(82 days)
The Biosense Webster PENTARAY® NAV eco High-Density Mapping Catheter is indicated for multiple electrode electrophysiological mapping of cardiac structures in the heart, i.e., recording or stimulation only. This catheter is intended to obtain electrograms in the atrial and ventricular regions of the heart. The PENTARAY® NAV eco High-Density Mapping Catheter provides location information when used with compatible CARTO® 3 EP Navigation Systems. (This catheter is not compatible with CARTO® 3 EP Navigation Systems prior to Version 3.x.)
The PENTARAY® High-Density Mapping Catheter family is a 7 Fr, multi-electrode electrophysiological mapping catheter designed for diagnostic electrogram mapping and pacing in all chambers (atria and ventricles) of the heart. Six proposed PENTARAY® NAV eco High-Density Mapping Catheters (D-1282-07-S to D-1282-12-S) will be similar to six predicate PENTARAY® NAV High-Density Mapping Catheters (D-1282-01-S to D-1282-06-S), modified only to implement an "eco" Handle. The modification to implement this "eco" Handle involves removing the Printed Circuit Board (PCB) that is associated with the Magnetic Location Sensor from the Handle Extension and connecting this catheter to an existing "eco" Interface Cable and a short, reusable cable component that now contains this PCB. This cable containing the PCB (known as the Dongle) is associated with the CARTO® 3 Electrophysiology (EP) Navigation System. This "eco" Handle is the only modification of the predicate catheters to create the six proposed Biosense Webster PENTARAY NAV eco Catheter models.
The working tip of the six proposed PENTARAY NAV eco Catheters is identical to the predicate catheters. The proposed catheters will continue to have five flexible spines, each with four ring electrodes for a total of 20 electrodes. There will continue to be three different electrode spacing configurations (4-4-4, 2-6-2, or 1-8-1 mm) and two different curves (F or D) in order to accommodate different clinical situations. The tip of the shaft of both the predicate and proposed catheters house a Magnetic Location Sensor that provides magnetic location information when used with the CARTO® 3 EP Navigation System. In addition to the 20 spine electrodes, two additional Ring Electrodes are located near the tip of the shaft for a total of 22 ring electrodes that provide location information via ACL technology when used with the CARTO 3 EP Navigation System. The shafts of the proposed and predicate catheters are the same as are the distal ends of the Handle. All the modifications are in the proximal portion of the Handle where the changes described in the previous paragraph were effected.
This document describes a 510(k) premarket notification for the PENTARAY® NAV eco High-Density Mapping Catheter. The submission aims to demonstrate substantial equivalence to a predicate device, the PENTARAY® NAV High-Density Mapping Catheter.
The key information regarding the acceptance criteria and the study performed is as follows:
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criteria Category | Reported Device Performance |
|---|---|
| Bench Testing | Passed all intended criteria in accordance with appropriate standards and test criteria. |
| ECG Signal Quality (Animal Test) | Confirmed that the proposed device had no change in the quality of the ECG signals compared to the predicate. |
2. Sample Size Used for the Test Set and Data Provenance
- Test Set Sample Size: Not explicitly stated in terms of a specific numerical count for a "test set" in the context of a clinical study. The device's performance was evaluated through bench testing and a single animal test.
- Data Provenance: The bench testing and animal testing were conducted internally by the manufacturer (Biosense Webster, Inc.). No information on country of origin for any external data, or whether it was retrospective or prospective, is provided as no human clinical trials were performed.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
Not applicable. No clinical test set requiring expert ground truth establishment was conducted. The evaluation relied on engineering bench tests and an animal study.
4. Adjudication Method for the Test Set
Not applicable. No clinical test set requiring human expert adjudication was conducted.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No. An MRMC comparative effectiveness study was not performed. The submission explicitly states, "No clinical testing was deemed necessary." The device is a hardware mapping catheter, not an AI diagnostic tool that would typically be evaluated with MRMC studies.
6. Standalone (Algorithm Only) Performance Study
Not applicable. The PENTARAY® NAV eco is a medical device (catheter), not a standalone algorithm. Its performance is evaluated as an integrated system with the CARTO® 3 EP Navigation System.
7. Type of Ground Truth Used
- For Bench Testing: Engineering specifications and established industry standards for catheter performance served as the ground truth.
- For Animal Test (ECG signal quality): The quality of ECG signals obtained using the predicate device likely served as the comparative "ground truth" to ensure no degradation with the modified device.
8. Sample Size for the Training Set
Not applicable. This device is a hardware catheter, not an AI algorithm that requires a training set.
9. How the Ground Truth for the Training Set was Established
Not applicable, as there was no training set for an AI algorithm.
{0}------------------------------------------------
MAR 0 5 2013
18. 510(K) SUMMARY
Applicant:
Contact Person:
Date:
Trade or Proprietary Name:
Common or Usual Name of Device:
Classification Name:
Predicate Devices:
Manufacturer:
Manufacturing Sites:
Biosense Webster, Inc. 3333 Diamond Canyon Road Diamond Bar, CA 91765 Phone: 800-729-9010 x8597 Fax: 909-839-8804
Wayne R. Hohman Project Manager Regulatory Affairs
December 11, 2012
PENTARAY® NAV eco High-Density Mapping Catheter
Deflectable Tip Electrophysiology Catheter-Diagnostic
Electrode recording catheter or electrode recording probe (21 CFR 870.1220, Product Code MTD)
PENTARAY® NA V High-Density Mapping Catheter 510(k): K120425
Biosense Webster, Inc. 3333 Diamond Canyon Road Diamond Bar, CA 91765
Biosense Webster, Inc 15715 Arrow Highway Irwindale, CA 91706
Cordis de Mexico, S.A., de C.V. Circuito Interior Norte 1820 Parque Industrial Salvarcar Cuidad Juarez, Chihuahua CP 32599 Mexico
Sterilization Sites:
Steris Isomedix 7685 St. Andrews Avenue San Diego, CA 92154
Sterigenics, Inc. 2400 Airport Road
{1}------------------------------------------------
Santa Teresa, NM 88008
18.1 Substantially Equivalent To
The Biosense Webster PENTARAY® NAV eco High-Density Mapping Catheter is substantially equivalent to the Biosense Webster PENTARAY® NAV High-Density Mapping Catheter (510(k) K120425, cleared May 24, 2013).
