to sale by or on the order of a physician.

Device Description

The TM-Oxi is a programmable electro medical system using one oximeter and one noninvasive blood pressure device (NIBP). The system comprises: a. Hardware: Non- invasive blood pressure, pump and valve boards and connections, tube, bladder and cuff 510k cleared (K093013) not modified. USB HUB DC converter with 3 features. Power supply of NIBP board via USB hub connected to USB port of the PC and microchip converter 5V to 6 V. microchip converter 510k cleared (K090671) not modified via USB hub > connected to USB port of the PC. connected to USB port of the connected to USB port of the PC. > b. Software: The used software are ESO software 510k K102442 for data analysis and ES Complex software 510k number k 113264 for data management. software 310k humber K 115204 for addinations of the data from oximeter (ESO) and blood pressure device (Contec 08A K110774). pressure device (Contect 687 ATT 10777). Therefore the TM-Oxi system will be a combination of the 2 separate devices.

AI/ML Overview

The provided text describes the TM-Oxi device, its intended use, and its equivalence to legally marketed predicate devices, but it does not contain a detailed study report that specifically outlines acceptance criteria and subsequently proves the device meets them with specific performance metrics from a study.

Instead, the document focuses on demonstrating substantial equivalence to predicate devices by comparing their technical specifications and by stating that the TM-Oxi uses cleared components and software from previously cleared devices.

Here's an analysis of the information provided, addressing your points where possible, and highlighting what is not available in the text:

1. A table of acceptance criteria and the reported device performance

The document provides a comparative table (page 4) between the TM-Oxi hardware and a predicate device (MD2000B Vital sign monitor K100740), which lists technical specifications. These can be inferred as performance criteria that the TM-Oxi is designed to meet or be equivalent to.

Acceptance Criteria (from Predicate Device/OEM Specifications)Reported Device Performance (TM-Oxi Hardware)
Intended UseThe TM-Oxi device indicated for use in non-invasively measuring and displaying functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate (PR), Non-invasive measurement of blood pressure (NIBP) of adult patients in medical facilities. The TM-Oxi device is intended for spot-checking of patients.
SpO2 %YES
Oximeter Wave form and Pulse rateYES
Systolic pressureYES
Diastolic pressureYES
Output Current100 mA
Output Power25VA
Voltage5V DC
Oximeter Technical Specifications
SpO2% Measuring range0 to 100%
SpO2% Accuracy70%~100%: ±2%, Below 70% unspecified
Pulse rate measuring range30 to 250bpm
Pulse rate accuracy± 2 bpm during the pulse rate range of 3099 bpm and 2% during the pulse rate range of 100250 bpm
Wavelength Red660 nm
Wavelength IR880 nm
Radiant Power Red1.8 mW
Radiant Power IR2 mW
BP Technical Specifications
Measuring principleOscillometric
Measurement modeControlled by software
Measuring range Systolic30-255 mmHg
Measuring range Diastolic15-220 mmHg
Other specifications
Target populationAdult
Application siteFinger
DisplayLCD (oximeter) and PC screen
Power supplyUSB
Safety ClassBF
Data transmissionUSB
Testing bench60601-1-1, 60601-1-2

Note regarding "Reported Device Performance": The document states that "The TM-Oxi hardware has same intended use and same performances and effectiveness" as the predicate. It then uses the technical specifications of the predicate and OEM components as a proxy for the TM-Oxi's performance. The detailed study proving these specific metrics for the TM-Oxi as a system is not presented in this summary.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

