K Number
K123977
Date Cleared
2013-11-21

(330 days)

Product Code
Regulation Number
862.1215
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Randox Liquid CK-MB: The Randox Liquid CK-MB test system is a device intended for the quantitative in vitro determination of CK-MB concentration in serum and plasma. Measurements of CK-MB are used in the diagnosis and treatment of myocardial infarction (MI). This product is suitable for use on the RX series instruments including the RX Daytona and the RX Imola.
Randox CK-MB Calibrator: The Randox CK-MB calibrator is an in vitro diagnostic product intended for use in the calibration of Randox CK-MB methods.
The Randox Liquid CK-MB test system for the RX Imola is a prescription use device intended to be used in clinical laboratories only.

Device Description

Liquid CK-MB is supplied in a kit containing:

  • 4 x 20.0 mL CK-MB Buffer
  • 4 x 6.0 mL CK-MB Substrate.
    The CK-MB calibrator is lyophilised, single analyte, human serum based product. The kit contains ten vials (single level) with 1.0 mL per vial. Double de-ionised water is required for reconstitution.
AI/ML Overview

Here's a summary of the acceptance criteria and the study details for the Randox Liquid CK-MB and Randox CK-MB Calibrator, based on the provided text:

1. Acceptance Criteria and Reported Device Performance

The document outlines performance characteristics rather than explicit "acceptance criteria" in a pass/fail table. However, implied acceptance is demonstrated by the reported empirical results.

Performance CharacteristicAcceptance Criteria (Implied by Study Design)Reported Device Performance (RX Daytona)Reported Device Performance (RX Imola)
Precision (Total CV%)No explicit criteria given, but results are expected to be low and consistent.10.9% (at 9.90 U/L), 2.9% (at 1011.98 U/L), 2.5% (at 244.85 U/L), 2.8% (at 437.45 U/L), 3.3% (at 47.53 U/L), 2.5% (Calibrator), 3.2% (Control)12.5% (at 10.38 U/L), 4.0% (at 1001.29 U/L), 3.4% (at 245.62 U/L), 3.3% (at 437.83 U/L), 5.0% (at 47.43 U/L), 3.3% (Calibrator), 2.8% (Control)
Linearity/Reportable RangeDeviation from linearity less than 5% within the claimed range.7 - 2000 U/L6 - 1100 U/L
Extended Recovery± 10%10200 U/L ± 10%10200 U/L ± 10%
Limit of Detection (LoD)Not explicitly stated, but lower values indicate better sensitivity.5.06 U/L2.41 U/L
Limit of Blank (LoB)Not explicitly stated, but lower values indicate better sensitivity.2.91 U/L0.87 U/L
Limit of Quantitation (LoQ)≤20% accuracy and ≤20% imprecision.7.00 U/L6.00 U/L
Analytical Specificity (Interference)% of Control ± 10%Met for all tested interferents (except Intralipid*)Met for all tested interferents (except Intralipid*)
Method Comparison (Correlation with Predicate)High correlation (strong linear relationship)r = 0.999 (Y = 0.95 + 0.59)r = 0.999 (Y = 0.96 + 2.36)
Matrix Comparison (Correlation with Serum)High correlation (strong linear relationship)r = 1.000 (Lithium Heparin), r = 0.999 (Potassium EDTA)r = 0.999 (Lithium Heparin), r = 1.000 (Potassium EDTA)
Expected Values/Reference RangeAll values reported in the expected range for Healthy Individuals (<25U/L).All values within <25U/LAll values within <25U/L

Note: Intralipid interfered, and the recommendation is to use clear, non-lipemic serum and plasma.