Description of the Device Subject to Premarket Notification 18.2
The PENTARAY® High-Density Mapping Catheter family is a 7 Fr, multi-electrode electrophysiological mapping catheter designed for diagnostic electrogram mapping and pacing in all chambers (atria and ventricles) of the heart. Six proposed PENTARAY® NAV eco High-Density Mapping Catheters (D-1282-07-S to D-1282-12-S) will be similar to six predicate PENTARAY® NAV High-Density Mapping Catheters (D-1282-01-S to D-1282-06-S), modified only to implement an "eco" Handle. The modification to implement this "eco" Handle involves removing the Printed Circuit Board (PCB) that is associated with the Magnetic Location Sensor from the Handle Extension and connecting this catheter to an existing "eco" Interface Cable and a short, reusable cable component that now contains this PCB. This cable containing the PCB (known as the Dongle) is associated with the CARTO® 3 Electrophysiology (EP) Navigation System. This "eco" Handle is the only modification of the predicate catheters to create the six proposed Biosense Webster PENTARAY NAV eco Catheter models.
The working tip of the six proposed PENTARAY NAV eco Catheters is identical to the predicate catheters. The proposed catheters will continue to have five flexible spines, each with four ring electrodes for a total of 20 electrodes. There will continue to be three different electrode spacing configurations (4-4-4, 2-6-2, or 1-8-1 mm) and two different curves (F or D) in order to accommodate different clinical situations. The tip of the shaft of both the predicate and proposed catheters house a Magnetic Location Sensor that provides magnetic location information when used with the CARTO® 3 EP Navigation System. In addition to the 20 spine electrodes, two additional Ring Electrodes are located near the tip of the shaft for a total of 22 ring electrodes that provide location information via ACL technology when used with the CARTO 3 EP Navigation System. The shafts of the proposed and predicate catheters are the same as are the distal ends of the Handle. All the modifications are in the proximal portion of the Handle where the changes described in the previous paragraph were effected.
18.3 Indications for Use
The Biosense Webster PENTARAY® NAV eco High-Density Mapping Catheter is indicated for multiple electrode electrophysiological mapping of cardiac structures in the heart, i.e., recording or stimulation only. This catheter is intended to obtain electrograms in the atrial and ventricular regions of the heart. The PENTARAY® NAV eco High-Density Mapping Catheter provides location information when used with compatible CARTO® 3 EP
{2}------------------------------------------------
Navigation Systems. (This catheter is not compatible with CARTO® 3 EP Navigation Systems prior to Version 3.x.)
Performance Data 18.4
The PENTARAY® NAV eco High-Density Mapping Catheter underwent Bench Testing and passed all intended criteria in accordance with appropriate standards and test criteria. The only animal test that was conducted was one to confirm that the proposed device had not change in the quality of the ECG signals. No clinical testing was deemed necessary.
18.5 Overall Performance Conclusions
The nonclinical studies demonstrate that the PENTARAY® NAV eco High-Density Mapping Catheter is safe and effective for anatomic mapping of the heart and establish equivalence of the PENTARAY® NAV eco High-Density Mapping Catheters to their predicate devices, the PENTARAY® NAV High-Density Mapping Catheters.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like symbol with three curved lines representing its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the symbol.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 5, 2013
Biosense Webster, Inc Wayne Hohman, Project Manager, Regulatory Affairs 3333 Diamond Canyon Road Diamond Bar, CA 91765
Re: K123837
Trade/Device Name: PENTARAY® NAV eco High-Density Mapping Catheter Regulation Number: 21 CFR 870.1220 Regulation Name: Electrode recording catheter or electrode recording probe Regulatory Class: Class II Product Code: MTD Dated: December 11, 2012 Received: February 4, 2013
· Dear Mr. Wayne Hohman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{4}------------------------------------------------
Page 2 – Mr. Wayne Hohman
forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHQffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
Owen P. Faris -S
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
7. INDICATIONS FOR USE STATEMENT
510(k) No (if known):
Device Name: PENTARAY® NAV eco High-Density Mapping Catheter
Indication for Use:
The Biosense Webster PENTARAY® NAV eco High-Density Mapping Catheter is indicated for multiple electrode electrophysiological mapping of cardiac structures in the heart, i.e., recording or stimulation only. This catheter is intended to obtain electrograms in the atrial and ventricular regions of the heart. The PENTARAY® NAV eco High-Density Mapping Catheter provides location information when used with compatible CARTO® 3 EP Navigation Systems. (This catheter is not compatible with CARTO® 3 EP Navigation Systems prior to Version 3.x.)
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Owen Pharis -S
§ 870.1220 Electrode recording catheter or electrode recording probe.
(a)
Identification. An electrode recording catheter or an electrode recording probe is a device used to detect an intracardiac electrocardiogram, or to detect cardiac output or left-to-right heart shunts. The device may be unipolar or multipolar for electrocardiogram detection, or may be a platinum-tipped catheter which senses the presence of a special indicator for cardiac output or left-to-right heart shunt determinations.(b)
Classification. Class II (performance standards).