K Number
K123871
Device Name
FINGERTIP PULSE OXIMETER
Date Cleared
2013-05-10

(144 days)

Product Code
Regulation Number
870.2700
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
The Fingertip Pulse Oximeter MD300CF315 is a portable, non-invasive device intended for spot checking of oxygen saturation of arterial hemoglobin (SpO2) and pulse rate of adult and pediatric patient at home, and hospital (including clinical use in internist/surgery, Anesthesia, intensive care etc). It is not for continuous monitoring.
Device Description
The applicant device of Fingertip Pulse Oximeter MD300CF315 is a battery powered fingertip device, which can detect, display and speak out the measured %SpO2 and pulse rate value. The device is normally applied to adult and pediatric patient at home, and hospital (including clinical use in internist/surgery, Anesthesia, intensive care etc). The applicant device consists of power supply module, detector and emitter LED, signal collection and process module, display module, user interface, voice module and button control circuit. The pulse oximeter works by applying a sensor to a pulsating arteriolar vascular bed. The sensor contains a dual light source and photo detector. The one wavelength of light source is 660 nm, which is red light; the other is 940 nm, which is ultra red light. Skin, bone, tissue, and venous vessels normally absorb a constant amount of light over time. The photo detector in finger sensor collects and converts the light into electronic signal which is proportional to the light intensity. The arteriolar bed normally pulsates and absorbs variable amounts of light during systole and diastole, as blood volume increases and decreases. The ratio of light absorbed at systole and diastole is translated into an oxygen saturation measurement. This measurement is referred to as SpO2. The applicant devices are not for life-supporting or life-sustaining, not for implant. The devices or transducers are not sterile and the transducers are reusable and do not need sterilization or re-sterilization. The devices are for prescription. The devices do not contain drug or biological products. The devices are software -driven and the software validation is provided in Section 11 Software.
More Information

Not Found

No
The device description outlines a standard pulse oximetry technology based on light absorption and signal processing, with no mention of AI or ML algorithms for data analysis or interpretation.

No.
The device is intended for spot checking of oxygen saturation and pulse rate, which are diagnostic and monitoring functions, not therapeutic. The description explicitly states it is "not for life-supporting or life-sustaining".

Yes

Explanation: The device is intended for "spot checking of oxygen saturation of arterial hemoglobin (SpO2) and pulse rate," which are physiological parameters used to assess a patient's health status and aid in clinical decision-making. Despite explicitly stating "It is not for continuous monitoring", the measurement of these parameters serves a diagnostic purpose by providing information about a patient's condition.

No

The device description explicitly states it is a "battery powered fingertip device" and lists hardware components such as "power supply module, detector and emitter LED, signal collection and process module, display module, user interface, voice module and button control circuit." While it is "software-driven," it is not solely software.

No, this device is not an IVD (In Vitro Diagnostic).

Here's why:

  • IVD Definition: In Vitro Diagnostics are medical devices used to examine specimens taken from the human body, such as blood, urine, or tissue, to provide information about a person's health. This testing is performed outside of the living body (in vitro).
  • Device Function: The description clearly states that the Fingertip Pulse Oximeter MD300CF315 is a "portable, non-invasive device intended for spot checking of oxygen saturation of arterial hemoglobin (SpO2) and pulse rate". It works by applying a sensor to the fingertip and measuring light absorption through the skin and tissue.
  • No Specimen Collection: The device does not require the collection of any biological specimens from the patient. It directly measures physiological parameters on the patient's body.

Therefore, based on the provided information, the Fingertip Pulse Oximeter MD300CF315 is a non-invasive medical device used for physiological monitoring, not an In Vitro Diagnostic.

N/A

Intended Use / Indications for Use

The Fingertip Pulse Oximeter MD300CF315 is a portable, non-invasive device intended for spot checking of oxygen saturation of arterial hemoglobin (SpO2) and pulse rate of adult and pediatric patient at home, and hospital (including clinical use in internist/surgery, Anesthesia, intensive care etc). It is not for continuous monitoring.

Product codes

DQA

Device Description

The applicant device of Fingertip Pulse Oximeter MD300CF315 is a battery powered fingertip device, which can detect, display and speak out the measured %SpO2 and pulse rate value. The device is normally applied to adult and pediatric patient at home, and hospital (including clinical use in internist/surgery, Anesthesia, intensive care etc).

The applicant device consists of power supply module, detector and emitter LED, signal collection and process module, display module, user interface, voice module and button control circuit.

The pulse oximeter works by applying a sensor to a pulsating arteriolar vascular bed. The sensor contains a dual light source and photo detector. The one wavelength of light source is 660 nm, which is red light; the other is 940 nm, which is ultra red light. Skin, bone, tissue, and venous vessels normally absorb a constant amount of light over time. The photo detector in finger sensor collects and converts the light into electronic signal which is proportional to the light intensity. The arteriolar bed normally pulsates and absorbs variable amounts of light during systole and diastole, as blood volume increases and decreases. The ratio of light absorbed at systole and diastole is translated into an oxygen saturation measurement. This measurement is referred to as SpO2.