2. Sample Size and Data Provenance for Test Set

  • Precision/Reproducibility:
    • Human Serum Samples (Spiked): 5 levels (10U/L, 50U/L, 250U/L, 450U/L, and 1100U/L).
    • Calibrator and Control Material: 1 level each.
    • Testing Protocol: 20 or 21 non-consecutive days, twice per day, 2 replicates per run. This totals 80-84 measurements per sample type.
    • Data Provenance: Not explicitly stated, but "unaltered human serum samples" and "control material" are mentioned. The origin (country, retrospective/prospective) is not provided.
  • Linearity/Assay Reportable Range:
    • Samples prepared at 11 levels.
    • Data Provenance: Not explicitly stated.
  • Detection Limit:
    • 300 determinations with 1 blank and 4 low-level samples for LoD.
    • Data Provenance: Not explicitly stated.
  • Analytical Specificity:
    • Two serum pools (20U/L and 415U/L) spiked with various interferents.
    • Data Provenance: Not explicitly stated.
  • Method Comparison:
    • RX Daytona: 90 serum samples (6 spiked)
    • RX Imola: 93 serum samples (4 spiked)
    • Data Provenance: "Patient samples," but no country of origin or retrospective/prospective information is given.
  • Matrix Comparison:
    • RX Daytona (Lithium Heparin): Minimum of 72 matched patient sample pairs.
    • RX Daytona (Potassium EDTA): Minimum of 71 matched patient sample pairs.
    • RX Imola (Lithium Heparin): Minimum of 71 matched patient sample pairs.
    • RX Imola (Potassium EDTA): Minimum of 71 matched patient sample pairs.
    • Data Provenance: "Patient samples," but no country of origin or retrospective/prospective information is given.
  • Expected Values/Reference Range:
    • 40 normal donors (16 Male, 24 Female; age 17-69).
    • Data Provenance: "human serum from 40 normal donors." No country of origin or retrospective/prospective information is given.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

There is no mention of experts being used to establish ground truth for the test set. The performance is based on analytical measurements and comparisons to a predicate device, which is typical for in vitro diagnostic (IVD) device submissions of this nature.

  • No number of experts mentioned.
  • No qualifications mentioned.

4. Adjudication Method for the Test Set

Not applicable. The study involves quantitative analytical measurements and comparisons rather than subjective assessments requiring adjudication.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, an MRMC comparative effectiveness study was not done. This type of study is typically for evaluating a diagnostic image interpretation aid rather than an in vitro diagnostic reagent.

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

Yes, the studies presented are all standalone performance evaluations of the analytical device (reagent used on the RX Daytona and RX Imola instruments). There is no human-in-the-loop component described for these performance characteristics.

7. The Type of Ground Truth Used

The "ground truth" for this in vitro diagnostic device is primarily established by:

  • Measured concentrations: For precision, linearity, detection limits, and analytical specificity, the "truth" is the actual or expected concentration of CK-MB in the prepared samples or controls.
  • Predicate device measurements: For method comparison, the "ground truth" or reference is the results obtained from the legally marketed predicate device (Roche CK-MB Kit on Hitachi 717).
  • Statistical analysis: For characteristics like precision, correlation coefficients serve as measures of agreement and performance.
  • Clinical consensus/literature: For expected values/reference range, the results are compared against an established range (<25 U/L) from cited literature.

8. The Sample Size for the Training Set

No explicit "training set" is mentioned for this device performance evaluation. IVD performance studies typically focus on validation of a pre-defined method rather than training an algorithm.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as no training set is mentioned in the provided text.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the word "RANDOX" in a bold, sans-serif font. The letters are all capitalized and evenly spaced. The letter "O" has a unique design, featuring a smaller circle inside the larger one, creating a target-like appearance.

1. SAFETY AND EFFECTIVENESS AS REQUIRED BY 21 CFR 807.92 STATEMENT

This summary of the 510(k) safety and effectiveness information is being submitted in accordance with the requirement 21 CFR 807.92.

2. SUBMITTER NAME AND ADDRESS

Name: Randox Laboratories Limited

Address: 55 Diamond Road, Crumlin, County Antrim, BT29 4QY, United Kingdom.

Telephone: +44 (0) 28 9442 2413 Fax: +44 (0) 28 9445 2912

Contact: Dr Pauline Armstrong E-mail: Pauline.Armstrong@randox.com

3. 510k NUMBER, DEVICE PROPRIETARY NAME, COMMON NAME, PURPOSE FOR SUBMISSION, REGULATORY CLASSIFCATION, PANEL. PRODUCT CODE AND 21 CFR NUMBER

510k No: K123977

Device Proprietary Name: Randox Liquid CK-MB, Randox CK-MB Calibrator

Common Name: Liquid CK-MB, CK-MB Calibrator

Purpose for Submission: New Device

Regulatory Classification: Class II

Panel: Clinical Chemistry

21 CFR Number: 21 CFR 862.1215

{1}------------------------------------------------

RANDOX

K123977 510(K) Summary Randox Liquid CK-MB, Randox CK-MB Calibrator

4. PREDICATE DEVICE PROPRIETARY NAME AND 510 (k) NUMBER

Predicate Device Proprietary Name: Roche Diagnostics CK-MB reagent, Roche Diagnostics CK-MB calibrator for automated systems.

