K Number
K123733
Manufacturer
Date Cleared
2013-04-01

(117 days)

Product Code
Regulation Number
878.4400
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Cardioblate CryoFlex Surgical Ablation System is intended for minimally invasive cardiac surgical procedures, including the treatment of cardiac arrhythmias. The Cardioblate CryoFlex 7 cm, 10 cm, and 10-S probes plus the Clamp and Surgical Ablation Console freeze target tissue and block the electrical conduction pathways by creating an inflammatory response and cryonecrosis.

Device Description

The Medtronic Cardioblate® CryoFlex® Surgical Ablation Probes, Console, and accessories are used together as a system. The system is composed of the Cardioblate CryoFlex Console, Cardioblate CryoFlex Console, Refurbished, Cardioblate CryoFlex 7cm Probe, Cardioblate CryoFlex 10cm Probe, Cardioblate CryoFlex 10S Probe, Cardioblate CryoFlex Clamp with 10cm Probe, Tank Carrier, Tank Regulator, Tank Regulator, Refurbished, Pressure Transducer Cable, 6ft. (1.8m), Gas Hose, 8ft (2.4m), and Power Cord, North American. The Medtronic Cardioblate CryoFlex Surgical Ablation System is a cryoablative surgical device that is used in the cryosurgical treatment of cardiac arrhythmias. The system utilizes an argon-based cryogen for fast, controlled freezing. The Cardioblate CryoFlex probes are single use, disposable cryoprobes that are designed for use with the system. Each probe has an integrated thermocouple for monitoring temperature at is ablation segment. The probes are supplied sterile, cannot be reused or re-sterilized, and are available in 7cm, 10cm, 10-S (slightly stiffer), and clamp with 10cm probe versions. The probes shaft and ablation segment are made of specially heat-treated stainless steel. The shaft was designed with enough malleability that the surgeon can easily shape it while still maintaining enough stiffness to ensure its stability in the operating field. The maileable ablation segment has a bellows configuration, which provides kink resistance and thermal performance. A movable insulative sleeve on the shaft allows the surgeon to vary the size of the ablation zone. The probe has a 3m (10ft) connection hose which is to be handed out of the sterile field to an operating room nurse for connection to the control panel.

AI/ML Overview

The provided text describes a 510(k) premarket notification for the Medtronic Cardioblate CryoFlex Surgical Ablation System. This is a submission for regulatory clearance, not a study evaluating device performance against specific acceptance criteria in the manner typically found in clinical trials or performance studies for diagnostic AI devices.

Instead, the documentation focuses on demonstrating substantial equivalence to a predicate device, which is the primary regulatory pathway for many medical devices. The "performance data" mentioned refers to verification and validation testing to ensure the new device meets its design specifications and functions as intended, rather than clinical performance metrics like sensitivity, specificity, or reader improvement.

Therefore, many of the requested categories (e.g., acceptance criteria, reported device performance metrics in a table, sample size for test set, ground truth by experts, adjudication method, MRMC study, standalone performance, training set details) are not applicable or explicitly stated in this type of regulatory submission.

Here's an attempt to answer the questions based only on the provided text, noting where information is not available:


Acceptance Criteria and Study for Medtronic Cardioblate CryoFlex Surgical Ablation System (K123733)

1. A table of acceptance criteria and the reported device performance

  • Not explicitly stated in this document in the format of specific acceptance criteria (e.g., "sensitivity > 90%") and reported performance metrics.
  • The document states: "Verification and validation testing has demonstrated that the Medtronic Cardioblate CryoFlex Surgical Ablation System is substantially equivalent to the predicates." This implies that the device met internal design specifications and functions comparably to the predicate, which would be the "acceptance criteria" from a regulatory perspective for substantial equivalence. However, no specific performance values are given.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Not specified. The document mentions "verification and validation testing" but does not detail the sample sizes, data provenance, or study design (retrospective/prospective) of these tests.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

  • Not applicable / Not specified. This type of information is relevant for studies evaluating diagnostic accuracy or clinical effectiveness, often involving human interpretation. This submission focuses on the safety and functional equivalence of a surgical ablation device. "Ground truth" in this context would likely refer to engineering specifications and physical performance measurements (e.g., temperature, ablation size), rather than expert medical interpretation.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • Not applicable / Not specified. Adjudication methods are typically used in clinical studies where expert consensus is needed to establish ground truth for ambiguous cases. This is not mentioned for the verification and validation tests for this device.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • Not applicable. This device is a cryosurgical ablation system, not a diagnostic AI device requiring human reader interpretation or assistance. Therefore, an MRMC study and AI assistance effect size are irrelevant to this submission.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Not applicable. This device is a surgical instrument, not an algorithm, so standalone algorithm performance is not relevant.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

  • Not explicitly stated for the "verification and validation testing." For a surgical device, ground truth would likely be based on engineering specifications, physical measurements (e.g., temperature achieved, lesion size in in-vitro or animal models), and compliance with material and safety standards.

8. The sample size for the training set

  • Not applicable / Not specified. This device is hardware for surgical ablation, not a machine learning or AI algorithm that requires a training set.

9. How the ground truth for the training set was established

  • Not applicable. As the device is not an AI algorithm, there is no training set or corresponding ground truth establishment process relevant to AI.

{0}------------------------------------------------

K123733 1/3

510(k) Summary

APR 0 1 2013

General Information l.

Submitter Information a.