The applicant devices are not for life-supporting or life-sustaining, not for implant. The devices or transducers are not sterile and the transducers are reusable and do not need sterilization or re-sterilization. The devices are for prescription. The devices do not contain drug or biological products.

The devices are software -driven and the software validation is provided in Section 11 Software.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

pulsating arteriolar vascular bed

Indicated Patient Age Range

adult and pediatric patient

Intended User / Care Setting

at home, and hospital (including clinical use in internist/surgery, Anesthesia, intensive care etc)

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

Bench tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:

  • IEC 60601-1:1988 + Al(1991) + A2(1995), Medical Electrical Equipment Part1: . General requirements for safety.
  • . IEC 60601-1-2:2007, Medical Electrical Equipment - Part 1: General requirements for safety-2, Collateral Standard: Electromagnetic compatibility - Requirements and tests.
  • ISO 9919:2005, Medical electrical equipment - Particular requirements for basic safety and essential performance of pulse oximeter equipment for medical use.
  • ISO10993-5: 2009, Biological evaluation of medical device - Part 5: Tests for in vitro cytotoxicity.
  • ISO10993-10: 2010, Biological evaluation of medical device - Part 10: Tests for irritation and delayed-type hypersensitivity.

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

Not Found

Predicate Device(s)

K070371

Reference Device(s)

Not Found

Predetermined Change Control Plan (PCCP) - All Relevant Information

Not Found

§ 870.2700 Oximeter.

(a)
Identification. An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.(b)
Classification. Class II (performance standards).

0

1123871

MAY 1 0 2013

Section 3 510(k) Summary

(As required by 21 CFR 807.92)

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

The assigned 510(k) Number is: Pending

3.1 Date of Submission: Dec. 14, 2012

3.2 Sponsor Information

Establishment Registration Number: 3005569927

Beijing Choice Electronic Technology Co., Ltd.

Room 320, West Building 4, No.83 Fuxing Road, Beijing 100039, P.R.China

Contact Person:

Mr. Lei Chen

Phone: +86-10-88798300 Ext 6020

Fax: 215-4052545

Email: cc@choicemmed.com

3.3 Proposed Device Information Device Common or Usual Name: Pulse Oximeter Device Trade or Proprietary Name: Fingertip Pulse Oximeter Model: MD300CF315 Classification Name: Oximeter Product Code: DQA Regulation Number: 870.2700 Panel: Anesthesiology

3.4 Predicate Device 510(k) Number: K070371 Common Name: Oximeter Device Trade or Proprietary Name: Fingertip Pulse Oximeter Model: MD300C Classification Name: Oximeter Device Class: II Product Code: DQA Regulation Number: 870.2700 Review Panel: Anesthesiology

Manufacturer: Beijing Choice Electronic Technology Co., Ltd.

3-1

1

3.5 Device Description

The applicant device of Fingertip Pulse Oximeter MD300CF315 is a battery powered fingertip device, which can detect, display and speak out the measured %SpO2 and pulse rate value. The device is normally applied to adult and pediatric patient at home, and hospital (including clinical use in internist/surgery, Anesthesia, intensive care etc).

The applicant device consists of power supply module, detector and emitter LED, signal collection and process module, display module, user interface, voice module and button control circuit.

The pulse oximeter works by applying a sensor to a pulsating arteriolar vascular bed. The sensor contains a dual light source and photo detector. The one wavelength of light source is 660 nm, which is red light; the other is 940 nm, which is ultra red light. Skin, bone, tissue, and venous vessels normally absorb a constant amount of light over time. The photo detector in finger sensor collects and converts the light into electronic signal which is proportional to the light intensity. The arteriolar bed normally pulsates and absorbs variable amounts of light during systole and diastole, as blood volume increases and decreases. The ratio of light absorbed at systole and diastole is translated into an oxygen saturation measurement. This measurement is referred to as SpO2.

The applicant devices are not for life-supporting or life-sustaining, not for implant. The devices or transducers are not sterile and the transducers are reusable and do not need sterilization or re-sterilization. The devices are for prescription. The devices do not contain drug or biological products.

The devices are software -driven and the software validation is provided in Section 11 Software.

3.6 Intended Use

The Fingertip Pulse Oximeter MD300CF315 is a portable, non-invasive device intended for spot checking of oxygen saturation of arterial hemoglobin (SpO22) and pulse rate of adult and pediatric patient at home, and hospital (including clinical use in internist/surgery, Anesthesia, intensive care etc). It is not for continuous monitoring.