510 (k) Number: K003158

5. INTENDED USE

The CK-MB test system is a device intended for the quantitative in vitro determination of CK-MB concentration in serum and plasma. Measurements of CK-MB are used in the diagnosis and treatment of myocardial infarction (MI). This product is suitable for use on the RX series instruments including the RX Daytona and RX Imola. The Randox liquid CK-MB is for clinical laboratory use and not for point-of-care use.

The Randox CK-MB calibrator is an in vitro diagnostic product intended for use in the calibration of Randox CK-MB methods.

The Randox CK-MB control is an in vitro diagnostic product intended for use in the quality control of Randox CK-MB methods. This product was approved previously with the following 510(k) number: K951223

The Randox Tri-level Cardiac Controls is an in vitro diagnostic product intended for use in the quality control of Cardiac Markers on clinical chemistry and Immunoassay systems. This product was approved previously with the following 510(k) number: K041361

These in vitro diagnostic devices are intended for prescription use only and can only be used by professionals.

6. DEVICE DESCRIPTION

Liquid CK-MB is supplied in a kit containing:

  • . 4 x 20.0 mL CK-MB Buffer
  • 4 x 6.0 mL CK-MB Substrate. .

The CK-MB calibrator is lyophilised, single analyte, human serum based product. The kit contains ten vials (single level) with 1.0 mL per vial. Double de-ionised water is required for reconstitution.

November 12, 2013

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image contains the word "RANDOX" in all capital letters. The font is bold and black. The letter "O" has a circle inside of it.

PREDICATE DEVICE COMPARISON TABLE 7.

Table 1 Predicate Comparison

CHARACTERISTICSRANDOX LIQUID CK-MBROCHE DIAGNOSTICSYSTEMS Inc. CK-MBREAGENT K003158
INTENDED USEThe CK-MB test system is adevice intended for thequantitative in vitro determinationof CK-MB concentration in serumand plasma. Measurements ofCK-MB are used in the diagnosisand treatment of myocardialinfarction (MI). This product issuitable for use on the RX seriesinstruments including the RXDaytona and RX Imola..The Roche Diagnostic Systems Inc.CK-MB reagent is an immunoinhibitionassay for the quantitative in vitrodetermination of the MB isoenzyme ofcreatine kinase in human serum andplasma on Roche automated clinicalchemistry analysers.
ASSAY PROTOCOLImmunoinhibition AssayUV assay with immunological inhibitionof CK-M
FORMATLiquid which are ready to useLiquids which are ready to use
KEY MATERIALMouse anti-human CK-Mmonoclonal antibodyMouse anti-human CK-M monoclonalantibody
STORAGE(unopened)Reagents are stable up to theexpiry date when storedunopened at +2 to +8°CReagents are stable up to theexpiry date when stored unopenedat +2 to +8°C
Sample TypePlasma and serumPlasma and serum
Sample TubesHeparin, EDTAHeparin, EDTA
Control(Frequency)-Randox CK-MB Control-Randox Tri-level CardiacControls levels 2 and 3-Two levels of control should be-Precinorm CK-MB Control-Precipath CK-MB-PreciControl ClinChem Multi 1-Control interval per institution
assayed at least once a dayrequirements

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the word "RANDOX" in a bold, sans-serif font. The "O" in "RANDOX" is stylized with a concentric circle design, resembling a target or a bullseye. The letters are all capitalized and evenly spaced, creating a balanced and symmetrical appearance.

K123977 510(K) Summary Randox Liquid CK-MB, Randox CK-MB Calibrator

Table 1 Predicate Comparison Continued.