This information is not provided in the document. The document primarily relies on the fact that existing 510(k) cleared components were used. It refers to ISO 9919 reports from the OEM oximeter and ANSI/AAMI SP10:2002/A1:2003/A2: 2006 reports from the OEM blood pressure device. These reports would contain the sample size and data provenance for the original component clearance, but these details are not included in this K130056 summary.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This information is not provided. The document states that the TM-Oxi system "does not report any diagnosis but provides values. It is the physician's responsibility to make proper judgments based on these numbers." The clinical assessment and ground truth establishment would likely have been part of the original OEM component clearances, but no details are given for the TM-Oxi system specifically.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This information is not provided.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This information is not provided. The TM-Oxi is a physiological patient monitor providing objective measurements (SpO2, PR, NIBP) and data management, and not an AI-assisted diagnostic tool that would typically involve human readers interpreting results for improved diagnostic effectiveness.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

The document mentions "Integration tests for oximeter and blood pressure device" and "Software structural testing using CRC (Cyclic redundancy check) Coding was performing to accurately capture, store, and analyze the data measured by the hardware." (Page 5).
It also states: "The TM-Oxi software uploads the data of the device and displays in clinical settings as an enhanced data management. The TM-Oxi software is intended for use in clinical settings as an aid for health care professionals to review, and evaluate the historical tests results." (Page 2).
This suggests the software performs data processing and display independently. However, a formal "standalone performance study" with specific metrics (e.g., accuracy against a gold standard) for the integrated system's measurements is not detailed. The reliance is heavily on the pre-cleared status of the hardware components which have their own performance data.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

This information is not provided for the TM-Oxi system specifically. For the original oximeter and NIBP components, the ground truth would typically be established against highly accurate reference devices (e.g., co-oximetry for SpO2, invasive blood pressure for NIBP) during their respective ISO 9919 and ANSI/AAMI SP10 validations.

8. The sample size for the training set

This information is not provided. The software "ESO software 510k K102442 for data analysis and ES Complex software 510k number k 113264 for data management" are pre-cleared components. Details on their training data (if any machine learning was involved, which is unlikely for basic data analysis/management software) are not included.

9. How the ground truth for the training set was established

This information is not provided.


Summary of the "Study" (or rather, justfication for substantial equivalence):

The document argues for substantial equivalence by stating that the TM-Oxi device is a combination of previously 510(k) cleared hardware components and software.

  • Hardware: It utilizes an oximeter and a non-invasive blood pressure device (NIBP) where the PCB boards, pump, valve, tube, bladder, cuff, and oximeter are from OEM subcontractors and were "cleared for use together" or are themselves 510(k) cleared devices (e.g., MD 200A K093013, CMS50E K090671).
  • Software: It uses pre-cleared software components: ESO Software (K102442) for data analysis and ES Complex Software (K113264) for data management, which are described as using the "same source codes" as the TM-Oxi software for data of the oximeter and blood pressure device.
  • Testing:
    • Integration tests for the oximeter and blood pressure device were performed.
    • Software structural testing (CRC) was done to ensure accurate data capture, storage, and analysis.
    • Laboratory testing for electrical safety (IEC 60601-1-1) and EMC (IEC 60601-1-2) was conducted for the components.
    • Software verification (SRS/SDS/STD/STR) was performed.
  • A Clinical Test was deemed NOT required: The document explicitly states: "Since the TM-Oxi is using 510K cleared devices clinically tested (ISO 9919 and SP10) and with regard of the testing describe above, the TM-Oxi do not required additional clinical test." This means that the clinical performance data for the individual components (oximeter and NIBP) from their original clearance processes (which included adherence to ISO 9919 and ANSI/AAMI SP10 standards) were sufficient, and no new clinical study was performed for the integrated TM-Oxi system to prove its performance against the listed acceptance criteria.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for LD Technology. The logo consists of a circular symbol on the left and the text "LD Technology" on the right. The circular symbol contains a stylized image of a tree or plant with multiple concentric circles around it. The text "LD Technology" is written in a stylized font, with "LD" in a larger, bolder font than "Technology".