Submitter's Name andAddressMedtronic, Inc.Medtronic Cardiovascular8200 Coral Sea Street NE, MVS83Mounds View, MN 55112
Contact PersonKari ChristiansonSenior Regulatory Affairs SpecialistTel: 763-514-9529FAX: 763-367-9827Email: kari.j.christianson@medtronic.com

Date of Summary

December 04, 2012

Cryosurgical System

Proprietary Name of Device

Cardioblate CryoFlex Surgical Ablation System

Surgical Device, for Ablation of Cardiac Tissue

Common/Usual Name

Classification Name

Class II, per 21 CFR 878.4350

OCL .

Product Code:

Classification

Predicate Device

Cardioblate® CryoFlex® Surgical Ablation System (K121878)

{1}------------------------------------------------

II. Device Description

System and Components a.

The Medtronic Cardioblate® CryoFlex® Surgical Ablation Probes, Console, and accessories are used together as a system. The system is composed of the following components:

ComponentsModel number
Cardioblate CryoFlex Console65CS1
Cardioblate CryoFlex Console, RefurbishedR65CS1
Cardioblate CryoFlex 7cm Probe60SF7
Cardioblate CryoFlex 10cm Probe60SF2
Cardioblate CryoFlex 10S Probe60SF3
Cardioblate CryoFlex Clamp with 10cm Probe60CM1
Tank Carrier65TC1
Tank Regulator67RAXNA
Tank Regulator, RefurbishedR67RAXNA
Pressure Transducer Cable, 6ft. (1.8m)67PS6
Gas Hose, 8ft (2.4m)67H08
Power Cord, North American671PCNA

b. System Overview

The Medtronic Cardioblate CryoFlex Surgical Ablation System is a cryoablative surgical device that is used in the cryosurgical treatment of cardiac arrhythmias. The system utilizes an argon-based cryogen for fast, controlled freezing.

The Cardioblate CryoFlex probes are single use, disposable cryoprobes that are designed for use with the system. Each probe has an integrated thermocouple for monitoring temperature at is ablation segment. The probes are supplied sterile, cannot be reused or re-sterilized, and are available in 7cm, 10cm, 10-S (slightly stiffer), and clamp with 10cm probe versions.

The probes shaft and ablation segment are made of specially heat-treated stainless steel. The shaft was designed with enough malleability that the surgeon can easily shape it while still maintaining enough stiffness to ensure its stability in the operating field.

The maileable ablation segment has a bellows configuration, which provides kink resistance and thermal performance. A movable insulative sleeve on the shaft allows the surgeon to vary the size of the ablation zone.

{2}------------------------------------------------

The probe has a 3m (10ft) connection hose which is to be handed out of the sterile field to an operating room nurse for connection to the control panel. Note that once the probe is connected to the control panel, it should not be disconnected until the end of the procedure because it cannot be reconnected. The probe is disabled once it is disconnected.

Indications for Use III.

The Cardioblate CryoFlex Surgical Ablation System is intended for minimally invasive cardiac surgical procedures, including the treatment of cardiac arrhythmias. The Cardioblate CryoFlex 7 cm, 10 cm, and 10-S probes plus the Clamp and Surgical Ablation Console freeze target tissue and block the electrical conduction pathways by creating an inflammatory response and cryonecrosis.

There is no change to the indications for use as compared to the predicate device.

Comparison to Predicate Devices IV.

The Medtronic Cardioblate CryoFlex Surgical Ablation system is substantially equivalent to the predicate surgical ablation system, has the same intended use and the same design with the exception of the addition of an alternate insulative sheath component on the 7cm, 10cm and 10S probes (Models 60SF7, 60SF2, and 60SF3, respectively) and on the clamp with 10cm probe (Model 60CM1).

Summary of Performance Data V.

Verification and validation testing has demonstrated that the Medtronic Cardioblate CryoFlex Surgical Ablation System is substantially equivalent to the predicates.

VI. Conclusion

Based on the accumulated technical information, intended use, verification tests and performance data provided, the Cardioblate CryoFlex Surgical Ablation system is substantially equivalent to the currently marketed predicate devices.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure with three curved lines representing its wings or feathers. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the bird symbol. The logo is printed in black and white.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

April 1, 2013

Medtronic Inc. c/o Ms. Kari Christianson Senior Regulatory Affairs Specialist 8200 Coral Sea Street NE, Mail Stop MVS83 Mounds View, MN 55112

Re: K123733

Trade/Device Name: Cardioblate CryoFlex Surgical Ablation System Regulation Number: 21 CFR 870.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories. Regulatory Class: II Product Code: OCL Dated: February 6, 2013 Received: February 8, 2013

Dear Ms. Christianson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2 – Ms. Kari Christianson

Please be advised that FDA's issuance of a substantial equivalence determination does not mean . that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820): and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours.

Bram D厚ùlokerman

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications for Use

510(k) Number (if known):

Device Name: Cardioblate CryoFlex Surgical Ablation System

Indications For Use: The Cardioblate CryoFlex Surgical Ablation System is intended for minimally invasive cardiac surgical procedures, including the treatment of cardiac arrhythmias. The Cardioblate CryoFlex 7 cm, 10 cm, and 10-S probes plus CryoFlex Clamp and Surgical Ablation Console freeze target tissue and block the electrical conduction pathways by creating an inflammatory response and cryonecrosis.

Prescription Use X

AND/OR

Over-The-Counter Use

(Part 21 CFR 801 Subpart D)

(21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Bram D. Zuckerman
2013.04.01 15:17:16
-04'00'

Page 1 of

§ 878.4400 Electrosurgical cutting and coagulation device and accessories.

(a)
Identification. An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.(b)
Classification. Class II.