CHARACTERISTICSRANDOX CK-MBCALIBRATORROCHE DIAGNOSTICSYSTEMS Inc. CK-MBCALIBRATORK003158
INTENDED USEThe Randox CK-MB calibrator isan in vitro diagnostic productintended for use in the calibrationof Randox CK-MB methods.C.f.a.s (Calibrator for automatedsystems) CK-MB is for use in thecalibration of Roche methods for thequantitative determination of CK-MB onRoche clinical chemistry analysers asspecified in the value sheets.
ANALYTECK-MBCK-MB
MATRIXHuman serumHuman serum
FORMLyophilisedLyophilised
STORAGE (opened)Stability of components in thereconstituted calibrator:At +25°C 8 hoursAt +4°C 5 daysAt -20°C 4 weeks when frozenonceStability of components in thereconstituted calibrator:At 15-25°C 24 hoursAt 2-8°C 2 daysAt (-15)-(-25)°C 4 weeks when frozenonce
(Unopened)Stable to the expiration date at 2-8oCStable to the expiration date at 2-8oC

{4}------------------------------------------------

8. TEST PRINCIPLE (1)

RANDOX

Immunoinhibition Assay: An antibody is incorporated into the CK reagent. This antibody will bind to and inhibit the activity of the M subunit of CK-MB. This means that only the activity of the B subunit in serum is measured. If the activity is multiplied by a factor of 2, it will give the activity of CK-MB in serum.

(1) WURBURG, U., ET AL., CLIN. CHEM. 1976; 54: 357.

The R1 reagent contains anti CK-MM antibody which binds the M subunit of CK in the serum sample thereby inhibiting the activity of the CK- M subunit. The CK-B activity is determined by the CK NAC method.

Creatine phosphate + ADP -> Creatine + ATP

Glucose + ATR > Glucose-6-P + ADP

Glucose-6-P + NADP =========================================================================================================================================================== - Gluconate-6-P + NADPH + H*

The rate of increase of absorbance at 340/700 nm due to the formation of NADPH is directly proportional to the activity of CK in the sample and if the activity is multiplied by a factor of two it will give the activity of CK-MB in the sample.

9. PERFORMANCE CHARACTERISTICS

Analytical performance:

a. Precision/Reproducibility:

Precision was evaluated consistent with C.L.S.I documents EP5 Precision studies were performed on the RX Daytona and RX Imola systems using control material and unaltered human serum samples that were spiked with CKMB concentrations of 10U/L, 50U/L, 250U/L, 450U/L and 1100U/L. Testing was conducted for one reagent lot of liquid CK-MB on one RX Daytona system and one RX Imola system, twice per day for 20 non-consecutive days. Two replicates per run was performed for each sample. No assay re-calibrations was required throughout the duration of the study. The results are summarized in the following tables.

{5}------------------------------------------------

K123977 510(K) Summary Randox Liquid CK-MB, Randox CK-MB Calibrator

Table 2 CKMB RX Daytona Precision Summary

All DataWithin RunBetween RunBetween DayTotal
SystemMethodProductData# Day# Run# RepMEANSDCVSDCVSDCVSDCV
Rx DaytonaCKMBHuman Serum 1All2140809.901.0110.20.000.00.394.01.0810.9
Rx DaytonaCKMBHuman Serum 5All2140801011.9817.601.715.611.517.191.729.142.9
Rx DaytonaCKMBHuman Serum 3All214080244.854.291.82.701.13.341.46.072.5
Rx DaytonaCKMBHuman Serum 4All214080437.457.491.75.161.28.021.812.132.8
Rx DaytonaCKMBHuman Serum 2All21408047.531.002.10.781.60.891.91.553.3
Rx DaytonaCKMBCKMB CalibratorAll214080189.653.201.71.780.92.961.64.712.5
Rx DaytonaCKMBCKMB controlAll214080137.452.792.03.112.31.431.04.423.2

Table 3 CKMB RX Imola Precision Summary

All DataWithin RunBetween RunBetween DayTotal
SystemMethodProductData# Day# Run# RepMEANSDCVSDCVSDCVSDCV
Rx ImolaCKMBHuman Serum 1All20408010.381.2412.00.000.00.494.71.3312.5
Rx ImolaCKMBHuman Serum 5All2040801001.2918.071.833.153.314.511.440.454.0
Rx ImolaCKMBHuman Serum 3All204080245.624.842.03.981.65.542.38.363.4
Rx ImolaCKMBHuman Serum 4All204080437.838.511.96.231.49.982.314.523.3
Rx ImolaCKMBHuman Serum 2All20408047.431.533.21.423.01.122.42.375.0
Rx ImolaCKMBCKMB CalibratorAll204080186.695.052.70.000.03.461.96.123.3
Rx ImolaCKMBCKMB controlAll204080135.573.702.70.390.30.390.33.742.8

b. Linearity/assay reportable range:

Linearity studies were performed to determine the analytical range of an assay - that is the range where the reported result is a linear function to the analyte concentration (or where deviation from linearity is less than 5%).