P1/7

510(k) Premarket Notification Number: K130056 Date of revision: 03/15/2013

APR 1 1 2013

510(k) Summary TM-Oxi device

This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of 21 CFR §807.92

  1. Submitter's Identification: Manufacturer: L.D TECHNOLOGY Address: 100 N.Biscayne Blvd, Suite 502 Miami, FL, 33132, USA Tel: 305-379-9900 E mail: albert.ldteck@gmail.com

2. Device Name / Classification

Trade name: Patient monitor Device Name and Model: TM-Oxi Regulation number: 21CFR 870.2300 Product Code: MWI Subsequent Product codes: DXN, DQA Classification: Class II Classification Name: monitor, physiological, patient

3. Predicate legally marketed devices

  • MD2000BVital Sign Monitor K100740.Applicant: Beijing Choice Electronic Technology Co., Ltd. Product . Code MWI. Subsequent Product codes: DXN, DQA
  • ESO Software K 102442. Applicant: LD Technology LLC .Product code DQA
  • ES Complex Software K1 13264. Applicant: LD Technology LLC. Product Code DXN, DQA, GZO, MNW ◆
  • CMS 50E K090671 Applicant: Contec Medical Systems Co., Ltd Product Code DQA .
  • MD 200A K093013 Applicant: Beijing Choice Electronic Technology Co., Ltd Product Codes DXN

4. Device Description

The TM-Oxi is a programmable electro medical system using one oximeter and one noninvasive blood pressure device (NIBP). The system comprises:

  • a. Hardware:
  • Non- invasive blood pressure, pump and valve boards and connections, tube, bladder ● and cuff 510k cleared (K093013) not modified.
  • USB HUB DC converter with 3 features .
    • ✔ Power supply of NIBP board via USB hub connected to USB port of the PC and

{1}------------------------------------------------

microchip converter 5V to 6 V.

  • microchip converter 510k cleared (K090671) not modified via USB hub > connected to USB port of the PC.
  • connected to USB port of the connected to USB port of the PC. >

b. Software:

Software:
The used software are ESO software 510k K102442 for data analysis and ES Complex software 510k number k 113264 for data management.

software 310k humber K 115204 for addinations of the data from oximeter (ESO) and blood pressure device (Contec 08A K110774).

pressure device (Contect 687 ATT 10777).
Therefore the TM-Oxi system will be a combination of the 2 separate devices.

5. Intended use and indications for use

  1. Intended use and indications for use in non-invasively measuring and displaying functional The I M-Oxi device indicated for use in non invasively measurement of the sive measurement of blood pressure (NIBP) of adult patients.

blood pressure (NIBP) of adult patients.
The TM-Oxi software uploads the data of the device and displays in olinical settings as an The TM-Oxi sollware uploads the data of the do rooms' in clinical settings as an
enhanced data management. The TM-Oxi software is intended for use in clinical settings as an enhanced data management. The TW On sentives to review, and evaluate the historical tests results. and for neath caree is intended for spot-checking of patients.

The TM-Oxi system does not report any diagnosis but provides values. It is the physician's THC TM-OXT Systeake proper judgments based on these numbers.

responsibility to make propel judgments based on areas any or on the order of a physician.

6. Performances, technical specifications and materials Performances

Block Diagram TM-Oxi /system

Image /page/1/Figure/15 description: The image shows a block diagram of a system with both software and hardware components. The software side includes modules for pulse rate, SpO2%, pulse wave, temporary storage, status report, data analysis and management, and blood pressure values. The hardware side includes components such as MD200A K093013, BP PCB + Pump, CMS50E K090671, USB Hub DC Converter, Microchip Converter (5V to 6V), and Hub USB ports.

2

{2}------------------------------------------------

Technical specifications I.e. Device description Patient contact materials:

Patient contact materials.
The materials in contact with the patient are the cuff, tube, blader and oximely from the first The materials in contact with the particul are PVC and they are latex free Cull material is inylon and Didustic may beauting in the material and latex free.

