K Number
K123743
Date Cleared
2013-03-26

(110 days)

Product Code
Regulation Number
880.5860
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Single use devices that are indicated for use as an accessory with pre-filled ISO Standard glass syringes to aid in the protection of healthcare professionals, patients who self-inject doctor prescribed medications and individuals that assist self-injecting patients, from accidental needlesticks. The intended patient population is unrestricted and may include children and adults, and parenteral methods of administration. Additionally, the PLUS device is designed with a larger viewing window indicated where pharma company offering is a low fill volume and instructions request visualization. The PLUS device is designed with a robust plunger and built in extended finger flanges indicated where pharma customer offering is viscous.

Device Description

The modification to the predicate device is the addition of the X100L PLUS device to the UltraSafe Passive Needle Guard X-Series family. The SSI UltraSafe PLUS Passive Needle Guard is an antineedlestick accessory for pre-filled ISO standard glass syringes. The device is composed of a guard, body, spring and plunger, is non-sterile and single use. The activation of the UltraSafe PLUS Passive Needle Guard device remains the same. Upon completion of the guard will slide forward, cover and lock over the needle of the syringe. It is a visual and/or audible recognition that the device safety feature has activated. The SSI devices are categorized as skin contact with a duration of category A- limited (< 24 h) as per ISO 10993 Biological evaluation of medical devices- Part 1: Evaluation and testing.

AI/ML Overview

The provided 510(k) summary describes the UltraSafe PLUS Passive Needle Guard, an accessory for pre-filled ISO standard glass syringes designed to prevent accidental needlesticks. The submission focuses on demonstrating substantial equivalence to a predicate device.

Here's an analysis of the acceptance criteria and study details based on the provided text:


1. Table of Acceptance Criteria and Reported Device Performance

Acceptance CriteriaReported Device Performance (UltraSafe PLUS Passive Needle Guard)
Activation Functionality: The safety device activates as intended upon completion of injection, covering and locking over the needle."The safety device activated in all 500 devices with zero failures... All 500 devices activated with zero failures meaning the device activated at the appropriate time under expected conditions of use and provided protection against unintended sharps injury..."
Non-Impedance to Clinical Performance: The device does not impede or adversely affect the intended clinical performance (e.g., successful injection, emptying of syringe, administering medication)."All injections were completed successfully; the safety device activated in all 500 devices with zero failures, consistent with the emptying of the syringe and administering all the medication, meaning the device did not impede the intended clinical performance."
Premature Activation Prevention: The device does not activate prematurely under expected conditions of use."All 500 devices activated with zero failures meaning the device activated at the appropriate time under expected conditions of use..."
Protection Against Unintended Sharps Injury: Provides protection against unintended sharps injury until disposal."All 500 devices activated with zero failures meaning the device activated at the appropriate time under expected conditions of use and provided protection against unintended sharps injury until disposal."
Failure Rate (Statistical Confidence): The true failure rate (for activation) is no higher than 0.7% with 97.5% confidence, and no higher than 1.1% with 99.5% confidence. (This is derived from the FDA Guidance cited and implied as the acceptance benchmark for the 0 failures observed)."testing 500 devices with zero failures for activation for a '97.5% confident that the true failure rate was no higher than 0.7% and 99.5% confidence that it is no higher than 1.1%'"

Study Details

  1. Sample size used for the test set and the data provenance:

    • Sample Size: 500 devices.
    • Data Provenance: The study was a "simulated use study" conducted internally by Safety Syringes, Inc. (as indicated by "Safety Syringes, Inc. maintains a Quality System... Test procedures, test protocols, and reports are maintained within the Quality System"). The country of origin is not explicitly stated, but the company is based in Carlsbad, CA, USA. The study design is prospective as it involves testing new devices to confirm performance.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable. This device is a mechanical safety accessory, and its "ground truth" (i.e., whether it activated correctly or not, or impeded injection) is determined by direct observation during simulated use, not by expert interpretation of complex data. No expert review or consensus was required for establishing the outcome of the mechanical activation.
  3. Adjudication method for the test set:

    • Not applicable. The "ground truth" was directly observed mechanical function (activation/non-activation, successful injection). There was no need for adjudication as there were zero failures observed which simplifies the outcome to a clear pass/fail for each device.
  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. This is a medical device (a needle guard) and not an AI-powered diagnostic or assistive technology for human readers. Therefore, an MRMC study is not relevant to its evaluation.
  5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • Not applicable. As noted above, this device is a mechanical safety accessory, not an algorithm. The study performed was a "simulated use study" that inherently involves human interaction with the device during the injection process. The performance reported is that of the device in this simulated use, which is its intended "standalone" operational mode.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • The ground truth was based on direct observation of mechanical function during simulated use, specifically:
      • Successful completion of injections.
      • Activation of the safety device (guard sliding forward, covering, and locking over the needle).
      • Absence of premature activation.
      • Protection against sharps injury.
    • Essentially, it's a functional performance ground truth, directly observable and measurable during the test procedure.
  7. The sample size for the training set:

    • Not applicable. This device is a mechanical one and does not involve machine learning or AI, so there is no "training set."
  8. How the ground truth for the training set was established:

    • Not applicable, as there is no training set for this type of device.

{0}------------------------------------------------

K123743

510(k) SUMMARY

UltraSafe Passive PLUS Needle Guard

MAR 2 6 2013

Company:Safety Syringes, Inc.2875 Loker Avenue EastCarlsbad, CA 92010 USA
Ph:(760) 918-9908
Fax:(760) 918-0565
Contact Person:Suzanne RichardsonVice President, Quality Assurance and Regulatory Affairs
Mary Stanners, RACSenior Regulatory Specialist
Date Prepared:December 5, 2012
Trade/Proprietary Name:UltraSafe PLUS Passive Needle Guard
Common Name:Anti Stick Syringe
Classification Name:Piston Syringe
Classification Number(s)/Product Code(s):21 CFR 880.5860 (MEG)

Legally Marketed Predicate Device: UltraSafe Passive Needle Guard (510(k) K011369, K060743, K122558)

Device Description:

The modification to the predicate device is the addition of the X100L PLUS device to the UltraSafe Passive Needle Guard X-Series family. The SSI UltraSafe PLUS Passive Needle Guard is an antineedlestick accessory for pre-filled ISO standard glass syringes. The device is composed of a guard, body, spring and plunger, is non-sterile and single use. The activation of the UltraSafe PLUS Passive Needle Guard device remains the same. Upon completion of the guard will slide forward, cover and lock over the needle of the syringe. It is a visual and/or audible recognition that the device safety feature has activated. The SSI devices are categorized as skin contact with a duration of category A- limited (< 24 h) as per ISO 10993 Biological evaluation of medical devices- Part 1: Evaluation and testing.

Intended Use/ Indications for Use:

The intended use remains the same. It is a safety mechanism to reduce occurrence of accidental needle sticks when using ISO standard glass prefilled syringes.

The indications for use remain the same with additional claims. The UltraSafe Passive Needle Guard is indicated for use as single use devices that are indicated for use as an accessory with pre-filled ISO Standard glass syringes to aid in the protection of healthcare professionals, patients who self-inject doctor prescribed medications and individuals that assist self-injecting patients, from accidental needlesticks. The intended patient population is unrestricted and may include children and adults, and

{1}------------------------------------------------

parenteral methods of administration. Additionally, the PLUS device is designed with a larger viewing window indicated where pharma company offering is a low fill volume and instructions request visualization. The PLUS device is designed with a robust plunger and built in extended finger flanges indicated where pharma customer offering is viscous.

Technical Characteristics Comparison Summary to Predicate Device:

The UltraSafe PLUS Passive Needle Guard is substantially equivalent to the predicate device in general technological features and principle of operation. The mechanism of the device does not change. Upon completion of the injection, the syringe will retract and guard will lock over the needle of the syringe.