2

emarket Notification 510(k) Submission – Sec 3 510(k) Summar

3.7 Comparison with the Predicate Device

ITEMProposed DevicePredicate Device
Product NameMD300CF315 Fingertip Pulse OximeterMD300C Fingertip Pulse Oximeter
Power2 AAA-size alkaline batteries2 AAA-size alkaline batteries
Low Power Beep Tip FunctionYesNo
Display PartOLEDOLED
User Interface4 directions for display6 directions for display
SpO2Display Range0~99%0~99%
Measurement Range70~99%70~99%
Accuracy70%~99%: ±3%;
0%~69% no definition70%~99%: ±3%;
0%~69% no definition
Resolution1%1%
PRDisplay Range0~254BPM0~254BPM
Measurement Range30~235BPM30~235BPM
Accuracy3099bpm, ±2bpm; 100235bpm,
±2%3099bpm, ±2bpm; 100235bpm,
±2%
Resolution1bpm1bpm
Sensor
WavelengthLED RED660±2nm660±2nm
IR940±10nm940±10nm
Environment
RequirementsOperation Temperature5~40℃5~40℃
Storage Temperature-20~55℃-20~55℃
Operation Relative
Humidity≤80%, no condensation≤80%, no condensation
Storage Relative Humidity≤93%, no condensation≤93%, no condensation

Table 3-1 Performance Specification Comparisor

3-3

3

emarket Notification 510(k) Submission – Sec 3 510(k) Summar

Construction Materials
1. ABS plastic1. ABS plastic
2. Medical silicone gel2. Medical silicone gel
Outline product
Dimension$60mm \times 36mm \times 39mm (L \times W \times H)$
$58mm \times 32mm \times 34mm (L \times W \times H)$
Weight46g
32g

able 3-2 Differences between the Proposed Device and Predicate Device

ITEMProposed DevicePredicate Device
Device NameFingertip Pulse Oximeter MD300CF315Fingertip Pulse Oximeter MD300C (K070371)
ComponentsPower supply module, detector and emitter LED,
signal collection and process module, display
module, user interface, voice module and button
control circuitPower supply module, detector and emitter
LED, signal collection and process module,
display module, user interface and button
control circuit
Low Power Beep Tip
FunctionYesNo
User Interface4 directions for display6 directions for display
Outline productDimension
Weight60mm×36mm×39mm (L×W×H)58mm×32mm×34mm (L×W×H)
46g32g

3-4

:

.

·

4

Premarket Notification 510(k) Submission - Sec 3 510(k) Summary

3.8 Test Conclusion

Bench tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:

  • IEC 60601-1:1988 + Al(1991) + A2(1995), Medical Electrical Equipment Part1: . General requirements for safety.
  • . IEC 60601-1-2:2007, Medical Electrical Equipment - Part 1: General requirements for safety-2, Collateral Standard: Electromagnetic compatibility - Requirements and tests.
  • ISO 9919:2005, Medical electrical equipment - Particular requirements for basic safety and essential performance of pulse oximeter equipment for medical use.
  • ISO10993-5: 2009, Biological evaluation of medical device - Part 5: Tests for in vitro cytotoxicity.
  • ISO10993-10: 2010, Biological evaluation of medical device - Part 10: Tests for irritation and delayed-type hypersensitivity.

The Fingertip Pulse Oximeter MD300CF315 share the same Blood Oxygen Module with the predicate device MD300C, so we believe the clinical test of MD300CF315 can be exempted.

3.9 Substantially Equivalent Conclusion

The proposed device, Fingertip Pulse Oximeter MD300CF315, is determined to be Substantially Equivalent (SE) to the predicate device, Fingertip Pulse Oximeter MD300C, K070371, in respect of safety and effectiveness.

5

Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles an abstract caduceus or a representation of human figures.

May 10, 2013

Mr. Lei Chen Beijing Choice Electronic Technology Company, Limited North Building 3F, No.9 Shuangyuan Road Badachu Hitech, Shijingshan District Beijing, China 100041

Re: K123871

Trade/Device Name: Fingertip Pulse Oximeter Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: II Product Code: DOA Dated: April 10, 2013 Received: April 12, 2013

Dear Mr. Chen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

6

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH0ffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely vours.

Kwame O. Ulmer for
-s FDA

Anthony D. Watson, B.S., M.S., M.B.A. Director

Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices

Office of Device Evaluation

Center for Devices and

Radiological Health

Enclosure

7

Section 2 Indications for Use

Indications for Use Form

510(k) Number (if known): k12387/

Device Name: Fingertip Pulse Oximeter

Indications for Use:

The Fingertip Pulse Oximeter MD300CF315 is a portable, non-invasive device intended for spot checking of oxygen saturation of arterial hemoglobin (SpO2) and pulse rate of adult and pediatric patient at home, and hospital (including clinical use in internist/surgery, Anesthesia, intensive care etc). It is not for continuous monitoring.

Prescription Use _ V

(Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

AND/OR

Concurrence of CDRH, Office of Device Evaluation (ODE)

Page_1_of_1

ivision Sign-Off) sion Sign-Off)
sion of Anesthesiology, General Hospital Division of Ancouroental Devices

510(k) Number: K123871

2-1