The linearity samples were prepared at 11 levels. The sponsor set a range from 0 analyte concentration (or other reasonable bottom of range level ) up to a high concentration approximately 10% greater than the upper level of linearity to be claimed for the method. The results are summarized in the following table:

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the word "RANDOX" in bold, black letters. The letters are sans-serif and evenly spaced. The word appears to be a logo or brand name. The letters are thick and have a slightly rough edge, suggesting they may have been printed or scanned.

K123977 510(K) Summary Randox Liquid CK-MB, Randox CK-MB Calibrator

Table 4 CKMB Linearity/ Reportable Range Summary

CKMBLinearityReportable Range
RX Daytona2000 U/L7 - 2000U/L
RX Imola1100 U/L6 - 1100U/L

Table 5 CKMB Validated Extended Recovery

CKMBExtended Recovery on event of Re-run
RX Daytona10200 U/L ± 10%
RX Imola10200 U/L ± 10%

c. Traceability, Stability, Expected values (controls, calibrators, or methods):

Table 6 Traceability Table Randox Liquid CK-MB Reagent and CK-MB Calibrator

ProductAnalyteSupplierProduct NoOriginSource
CK-MBCalibratorCK-MBLeeBiosolutions190-24HumanHumanHeart
LiquidCK-MBCK-MBRoche04688457AnimalMouse

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the word "RANDOX" in a bold, sans-serif font. The letters are large and black, creating a strong visual impact. The word appears to be a logo or brand name, presented in a clear and legible manner.

d. Detection limit:

Sensitivity studies have been carried out in accordance with C.L.S.I. guideline EP17-A 'Protocols for Determination of Limits of Detection and Limits of Quantification; Approved Guideline'. A Limit of Blank (L.o.B.), a Limit of Detection (L.o.D.) and a Limit of Quantification were performed on the RX Daytona and RX Imola systems.

CKMB on the RX Daytona

The Limit of Detection (LoD) for CKMB on the RX Daytona is 5.06 U/L based on 300 determinations, with 1 blank and 4 low level samples.

The Limit of Blank (LoB) is 2.91 U/L.

The Limit of Quantitation (LoQ) is 7.00 U/L as determined by the lowest concentration at which precision and accuracy are still met. Acceptable criteria ≤20% accuracy and ≤20% imprecision.

CKMB on the RX Imola

The Limit of Detection (LoD) for CKMB on the RX Imola is 2.41 U/L based on 300 determinations, with 1 blank and 4 low level samples.

The Limit of Blank (LoB) is 0.87 U/L.

The Limit of Quantitation (LoQ) is 6.00 U/L as determined by the lowest concentration at which precision and accuracy are still met. Acceptable criteria ≤20% accuracy and ≤20% imprecision.

e. Analytical Specificity:

The effects of potential interferents were determined by calculating the mean value of the spiked interferent with the corresponding control solution. Potential interferents and some commonly used drugs and exogenous susbstances were spiked into two serum pools with a CK-MB concentration of approximately 20U/L and 415U/L. The spiked sample results were compared to control samples prepared without the potential interferents.

Acceptance Criteria: % of Control ± 10%

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the word "RANDOX" in all capital letters. The letters are bold and black. The letter "O" has a double circle in the middle.

K123977 510(K) Summary Randox Liquid CK-MB, Randox CK-MB Calibrator

The results are summarized in the following table. No interference was observed at the concentrations listed in the table.