7. Contra-indications

  • Patients undergoing external defibrillation .
  • r atients undergoing external delife support devices, or any implanted electronic device.
  • Patients moving or long term monitoring
  • Do not use this device in the presence of :
    • Ilor use inis device in the praging (MR or MRI) equipment. MRI equipment may cause induced current to the device
    • Strong electromagnetic sources, such as electro surgery equipment.
    • Computed tomography (CT) equipment.

    • When using the oximeter finger probe, utilize the arm not in use for blood pressure,
  • arterial catheter, or having an AV fistula or pressure dressing.
  • arterial catherer, of having an Arream, such as methylene blue, indocyanine green, indigo carmine, and fluorescein
  • earmine, and freerer of dysfunctional hemoglogins, such as carboxyhemoglobin or . methemoglobin, will affect the accuracy of the SpO2 measurement.
  • mentemogiobin, will arrestricts blood flow, such as use of a blood pressure cuff or extremes Ally condition that resistance, may cause an inability to determine accurate pulse rate and SpO2 readings
  • and Spolish or false fingernails: Fingernail polish or false fingernails may cause . inaccurate SpO2 readings
  • maccurate 5px2 readings . regurgitation)
  • Caution with low perfused patient. Using the blood pressure device may cause skin erosion and/or pressure necrosis.

8. Undesirable side effects:

Side effects or adverse reactions are none known to date.

{3}------------------------------------------------

K130056 P4/7

9. Substantial equivalence Predicate legally marketed devices

  • are legally marketed devices
    MD2000BVital Sign Monitor K100740.Applicant: Beijing Choice Electronic Technology
    MD2000BVital Sign Monitor K100740.Applicant: Beijing DNDL DOA . MD2000D Vital Bigh Monitor . Subsequent Product codes: DXN, DQA
  • Co., Ltd. Product 00042. Applicant: LD Technology LLC .Product_code DQA .
  • ESO Software K113264. Applicant: LD Technology LLC. Product Code DXN, DQA, GZO, MNW

Similarities

  • Illarities
    ✔ The TM-Oxi hardware has same intended use and same performances and effectiveness
  • V The TM-Oxi software used the same source codes than ESO and ES Complex software
    ✔ The TM-Oxi software used the same source codes than ESO and ES Complex software The TM-OXI Software used the same source of the data of the oximeter and blood . pressure device.
  • And Children Children Children

Hardware Differences:

Hardware Technological characteristics

  • V The predicate device has an internal Oximeter PCB board connected to a SpO2 probe. The TM-Oxi uses 510k cleared oximeter which has PCB board and Spx2 probe in one The TM-OXF ases of the reast of the other provided by the OEM subcontractor.
  • The blood pressure device is powered by the USB port of a PC and microchip The blood pressure do rooms in and AC DC converter, measurements are managed by software and the results are directly display to a PC screen.

