K Number
K123822
Manufacturer
Date Cleared
2013-09-27

(289 days)

Product Code
Regulation Number
880.5725
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The KleinTouch Pump is an infiltration pump used to cause a flow of fluid from an IV bag into a patient in a manner controlled manually by a health care professional. The KleinTouch pump is not intended to be used as an IV infusion pump.

Device Description

Device Functionality: The KleinTouch Pump is an electrically or battery operated, manually controlled, peristaltic pump using three rollers to create pumping action in a section of sterile tubing.
Scientific Concept(s): The KTP is designed to specifically function with sterile single use only Klein Touch Tubing (KT20). Tubing must be primed before use. Please refer to the KT20 Instructions for Use for further details.
Physical And Performance Characteristics: Device Design Display: Color Graphic LCD Dual Foot Switches With Hose - Air Activated Dimensions: 11" x 10 ¾" x 5" Weight: 8.4 lbs.
Components Used: Motor Fluid Pathway Materials Silicone Inlet Hospital Power Cord Hospital Grade Plugs Barcode Reader Base, Plastic Battery Instrument Foam PCBA - Power Board PCBA, Main Circuit Board Pump, Peristaltic, 3-Roller, Fixed Clamp Switch, Rocker, Lighted
Physical Properties: The Pump is intended for use in combination with the KT20 tubing. The peristaltic pump using three rollers will create pumping action in KTP20 tubing set specifically designed for the KleinTouch Pump

AI/ML Overview

The provided text describes the 510(k) Notification for the HK Surgical KleinTouch Pump and Tubing Set. However, it does not contain specific acceptance criteria tables or detailed study results that typically outline performance metrics for AI/ML-based medical devices. The document focuses on regulatory compliance, intended use, and general safety testing for an infusion pump, which is a physical device, not an AI/ML algorithm.

Therefore, it is not possible to extract the requested information regarding acceptance criteria and a study proving device performance in the context of AI/ML, as the document describes a traditional medical device (an infiltration pump).

Here's an analysis of what can be extracted and why other parts of your request cannot be fulfilled:

1. A table of acceptance criteria and the reported device performance:

  • Cannot be provided. The document lists general performance testing categories (UL Standards, EMC, Biocompatibility, Sterilization, Human Factors) but does not provide specific quantitative acceptance criteria or reported performance values for these tests. For example, it states "Performance testing demonstrates the KleinTouch Pump meets the specifications" but doesn't detail what those specifications are or the exact results obtained.

2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

  • Cannot be provided. This information is relevant for studies involving data-driven devices or algorithms. The KleinTouch Pump is a physical device, and its testing involves physical and electrical safety standards, biocompatibility, and human factors, not data samples in the context of AI/ML.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

  • Cannot be provided. "Ground truth" in the context of expert consensus is typically for evaluating diagnostic or predictive AI/ML models. For a physical pump, testing involves adherence to engineering standards and safety protocols by qualified testing labs (MET LABS is mentioned), not expert consensus on medical image interpretation or similar. The "Human Factors evaluation" implies expert observation, but no specific numbers or qualifications are given for that.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

  • Cannot be provided. Adjudication methods are specific to scenarios where multiple experts might disagree on ground truth, which is not applicable to the testing described for this device.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • Cannot be provided. MRMC studies are for evaluating diagnostic accuracy, particularly with AI assistance, which is not relevant to a physical infiltration pump.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

  • Cannot be provided. This device is a physical pump, not an algorithm. Therefore, "standalone" performance in the AI/ML sense is not applicable.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

  • Cannot be provided. The "ground truth" for this device's performance is adherence to established international and national safety and performance standards (e.g., UL 2601-1, IEC/EN 60601-1, ISO10993-1, ISO11135). The testing verifies that the device meets these engineering and safety requirements.

8. The sample size for the training set:

  • Cannot be provided. The device is not an AI/ML model, so there is no "training set."

9. How the ground truth for the training set was established:

  • Cannot be provided. As there is no training set, this question is not applicable.

In summary: The provided text is a 510(k) summary for a physical medical device (an infiltration pump) and details its design, intended use, and compliance with various established safety and performance standards. It does not describe an AI/ML device or studies involving data-driven performance metrics, ground truth establishment, or human-AI interaction.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows a logo for HK Surgical. The logo features the letters "HK" in a stylized font, with the word "SURGICAL" underneath. Below that is the website "HKSurgical.com". The address "1271 Puerta del Sol San Clemente, CA 92673" is printed below the website.