RX daytonaRX imola
InterferentCKMBLow PoolCKMBLow Pool
Haemoglobin1000 mg/dl1000 mg/dl
Bilirubin (F)30 mg/dl30 mg/dl
Bilirubin (C)60 mg/dl60 mg/dl
IntralipidⓇInteferes*Inteferes *
Triglycerides500 mg/dl500 mg/dl
Acetominophen1660µmol/l1660µmol/l
Caffeine308µmol/l308µmol/l
Phenytoin198µmol/i198µmol/l
Salicylic Acid4.34µmol/l4.34µmol/l
Digoxin6.15µmol/l6.15µmol/l
Nicotine6.2µmol/l6.2µmol/l
Theophylline222µmol/l222µmol/l
Acetyl Salicylic Acid3333µmol/l3333µmol/l
Ascorbic Acid227µmol/l227µmol/l
Chloramphenicol155µmol/l155µmol/l
Furosemide181 µmol/l181 µmol/l
Ibuprofen2425µmol/l2425µmol/l

Table 7 CKMB Interference Summary

*Recommend the use of clear, non-lipemic serum and plasma

Comparison Studies:

a. Method comparison with predicate device:

Correlation studies were carried out in accordance with C.L.S.I. guideline EP9-A2 'Method Comparison and Bias Estimation Using Patient Samples: Approved Guideline - Second Edition'.

90 serum samples (6 spikes) were tested on the RX Daytona analyzer and 93 serum samples (4 spikes) were tested on the RX Imola across 5 working days with each sample tested in singlicate. The test method was compared to the predicate device Roche CKMB Kit (ref. 12132893) that was tested on the Hitachi 717. One RX Daytona, one RX Imola and one Hitachi 717 system were used to carry out the testing.

{9}------------------------------------------------

RANDOX

K123977 510(K) Summary Randox Liquid CK-MB, Randox CK-MB Calibrator

The results of the studies are shown below.

CKMB on the RX Daytona v the predicate device

This method (Y) was compared with another commercially available method (X).

90 serum patient samples were analyzed spanning the range 8.88 to 1931.75 U/L and the following linear regression equation was obtained:

Y = 0.95 + 0.59

Correlation coefficient of r = 0.999

CKMB on the RX Imola v the predicate device

This method (Y) was compared with another commercially available method (X).

93 serum patient samples were analyzed spanning the range 6.64 to 992.31 U/L and the following linear regression equation was obtained:

Y = 0.96 + 2.36

Correlation coefficient of r = 0.999

b. Matrix comparison:

Matrix method comparisons for the CKMB assay on the RX Daytona and RX Imola systems were assessed. Both serum, lithium heparin and potassium EDTA plasma were conducted to determine whether method accuracy with lithium heparin and potassium EDTA plasma specimens are equivalent to serum results and that lithium heparin and potassium EDTA plasma does not interfere with either the method or the system.

CKMB Matrix comparison on the RX Daytona (Lithium Heparin)

Patient samples were drawn in matched pairs - one sample serum (x) and the second sample lithium heparin plasma (y). A minimum of 72 matched patient sample pairs were analyzed spanning the range 9.09 to 1931.8U/L and the following linear regression equation was obtained:

Y = 1.01x + -0.19

Correlation coefficient of r = 1.000

November 12, 2013

{10}------------------------------------------------

K123977 510(K) Summary Randox Liquid CK-MB, Randox CK-MB Calibrator

CKMB Matrix comparison on the RX Daytona (Potassium EDTA)

Patient samples were drawn in matched pairs -- one sample serum (x) and the second sample potassium EDTA plasma (y). A minimum of 71 matched patient sample pairs were analyzed spanning the range 9.09 to 1931.9 U/L and the following linear regression equation was obtained:

Y = 1.03x -1.92

RANDOX

Correlation coefficient of r = 0.999

CKMB Matrix comparison on the RX Imola (Lithium Heparin)

Patient samples were drawn in matched pairs - one sample serum (x) and the second sample lithium heparin plasma (y). A minimum of 71 matched patient sample pairs were analyzed spanning the range 6.81 to 992.31U/L and the following linear regression equation was obtained:

Y = 1.00x + 0.37

Correlation coefficient of r = 0.999

CKMB Matrix comparison on the RX Imola (Potassium EDTA)

Patient samples were drawn in matched pairs - one sample serum (x) and the second sample potassium EDTA plasma (y). A minimum of 71matched patient sample pairs were analyzed spanning the range 6.64 to 992.31 U/L and the following linear regression equation was obtained:

Y = 1.01x + 0.90

Correlation coefficient of r = 1.000

Clinical studies:

  • a. Clinical Sensitivity
    Not applicable

  • b. Clinical specificity:
    Not applicable

November 12, 2013

{11}------------------------------------------------

Image /page/11/Picture/0 description: The image shows the word "RANDOX" in a bold, sans-serif font. The letters are black against a white background. The "O" in "RANDOX" is stylized with two concentric circles, creating a distinctive visual element.