Performance Data:

Bench testing was performed on the UltraSafe PLUS Passive Needle Guard and confirms that the PLUS functioned as intended and is substantially equivalent to the predicate X-Series device. Safety Syringes, Inc. maintains a Quality System compliant with 21 CFR 820, Quality System Regulation. The firm's Ouality System is registered by TUV SUD America to ISO 13485:2003. Medical Devices-Ouality Management Systems-Requirements for Regulatory Purposes. Test procedures, test protocols, and reports are maintained within the Quality System. The firm uses the standard elements of design control compliant to 21 CFR Part 820 and ISO 13485:2003 to develop its new products. Risk Management was conducted as per ISO 14971 Medical device-Application of risk management to medical devices. Biocompatibility testing performed demonstrates that the additional plungers met the requirements of Blue Book Memorandum G95-1 Use of International Standard ISO 10993 Biological Evaluation of Medical Devices Part 1: Evaluation and Testing.

Clinical Testing:

As per FDA guidance and ISO 23908: Sharps Injury Protection-Requirements and Test Methods-Sharps Protection Features for Single-Use Hypodermic Needles. Introducers for Catheters and Needles Used for Blood Sampling, simulated use studies were conducted on the PLUS device, testing 500 devices with zero failures for activation for a "97.5% confident that the true failure rate was no higher than 0.7% and 99.5% confidence that it is no higher than 1.1%" as per the FDA Guidance to ensure that the UltraSafe PLUS Passive Needle Guard did not impede or adversely affect the intended clinical performance of the device, did not activate prematurely under expected conditions of use and provided protection against unintended sharps injury until disposal. All injections were completed successfully; the safety device activated in all 500 devices with zero failures, consistent with the emptying of the syringe and administering all the medication, meaning the device did not impede the intended clinical performance. All 500 devices activated with zero failures meaning the device activated at the appropriate time under expected conditions of use and provided protection against unintended sharps injury until disposal. (Reference Guidance for Industry and FDA: Medical Devices with Sharps Injurv Prevention Features and ISO 23908: Sharps injury protection-Requirements and test methods-Sharps protection features for single-use hypodermic needles, introducers for catheters and needles used for blood sampling).

Conclusion:

Based upon the design, technology, performance and functional testing, the UltraSafe PLUS Passive Needle Guard device is substantially equivalent to predicate device previously cleared as 510(k)s K011369, K0607043 and K122558.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a stylized eagle with three stripes forming its body and wing. The eagle faces right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

March 26, 2013

Ms. Suzanne Richardson Vice President, Quality Assurance and Regulatory Affairs Safety Syringes, Incorporated 2875 Loker Avenue East CARLSBAD CA 92010

Re: K123743

Trade/Device Name: UltraSafe PLUS Passive Needle Guard Regulation Number: 21 CFR 880.5860 Regulation Name: Piston Syringe Regulatory Class: Class II Product Code: MEG Dated: February 12, 2013 Received: February 21, 2013

Dear Ms. Richardson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 - Ms. Richardson

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Image /page/3/Picture/6 description: The image shows two stamps with the text "Anthony D. Watson -S" on each. The stamp on the left is larger and more bolded than the stamp on the right. The stamp on the right also includes the date and time "2013:03.26 16:38:47" and "04'00'-".

Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

INDICATIONS FOR USE

510(k) Number (if known): TBD Ki23743

Device Name: UltraSafe PLUS Passive Needle Guard

Indications for Use:

Single use devices that are indicated for use as an accessory with pre-filled ISO Standard glass syringes to aid in the protection of healthcare professionals, patients who self-inject doctor prescribed medications and individuals that assist self-injecting patients, from accidental needlesticks. The intended patient population is unrestricted and may include children and adults, and parenteral methods of administration. Additionally, the PLUS device is designed with a larger viewing window indicated where pharma company offering is a low fill volume and instructions request visualization. The PLUS device is designed with a robust plunger and built in extended finger flanges indicated where pharma customer offering is viscous.

Prescription Use X (Per 21 CFR 801.109)

OR

Over-The Counter Use

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

(PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

FDA

Richard C. Chapman 2013.03.25 16:16:16 -04'00'

Oivision Sign-Off) Division of Anesthesiology, General Hospital fection Control, Dental Devices

510(k) Number. K123743

§ 880.5860 Piston syringe.

(a)
Identification. A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.(b)
Classification. Class II (performance standards).