Hardware table of comparison

{4}------------------------------------------------

Name device (510knumberTM-Oxi hardwareMD2000B Vital sign monitor K100740
Intended use and features
Intended useThe TM-Oxi device indicated for use in non-invasively measuring and displaying functionaloxygen saturation of arterial hemoglobin (SpO2),pulse rate (PR), Non-invasive measurement ofblood pressure (NIBP) of adult patients inmedical facilities.TheTM-Oxi device is intended for spot-checkingof patients.The vital signs monitor is a portable deviceindicated for use in non-invasivelymeasuring and displaying functional oxygensaturation of arterial hemoglobin(SpO2), pulse rate (PR), Non-invasivemeasurement of blood pressure (NIBP) ofadult and pediatric patients in hospitals,medical facilities, and sub-acuteenvironments. The vital sign monitor isintended for spot-checking and/or continuousmonitoring of patients.
SpO2 %YESYES
Oximeter Wave formand Pulse rateYESYES
Systolic pressureYESYES
Diastolic pressureYESYES
Output
Output current100 mA75 mA
Output Power25VA25VA
Voltage5V DC7.2 V DC
Oximeter Technical Specifications
SpO2% Measuringrange0 to 100%0 to 100%
SpO2% Accuracy70%~100%: ±2%, Below 70% unspecified80-100%:±2%; 70-79%±3%;0-69%Unspecified
Pulse rate measuringrange30 to 250bpm30 to 235bpm
Pulse rate accuracy± 2 bpm during the pulse rate range of 3099bpm and 2% during the pulse rate range of100250 bpm30-100, ±2bpm; 101-235±2%;0-29Unspecified
LEDSWavelength Red : 660 nmWavelength IR : 880 nmRadiant Power Red : 1.8 mWRadiant Power IR : 2 mWWavelength Red : 660 nmWavelength IR : 940 nmRadiant Power Red : 1.8 mWRadiant Power IR : 2 mW
BP Technical Specifications
Measuring principleOscillometricOscillometric
Measurement modeControlled by softwareManual/Automatic/STAT
Measuring rangeSystolic30-255 mmHg30-255 mmHg
Measuring rangeDiastolic15-220 mmHg15-220 mmHg
Other specifications
Target populationAdultAdult and pediatric
Application siteFingerFinger
DisplayLCD (oximeter) and PC screenLCD, LED PC Screen
Power supplyUSBBattery, Converter
Safety ClassBFBF
Data transmissionUSBUSB or flash key (optional)
Testing bench60601-1-160601-1-260601-1-160601-1-2ISO 9919

{5}------------------------------------------------

K130056 P6/7

10. Performances and Effectiveness

    1. The OEM subcontractor for MD 200A K093013 provided the PCB boards (blood pressure, The OEM subcomractor to: ladder, cuff and they were cleared for use together.
  • punip allu valve), tuoc and bladder, can bladel, career sensor, USB cable The OLM subcomments and they were cleared for use together.

Therefore, ISO 9919 report from OEM oximeter and ANSI/AAMI SP10:2002 I herefore, ISO 9919 report nom OEM onlines. and read.
& ANSI/AAMI SP10:2002/A1:2003/A2: 2006 report from OEM blood pressure can be used for TM-Oxi

Testing comprises:

    1. Integration tests for oximeter and blood pressure device
    1. Integration tests for oxinceler and over (Cyclic redundancy check) Coding was performing to Soltware structurar testing using Gree ( ) ) ecurately capture, store, and analyze the data measured by the hardware.
    1. Laboratory testing for electrical safety (60601-1-1) and EMC (60601-1-2)
    1. Software verification (SRS/SDS/STD/STR)

Discussion of requirements of additional Clinical Test

Since the TM-Oxi is using 510K cleared devices clinically tested (ISO 9919 and SP10) and with regard of the testing describe above, the TM-Oxi do not required additional clinical test.

. .

The following facts:

  • The predicate device has an internal Oximeter PCB board connected to a SpO2 probe. The products as not the cleared oximeters which have PCB board and SpO2 probe in
  • one block, USB port, specific USB cable and software provided by the OEM subcontractor.
  • The blood pressure device is powered by the USB port of a PC and microchip The blood pressure are active and not AC DC converter, measurements are managed by software and the results are directly display to a PC screen.

Do not affect the performances and the effectiveness of the TM-Oxi system as shown in the Integration tests.

11. General Safety Concerns

General Safety Concerns

General Salety Collectilis
The laboratory tests reports of the components of TM-Oxi System (IEC 60601-1-2 and IEC The laboratify tosts reports of the componsible of the system comparing to the legally marketed predicate device.

6

{6}------------------------------------------------

K130056 P7/7

7

The following facts:

  • The predicate device has an internal Oximeter PCB board connected to a SpO2 probe. > The predioute de roo simeters which have PCB board and SpO2 probe in The TM-OXI uses broxesific USB cable and software provided by the OEM : subcontractor. .
  • subcomination.
    The blood pressure device is powered by the USB port of a PC and microchip The blood pressure are act not AC DC converter, measurements are managed by software and the results are directly display to a PC screen.