HK Surgical KleinTouch Pump and Tubing Set 510(k) Notification

5. 510(k) Summary

Submitter:HK Surgical, Inc.
Address:1271 Puerta del Sol San Clemente, CA 92673
Phone number:949-369-0101
Fax number:949-369-9797, email peter@hksurgical.com
Contact person:Pejman Farivar
Date prepared:12/10/2012
510(k) Number:To be Determined
Trade name:KleinTouch Pump (KTP)
Common name:Infiltration Pump
Classification name:Pump, Infusion
Classification:Product Code: FRN,Regulation Number: 880.5725Class II
Substantial equivalenceclaimed to:Klein Infiltration Pump - K031432
Description:Device Functionality:The KleinTouch Pump is an electrically or battery operated, manuallycontrolled, peristaltic pump using three rollers to create pumpingaction in a section of sterile tubing.Scientific Concept(s)The KTP is designed to specifically function with sterile single use onlyKlein Touch Tubing (KT20). Tubing must be primed before use.Please refer to the KT20 Instructions for Use for further details.Physical And Performance Characteristics
Device Design Display: Color Graphic LCD Dual Foot Switches With Hose - Air Activated Dimensions: 11" x 10 ¾" x 5" Weight: 8.4 lbs.
Components Used Motor Fluid Pathway Materials Silicone Inlet Hospital Power Cord Hospital Grade Plugs Barcode Reader Base, Plastic Battery Instrument Foam PCBA - Power Board PCBA, Main Circuit Board Pump, Peristaltic, 3-Roller, Fixed Clamp Switch, Rocker, Lighted
Physical Properties The Pump is intended for use in combination with the KT20 tubing. The peristaltic pump using three rollers will create pumping action in KTP20 tubing set
specifically designed for the KleinTouch Pump
Intended use:The KleinTouch Pump is an infiltration pump used to cause a flow offluid from an IV bag into a patient in a manner controlled manually by ahealth care professional. The KleinTouch pump is not intended to beused as an IV infusion pump.
Safety and LabelingWARNINGS AND CAUTIONSThe device labeling contains cautions and warnings which areconsidered essential to the safety of personnel, patients, equipment,and property.
This device must be used by a trained professional with currentknowledge regarding safety of total dosage. Orders for dosage mustbe by a physician and legible.Protection against electric shock: Class I, Type BRefer to the Instructions for Use for the Pump and Infusion Set forWarnings and cautions.
Performance TestingMET LABS tested per UL Standards 2601-1; 2nd ed. MedicalElectrical Equipment- Part 1 General Requirements for Safety• CSA C22.2• UL 60601-1• EN/IEC 60601-1
EMC:• IEC/EN 60601-1-2:2007-ED30• JSA-JIS T0601-1-2:2002• AS/NZ 3200.1.2:2005Biocompatibility Testing demonstrates all fluid path materials arebiocompatible in accordance with ISO10993-1.Sterilization Validation to SAL 10-6 in accordance with ISO11135.Human Factors evaluation demonstrates the KleinTouch Pump can beused safely with no critical task errors.Performance testing demonstrates the KleinTouch Pump meets the specifications.

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the logo and contact information for HK Surgical. The logo features the letters 'HK' in a stylized font, with the word 'SURGICAL' written below. The website 'HKSurgical.com' is also listed. The address '1271 Puerta del Sol, San Clemente, CA 92673' is provided, indicating the company's location.

Substantial Equivalence Conclusion:

HK has provided test data to assess the technological differences and demonstrate Subject Device is substantially equivalent to the predicate device (K031432) and at least is safe and effective.

Page 15 of 92HK Surgical KleinTouch Pump and Tubing Set 510(k) NotificationConfidential
---------------------------------------------------------------------------------------------

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three legs, representing the department's mission to protect the health of all Americans and provide essential human services. The eagle is encircled by the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

September 27, 2013

HK Surgical, Inc. Ms. Clare Bennett General Manager 1271 Puerta del Sol SAN CLEMENTE, CA 92673

Re: K123822

Trade/Device Name: KleinTouch Pump (KTP) Regulation Number: 21 CFR 880.5725 Regulation Name: Infusion Pump Regulatory Class: II Product Code: FRN Dated: February 1, 2013 Received: February 1, 2013

Dear Ms. Bennett:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must

{3}------------------------------------------------

Page 2 - Ms. Clare Bennett

comply with all the Act 's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events} (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

· Sincerely yours.

Mary S. Runner -S

Kwame Ulmer M.S. Acting Division Director Division of Anesthesiology, General Hospital. Respiratory, Infection Control and Dental Devices Office of Device Evaluation -Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows a logo for HK Surgical. The logo features the text "HK SURGICAL" in a bold, sans-serif font, with "HKSurgical.com" written below it. The address "1271 Puerta del Sol, San Clemente, CA 92673" is printed below the website address. The logo is simple and professional, and it is likely used to identify the company on its products and marketing materials.

K123822

HK Surgical KleinTouch Pump and Tubing Set 510(k) Notification

4. Indications for Use Statement

The KleinTouch Pump is an infiltration pump used to cause a flow of fluid from an IV bag into a patient in a manner controlled manually by a health care professional. The KleinTouch pump is not intended to be used as an IV infusion pump.

. . Prescription Use

(21 CFR 807 Subpart C)

AND/OR Over-The-Counter Use

(Part 21 CFR 801 Subpart D)

(Please DO NOT WRITE BELOW THIS UNE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Cancurrence of CDRH, Office of Dovice Evaluation (ODE)

Image /page/4/Picture/10 description: The image shows the logo of the Food and Drug Administration (FDA). The logo is composed of the letters "FDA" in a stylized font. The letters are made up of small, geometric shapes, giving the logo a textured appearance. The logo is black and white.

Richard C. Chapman 2013.09.27 15:31:41 -04'00'

§ 880.5725 Infusion pump.

(a)
Identification. An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.(b)
Classification. Class II (performance standards).