  • c. Other clinical supportive data (when a. and b. are not applicable):
    Not applicable

Clinical cut-off:

Not applicable

Expected values/Reference range:

The Reference range was evaluated consistent with C.L.S.I guideline C28-A2. How to Define and Determine Reference Intervals in the Clinical Laboratory. In a study, human serum from 40 normal donors (16 Male, 24 Female with age ranging from 17 to 69) were tested in singlicate on both the RX imola and RX daytona. The results obtained from both systems were ordered from lowest to highest before being examined for outliers using the Dixon test.

The following references have been cited on the package insert:

Klein G, Berger A, Bertholf R et al. Abstract: Multicenter Evaluation of Liquid Reagents for CK, CK-MB and LDH with Determination of Reference Intervals on Hitachi Systems. Clin Chem 2001; 47:Suppl. A30.

Thomas L, Müller M, Schumann G, Weidemann G et al. Consensus of DGKL and VDGH for interim reference intervals on enzymes in serum. J Lab Med 2005;29:301-308.

Upon confirmation there were no outliers in either data set, the values were compared to the quoted range of CK-MB <25U/L. Results of the study indicate that all values reported in the range for Healthy Individuals.

10. CONCLUSION

Testing results indicate that the proposed device is safe and effective for the stated intended use and is substantially equivalent to the predicate devices.

{12}------------------------------------------------

Image /page/12/Picture/1 description: The image shows the seal of the Department of Health & Human Services USA. The seal is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. In the center of the seal is a stylized eagle with its wings spread, symbolizing the department's mission to protect and promote the health and well-being of the nation.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

November 21, 2013

RANDOX LABORATORIES, LTD. DR. PAULINE ARMSTRONG 55 DIAMOND RD. CRUMLIN, COUNTY ANTRIM BT29 4QY UK

Re: K123977

Trade/Device Name: Randox Liquid CK-MB, Randox CK-MB Calibrator Regulation Number: 21 CFR 862.1215 Regulation Name: Creatine phosphokinase/creatine kinase or isoenzymes test system Regulatory Class: II Product Code: CGS, JIT Dated: November 13, 2013 Received: November 18, 2013

Dear Dr. Armstrong:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{13}------------------------------------------------

If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Carol Ei Benson -S for

Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{14}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: December 31, 2013 See PRA Statement on last page.

510(k) Number (if known) K123977

Device Name

Randox Liquid CK-MB, Randox CK-MB Calibrator

Indications for Use (Describe)

Randox Liquid CK-MB

The Randox Liquid CK-MB test system is a device intended for the guantitative in vitro determination of CK-MB concentration in serum and plasma. Measurements of CK-MB are used in the diagnosis and treament of myocardial infarction (MI). This product is suitable for use on the RX series instruments including the RX Daytona and the RX Imola.

Randox CK-MB Calibrator

The Randox CK-MB calibrator is an in vitro diugnostic product intended for use in the callbration of Randox CK-MB methods

The Randox Liquid CK-MB test system for the RX Imola is a prescription use device intended to be used in clinical laboratories only.

Type of Use (Select one or both, as applicable)

X Prescription Use (Pan 21 CFR 801 Subpart D)

[] Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FOR MENT FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR oncurrence of Center for Devices and Radiological Health (CDRH) (Signature)

FORM FDA 3881 (9/13)

§ 862.1215 Creatine phosphokinase/creatine kinase or isoenzymes test system.

(a)
Identification. A creatine phosphokinase/creatine kinase or isoenzymes test system is a device intended to measure the activity of the enzyme creatine phosphokinase or its isoenzymes (a group of enzymes with similar biological activity) in plasma and serum. Measurements of creatine phosphokinase and its isoenzymes are used in the diagnosis and treatment of myocardial infarction and muscle diseases such as progressive, Duchenne-type muscular dystrophy.(b)
Classification. Class II.