Do not affect the general safety of the TM-Oxi.

12. Standards

ndards
ANSI/AAMI SP10:2002/A1:2003/A2:2006/(R)2008 Issued 2008/12/18.Consolidated ANSWARNHT Of Feetronic or Automated Sphygmomanometers.

version. Manual, Licetionic of Pracedia.org +1 -1. General requirements for safety -ILC00001 - 1 1 155de Safety requirements for medical electrical systems Conateral standard: Safecy requirement 2 Medical Electrical Equipment Part 1-2: General

IEC 60001-1-2 133ded: 2001/07/90 Standard: Electromagnetic Compatibility - Req. and Tests -- Including Section 6 manual review

Tests = Including Section o manadi 107.07
IEC 60601-2-30: Issued: 1999/12/22 Ed: 2 Medical electrical equipment: Part 2: IEC 60001-2-30. Issued: 1777 12225, including essential performance, of automatic cycling non- invasive blood pressure monitoring equipment.

cycling noti- first.ive blood pressure montaining on including was conducted to validate and Perify that the Blood Pressure Monitor met all design specifications. (OEM subcontractor) vertly that the Diood Fressure Medical electrical equipment Particular requirements
ISO9919 Issued: 2005/03/15 Ed: 2 Medical electrical equipment Particular requirements ISO9919 ISSued. 2005/05/15 Date: 2 Nefarmance of pulse oximeter equipment for medical use (OEM subcontractor)

use (OEM Subcontractor)
Guidance for: Industry; FDA staff; and Content of Premarket Submissions for Software Contained in Medical Devices Document issued on: May 11, 2005

Contained in Medical devices -------------------------------------------------------------------------------------------------------------------------------------------------. March 01 2007

Conclusions

Conclusions
TM-Oxi is equivalent in performances, technology, safety and efficacy to the legally marketed predicate device.

{7}------------------------------------------------

Image /page/7/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle-like symbol with three wavy lines extending from its body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA" is arranged in a circular fashion around the symbol.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

April 11, 2013

LD Technology, Inc. c/o Mr. Albert Maarek Regulatory Consultant 100 North Biscayne Ave, Suite 502 Miami, FL 33132

Re: K130056

Trade/Device Name: TM-Oxi System Regulatory Number: 21 CFR 870.2300 Regulation Name: Physiological Patient Monitor (without arrhythmia detection or alarms) Regulatory Class: II (two) Product Code: 74 MWI Dated: March 4, 2013 Received: March 5, 2013

Dear Mr. Maarek:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act

{8}------------------------------------------------

Page 2 - Mr. Albert Maarek

or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely vours.

Image /page/8/Picture/6 description: The image contains the text "Owen 2 Earis -S". The text is written in a bold, sans-serif font. The letters are black, and the background is white. The text is slightly distorted, as if it were scanned from a document.

for

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{9}------------------------------------------------

K130056

Indications for Use

510(k) Number:

Device Name: TM-Oxi system

Indications for Use:

The TM-Oxi device indicated for use in non-invasively measuring and displaying functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate (PR), Non-invasive measurement of blood pressure (NIBP) of adult patients.

The TM-Oxi software uploads the data of the device and displays the data into a computer for enhanced data management. The TM-Oxi software is intended for use in clinical settings as an aid for health care professionals to review, analyze and evaluate the historical tests results. TheTM-Oxi device is intended for spot-checking of patients.

The TM-Oxi system does not report any diagnosis but provides values. It is the physician's responsibility to make proper judgments based on these numbers.

Prescription Use: Federal law restricts this device to sale by or on the order of a physician.

Prescription Use X Over-The-Counter Use AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Page 1 of 1

Owen P. Faris -S
2013.04.11 10:32:16
-04'00'

§ 870.2300 Cardiac monitor (including cardiotachometer and rate alarm).

(a)
Identification. A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.(b)
Classification. Class II (performance standards).